Professional Documents
Culture Documents
Ventilation Handbook
Ventilation Handbook
This review provides an update on the various techniques that are available to monitor patients
during mechanical ventilation with an emphasis on clinical observations and applications in
critically ill patients. (CHEST 1999; 116:1416 –1425)
Key words: elastance; pressure-time product; pulse oximetry; resistance
Abbreviations: ABG 5 arterial blood gas; DR 5 additional resistance; Edyn,L 5 dynamic elastance of the lung;
Edyn,rs 5 dynamic elastance of the respiratory system; Est;rs 5 static elastance of the respiratory system;
f 5 respiratory frequency; Hb 5 reduced hemoglobin; O2Hb 5 oxyhemoglobin; P0.1 5 mouth occlusion pressure at
0.1 s after onset of inspiratory effort; Paw 5 airway pressure; PEEP 5 positive end-expiratory pressure;
PEEPi 5 intrinsic positive end-expiratory pressure; Pes 5 esophageal pressure; PETco2 5 end-tidal Pco2 concentra-
tion; Pga 5 gastric pressure; Pinit 5 initial pressure; Pl 5 transpulmonary pressure; Ppeak 5 peak pressure;
Pplat 5 plateau pressure; PSV 5 pressure support ventilation; PTP 5 pressure-time product; Rmax 5 maximum
resistance; Rmin 5 minimum resistance; Sao2 5 arterial oxygen saturation; Spo2 5 pulse oximeter estimate of arterial
oxygen saturation; Vt 5 tidal volume; WOB 5 work of breathing
and it increased to 5.1 6 2.7% when Sao2 was ing standardized motion, Masimo SET exhibited
# 90%.5 much lower error rates (defined as percentage of
time that the oximeter error exceeded 5, 7, and 10%)
Limitations: Pulse oximeters employ only two and dropout rates (defined as the percentage of time
wavelengths of light, and thus can distinguish only that the oximeter provided no Spo2 data) than did
two substances, Hb and O2Hb. Elevated carboxyhe- the Nellcor N-200 and Nellcor N-3000 oximeters
moglobin and methemoglobin levels can cause inac- (Nellcor Puritan-Bennett Corp; Pleasanton, CA).11
curate oximetry readings.1 Anemia does not appear In 50 postoperative patients, the pulse oximeter’s
to affect the accuracy of pulse oximetry: in nonhy- alarm frequency was decreased twofold with a
poxemic patients with acute anemia (mean Hb, Masimo SET system vs a conventional oximeter
5.2 6 0.3 [SEM] g/dL), pulse oximetry was accurate (Nellcor N-200).12
in measuring O2 saturation with a bias of only
0.53%.6 Moreover, in patients with sickle cell anemia Clinical Applications: Moller et al13 conducted the
who presented with acute vaso-occlusive crisis, Ortiz first prospective, randomized study of pulse oximetry
et al7 found that pulse oximetry overestimated Sao2 on the outcome of anesthesia care in 20,802 surgical
by an average of 3.4%; the error of Spo2 was never patients. A 19-fold increase in the detection of
enough to misdiagnose either hypoxemia or normox- hypoxemia (defined as an Spo2 , 90%) was noted in
emia in such patients. the oximeter group vs the control group. Myocardial
Motion artifact continues to be a significant source ischemia was more common in the control group
of error and false alarms.1,8 In a recent prospective than in the oximetry group (26 and 12 patients,
study in an ICU setting, Spo2 signals accounted for respectively). However, pulse oximetry did not de-
almost half of a total of 2,525 false alarms9 (Fig 2). crease the rate of postoperative complications or
Various methods have been employed to reject mortality.
motion artifact, but have met with little success. An Pulse oximetry can assist with titration of the
innovative technologic approach, termed Masimo fraction of inspired oxygen concentration (Fio2) in
signal extraction technology (Masimo SET; proto- ventilator-dependent patients, although the appro-
type), was recently introduced to extract the true priate Spo2 target depends on a patient’s pigmenta-
signal from artifact due to noise and low perfusion.10 tion.5 In white patients, a Spo2 target value of 92%
This technique incorporates new algorithms for pro- predicts a satisfactory level of oxygenation, whereas
cessing the pulse oximeter’s red and infrared light in black patients, this target may result in significant
signals that enable the noise component, which is hypoxemia. While a higher target Spo2 value of 95%
common to the two wavelengths, to be measured and avoids hypoxemia in black patients, some will have
subtracted. When tested in healthy volunteers dur- Pao2 values as high as 198 mm Hg; if such patients
Respiratory Mechanics
Measurements of respiratory mechanics in a re-
laxed ventilator-dependent patient can be obtained
using the technique of rapid airway occlusion during
constant flow inflation.28 Rapid airway occlusion at
the end of a passive inflation produces an immediate
drop in both airway pressure (Paw) and transpulmo-
nary pressure (Pl) from a peak value (Ppeak) to a
lower initial value (Pinit) followed by a gradual
decrease until a plateau (Pplat) is achieved after 3 to
5 s29,30 (Fig 3). Pinit is measured by back extrapola-
tion of the slope of the latter part of the pressure
tracing to the time of airway occlusion.28 Pplat on the
Paw, Pl, and pleural pressure (Pes) tracings repre-
sents the static end-inspiratory recoil pressure of the
total respiratory system, lung, and chest wall, respec-
tively. Figure 3. Flow (inspiration upwards), Paw, Pl, and Pes tracings
