Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

ORIGINAL ARTICLE

Miniscrew stability evaluated with computerized


tomography scanning
Jung-Yul Cha,a Jae-Kyoung Kil,b Tae-Min Yoon,c and Chung-Ju Hwangd
Seoul, Korea

Introduction: In this study, we aimed to determine the effect of bone mineral density (BMD), cortical bone
thickness (CBT), screw position, and screw design on the stability of miniscrews. Methods: Ninety-six mini-
screws of both cylindrical and tapered types were placed in 6 beagle dogs. The BMD and CBT were measured
by computerized tomography and correlated with the placement and removal torque and mobility. A regres-
sion equation to predict the placement torque was calculated based on BMD, CBT, screw type, and screw
position. Results: The placement torque showed a positive correlation in the order of removal torque
(0.66), BMD of the cortical bone (0.58), and CBT (0.48). Placement and removal torque values were signifi-
cantly higher in the mandible compared with the maxilla. Tapered miniscrews had higher placement torque
than did the cylindrical type (P \0.001). However, the removal torque was similar in both groups. Placement
torque was affected by screw position, screw type, and BMD of cortical bone, in that order. Conclusions:
BMD of cortical bone, screw type, and screw position significantly influence the primary stability of mini-
screws. (Am J Orthod Dentofacial Orthop 2010;137:73-9)

quantity and quality.4,5 The initial stability of minis-

E
asy placement is an advantage of orthodontic
miniscrews. In addition, the tooth can be moved crews is considered essential in clinical use because of
without the patient’s cooperation; this enhances immediate or early loading in many patients.6
treatment efficiency. Some success has been reported Two factors affect the initial stability of a screw: the
regarding the use of miniscrews as orthodontic anchor- screw factor and the host factor. The screw factor is
ages. However, the clinical application of a miniscrew related to the characteristics of the screw design, includ-
does not guarantee treatment success, and its stability ing diameter and length.7 Various screw designs have
is essential before it can be used as orthodontic been introduced to enhance initial stability.8 The goal
anchorage.1-3 is to increase initial fixation by inducing controlled
It is important to ensure the initial stability of an compressive forces in the cortical bone layer. In the
orthodontic miniscrew because most failures occur prosthodontic field, it was reported that a tapered shape
during the initial stage. Furthermore, stability of the can enhance stability for immediate loading by increas-
miniscrew in the initial stage reduces its micromove- ing the mechanical contact between the dental implant
ment, thus allowing for an appropriate environment and the surrounding bone.9
that supports the healing process surrounding the bone. The host factor is related to the quantity and quality
The success of an implant or miniscrew is determined of the bone where the screw is placed. Cortical bone
by the patient’s general condition, the biocompatibility thickness (CBT) (quantity) can affect the initial stability
of the materials, the placement procedure, and bone of a screw.10 Finite element analysis showed that, when
a lateral force is applied to the screw, most of the force is
a
Assistant professor, Department of Orthodontics, College of Dentistry, Yonsei concentrated on the cortical bone.11,12 For this reason,
University, Seoul, Korea. the previous studies focused on the anatomic back-
b
Private practice, Seoul, Korea.
c
Postgraduate student, Department of Orthodontics, College of Dentistry, ground related to CBT rather than bone quality, to
Yonsei University, Seoul, Korea. improve the stability of the miniscrew.
d
Professor, Department of Orthodontics, Dental Science Research Institute, The On the other hand, several studies have used com-
Institute of Craniofacial Deformities, College of Dentistry, Yonsei University,
Seoul, Korea. puted tomography (CT) in prosthodontics and orthope-
The authors report no commercial, financial, or proprietary interest in the prod- dics to study factors related to bone density.13-16 It was
ucts or companies described in this article. reported that bone density differs according to sex, age,
Reprint requests to: Chung-Ju Hwang, Department of Orthodontics, College of
Dentistry, Yonsei University, 134 Shinchon-dong, Seodaemun-gu, Seoul 120- and physical condition. CT was recently suggested to
752, Korea:; e-mail, hwang@yuhs.ac. successfully measure bone mineral density (BMD) in
Submitted, August 2007; revised and accepted, March 2008. orthopedics. BMD has been used as a parameter to
0889-5406/$36.00
Copyright Ó 2010 by the American Association of Orthodontists. establish a treatment plan to ensure the stability of
doi:10.1016/j.ajodo.2008.03.024 implants in dentistry.17
73
74 Cha et al American Journal of Orthodontics and Dentofacial Orthopedics
January 2010

