KMC650 Troubleshooting (X-Ray)
KMC650 Troubleshooting (X-Ray)
Table of Contents
Unit 1. Tube(Monotank)-----------------------------------------------------------------------------------------------------------------------------------------
2
Unit 1. Tube(Monotank)
WARING
You MUST switch off the Unit before replacing the part and ONLY an approved GEMSS Medical technician is qualified to carry out any intervention
on the Unit while it is switched on.
Definition :
This device emits X-ray by creating electron beam which moves fast and collides with the material.
How to check :
TP9(IP) - mA Measurement
1. kV, mA Feedback
Scope Value
- Small: 250mV 1mA
- Large: 100mV 10mA
Scope Value
1V DC 25KV
GND
2. Tube Filament
Definition :
Drive Board is a negative circuit that changes direct current (DC) to alternating current (AC)
How to check :
1. IGBT terminal short circuit test
1) Must remove the input power(DC220V) cable from Drive board before checking
+220VDC
Definition :
PWR Board is a electronic power supply circuit that changes alternating current (AC) to direct current (DC)
How to check :
1. PWR Board check point
1. Input voltage:
110VAC
3. Output voltage
+220VDC
Definition :
Main controller is a control board. It is responsible for varies electrical control by combining control switch, relay, electric contractor,
and controller.
How to check :
1. Measure voltage of TP2, TP4 to check the kV and mA signal data