Lab Mscope

You might also like

Download as doc, pdf, or txt
Download as doc, pdf, or txt
You are on page 1of 2

KNOW YOUR MICROSCOPE

Basic terminology Three rules for cleaning a microscope:


Light source 1. Don't! It easy to cause permanent damage
The light source, or "illuminator", should have by injudiciously cleaning a microscope.
adjustable brightness and may include an adjustable field
diaphragm to minimize scattered light.  Clean eyepieces and other optical surfaces only
when absolutely necessary. It is better to ignore
Condenser (light-gathering lens) dirt than to cause permanent damage to the
The condenser is located beneath the stage and microscope.
includes an adjustable aperture, controlled by a diaphragm  Never the touch the objective lens surface, unless
within the condenser. Proper condenser adjustment, or is you own the microscope and have been properly
essential for optimal resolution (see "Kohler illumination" trained.
on the other side of this page).

Stage (specimen support platform) 2. If you see dirt when you look through the
microscope, determine by logic and
The stage supports the microscope slide and allows
specimen movement in x-y plane. The stage may move up
experiment where the dirt is located.
and down for specimen focus.  Dirt on a slide moves when you move the slide.
 Dirt on the condenser goes in and out of focus
Specimen
when you focus the condenser.
Note that the slide and coverslip are both parts of the
 Dirt on an eyepiece will move when you rotate the
optical system. At high magnification, optimal optical
eyepiece.
function requires that coverslip thickness be appropriate
for the objective lens.  Dirt on an objective cannot be seen while viewing
through the microscope, but manifests itself as a
Objective lens (image-forming lens) decrease in image quality.
Functionally, the objective lens is the most critical (and
most expensive) element of the microscope. Objectives 3. If dirt is objectionable, use the least-
may move up and down for specimen focus. Never touch intrusive procedure to make it go away.
the optical surface of the objective lens.
 Some dirt can blown away without touching the
Eyepiece (enlarging lens) surface. Use this procedure freely, but blow gently
and be careful not to spit.
If eyepieces are paired, then interpupillary distance is
adjustable. At least one eyepiece should be independently  Dirt on slides may be wiped off at will, using the
focussable. Eyepieces are removeable; when held same care you would use for eyeglasses.
upsidedown, an eyepiece can serve as a hand lens.  Dirt on an eyepiece is often accentuated by having
_______________________________________ the condenser aperture too small; such dirt may
become much less conspicuous when the
condenser diaphragm is set correctly (see "Kohler
illumination" on the other side of this page).
Three rules for carrying a microscope:
 Dirt on the condenser may become inconspicuous
1. Always carry the microscope upright if you lower the condenser (i.e., move the dirt out
(otherwise the eyepieces can fall out). of focus).
2. Use both hands.  If dirt must be removed manually, use clean lens
paper and moisture from breath. Wipe gently in
3. Give the job your full attention, just as if you one direction. Do not scrub with a circular
were carrying a baby. motion. Do not try to clean an objective lens.
Kohler Illumination Magnification
The following instructions apply to most research  Magnification is determined in principle by
microscopes, including the multi-viewer- multiplying the power of the Objective Lens
microscopes in Lindegren Room 206. Condenser by the power of the Eyepiece.
adjustment may be simplified on student  Magnification is determined in practice by
microscopes.
calibration with a specimen of known size.
Initial set up: Turn on the lamp, place a slide on the o For research purposes, calibration is done with a
stage, focus on the specimen, adjust interocular distance to reference standard, such as a "calibration slide",
match that of your own eyes. engraved with a very finely-calibrated ruler.
o An "eyepiece micrometer" (a small, transparent ruler
1. Focus the condenser. mounted within the eyepiece) can be used to measure
Close the field diaphragm located on the base of the specimens, after calibration with a specimen of
microscope. This will reduce the illuminated field of known size (see preceding item).
view, as seen through the eyepieces, to a small circle of
light. Then adjust the substage condenser up or down
 In routine practice, magnification is estimated
until the edge of this circle appears sharp. with reference to familiar objects of known
size, such as red blood cells (the diameter of
2. Center the condenser. human RBCs is approximately 6-8µm).
Open the field diaphragm until the illuminated
circle almost fills the viewing area. Then adjust the
screws at either side of the condenser until this circle is CALIBRATE YOUR OWN MICROSCOPE.
precisely centered in your field of view. Now open the  A good calibration standard can be made by
field diaphragm completely to restore illumination over using a size-reducing photocopier to make a
the entire field of view.
miniature copy of a millimeter ruler. Measure
3. Adjust the condenser diaphragm. the actual size of this miniature ruler, so you
Remove one eyepiece and look into the open tube. know the length between marks.
You should see an illuminated field of view whose size  Put your calibration standard on a microscope
is controlled by the diaphragm within the condenser.
Adjust the condenser diaphragm until its edge just slide and view it through the microscope.
barely encroaches on the illuminated field of view. Measure edge-to-edge across measure the field
Then replace the eyepiece. of view. Repeat for each magnification (i.e.,
This adjustment should be repeated whenever the each different objective lens). Or, measure
objective lens is changed (i.e., whenever magnification once with lowest power and use arithmetic to
is changed).
compute the change when magnification is
4. Adjust illuminator brightness. increased. (For example, if magnification is
Use the voltage regulator on the base of the microscope increased from 4x to 40x, which is a factor of
to adjust image brightness to suit your eyes.
10, the real area visible in the field of view
5. Repeat steps 3 and 4 whenever you change will decrease by a factor of 10.)
objectives.  If your microscope has an eyepiece pointer,
you can also measure the thickness of the
pointer and use that number as a standard for
Congratulations. Your microscope should now estimating the size of objects much smaller
be adjusted for proper Kohler illumination. than the field of view.

___________________________________________________________________________________________________

Additional details may be found at http://www.siumed.edu/~dking2/intro/kohler.htm

For yet more information . . .


http://www.micrographia.com/tutoria/micbasic/micbpt01/micb0100.htm
http://www.microscopy.fsu.edu/primer/anatomy/anatomy.html

You might also like