Professional Documents
Culture Documents
Axios MAX
Axios MAX
Designed to perform
ZETA Technology
SST-mAX stability in measuring boron
ZETA Technology fully eliminates
102
tube aging and ensures stable
100
tube output during the lifetime of
Rel. Intensity (%)
Flexible Automation
Sample handling with AxiosmAX is AxiosmAX is easily integrated in
highly flexible. A single batch queue automated laboratory systems. Sample
can contain a wide variety of sample loading access is possible from either
types and sizes and high-priority side of the instrument with a specifically
measurements can be added easily to an designed sample inverter or from the
ongoing batch. rear with a backfeed unit.
The features making this possible are: Easy and pleasant to work with
• Integrated sample changer The AxiosmAX system is extremely quiet
• Priority sample because special attention has been paid
position with to reduce the noise level generation
sample detection by fans, water pumps and generators.
allowing Furthermore, it dissipates little heat
QuickStart into the room. Its small footprint and
• Programmable mobility make it easy to maneuver and
collimator masks position in confined laboratory spaces.
(6 – 37 mm)
Serviceability Quality
AxiosmAX is designed with AxiosmAX is designed, built and
serviceability in mind. Components tested according to the ISO9001
on special service racks make regular quality standards. PANalytical is
service tasks easy. For example, certified according to ISO14001
exchanging the internal cooling water for environmental management.
or topping up the oil level of the In this way PANalytical accepts its
vacuum pump. responsibility to strive for continuous
Furthermore, improvement of its performance
accessibility from towards the environment with respect
all sides allows to its processes and products.
unrestricted
access for
maintenance,
speeding up
servicing and
minimizing
downtime.
AxiosmAX technical specifications
Sample handling Optical path (continued)
Types Solid, fused beads, filters, pressed/loose powders Primary collimators 3 max.: 100, 150, 300, 550, 700 or 4000 µm
and liquids Primary beamfilters 4 max.: One may be used as beam stop
Handling All samples are fitted in sample holders during Crystals 8 max.: LiF420, LiF220, LiF200, Ge111 (flat/
measurement. Liquids and loose powders in curved), PE002 (flat/curved), InSb (flat/curved),
disposable (P2) liquid cells, which are fitted in TlAP coated, PX1, PX4, PX5, PX6, PX7, PX10
liquid sample holders
Detectors Flow, scintillation, sealed Xe in tandem with flow
Dimensions 51.5 mm Ø x 40 mm height, maximum counter as duplex counter (optional), Hi-Per Scint
Weight Max. 400 g including sample holder (optional)
Sample changer X-Y changer with priority position for 1 sample Hi-Per channels Maximum of 2: for B to Mg
(detected) and 2 empty holders with capacity Beam path Vacuum: <10 Pa; He, N2 optional
for 8 trays for 8 sample holders and 1 tray for 4
Counting electronics
sample holders.
Optional: High-capacity changer up to 209 Type Dual multi channel analyser with digital signal
uncupped samples of 32 mm or 140 uncupped processor
samples of 40 mm diameter Maximum count rate Flow counter: 3000 kcps
Changer trays for: 8 sample holders, 8 flex positions, 8 steel Scintillation and sealed counter: 1500 kcps
rings (Ø 51.5 mm), 21 samples (Ø 25 mm), Duplex counter: 4500 kcps
12 samples (Ø 32 mm), 10 samples (Ø 41 mm), or Hi-Per Scint: 3500 kcps
4 sample holders (SPC/monitor tray) (non-linearity ≤ 1 %)
Loading Air lock with programmable pumping time, one Pulse shift correction Automatic, dynamic (selectable)
or two position turret mechanism, sample-surface Dead time correction Automatic
down, direct loading of unmounted samples in Automation interfacing
holder in loading position
Inverter
Spinner 0.5 rev/s Transfer point at side 41.1 cm (16.18 in) from front side, protruding
X-ray tube SST-mAX 28.95 cm (11.4 in)
Type Super Sharp End Window tube featuring ZETA Transfer point heights 85, 88.5, 91.2, 111.2 cm
Technology (33.46, 34.84, 35.91, 43.78 in)
Although diligent care has been used to ensure that the information herein is accurate, nothing contained herein can be construed to imply any
further notice.Please contact us for the latest version of this document or further information. © PANalytical B.V. [2009]. 9498 707 29713 PN7213
representation or warrantee as to the accuracy, currency or completeness of this information. The content hereof is subject to change without
Stability 0.00006 % / 1 % mains variation Electrical Installation category class II
Stability mains ± 10 % Protection Pollution degree 2 Class IP40
Rating Installation
AxiosmAX 2.4 kW 20 - 60 kV, 10 - 100 mA Dimensions 84 x 98 x 151 cm (33.1 x 38.6 x 59.4 in) (WxDxH)
AxiosmAX 3.0 kW 20 - 60 kV, 10 - 125 mA Floor space 134 (25 + 84 + 25) x 223 (25 + 98 + 100) cm2
AxiosmAX 4.0 kW 20 - 60 kV, 10 - 160 mA (WxD, in use) (53 x 88 in2)
Goniometer Floor space Min. 2 x 2 m2 (6.6 x 6.6 ft2), i.e. 1 m (39 in) extra
Type θ/2θ decoupled with Direct Optical Position (WxD, maintenance) on service side
Sensing (DOPS) Weight 550 kg
Angular accuracy 0.0025° θ and 2θ Power consumption 5.5 kW, 7.5 kVA, 33 A (@ 230 V, 50 Hz)
Angular 0.0001° θ and 2θ Mains requirements Single phase, 50-60 Hz, 188-253 V, 38 A max.
reproducibility
Environment temp. 10° to 35° C (50° to 95° F)
Scanning speed Up to 10° 2θ /s Cooling water 12° to 20° C (54° to 68° F), 3.5 to 8 bar,
Slewing speed 40° 2θ/s 5 to 11 l/min
Temp. stabilization ± 0.05°C at 31° cabinet temperature Compressed air 4 to 5 bar
Optical path
Channels masks Single mask (fixed 27, 30 or 37 mm),
switchable 3 positions (27, 30, 37 mm) or
6 positions (6, 10, 20, 27, 30, 37 mm)
Asia Pacific
T +65 6741 2868
F +65 6741 2166