Download as pdf or txt
Download as pdf or txt
You are on page 1of 192

Spectro Scientific MiniLab 53

User's Guide and


Integration with
AMS Machinery Health Manager
Doc. P/N: 100-00087
Rev. A | June 2015

Spectro Scientific | An ISO 9001:2008 company


One Executive Drive, Suite 101, Chelmsford, MA 01824-2563
(+1) 978-431-1120 | www.spectrosci.com

Copyright © Spectro Scientific 2014. All rights reserved


Preface
Copyright Notice
© 2014 Spectro Scientific. All rights reserved.
The information and descriptions in this document are the sole property of Spectro Scientific and may
not be copied, reproduced, disseminated, or distributed without explicit written permission from Spec-
tro Scientific.
The statements, configurations, technical data, and recommendations in this document are believed to
be accurate and reliable at the time of this publication, but Spectro Scientific assumes no responsibility
or liability for any errors or inaccuracies that may appear in this document.
Spectro Scientific is not responsible for any infringement of patents or other rights of third parties that
may result from the use of this document. The content of this document is furnished for informational
purposes only, is subject to change without notice, and does not represent a commitment or guaranty
by Spectro Scientific.
Spectro Scientific makes no warranty, express or implied, with respect to the content of this doc-
ument. Spectro Scientific shall not be liable for any loss or damage, including consequential or special
damages, resulting from any use of this information, even if loss or damage is caused by Spectro
Scientific as a result of negligence or any other fault.

Trademarks
Spectroil, SpectroVisc, SpectroTrack, SpectroLNF, SpectroFTIR, and SpectroFDM are trademarks of
Spectro Scientific.
All other trademarks are the property of their respective holders.

Safety Information
The safe and proper operation of all our devices is our highest priority.
Personal Precautions
Wear appropriate protective equipment.
Ensure there is adequate ventilation.
Environmental Precautions
Use an absorbent material on the spill.
Prevent contaminated fluids from entering drains or groundwater if possible.
Dispose of in accordance to local regulations.

Warranty
Spectro Scientific warrants to the original purchaser only, that all Spectro Scientific portable instru-
ments will be free from defects in material or workmanship for a period of one year from date of
invoice. SPECTRO SCIENTIFIC'S SOLE OBLIGATION AND YOUR EXCLUSIVE REMEDY under this Lim-
ited Warranty and, to the extent permitted by law, any warranty or condition implied by law, shall be
the repair or replacement of parts, without charge, that are defective in material or workmanship and
that have not been misused, carelessly handled, abnormal conditions or operation, accidents or acts of

i
God, or misrepaired by persons other than Spectro Scientific or Authorized Service Provider. To make a
claim under this Limited Warranty, you must return the complete portable instrument, transportation
prepaid, to Spectro Scientific Factory Service Center or Authorized Service Location.
The determination of whether any product has been subject to misuse or abuse will be made solely by
Spectro Scientific. If a hardware defect arises and a valid claim is received within the limited warranty
period, at its option and to the extent permitted by law, Spectro Scientific will either (1) repair the hard-
ware defect at no charge, using new parts or refurbished parts that are equivalent to new in per-
formance and reliability, (2) exchange the product with a product that is new or refurbished that is
equivalent to new in performance and reliability and is at least functionally equivalent to the original
product, or (3) refund the purchase price of the product.
Spectro Scientific may request that you replace defective parts with user-installable new or refurbished
parts that Spectro Scientific provides in fulfillment of its warranty obligation. A replacement product or
part, including a user-installable part that has been installed in accordance with instructions provided
by Spectro Scientific, assumes the remaining warranty of the original product.
After an issue with the unit has been verified, the offending module where the issue originates should
be returned to the factory for repair by authorized Spectro personnel.
For Authorized Service Locations, please contact Spectro Scientific Technical Support, +1- 978-431-
1130 or support@spectrosci.com.

Manual Conventions
These conventions are used throughout this manual to call attention to the adjacent text:
The Note paragraph indicates special comments or instructions.
The Important paragraph alerts you to actions that could have an effect on the equipment or the col-
lected data.
The Warning paragraph warns you of actions that could cause serious injury or death.

OilView Disclaimer
The OilView instrumentation and software helps with diagnosing oil lubricating and machine wear
problems. Because of the complexities of real-world problems, there are no claims or guarantees that
OilView will correctly solve all of the problems encountered. Questionable results should be sent to a
reputable oil analysis laboratory for a more comprehensive analysis than OilView is able to perform.
All results must be considered as advisory information only and should be weighted carefully by the
personnel who are responsible for maintenance actions which are taken. Spectro does not accept any
liability with regard to the actions performed or omitted by those using this package.
The user is responsible for the safety of personnel and machinery. The user should consult their safety
department for the recommended methods of handling, storing, and ventilating volatile and lubricating
materials.

Firmware Revisions and Updates


AMS Suite: Machinery Health Manager requires version 5.61 or later. Contact Spectro Scientific Tech-
nical Support for the upgrade from version 5.6.
All LNF Q230 software versions released in 2014 can be upgraded; versions prior to 2014 are not sup-
ported.
Any FluidScan Q1100 device can be upgraded.

ii Doc. p/n 100-00087 Rev. A


Any SpectroVisc Q3000 Series Viscometer running the correct software version can be upgraded -
please contact Technical Support for information.

Document Revision Information


Refer to the following table for information regarding changes for this release:

Revision Date Description


A June 2015 Initial Release

iii
Table of Contents
Copyright Notice i

Trademarks i

Safety Information i

Personal Precautions i

Environmental Precautions i

Warranty i

Manual Conventions ii

OilView Disclaimer ii

Firmware Revisions and Updates ii

Document Revision Information iii


1 Introduction 1
User Guides 1

Machinery Health Manager Help Files 2

Online Help 2

Languages Other Than English 2

Licensing 2

Shipping Carton Contents 3

OilView GUI Overview 4

A\E\P Database Hierarchy 4

Customer Option 4

Primary, Secondary and Database Tab Overview 5

Primary and Secondary Tabs 5

Database Tabs 6

Fluid Sample Processing Options 8

Single Test 8

Sample List 8

Batch List 8

Routes 8

Advanced User Status 8

iv
2 Quick Start - Creating and Testing Samples 9
Enabling Export Control 9

Opening OilView 10

Creating an Area-Level Entry 11

Creating an Equipment-Level Entry 11

Creating a Point-Level Entry 12

Assigning a Substance ID to a Reference Fluid 15

Associating a Reference Fluid with a Point 16

Creating a Sample-Level Entry 16

Sample Testing 18

LNF Q230 Tests for Particle Contamination 19

Creating a Diluted Sample 19

Testing Samples 19

Data Analysis 20

Q3000 Series Viscometer Tests for Viscosity 21

Data Analysis 22

FluidScan Tests for Contamination and Degradation 23

Data Analysis 23
3 Sample List Testing For FluidScan and Viscometer 24
Creating Samples in the User Database 24

Creating the Sample List 25

FluidScan Sample List Testing 27

Viscometer Sample List Testing 28


4 Batch and Route Testing 29
Batch Testing Overview 29

Organizing Samples into Batches 30

Testing Batches 32

Batch Testing with a FluidScan 33

Batch Testing with a Viscometer 33

Batch Testing with a Q230 Particle Counter. 33

v Doc. p/n 100-00087 Rev. A


Batch Info Secondary Tab 34

Printing Reports 35

Result Options Secondary Tab 35

Contacts Secondary Tab 36

Assign Contacts Secondary Tab 37

Manual Entry Secondary Tab 39

Serial Communications Tabs 39

Route Secondary Tab 39

Printing Lab Sample Labels 39

Imported Batches Folder 40

Deleting Batches 40

Route Testing 41

Exporting Route Data 42

Testing Routes with FluidScan 44

Importing Route Test Data from a Data Stick 45

LIMS Management Reports 46

Contact Statistics 46

Lab Statistics 47
5 Displaying Test Results 48
Spectro Scientific Sample Trend Report 48

Trend Info Tab 51

Detailed Parameter Plots 53

Displaying and Printing Trend Info Reports 54

Sample Info Tab 55

TriVector Plot 55

Displaying and Printing Sample Info Reports 58


6 Analysis Parameter (AP) Sets and Alarm Limit (AL) Sets 59
Spectro Default AP Set and AL Sets 59

Creating Custom AP Sets and AL Sets 60

Creating a New AP Set 60

Selecting AP Set Parameters 60

vi
Configuring Custom AP Set Parameters 61

Creating a Custom AL Set 62

Alarm Templates 64

AP Sets and AL Sets Folder Options 65

Other AP Set Options 66

AP Sets - Report Groups Tab 66

AP Sets - Statistics Tab 66

Overwriting an AP Set 67

AP Set Right-Click Options 67

Other AL Set Tabs 68

AL Set Statistics Tab 68

AL Set Histogram Plot Tab 69

Baseline Adjusted Plot Sub-Tab 70

Raw Data Plot Sub Tab 71

Raw Data Sub-Tab 72

AL Set Scatter Plot Tab 72

AL Set Alarm Analysis Tab 73

AL Set Alarm List Tab 74


7 Managing the User and Reference Databases 76
File Menu Top-Level Database Functions 76

Database Change Marks 77

Database Tabs 78

User Database and Reference Database Levels 78

Creating a Custom User Database 78

Adding a Customer-Level Entry 79

Adding an Area-Level Entry 82

Adding Equipment, Point and Sample-Level Entries 82

Right-Click Options in the User Database 83

Folder-Level Options 83

vii Doc. p/n 100-00087 Rev. A


Customer-Level Options 83

Area-Level Options 83

Equipment-Level Options 83

Point-Level Options 84

Sample-Level Options 84

Common Right-Click Options in the User Database 85

Reference Fluid Database Options 89

Creating a Customized Reference Database 89

Setting the Preferred Reference Database 89

Adding Reference Fluids to the Reference Database 90

Creating the Reference Sample 91

Associating a Reference Fluid with a Point in the User Database 93

Assign Ref Secondary Tabs 94

Data Available Tab 94

Data Tab 95

Oil Type Tab 96

Properties Tab 97

Purpose Tab 98

Stat Plot Tab- Histograph Option 99

Stat Plot Tab - Stat Data Option 100

Changing a Reference Fluid Associated with a Point 101

Copying and/or Moving Reference Fluids 102

Maintaining Historical Sample Data While Updating Reference Fluids 102

Overwriting Reference Fluids for Point and Sample 103

Show Reference Samples Used by this Database 103

Querying the Database 104

Customer Sample Query 104

Generating a Schedule 104


8 Point Info Primary Tab 106
General Tab 106

Scheduling Tab 109

viii
Printing Sample Labels for Scheduled Samples 109

Test and Display Filter Tab 110

Internal Tab 111

Process Tab 112

External Tab 113

P1 Tab 114

P2 Tab 115

P3 Tab 116
9 Data I/O Primary Tab 117
Import Lab E-mail Tab 117

Import Lab File Tab 118

Database Export Tab 118

Statistics Export Tab 119


10 Setup Primary Tab 120
System Tab 120

Trend Tab 121

Translations Tab 122

Severity Codes Tab 123

File Locations Tab 123

DB Level Names Tab 124

Labels Tab 125

Lab Labels Tab 126

Reports Tab 127

E-mail Tab 128


11 WDA Primary Tab 130
WDA Overview 130

Microscope Setup and Calibration 130

Recommended Video Specifications 130

Sample Tab 132

Capturing Images Through the Microscope 133

ix Doc. p/n 100-00087 Rev. A


Measuring Image Size and Particle Density 135

Measuring Particle Density by Counting Particles 136

Sample Atlas Tab 137

Atlas Setup Tab 138

Analyst Tab 139

Working with Filter Patches 140

Using Stacked Filter Patches 140

Ferrous / Non-Ferrous Separation 140

Diluting Samples 140

Making Patches 141


12 Troubleshooting 147
Send to Instrument and Import Results Buttons are Disabled 147

<Device Name> is Performing Another Operation 147

Error Tracing 147

Test Modes 147

Data Edit/Entry 147

GUI Does Not Display Properly for Q230 147

"Cannot Find Reference 0" Message 148

Too Many Frames Message 148


13 Oil Properties Table 149
14 Default Spectro AP Set Parameters 162
15 Reference Fluid Database 165
A Replacement Parts and Consumables List 179
B Specifications 180

x
1 Introduction
Spectro Scientific has a direct interface between AMS 5.61, MiniLab 53, and the individual devices for
testing, analyzing and reporting data through the OilView software:
• SpectroVisc Q3000 Series Viscometer, used to measure the kinematic viscosity of a lubricant.
• SpectroVisc FluidScan Q1100, used to provide a quantitative measurement of fluid condition,
detecting any water contamination and degradation in a lubricant.
• Spectro LNF Q230, a particle counter, wear particle classifier, and ferrous monitor: tests which
measure particle contamination and machine wear.

The purpose of this document is to provide instructions to a user who wants to set up the software,
create and run fluid samples, and view and/or store test results using OilView software.
Important: Regardless of how many test instruments you own, you can only run tests with one Flu-
idScan, one Viscometer, and one Particle Counter.
This chapter describes the following:
User Guides, below
Machinery Health Manager Help Files on page 2
Languages Other Than English on page 2
Licensing on page 2
Shipping Carton Contents on page 3
OilView GUI Overview on page 4
Introduction on page 1
Primary, Secondary and Database Tab Overview on page 5
Fluid Sample Processing Options on page 8
Advanced User Status on page 8

User Guides
Each Spectro Scientific device has its own user guide for standalone operation that contains the
information you need to prepare samples and introduce them to each test device.
Unless you are already very familiar with these devices in standalone mode, you should have the fol-
lowing User Guides to hand:
• SpectroVisc Q3000 Series User's Guide, p/n 100-00070
• FluidScan Q1100 User's Guide, p/n 100-00080
• LNF Operators Guide, p/n LNF-211

1
Machinery Health Manager Help Files
Pressing F1 on the keyboard launches a special version of Microsoft Explorer. Explorer displays the
same information as is found in the Machinery Health Manager user’s manuals in HTML (.chm ) format.
These files are stored within their own directory.

Online Help
There is also a Help drop-down menu accessible from the top menu bar that includes the following
functions:
Help – accesses the online help system.
About – displays the current software version number and build date.

Languages Other Than English


AMS OilView supports several languages. If you use OilView in a language other than English, contact
Spectro Scientific to see if a language file is available that would provide a version of the Spectro Sci
tab in your preferred language.
Also, refer to the user's guide for each Spectro Scientific device for instructions on how to change the
GUI language.

Licensing
A single-user license is included with the purchase of a LIMS module.

2 Doc. p/n 100-00087 Rev. A


Shipping Carton Contents
The shipping carton contains the following items. You should review the contents of the shipping
package(s) to confirm the shipment is complete. Notify Spectro Scientific immediately in the case of
discrepancies.
Refer to Replacement Parts and Consumables List on page 179 for re-ordering information.

Description Quantity
Spectro Q1100 FluidScan 1

Spectro Q3050 Viscometer 1

Spectro LNF Q230 Particle Monitor with License Key 1

Software and User Guide Pack 1

USB A-B Male Cable, 3 meter 1

Universal Cable Kit 2

USB Adapter A-Male to Mini-B Male, 2 meter 1

USB Adapter USB to Mini-B Male, 2 meter 1

Ultrasonic Cleaner 1

Tabletop Homogenizer with Complete Water Solution License 1

Transport Case 1

Plug-in AC Adapter - Viscometer 1

Plug-in AC Adapter - FluidScan 1

Standard Accessories Kit 1

USB Flash Drive 1

3
OilView GUI Overview
This section provides a brief overview of the work areas you can expect to encounter as you perform
the tasks described in this guide.

A\E\P Database Hierarchy
Spectro Scientific sample fluid tests and analyses follow the familiar A\E\P format. Fluid test activity
takes place in nested database levels for Area, Equipment, Point, and Sample similar to this example:

You need at least one Area-level entry. Below that you can create one or more pieces of Equipment,
each with one or more measurement Points, and each with one or more Samples to test.
Customer Option
There is a "hidden" Customer-level option above the Area-level. You can create multiple Customers
holding multiple Areas, but you also need this level in order to create and distribute a number of
reports.
See Printing Reports on page 35
See Adding a Customer-Level Entry on page 79

4 Doc. p/n 100-00087 Rev. A


Primary, Secondary and Database Tab Overview
All OilView screens are created by selecting a combination of Primary and Secondary tabs. Depending
on their function, Primary tabs have different Secondary tab options.
Primary and Secondary Tabs
The Primary tabs run across the top of the OilView GUI and they are used to select the group of func-
tions that you wish to perform:

Note: the tabs for 5100, 5xDV, 51PC, 51FW 5200 and WAN are discussed in the
AMS documentation.
The Secondary tabs run across the bottom of the OilView GUI and they are used to select a variety of
options available under each primary tab.
For example, these are the secondary tabs for the Spectro Scientific primary tab:

5
Database Tabs
Some screen combinations are also dependent on selecting the correct database using the Database
tabs in the lower left corner.
There are four tabs, one each for the User, Reference and LIMS databases, plus a Q(uery) tab, which is
used with the LIMS database to run customer queries.
The User database opens with OilView by default:

If you select any sample from the User database, and click the Spectro Scientific primary tab, you will
see a screen similar to this:

6 Doc. p/n 100-00087 Rev. A


Sample field - Information about the Area, Equipment and Point for the selected sample.
Reference field - Information about the reference fluid associated with the Point.
LNF Dilution field - Used if a fluid has been diluted for tests with the Q230.
Measurement Results field - displays the raw test results for the the sample.
Status field - displays a variety of status messages depending on the test device.

7
Fluid Sample Processing Options
There are four distinct processing options available:
Single Test
Here you create a single entry in the User database at the Sample level - engine reservoir oil that may
have been contaminated in a flood for instance - and test the sample for particle contamination, and/or
fluid degradation and/or viscosity.
Refer to Quick Start - Creating and Testing Samples on page 9
Sample List
This option is available only for the Viscometer and FluidScan. Their portability makes this an ideal
option for users who want to create a list of samples in the field and test each sample just once as a
spot-check.Quick Start - Creating and Testing Samples on page 9
Quick Start - Creating and Testing Samples on page 9
Batch List
Batch lists are set up under the LIMS primary tab; refer to Organizing Samples into Batches on page 30
for more information.
A batch list can be run as often as necessary. This allows you to create a record of changes over time
that can be viewed using the Trends tab.
For example, you could add new engine oil today, then test it every 30 days to monitor the progressive
rate of deterioration over time.
Routes
Routes are functionally similar to Sample lists because the results are not retained in memory and so
cannot be trended. They are set up from the LIMS database and primary tabs, and the Routes sec-
ondary tab.
Note: Routes are tested with the FluidScan only.
Refer to Batch and Route Testing on page 29 for more information.

Advanced User Status


Some options can only be accessed if you have been granted Advanced User status. To do this, click
the Setup primary tab and the System secondary tab and click the Advanced User check box:

8 Doc. p/n 100-00087 Rev. A


2 Quick Start - Creating and Testing
Samples
This chapter describes the minimum number of steps and processes needed to create and test one
sample with the three Spectro Scientific devices that comprise the Minilab 53.
Note: Right-clicking on any of the database sub-level entries described in this chapter opens many addi-
tional create and/or edit options, particularly in the User database. For full information regarding these
options, refer to Managing the User and Reference Databases on page 76
Refer to the user's guide for each test device for information on how to power up each device, connect
it over the USB interface to a laptop running the AMS software, and how to perform any other basic
sample preparation and measuring tasks.
This chapter describes the following:
Enabling Export Control , below
Opening OilView on page 10
Creating an Area-Level Entry on page 11
Creating an Equipment-Level Entry on page 11
Creating a Point-Level Entry on page 12
Assigning a Substance ID to a Reference Fluid on page 15
Associating a Reference Fluid with a Point on page 16
Creating a Sample-Level Entry on page 16
Sample Testing on page 18

Enabling Export Control


Important: Open the LNF management software either from the Start menu or from a desktop icon
and navigate to the Export Control tab. If necessary, click the Enable Q5300 Interface check box. If
this is not checked, there is no communication link established with the LIMS module and the GUI will
not display correctly.
Note: Keep the management software open in the background while you conduct tests involving the
Q230 particle counter.

9
Opening OilView
Follow these steps to open up the OilView GUI.
1. Open Machinery Health Manager (MHM) from the Start menu or desktop icon.
2. Click the Tools tab:

3. Click the OilView Analysis icon.


Important: If the OilView Analysis icon is not visible in this toolbar by default, look for the Tribology
icon in the lower left corner and click it to add the icon to the toolbar:

This opens the OilView home page:

The User database opens by default.


We strongly advise you read OilView GUI Overview on page 4 before continuing.

10 Doc. p/n 100-00087 Rev. A


Creating an Area-Level Entry
There are two options to add an Area depending on whether or not you created a Customer-level entry
in the User database. The following example assumes you did not create a Customer-level entry.
Follow these steps to create an Area:
1. Right click the Example.rbm folder icon at the top of the User database tree.
2. Select Add Area to open the following dialog box:

3. Enter up to 4 characters to create an ID for the Area. For example, if you have gold mines in five
states, you might want to identify it using the postal abbreviation AK as the ID for the Alaska
mine.
4. Enter up to 32 characters to create a description; for this example call it Gold Mine.
5. Click OK to add Alaska Gold Mine to the database:

Creating an Equipment-Level Entry


Follow these steps to create an entry at the Equipment level:
1. Right-click on AK - Gold Mine and select Add Equipment
2. Enter an identifier for your equipment. For the purposes of this example, accept the default,
E1, and call it Rock Truck 1.
Now the database looks like this:

11
Creating a Point-Level Entry
The Point is a subset of your Equipment, such as a gearbox or an engine.
Follow these steps to create an entry at the Point level:
1. Right click on the Rock Truck 1 entry and select New Point. This opens the Point Info primary
tab at the General secondary tab ready for your selections:
Note: If you need more information about fields not fully described here, refer to Point Info
Primary Tab on page 106

2. Point ID field: Use up to three alpha/numeric characters to create an identifier for the Point. For
this example, accept the default ID: P1.
3. Description field: Enter a name for the Point. For this example, call it Engine.
4. Lab Unit ID field: For this example, leave this field blank.
5. Report Language field: For this example, accept CSI Default, which is American English.
6. Report Group field: This is defined in Analysis Parameter (AP) Sets and Alarm Limit (AL) Sets
on page 59
7. Usage Units field: For this example, accept the default Hours.
8. Target Cleanliness field: For this example, accept the default 21/18/15.
9. Lubrication System field: For this example, leave the Capacity field blank and accept the default
Units- (Gallons US).

12 Doc. p/n 100-00087 Rev. A


10. Analysis Parameter (AP) Set field: AP Sets are described fully in Analysis Parameter (AP) Sets
and Alarm Limit (AL) Sets on page 59. Select an AP Set by clicking the down-arrow in the
Analysis Parameter AP Set field and selecting from the drop-down list. For this example, use
Spectro.
11. Equipment Type field: Describes the Point hardware. Select an entry by clicking the Browse
button to the right of the Equipment Type field and scrolling to the appropriate entry. For this
example, scroll down to Engine\Other\Diesel:

12. Alarm Limit (AL) Set field. AL Sets are described fully in Analysis Parameter (AP) Sets and
Alarm Limit (AL) Sets on page 59. Select an AL Set by clicking the down-arrow in the Alarm
Limit (AL) Set field and selecting from the drop-down list. For this example, use Diesel Engine.
This is one of the default AL Sets created as part of the Spectro AP Set.
Now the General Information tab looks like this:

13
13. The Point adds itself to the database automatically after you select the AL Set and click off the
screen - there is no Add or Save button.
Now the database looks like this:

Notice that the Point icon is blue. This indicates that a Reference fluid has not yet been asso-
ciated with the Point, which is described below.