There are 2 methods for measuring BMD. One is trated into the placement area. Before placement,
with Hounsfield units (HU), and the other is to measure a 5-to-10 mm gingival incision was made under saline-
BMD indirectly by using a hydroxyapatite block. Since solution irrigation, and complete placement of the screw
Hounsfield units can be affected by the loaded voltage into the alveolar bone was confirmed.
of x-rays and the protocol used, bone density analysis Screw placement was performed manually with 70
with a BMD calibration standard (hydroxyapatite to 90 angulations to gingival surface in consideration
block) has greater advantages for evaluating bone qual- of the buccolingual width of alveolar bone for each
ity more accurately.18 experiment. In all miniscrews, a force of 250 to 300 g
One assessment method of initial stability is to was applied with an elastomeric chain engaged recipro-
measure the torque during placement.9,13,19 Cheng et cally from a cylindrical miniscrew to a tapered mini-
al19 proposed that placement torque could be used as screw immediately after the placement of miniscrews,
an indicator for initial stability, and, to obtain initial sta- and the elastic chain was exchanged every 3 weeks.
bility, a certain amount of torque is necessary. Place- Screw mobility was measured twice on each mini-
ment torque is the measurement of the resistance at screw by using a periotest (Simens AG, Bensheim, Ger-
the screw-bone interface; it reflects the level of bone many) before removing the screw. The sleeve of the
deformational strain caused by the miniscrew. handpiece of the periotest was positioned with 1 to 2
In this study, we hypothesized that BMD values can mm from the screw head perpendicularly after an elas-
be used to predict the mechanical stability of minis- tomeric chain was removed. The average of 2 measure-
crews by evaluating placement torque. Our aim was to ments for a miniscrew was recorded as the mobility
determine the effect of BMD, CBT, screw position, value. During the test period, a chlorhexidine solution
and screw design on the stability of miniscrews. was applied daily to maintain the animals’ oral hygiene
(Fig 3).
The placement torque was the highest (in newtons
MATERIAL AND METHODS per square centimeter) when the miniscrew was placed
For this study, miniscrews were placed in 6 beagle completely into the bone. The highest removal torque
dogs (age, 1 year; weight, 12 kg). Their purchase, was measured during a quarter initial turn by using a tor-
selection, and management, and the experimental pro- que sensor (MGT50, Mark-10 Co, New York, NY).
cedures were carried out according to prescribed condi- To measure the BMD of the implanted area, CT
tions of the institutional review board, the Animal scanning was done 3 days before placing and removing
Experiment Committee of Yonsei Hospital, Seoul, Ko- the miniscrews. Before scanning, each dog was placed
rea. A nondrilling type of miniscrew, 1.4 mm in diam- under intramuscular sedation. The gantry of the CT
eter and 7 mm in length, was used. Both the cylindrical device was placed parallel to the occlusal plane, and
type (OAS-1507C) and tapered type (OAS-1507 T, the dog’s head was fixed with a strap. The CT scan
Biomaterials Korea, Seoul, Korea) were selected; 96 was performed with a high-speed advantage CT scanner
screws were used (Fig 1). (GE Medical System, Milwaukee, Wis) by using
The location of the orthodontic miniscrew place- standard CT protocol (high-resolution algorithm,
ment was determined after evaluating the reconstructed 512 3 512 matrix, 120 Kv, 200 mA) at a table feed of
CT image, taken before placement, to determine 6 mm per second. The CT data were reconstructed to
whether there was sufficient interdental space. Sites se- 1-mm thick transaxial images.
lected were between the roots of the second, third, and A calibration standard (Dental Phantom, Image
fourth premolars, and the first molar in the mandible, Analysis, Columbia, Ky) was applied to calibrate the
between the roots of the second and third premolars, level in Hounsfield units in the dogs during the CT scan-
the first molar, and between the second and third premo- ning. The calibration standard containing 3 compart-
lars in the maxilla (Fig 2). ments with 0, 75, and 150 mg of hydroxyapatite per
The animals were injected subcutaneously with 0.05 cubic centimeter was attached to the phantom for
mg per kilogram of atropine followed by an intravenous BMD calibration. Calibration was performed by mea-
injection of rompun, 2 mg per kilogram, and ketamine, suring the Hounsfield unit values in the 3 compartments
10 mg per kilogram, to induce general anesthesia. The of the dental phantom and relating these values to deter-
anesthesia was maintained with 2% enflurane, and mine the BMD with a linear equation. The calibrations
each animal’s temperature was maintained with a heat- were performed for each dog.
ing pad and an electrocardiogram, and monitored. The scanned images were analyzed by V-implant
When placing the mini-implant, 2% hydrochloric acid (CyberMed, Seoul, Korea). The CT data taken before re-
lidocaine containing 1:100,000 epinephrine was infil- moving the screws were reconstructed to a transaxial
American Journal of Orthodontics and Dentofacial Orthopedics Cha et al 75
Volume 137, Number 1

