Nr. : 126
Datum / Date: 08/02/1999
Pag./page: 1/9
Introduction
Picture
Introduction In this Packo Flash we introduce the temperature controller PCV2-2, the new
execution of the PCV2 controller.
Functions The PCV2-2 has the same functions as the PCV2 controller:
• digital thermometer
• temperature controller;
• interval agitation..
The sensor
The sensor of the PCV2-2 is NOT the same as for the PCV2. When a PCV2-
2 is ordered, the correct sensor is delivered. The PCV2-2 sensor can also be
ordered separately.
Remark: the PCV2-2 cable carries a label with the mention S10-6P.
Technical data
Overview Below you will find an overview of the technical date of the PCV2-2:
Description Value
Tension 230 V AC, 50/60 Hz
Relay contact 2
Maximum current 16A AC1 / 250V ~
Display 13 mm LED, 3 positions
Type of controller two point controller
Measurement range -30°C/ +70°C
Length of the sensor cable 2 metre
Cut away for front panel 90 x 42 mm
Degree of protection (front panel IP64
housing)
Connection clamping screw
Working temperature 0°C/+ 50°C
Maximum humidity 75% (no vapour)
FLASH
Nr. : 126
Datum / Date: 08/02/1999
Pag./page: 4/9
1
6
2
3 7
4
5 8
Parts Below you will find an overview of the parts of the control panel:
Nr Part Function
1 LED Is lit when the agitator is running.
2 LED Is lit when the cooling unit (DX) or the ice water pump (IB)
is running.
3 LED Is lit when temperature setpoint T1 is selected.
4 LED Is lit when temperature setpoint T2 is selected.
5 Button Modification of the temperature setpoints T1/T2, in
Circuit diagram
Parts
Below you will find an overview of the parts of the circuit diagram:
Part Function
RW Agitator
Vo Sensor
O1 Option 1: external switch for the selection of the
temperature setpoint T1/T2.
O2 Option 2: external button for manual agitation.
S Tension
K Cooling
FLASH
Nr. : 126
Datum / Date: 08/02/1999
Pag./page: 6/9
Overview
or .
1 = locked: the temperature setpoints can only be modified by setting
parameters C1 & C2.
Step Action
1
Press and simultaneously for ± 3 seconds. The display shows C1.
2 Press the SET button to see the set value on the display.
3
Press the SET button and (to raise the value) or (to lower the value)
simultaneously...
4
Press to proceed with the next parameter.
5 Repeat steps 2 to 4 until all parameters are set.
6
• Press and simultaneously for 5 seconds or
• wait for 60 sec.
to return to the temperature display.
FLASH
Nr. : 126
Datum / Date: 08/02/1999
Pag./page: 7/9
indicative range
P20 Minimum value temperature setpoint T1 0 - 3,1 2,0
P21 Maximum value temperature setpoint T1 3,3 - 50,0 10
P22 Minimum value temperature setpoint T2 0 - 3,1 2,0
P23 Maximum value temperature setpoint T2 3,3 - 50 20
P30 Minimum value differential T1 0,1 - 0,6 0,5
P31 Maximum value differential T1 0,8 - 100 1,5
P32 Minimum value differential T2 0,1 - 0,6 0,5
P33 Maximum value differential T2 0,8 - 100 1,5
P70 Minimum duration of cooling 0 - 999 60 sec.
P71 Minimum rest period between 2 cooling 0 - 999 180 sec.
cycles
P80 Selection T1/T2: 2
0 = not possible
Error messages
Overview Below you will find an overview of the possible error messages: