Background:
User programs provide the opportunity to run own methods besides the
pre-calibrated methods from the original supplier. It does not play a role if
these are own developed methods or methods from other suppliers of
test kits.
For more details please look into chapter 7.1 of the DR 4000 instrument
manual.
All input steps and their significance and options are explained in the
following sections. It will explain how to create and save a new “User
Program” for the Spectroquant® Test Kit.
Merck KGaA, Frankfurter Straße 250, 64293 Darmstadt (Germany): +49 6151 72-0
EMD Millipore Corporation – A division of Merck KGaA, Darmstadt, Germany
290 Concord Road, Billerica, MA 01821, USA, Phone: (781) 533-6000
Note:
Each user-entered calibration may be assigned a
unique number between 1 and 999. This number is
displayed in front of the specified program as part of
the user program.
Four-digit program numbers are reserved for Hach
user calibration templates. The prompt line will flash
INVALID ENTRY if the number is already in use.
Note:
Press the exit key at any time to return to the main
menu.
A press of the SINGLEλsoft key brings up the
create program screen where parameter and
setting information can be entered.
Note:
If a negative value is to be entered, press the SIGNsoft
key before the numbers are entered.
Note:
Enter a lower limit to enable an under range warning
(UNDER!) if a measurement falls below the lower limit.
Enter an upper limit to enable an over range warning
(OVER!) if a measurement exceeds the upper limit.
The user can thus be warned of values outside the
working range of the test.
Upper limit: Select e.g. 8.00 Press the ENTER key to accept.
Wavelength: Enter the wavelength by keying in the number
with the numeric keypad. Accept with a press of
the ENTER key or use the DOWN arrow key to
move to the next line.
Note:
C=Concentration; a, b, c, and d = coefficients;
A=absorbance
Note:
If an error is made while entering numbers on the
prompt line, press the CE key to clear and start over.
Note:
The EDIT FACTOR soft key does not appear until a
chemical form is entered. If more than one chemical
form exists for the test, enter these forms in the same
manner as the first chemical form, using the character
table.
Timer 1: Press the PAGE DOWN arrow key to move
to the line to specify the Timer 1.
LABEL
Gives options for instructing the user during the
timer period. The options include: Wait, Shake,
Invert, Swirl, Boil, Heat, and Timer.
EDIT TIMER
Allows the user to change the numeric portion of
the timer.
Note:
The options LABEL, EDIT TIMER are not available
unless a timer value is entered.
Note:
press the NOsoft key to exit without storing changes to
the calibration.
Now you can start with the measurement of the Note: Please keep in mind that the DR 4000 requires
® always a blank. This is the same procedure when
Spectroquant test kits by using the DR 4000 ®
photometer. using the Spectroquant test kits as if you would use a
Hach test kit
When you want to run the programmed test kit just ®
In the list “Programming data for Spectroquant
start the program. If you require any assistants or Test Kits for DR 4000 photometer from Hach” it is
help, please contact your local Merck Millipore / specified what kind of blank can be used for each
EMD Millipore organization or our local Merck individual parameter and measuring range.
Millipore / EMD Millipore dealer. You will find them
in most countries around the world.
®
Details about the Spectroquant test kits, which can be measured in the DR 4000 can be found in the
Internet. The package inserts, which describe the handling of the test kits, can be found also in our
Internet homepage.
As we are introducing a lot of test kits each year we are updating the list “Programming data for
Spectroquant test kits for DR 4000 photometer” regularly. Please have a look every 3-6 month into our
Internet homepage under: http://www.merckmillipore.com/photometry.