Fig 1. Features and measurements of the miniscrews tested (mm).

was based on a linear relationship: BMD 5 a HU 1


b, where a and b are calibration coefficients.
Pearson correlation analysis was used to determine
the association between placement torque, removal tor-
que, mobility, BMD, and CBT. The categorical vari-
ables (screw type and position) were statistically
evaluated by using independent t tests to determine
any differences in placement torque. The results suggest
that the variables were linearly correlated. Thus, a mul-
tiple regression model was used with placement torque
as the dependent variable, and CBT, BMD, screw type,
and screw position as the independent variables.

RESULTS
Fig 2. Schematic diagram indicating the approximate
A regression equation was obtained by using the av-
locations of screws. Circles indicate screw heads, and
erage Hounsfield value (Table I, Fig 5) for the 3 parts in
orthodontic forces were loaded reciprocally between
a cylindrical miniscrew (CS) and a tapered miniscrew the standard calibration of each beagle. The parameter
(TS). a (inclination) was calculated within the range of
0.813 to 0.935, and the intercept b ranged from 3.80
to 24.47. The correlation coefficient of all regression
image at 1-mm intervals for each occlusal plane of the equations was .0.98.
maxilla and mandible (Fig 4, A). The vertical location A correlation was observed between placement and
and inclination of each screw was localized by using a re- removal torque, mobility, screw type, and BMD corti-
constructed image at the coronal plane from the postop- cal. The mobility showed a negative correlation with
erative scans. This information was transferred to the placement torque (–0.577) and a positive correlation
preoperative images. From the preoperative scanning with BMD cortical (0.575) (Table II).
data, the average BMD of a cylindrical area (diameter, The values of placement and removal torque, CBT,
2 mm; depth, 5 mm) for each screw site was measured and BMD were significantly higher in the mandible
as the BMD total by Hounsfield unit calibration. The compared with the maxilla (P \0.05). The tapered
CBT and average bone mineral density in the cortical screw (14.57 N per square centimeter) had higher place-
area (BMD cortical) were also calculated (Fig 4, B). ment torque than the cylindrical screw (9.69 N per
square centimeter); this was statistically significant in
the mandible (P \0.001) (Table II). There were no sig-
Statistical analysis nificant differences in the removal torque between the
To determine the relationship between Hounsfield cylindrical and tapered screws. The values of CBT and
units and BMD values, a calibration regression equation BMD were similar in the tapered and cylindrical screw
was derived individually for all experimental dogs and groups (Table III).
76 Cha et al American Journal of Orthodontics and Dentofacial Orthopedics
January 2010

Fig 3. Experimental protocol used in this study.

