Professional Documents
Culture Documents
Jurnal 1
Jurnal 1
Barbiturates
Application Note
Pharmaceuticals
O O CH3 O N O
H
R groups attached to the
number 5 carbon atom
CH3
O
O
CH2 CHCH 3
H3 C
N
HN CH 3
CH2 CH CH 2
O N O
O N O H
H
Butalbital Hexobarbital
Figure 1. Structures of barbiturates used in this study and barbituric acid.
Start with a Proven HPLC Method They maintain column efficiency and resolution
because the particle size decreases in proportion
The objective of this investigation was to develop a to the column length. To demonstrate the scalabil-
high throughput HPLC method for the analysis of ity of three different column lengths and particle
barbiturates based on an existing slower conven- sizes for the test barbiturates, Figure 2 shows an
tional HPLC method. Rapid Resolution (RR) and overlay of three chromatograms. The top chro-
Rapid Resolution High Throughput (RRHT) HPLC matogram is the original USP method for pheno-
column technology allows one to convert the exist- barbital with the internal standard (caffeine) but
ing method into a high throughput method easily with the three additional barbiturates. The USP
and straightforwardly to provide a gain in method specifies a 4.6-mm × 250-mm column with
productivity. L1 type stationary phase (C18 phase bonded to
silica particles) with a minimum resolution of 1.2
For established pharmaceuticals, the starting point
[3]. A ZORBAX Eclipse XDB-C18 column with
for HPLC methods is the United States Pharma-
these dimensions and 5 µm particles was selected
copoeia 27 (USP). The USP method for phenobarbi-
for this separation and produced an analysis time
tal can be found in reference 3. In addition to the
of about 32 minutes. Simply replacing this column
long acting barbiturate, phenobarbital, we added
for a shorter one, 4.6-mm × 100-mm with 3.5 µm
an ultra-short (hexobarbital), a short (allobarbital),
particles, reduced analysis time by about a factor
and an intermediate (butalbital) acting barbiturate
of 2.5 (or 13 min), or by a factor of five using a
as standards to demonstrate the power of RR and
4.6-mm × 50-mm column with 1.8-µm particles (or
RRHT columns.
7 min). These latter columns provided faster sepa-
rations with only a small loss of resolution and are
Scalability of Barbiturate Reversed-Phase HPLC Method
referred to as RR and RRHT columns, respectively.
Scalability refers to the ability of an HPLC method These data show that older methods done on larger
to use columns of different diameters and/or particle columns can be easily transferred to RR
lengths and particle sizes and still maintain the and RRHT columns without sacrificing separation
separation characteristics of the method. Earlier, it integrity. As can be noted in Figure 2, the resolu-
was demonstrated that ZORBAX columns with tion of both the 3.5- and the 1.8-µm columns is
smaller particles and shorter lengths provide simi- easily within the minimum specification of the USP
lar chromatography in a fraction of the time com- method.
pared with longer columns and larger particles [4].
2
Eclipse XDB-C18
1 4.6 × 250 mm, 5 µm
mAU
60 P/N 990967-902
Pressure = 174 bar R2,3 = 5.99 32 min.
40 R4,5 = 4.43
3 5
20 2
4
0
5 10 15 20 25 30 min
Eclipse XDB-C18
4.6 × 100 mm, 3.5 µm (Rapid Resolution, RR)
mAU P/N 961967-902 1 Caffeine (ISTD)
40 Pressure = 167 bar 2 Allobarbital
30 13 min. 3 Phenobarbital
20 R2,3 = 4.88 R4,5 = 3.72 4 Butalbital
10 5 Hexobarbital
0
Mobile phase: 60% A, 40% B
5 10 15
A = pH 4.5 Na Acetate
B = MeOH
Eclipse XDB-C18 Flow: 1 mL/min
4.6 × 50 mm, 1.8 µm (Rapid Resolution HT, RRHT) Injection volume: 2 µL
mAU P/N 927975-902 Detector: UV 254 nm
30 Pressure = 291 bar Flow cell: 2 µL, 3 mm flow path
20
R2,3 = 4.59
10
R4,5 = 3.32
7 min.
0
Figure 2. RR and RRHT column configurations increase speed and maintain sufficient resolution.
Stationary Phase Selectivity as a Function of Particle Size Figure 3 depicts an overlay of the high throughput
separation of barbiturates by equivalent 4.6-mm ×
The productivity increase by changing column 50-mm Eclipse XDB-C18 columns packed with dif-
dimensions and particle size is achieved due to the ferent particle sizes. Note selectivity (α) is the
reproducibility of the spherical silica particles and same for all three columns, independent of particle
bonded phase. The highly uniform manufacturing size. The same selectivity indicates the three differ-
process of the ZORBAX base silica, of the organosi- ent sized particles are chemically very similar. The
lane bonding process, and of the standardized test- different sized particles can be packed in different
ing of the column leads to the ability of the user to column dimensions with predictable and scalable
freely substitute columns without suffering selec- results, because they exhibit the same chromato-
tivity changes which would ultimately affect reso- graphic characteristics. Figure 3 also demonstrates
lution. One way to measure this uniformity is to that as particle size decreases, efficiency increases.
compare the selectivity factors among different Note that peak widths decrease, producing better
particle sizes. Selectivity (α) is the ratio of the sensitivity (taller peaks), as the particle size
retention factors: α2,1 = k'2/k'1 where k'1 is the decreases from 5 to 3.5 to 1.8 µm.
adjusted capacity factor for compound 1 and k'2 is
the adjusted capacity factor for compound 2.
