Ensayo de Tension Probeta redonda fluencia Report No. 4884


Test Date 09-Aug-18 Testing Machine Smart Series

CLIENTE grupo 11 a 12
LABORATORISTA Luis B

Longitud Calibrada

Load Cell S/N (H773), Units (LBS ) 224810 Crosshead Speed ( mm / min ) or Rate 30
Displacement Sensor XHD100 ( XHD100 )

Diametro Fluencia Resistencia Elongacion


Test No Specimen ID (mm) (OFFSET) (MPa) (%)
(MPa)

10825 5 9.18 467.36 697.599


10826 10 9.18 439.06 733.556
10827 15 9.18 303.50 717.652

Mean 9.18 403.30 716.269


Median 9.18 439.06 717.652
Std Dev 0.00 87.59 18.019
Maximum 9.18 467.36 733.556
Minimum 9.18 303.50 697.599
Range 0.00 163.86 35.957

By:________________________ Date:_______________

Test Performed in Accordance with ASTM E-8, Rev. 88