14 Doc. p/n 100-00087 Rev. A


Assigning a Substance ID to a Reference Fluid
For this example, assume you are using Castrol GTX 10W-30 engine oil. This oil is included in the Refer-
ence fluid database, but you must make sure that it has a Substance ID associated with it.
Substance IDs and Reference fluids are fully described in Managing the User and Reference Databases
on page 76.
Follow these steps to assign the Substance ID to Castrol GTX 10W-30.
1. Click the Ref DB tab to open the Reference fluid database.
2. Drill down the Castrol entries until you reach the Oil Description level:

3. Open the Reference Fluid Database on page 165 and look up the Substance ID for Castrol
GTX 10W30. It is 256.
4. Right-click the Oil Description and select Add Ref Sample.
5. Accept the default Sample Date. For this example, use the Substance ID itself - 256 - as the
Sample ID.
6. Click OK to add the new Sample-level entry to the database.
7. The Trend Info primary tab now opens by default. Scroll down to the bottom and insert the
Substance ID - 256:

15
Associating a Reference Fluid with a Point
You must assign a Reference fluid to a Point Managing the User and Reference Databases on page 76
Refer to Managing the User and Reference Databases on page 76
Follow these steps to associate the Reference oil with the P1 - Engine point you just created.
1. Click the User database tab.
2. Click on the P1 - Engine point and click the Assign Ref primary tab. This places the User
database on the left, and the Reference database on the right:

3. In the Reference database, drill down to the Castrol GTX 10W-30 Sample you created to
activate the Set Reference button at the top of the column:

4. Click the Set Reference button to change the Point from blue to white:

Creating a Sample-Level Entry


Follow these steps to create an entry at the Sample level.

16 Doc. p/n 100-00087 Rev. A


1. Right click the P1 - Engine point and select New Sample. The sample will automatically add
itself below the Point using a default ID:

2. To change the ID to something more appropriate, right click the sample and select Edit.
3. The Add Sample option pops up a dialog box so you can change the Sample ID. For this
example, call it P1 test 1:

Optional: You have the opportunity to record any physical observations that you made when
the sample was collected by right-clicking each question and selecting the appropriate answer
from the drop-down options: Yes, No, Unknown. Answer from 0 to all questions as you want.
The information will appear in the Sample Info tab sidebars. For this example, accept the
default: Unknown for all entries.
4. Accept the Sample Date - default is to today, although you can edit the date by highlighting
and changing it.
5. Click OK. The database now looks like this:

17
Sample Testing
Follow these steps to test your sample.
1. Click the P1 test 1 sample you just created and click the Spectro Scientific primary tab to open
a screen similar to this:

Sample field - Shows the parameters for the sample you selected from the User database.
Reference field - Shows the chemical properties of the Reference oil that you associated with the Point.
LNF Dilution field -
This field has no effect on samples for the FluidScan and Viscometer - leave the field blank or at 0.
For the Q230 only , you can create a diluted sample for fluids with a viscosity equal to or greater than 30
cSt. The recommended process is described in LNF Q230 Tests for Particle Contamination on page 19.
If you are not diluting a sample for the Q230, this field must be set to 0 or left blank for each sample.
Connected/Disconnected: Each device must report that it is connected via the USB cable by displaying
a green bar as in the illustration above.
If you see a red bar for one or more devices, disconnect the USB cable then reconnect it and wait for 5 -
15 seconds for the bar to turn green.
If it still does not connect, reboot the device and the computer before contacting Technical Support.
Status field - Displays a variety of test status information. The status field will show "Ready" when you
can begin tests.
Note: The most common status message is < Device Name> is Performing Another Operation, but
processing is actually hanging. Make sure that the physical FluidScan and Viscometer GUI and the
LNF management software GUI is set to the Main Menu to continue.

18 Doc. p/n 100-00087 Rev. A


LNF Q230 Tests for Particle Contamination
This section explains how to analyze a fluid sample using the LNF Q230 Particle Counter, Wear Clas-
sifier, and Ferrous Monitor.
Note: Refer to the "Q200-Series Particle Counter User's Guide" for basic information on setting up the
device and testing fluid samples.
Creating a Diluted Sample
You can create a diluted sample for fluids over 30 cSt. This speeds up processing time.
Important: You must create a diluted sample for fluids >320 cSt
The recommended procedure to create a diluted sample is as follows:
1. Fill a syringe with 10 ml of diluent (lamp oil or kerosene for example).
2. Take a Spectro sample bottle and fill it up to the second line marker with sample fluid:

3. Add the diluent from the syringe.


4. Shake the bottle for ~30 seconds.

Testing Samples
Make sure the fluid sample is shaken vigorously for 30 seconds, then placed in an ultrasonic bath for
30 seconds to clear any bubbles. You can also use the homogenizer as described in the "Q1100
FluidScan User's Guide."
Begin tests immediately after de-gassing to avoid settling.
1. Select the P1 test 1 sample you just created.
2. If you are using a diluted sample, enter the dilution ratio in the LNF Dilution field: 4.5 if you use
the recommended method. Otherwise, enter a customer-determined ratio.
3. Press the Start button and insert the sipper tube when prompted:

4. Press the Continue button.

19
5. When the sample test is complete, you are prompted to begin the flush cycle:

6. Switch the fluid sample to a vial of cleaning fluid (pure lamp oil or kerosene for example) and
press Continue to start the flushing cycle:
7. After the flush cycle, the results are displayed in the Measurement Results panel:

The sample icon in the User database has turned red, so at least one parameter is in an extreme alarm:

Data Analysis
For full details on how to analyze data, refer to Displaying Test Results on page 48.

20 Doc. p/n 100-00087 Rev. A


Q3000 Series Viscometer Tests for Viscosity
This section explains how to test a fluid sample for viscosity. Refer to the "SpectroVisc Q3000 Series
User's Guide" for basic information on setting up the device and running the tests.
Check the user's guide to see if a calibration is required. If it is, you must perform the calibration pro-
cess before continuing with this section.
Follow these steps to run the viscosity test:
1. Select the Sample 1 sample from the User database.
2. Make sure the Viscometer GUI is at the Main Menu; this activates the Start button in the
OilView GUI.
3. Press Start. The Status message changes to:

It is not yet processing a sample. To actually start processing, you need to focus on the Viscometer
GUI and follow the instructions there.
By default, only the result of the test at 40C is displayed in the Viscometer GUI.
If you want a result at 100C, refer to the "Q3000 Series Viscometer User's Guide" for full information
about the Viscosity Index parameters.
For the purposes of this example, follow these steps to create a 100C result:
1. Press the back arrow in the Viscometer GUI.
2. Press the VI button.
3. Use the arrow buttons to adjust the Viscosity Index. For this example, accept the default.
4. Press the right arrow. The 100C value is now shown along with the 40C measurement.
5. Press the Right arrow again to return the Viscometer to the Main Menu and export the test
result to the OilView GUI:

The sample icon has turned red, so at least one parameter is in an extreme alarm state:

21
Data Analysis
For full details on how to analyze data, refer to Displaying Test Results on page 48.

22 Doc. p/n 100-00087 Rev. A


FluidScan Tests for Contamination and Degradation
This section explains how to test a fluid sample using the Q1100 FluidScan.
Refer to the "Q1100 FluidScan User's Guide" for basic information on setting up the device and run-
ning tests.
Follow these steps to run samples:
1. Click the Sample 1 sample in the User database.
2. Make sure the FluidScan GUI is open at the Main Menu to activate the Start button in OilView.
3. Press the Start button. The OilView Status message changes to:

4. Follow the steps in the FluidScan GUI to measure the sample.


5. Once the measurement is complete, FluidScan analyzes the results and displays them in the
FluidScan GUI.
6. Press the FluidScan Save button to save results to the FluidScan database and export the
result to OilView:

7. Press the FluidScan OK button to return the FluidScan GUI to the Main Menu. This makes the
device ready for the next sample.
The sample icon has turned red, so at least one parameter is in an extreme alarm:

Data Analysis
For information about analyzing data, see Displaying Test Results on page 48.

23
3 Sample List Testing For FluidScan and
Viscometer
This function allows you to set up a list of samples you can run consecutively and upload the results as
a group rather than testing each sample individually.
You can run the same list through both devices, or create a separate list for testing on one device or
the other.
Sample list results are not retained in memory. It can be considered an information snapshot. If you
want to create a list that you can save and re-run multiple times to establish a trend analysis, then you
need to create a Batch list. This is described in Batch and Route Testing on page 29
This chapter describes the following:
Creating Samples in the User Database, below
FluidScan Sample List Testing on page 27
Viscometer Sample List Testing on page 28

Creating Samples in the User Database


For this example, you want to test the reservoir oil for your mining trucks: Rock Truck 1, Rock Truck 2
and Rock Truck 3.
The Equipment, Point and Sample level entries for Rock Truck 1 were created earlier; see Creating an
Equipment-Level Entry on page 11

Create Equipment, Point and Sample entries for Rock Truck 2 and Rock Truck 3 using the following
parameters:

Equipment ID Point ID Sample Name

E2 - Rock Truck 2 P2 - Engine P2 test 1

E3 - Rock Truck 3 P3 - Engine P3 test 1

The database now looks like this:

24
Creating the Sample List
Follow these steps to create a Sample List:
1. Click the P1 test 1 Sample.
2. Click the Spectro Scientific primary tab and the Sample List secondary tab to open a screen
similar to this:

3. In the User database, select the P1 test1 sample for Rock Truck 1. This will display in the
Sample field:

25 Doc. p/n 100-00087 Rev. A


4. Press the >> button to move the sample to the Sample List field:

As soon as you add your first sample to the list, the Send to Instrument and Import Results
buttons should become active. If they do not, check the FluidScan GUI and/or the Viscometer
GUI and make sure it is showing the Main Menu. This will activate the buttons.
5. Add the samples for Rock Truck 2 and Rock Truck 3 so your Sample List looks like this:
Note: If you get an error message saying that the Sample has already been added, check the
Point ID matches the Equipment ID - P1 is the default when you create a Point, but you cannot
have multiple P1 Point IDs in a Sample List.

6. Click the Send To Instrument button for both devices. The Status screen changes to Sample
List accepted by < device name>.
7. Refer to the device GUI and follow the instructions, as follows:

26
FluidScan Sample List Testing
Refer to the "Q1100 FluidScan User's Guide" for basic information on device setup and testing
procedures described in this section.
1. From the FluidScan GUI Main Menu, select Manage Samples.
Note: It may be that there are results in the FluidScan database from samples that were previously
tested and saved to the database and are now in the way of managing your untested samples. You can
hide these by pressing the Hide Done button on the FluidScan. This does not delete them from the Flu-
idScan, and you can unhide them at any time.
For this exercise, assume there are no previous results in the database.
2. Sample List Testing For FluidScan and Viscometer on page 24
3. Place 2 -4 drops of your test fluid onto the cell window - just enough to cover the glass. Return
the cell to the FluidScan housing and press OK.
4. After the measurement, the device analyzes the sample and presents the result in the FluidScan
GUI.
5. Press Save, then press OK to measure the next sample in the list.
6. Repeat the process for each sample in the Sample List.
7. Press the Back button to return the FluidScan to the Main Menu. This will activate the Import
Results button in the OilView GUI.
8. Click the Import Results button in OilView.
9. Refer to Displaying Test Results on page 48 for information on how to display and analyze the
results.

27 Doc. p/n 100-00087 Rev. A


Viscometer Sample List Testing
Refer to the "Q3000 Series Viscometer User's Guide" for basic information on device setup and test-
ing procedures.
1. From the Viscometer GUI Main Menu, press the list icon:

2. The List screen opens at the first sample:

If you want to see all the samples in the list and rearrange them in some way other than how the list
was imported, click the DN button. For this example, leave the list as it was imported, starting with P1
test 1.
3. Select the sample you want to measure and click the Right arrow.
4. Follow the instructions in the viscometer GUI to test the sample and display the results.
5. Press the Return arrow to return the device to the Main Menu and press the List icon to select
the next sample.
6. When all samples have been tested, make sure the viscometer GUI is at the Main Menu, then
click the Import Results button in OilView.
7. Press the Clear List on Instrument button to clear the samples from the viscometer:

Refer to Displaying Test Results on page 48 for information on how to display and analyze the results.

28
4 Batch and Route Testing
The Laboratory Information Management System (LIMS) module is licensed software that enables
graphical interfaces for testing and reporting.
Its function here is to support Batch and Route testing, plus a number of reporting functions.
This chapter describes the following:
Batch Testing Overview, below
Organizing Samples into Batches on page 30
Testing Batches on page 32
Printing Reports on page 35
Route Testing on page 41

Batch Testing Overview


Batches are created in the LIMS database and consist of samples grouped together under a Point-level
entry for testing, analysis, and reporting. Samples remain linked to the master database so that data,
analysis, and diagnostic comments are automatically stored in the proper location in the database.
Each of the required activities – tests, analyses, and reports – is checked off for each sample in the
batch as it is performed. This makes it easy to keep track of the work that remains to be done. LIMS
allows you to establish a directory of report recipients.
Batches can be re-run on any suitable schedule. For example, you could test transmission fluid for Red
Truck and Blue Truck in January and July, and Green Truck and Yellow Truck in March and September.
Data and reports can be emailed, faxed, or printed according to preferences that you have established
in the secondary tabs for the LIMS primary tab; see Printing Reports on page 35 for more information.

29
Organizing Samples into Batches
Follow these steps to organize the samples to be tested into batches according to similar oils, test pro-
files, customers, or whatever is most appropriate for your particular situation. The maximum batch size
is 99, and there can be multiple batches for a given day.
For this example, the batch list will be created from samples that were created in the User database for
two mine trucks: Rock Truck 1 and Rock Truck 2.
1. Click the LIMS database tab.
2. Click the LIMS primary tab and the Assign Samples secondary tab.
3. Right click the Test Batches folder and select Add Batch. The date will default to today, but
that can be changed by right-clicking the batch.
The Batch ID (600 in this example) is assigned by default - it becomes part of the Lab # (see
below).
4. In the User database on the right, drill down to the two samples P1 test 1 and P2 test 1.
The screen now looks like this with the <<<Add button inactive:

5. Click on P1 test 1 to activate the <<<Add button at the top of the database.
6. Click the <<<Add button to add the sample to the Batch.
7. Click on P2 test 1 and click the <<<Add button.

The screen looks like this, with the two samples now part of the batch in the left pane, and the sample
detail in the center pane:

30 Doc. p/n 100-00087 Rev. A


You now have a batch list that is easily identified by date and\or Batch ID.
Note: The Batch ID forms the root of the Lab #.

31
Testing Batches
Click the Spectro Scientific primary tab if it is not already open. Allow up to 15 seconds for the devices
to connect to the laptop.
Important: Each device needs to display a green "Connected" message before you can perform meas-
urements with it; otherwise, refer to the Troubleshooting chapter for guidance:

When your devices are all connected, you can begin testing. Tests can be performed in any order, or
concurrently with each other.
Note: At any time in the testing process, if you see a Status box message "<device name> is per-
forming another operation..." make sure the device GUI and\or the Q230 management software GUI is
at the Main Menu to clear this message and continue testing.
Refer to Batch and Route Testing on page 29 to record the progress of each sample as batch testing is
completed.

32 Doc. p/n 100-00087 Rev. A


Batch Testing with a FluidScan
Sample preparation and test fluid introduction are the same as described in the "Q1100 FluidScan
User's Guide." Detailed steps can also be found in Quick Start - Creating and Testing Samples on page
9
1. Make sure the FluidScan is at the Main Menu and the Status field in OilView says Ready.
2. Click the Start button in the OilView GUI. The Status field message will change to Measuring
Sample, please wait.
3. Run the test following the instructions in the FluidScan GUI.
4. The test result displays in the FluidScan GUI.
5. Press the Save button on the FluidScan to save the data and upload the results to OilView.
6. Test the next sample(s) in the same way.
Batch Testing with a Viscometer
Sample preparation and test fluid introduction are the same as described in the "Q3000 Series Vis-
cometer User's Guide." Detailed steps can also be found in Quick Start - Creating and Testing Samples
on page 9
1. Make sure the viscometer is at the Main Menu and the Status field in OilView says Ready.
2. Click the Start button in the OilView GUI. The Status field message will change to Measuring
Sample, please wait.
3. Run the test following the instructions in the viscometer GUI.
4. When the test results display, press the forward arrow icon to upload the results to OilView.
5. Test the next sample(s) in the same way.
Batch Testing with a Q230 Particle Counter.
Sample preparation and test fluid introduction are the same as described in the "Q200 Series Particle
Analyzer User's Guide." Detailed steps can also be found in Quick Start - Creating and Testing Samples
on page 9
1. Make sure the LNF management GUI is at the Main Menu and the Status field in OilView says
Ready.
2. Click the Start button in OilView and follow the instructions in the Status field.

33
Batch Info Secondary Tab
This shows the progress of each sample as activities are completed. As each test is run, the device
check-box is automatically marked.
You can also finish a sample by clicking the Trend Info primary tab. Right click on a sample, then select
Finish from the menu. This is used to confirm that the sample diagnosis is complete.
The Wear, Contamination, and Chemistry status of each sample is automatically displayed for each
sample, making it easy to identify problem samples:

Diag column - is populated when a diagnostic message is entered.


RVWD column checkbox - is checked when the Reviewed button is clicked on the Sample Info primary
tab.
REPT and EXP checkboxes - are automatically marked when the reports are printed and the data have
been exported.
On the extreme right of the Batch Info table are date columns reflecting the date on which each pro-
cess was completed. The analyst and reviewer user IDs are also recorded with each sample:

Note: The RVWD, REPT, Exp, Imp, and Diag flags are used to control which samples are reported on.
They can be cleared using the right-click Clear Flag option:

Depending on which of the flags are set on the options specified for reporting, you can then generate a
variety of reports for printing and automatic E-mailing. You can also export the data for importing at a
remote site using the report results menu option shown below.

34 Doc. p/n 100-00087 Rev. A


Printing Reports
Important: Before any reports can be generated, each Area and/or Equipment level in your database
must have at least one Customer assigned to it.
Refer to Adding a Customer-Level Entry on page 79 for more information.
Each customer has an address that is printed on the report, and a profile specifying the kinds of
reports that particular customer requires.
Test reports can be created in a variety of ways using the secondary tabs found under the LIMS
primary tab:
Result Options Secondary Tab
The Result Options tab serves two purposes: to establish the conditions under which reports can be
generated, and to override the normal reporting process.

The General Options field is used to set the conditions for generating reports based on the flags that
are set as each sample is processed. For each of the flags shown, you can specify Ignore, Checked, or
Unchecked.
Reports will be generated only for those samples in the batch that match these conditions. The actual
reports that are generated depend on the Customer report profile for each sample.
The file name defaults to the date format that is chronologically listed in the directory. For data export,
this file is appended to it. For report export, this file is overwritten.
Report Override Options field is used to generate ad hoc reports that will ignore or override the specific
setup for each customer assigned to the sample. The options here are the same as those in the cus-
tomer report profile that will be described in the next section.

35
If Duplicate File Name field. Each day data and reports are temporarily exported into each contact’s
sub-directory where they are assembled. A temporary file with the day’s date is created to store this
information.
It is also used to specify the action to be taken if a report or data for the contact is generated more than
once per day. By default new reports and data are appended to the existing file. These files can also be
manually renamed or automatically renamed.
Contacts Secondary Tab
This tab is used to create a contact address book and reporting profile for each contact. The Add New,
Delete, and Update buttons are used to manage this list. The E-mail address can be used to auto-
matically e-mail reports to contacts.
These e-mailed reports can be viewed and printed.

To generate a specific report, the severity level at which that report will be created must be set. The low-
est severity level is Normal meaning the report will always be generated:

36 Doc. p/n 100-00087 Rev. A


If the severity level is set higher, then reports will only be created if the severity for the sample is
greater than or equal to that level.
The Fax reports operate in the same manner. You must setup a fax printer to enable automatic faxing
of reports. (See E-mail/Fax tab in the Setup tab.)
The check boxes are used to specify in which format the reports are to be sent. To automatically E-mail
a report directly to a contact, check the E-mail box and the Format pdf (Adobe Acrobat) box.
In a network environment, it may be more convenient to place the reports in a directory that the con-
tact can access. To use this method, check the Directory box instead of the E-mail box. The directory is
specified in the Contact Sub-Directory field.
Each customer is required to have a unique sub-directory that is used as a working directory by the
software.
Assign Contacts Secondary Tab
Note: There is a distinction between customers and contacts. Customers own the data in a database
and only they can import the data, whereas contacts receive reports either by e-mail or in printed form.
There can be only one customer assigned at the Area level, but there can be many contacts at either
the Area or Equipment levels.
Once the contact list has been created, the next step is to assign contacts to the User databases so
that contact-specific reports can be generated. Multiple contacts can be assigned at the Customer,
Area, and Equipment levels of the User database. This means that more than one contact can receive
reports and data for each sample.
To assign a contact, select the relevant node on the database tree and then select the contact from the
bottom list. Use the Add button to assign the contact. For example, this is the screen at the Customer
level:

37
For those customers that require the sample data to be sent directly to them, the Data Export section
must be completed (see above). This information is completely independent of the contacts assigned
to the Customer and Area levels.
At the Customer level you must specify the sub-directory into which temporary information is stored
prior to the data being sent. Each customer must have their own unique sub-directory to avoid data
being sent to the wrong customer. You must also specify the e-mail address to which the data is sent.
If you do not select the e-mail option then the data remain in the sub-directory and can be accessed dir-
ectly from a WAN. The data format defaults to the Area, Equipment, Point (AEP) format. The altern-
ative uses the Unit ID format.
The data export setup only applies to the Area level. You must specify the customers database name
where the data is to be imported, as well as the temporary sub-directory on the local system where the
data is written.
There must be a different sub-directory for each customer database. In general, the local lab database
level structure may be different from the customers at the top (Area) level. If this is the case, then the
Cust Area ID field must show the Customer ID.
To automatically E-mail the data to the customer, the E-mail checkbox must be checked and an E-mail
address entered. For the export process as a whole to be activated, the Export flag must be checked.
The AEP export specifies the Area, Equipment, and Point ID’s for each sample in the export file. This is
the most versatile method as it will automatically create the area, equipment, and point levels in the tar-
get database if they do not exist.
The Unit ID method requires that the Lab Unit ID in the target database point setup be specified. Cus-
tomer DB Name and Customer Area ID are not used with Unit ID method.
At each Area level of the database you must enter the database name to which the data is being sent
(in general this will be different from the current database) as well as the equivalent Area ID to which
the data is being sent:

38 Doc. p/n 100-00087 Rev. A


This too may be different from the current Area ID. This feature enables a single LIMS/User (RBM) data-
base to store data from multiple user databases. To export data, select the Export check box.
Manual Entry Secondary Tab
Not used with the Minilab 53.
Serial Communications Tabs
Not used with the Minilab 53.
Route Secondary Tab
See Route Testing, below.

Printing Lab Sample Labels


You can print lab sample labels for all samples in a batch, or for multiple batches, depending on the
level in the LIMS database tree (selected from the right clickpop-up menu).

The lab labels, the Lab# printed on it, and which tests are to be performed on the sample are used
internally. This information is derived from the Point setup for the sample.
For more information, refer to Setup Primary Tab on page 120 for more information on setting up and
printing labels.

39
Imported Batches Folder
This batch type is automatically created when you import results data using the Data I/O tab.
This groups together in batches all samples that are imported at any one time so that you can easily
find the individual samples without having to search through the user databases. It also enables you to
generate reports for all the samples at one time.
Refer to the Online Help for more information about the Data I/O tab.

Deleting Batches
Batches can be deleted by right-clicking the batch and selecting Delete.
Important: When a batch is deleted, the samples and their data are not deleted from your User data-
base, they are only removed from the batch list.

40 Doc. p/n 100-00087 Rev. A


Route Testing
Important: Route testing can only be done with a FluidScan.
The Routes secondary tab allows you to:
• Create a route
• Add one or more Points from the User database.
• Export the point database to a memory stick.
• Run sample tests in FluidScan.
• Re-import the data back into OilView.
Follow these steps to create a Route and export/import the data:
1. Insert a data stick into the laptop.
2. Click the LIMS database tab.
3. Click the LIMS primary tab:
4. Click the Routes secondary tab.
5. Right-click the Routes folder in the LIMS database and select Add Route; for this example, call
it - what else - Route 66. The screen changes to show the route on the left and the User
database on the right:

6. From the User database, drill down to the Point for Rock Truck 1 and click the Add button. Do
the same for Rock Truck 2 and for Rock Truck 3:

41
Exporting Route Data
You can either export data for all Points in a route, or you can select one particular Point. Except for
Step 1, the steps are the same for both:
1. Right-click either the Route 66 icon or an individual Point.
2. Click Export Spectro Scientific csv. This will change the screen to the Data I/O primary tab and
the Database Export secondary tab.
3. Enter a file name with a .csv extension in the Save As field. For this example, use Route 66.csv:

4. Optional: you can filter by Start Date and End Date and/or by Change Marks (see Managing the
User and Reference Databases on page 76 for more information about change marks).
5. Make sure the Spectro Scientific Route/Batch (csv) button is selected.
6. Click the Save As button.
7. Click the Start button to begin the data export:

42 Doc. p/n 100-00087 Rev. A


8. When you see the Export Complete notice, remove the memory stick and plug it into the
USB port on the FluidScan.
You are now ready to begin testing your route.