BMD cortical had significant influences on placement


torque (R2 5 0.801) (Table IV).

DISCUSSION
Recently in orthodontics, the use of miniscrews has
been generalized for the preparation of orthodontic
anchorage in clinical patients. However, studies on the
stability of orthodontic miniscrews are still in their early
stages compared with studies of prosthodontic implants.
Factors that can affect a miniscrew’s stability are
design,20 placement procedures,21,22 and quantity and
quality of the bone related to the host factor.5,23 This
study focused on the assumptions that the quantity
(thickness) and quality (density) of the bone have signif-
icant effects on a miniscrew’s stability.
Information on bone density can be obtained from
the measurement of the Hounsfield units, but there are
issues regarding the reliability of many CT scanners,
since CT values are affected by changes in the effective
energy of the x-ray source, the so-called beam-harden-
ing effect.18 To calibrate Hounsfield unit values, refer-
ence calibration standards for these CT values are
needed for each subject. Calibration coefficients had
a range of 0.813 to 0.935 in this study. The same
1000-HU value for 2 subjects can have a significant
discrepancy of about 120 BMD units. Even though 2
subjects might have the same Hounsfield unit value,
Fig 4. Procedure of BMD measurement in the region of they can have different BMD values, and calibration
interest: A, position of calibration phantom and recon-
coefficients must be used to calculate the BMD values.
structed image according to the occlusal plane; B, mea-
We therefore used a hydroxyapatite phantom to provide
surement of Hounsfield units for the region of interest
(red box). accurate calibration.
Regarding the relationship between BMD and
placement torque, there was a positive correlation in
Multiple regression analysis showed that, among all the cortical part of the alveolar bone but no similar
the variables (CBT, BMD cortical, BMD total, screw correlation in the total bone. Previous studies15,17
type, screw position), screw position, screw type, and of prosthodontic implants reported high correlation
American Journal of Orthodontics and Dentofacial Orthopedics Cha et al 77
Volume 137, Number 1

Table I. Calibration of BMD by using Hounsfield units (HU) in the 3 compartments of the dental phantom
Calcium hydroxyapatite
3
0 mg/cm 75 mg/cm3 150 mg/cm3

HU HU HU Calibration coefficients

Dog Mean SD Mean SD Mean SD Parameter a Parameter b Correlation coefficient

1 4.2 0.8 81.6 0.7 168.7 1.1 0.868 3.80 0.99


2 6.4 6.2 79.4 2.6 176.9 6.3 0.813 7.24 0.99
3 3.9 1.8 82.3 3.0 178.8 3.4 0.819 4.76 0.99
4 10.8 5.3 74.1 2.1 169.8 4.0 0.827 10.71 0.99
5 7.6 2.0 53.0 4.9 149.8 1.4 0.935 14.16 0.99
6 21.0 5.7 47.8 2.3 160.4 1.4 0.810 24.47 0.98

Table II. Pearson correlation test and independent t test


for placement torque
Variable Correlation coefficient P value

Removal torque .661 0.001


Mobility .577 0.001
CBT .476 0.001
BMD cortical .575 0.001
BMD total .078 0.504

Mean 6 SD

Screw type
Cylindrical 9.69 6 5.32 0.001
Tapered 14.57 6 8.13
Screw
Maxilla 6.06 6 3.03 0.001
position
Mandible 17.84 6 5.03

