Professional Documents
Culture Documents
Interference-Free Flame Photometry of Calcium in Serum and Urine
Interference-Free Flame Photometry of Calcium in Serum and Urine
Data are given for the rapid determination of calcium in biologic material by flame
photometry. An “automatic background-subtraction,” photometer added to the
available instrumentation eliminated cation interference, while the use of a chelating
agent (EDTA) prevented interference from anions (chiefly phosphate). The pro-
cedure shows excellent precision and accuracy in comparison with standard chemical
procedures and in the analysis of control sera.
From the Clinical Chemistry Laboratories of the Department of Biochemistry, Duke 15111-
versity Medical Center, Durham, N. C., and the Peter Bent Brigham Hospital, Boston, Mass.
The work reported here was supported in part by an Institutional Grant to Duke University
Medical Center by the American Cancer Society.
Received for publication Jan. 3, 1961.
35
36 THIERS & HVIID Clinical Chemistry
Method
Reagents and Apparatus
ROWLAND CIRCLC
Procedure
Samples of serum or urine are diluted 50-fold with 0.15M EDTA and
aspirated into the flame of the background-correcting photometer. The
reading of the microammeter on the photometer is directly propor-
tional to the concentration of calcium in the sample, provided that the
instrument has been adjusted by the use of water and background cor-
rection solutions, as described in the following. With the instrument
employed in this work, the slope of the calibration curve is about 50
tAmp./mEq./L. of Ca + in the original undiluted sample.
All of the reagents from the list above are not required for this pro-
cedure. Only EDTA, background correction solution, and calibrated
calcium standards are used routinely. The other reagents were em-
ployed in the experimental verification of t.he method, as described be-
low. Stability of the instrument is such that the adjustment of back-
groulld correction and the calibration curve need to be checked only
once each time the flame is turned on.
Aa+bI (1)
Vol. 8, No. I, 1962 FLAME PHOTOMETRY OF CALCIUM 39
lL+il_af+B+bL-1-BIL+n (2)
and
An=an+bnln (3)
;b b*
-J
z
(()
Fig. 2. Adjustment of photo
multiplier tube response curves
for automatic background cor-
rection.
I
0
Ui
U.
-J
0
12
LIGHT INTENSITY, I
40 THIERS & HVIID Clinical Chemistr,
rents on the output meter of the instrument allows very precise ad just-
ment of their superposition, which simultaneously compensates for
all linear variables in the method.
Practice
aL+B_aB (4)
and
(5)
Signa1 (VAmp.)
Coneentrations of ealiono*
(mEq./L.) I II Iti
(L+B) (5) (L+B-
Sampte Ca Atg Na K BL)
1. Water 0 0 0 0 0 0 0
2. Background correction solution 0 0.050 3.0 0.10 290 290 0
3. Calcium alone 0.100 0 0 0 265 100 165
4. Metals of serum 0.100 0.050 3.0 0.10 470 305 165
8During preparation of this report, work by Herrman and Rick on the application of the
ammonium salt of EDTA to flame photometry of calcium, without background correction
came to our attention.
42 THIERS & HVIID Clinical Chemistry
300
0 0.2
ter (or isotonic NH4C1) is used as the diluent, and rises with increasing
concentrations of EDTA until it reaches a plateau between 0.15 M and
0.2 M. This phenomenon was observed with all samples of serum or
urine tested.
The accuracy of the prevention of anion interference and of auto-
matic background subtraction was tested directly by comparing the
intensity of the calcium line from two simulated urine solutions of
equal calcium concentration, one of which contained EDTA and phos-
phate, the other of which did not. As shown in Table 2, the presence of
phosphate depressed the calcium signal by a large factor, but the addi-
tion of EDTA prevented any decrease whatsoever. Neither the EDTA,
nor the ]IDTA plus phosphate, showed any signal. Thus, the accuracy
and effectiveness of simultaneous correction technics for both cation
and anion interferences on the flame photometry of calcium in biologi-
cal samples, is clearly demonstrated.
Precision
A sample of pooled serum was divided into 18 aliquots, and each
aliquot was diluted separately for analysis by the procedure described.
Vol. 8, No. I, 1962 FLAME PHOTOMETRY OF CALCIUM 43
Table 2. COMPARISON OF CALCIUM LINE OF SIMULATED URINE SOLUTION WITH AND WITHOUT
1 150 0 0 0 0 0 0
2 150 1.0 0 0 0 0 0
3 0 1.0 0.130 5.8 0.87 95 95
4 0 0 0.130 5.8 0.87 255 260
5 150 1.0 0.130 5.8 0.87 260 255
A sample of urine was treated in the same manner. The average read-
ing for the serum samples was 210 VAmp. (corresponding to 4.2
mEq./L.) and, for the urine samples, was 95 jiAmp. (1.9 mEq./L.).
The over-all range of variation of the readings obtained was ±5 pAmp.
in both cases. Since the meter could be read only to the nearest 5
VAmp., the precision can best be expressed as a range of ±5 1LAmp. or
±0.1 mEq./L. (13). Calculation of average or standard deviation
would be misleading in this case, where the data are not distributed ac-
cording to the Gaussian or “normal” relationship, and, in fact, would
give a result far less than the smallest readable meter division. The
precision of all measurements subsequently undertaken proved to be
the same, ±0.1 mEq./L., and was limited by the readability of the
meter.
Accuracy
more than the accuracy of the over-all method. They confirm for real
serum what the data of Table 2 checked for simulated urine, that is, the
accuracy of the background subtraction technic applied to a complex
system.
In addition, blind analyses were performed on 10 randomly selected
sera from the routine load of the clinical chemistry laboratory over a
period of 2 weeks. These sera had been analyzed by a standard method
of calcium determination (1) which involves a precipitation of calcium
oxalate and a titration. The calcium in these samples ranged in con-
centration from 2.9 to 7.0 mEq./L. The average difference between
the two methods was 0.01 mEq./L., and no difference greater than
±0.1 mEq./L. was observed.
A known sample of urine, comparable to Versatol for serum, is not
available, and the standard methods for determining calcium are far
less reliable when applied to urine than to serum. While this situation
is tolerable in clinical chemistry, because clinically meaningful altera-
tions in calcium concentration are relatively much larger in urine than
in serum, it precludes a direct check on the accuracy of the flame meth-
od with real urine. The results of recovery experiments that were per-
formed (Table 4) are self-consistent and show the same precision as
was observed with serum samples. Accuracy is demonstrated with
simulated urine (Table 2). It is, therefore reasonable to presume that
the same high accuracy prevails with urine as with serum.
Discussion
The precision of this method and its accuracy can both be expressed
as a range of ±0.1 mEq./L. (13). For normal serum containing 5.0
mEq./L. this is a range of ±2%, which makes the method excellent for
clinical use.
Table 4. RECOVERY OF CALCIUM ADDED TO URINE SAMPLES
Duplicate readings.
Vol. 8, No. I, 1962 FLAME PHOTOMETRY OF CALCIUM 45
References
1. Reiner, M., Ed., Standard Method.o of Clinical Chemistry Vol. 1, Acad. Press, New York
(]953, p. 16.
46 THIERS & HVIID Clinical Chemistry