43
Testing Routes with FluidScan
Follow these steps from the FluidScan GUI:
1. From the Main Menu, select Run Route.
2. Select Import New Route and press the Import button.
3. Select the route from the memory stick and press Open.
4. Verify the route information and press the Measure button.
5. Test each sample as described in the "Q1100 FluidScan User's Guide."
6. After you have tested all samples, select Run Route from the Main Menu. The route you just
tested will show as Completed:

7. Click the Export button. You will get the following data transfer confirmation message:

8. Remove the memory stick and insert it into the laptop.

44 Doc. p/n 100-00087 Rev. A


Importing Route Test Data from a Data Stick
Follow these steps to import the result data from the data stick:
1. Click the Data I/O primary tab.
2. Click the Import Lab File secondary tab:

3. Click the Data File button:

4. Select the memory stick and import the file. The file extension will be .txt.
5. Click the Start button.
6. The data are automatically imported and displayed in the User database against the samples
for each Point in the route:

You can display sample test results by selecting the Sample and clicking the Sample Info primary tab.
Refer to Displaying Test Results on page 48 for more information.

45
LIMS Management Reports
The following reports are available by clicking on a batch and selecting from the drop-down menu:

Contact Statistics
1. From the drop down menu, select Contact Statistics.
2. Click on the Report Preview magnifier icon.
3. A preview of the Contact Statistics report for contact dick Ensian opens for the selected batch:

46 Doc. p/n 100-00087 Rev. A


Lab Statistics
1. From the drop down menu, select Lab Statistics.
2. Click on the Report Preview magnifier icon.
3. A preview of the Lab Statistics report opens:

47
5 Displaying Test Results
There are several ways to display and analyze test results in the OilView GUI.
This chapter describes the following:
Spectro Scientific Sample Trend Report, below
Trend Info Tab on page 51
Sample Info Tab on page 55

Spectro Scientific Sample Trend Report


You can display a comprehensive sample report directly from the Spectro Scientific tab that will dis-
play up to five sample test results so you can track trends over time.
Important: we strongly suggest you read the chapter on setting up AP and AL Sets first so the correct
data are displayed properly.
1. From the Oilview GUI, click the Spectro Scientific primary tab. Then click the Report secondary
tab to open the report. The User database is displayed, plus the Point information in the
header:

2. Typically users add samples from the newest to oldest to see the fluid condition as of the most
recent sample. As you add new samples to the database, you would need to delete the oldest
sample from the list first.
Locate the Point containing the samples to display, click the newest sample - Test 5, and click
the Add button:

3. Click the remaining samples in descending age order until you have all five samples in the list:

48
As you have been adding the samples, test result data have been imported.
Note: Some of the data used have been manipulated to show extreme results.
The left side of the data output screen displays raw data and alarm severity colors from the AL Set.
The right side displays a series of graphs.
Note: These graphs are hard-coded and display regardless of any settings in the Plot column in the
AP Set or if there are any data from test results.
• Total ferrous particles.
• Ferrous particles bigger than 25 microns
• ISO particle size greater than 4/6/14 microns
• Water ppm
• Viscosity at 40C
• MxD particles
• Total acid number (TAN)

49 Doc. p/n 100-00087 Rev. A


Wear particles, if there are any, can be displayed at the bottom similar to this example:

50
Trend Info Tab
The Trend Info tab is used to view historical data and alarm status for all samples defined for a Point-
level entry.
Important: If you have a high number of sample results for a given Point but you are only interested in
seeing or reporting on a limited number of results, the number of sample columns can be limited to
either a fixed number or for the period since a particular date. This is configured on the Setup primary
tab, System secondary tab; see for more information.
Follow these steps to view the Trend Info table.
1. Open the Oilview GUI and navigate to a Point-level entry in the User database and select it:
2. Click the Trend Info primary tab. The All Data secondary tab opens by default:

Alarms are reflected in the color shading. The color corresponds to the severity criteria created in the
Setup primary tab - 3 colors or 5. 5 will give more granularity in your result displays, as there are more
threshold points.
Column 1 - shows the AP Set parameters - Note: these are not in the same order as in the AP Sets Pro-
file Setup tab.
Column 2 - click until Oil Ref appears to view parameters for the Reference oil associated with this
Point.
Other columns - shows test result data for each Sample.
The top three rows indicate the overall test fluid condition for Wear, Contamination and Chemistry.
Scroll down the table to see individual test results; for example, ISO particle results:

You can add data to any cell that does not have actual test result data in it. You cannot change derived
sample test data.
Important: Making edits will not update the overall Trivector status for wear, contamination, and chem-
ical analysis. The Trivector is displayed on the Sample Info primary tab, discussed below.
It will not update the content of any Trend reports.
After typing in or editing data, you must
1. Remove the samples from the Trend report by selecting them and clicking the Remove button:

51 Doc. p/n 100-00087 Rev. A


2. Right click every Sample in the User database tree that was changed, select Recalculate Alarms
and click OK in the popup box.
3. Add the samples back into the report to recreate the display.
You can edit alarm thresholds on the fly by right clicking in the cell and selecting from the options in
the drop-down menu (below). The cell color will change accordingly.

No Alarm will take off the alarm threshold, but leave the data in the cell.
No Data will remove the data from the cell.

52
Detailed Parameter Plots
In the Trend Info screen, you can right-click on any parameter with test result data and you will get a
basic line graph for that parameter.
In this example for the Chemical Idx parameter, you can see the basic line graph, but it has been
enhanced by selecting additional information to display by using the check boxes in the left column.
These include showing the alarm levels, the reference value, local median, mean and standard devi-
ations, and symbols to indicate what observations are applicable.

Click the Log Scale button to change the line graph display.
Click the Finish button to exit

53 Doc. p/n 100-00087 Rev. A


Displaying and Printing Trend Info Reports
There is a drop-down box on the top bar of the OilView GUI with a number of report options based on
the Trend Info test data that you can view and print.
Click the magnifier to display the report, click the printer icon to print the report, and click the X icon to
close it.

The following reports are discussed in this section:


• Sample Trend
• Sample Trend with Reference
• Standard Sample Report
• Vibration Analysis (not used)
• Spectro Sci Trend (not used)
For more information regarding these reports, refer to the MHM Online Help (not Spectro).

54
Sample Info Tab
The Sample Info tab is used to view graphs, observations, and notes for individual oil samples rather
than multiple samples collected under a Point-level.The sample data will not be used for any kind of
trend over elapsed time monitoring.
TriVector Plot
Press the Sample Info primary tab. This will open at the Trivector secondary tab.
The Trivector Plot combines multiple test results into one graph. It consists of three evaluation cat-
egories: Wear, Contamination, and Fluid Chemistry. The plot reports oil condition in each category from
Normal (green) to Extreme (red).
A plot like this sample means there is acceptable wear and the fluid chemistry is normal, but the fluid is
showing signs of contamination:

You can also choose to display the results in 5 colors for more granularity in the results; refer to Alarm
Display Option in the Managing the User and Reference Databases, below chapter for instructions
how to set this.

55 Doc. p/n 100-00087 Rev. A


There are three buttons on this screen:
Reviewed - This button is used in conjunction with the LIMS.
Schedule - Once a sample has been diagnosed and it is determined another sample will be required,
you can go directly to the schedule for the Point-level entry (i.e. gearbox) and schedule another sample.
Analyze - If there is any additional data available, it will appear in the Observations, Actions and/or
Sample Diagnosis boxes, as shown here:

If you want to add any entries now, you can click on an empty row and right click to open a drop-down
options box and make your selections.
Selecting an observation automatically adds one or more items to the Actions field.

56
Important: Any entry in the Sample Diagnosis field is your text description of the status and recom-
mendation for this sample.
For example, if the result of test indicated that the sample contained a Contaminant A component of
12345 ppm, you can enter this information using the Notes secondary tab, with a recommendation to
change the oil - but none of this information comes directly from the software - it is not advising you
what you must do in a given situation.

57 Doc. p/n 100-00087 Rev. A


Displaying and Printing Sample Info Reports
There is a drop-down box on the top bar of the OilView GUI with a number of report options based on
the Sample Info test data that you can view and print.
Click the magnifier to display the report, click the printer icon to print the report, and click the X icon to
close it.

The following reports are available:


• Sample PPM
• Sample ISO
• Sample NAS
• Sample Analyzer
• Sample WDA
• Minilab Report
• Trivector Report
For more information regarding these reports, refer to the MHM Online Help (not Spectro).

58
6 Analysis Parameter (AP) Sets and Alarm
Limit (AL) Sets
Analysis Parameter (AP) Sets and Alarm Limit (AL) Sets are created or edited by right-clicking on the
AP Sets and AL Sets folder in the User database and making selections from the drop-down options.
They define which test result parameters are displayed in OilView for analysis, and which are not.
This chapter describes the following:
Spectro Default AP Set and AL Sets, below
Creating Custom AP Sets and AL Sets on page 60
Alarm Templates on page 64
Other AP Set Options on page 66
Other AL Set Tabs on page 68

Spectro Default AP Set and AL Sets


When you create a new Point-level entry in the User database, you are required to select an AP Set
from a drop-down list in the Point Info primary tab. Rather than create a new customized AP Set, we
recommend you use the default Spectro AP Set.
It consists of 81 parameters and it was created by us after testing and experience revealed the most
likely useful parameters for general engineering applications. It still leaves room for customer fine-tun-
ing if necessary - up to a total maximum of 96 parameters.
Refer to Default Spectro AP Set Parameters on page 162 if you want to know the default parameters
in the Spectro AP Set.
The Spectro AP Set also comes with four default AL Sets suitable for most industrial/engineering
applications:

This is an important consideration as custom AP Sets do not generate any AL Sets by default.
The four Spectro AL Sets are based on the same 8 parameters but with different alarm thresholds
• PPM Water
• Fe>25
• ISO >6
• IR Oxidation

59
• Visc 40C
• Total Acid
• Total Base
• Total Fe
The following example shows the thresholds for the Gear AL Set:

Creating Custom AP Sets and AL Sets


This section describes the minimum steps to create a Custom AP Set and a custom AL Set. All other
screens and options are described in Other AP Set Options on page 66.
Creating a New AP Set
Follow these steps to create a custom AP Set:
1. Right click the AP Sets and AL Sets Folder icon and select Add AP Set from the pop-up menu.
2. Enter an ID Number and a name in the dialog box. For this example, call it WilsonOil, and click
OK to add it to the list of AP Sets:

The screen automatically opens at the AP Sets primary tab and the Profile Setup secondary tab
so you can select the AP Set parameters.
Selecting AP Set Parameters
Select your AP Set, then click the AP Sets primary tab and the Profile Setup secondary tab to display
the list of available AP Set parameters. An AP Set is defined by selecting up to 96 parameters, so
select/deselect the check boxes as necessary in order to define parameters for a new AP Set.

60 Doc. p/n 100-00087 Rev. A


Parameters in AP Set field indicates how many parameters are selected. This changes up/down in real
time to help you keep count:
Update Changes button – saves the database to the AP Set list.
Important: If this is an edit to an existing AP Set, you must backup the database before making
changes.
Highlight Used Parameters – This option is not used.

Configuring Custom AP Set Parameters


Every parameter you select for a new AP Set is defined by a series of options that you can either accept
or edit.
Follow these steps to configure the WilsonOil AP Set you created earlier.
1. Click the WilsonOil AP Set in the AP Sets and AL Sets folder.
2. Click the AP Sets primary tab and the Current Views secondary tab:

Parameter column - is the list of all the parameters in the AP Set.


Units column - shows the measurement value for each parameter. Click in the cell for a drop-down
option box. Some parameters have only one option; others have more. We recommend the defaults.
Wear, Chemistry, and Contamination columns - defines which test results contribute to the overall
sample condition report in the Trend Info and Sample Info tabs.
Click in the cell and select Yes or No. In the example below, the PPM Water result would be part of the
overall condition for both Contamination and Chemistry, but aluminum and antimony will only become
part of the overall Wear condition:

61
The Minilab and Lab columns indicate which parameters are associated with OilView minilab instru-
ments and which come from oil laboratories.
Plot column - Click Yes for each parameter (maximum of 6) you want to appear as a graph on the
Sample Trend report (see Displaying Test Results on page 48).
We recommend the following six parameters:
• Total Fe
• Fe>25
• ISO>4
• IR Ox
• Acid
• Water PPM
Alarm column - See Analysis Parameter (AP) Sets and Alarm Limit (AL) Sets on page 59.
Manual column - is used to mark those parameters that appear on the manual data entry form in the
LIMS module.
Digits column - shows the number of display digits for each parameter and can be set from 0 to 4. For
example, a percentage with 2 in the Digits column would output the result to 2 decimal places:
"87.42%."

Import Units column - is used to specify the units for importing lab data. For this example, accept the
defaults.
Creating a Custom AL Set
Before creating a customized AL Set, consider using a template; see Alarm Templates on page 64.
For this example, follow these steps to create a new AL Set for the custom AP Set WilsonOil:
1. Right-click the WilsonOil AP Set in the AP Sets and AL Sets folder and select Add User Alarm.
2. Select the Application Type from the list of options. For this example, select Generator and
click OK.
3. Create an ID and add a description; for this example call it Generator 1 and click OK to add the
AL Set.
4. Click the WilsonOil AP Set, then click the AP Sets primary tab and the Current Views secondary
tab.
5. In the Alarm column, select Yes for every parameter you want to add to the AL Set.

62 Doc. p/n 100-00087 Rev. A


6. Click the new Generator 1 AL Set, then click the AL Sets primary tab and Current Alarms
secondary tab to see the new AL Set:

7. Accept the default alarm values, or edit them as you like by clicking in each cell. You can adjust
the Base Type, Delta Type and Alarm Type column threshold parameters based on this table:

Base Type Delta Type Alarm Type


Value
Zero Up
Absolute Value

Oil Ref
Value Up
AL Base
Absolute Value Down
Pnt Base
Percent Up and Down
Ref or Pnt

63
Alarm Templates
You might consider using the available alarm templates as an option to creating a customized AL Set.
These standard alarms have been established using data from many customers and are appropriate for
general industrial machinery:
1. Select an AP Set in the User database and create an AL Set as described above.
2. Click on the new AL Set, then click the AL Set primary tab and Alarm Templates secondary tab.
3. Select the appropriate template and then press the <<< button to assign the template values.

64 Doc. p/n 100-00087 Rev. A


AP Sets and AL Sets Folder Options
This section describes the options available by right-clicking the folder that were not described above.
Set Default AP/AL IDs -
Delete Unused Alarms - This option removes all unused AL Sets from the database. This not only frees
up space in the database for new AL Sets (there is a limit of 512), it also makes it a lot easier to view the
statistics for each AL Set as the only AL Sets available are those that are actually used. You can use
this option to clear all the unused AL Sets.
Set Default Alarm Type -
Update Statistics - Exports all the parameter data for generating statistical information. You must first
set both the correct alarm type and which are to be alarmed. Also, you must update the Point Baseline:
Update Alarm Baseline - Once the statistics have been exported, you can use them update the AL Set
baseline. This is necessary if you have set an alarm to use the AL Set baseline.

65
Other AP Set Options
This section describes other AP Set tabs and options not described above.
AP Sets - Report Groups Tab
This screen is used to setup the parameters and trend plots that appear on the Sample Trend report.
To access it, select an AP Set then click the AP Sets tab, and Report Groups secondary tab:
The specific group used is defined in the Report Group field on the Point Info primary tab, or when you
create a Point in the User database.

The Table column is used to mark those parameters that appear in the table on the report and the Plot
column is used to mark those parameters for which a trend plot is provided. There is a maximum of 6
trend plots and 70 parameters on a report.
The different groups enable you to define different combinations of parameters and plots for different
types of equipment. For example, you could have one group for engines and another for gearboxes.
AP Sets - Statistics Tab
This information is accessed through the AP Sets primary tab and Statistics secondary tab.
It shows the coverage weeks between an alert severity happening (i.e., severity of 1) where 0 implies
normal.

66 Doc. p/n 100-00087 Rev. A


This can be referred to as Mean Time Between Alerts (MTBA). Best results depend on alarm limits being
correct and uniformly applied to the statistics. The Recalculate option is useful in achieving this. Once
this has been done, it is also necessary to Update Statistics.
The MTB can be used to establish sampling intervals, which should be more frequent than the MTBA
to ensure alert situations are not missed due to infrequent sampling.
The breakdown into Wear, Contamination, and Chemistry MTBA’s will give an indication of the most
frequent types of problems.
Overwriting an AP Set
You can overwrite an AP Set with another by dragging and dropping.
Important: In addition to just merging two AP Sets, the AL Sets from the AP Set being dragged and
dropped are also added to the AL Sets of the target AP Sets.
Also, an AL Set may also be dragged and dropped or copied from one AP Set to another. In the pro-
cess all data are automatically updated. This can be a slow process and there is a progress bar show-
ing status of the merge operation.
You can also drag one parameter on top of another thereby superseding one with the other. This is
essential to replace the Special Test parameters with actual parameters or if a parameter has been
incorrectly used.
AP Set Right-Click Options
The following options not described above are available by right-clicking on the AP Set:
Edit AP Desc - Use this option to edit the AP Set name and AP Set ID only. This ID is used to export AP
Sets and then import them to the correct location within a database.
Delete AP - You can delete an AP Set and all its associated AL Sets. The software first checks to ensure
that neither the AP set nor any of the AL sets are currently being used. If they are, you cannot delete
the AP Set until the AL Set has been deleted.
Generate Default Alarm Sets – Generate a complete set of CSI Default AL Sets. We do not recommend
using this.
Show Used AL Sets – The used AL Sets are highlighted. This is useful for reviewing only those AL Sets
that are used.

67
Other AL Set Tabs
This section describes other AL Set tabs and options not described above.
AL Set Statistics Tab
The table below shows the average, minimum, maximum, count and standard deviation of the set of
samples to which this AL Set has been assigned in the database. This information is useful for estab-
lishing alarm limits.

This statistical information must be created by first exporting all the data into a special database using
the Data I/O tab as shown below. Click the book icon to open a calendar to select the start/end date for
statistic sampling:

The export process is confirmed:

68 Doc. p/n 100-00087 Rev. A


The annual average values for each parameter as calculated based on a simple average, or on the 80%,
90%, 95%, or 98% percentile values. This information is used to establish whether the oil analysis pro-
gram is having a beneficial impact on the equipment, and is based on a steady decline in contamination
and wear.

AL Set Histogram Plot Tab


Right click on a particular parameter to display a histogram over time:

This is a useful statistical plot for setting alarm limits. The x-axis plots the parameter values and the y-
axis the percentage of samples for each parameter value. The basic concept in establishing alarm limits
is that a maintenance department can only handle a certain percentage of problems at any one time.
Typically, you assign the extreme alarm limit at 95% (i.e. put only those samples whose parameter val-
ues fall in the top 5% into this category). Similarly the next alarm limit down is set at 90% (top 10%),
then at 80% (top 20%) and the lowest alarm limit at 70% (top 30%).

69
The vertical bars representing the alarm limit values can be dragged with the mouse to the desired
value. Once all the alarm limit positions have been set then press the Save Changes button and the
alarm limit value will be automatically written back to the AL Set. Using this method you can set the
alarm limits for all the parameters.
There should be at least 100 samples in the statistics before you can reliably use this method to set the
alarm limits. The more samples the better. Remember to update the statistics at regular intervals to
ensure you are using the latest samples collected.
Baseline Adjusted Plot Sub-Tab
This graph is accessed as part of the Histogram Plot secondary tab:

Set/Save All - automatically sets all alarm limits.


Set Alarm Limits - automatically sets alarm limits for selected parameters.
Save Changes - save manual changes by dragging vertical alarm limit lines, and then pressing the Save
Changes button.
Sometimes the plot has several “bumps” that are difficult to interpret. This is most often when a Refer-
ence baseline is being applied.

70 Doc. p/n 100-00087 Rev. A


Raw Data Plot Sub Tab
This graph is accessed as part of the Histogram Plot secondary tab:
The raw data plot ignores any reference or other baseline and only plots the raw data. This is useful for
diagnosing the source of obscure plots.
The solid vertical line shows the mean value and the dashed vertical line one standard deviation from
the mean value.
Most often these plots with “bumps” are due to incorrect references being applied.

71
Raw Data Sub-Tab
This graph is accessed as part of the Histogram Plot secondary tab:
To further diagnose obscure plots the raw data itself can be viewed here to find the specific data that is
causing inconsistent plots. Using this information you can then usually correct the problem much as
fixing an incorrect oil reference.
One way of minimizing these sorts of issues is to run the Point Configuration Summary report, which
is designed to find incorrect oil reference and other Point Setup issues.

AL Set Scatter Plot Tab


The scatter plot is used to see if there is a correlation between any two parameters. Simply select the
appropriate two parameters from the X and Y drop down lists and then press the Plot button.

72 Doc. p/n 100-00087 Rev. A


If the two parameters correlate well, they will fall on a relatively straight line unlike the one shown here.
The slope and intercept of the best fit line are shown on the display together with the correlation coef-
ficient. A correlation coefficient of 1 represents a perfect correlation.
This feature is extremely useful for developing rules for the expert system when it is released. The scat-
ter plot can either use linear or log axis depending on the scale of the parameters.
AL Set Alarm Analysis Tab
This data is used to assist with analysis of alarms. For the case shown, 15 samples use this alarm set
with 6 samples in alarm condition. There is 1 sample with a severe alarm caused by only 1 parameter:
Manganese.

73
AL Set Alarm List Tab
This tab is useful for quickly checking and changing configuration information. It is a very wide display
and is shown here in three parts:

Lines can be sorted by clicking a caption at the top of the column. Additional sort criteria may be
entered in the text boxes at the top. Columns can be hidden and made visible by clicking the buttons at
the top.
The Type ID, AL Set, RefID, TCL’s, and Report Group (Grp) can all be modified from this screen:

74 Doc. p/n 100-00087 Rev. A


You can also select Reference oils from a drop down list on the Ref Oil Desc field:

75
7 Managing the User and Reference
Databases
This chapter describes the two primary databases and explains how you can manipulate them in a vari-
ety of ways for a variety of outcomes.
It also explains all the right-click menu options at all levels - if you are looking for a Quick Start guide
with the minimum of options to get up and testing, refer to Quick Start - Creating and Testing Samples
on page 9.
This chapter describes the following:
Database Change Marks on page 77
User Database and Reference Database Levels on page 78
Creating a Custom User Database on page 78
Adding a Customer-Level Entry on page 79

File Menu Top-Level Database Functions


The File drop-down menu provides access to several important top-level database functions for both
databases but only when you have the User database open:

Open Database – Switch to another User database if one exists.


Create Database – Create a new User database.
Create Oil Reference Database – Create a new Reference database. This creates the database, but in
order to use it, you need to select it as the default Reference database from the Set Oil Reference Data-
base option in this drop-down list.
Print – Prints the report (if available) for the current data view.
Printer Setup – Sets up the printer for printing reports.
Set Oil Reference Database – Selects the active Reference database. You can create as many Refer-
ence fluid databases as you need, but you can only activate one per User database at any time.
Compact WDA & LIMS Databases – These are MS Access databases that can become very large with
use. Compacting them reduces their size and makes the data access more efficient.

76
Database Change Marks
Whenever any entry is updated, the database tree places Change Marks on the left side of the data-
base:

These can be cleared by right-clicking on any entry at the Point-level or above.

77 Doc. p/n 100-00087 Rev. A


Database Tabs
Open the OilView GUI by clicking on the OilView Analysis icon:

Check in the lower left corner for the Database tabs:

There are three database tabs: User Database, Reference Database, and LIMS Database plus a Query
tab.
The Query tab is used in conjunction with the LIMS module to query the database; refer to Managing
the User and Reference Databases on page 76 for full details.
Refer to Batch and Route Testing on page 29 for more information about creating and using Batches
and Routes.

User Database and Reference Database Levels


The default User database is Example.rbm . The default Reference database is Example.ref.
There is an option to add a Customer-level beneath the Example.rbm folder in the User database only.
This is normally hidden, but it is necessary for automatically forwarding analysis reports.
There are three sub-levels for each database that correspond with each other based on the A/E/P
model, plus a Sample level.