3
1 t0 = 0.44 min
5 µm α3.2 = 1.20 α4.3 = 1.97 α5.4 = 1.13
P = 73 bar 3 5
2
4
1 2 3 4 5 6 min
1
t0 = 0.44 min
α3.2 = 1.20 α4.3 = 1.97 α5.4 = 1.14
3.5 µm
P = 111 bar 2 3 5
4
1 2 3 4 5 6 min
1
t0 = 0.46 min
α3.2 = 1.22 α4.3 = 1.94 α5.4 = 1.16
1.8 µm
P = 291 bar 2 3 5
4
1 2 3 4 5 6 min
Optimization of Resolution with RRHT technology diffusion. At lower flow velocities, the contribution
from axial diffusion to the H value may be quite
A van Deemter plot is often used to depict the high and analysis time quite long. The “C” term
changes in column efficiency, usually expressed as relates solute mass transfer from the mobile phase
H (or Height Equivalent to a Theoretical Plate, to the stationary phase and vice-versa. It is both
HETP), as a function of linear velocity (propor- flow rate and particle size dependent. For a typical
tional to flow rate). A typical van Deemter plot is column packed with larger particles, say 5- or
shown in Figure 4a. 10-µm, the column will behave like the van
Deemter plot in Figure 4a. However, for smaller
The shape of the plot can be described by Equation 1:
particles such as the 1.8-µm packings, the slope of
H = A + B/u + Cu (Equation 1) the van Deemter curve at high linear velocities is
much flatter, as depicted in Figure 4b. This flat-
In chromatography, one strives for low values of H ness is due to the superior solute mass transfer
that means high values of plates (N). The “A” term into and out of the smaller particles, even at high
in the van Deemter equation, that represents eddy flow rates. Thus, small particle columns may be
diffusion, is particle size dependent and is mini- run at these higher linear velocities and retain
mized for small particles. The “B” term is flow rate their efficiency while dramatically increasing the
dependent and is governed by longitudinal (axial) separation speed.
4
a) A hypothetical Van Deemter Plot b) A typical 1.8 µm Van Deemter Plot
6
1.0 mL/min
Minimum plate 2
height
0
Mobile phase volcity 0.0 2.0 4.0 6.0 8.0 10.0 12.0
Mobile phase velocity, u (mm/s)
Figure 4. Van Deemter plots - plate height versus mobile phase velocity.
Based on the shape of the van Deemter curve in higher linear velocity the pressure drop will
Figure 4b, in order to achieve the best overall effi- increase, especially for longer columns at room
ciency, the 1.8-µm column should be run at higher temperature. The column back pressure was 291
flow rates (greater than 2-mL/min). Note the com- bar at 1 mL/min and 550 bar at 2 mL/min. Thus,
parative chromatograms of Figure 5 where the higher pressure capability for the HPLC instru-
flow rate was increased from 1-mL/min to 2-mL/min mentation, such as can be achieved with the Agi-
resulting in a more rapid separation and better lent 1200 Rapid Resolution system, is useful. In
resolution since the linear velocity was closer to addition, Agilent's proprietary engineered particle
the optimum and column efficiency was greater. Of size distribution, unique to the RRHT columns,
course, since the column pressure is proportional generates lower system pressure than would be
to the inverse of particle diameter squared, at expected for a 1.8-µm particle.
Flow: 1 mL/min
P = 291 bar
1
1 2 3 4 5 6 7 min
Flow: 2 mL/min
P = 551 bar
1 Caffeine (ISTD) Columns: Eclipse XDB-C18, 4.6 × 50 mm, 1.8 um
2 Allobarbital Mobile phase: 60% A, 40% B
3 Phenobarbital A = pH 4.5 Na Acetate
4 Butalbital B = MeOH
5 Hexobarbital Injection volume: 2 µL
R2,3 = 4.66 R4,5 = 3.84
Detector: UV 254 nm
Flow cell: 2 µL, 3 mm flow path
3.5 min
1 2 3
5
www.agilent.com/chem
Conclusions
In this study, an established HPLC method for USP
Phenobarbital was sequentially and quickly
improved to a high throughput method using RR
and RRHT techniques. New RRHT column technol-
ogy is an especially powerful tool for improving lab
productivity by reducing HPLC analysis time dra-
matically. Starting with a proven rugged method,
conversion to a high throughput method can be
achieved simply by replacing the original analytical
-sized column with either a RR or RRHT column.
The reasons for this direct conversion include the
similar selectivity of smaller particles (1.8- and
3.5-µm) compared to larger 5 µm particles, and an
engineered particle size distribution that reduces
unacceptably high system pressures that may be
experienced with other sub-two micron packings.
Additional method development techniques worth
considering are use of even faster flow rates, ele-
vated temperature and different bonded phases.
The new Agilent 1200 series HPLC system is
designed especially for using ZORBAX RRHT
columns to take advantage of higher flow rates and
elevated temperatures.
References
1. http://www.dea.gov/concern/barbiturates.html
2. http://www.elmhurst.edu/~chm/vchembook/
6673barbit.html
3. USP27-NF22 Page 1461; U.S.
Pharmacopeia/National Formulary, 29(6), 2004.
4. J. W. Henderson, Jr., “Plug & Play” Fast and
Ultra-Fast Separations Using 3.5-µm Rapid
Resolution and 1.8-µm Rapid Resolution High
Throughput Columns, Agilent Technologies,
publication 5989-2908EN,
www.agilent.com/chem
Agilent shall not be liable for errors contained herein or for incidental or consequential
damages in connection with the furnishing, performance, or use of this material.