in a representative patient during passive ventilation. An end-
inspiratory occlusion produced a rapid decline in both Paw and Pl
from Ppeak to a lower Pinit, followed by gradual decrease to
Elastance Pplat. Reprinted with permission from Jubran and Tobin.29
The end-inspiratory airway occlusion method is
clinically used to measure the static compliance of
the respiratory system or its reciprocal, elastance of monary edema than in patients with COPD, whereas
the respiratory system (Est,rs), according to the static chest wall elastance was similar in both patient
following equation31: groups.30
Dynamic Compliance
Est,rs 5 (Pplat 2 PEEPi)/Vt
An index that commonly is referred to as effective
where Pplat is plateau pressure obtained after oc- dynamic compliance, or the dynamic characteristic,
cluding the airway, PEEPi is intrinsic positive end- can be derived by dividing the ventilator-delivered
expiratory pressure (PEEP), and Vt is tidal volume. Vt by [peak Paw 2 PEEP]. This index is not a
Using an esophageal balloon catheter, Est,rs can be measure of true thoracic compliance because peak
partitioned into its lung and chest wall components Paw includes all of the resistive and elastic pressure
by dividing [Pplat 2 PEEPi] by Vt on the Pl and Pes losses of the respiratory system and endotracheal
tracings, respectively (Fig 3). tube. Alternatively, dynamic elastance of the respi-
In mechanically ventilated patients with acute ratory system (Edyn,rs) can be obtained by dividing
respiratory failure secondary to COPD or pulmonary the difference in Paw at points of zero flow by the
edema, Est,rs is higher than in normal subjects.30 delivered Vt.31 Accordingly, Edyn,rs can be com-
Static lung elastance is higher in patients with pul- puted according to the formula:
Figure 5. Relationship between wasted effort (quantitated as wasted PTP) and resistance, elastance,
and PEEPi in 11 ventilator-dependent patients. Wasted PTP significantly correlated with resistance,
elastance, and PEEPi. On multiple linear regression analysis, 93% of the variance in wasted PTP among
patients could be explained by these three variables. Based on data from Leung et al.40
Pressure-Time Product
A significant limitation of measurements of respi-
ratory work is that they underestimate energy expen-
diture during isometric contractions. To overcome
this problem, many investigators have measured
pressure-time product (PTP) during mechanical
ventilation.49,50 This is calculated as the time integral
of the difference between esophageal pressure (Pes)
measured during assisted breathing and the recoil
pressure of the chest wall measured during passive
ventilation with Vt and flow settings that are iden- Figure 6. Pes (continuous line) and estimated recoil pressure of
tical to the assisted breaths. While this can be the chest wall (Pescw, interrupted line) tracings in a patient
achieved satisfactorily during assist-control ventila- receiving PSV. Top, A, pressure tracings have been superimposed
so that Pescw is equal to Pes at the onset of the first inspiratory
tion and intermittent mandatory ventilation, a prob- effort, and the integrated difference (hatched area) represents
lem arises during PSV because lung volume and upper-bound inspiratory PTP. Bottom, B, pressure tracings have
inspiratory flow vary from breath to breath in this been superimposed so that Pescw is equal to Pes at the first
moment of transition from expiratory to inspiratory flow, and the
mode. To overcome this problem, a modified ap- integrated difference (hatched area) represents lower-bound
proach in the calculation of PTP has been de- inspiratory PTP. Reprinted with permission from Jubran et al.47
Conclusions
Several devices can be used to monitor a patient’s
gas exchange function, respiratory neuromuscular
capacity, respiratory mechanics, and breathing effort
during mechanical ventilation. Use of the derived
information permits the physician to better tailor
ventilator settings to an individual patient’s require-
ments with the promise of enhancing patient com-
fort. In addition, such measurements are helpful in
characterizing the pathophysiology of a patient’s
respiratory disorder, minimizing the risk of ventila-
tor-induced complications, and determining the pa-
tient’s readiness for the discontinuation of ventilator
support.
References
1 Jubran A. Pulse oximetry. In: Tobin MJ, ed. Principles and
practice of intensive care monitoring. New York, NY:
McGraw-Hill, 1998; 261–287
2 Bowton DL, Scuderi PE, Haponik EF. The incidence and
effect on outcome of hypoxemia in hospitalized medical
patients. Am J Med 1994; 97:38 – 46
3 Wukitisch MW, Peterson MT, Tobler DR, et al. Pulse
Figure 7. Recordings of flow, Paw, and transversus abdominis
electromyogram (EMG) in a patient with COPD receiving oximetry: analysis of theory, technology, and practice. J Clin
pressure support of 20 cm H2O. The onset of expiratory muscle Monit 1988; 4:290 –301
activity (vertical line) occurred when mechanical inflation was 4 Webb RK, Ralston AC, Runciman WB. Potential errors in
only partly completed. Reprinted with permission from pulse oximetry: II. Effects of changes in saturation and signal
Parthasarathy et al.45 quality. Anesthesia 1991; 96:207–212