Database
User Database Reference Database
Level
---- Customer
A Area (Gold Mine) = Manufacturer (Castrol)
E Equipment (Truck) = Brand (GTX)
P Point (Engine) = Oil Description (10W-30)
---- Sample = Reference Sample

Creating a Custom User Database


The default User database is Example.rbm . You can't edit the name to something more descriptive, so
you might want to create your own User database(s).

78
Follow these steps to create a new User database:
1. From the File menu, click Create RBM Database:

2. Select the Server from the drop-down list and create a name for the database. For this example,
call it Tribology Labs and click OK.
3. Enter Tribology Labs in the File Name field and click Create New.
Select a Reference fluid database to associate with this User database. There are two data-
bases supplied with OilView: Example.ref and UserOil.ref.
4. Example.ref is the default, but because Example.ref is overwritten whenever the AMS software
is updated, we recommend you select any other database. For this example, select UserOil.ref
and click OK.
This will add the Tribology Labs database and include a default AP and AL Set:

Adding a Customer-Level Entry


There is a "hidden" Customer-level available immediately below the Folder-level in the User database.
Before any reports can be generated, each Area and/or Equipment level in your database must have at
least one customer assigned to it. Each customer has an address that is printed on the report, and a
profile specifying the kinds of reports that particular customer requires. Refer to Batch and Route Test-
ing on page 29
Follow these steps to create a Customer-level entry:
1. Click the Setup primary tab.
2. Click the System secondary tab.
3. Locate the LIMS area and click the Display Customer Level box:

79 Doc. p/n 100-00087 Rev. A


This creates a default entry called Unknown Customer.
If there are any Areas already in this database, they are now grouped under the Unknown Cus-
tomer, like so:

Note - you will not be able to delete the Unknown Customer, or edit the name, but you will be able to
drag and drop any areas grouped within it to other Customers you create at this level.
You can also hide all Customers without deleting them by de-selecting the Display Customer Level
checkbox.
Any Areas that exist below a Customer-level when you hide it will remain visible in the database, but
they will be unsorted, as in this example:

4. Right click on the Folder icon and select Add Customer to open the Add Customer dialog box.
5. Add a Customer ID; for this example, use Mine.
6. Add a description; for this example, call it Pardner Gold Mining:

80
7. Optional: complete the Customer Export Info fields similar to the above example.
8. Click OK to add the Customer to the User database tree:

81 Doc. p/n 100-00087 Rev. A


Adding an Area-Level Entry
There are two ways to do this.
Either,
If you did not create a Customer level, see Creating an Area-Level Entry on page 11
Or,
If you created a Customer level, follow these steps:
1. Right click the Customer icon and select Add Area to open the Add Area dialog box.
For this example, assume you have five gold mines. Use the state postal abbreviation to ID this
one in Washington, and call it Gold Mine 1:

2. Complete the Area Export Info field and click OK. The Area is now added below the Customer
level. You can add as many as you need:

See also Managing the User and Reference Databases on page 76

Adding Equipment, Point and Sample-Level Entries


The instructions to add these levels are the same as the description in the Quick-Start section of this
guide:
Creating an Equipment-Level Entry on page 11
Creating a Point-Level Entry on page 12
Creating a Sample-Level Entry on page 16

82
Right-Click Options in the User Database
There are many right-click options for all levels in the User database that were not discussed in Quick
Start - Creating and Testing Samples on page 9. This section provides top descriptions for each
database level, plus a separate section describing common options that appear at more than one level;
see Common Right-Click Options in the User Database on page 85.
Folder-Level Options
Save Severity - Assists with cleaning up alarm levels. Saved severity levels will allow you to undo any
future changes to the Sample alarm by using the Restore Severity option (below).
Note: Saved severity values will replace lab alarm severities. Lab alarm severities cannot be recovered
without re-importing the lab data.
Restore Severity - Assists with cleaning up alarm levels. It is assumed you previously performed a Save
Severity (above).
Note: Saved alarm severities will replace User alarm severities. Existing User alarm severities cannot be
recovered, but they can be made consistent by selecting Recalculate Alarms. If you change alarm lim-
its, you can recalculate the alarms by right-clicking any sub-level entry.
Add Customer - When the Customer-level is set to display, opens the Add Customer dialog box.
Add Area - When the Customer-level is hidden, opens the Add Area dialog box.
Find Sample # - Not limited to numbers only; enter any alpha-numeric string you might have used as a
Sample ID into the dialog box and click OK to locate the sample.
Calculate PC Volume - Pops up a dialog box while calculating.
Refer to Common Right-Click Options in the User Database on page 85 for more.
Customer-Level Options
Add Area - Opens the Add Area dialog box. Create an Area as described in this chapter.
For all user-created Customers only:
Edit Customer - Edit Customer ID, Description and/or Customer Export info.
Delete Customer - Asks you to confirm delete.
Refer to Common Right-Click Options in the User Database on page 85 for more.
Area-Level Options
Add Equipment - Opens the Add Equipment dialog box. Refer to Creating an Equipment-Level Entry on
page 11 for procedure.
Edit Area - Edit Area ID and Name.
Delete Area - Allows you to delete an Area provided there are only Points beneath it.
Set Report Language - Select the appropriate language from the drop-down options. Default is English.
Refer to Common Right-Click Options in the User Database on page 85 for more.
Equipment-Level Options
Add Point - Opens the Point Info primary tab and the General secondary tab. Refer to Creating a Point-
Level Entry on page 12
Edit Equipment - Edit Name and ID.

83 Doc. p/n 100-00087 Rev. A


Refer to Common Right-Click Options in the User Database on page 85 for more.
Point-Level Options
Add Sample - Creates a sample with default Date and ID. See Creating a Sample-Level Entry on page 16
Edit Point - Opens the Point Info primary tab and General secondary tab - all options are available to
edit.
Delete Point - Asks for confirmation.
Copy Point - Copies the Point which can then be pasted into another Equipment-level entry with a new
Point ID.
Refer to Common Right-Click Options in the User Database on page 85 for more.
Sample-Level Options
Edit Sample - Edit Name, Date and any optional sample-taking questions.
Delete Sample - Asks for confirmation.
Delete WDA - Opens a dialog box asking for confirmation to delete Wear Data Analysis information
associated with a Sample (as indicated by the "flake" symbol in the Sample icon):

Refer to Common Right-Click Options in the User Database on page 85

84
Common Right-Click Options in the User Database
The following right-click options appear at two or more sub-levels.
Recalculate Alarms – you can recalculate the alarms if you change alarm limits.
Update Point Baseline – Alarms can have a baseline relative to the median at each point. This option
updates the point baseline for all Points in the database.
Clear Marks - Clears database Change Marks.
Schedule - Pops up a calendar. Right-click on calendar to add or remove scedule. Depending on where
you apply this, it will override selections made at a lower level.
Print and Preview Sub-Sets are the same, as follows:
Point Labels – Create sample bottle labels for the entire database.
Sample Summary – Sample summary report for entire database.
Parameter Exception – A report similar to the parameter exception report created using the reporting
module. This report has been customized for oil data giving more detailed information.
Point Configuration Summary – This report checks for inaccurate or incomplete Point-level setup
information.
Alarm Setup Exception – Once the Point has been correctly setup then this report is used to check that
the alarms have been correctly applied to the sample data.
Oil Usage – A report showing oil consumption per Point and by oil type.
Schedule Summary – A report showing annual sampling schedule calendar.
Alarm Tree Filters - We recommend you uncheck all except the Oil (Tribology) box to eliminate
unwanted entries in the Alarm Details screen.

Alarm Details - There are two options:


When clicked at the Area level, opens the Alarm Details dialog box for the Area only. If you set the
Alarm Tree Filters to only show Oil (Tribology) as recommended, click the Use Alarm Tree Technology
Filters box at the bottom of the screen to show just Tribology alarms at the Area level.
Click the Select box at the right of the Area field to select from all Areas in the User database tree:

85 Doc. p/n 100-00087 Rev. A


When clicked at the Equipment level, opens the Alarm Details dialog box for both the Area and
Equipment. If you set the Alarm Tree Filters to only show Oil (Tribology) as recommended, click the
Use Alarm Tree Technology Filters box at the bottom of the screen to show just Tribology alarms at
the Equipment level.
Click the browse box at the right of the Equipment field to select from all Equipment entries for this
Area:

The Alarm Details right-click is not an option at the Point-level. If you want to show the alarm details
for a Point, open the Alarm Details screen at the Equipment-level and select the correct Point by click-
ing the browse button and selecting from the Point list:

Alarm Display Options - You can set the following preferences:

86
Display Color Scheme – Select either 3 or 5 levels for the Sample Info output and all other alarms. 5 col-
ors provides more granularity in the Trend Info and Sample Info primary tab output by providing more
alarm threshold points:

Show Alarm Severity icons – Select to enable or disable icon display. The default is enable.
Show an icon for no alarm or normal – Select to show or hide the No Alarm icon. The default is hide.
Show an icon when no analysis has been performed – Select to show or hide the Not Set icon. The
default is hide.
Automatic repaint interval in minutes – This feature is most useful on network installations. It controls
the frequency with which the icons on the database tree are automatically updated by reading the val-
ues from the database.

87 Doc. p/n 100-00087 Rev. A


This will allow changes made by one user to be seen on another user’s computer without the second
user having to close and re-open AMS Machinery Manager programs. The default is 0, which means
automatic repaint is disabled.

88
Reference Fluid Database Options
Each FluidScan has a substantial database of lubricants each identified by a unique Substance ID. It
uses the Substance ID to compare the chemistry of a pristine fluid against the chemistry of a used and
possibly contaminated sample of that same fluid and generates test results based on the comparison.
Refer to Reference Fluid Database on page 165
Any fluids not in the database can be created as "User Fluids." This process is described in the "Q1100
FluidScan User's Guide." The Substance ID is created by the FluidScan performing the fluid sample
analysis, and the four digit serial number of the FluidScan is embedded as part of the Substance ID. It is
important that you make a note of this Substance ID because you will need to enter it when you create
the Reference fluid and click the Trend Info primary tab - see Creating a Custom User Database on page
78
"User Fluids" could also be commercial fluids that a customer is adapting for their own purposes; use
in extreme environments, for example.
OilView ships with two databases: Example.ref and UserOil.ref, with Example.ref set as the
default.The Example.ref database is overwritten whenever AMS software is updated. To protect
against this, we recommend you do one of two options.
• Switch to UserOil.ref as described in Managing the User and Reference Databases on page 76
• Create a customized Reference database as described in the next section.

Creating a Customized Reference Database


Regardless of how many or how few oils and other lubricants you use, you can create a customized
Reference database. You can create as many custom databases as you need.
Follow these steps to create a customized Reference database and add a Reference fluid to it.
1. Click on the User DB tab.
2. From the File menu, click Create Oil Reference Database:

3. Enter the server name, or select one from the dropdown list.
4. Create a name for the database. For this example, call it Oils and Fluids.
5. Click OK at the confirmation popup.

Setting the Preferred Reference Database


Follow these steps to set the preferred Reference database:
1. Click the UserDB tab.

89 Doc. p/n 100-00087 Rev. A


2. From the File menu, click Set Oil Reference Database:

3. For this example, select Oils and Fluids.ref and click OK.

Adding Reference Fluids to the Reference Database


Follow these steps to add Chevron Delo 400 15W40 engine oil to the database. The Substance ID from
the Reference Fluid Database on page 165is 661.
1. Click the Ref DB tab.
2. Right-click the Oils and Fluids.ref folder icon and select Add Supplier.
3. Accept the default ID - A1, name it Chevron, and click OK to add it to the database.
4. Right-click Chevron and select Add Brand Name.
5. Accept the default ID - E1, and name it Delo 400, and click OK to add it to the database.
6. Right-click Delo 400 and select Add Oil Description. This opens the Ref Info primary tab at the
General secondary tab:

90
7. Accept the default Point ID - P1, and change the Point Description to 15W-40.
8. Place a check mark in the appropriate Base Stock. For this example, assume it is a Mineral
stock.
Note - If you have a Synthetic stock, it will activate the options in the Synthetic field in this
screen.
Optional: Click the Application secondary tab and select any Properties and Purpose check
boxes you want.
The Reference database now looks like this:

Creating the Reference Sample


9. Right-click P1-15W-40 and select Add Ref Sample. This is the sample-level that you will use to
assign the Reference fluid to the Point-level in the User database. Accept the default date and
we recommend you use the Substance ID as the Sample # - 661.
10. Click the sample you just created and click the Trend Info primary tab:

11. Scroll to the bottom of the Sample column and type the substance ID in the Substance ID cell:

91 Doc. p/n 100-00087 Rev. A


You now have a Reference fluid that you can use to associate with a Point in the User database
(below).

92
Associating a Reference Fluid with a Point in the User Database
Important: Associating a Reference fluid with a Point before you create any Samples below the Point
means that the chemical properties for the Reference fluid will be bound to every Sample you create for
this Point.
Important: You cannot create a Point, create your Samples for that Point, then go back and assign the
Reference fluid to the Point, because the Reference properties will not bind to the Sample-level after
the fact. There will be no Reference properties to compare against your test results.
When you try to test a sample with no Reference fluid properties, you will see an error message in the
FluidScan GUI: "Cannot find Reference 0."
If this happens, either delete the sample from the Point, associate a Reference fluid to the Point and
recreate the sample(s) or refer to the "Q1100 FluidScan User's Guide" and add the fluid as a "User
Fluid."
If you later change the Reference fluid associated with a Point, the historical samples will retain the
parameters of the original Reference fluid. New Samples will be bound to the properties of the new
Reference fluid.
Follow these steps to associate the Chevron Reference fluid you just created with a Point in the User
database.
1. From the User database, drill down to a Point that does not have a Reference fluid associated
with it. The Point icon will be blue.
2. Click the Assign Ref primary tab to display the Reference oil database:

3. Click the Reference sample to activate the Set Reference button:

93 Doc. p/n 100-00087 Rev. A


4. Click the Set Reference button to associate the Reference fluid with the Point. This will cause
the Point icon to turn white.

Assign Ref Secondary Tabs


Reference information from a number of secondary tabs is displayed whenever you are assigning a
Reference fluid to a Point, because they are associated with the Assign Ref primary tab:

Data Available Tab


The Data Available tab is a display of the instruments and tests for which Reference oil information has
been stored against a specific Reference oil sample. This display is for information only and con-
sequently greyed out.

94
Data Tab
The Data tab is a single column of Reference oil parameters and associated measurement values.
This list is the same as the Data tab under Trend Info primary tab when you select a Sample from the
User database:

95 Doc. p/n 100-00087 Rev. A


Oil Type Tab
The Oil Type tab shows whatever you selected from the Base Stock column of the Ref Info primary tab
when you were creating the Reference oil description.

96
Properties Tab
The Properties tab shows whatever you created in the Properties column of the Ref Info primary tab
when you were creating the Reference oil description.

97 Doc. p/n 100-00087 Rev. A


Purpose Tab
The Purpose tab shows whatever you created from the Purpose column of the Ref Info primary tab
when you were creating the Reference oil description.

98
Stat Plot Tab- Histograph Option
This plot shows the reference value relative to all samples that use this reference.
This is used for verifying that the reference value is appropriate for the samples to which it is being
applied.

This feature can also be used to set a reference value by either dragging the vertical line to the appro-
priate value of by pressing the Set Ref Value button, which would set it at the median value of 50%.
Save the new value.
You can also net all reference parameters using the Set/Save All button and establish reference values
when no Reference oil is available.

99 Doc. p/n 100-00087 Rev. A


Stat Plot Tab - Stat Data Option
This displays the data used to create the histogram plot and is used to identify these data points that
appear to be incorrectly assigned:

100
Changing a Reference Fluid Associated with a Point
For this example, assume you have a truck working exclusively in Florida and running happily on one
type of engine oil. Then it is sent to work in Alaska, so the likelihood is that you would want to fill the
reservoir with an entirely different engine oil.
Or, as sometimes happens, a manufacturer changes the formula of a particular oil.
You can associate the new Reference oil by following the procedure described above, but, the
properties of the new Reference oil will only propagate down to samples that you create after the
change.
Any historical samples that were created before the change will retain the properties of the old
Reference oil.
The only Reference oil that matters is the one assigned at the Sample level. If you change a Reference
oil value or an alarm threshold and re-run a test on a sample assigned to the old Reference oil, you will
get new test results.
For example, suppose the dielectric on a Reference oil is reformulated down by 0.10 units. When you
recalculate the alarms for a database, then all the places where that Reference oil is assigned will imme-
diately show an increase of 10.0 in chemical index (Chen index = delta dielectric x 100) and almost
every test result will show an increased Chemical Alarm Status.
However, if you add a new Reference oil and assign it to the Point level but you do not change the Refer-
ence oil assigned to all the historical samples, then nothing will be changed when you recalculate the
alarms.
To correct this,
1. Click the Point you want to change and click the Assign Ref primary tab.
2. Locate the new Reference sample and click the << button.
3. Click the first historical sample to be changed to the new Reference oil.
4. Locate the new Reference oil and click the << button.
5. The Recalculate Alarms dialog box opens.

6. Click the Start button to recalculate the alarms for this sample.

101 Doc. p/n 100-00087 Rev. A


Copying and/or Moving Reference Fluids
A Reference fluid created at the Oil Description levelcan be associated with another Brand Name with a
simple copy and paste or drag and drop operation.
Click and hold the left mouse button on the new Reference oil. Then move the highlight data to another
Point and release the mouse button. You will be prompted with a message to continue; click OK.

Maintaining Historical Sample Data While Updating Reference Fluids


Suppose you switch from using Chevron to Mobil and you want to re-assign Reference fluids for
Points which had previously used Chevron. In this case, the historical data samples must keep the old
Chevron Reference oil.
However, all new Samples added to the Point should use Mobil as the Reference oil. Instead of manu-
ally assigning the Mobil Reference oil to each Point, the software allows the old Chevron Reference oil
to be superseded by the new Mobil Reference oil on every Point in the current user database where
Chevron was assigned.
After adding the new Reference oil and saving the data, simply use the mouse to drag the new Refer-
ence oil sample and drop it onto the old Reference oil sample in the Reference tree on the Asssign Ref
primary tab.

102
Overwriting Reference Fluids for Point and Sample
A somewhat different situation arises when you merge Reference fluid databases and you now have
duplicate Reference samples. In this case, you can supersede the Reference fluids assigned at the Point
level and Sample level for future samples.
One of the duplicate References will be assigned and the other duplicate references will be deleted.
To do this, right-click at the top level of the Reference database tree on the Assign Ref tab, and then
choose the Enable Global Supersede and Lock DB menu item.
You will see an information dialog box similar to this:

After this, drag and drop a Reference oil sample to assign that particular fluid to all Samples and Points
where the old Reference oil was previously assigned. After completing the global supersede function,
right-click at the top level again to Enable Default Supersede and Unlock DB.

Show Reference Samples Used by this Database


Right click the top level icon of the Reference database tree on the Assign Ref primary tab to open the
Show Ref Samples used by this Database option.
Once the tree has been expanded, then all Reference fluid samples used by the current database will be
highlighted in green.

103 Doc. p/n 100-00087 Rev. A


Querying the Database
This option requires a license for the LIMS module.
To query the database, select the Q tab at the bottom of the database column. Right-click on Cus-
tomer Query and select either Customer Sample Query or Generate Schedule:

Customer Sample Query


Select Cust IDor Cust Desc from the drop-down menus. You can also select various reporting con-
ditions as Show and a Start and an End date. Press the Start button to run the query:

®
The results of a query are shown as a list similar to the LIMS Batch Info display. You have full access to
all the reporting options for the list; refer to Batch and Route Testing on page 29

Generating a Schedule
This menu option generates a list of samples that are due based on the Current date and Look Ahead
date.
Press Start to generate the list:

104
105 Doc. p/n 100-00087 Rev. A
8 Point Info Primary Tab
This is the default primary tab that you would use to create a new Point-level entry beneath any piece
of Equipment defined in the User database.
You can also use it to edit an existing Point-level entry, as it will open by default whenever you click on
a Point in the User database.
This chapter describes the following:
General Tab, below
Scheduling Tab on page 109
Test and Display Filter Tab on page 110
Internal Tab on page 111
Process Tab on page 112
External Tab on page 113
P1 Tab on page 114
P2 Tab on page 115
P3 Tab on page 116

General Tab
Provides general information about a particular Point.

106
Point ID - A three-character designator you select to differentiate between all the Points below the
Equipment-level.
Description - This field of up to 28 alphanumeric characters describes the current point and should be
as descriptive and unique as possible.
Lab Unit ID - This field can be used to enter an alphanumeric identification when using an oil lab other
than the CSI Trivector Laboratory. Note: This identification must be unique throughout all of your data-
bases.
Report Language - The language used to translate the test on the Sample Trend and Minilab reports.
The diagnostic must be written in the selected language. The other test is translated on the Setup
primary tab - Translation secondary tab.
Report Group - This field selects the report group (1 - 8) that determines the layout of the sample
report. Refer to Analysis Parameter (AP) Sets and Alarm Limit (AL) Sets on page 59
Usage Units – These units are used to specify the amount the unit (measurement point) has been in
operation (for example, the number of hours, days, distance it has been running). This unit of measure
applies to both the quantity of usage for the unit as well as the quantity of usage for the oil in the unit.
This should not be confused with the units below that refer to the quantity of oil in the unit.
Target Cleanliness - The ISO cleanliness code - 2 micron/5 micron/15 micron - that this equipment
should be operating at.
Lubrication System, Capacity – This is the oil capacity of the system and will be used to calculate oil
usage and system debris.
Lubrication System, Units – These units are used to specify the volume of the oil compartment as well
as the units used when entering the amount of oil added when a new sample is taken.
Analysis Parameter Set - This is the listing of all of the measurement parameters for which you intend
to collect data. Click on this field to assign an AP Set from the list of predefined sets to the selected
measurement point.
We recommend that you use the Spectro AP Set as we have created list AP Set of best parameters
based on experience and testing. You can have up to 96 parameters in an AP set, so you have room to
customize, or even create a totally custom AP Set if necessary. Spectro supplies default AL Sets
matched to the Spectro AP set; this is not the case if you create your own AP Set. Refer to Analysis
Parameter (AP) Sets and Alarm Limit (AL) Sets on page 59 for more information.
Equipment Type - Clicking on the button on the right side of this field brings up a tree list of Equipment
types. Once you have selected the Equipment type, you must then select the corresponding AL Set.
Alarm Limit (AL) Set - Used to assign an AL Set from the pop-up window containing a list of predefined
sets to the current Point.
The AL Set is the list of alarms for the Point. If you have selected the Spectro AP Set, the AL Set will
automatically be generated when you select an Equipment Type:

107 Doc. p/n 100-00087 Rev. A


If this relevant AL Set does not exist you must go to the AP and AL Set branch of the database tree and
add the required AL Set:

Note: We suggest when specifying a new Equipment-level entry, you begin by using the default AP
and AL Set for the first three months or so at your discretion. Then gradually increase or decrease
alarm limits so that approximately 10% of your equipment is in the highest alarm level and another
10% is in the second alarm level.

108
Scheduling Tab

Set up sampling schedules with a right click on the appropriate date. You should setup the schedule
for at least one year in advance, although you can set one up for as far in the future as you want.
Printing Sample Labels for Scheduled Samples
Printing of sample labels for due samples is done on the Q (Database Queries) tab at the bottom of the
database tree.

109 Doc. p/n 100-00087 Rev. A


Test and Display Filter Tab
This filter can be used to filter out unwanted analysis parameters from being displayed in the test ana-
lysis output. Only the analysis parameters with a checked box will display in the report.
The filter is organized by measurement device. This way you can turn off the display of blank inform-
ation when you are not conducting particular tests. If you begin collecting these data later, simply turn
the display back on.
Refer to Displaying Test Results on page 48

Mini-lab Tests -
Criticality - The relative impact of a failure of this equipment on overall plant operation. This function
enables you to specify one of five levels of criticality for this piece of equipment.
All Other Tests-

110
Internal Tab
You can use these fields to document reference information that may be useful in detailed oil analysis
and root cause diagnosis.

Oil Pressure - Indicate the system oil pressure as:


Low = <500 psig
Medium = <2000 psig
High = >2000 psig
Oil Temperature - Oil temperature.
Separator - The type of water/oil separator that the piece of equipment uses.
Oil Change - Used to decide when to do an oil change.
Coolant - Indicate whether coolant is used and if so, what type.
Filter Category - Select either Filtered or Not Filtered.
Actual - If known, enter the actual filter specification in terms of micron size and beta ratio.
Oil Capacity: Estimate - Enter the approximate oil capacity of the system.
Oil Capacity: Actual - If known, enter the actual oil capacity of the system.
Oil Capacity: Breather - Indicate if there is a breather on the oil compartment.
Usage Units - These units are used to specify the volume of the oil compartment as well as the units
used when entering the amount of oil added when a new sample is taken.

111 Doc. p/n 100-00087 Rev. A


Process Tab

General - If process contamination of the oil is possible, identify if this can cause the oil to become
chemically neutral, acidic, caustic, or unknown.
Gas - If the processed material is a gas, enter the kind.
Liquid - If the processed material is a liquid, enter the kind.
Solid - If the processed material is a solid, enter the kind.
Transformer - If the Equipment is a transformer, indicate the rated features.

112
External Tab
This allows you to select parameters for the physical working environment.

Type - Indicate how corrosive the external environment is.


Lube System - Specify whether the system is opened or closed.
External Temperature - Indicate the ambient temperature:
Hot = >140oF (60oC)
Temperate = 40oF (5oC) to 140oF (60oC)
Cold = <40oF (5oC)
Particulate - Indicate if the area contains abrasives that may be problematic.
Humidity - Indicate the ambient relative humidity:
Dry = <40%
Moderate = 40% to 89%
Very Humid = 80% to 95 %
Wet = 100%

113 Doc. p/n 100-00087 Rev. A


P1 Tab
Displays different parameters applicable to oil-wetted (specifically bearing) parts:

Bearings - Indicate if bearings are used in the system.


Speed - Enter the speed range of the piece of equipment.
Sleeve (Conformal) - Indicate what type of conformal sleeves (if any) are used.
Roller Cage - If a roller cage is in the bearing, indicate it’s material.
Other - Select any other elements that may be present in the oil system.

114
P2 Tab
Displays different parameters applicable to oil-wetted parts:

Gears - Enter the composition of any gears.


Seals - Enter the composition of any seals.
Cylinders/Pistons/Rings - Enter the composition of any of these components.
Shafts/Slides/Cams - Enter the composition of any of these components.

115 Doc. p/n 100-00087 Rev. A


P3 Tab
Displays different parameters applicable to oil-wetted parts:

Turbine Blades - Enter the composition of any turbine blades.


Oil Pump - Enter the composition of an oil pump.
Other Parts - Enter the composition of any other parts as required.

116
9 Data I/O Primary Tab
This section describes various ways to import and export information using the Data I/O secondary
tabs.
Note: You must configure your E-mail connection in the Setup -> E-mail tab before you can connect.
This chapter describes the following:
Import Lab E-mail Tab, below
Import Lab File Tab on page 118
Database Export Tab on page 118
Statistics Export Tab on page 119

Import Lab E-mail Tab

Logon - Press this button to log onto your E-mail system. Once you have logged onto the E-mail then
this button will change to Logoff so that you can disconnect once you have completed importing data.
Import - Once logged-on, the system will search for all E-mail messages whose subject contains the
phrase Lab Data. This enables the system to distinguish E-mail messages containing lab data file attach-
ments from other E-mail messages. A list of these E-mail messages is then displayed. To import the
data from each E-mail, click on the E-mail message to select it and then press the Import button.
Once the data have been imported, the attached file containing the data is automatically archived in the
Lab Export/Import directory which is setup under File Locations on the Setup tab. The E-mail message
is then automatically deleted. If for some reason you need to re-import that data then use the Import
Lab File tab to import the data from the file in this location.
Delete - Use this button to manually delete an E-mail message.
Cancel - This button stops an import before it is complete. There may be a short delay before the
import stops.
Lab File Type - See next section for details.
Import Into - See next section for details.

117
Import Lab File Tab
This tab is used to import lab data from oil laboratories. Before using this function, you must have
already obtained the file to be imported. Sources might be the internet, an E-mail, a USB drive, etc.

Lab Type - These are the currently supported download formats. If your oil laboratory is not rep-
resented, Spectro can provide them with a specification for the Generic file format.
Data File - allows you to browse the computer to find the file that is to be imported.
Import Into - This option either selects the current database (fastest) or all the databases in the RBMad-
min group.
Start - Click to begin the import process.

Database Export Tab


This tab is used to export data from the AMS Machinery Manager database to a file that can be impor-
ted into a spreadsheet or relational database. The data exported will depend on the current position in
the database tree.
For example: to export the entire database, the database level would be selected. To export just one
sample, that sample would be selected.

Save As - The name of the file that will contain the exported data.
CSI Generic (.zip) using the Lab Unit ID as the key for data import

118 Doc. p/n 100-00087 Rev. A


Area, Equipment, Point (.zip) - codes as keys for import. This option automatically creates the Area,
Equipment or Point in the target database if they do not exist.
Comma Delimited - output file.
Entek IRD format.
A/E/P with Cust (.zip). This option includes the customer level in the export.
DB Structure This option exports only the database structure down to the point level without any
sample information. This is useful for duplicating database structures.
AP & AL Sets this option exports AP and AL Set data. You must first assign AP and AL Set ID’s
because they are used to identify the AP and AL Sets when importing.
Important: When exporting sample data using the A/E/P option, the reference oil and AP &AL Set ID’s
are included in the generic file. This enables the correct assignment of reference oils and AP and AL
Sets on import. It does depend on the target database having the same ID’s.
Date Filter - Check the Enabled box to limit the export to data that falls on or within the selectable Start
Date and End Date.
Change Mark - Only exports sample for which the change mark has been set.
DB Name – The name of the .RBM (User) database into which the data is to be imported.
RBM DB Target – Data will be imported into the .RBM (User) database.
REF DB Target – Data will be imported into the Reference database.

Statistics Export Tab


This tab is used to export data for the Alarm Limit statistics option.

119
10 Setup Primary Tab
The Setup primary tab and its associated secondary tabs allow you to specify operating characteristics
of the OilView hardware and software.
This chapter describes the following:
System Tab, below
Trend Tab on page 121
Translations Tab on page 122
Severity Codes Tab on page 123
Setup Primary Tab on page 120
DB Level Names Tab on page 124
Labels Tab on page 125
Lab Labels Tab on page 126
Reports Tab on page 127
E-mail Tab on page 128

System Tab
This allows the system administrator to set the security status for data editing and data entry within
the OilView software. This is also where you assign someone Advanced User status.
The remaining functions are not used in the normal operation of the OilView software. They are tools
that you may need to use when communicating with customer support.
Read Only - You are allowed to download data into the OilView database, and the ability to view and
print plots and reports. .
Advanced User - This option places the software in the Advanced User mode and allows the admin-
istrator of the OilView software to set up or edit Areas Equipment, Points, and other technical inform-
ation. When this option is selected, the “grayed out” primary tabs that are not available to the “read
only” user are now accessible.
Pause on Daily Tip – If you do not have time to read the daily tip this option will pause the software for
you.
Mini-Lab Demo Mode – Switches the software into demo mode whereby the running of the MiniLab
instruments can be simulated without actually having them connect to the computer.
Let lab alarm levels override user alarms on import – By default when importing lab data the lab sever-
ity is stored with the user severity, which is calculated based on alarm limits. This option disables the
calculation of user severity based on stored alarm limits and instead sets the user severity equal to the
severity imported from the lab. This is essential if the alarm limits have not been correctly set up.
Let user alarms override lab alarms at overall level – When importing data the user set alarms are used
to override the overall (Wear, Contamination and Chemistry) alarm severities.
Disable RBMview connection (restart after change) – On slow networks and if not using RBMview for
reporting, then this option disables the connection. You have to restart OilView for this to take effect.

120
Disable auto problem creation in RBMview – When diagnosing a sample in Sample Info tab , the obser-
vations and actions are automatically sent to RBMview. This option disables auto sending and requires
the user to manually post the diagnosis to RBMview if required.
Disconnect Actions from Observations – There are times when the actions associated with obser-
vation may not be appropriate. This option disables the relationship. The user must then manually add
the required actions for the selected observations.
Print WDA Analyst diagnosis grid – The WDA analyst report will append the WDA matrix grid if this
option is selected.
Skip Edit on Sample Add – The sample will be added without displaying the edit sample screen.
Rename Log File - This function allows you to rename the current log file. This feature can be used, for
example, to create multiple log files during a software troubleshooting session.
Clear Log File - This function clears the existing log file and begins writing new data.

Trend Tab
All Data – plot all data on trend plots.
Last – plot the last “n” samples on trend plots. You select the number of samples.
Start Date –Plot all sample data after the start date on trend plots.

121 Doc. p/n 100-00087 Rev. A


Translations Tab
Reports can be printed in multiple languages. The specific language is setup at each point. Different
points can have different languages assigned to them.
Note: These reports require that the diagnostic message be written in the appropriate language for
each point.

122
Severity Codes Tab
During the diagnosis process there are up to five levels of severity ranging from Normal to Extreme.
Note: Default terms for these levels are provided; however, you can use this window to specify your
own terminology.
You can also choose to use only three severity levels.
When renaming the severity codes with this option, the actual levels are not changed, only the name
associated with the particular severity code.
Severity is ranked from 0 = Normal to a value between 1 - 100:

File Locations Tab


This tab is used for WAN only; refer to the AMS Online Help for more information.

123 Doc. p/n 100-00087 Rev. A


DB Level Names Tab
These options enable you to globally substitute other descriptive names used in the databases.
Note: Do not use a keyword to replace another keyword. For example, Equipment should not be sub-
stituted for Area.
If you decide not to keep the new names you have selected, click on the Restore Default Level Names
button to restore the names to factory setting.

124
Labels Tab
Use this function to set up the dimensions of the labels being used on sample bottles.

Label Printer - Select the appropriate printer. Labels can either be printed form sheets or single labels
depending on the printer selected.
This enables printing labels on a specific printer rather than the system default printer.
Default Labels - Select a default label size.

125 Doc. p/n 100-00087 Rev. A


Lab Labels Tab
The lab labels identify the sample and are used during the testing phase in the laboratory.
The Test Matrix check box lists the tests to be performed based on the Point Setup

Default Labels - Select a default label size.


Label Printer - Select the appropriate printer. Labels can either be printed form sheets or single labels
depending on the printer selected.
This enables printing labels on a specific printer rather than the system default printer.

126
Reports Tab
The name and address appearing on several reports may be customized as shown below. You may also
use your own logo by specifying the location of a graphic.

Report Colors - This sliding color level is used to tone down the colors on trend plots or on printed
reports.

127 Doc. p/n 100-00087 Rev. A


E-mail Tab
The LIMS Module is required to interface with Spectro Scientific instruments. It requires an internet
connection for E-mailing data and reports. The Data I/O primary tab also requires an internet con-
nection to download data directly using E-mail; refer to WAN Setup for OilView on page 1 for more
information.

Your internet service provider will provide you with a user name, E-mail address, password, SMTP
host name and POP3 host name. Enter these in the appropriate fields and then test the connections
using the Test SMTP and Test POP3 buttons.
Select Fax Printer - LIMS enables the automatic faxing of reports. This is enabled by setting up a fax
printer.

By default, the tree will show the status for all available technologies. However, applications may set a
filter to limit the number of technologies for which status is displayed. For example, OilView may
choose to set a filter to only display Oil Technology statuses.
You can also click Use these filters for all programs to turn on the filters for all the technologies. Doing
so changes the dialog box heading to Global Alarm Tree Filters.
LIMS
Lab Options – Reserved for the Trivector lab.

128
Display Customer Level – The customer level in the RBM database groups areas under a given cus-
tomer. This option is activated here.
Print lab label @ login – This option activates the automatic printing of lab labels one at a time as
samples are logged into the LIMS batch system. You need a label printer to use this option.

129 Doc. p/n 100-00087 Rev. A


11 WDA Primary Tab
The Wear Debris Analysis (WDA) tab is used to assist in analysis and documentation of visual micro-
scopic observations of wear debris and contamination extracted from oil samples. Wear debris ana-
lysis is implemented using a patch maker to collect debris which is then viewed under a microscope.
Various video cameras and frame grabbers may be used, but the key to the software interface is that
the frame grabber software must provide an industry standard TWAIN driver dll along with the card.
The software can import JPG image files regardless of their source.
This chapter describes the following:
WDA Primary Tab, below
Microscope Setup and Calibration on page 130
Recommended Video Specifications on page 130
Sample Tab on page 132
Capturing Images Through the Microscope on page 133
Measuring Image Size and Particle Density on page 135
Measuring Particle Density by Counting Particles on page 136
Atlas Setup Tab on page 138
Analyst Tab on page 139
Working with Filter Patches on page 140
Making Patches on page 141

WDA Overview
Wear debris analysis is performed on an oil sample and may include one or more photomicrograph
images. Some of the actions within this tab are pertinent to a specific image and others are pertinent
to the oil sample which may have no images, one image, or several images.
In effect, wear debris analysis introduces another level below the Sample level on the User Database
tree—the Image level.
Important: You need to pay particular attention when you are recording information in the WDA
primary tab as to whether it is being applied to one specific image or to the sample.

Microscope Setup and Calibration


Documentation on the CD that came with your equipment includes the "Stereoscopic Zoom Micro-
scope Setup and Calibration Guide" for information on how to set up and calibrate the microscope
(Spectro Scientific document p/n 100-00088).

Recommended Video Specifications


The following are recommended video specifications:
• Composite NTSC, PAL, or S-Video input
• 24/16/15 bit RGB video digitizing
• 640x480 NTSC (768x576 PAL) digitizing resolution
• 30 frames/sec, 60 fields/sec

130
• Full bandwidth PCI bus mastering read and write (up to 132 Mb/sec)
• One available PCI 32-bit bus master expansion slot
• Interpolated scaling in x and y directions
• Digital control of offset, gain, brightness, contrast, hue, and saturation

131 Doc. p/n 100-00087 Rev. A


Sample Tab
The Sample secondary tab is used to analyze and document each individual image as well as the overall
oil sample.

These data pertain to the image and to the sample:

Feature Image Sample

Photograph with tool bar X

Image Caption X

Image Note X

Wear Debris Analysis Sample Dia-


X
gnosis

Analysis Grid with 5 Groups X

132
Capturing Images Through the Microscope
This diagram shows the options available on the toolbar. Add, Capture and Annotate are described in
this section; the others are described in the Measuring Image Size section that follows.

Photographs taken through the microscope are captured using the Capture Image button.
Photographs can also be imported using the Add Image from External File button.
Multiple images can be scrolled through using the left and right arrow buttons.
Each image can be given a caption and note with the Annotate button. These notes and captions
remain associated with the image and are not passed on to the sample record.
Notes added to the WDA Sample Diagnosis field are copied to the sample record under the Sample Info
tab. These notes apply to any or all images for a given sample.
The WDA grid includes five pages. Each page represents an analysis of a different group of particles.
To do this analysis you can select one item from each column for each group which most closely
describes the group.
There are ten categories of analysis for each group of particles:
Concentration - A relative comparison of the concentration or amount of particles in the group.
Average Size - The average size in microns for the particles represented in this group.
Max Size - The maximum size particle in the group.

133 Doc. p/n 100-00087 Rev. A


Aspect Ratio - The maximum to minimum dimensions ratio for a particle (e.g., length to width).
Shape - The three dimensional shape of the particles.
Texture - The surface texture of the particles.
Composition - The physical composition of the particles.
Classification - The wear mechanism classification of the particles.
Severity - The overall severity of wear condition based on this group of particles.
Concentration, average size, maximum size, and severity are all trend parameters that are in the CSI
Default AP Set. We recommend that you do not apply alarms to any of these except the severity level.
The others are used only for reference during analysis.
You can use the Assist button to automatically add selected categories to the WDA Sample Diagnosis
note after the items are selected. You are free to edit this note field. These automatic entries are
provided for your convenience.
Instead of selecting individual items from the WDA grid, many people choose to use the wear debris
analysis atlas to select an image that most closely represents the one being analyzed. The analysis of
the image in the atlas can be copied over to any of the five groups automatically.
You are then free to edit the results to correct any differences between the atlas image and your actual
image.
This is described in the following sections:

134
Measuring Image Size and Particle Density
Particle sizes as defined as follows: Fine <6u, Small 6-14u, Medium 14-40u, Large 40-100u, and Huge
>100u.
Important: Before you can measure the size of particles in the captured image, you must calibrate the
microscope as described in the microscope setup and calibration guide on the document CD.

Important: You must enter the magnification and zoom settings when the image is captured.
To measure the particle size,
1. Click the Measure particle size button.
2. Position the cursor near the particle, click the left mouse button, and then draw a rectangle
around the particle. The size of the particle is shown in the popup dialog box.

135 Doc. p/n 100-00087 Rev. A


Measuring Particle Density by Counting Particles
To measure the particle density (particles per mm 2),
1. Press the Measure particle size button and then draw a rectangle around the region containing
the particles.
2. Press the Count particles with mouse click button. Now you can visually count the particles by
pointing the mouse cursor to each particle and clicking. This automatically counts the particles.
Or you can count the particles and manually enter the number to get the particle diversity.
When finished, this information can be used to fill in the WDA matrix.

136
Sample Atlas Tab
This tab is used to compare images from the WDA Atlas with sample images and to import analysis res-
ults accompanying the atlas image over to a selected group in the sample image:

The left photograph is functionally the same as it is on the Sample screen, except that you cannot edit
the caption, notes, or select through the wear debris analysis grid.
The right photograph is one from the atlas. If no atlas image is present then you may scroll through the
tree on the lower right to select and view until you see one that looks like the sample image.
The Show Analysis button displays the WDA matrix settings for the selected atlas image.
Once you have found an image that has features you want ported over to the WDA, select one of the
five groups on the lower left and then select the <<< button. This automatically replaces entries in
that analysis group with those from the atlas image.
Go back to the Sample tab to edit these entries to exactly fit your analysis of the sample.

137 Doc. p/n 100-00087 Rev. A


Atlas Setup Tab
This tab is much like the Sample tab except that it is used to add or edit the wear debris analysis atlas.
Note: You must have Advanced User status to edit the atlas.

The tree on the upper left is used to select the image to be viewed, and the note field below the tree
contains information relevant to this image.
The wear debris analysis grid below the image represents the most obvious group of particles in the
atlas image. There is only one group of particles represented in this grid.
To add images to the atlas, first use a right mouse click on a tree folder and select the Add Child Folder
option. Then use either the Add file or Capture buttons below the photograph to add an image to the
folder. These work the same as they do for the Sample images.

138
Analyst Tab
The Analyst tab displays multiple sample images simultaneously for easier classification and diagnosis:

139 Doc. p/n 100-00087 Rev. A


Working with Filter Patches
Select the pore size(s) for the filter patch(es) based upon the estimated or measured particle size range.
Standard filter patch pore sizes are 3 micron and 8 micron. Other pore sizes ranging from 0.8 micron to
30 micron are readily available and may be used.
A few tests will quickly reveal what pore sizes tend to plug and which ones yield meaningful spe-
cimens. The goal is to produce wear debris analysis patches with particles distributed so they can be
easily viewed with a microscope.
Using Stacked Filter Patches
You can separate large particles from smaller particles using stacked filter patches. For example, if the
sample is highly contaminated you may choose to use a 30 micron filter above a 10 micron filter. Keep
in mind that the small pore size filter is most often the one that is plugged from excess contamination.
Ferrous / Non-Ferrous Separation
If you wish to view ferrous and non-ferrous particles on separate patches, then use the sheathed mag-
net to extract ferrous particles from the diluted sample mixture before ever adding it to the patch
maker chamber. After removing the ferrous particles, you can pour the remaining diluted sample into
the chamber and prepare the non-ferrous patch. Remove the non-ferrous patch and insert a new patch
for the ferrous particles. Then remove the magnet from the sheath, and rinse the ferrous particles off
of the sheath into the chamber with solvent. Flushing this time will create the ferrous patch.
Diluting Samples
Important: All samples must be diluted before they are put into the patch maker. A very small amount
of actual sample is required. About 2 ml of sample should be used consistently, unless a sample is
extremely clean or extremely dirty. Shake the sample thoroughly and then extract 2 ml using a clean
pipette. Add this to a clean mixing bottle or beaker, and then add 10 to 20 ml of non-volatile mild
solvent (lamp oil will suffice).

140
Making Patches
Follow these steps to make a patch:
1. Prepare filter for rinse solvent - Unscrew the two sections of the 25 mm filter holder, and use
forceps to center a 25 mm white filter disc on the perforated plastic support in the lower
section of the filter holder. Place a stainless steel screen on top of the filter, and reassemble the
holder:

2. Prepare filtered solvent dispenser - Remove the plastic cap from the spout of the wash bottle
and twist a 25 mm filter holder onto the spout. Fill the bottle with the proper solvent for the
hydraulic fluid to be tested, and close the bottle with a vented screw cap. Use the spout cap
and the solid screw cap to prevent spillage when storing the filled bottle:

3. Assemble test filtration apparatus - Remove the 47 mm filter holder assembly from the vacuum
filtration flask if necessary, using the butt end of the forceps to pry against the flask rim at the
cutout slot. Invert the holder assembly and install it on theflask, with the funnel up and the
seating holder base flange resting evenly against the flask rim:

141 Doc. p/n 100-00087 Rev. A


4. Wash inside the funnel - Close the vent hole in the solvent dispenser cap by covering it with a
finger tip, and squeeze the bottle to direct a jet of filtered solvent against all the inner surfaces
of the funnel. Use enough solvent to carry all the contaminants through the filter holder screen
and into the flask:

5. Install the test filter - Remove the funnel from the test filter holder by turning the knurled ring
from right to left until free, then lifting. Using forceps, center a 47 mm white test filter disc on
the wire mesh support screen. Replace the funnel and retighten the knurled ring to seal the
filter in place:

142
6. Attach syringe vacuum pump - The free end of the flexible tube attached to the syringe is fitted
with a small nylon adapter. Insert the adapter firmly into the small hole on the side of the filter
holder base. The tube and adapter are normally left attached to the syringe for storage, but can
be removed for cleaning:

7. Rinse sample bottle and collect sample - Rinse out the inside of a clear plastic bottle with
filtered solvent, then discard the used solvent. Fill the sample bottle to the shoulder point with
the test fluid and cap securely:

143 Doc. p/n 100-00087 Rev. A


8. Filter test sample - Shake the sample bottle. Empty the contents into the funnel, pouring
against the funnel walls to distribute the fluid evenly over the surface of the filter. Then, draw a
vacuum inside the flask by working the syringe plunger until the level of fluid in the funnel
begins to drop, indicating the start of filtration:

9. Rinse the contamination from the sample bottle - Fill the test sample bottle one third full with
filtered solvent, shake thoroughly, and pour the solvent into the funnel as soon as the level of
fluid in the funnel permits. Operate the syringe again to sustain filtration if necessary, but do
not allow the level of fluid to drop to the filter surface before you do Step 11 below:

144
10. Rinse contamination from funnel walls - When the sample level in the funnel drops to the
narrow neck of the funnel, wash down the inner surfaces of the funnel with a stream from the
filtered solvent dispenser. Note: Do not let the solvent stream directly strike the test filter:

11. Check rinse-through of sample fluid - Continue filtration until all the fluid has passed through
the filter. Remove the funnel, leaving the filter in place, and inspect the filter surface. If the
central area of the filter is pink, indicating insufficient rinsing, replace the funnel and aim a
stream of filtered solvent against the funnel walls until fluid reaches the top of the funnel neck.
Work the syringe again to draw all the fluid through the filter.

145 Doc. p/n 100-00087 Rev. A


12. Check the filter patch by placing the filter in a microscope and importing the image into OilView
by clicking the microscope icon in the WDA tab:

146
12 Troubleshooting
This chapter provides information that may help you troubleshoot issues.

Send to Instrument and Import Results Buttons are Disabled


This can happen when you have created a sample list and are ready to export the list to the device
and/or import the results back into OilView.
To resolve it, make sure the device GUI for the FluidScan and\or Viscometer, and\or the
LNF management software GUI for the Q230 is reset to the Main Menu. This will activate these but-
tons.

<Device Name> is Performing Another Operation


At various points during the testing phases for all three devices, you might see this message:

The device is still working correctly, but it is hanging and waiting for a response from you.
To resolve it, make sure the device GUI for the FluidScan and Viscometer, and the LNF management
software GUI for the Q230 is reset to the Main Menu. This will clear the message and allow you to con-
tinue.

Error Tracing
These options are for use only when working with Technical Support and you are resolving a problem.
Test Modes
Diagnostics – Enables MiniLab 53 instrument diagnostics. These features are used for diagnosing prob-
lems in conjunction with Technical Support.
Add Sample With Start - When a new test is started with any of the MiniLab 53 instruments, you are
automatically prompted to enter a new sample number. When this box is checked, if the sample
already exists then it is used; otherwise, a new sample is added.
Verbose – More detailed errors and warnings. These features are used for diagnosing problems in con-
junction with Technical Support.
Data Collection Mode – Used in a WAN installation
Data Analysis Mode – Used in a WAN installation
Data Edit/Entry
This function is used to change the security level in the OilView software.

GUI Does Not Display Properly for Q230


If the GUI does not display properly, open the LNF management software and navigate to the Export
Control tab. Confirm the Enable Q5300 Interface check box is checked.

147
"Cannot Find Reference 0" Message
If you see this message in FluidScan, it is because you created the Sample before you assigned a Refer-
ence fluid at the Point level.
You can follow the instructions in the GUI to locate the appropriate Reference oil in the internal data-
base and continue the test. but this is only recommended for a very small number of samples and it will
not work if the sample is part of a batch that will be re-tested over and over or is likely to become part
of a batch.
The best alternative is to locate the Point, delete all samples created beneath it, assign the Reference oil
to the Point, then recreate the samples. This will propagate the Reference oil to the sample level.
Refer to Managing the User and Reference Databases on page 76 and

Too Many Frames Message


In some cases, the test sample can be so contaminated that it overwhelms the Q230 and the following
message is displayed in the LNF management GUI:

Follow the instructions in the popup to resolve the problem.

148 Doc. p/n 100-00087 Rev. A


13 Oil Properties Table
Property UOM Keyword Description Generic Tag C OV Params
SUBID 1 Substance ID
V40 cSt Visc40 Visc at 40 degrees VISC40 Visc40C
ASTM 
Oxi abs/mm2 IROXIDAT IROXIDAT IR Oxidation
Oxidation
Nit abs/mm2 IRNITRAT ASTM Nitration IRNITRAT IR Nitration
Sul abs/mm2 IRSULFAT ASTM Sulfation IRSULFAT IR sulfation
TBN mgKOH/g TBN TBN TBN Total Base
TAN mgKOH/g TAN TAN TAN Total Acid
Water ppm WATERKF Water WATERKF WaterK.Fish
Gly %vol FS_GLYCOL Glycol FS_GLYCOL Glycol Vol
Additive
AWAdd %depleted FS_AW Additive depletion FS_AW
Depletion
Turbo Antioxidant AntiOxid Deple-
AOx %remain FS_AOX FS_AOX
Depletion Status tion
FRW abs/mm2 FS_FRW Free Water FS_FRW Free Water
Bub Abs FS_BUB Bubbles in Oil
ASTM Antiwear
AWc abs/mm2 FS_AWC FS_AWC Antiwear Comp
Components
WatA abs/mm FS_WAT ASTM Water IRWATER IR Water
GlyA abs/mm2 FS_GLYA ASTM Glycol IRGLYCOL IR Glycol
Awa abs/mm2 FS_AWA ASTM Antiwear IRANTIWR IR Anti-Wear
StA Abs*100 FS_STA ASTM Soot IRSOOT IR Soot
Ar1 % FS_AR1 Area I Check FS_AR1 Area I Check
Ar2 % FS_AR2 Area II Check FS_AR2 Area II Check
Soot %Wt FS_SOOT Soot SOOT Soot LEM/TGA
Fluid Fluid
Disp ml FPQ_D FPQ_D
Dispensed in ml Dispense
SampOrderSeq n/a SampOrderSeq Sequence # LNF_SEQ 2
SampNum n/a SampNum Sample # SAMPLEID
Software LNF SW 
n/a SoftwareVer LNF_VER 2
Ver Version #
LNFNum n/a LNFNum LNF S/N LNF_SN 2
Operator Making
OperatorID n/a OperatorID LNF_OPID 2
Meas
Notes n/a Notes Notes Comments
Concentration Concentration Fraction
n/a LNF_FDIL Fraction Dilut
Mult Mult Dilution
#/ml>5 use
MDiamCum5 #/ml MDiamCum5 MDIAMCUM5 5 MxD>5
MxDia
#ml>10 use
MDiamCum10 #/ml MDiamCum10 MDIAMCUM10 5 MxD>10
MxDia

149
Property UOM Keyword Description Generic Tag C OV Params
#ml>15 use
MDiamCum15 #/ml MDiamCum15 MDIAMCUM15 5 MxD>15
MxDia
#ml>20 use
MDiamCum20 #/ml MDiamCum20 MDIAMCUM20 5 MxD>20
MxDia
#ml>25 use
MDiamCum25 #ml MDiamCum25 MDIAMCUM25 5 MxD>25
MXDia
#ml>50 use
MDiamCum50 #ml MDiamCum50 MDIAMCUM50 5 MxD>50
MxDia
#ml>100
MDiamCum100 #ml MDiamCum100 MDIAMCUM100 5 MxD>100
use MxDia
Max diam in
MDiamMaxSize microns MDiamMaxSize microns use MDIAMMAXSIZE Mx D Sz
MxDia
Mean diam in
MDiamMeanSize microns MDiamMeanSize microns use MDIAMMEANSIZE Me D Sz
MxDia
StDev all diams in
MDiamStdDev microns MDiamStdDev microns use MDIAMSTDDEV StD D Sz
MxDia
ISO44406>4
ISO4 n/a ISO4 PC_R2 3 ISO>2
microns
ISO44406>6
ISO6 n/a ISO6 PC_R5 3 ISO>5
microns
ISO44406>6
ISO14 n/a ISO14 PC_R15 3 ISO>15
microns
NAVAIR n/a NAVAIR NAS1638 code PC_NAS NAS 1638
CHA(RN) n/a CHA(RN) CHA(RN) code CHARN CHA(RN)
HAL n/a HAL HAL code HAL HalCode
#/ml>5 use equiv
CDiamCum5 #/ml CDiamCum5 CNT>5 Cnts>5
cir
#/ml>15 use
CDiamCum15 #/ml CDiamCum15 CNT>15 Cnts>15
equiv cir
#ml>20 use
CDiamCum20 #/ml CDiamCum20 CNT>20 Cnts>20
equiv cir
#ml>21 use
CDiamCum21 #ml CDiamCum21 ? 4 ?
equiv cir
#ml>25 use
CDiamCum25 #ml CDiamCum25 CNT>25 Cnts>25
equiv cir
#ml>50 use
CDiamCum50 #ml CDiamCum50 CNT>50 Cnts>50
equiv ci
#ml>100 use
CDiamCum100 #ml CDiamCum100 CNT>100 Cnts>100
equiv cir
Free water in
WaterIndex ppmv WaterIndex WATERIDX Water Index
ppmv

150 Doc. p/n 100-00087 Rev. A


Property UOM Keyword Description Generic Tag C OV Params
#ml>20 use
CutCum20 #ml CutCum20 CUTCUM20 5 MxD Ct>20
MxDia Ct Particles
#ml>25 use
CutCum25 #ml CutCum25 CUTCUM25 5 MxD Ct>25
MxDia Ct Particles
#ml>50 use
CutCum50 #ml CutCum50 CUTCUM50 5 MxD Ct>50
MxDia Ct Particles
#ml>100 use
CutCum100 #ml CutCum100 CUTCUM100 5 MxD Ct>100
MxDia Ct Particles
MxDia Ct particle
CutMaxSize microns CutMaxSize CUTMAXSIZE Ct Mx Sz
in microns
#ml>20 use
SlideCum20 #ml SlideCum20 MxDia slide wear SLIDECUM20 5 MxD Sl>20
particles
#ml>25 use
SlideCum25 #ml SlideCum25 MxDia slide wear SLIDECUM 5 MxD Sl>25
particles
#ml>50 use
SlideCum50 #ml SlideCum50 MxDia slide wear SLIDECUM 5 MxD Sl>50
particles
#ml>100 use
SlideCum100 #ml SlideCum100 MxDia slide wear SLIDECUM 5 MxD Sl>100
particles
MxDia slide wear
SlideMaxSize microns SlideMaxSize particles in SLIDEMAXSIZE Sl Mx Sz
microns
#/ml > 20 use
FatCum20 #ml FatCum20 MxDia fatigue FATCUM20 5 MxD Fg >20
wear particles
#/ml > 25 use
FatCum25 #ml FatCum25 MxDia fatigue FATCUM25 5 MxD Fg >25
wear particles
#/ml > 50 use
FatCum50 #ml FatCum50 MxDia fatigue FATCUM50 5 MxD Fg >50
wear particles
#/ml > 100 use
FatCum100 #ml FatCum100 MxDia fatigue FATCUM100 5 MxD Fg >100
wear particles
MxDia fatigue
FatMaxSize microns FatMaxSize wear particles in FATMAXSIZE Ft Mx Sz
microns
MxDia fatigue
FatCntMean microns FatCntMean wear particle in FATCNTMEAN Ft Me Sz
microns
Mean dia fatigue
wear particle in
FatCntStdDev microns FatCntStdDev FATCNTSTDDEV Ft StD
microns use
MxDia

151
Property UOM Keyword Description Generic Tag C OV Params
#/ml > 20 use
OxideCum20 #ml OxideCum20 MxDia oxide OXIDECUM20 5 MxD Ox >20
particles
#/ml > 25 use
OxideCum25 #ml OxideCum25 MxDia oxide OXIDECUM25 5 MxD Ox >25
particles
#/ml > 50 use
OxideCum50 #ml OxideCum50 MxDia oxide OXIDECUM50 5 MxD Ox >50
particles
#/ml > 100 use
OxideCum100 #ml OxideCum100 MxDia oxide OXIDECUM100 5 MxD Ox >100
particles
MxDia oxide
OxideMaxSize microns OxideMaxSize particle in microns OXIDEMAXSIZE Ox Mx Sz
use MxDia
#/ml > 20 use
OtherCum20 #ml OtherCum20 MxDia unclassed OTHERCUM20 5 MxD Ot >20
particles
#/ml > 25 use
OtherCum25 #ml OtherCum25 MxDia unclassed OTHERCUM25 5 MxD Ot >25
particles
#/ml > 50 use
OtherCum50 #ml OtherCum50 MxDia unclassed OTHERCUM50 5 MxD Ot >50
particles
#/ml > 100 use
OtherCum100 #ml OtherCum100 MxDia unclassed OTHERCUM100 5 MxD Ot >100
particles
MxDia unclassed
particles in
OtherMaxSize microns OtherMaxSize OTHERMAXSIZE Oth Mx Sz
microns use
MxDia
mean dia
unclassed
OtherCntMean microns OtherCntMean particles in OTHERCNTMEAN Oth Me Sz
microns use
MxDia
StDev dia
unclassed
OtherCntStdDev microns OtherCntStdDev particles in OTHERCNTSTDDEV Oth StD sz
microns use
MxDia
4 micron SAE
SAE4 n/a SAE4 SAE4 SAE 4
AS4059 code
6 micron SAE
SAE6 n/a SAE6 SAE6 SAE 6
AS4059 code

152 Doc. p/n 100-00087 Rev. A


Property UOM Keyword Description Generic Tag C OV Params
14 micron SAE
SAE14 na/ SAE14 SAE14 SAE 14
AS4059 code
21 micron SAE
SAE21 na/ SAE21 SAE21 SAE 21
AS4059 code
38 micron SAE
SAE38 na/ SAE38 SAE38 SAE 38
AS4059 code
70 micron SAE
SAE70 na/ SAE70 SAE70 SAE 70
AS4059 code
Max SAE
SAEmax na/ SAEmax AS4059 code for SAEMAX SAE Mx
ranges tested
PctSoot %wt PctSoot %soot in oil PCTSOOT Pct Soot
Kinematic Vis-
Kinematic Vis- Kinematic Vis-
cSt cosity(cSt) at 40 VISC40 Visc 40C
cocity (cSt) cocity (cSt)
C derived
Dynamic Vis-
Dynamic Vis- Dynamic Vis-
cP cosity(cP) at DVISC40 Dyn Visc 40C
cosity(cP) cosity(cP)
40C measured
BSW n/a BSW BSW BSW Btm Sed &Wat
CFD n/a CFD CFD CFD Cont Fuel Dep
Fibers #/ml Fibers #fibers/ml of fluid FIBERS Fibers
GOST n/a GOST GOST code GOST Gost Code
Total Ferrous ppm TFerrous Total Ferrous ppm TFERROUS Total Fe
Total Non Total Non-Ferrous
ppm TNon-Ferrous TNONFERROUS Total Non-Fe
Ferrous ppm
Ferrous #/ml > 5
FMDiamCum5 #/ml FMDiamCum5 FMDIAMCUM5 5 Fe MxD > 5
use MxDia
Ferrous #/ml >
FMDiamCum10 #/ml FMDiamCum10 FMDIAMCUM10 5 Fe MxD > 10
10 use MxDia
Ferrous #/ml >
FMDiamCum15 #/ml FMDiamCum15 FMDIAMCUM15 5 Fe MxD > 15
15 use MxDia
Ferrous #/ml >
FMDiamCum20 #/ml FMDiamCum20 FMDIAMCUM20 5 Fe MxD > 20
20 use MxDia
Ferrous #/ml >
FMDiamCum25 #/ml FMDiamCum25 FMDIAMCUM25 5 Fe MxD > 25
25 use MxDia
Ferrous #/ml >
FMDiamCum25 #/ml FMDiamCum25 FMDIAMCUM25 5 Fe MxD > 50
25 use MxDia
Ferrous #/ml >
FMDiamCum25 #/ml FMDiamCum25 FMDIAMCUM25 5 Fe MxD > 100
25 use MxDia
Ferrous Max diam
FMDiamMaxSize microns FMDiamMaxSize in microns use FMDIAMMAXSIZE Fe Mx Sz
MxDia
Ferrous Mean
FMDiamMeanSize microns FMDiamMeanSize diam in microns FMDIAMMEANSIZE Fe Me Sz
use MxDia

153
Property UOM Keyword Description Generic Tag C OV Params
Ferrous StDev all
FMDiamStdDev microns FMDiamStdDev diams in microns FMDIAMSTDDEV Fe StD Sz
use MxDia
Ferrous
FISO4 n/a FISO4 ISO44406 > 4 FISO4 Fe ISO >4
microns
Ferrous
FISO6 n/a FISO6 ISO44406 > 6 FISO6 Fe ISO >6
microns
Ferrous
FISO14 n/a FISO14 ISO44406 > 14 FISO14 Fe ISO >14
microns
Ferrous
FNAVAIR n/a FNAVAIR FNAVAIR Fe NAS 1638
NAS1638 code
Ferrous CHA(RN)
FCHA(RN) n/a FCHA(RN) FCHARN Fe CHA(RN)
code
FHAL n/a FHAL Ferrous HAL Code FHAL Fe HAL Code
Ferrous #/ml > 5
FCDiamCum5 #/ml FCDiamCum5 FCDIAMCUM5 Fe >5
use eqiv cir
Ferrous #/ml >
FCDiamCum15 #/ml FCDiamCum15 FCDIAMCUM15 Fe >15
15 use eqiv cir
Ferrous #/ml >
FCDiamCum20 #/ml FCDiamCum20 FCDIAMCUM20 Fe >20
20 use eqiv cir
Ferrous #/ml >
FCDiamCum21 #/ml FCDiamCum21 FCDIAMCUM21 4 Fe >21
21 use eqiv cir
Ferrous #/ml >
FCDiamCum25 #/ml FCDiamCum25 FCDIAMCUM25 Fe >25
25 use eqiv cir
Ferrous #/ml >
FCDiamCum50 #/ml FCDiamCum50 FCDIAMCUM50 Fe >50
50 use eqiv cir
Ferrous #/ml >
FCDiamCum100 #/ml FCDiamCum100 FCDIAMCUM100 Fe >100
100 use eqiv cir
Ferrous #/ml >
Fe MxD Ct >
FCutCum20 #/ml FCutCum20 20 use MxDia Ct FCUTCUM20 5
20
particles
Ferrous #/ml >
Fe MxD Ct >
FCutCum25 #/ml FCutCum25 25 use MxDia Ct FCUTCUM5 5
25
particles
Ferrous #/ml >
Fe MxD Ct >
FCutCum50 #/ml FCutCum50 50 use MxDia Ct FCUTCUM50 5
50
particles
Ferrous #/ml >
Fe MxD Ct >
FCutCum100 #/ml FCutCum100 100 use MxDia Ct FCUTCUM100 5
100
particles
Ferrous MxDia Ct
FCutMaxSize microns FCutMaxSize FCUTMAXSIZE Fe Ct Mx Sz
particle in microns

154 Doc. p/n 100-00087 Rev. A


Property UOM Keyword Description Generic Tag C OV Params
Ferrous #/ml >
20 use MxDia Fe MxD Sl >
FSlideCum20 #/ml FSlideCum20 FSLIDECUM20 5
slide wear 20
particles
Ferrous #/ml >
25 use MxDia Fe MxD Sl >
FSlideCum25 #/ml FSlideCum25 FSLIDECUM25 5
slide wear 25
particles
Ferrous #/ml >
50 use MxDia Fe MxD Sl >
FSlideCum50 #/ml FSlideCum50 FSLIDECUM50 5
slide wear 50
particles
Ferrous #/ml >
100 use MxDia Fe MxD Sl >
FSlideCum100 #/ml FSlideCum100 FSLIDECUM100 5
slide wear 100
particles
Ferrous MxDia
FSlideMaxSize microns FSlideMaxSize slide wear particle FSLIDEMAXSIZE Fe Sl Mx Sz
in microns
Ferrous #/ml >
20 use MxDia Fe MxD Fg >
FFatCum20 #/ml FFatCum20 FFATCUM20 5
fatigue wear 20
particles
Ferrous #/ml >
25 use MxDia Fe MxD Fg >
FFatCum25 #/ml FFatCum25 FFATCUM25 5
fatigue wear 25
particles
Ferrous #/ml >
50 use MxDia Fe MxD Fg >
FFatCum50 #/ml FFatCum50 FFATCUM50 5
fatigue wear 50
particles
Ferrous #/ml >
100 use MxDia Fe MxD Fg >
FFatCum100 #/ml FFatCum100 FFATCUM100 5
fatigue wear 100
particles
Ferrous MxDia
FFatMaxSize microns FFatMaxSize fatigue wear FFATMAXSIZE Fe Fg Mx Sz
particle in microns
Ferrous MxDia
FFatCntMean microns FFatCntMean fatigue wear FFATCNTMEAN Fe Fg Me Sz
particle in microns
Ferrous mean dia
fatigue wear
FFatCntStdDev microns FFatCntStdDev FFATCNTSTDDEV Fe Fg StD Sz
particle in microns
use MxDia
Ferrous #/ml >
Fe MxD Ox
FOxideCum20 #/ml FOxideCum20 20 use MxDia FOXIDECUM20 5
>20
oxide particles

155
Property UOM Keyword Description Generic Tag C OV Params
Ferrous #/ml >
Fe MxD Ox
FOxideCum25 #/ml FOxideCum25 25 use MxDia FOXIDECUM25 5
>25
oxide particles
Ferrous #/ml >
Fe MxD Ox
FOxideCum50 #/ml FOxideCum50 50 use MxDia FOXIDECUM50 5
>50
oxide particles
Ferrous #/ml >
Fe MxD Ox
FOxideCum100 #/ml FOxideCum100 100 use MxDia FOXIDECUM100 5
>100
oxide particles
Ferrous MxDia
oxide particle in
FOxideMaxSize microns FOxideMaxSize FOXIDEMAXSIZE Fe Ox Mx Sz
microns use
MxDia
Ferrous#/ml > 20
use MxDia Fe MxD Ot >
FOtherCum20 #/ml FOtherCum20 FOTHERCUM20 5
unclassed 20
particles
Ferrous#/ml > 25
use MxDia Fe MxD Ot >
FOtherCum25 #/ml FOtherCum25 FOTHERCUM25 5
unclassed 25
particles
Ferrous#/ml > 20
use MxDia Fe MxD Ot >
FOtherCum50 #/ml FOtherCum50 FOTHERCUM50 5
unclassed 50
particles
Ferrous#/ml >
100 use MxDia Fe MxD Ot >
FOtherCum100 #/ml FOtherCum100 FOTHERCUM100 5
unclassed 100
particles
Ferrous MxDia
unclassed
FOtherMaxSize microns FOtherMaxSize particles in FOTHERMAXSIZE Fe Ot Mx Sz
microns use
MxDia
Ferrous mean dia
unclassed
FOtherCntMean microns FOtherCntMean particles in FOTHERCNTMEAN Fe Ot Me Sz
microns use
MxDia
Ferrous StDev dia
unclassed
FOtherCntStdDev microns FOtherCntStdDev particles in FOTHERCNTSTDDEV Fe Ot StD Sz
microns use
MxDia

156 Doc. p/n 100-00087 Rev. A


Property UOM Keyword Description Generic Tag C OV Params
Ferrous 4 micron
FSAE4 n/a FSAE4 SAE AS4059 FSAE4 Fe SAE 4
code
Ferrous 6 micron
FSAE6 n/a FSAE6 SAE AS4059 FSAE6 Fe SAE 6
code
Ferrous 14 micron
FSAE14 n/a FSAE14 SAE AS4059 FSAE14 Fe SAE 14
code
Ferrous 21 micron
FSAE21 n/a FSAE21 SAE AS4059 FSAE21 Fe SAE 21
code
Ferrous 38 micron
FSAE38 n/a FSAE38 SAE AS4059 FSAE38 Fe SAE 38
code
Ferrous 70 micron
FSAE70 n/a FSAE70 SAE AS4059 FSAE70 Fe SAE 70
code
Ferrous Max SAE
FSAEmax n/a FSAEmax AS4059 code for FSAEMAX Fe SAE Mx
ranges tested
Ferrous #fibers/ml
FFibers #/ml FFibers FFIBERS Fe Fibers
fluid
Ferrous GOST
FGOST n/a FGOST FGOST Fe GOST Code
code
Non-ferrous #/ml Nfe MxD Cnts
NMDiamCum5 #/ml NMDiamCum5 NMDIAMCUM5 5
> 5 use MxDia >5
Non-ferrous #/ml Nfe MxD Cnts
NMDiamCum10 #/ml NMDiamCum10 NMDIAMCUM10 5
> 10 use MxDia >10
Non-ferrous #/ml Nfe MxD Cnts
NMDiamCum15 #/ml NMDiamCum15 NMDIAMCUM15 5
> 15 use MxDia >15
Non-ferrous #/ml Nfe MxD Cnts
NMDiamCum20 #/ml NMDiamCum20 NMDIAMCUM20 5
> 20 use MxDia >20
Non-ferrous #/ml Nfe MxD Cnts
NMDiamCum25 #/ml NMDiamCum25 NMDIAMCUM25 5
> 25 use MxDia >25
Non-ferrous #/ml Nfe MxD Cnts
NMDiamCum50 #/ml NMDiamCum50 NMDIAMCUM50 5
> 50 use MxDia >50
Non-ferrous #/ml
Nfe MxD Cnts
NMDiamCum100 #/ml NMDiamCum100 > 100 use MxDia NMDIAMCUM100 5
>100

Non-ferrous Max
NMDiamMaxSize microns NMDiamMaxSize diam in microns NMDIAMMAXSIZE Nfe Mx D
use MxDia
Non-ferrous Mean
NMDiamMeanSize NMDiamMeanSize
microns diam in microns NMDIAMMEANSIZE Nfe Me D
use MxDia

157
Property UOM Keyword Description Generic Tag C OV Params
Non-ferrous
StDev all diams in
NMDiamStdDev microns NMDiamStdDev NMDIAMSTDDEV Nfe Std D
microns use
MxDia
Non-ferrous
3
NISO4 n/a NISO4 ISO44406 > 4 NISO4 Nfe ISO >4
microns
Non-ferrous
3
NISO6 n/a NISO6 ISO44406 > 6 NISO6 Nfe ISO >6
microns
Non-ferrous
3
NISO14 n/a NISO14 ISO44406 > 14 NISO14 Nfe ISO >14
microns
Non-ferrous
NNAVAIR n/a NNAVAIR NNAVAIR Nfe NAS 1638
NAS1638 code
Non-ferrous CHA
NCHA(RN) n/a NCHA(RN) NCHARN Nfe CHA(RN)
(RN) code
Non-ferrous HAL
NHAL n/a NHAL NHAL Nfe HAL Code
code
Non-ferrous #/ml
NCDiamCum5 #/ml NCDiamCum5 NCDIAMCUM5 Nfe Cnts >5
> 5 use eqiv cir
Non-ferrous #/ml
NCDiamCum15 #/ml NCDiamCum15 NCDIAMCUM15 Nfe Cnts >15
> 15 use eqiv cir
Non-ferrous #/ml
NCDiamCum20 #/ml NCDiamCum20 NCDIAMCUM20 Nfe Cnts >20
> 20 use eqiv cir
Non-ferrous #/ml
NCDiamCum21 #/ml NCDiamCum21 NCDIAMCUM21 4 Nfe Cnts >21
> 21 use eqiv cir
Non-ferrous #/ml
NCDiamCum25 #/ml NCDiamCum25 NCDIAMCUM25 Nfe Cnts >25
> 25 use eqiv cir
Non-ferrous #/ml
NCDiamCum50 #/ml NCDiamCum50 NCDIAMCUM50 Nfe Cnts >50
> 50 use eqiv cir
Non-ferrous #/ml
NCDiamCum100 #/ml NCDiamCum100 > 100 use eqiv NCDIAMCUM100 Nfe Cnts >100
cir
Non-ferrous #/ml
Nfe MxD Ct
NCutCum20 #/ml NCutCum20 > 20 use MxDia NCUTCUM20 5
>20
Ct particles
Non-ferrous #/ml
Nfe MxD Ct
NCutCum25 #/ml NCutCum25 > 25 use MxDia NCUTCUM25 5
>25
Ct particles
Non-ferrous #/ml
Nfe MxD Ct
NCutCum50 #/ml NCutCum50 > 20 use MxDia NCUTCUM50 5
>50
Ct particles
Non-ferrous #/ml
Nfe MxD Ct
NCutCum100 #/ml NCutCum100 > 100 use MxDia NCUTCUM100 5
>100
Ct particles

158 Doc. p/n 100-00087 Rev. A


Property UOM Keyword Description Generic Tag C OV Params
Non-ferrous
NCutMaxSize microns NCutMaxSize MxDia Ct particle NCUTMAXSIZE Nfe Ct Mx Sz
in microns
Non-ferrous #/ml
> 20 use MxDia Nfe MxD Sl >
NSlideCum20 #/ml NSlideCum20 NSLIDECUM20 5
slide wear 20
particles
Non-ferrous #/ml
> 25use MxDia Nfe MxD Sl >
NSlideCum25 #/ml NSlideCum25 NSLIDECUM25 5
slide wear 25
particles
Non-ferrous #/ml
> 50 use MxDia Nfe MxD Sl >
NSlideCum50 #/ml NSlideCum50 NSLIDECUM50 5
slide wear 50
particles
Non-ferrous #/ml
> 100 use MxDia Nfe MxD Sl >
NSlideCum100 #/ml NSlideCum100 NSLIDECUM100 5
slide wear 100
particles
Non-ferrous
NSlideMaxSize microns NSlideMaxSize MxDia slide wear NSLIDEMAXSIZE Nfe Sl Mx Sz
particle in microns
Non-ferrous #/ml
> 20 use MxDia Nfe MxD Fg >
NFatCum20 #/ml NFatCum20 NFATCUM20 5
fatigue wear 20
particles
Non-ferrous #/ml
> 25 use MxDia Nfe MxD Fg >
NFatCum25 #/ml NFatCum25 NFATCUM25 5
fatigue wear 25
particles
Non-ferrous #/ml
> 50 use MxDia Nfe MxD Fg >
NFatCum50 #/ml NFatCum50 NFATCUM50 5
fatigue wear 50
particles
Non-ferrous #/ml
> 100 use MxDia Nfe MxD Fg >
NFatCum100 #/ml NFatCum100 NFATCUM100 5
fatigue wear 100
particles
Non-ferrous
MxDia fatigue
NFatMaxSize microns NFatMaxSize NFATMAXSIZE Nfe Fg Mx Sz
wear particle in
microns
Non-ferrous
MxDia fatigue
NFatCntMean microns NFatCntMean NFATCNTMEAN Nfe Fg Me Sz
wear particle in
microns

159
Property UOM Keyword Description Generic Tag C OV Params
Non-ferrous mean
dia fatigue wear
NFatCntStdDev microns NFatCntStdDev NFATCNTSTDDEV Nfe Fg StD Sz
particle in microns
use MxDia
Non-ferrous #/ml
Nfe MxD Ox >
NOxideCum20 #/ml NOxideCum20 > 20 use MxDia NOXIDECUM20 5
20
oxide particles
Non-ferrous #/ml
Nfe MxD Ox >
NOxideCum25 #/ml NOxideCum25 > 25use MxDia NOXIDECUM25 5
25
oxide particles
Non-ferrous #/ml
Nfe MxD Ox >
NOxideCum50 #/ml NOxideCum50 > 50 use MxDia NOXIDECUM50 5
50
oxide particles
Non-ferrous #/ml
Nfe MxD Ox >
NOxideCum100 #/ml NOxideCum100 > 100 use MxDia NOXIDECUM100 5
100
oxide particles
Non-ferrous
MxDia oxide
NOxideMaxSize microns NOxideMaxSize NOXIDEMAXSIZE Nfe Ox Mx Sz
particle in microns
use MxDia
Non-ferrous #/ml
> 20 use MxDia Nfe MxD Ot >
NOtherCum20 #/ml NOtherCum20 NOTHERCUM20 5
unclassed 20
particles
Non-ferrous #/ml
> 25 use MxDia Nfe MxD Ot >
NOtherCum25 #/ml NOtherCum25 NOTHERCUM25 5
unclassed 25
particles
Non-ferrous #/ml
> 50 use MxDia Nfe MxD Ot >
NOtherCum50 #/ml NOtherCum50 NOTHERCUM50 5
unclassed 50
particles
Non-ferrous #/ml
> 100 use MxDia Nfe MxD Ot >
NOtherCum100 #/ml NOtherCum100 NOTHERCUM100 5
unclassed 100
particles
Non-ferrous
MxDia unclassed
NOtherMaxSize microns NOtherMaxSize particles in NOTHERMAXSIZE Nfe Ot Mx Sz
microns use
MxDia
Non-ferrous mean
dia unclassed
NOtherCntMean microns NOtherCntMean particles in NOTHERCNTMEAN Nfe Ot Me Sz
microns use
MxDia

160 Doc. p/n 100-00087 Rev. A


Property UOM Keyword Description Generic Tag C OV Params
Non-ferrous
StDev dia
NOtherCntStdDev unclassed
microns NOtherCntStdDev NOTHERCNTSTDDEV Nfe Ot StD Sz
particles in
microns use
MxDia
Non-ferrous 4
NSAE4 n/a NSAE4 micron SAE NSAE4 Nfe SAE 4
AS4059 code
Non-ferrous 6
NSAE6 n/a NSAE6 micron SAE NSAE6 Nfe SAE 6
AS4059 code
Non-ferrous 14
NSAE14 n/a NSAE14 micron SAE NSAE14 Nfe SAE 14
AS4059 code
Non-ferrous 21
NSAE21 n/a NSAE21 micron SAE NSAE21 Nfe SAE 4
AS4059 code
Non-ferrous 38
NSAE38 n/a NSAE38 micron SAE NSAE38 Nfe SAE 38
AS4059 code
Non-ferrous 70
NSAE70 n/a NSAE70 micron SAE NSAE70 Nfe SAE 70
AS4059 code
Non-ferrous Max
SAE AS4059
NSAEmax n/a NSAEmax NSAEMAX Nfe SAE Mx
code for ranges
tested
Non-ferrous #fiber-
NFibers #/ml NFibers NFIBERS Nfe Fibers
s/ml fluid
Non-ferrous
NGOST n/a NGOST NGOST Nfe GOST Code
GOST code

161
14 Default Spectro AP Set Parameters

A maximum of 96 parameters can be included in an AP Set. The following list shows the 81 para-
meters in the default Spectro AP Set.
Note: this is not all available parameters, only pages where a Spectro AP Set parameter is selected.
Read top to bottom, left to right:

Continued...

162
Continued..

163 Doc. p/n 100-00087 Rev. A


164
15 Reference Fluid Database

The following is a list of current Reference fluids and their Substance ID with the corresponding Fluid
Category.

ID # Substance Name Category


201 Generic MIL-PRF-6083 HYDRAULIC-AERO
202 BP Turbo Oil 2197 (MIL-PRF-23699F HTS) TURBINE-AERO
203 Mobil Jet Oil II (MIL-PRF-23699) TURBINE-AERO
204 AeroShell Turbine Oil 500 (MIL-L-23699) TURBINE-AERO
205 Anderol Royco Turbine Oil 500 (MIL-PRF-23699F HTS) TURBINE-AERO
206 BP Turbo Oil 2380 (MIL-PRF-23699F STD) TURBINE-AERO
207 Generic MIL-PRF-23699 TURBINE-AERO
208 Mobil Jet Oil 254 (MIL-PRF-23699 HTS) TURBINE-AERO
209 NYCO Turbonycoil 400 (MIL-PRF-7808L Gr. 4) TURBINE-AERO
210 HATCO Corp MIL-PRF-7808L Gr. 3 TURBINE-AERO
211 Mobil Turbo 284 (MIL-PRF-7808 Gr. 4) TURBINE-AERO
212 AeroShell Turbine Oil 560 (MIL-L-23699D) TURBINE-AERO
213 Anderol Royco 808 (MIL-PRF-7808L Gr. 3) TURBINE-AERO
214 BP Turbo Oil 2389 (MIL-PRF-7808L Gr. 3) TURBINE-AERO
215 Amsoil Synthetic Anti-Wear Hydraulic Oil ISO 32 (AWH) HYDRAULIC
216 Mobilgear SHC 220 GEAR-SPLASH
217 Mobilgear SHC 680 GEAR-SPLASH

219 Skydrol 500 B4 HYDRAULIC-FIRE RESISTANT


220 Skydrol 5 HYDRAULIC-FIRE RESISTANT
221 Chevron Ultra Gear 220 GEAR-SPLASH
222 Mobil Delvac SAE 50 GEAR-PRESSURE
223 Generic Synthetic Gear/Optigear GEAR-SPLASH
224 Mobil ATF TRANSMISSION
225 Amalie Chrysler ATF+3 Type7176E TRANSMISSION
226 Federated Dexron 3 ATF TRANSMISSION
227 Mobil Multi-Vehicle ATF TRANSMISSION
228 Caltex Dexron III TRANSMISSION
229 Generic Transmission 2 TRANSMISSION
230 Mobil Multipurpose ATF TRANSMISSION
231 Shell Donax TC 30 GEAR-PRESSURE
232 Shell Donax TC 50 GEAR-PRESSURE
233 Caltex Torque Fluid 454 GEAR-PRESSURE
234 Shell Tellus T32 HYDRAULIC
235 Shell Tellus T68 HYDRAULIC
236 Shell Valvata J680 HYDRAULIC

165
237 Mobil 600w Super Cylinder Oil GEAR-SPLASH
238 Amalie Mercon V ATF Synthetic Blend GEAR-PRESSURE
239 Generic Industrial GEAR-PRESSURE
240 Texaco Rando HDZ 68 HYDRAULIC
241 Shell Tellus Oil S 68 HYDRAULIC
242 Shell Tellus Oil 46 HYDRAULIC
243 Mobil Delvac 80W-140 GEAR-SPLASH
244 Esso Gear Oil GX 85W-140 GEAR-PRESSURE
245 Jomo Dreamer ATF Special GEAR-PRESSURE
246 Shell Omala Oil 220 GEAR-SPLASH
247 Generic Petroleum Gear GEAR-SPLASH
248 Essolube Hydraulic 30W HYDRAULIC
249 Essolube Hydraulic 10W HYDRAULIC
250 Esso Gear Oil GX 80W-90 GEAR-SPLASH
251 Jomo Torcon AW18 HYDRAULIC
252 Shell Helix 40 ENGINE
253 Shell Rimula D 30 ENGINE
254 Exxon XD-3 ENGINE
255 Castrol Assuron 10W ENGINE
256 Castrol GTX High Mileage 10W-30 ENGINE
257 Mobil 1 0W-20 ENGINE
258 Mobil 1 Turbo Diesel Truck 5W-40 ENGINE
259 Mobil 1 High Mileage 10W-30 ENGINE
260 Pennzoil SAE 5W-30 Advanced Protection ENGINE
261 PittPenn Motor Oil SAE 40 ENGINE
262 Caltex 400 Multigrade 15W-40 ENGINE
263 Caltex 6000 SAE 40 ENGINE
264 Caltex Delo Gold 15W-40 ENGINE
265 Caltex Havoline Formula SAE 20W-50 ENGINE
266 Caltex Supreme Five Star SAE 40 ENGINE
267 Chevron Veritas 800 Marine 30 ENGINE
268 Kixx PAO RV 5W-40 ENGINE
269 Generic Reciprocating Engine ENGINE
270 Mobil Super 2000 5W-40 ENGINE
271 My Total Rubia ENGINE
272 Esso Racing 20W-50 ENGINE
273 Apolloil EX DH-1 10W-40 ENGINE
274 Jomo Delstar FX 15W-40 ENGINE
275 Shell Clavus Oil 68 CHILLER
276 Shell Tellus Oil 68 HYDRAULIC
277 Valvoline SYNPower 5W-30 ENGINE
278 Shell Vitrea Oil 68 TURBINE-STEAM
279 Exxon NutoH 68 HYDRAULIC

166 Doc. p/n 100-00087 Rev. A


280 Exxon Teresstic 320 GEAR-SPLASH
281 Mobil Spartan EP 68 GEAR-PRESSURE
282 Caltex RPM Compresor Oil 68 COMPRESSOR
283 Caltex GS Turbine Oil 68 TURBINE-STEAM
284 Caltex Regal Oil R&O 46 HYDRAULIC
285 Caltex RPM Compresor Oil 100 COMPRESSOR
286 Mobil Delvac 1300 Super 15W-40 ENGINE
287 SK TURBINE 32 TURBINE-STEAM
288 Generic Multipurpose Type 2 GEAR-SPLASH
289 Shell Turbo Oil T32 TURBINE-STEAM
290 Caltex Regal Oil R&O 32 HYDRAULIC
291 ZIC Supervis AW 46 HYDRAULIC
292 Shell Turbo Oil 32 TURBINE-STEAM
293 Sunoco Suniso 4GS CHILLER
294 Mobil DTE Oil Heavy Medium GEAR-SPLASH
295 BP Energol TH-HT 32 TURBINE-STEAM
296 Mobil DTE Heavy GEAR-SPLASH
297 Shell Alexia 50 ENGINE-HEAVY DUTY
298 Shell Argina T30 ENGINE-HEAVY DUTY
299 Shell Caprinus HPD 40 ENGINE-HEAVY DUTY
300 Shell Gadinia 40 ENGINE-HEAVY DUTY
301 Shell Gadinia AL 40 ENGINE-HEAVY DUTY
302 Shell Melina S 30 ENGINE-HEAVY DUTY
303 Generic Heavy Duty Diesel ENGINE-HEAVY DUTY
304 Caltex Delo 6200 SAE 40 ENGINE-HEAVY DUTY
305 Chevron Taro 30 DP SAE 40 ENGINE-HEAVY DUTY
306 Chevron Taro 40 XL SAE 40 ENGINE-HEAVY DUTY
307 Chevron Taro Special 70 ENGINE-HEAVY DUTY
308 Caltex Valor 320 Sae 40 ENGINE-HEAVY DUTY
309 Idemitsu Kosan Daphne Out Board 30 ENGINE-HEAVY DUTY

312 Generic Phosphate Ester HYDRAULIC-FIRE RESISTANT


313 Anderol Royco 782 (MIL-PRF-83282D) HYDRAULIC-AERO
314 Generic Biodiesel B100 FAME
315 Biodiesel B100 B FAME
316 Chevron Texaco Meropa 100 GEAR-SPLASH
317 Texaco Rando HD 100 GEAR-SPLASH
318 ZIC Supervis AW 150 GEAR-SPLASH
319 ZIC Super Gear EP 220 GEAR-SPLASH
320 Mobil DTE Extra Heavy GEAR-SPLASH
321 Shell Corena Oil P 100 COMPRESSOR
322 ZIC Super Gear EP 320 GEAR-SPLASH
323 Mobilgear 600 XP 460 GEAR-SPLASH
324 Chevron Texaco Meropa 320 GEAR-SPLASH

167
327 Generic Petroleum ASTM Petroleum Crankcase
328 Generic Gear/Hydro ASTM Polyol Ester
329 Generic Ester ASTM EP Gear/Hydro
330 Anderol Royco 555 (MIL-PRF-85734) TURBINE-AERO
331 Castrol Aero 40 (MIL-PRF-5606) TURBINE-AERO
332 Generic MIL-PRF-5606 TURBINE-AERO
333 AeroShell Fluid 4 (MIL-PRF-5606) TURBINE-AERO
334 PittPenn SAE 30 (MIL-PRF-2104G) ENGINE
335 PittPenn Gear Oil 80W90 GEAR-SPLASH
336 HATCO Corp MIL-PRF-23699 C/I TURBINE-AERO
337 Caltex Delo 1000 Marine SAE 30 ENGINE-HEAVY DUTY
338 Caltex Delo Silver SAE 10 ENGINE
339 Caltex Delo Silver SAE 30 ENGINE
340 Caltex Delo Silver SAE 40 ENGINE
341 Caltex Delo Silver SAE 50 ENGINE
342 Caltex Havoline Formula SAE 10W-30 (SL) ENGINE
343 Caltex Regal Marine 77 GEAR-PRESSURE
344 Caltex RPM XD SAE 40 ENGINE

346 Chevron Taro 20 DP SAE 40 ENGINE-HEAVY DUTY


347 Chevron Thuban GL4 SAE 140 GEAR-SPLASH
348 Chevron Thuban GL4 SAE 90 GEAR-SPLASH
349 Chevron Thuban GL5 EP SAE 140 GEAR-SPLASH
350 Chevron Thuban GL5 EP SAE 80W-90 GEAR-SPLASH
351 Chevron Thuban GL5 EP SAE 85W-140 GEAR-SPLASH
352 Chevron Gas Turbine Oil 32 TURBINE-CCGT
353 Total Transmission TM 80W-90 GEAR-PRESSURE
354 Royal Purple SAE 15W-40 ENGINE
355 Chevron Meropa 150 GEAR-SPLASH

358 Mobil DTE 24 HYDRAULIC


359 Royal Purple SAE 5W-30 ENGINE
360 GS Turbine Oil 46 TURBINE-STEAM
361 Royal Purple SAE 0W-40 ENGINE
362 GS Turbine Oil 32 TURBINE-STEAM
363 Chevron Meropa 220 GEAR-SPLASH
364 Royal Purple SAE 5W-40 ENGINE
365 Mobil DTE Oil Light GEAR-SPLASH
366 Shell Turbo Oil T 46 TURBINE-STEAM
367 BP Energol THB 32 TURBINE-STEAM
368 ZIC Supervis AW 32 HYDRAULIC
369 Texaco Rando 46 HYDRAULIC

168 Doc. p/n 100-00087 Rev. A


370 Texaco Rando 32 HYDRAULIC
371 Chevron Regal Oil R&O 100 GEAR-SPLASH
372 Texaco Rando 68 HYDRAULIC
373 Texaco Rando 150 GEAR-SPLASH
374 Mobil DTE 26 HYDRAULIC
375 Shell Turbo Oil T 68 TURBINE-STEAM
376 Texaco RANDO ALP 68 HYDRAULIC
377 Shell Spirax HD 80W-90 GEAR-SPLASH
378 Mobil DTE 25 HYDRAULIC
379 Mobilgear 632 GEAR-SPLASH
380 Mobil SHC 639 GEAR-SPLASH
381 Shell Argina T 40W ENGINE-HEAVY DUTY
382 Shell Argina X 40W ENGINE-HEAVY DUTY
383 Shell Argina XL 40W ENGINE-HEAVY DUTY
384 Shell Dentax 140W GEAR-SPLASH
385 Shell Gadinia 30W ENGINE-HEAVY DUTY
386 Shell Omala 100 GEAR-SPLASH
387 Shell Omala 150 GEAR-SPLASH
388 Shell Omala 320 GEAR-SPLASH
389 Shell Omala RL 460 GEAR-SPLASH
390 Shell Omala 680 GEAR-SPLASH
391 Mobil Spartan EP 150 GEAR-SPLASH
392 Mobil Spartan EP 320 GEAR-SPLASH
393 Mobil Spartan EP 220 GEAR-SPLASH
394 Shell Tonna V 68 SLIDEWAY
395 Amalie Hypoid Gear Lubricant 85W-140 GEAR-SPLASH
396 Lucas 15/40 Magnum ENGINE
397 Pittpenn Premium 15W-40 ENGINE
398 Spectrum Plus 10W-30 High Mileage ENGINE
399 Spectrum Plus 10W-40 High Mileage ENGINE
400 Spectrum Plus 30W ENGINE
401 Valvoline 10W-30 Dura Blend Tough Driving Formula ENGINE
402 Valvoline 10W-30 Dura Blend ENGINE
403 Valvoline 10W-30 Max Life ENGINE
404 Valvoline 10W-40 Preimum ENGINE
405 Valvoline 10W-40 Dura Blend ENGINE
406 Valvoline 20W-50 Max Life ENGINE
407 Valvoline 5W30 ENGINE

409 Mobil 1 0W-30 ENGINE


410 Mobil 1 0W-40 ENGINE
411 Mobil 1 10W-30 ENGINE
412 Mobil 1 10W-30 Extended Performance ENGINE
413 Mobil 1 15W-50 ENGINE
414 Mobil 1 5W-20 ENGINE

169
415 Mobil 1 5W-30 ENGINE
416 Mobil 1 5W-30 Extended Performance ENGINE
417 Mobil 1 Turbo diesel 5W40 ENGINE
418 Mobil Clean 5000 10W-30 ENGINE
419 Mobil Clean 5000 10W-40 ENGINE
420 Mobil High Mileage 10W-30 ENGINE

422 Pennzoil Advanced Protection 10W-30 ENGINE


423 Pennzoil Advanced Protection 5W-20 ENGINE
424 Pennzoil Advanced Protection 5W-30 ENGINE
425 Pennzoil High Mileage 10W-30 ENGINE
426 Pennzoil High Mileage 10W-40 ENGINE
427 Pennzoil High Mileage 5W-30 ENGINE
428 Pennzoil Platinum Synthetic 10W-30 ENGINE
429 Pennzoil Platinum Synthetic 5W-30 ENGINE
430 Penzoil Motor Oil 10W-40 ENGINE
431 Penzoil Motor Oil 20W-50 ENGINE
432 Penzoil Motor Oil HD-30 ENGINE
433 Apolloil Tough Runner DH-1 15W40 ENGINE
Placeholder for 434 through 439
440 Generic Diesel/Biodiesel Mix FAME in DIESEL
441 Generic Biodiesel Feedstock BIODIESEL FEEDSTOCK
442 Mobil SHC PM 320 GEAR-SPLASH
Placeholder for 443 through 605
606 Total Azolla ZS 68 HYDRAULIC
607 Total Azolla ZS 100 GEAR-PRESSURE
608 Total Rubia TIR 7400 15W40 ENGINE
609 Total Carter EP 68 HYDRAULIC
610 Raloy Racing Monograde 40W SL ENGINE
611 Raloy Honda Aceite Genuino 10W-30 SM/GF-4 ENGINE
612 Raloy Max Raloy Diesel 3277 M 10W-40 ENGINE
613 Raloy Super G-1 Multigrade 25W-60 TIPO SL ENGINE
614 Raloy Diferenciales EP 85W-140 GL-5 GEAR-SPLASH
615 Raloy Racing Oil Monograde 40W SL ENGINE
616 Raloy Genuine VW Engine Oil 5W-40 ENGINE
617 Raloy Diesel Supreme VDS-3 15W-40 CI-4/SL ENGINE
618 Raloy New Experience Synthetic Oil 5W-30 SM/GF4 ENGINE
619 Raloy AK Power 25W-50 Tipo SL ENGINE
620 Raloy Diesel Supreme VDS-3 20W-50 CI-4 /SL ENGINE
621 Raloy Diesel Advanced CJ-4/SM 15W-40 ENGINE
622 Raloy Racing Monograde Oil 20W-50 SL ENGINE
623 Raloy Diesel Multigrade 20W-50 CF-4/SG ENGINE
624 Raloy Diesel Premium 25W-50 CF ENGINE

170 Doc. p/n 100-00087 Rev. A


625 Raloy Racing Multigrade Oil 10W-30 SL ENGINE
626 Raloy Super 15W-40 MB 227.1 ENGINE
627 Raloy Racing Multigrade Oil 15W-40 SL ENGINE
628 Raloy Racing Turbo Oil 5W-30 SM/GF-4 ENGINE
629 Raloy Diesel Series 5 Plus 50W ENGINE
630 Raloy Diesel Series 5 Plus 40W ENGINE
631 Raloy Hildraloy 300 ISO-68 HYDRAULIC
632 Apolloil Widegear LW 80W-90 GEAR-PRESSURE
633 Shell Corena AP 100 COMPRESSOR
634 Generic Synthetic Gear/Hydraulic GEAR-SPLASH
635 Shell Sirius X 40W ENGINE-HEAVY DUTY
636 Mobil Vacuoline 525 GEAR-PRESSURE
637 Shell Rimula X 15W40 ENGINE
638 Shell Tellus 22 HYDRAULIC
639 Mobilgear 600XP 150 GEAR-SPLASH
640 Mobil DTE Oil BB GEAR-SPLASH
641 SK 5W20 GF4 ENGINE
642 Shell 5W30 A5 ENGINE
643 S-Oil 5W30 A5 ENGINE

645 Repsol Extra Vida Plus 15W-40 ENGINE


646 Castrol Aircol PD 100 COMPRESSOR
647 Castrol Hyspin AWH-M32 HYDRAULIC
648 Castrol MLC 40 ENGINE-HEAVY DUTY
649 Repsol EVP 15 ENGINE
650 Total Aurelia TI-EX 4040 ENGINE-HEAVY DUTY
651 Total Rubia TIR 6400 15W-30 ENGINE
652 Castrol Alpha SP 150 GEAR-SPLASH
653 Total Aurelia XL 4040 ENGINE-HEAVY DUTY
654 Total Aurelia TI 4030 ENGINE-HEAVY DUTY
655 Guascor Motoroil 4000M ENGINE
656 Repsol Turbo THPD 15W40 ENGINE
657 Repsol Extra Vida ENGINE
658 AGIP SPA Sigma Truck ENGINE
659 Atlas Copco Roto-Inject 46 COMPRESSOR
660 Chevron Gear Comound EP 220 GEAR-SPLASH
661 Chevron Delo 400 LE 15W40 ENGINE
662 Atlas Copco Roto 268 COMPRESSOR
663 Sunoco Ultra GL5 85W140 GEAR-SPLASH
664 Conoco HD Fleet Supreme 15W40 ENGINE

666 Conoco PowerDrive 10W ENGINE


667 Conoco PowerDrive 50/60 ENGINE
668 Chevron Regal R&O 32 GEAR-SPLASH

171
670 Atlas Copco Roto Z COMPRESSOR
671 Mobil SHC 526 GEAR-SPLASH
672 Bel-Ray 100 Gear Oil 90 GEAR-SPLASH
673 Shell Rotella T 15W40 ENGINE
674 Terpel SAE 85W140 API GL-5/MT-1 GEAR-SPLASH
675 Terpel Marino TBN 70 ENGINE-HEAVY DUTY
676 Terpel Marino TBN 30 ENGINE-HEAVY DUTY
677 Terpel SAE 80W90 API GL-5/MT-1 GEAR-SPLASH
678 Terpel ATF Dexron III TRANSMISSION
679 Terpel Turbina O&R ISO 46 TURBINE-STEAM
680 Terpel Turbina O&R ISO 32 TURBINE-STEAM
681 Terpel Hidraulico ISO 68 HYDRAULIC
682 Terpel Maxter Progresa API CJ-4/SM, 15W40 ENGINE
683 Terpel Maxter Progresa API CI-4/SL, 15W40 ENGINE
684 Terpel FD-1 60W GEAR-SPLASH
685 Terpel 15W40 CI-4/SL ENGINE
686 Terpel DT CAT TO-4 10W ENGINE
687 Terpel DT CAT TO-4 30W ENGINE
688 Terpel DT CAT TO-4 50W ENGINE
689 ExxonMobil Mobilgard M50 ENGINE-HFO
690 Shell Spirax MA 80W GEAR-SPLASH
691 Shell Spirax G 140W GEAR-SPLASH
692 Shell Spirax 85W-140 GEAR-SPLASH
693 Shell Spirax G 90W GEAR-SPLASH

695 Shell Donax TG TRANSMISSION

697 Generic Heavy Duty Transmission TRANSMISSION

700 Shell Tellus 46 HYDRAULIC


701 Shell Tellus T15 HYDRAULIC
702 Shell Tellus T46 HYDRAULIC
703 Shell Omala HD 150 GEAR-SPLASH
704 Shell Omala HD 220 GEAR-SPLASH
706 Shell Helix Super 20W50 ENGINE
707 Shell Thermia B HEAT TRANSFER
709 Shell Rimula X 40W ENGINE
710 Shell Rimula Ultra 10W40 ENGINE
711 Shell Rimula X 30W ENGINE
712 Shell Rimula X 10W ENGINE
713 Addinol Diesel Longlife MD 1548 ENGINE
714 Addinol Superstar MX 1547 ENGINE

172 Doc. p/n 100-00087 Rev. A


715 Alpine RSL 5W40 ENGINE
717 Total Fluid AT 42 GEAR-PRESSURE
718 Petronas Hidroback ISO 32 HYDRAULIC
719 Unotec Matrix EP/TK HYDRAULIC
720 Lion Petroleo Lionmult Gear HD 80W90 GEAR-SPLASH
721 Power Texco 40W ENGINE
722 Keplen Oil Keen Press ISO 150 GEAR-SPLASH
723 Keplen Oil Keen Supra AW ISO 68 HYDRAULIC
724 Unotec Paragon TM 75W80 GEAR-PRESSURE
725 Lubri Motor's Multigear 85W140 GEAR-SPLASH
726 Kelpen Oil Keen Slide ISO 68 HYDRAULIC
728 Kelpen Oil Keen HBP ISO 46 HYDRAULIC
731 Unotec Surtac 2000 85W140 GEAR-SPLASH
732 Summit SH ISO 150 GEAR-SPLASH
733 Kelpen Oil Keen HBP ISO 68 HYDRAULIC
734 Kelpen Oil Keen Supra ISO 68 HYDRAULIC
735 GTOIL 40W ENGINE

738 Power Texco 30W ENGINE


739 Agecom SE/CD 40W ENGINE
740 Kelfus 001 10W30 API GL4 ENGINE

742 Unotec Decatlhon TM HTC 90W GEAR-PRESSURE


743 Agecom Text ISO 68 HYDRAULIC
744 GTOIL 50W ENGINE
745 Agecom LG ISO 68 HYDRAULIC
746 Power Texco 90W GL4 ENGINE
747 Kelpen Oil AW ISO 68 HYDRAULIC
748 Unotec Decatlhon TM Gold CCL 85W140 GEAR-SPLASH

750 Kelpen Oil Keen Slide ISO 150 GEAR-SPLASH


751 PolyOil PolyLub H ISO 68 HYDRAULIC
752 Agecom Hercon ISO 68 HYDRAULIC
753 Lion Petroleo Lion Gear EP 90W GEAR-SPLASH
754 Lubri Motor's Redux ISO 220 GEAR-SPLASH
755 Idemitsu Daphne Super Turbine Oil SP32 TURBINE-STEAM
756 Idemitsu Daphne Super Turbine Oil SP46 TURBINE-STEAM
757 Idemitsu Daphne Super Turbine Oil SP68 TURBINE-STEAM
758 Idemitsu Daphne Super Gear Oil 100 GEAR-SPLASH
759 Idemitsu Daphne Super Gear Oil 150 GEAR-SPLASH
760 Idemitsu Daphne Super Gear Oil 220 GEAR-SPLASH
761 Idemitsu Daphne Super CS 68 HYDRAULIC

763 Cat DEO 15W40 ENGINE

173
764 Cat DEO 10W30 ENGINE
765 Cat Hydo Advanced 10 GEAR-PRESSURE
766 Cat TDTO ENGINE
767 Cat NGEO EL350 ENGINE
768 Castrol Enduron Low SAPS 10W40 ENGINE-NAT GAS

770 Cat 10W ENGINE


771 Cat Hydo 30W GEAR-PRESSURE
772 Cat TDTO 50W ENGINE
773 Cat TDTO 60W ENGINE
Placeholder for fluids 774 through 780
781 Tribology Inc. T-HYD 150 GEAR-SPLASH

783 Shell Tegula V32 GEAR-PRESSURE


784 Mobil DTE FM 32 HYDRAULIC
785 Mobil DTE FM 220 GEAR-SPLASH
786 Mobil DTE FM 320 GEAR-SPLASH
787 Mobil DTE FM 460 GEAR-SPLASH
788 Mobil Vacuoline 533 GEAR-PRESSURE
801 Mobil Rarus 827 COMPRESSOR
802 Sharples 9CH28B COMPRESSOR
803 Quaker Quinto Lubricant 888-46 HYDRAULIC
804 Ingersol Rand OCU-0 Centacs COMPRESSOR
805 Mobil SHC 627 GEAR-SPLASH
806 Mobil SHC 626 GEAR-SPLASH
807 Mobil SHC 630 GEAR-SPLASH
808 Mobil SHC 634 GEAR-SPLASH
809 Mobil SHC 629 GEAR-SPLASH
810 Startex Turbo Plus 15W40 API CH-4 ENGINE
811 Startex Super Plus 10W30 API SL ENGINE
812 Startex Super Plus 20W50 API SL ENGINE
813 Shell Turbo Oil CC 32 TURBINE-CCGT
814 Shell Turbo Oil CC 46 TURBINE-CCGT

817 Total Carter EP 320 GEAR-SPLASH


818 Total Preslia 32 TURBINE-STEAM
819 Total Preslia 46 TURBINE-STEAM
820 Total Cirkan ZS 220 GEAR-PRESSURE
821 Total Dacnis VS 46 COMPRESSOR
822 Total Azolla ZS 32 HYDRAULIC
823 Total Azolla ZS 46 HYDRAULIC
824 Total Azolla ZS 150 GEAR-PRESSURE

174 Doc. p/n 100-00087 Rev. A


825 Wolf's Head 100% Syn 5W30 ENGINE-NAT GAS
Place holder for fluids 826 through 873
874 Mobil DTE 832 TURBINE-CCGT
875 Galco 2009M ENGINE-HEAVY DUTY
876 Castrol Duratec L ENGINE
877 Castrol Duratec MX ENGINE
880 14% ETOH/Gasoline ETHANOL IN GASOLINE
881 Mobil SHC Cibus 46 ISO 46 HYDRAULIC
882 Yancey Elite 50W Transmission Oil GEAR-PRESSURE
883 Chevron Tegra Synthetic Gear Oil 680 GEAR-SPLASH
884 Chevron Hypersyn 68 HYDRAULIC
885 Mobil SHC 636 GEAR-SPLASH
886 MobilGear 600XP 320 GEAR-SPLASH
887 Mobil Nuto H 46 HYDRAULIC
888 Chevron Rykon 32 HYDRAULIC
889 Chevron Tegra Synthetic Barrier Fluid SLIDEWAY
890 Chevron Regal 68 HYDRAULIC
891 MobilGear SHC 150 GEAR-SPLASH
892 Chevron Gear ISO 1000 GEAR-SPLASH
893 Concoco Gear 80W90 GEAR-SPLASH
894 Concoco Hydroclear Multipurpose R&O 68 HYDRAULIC
895 Chevron Clarity 220 GEAR-SPLASH
896 Schaeffer 280 GEAR-SPLASH
898 Chevron Meropa 460 GEAR-SPLASH
899 Conoco Gear 460 GEAR-SPLASH
900 Conoco Hydroclear Multipurpose R&O 150 GEAR-SPLASH
901 Chevron Rando HD 68 HYDRAULIC
902 Chevron Texaco Clarity 220 GEAR-SPLASH
903 Schaeffer H-1 Food Grade 280 GEAR-SPLASH
904 Chevron Rando HD 220 GEAR-SPLASH
905 Mobil Spartan EP 460 GEAR-SPLASH
906 Conoco Gear 220 GEAR-SPLASH
907 Mobil DTE 732 TURBINE-CCGT
908 Yancey Hydraulic 10W HYDRAULIC
909 Chevron Hypersyn 46 HYDRAULIC
910 Yancey Elite 30W Transmission Oil GEAR-PRESSURE
911 Conoco Hydroclear Multipurpose R&O 32 HYDRAULIC
912 Mobil SHC 524 GEAR-SPLASH
913 Mobil SHC 625 GEAR-SPLASH
914 Chevron Cylinder 680 GEAR-SPLASH
915 Chevron Tegra Synthetic 460 GEAR-SPLASH
916 Chevron GST 68 TURBINE-STEAM
917 Chevron GST 46 TURBINE-STEAM
918 Chevron GST 32 TURBINE-STEAM
919 Chevron Tegra Synthetic Compressor Oil 32 COMPRESSOR

175
920 Yancey Super HD 80W90 GEAR-PRESSURE
921 Mobil SHC 680 GEAR-SPLASH
922 Conoco Hydroclear Multipurpose R&O 220 GEAR-SPLASH
923 Mobil SHC 632 GEAR-SPLASH
924 Chevron Tegra Synthetic Gear Lube 460 GEAR-SPLASH
925 Chevron Rando HD 150 GEAR-SPLASH
926 Mobilgear SHC 320 GEAR-SPLASH
927 Chevron Rando HD 46 HYDRAULIC
928 Yancey AW 68 HYDRAULIC
929 Yancey Complex 150 GEAR-SPLASH
930 Shell Rimula Super 15W40 ENGINE
931 JP-8 N/A
932 Jet A N/A
933 OSP-HF-46 HYDRAULIC
934 BP Energol SHF-HV 100 HYDRAULIC
935 BP Energol SHF-HV 32 HYDRAULIC
936 BP Energol SHF-HV 68 HYDRAULIC
937 Castrol Hyspin AWH-M 100 HYDRAULIC
938 Castrol Hyspin AWH-M 68 HYDRAULIC
939 Castrol Hyspin AWH-M 32 HYDRAULIC
940 Castrol Alpha SP 220 GEAR-SPLASH
941 BP Energol OE-HT 30 ENGINE-HEAVY DUTY
942 Castrol Tection Global 15w-40 ENGINE
943 BP Energol GR-XP 150 GEAR-SPLASH
944 BP Energol GR-XP 100 GEAR-SPLASH
945 BP Energol IC-HFX 303 ENGINE-HEAVY DUTY
946 BP Energol IC-HFX 404 ENGINE-HEAVY DUTY
947 Chevron Oil FM ISO 460 GEAR-SPLASH
948 Chevron Tegra Synthetic Compressor Oil 68 COMPRESSOR
949 Milroyal AGMA Compounded Gear 7 GEAR-SPLASH
950 Chevron Oil FM ISO 100 GEAR-PRESSURE
951 Napa UFMO 15W40 ENGINE
952 Chevron Oil FM ISO 32 HYDRAULIC
953 Chevron Oil FM ISO 68 HYDRAULIC
954 Milton Roy EP 95 Lubricant GEAR-SPLASH
955 Milton Roy Hydraulic Fluid HYDRAULIC
956 Pulsalub #1 Hydraulic HYDRAULIC
957 WSP Duoprime Oil 90 HYDRAULIC
958 Micare TO-04 SAE10W ENGINE
959 Micare TO-04 SAE30W ENGINE
960 Micare TO-04 SAE50W ENGINE
961 Conoco-Phillips Syncon R&O 32 HYDRAULIC
962 Ingersoll Rand Compressor COMPRESSOR

176 Doc. p/n 100-00087 Rev. A


963 B5 FAME in DIESEL
964 B10 FAME in DIESEL
965 B15 FAME in DIESEL
966 B3 FAME in DIESEL
967 B20 FAME in DIESEL
968 B0.25 FAME in DIESEL
969 B30 FAME in DIESEL
970 B80 FAME in DIESEL
971 Shell Morlina S4 B680 GEAR-SPLASH
972 Shell Spirax HD 85W-140 GEAR-SPLASH
973 Shell Morlina S4 B220 GEAR-SPLASH
974 Shell Morlina S4 B460 GEAR-SPLASH
975 Shell Paper Machine S3 220 GEAR-PRESSURE
976 Shell Morlina S4 B150 GEAR-SPLASH
977 Shell Spirax S4 ATFX GEAR-SPLASH
978 Shell Spirax S4 TXM TRANSMISSION
979 Shell Omala S4 GX220 GEAR-SPLASH
980 Brad Penn DDS Plus 9124 ENGINE-HEAVY DUTY
981 Castrol Optigear Synthetic X 320 GEAR-PRESSURE
982 Mobil DTE PM 150 GEAR-PRESSURE
983 Mobil DTE 10 Excel 46 HYDRAULIC
984 Mobil DTE PM 220 GEAR-PRESSURE
985 John Deere Hy-Gard Hydraulic/Transmission Oil HYDRAULIC
986 Castrol Dexron VI Transmax TRANSMISSION
987 Chevron Gear Compound Iso 220 GEAR-PRESSURE
988 Gardner Denver Aeon CF 150 COMPRESSOR
990 Valvoline Premium Blue SAE 15W-40 Diesel ENGINE-NAT GAS
991 Sullair 24KT COMPRESSOR
992 Conoco Iso 32 AW HVI Hidraulico HYDRAULIC
993 Lubriquem Redlub Iso 220 EP GEAR-PRESSURE
994 Lubriquem Redlub Iso 460 EP GEAR-PRESSURE
995 Autolub II Dexron HYDRAULIC
996 Lubriquem Tramilub 80W90 GL5 GEAR-PRESSURE
997 Conoco Megaflow AW HVI Iso 46 HYDRAULIC
998 Whitmore Paragon Iso 680 GEAR-PRESSURE
999 Dialube SAE 40 ENGINE-HEAVY DUTY
1000 Hidralub Iso 680 HYDRAULIC
1001 Synlube WS 460 GEAR-PRESSURE
1002 Mobil Glygoyle 220 GEAR-PRESSURE
1003 Purple Crown 220 GEAR-PRESSURE
1004 CAT NGEO EL250 SAE 40 ENGINE-NAT GAS
1005 Shell Rotella 30 ENGINE-NAT GAS
1006 Shell Rimula R4X 15W-40 ENGINE-NAT GAS
1007 Shell Rotella T6 5W-40 ENGINE-NAT GAS
1008 Shell Rotella T5 10W-40 ENGINE-NAT GAS

177
1009 Shell Rotella T Triple Protection 15W-40 ENGINE-NAT GAS
1010 Chevron Delo 400 LE SAE15W-40 ENGINE-NAT GAS
1011 Mobil 1 Turbo Diesel Truck 5W-40 ENGINE-NAT GAS
1012 Mobil Delvac 1300 Super 15W40 ENGINE-NAT GAS
1013 Mobil Delvac 1 LE 5W-40 ENGINE-NAT GAS
1014 Mobil EAL 224H ENGINE-NAT GAS
1015 Royal Purple Synergy 150 ENGINE-NAT GAS
1016 Royal Purple ThermylGlyde 680 ENGINE-NAT GAS
1017 Royal Purple ThermylGlyde 150 ENGINE-NAT GAS

1019 Sullair Sullube ENGINE-NAT GAS


1020 Shell Rimula R3 X 15W-40 ENGINE-NAT GAS

1022 Valvoline Premium Blue GEO-MA 40 ENGINE-NAT GAS

1024 EcoSafe EHC 46 ENGINE-NAT GAS


1025 Jomo EP 3080 GEAR-SPLASH

1028 Chevron Hipersyn ISO 32 COMPRESSOR


1029 Chevron Tegra Syn ISO 320 GEAR-SPLASH
1030 Chevron Hydraulic Oil AW 32 HYDRAULIC
1031 Chevron Hydraulic Oil AW 68 HYDRAULIC
1032 Chevron Hydraulic Oil AW 100 HYDRAULIC
1033 Shell Mysella S5 N40 ENGINE-NAT GAS

1035 Guardoll SAE 10W-30 ENGINE


1036 Guardoll SAE 15W-40 ENGINE
1037 Mobil Delvac 75W-90 GEAR-SPLASH
1038 Red Line D6 ATF TRANSMISSION
1039 Red Line D4 ATF TRANSMISSION
1040 Red Line DCTF TRANSMISSION
1041 Red Line 75W-140 GEAR-SPLASH
1042 Red Line 75W90 GEAR SPLASH
1043 Red Line MTL TRANSMISSION
1044 Red Line MT-90 TRANSMISSION
1045 G-Profi GT LA 10W-40 ENGINE
1046 G-Profi MSJ 5W-30 ENGINE

178 Doc. p/n 100-00087 Rev. A


A Replacement Parts and Consumables List

The following replacement parts and consumables are available from Spectro Scientific:

P/N Description
800-00019 MiniLab 53 Standard Accessories Kit:
LNF-509: LNF Calibration Check Fluid 2806
LNF-903: Q200 Series Preventive Maintenance Kit
LNF-545: Ferrous Calibration Fluid for Magnetometer
FL310: IR Check Fluid 5ml
P-11177: Positive Displacement Pipette 10-100 ul, box of192
P-11178: Pipette Tips 10-100 uL box of192
PV1026: Visc 120 Standard, 20 mL
PV1027: Visc 30 Standard, 20 mL
P-10739: Fuse SMD 1.5A 125V Fast Acting
MHM-10605: 10 mL plastic syringe, quantity 100
M99919: 4 oz ASP bottles, quantity 500

MHM-92631 Filter patch, 5µ, Durapore PVDF, 25 mm dia, quantity 100

MHM-92640 Petrislides w/pad for filter patches, 47 mm dia, quantity 100

P-11101 Cable USB A-B Male 3 Meter

P-11104 USB Adapter A-Male to Mini B Male 2 meter

P-11153 Universal Cable Kit

P-11058 USB to Mini A Plug Adapter

PV1020 Plug-In AC Adapter 18 VDC 2.5A 2.1 mm

179
B Specifications

This section provides the MiniLab 53 specifications:

Product Information

Mineral and synthetic lubricants, including gear, engine, hydraulic, turbine,


Applications
and distillate fuels

Particle count, size distribution, and ISO codes per ISO 4402/4406


Wear particle counts and size distribution by wear mode:
cutting, fatigue, sliding, non-metallic, fibers
Total ferrous, ppm
Output Ferrous particle count and size distribution
Total Acid Number (TAN), mg KOH/g
Oxidation, Abs/mm 2
Water, ppm
40C Kinematic viscosity, cSt

Methodology ASTM D7596

Standard Analytical Particles: 4 µm-100 µm


Range Viscosity: 1-700cSt at 40C
Water: 0.1-6.5% (100-65,000 ppm)
TAN: 0.25-6 mg KOH/g (oil specific)

Accuracy Particle count:+/- 1/2 ISO Code


Viscosity:
≤±3% RSD of value over range 1-350 cSt
≤±5% RSD >350 cSt
TAN <±3% RSD typical
Total ferrous <±20% RSD

Calibration Factory calibrated; field calibration not required. Validation standards sup-
plied

Operational Specifications

Environmental Oper- 5-40C ambient temperature,


ating Requirements 10-80% RH non-condensing;
2000 m maximum altitude

180
Sample Volume 5-30 mL varies with viscosity

Solvents Lamp oil, odorless kerosene or Electron 22

User Interface Specifications

Software/Operating Requires AMS OilView, LIMS module, Monitor and PC/laptop with Windows


System 7 Pro, 32 or 64-bit, US English version

Power Requirements

Power 1-Phase, 110 VAC/60Hz or 220 VAC/50Hz, 110 W

Mechanical Specifications
MiniLab 53:
35 cm x 50 cm x 53 cm (13.8" x 19.7" x 21")
Dimensions (HxWxD)
Homogenizer and Stand:
74 cm x 48 cm x 28 cm (29" x 18" x 11")

Weight (Minilab 53
14 kg (31 lbs)
with homogenizer)

Compliance

CE Mark EMC directive, RoHS, CSA

181 Doc. p/n 100-00087 Rev. A

You might also like