Pearson correlation test was done between removal torque, mobility,


CBT, and BMD, and the independent t test was performed for screw
type and screw position.
Fig 5. Graph depicting calibration of BMD values from
Hounsfield units for dog 1. The linear regression equa- placed near cancellous bone, and the quality of cancel-
tion is shown on the graph, and the values are listed in lous bone is important in the stability of the implant.27
Table I. mg HA/cm3 equal milligrams (0, 75, 150) of
However, our results indicated a poorer correlation be-
calcium hydroxyapatite per cubic centimeter.
tween total bone density and placement torque com-
pared with cortical bone density.
It was observed that placement torque is related to
between BMD and placement torque14 and significant screw position. This is due to thicker cortical bone
correlation between BMD and pull-out strength.15,22,24 and better bone quality in the mandible compared
Seebeck et al25 reported that in an axial pullout strength with the maxilla.27 Despite the implant position, the sta-
and cantilever bending test to the implanted screw, the bility of a prosthodontic implant is more influenced by
screw’s stability was mostly related to the CBT and can- its length and diameter in alveolar bone, rather than
cellous bone density. It is also known that the strength the thickness of cortical bone.27 However, the stability
and stiffness of trabecular bone are in linear proportion of orthodontics miniscrews, with their relatively smaller
to bone density on the axial and bending loads.26 This diameter and length, is related to CBT. This study
demonstrates that the density of trabecular bone and showed a significant correlation between placement
cortical bone makes a significant contribution to the sta- torque and cortical bone density.
bility of an implant. In a prosthodontic implant, most of Placement and removal torque also showed a posi-
the implant, either in the maxilla or the mandible, is tive correlation, which supports previously reported
78 Cha et al American Journal of Orthodontics and Dentofacial Orthopedics
January 2010

Table III. Comparison of variables according to screw location and screw type
Screw type Cylindrical Tapered

Variable Position Mean SD Mean SD P value

Removal torque (Ncm) Maxilla 2.29 1.24 1.83 0.81 0.480


Mandible 4.83 2.07 5.89 4.59 0.337
P value 0.001 0.001
Mobility (PTV) Maxilla 3.21 4.41 1.61 3.94 0.203
Mandible 0.12 2.77 2.94 4.03 0.012
P value 0.073 0.014
CBT (mm) Maxilla 1.38 0.6 1.42 0.51 0.510
Mandible 1.98 0.48 2.02 0.33 0.403
P value 0.001 0.001
BMD cortical (mg Maxilla 867.46 242.99 866.46 196.2 0.988
of hydroxyapatite/cm3) Mandible 1030.17 167.81 1049.51 202.3 0.722
P value 0.01 0.004
BMD total (mg Maxilla 579.41 197.73 583.26 172.38 0.945
of hydroxyapatite/cm3) Mandible 697.32 149.34 787.91 191.24 0.076
P value 0.025 0.001

Table IV.Regression equation model to predict place- was found in removal torque. This suggested that the
ment torque with CBT, BMD cortical, BMD total, screw strain generated between the screw and the bone after
type, and screw position placement decreased during healing of the surrounding
bone. Further research involving histologic analysis is
Unstandardized
coefficient Standardized required to confirm this.
coefficient A regression equation between placement torque,
Variable Beta SE Beta t P value BMD, and screw type was derived. Screw type and
CBT 0.053 0.879 0.004 0.060 0.952 position were major variables for the prediction of
BMD cortical 0.005 0.002 0.159 2.781 0.007 placement torque. This results from the difference in
BMD total 0.003 0.002 0.068 1.092 0.278 bone density and structure in the cortical bones of the
Screw type 4.692 0.680 0.326 6.897 0.000 maxilla and the mandible. The maxilla consists of
Screw position 10.428 0.922 0.725 11.310 0.000
more trabecular bone than the mandible, and its cortical
Corrected R2 0.801
bone is relatively thinner. The BMD cortical was also
Dependent value was placement torque (Ncm). chosen as a significant variable, but the impact of the
variables on the regression equation was smaller than
those of screw type and position. These results suggest
studies, and they also had a correlation with the BMD of that it might be difficult to predict the primary stability
cortical bone.28 However, the BMD of the total bone did of a miniscrew solely with BMD values, and screw
not have a major effect on placement and removal torque. design could play a major role in increasing the place-
On the other hand, mobility was associated with the BMD ment torque despite the quantity of bone.
of the total bone and CBT. Therefore, this suggests that Because of the high radiation dose during CT
BMD and CBT could affect a miniscrew’s stability. scanning, there are limitations in using it for clinical
Tapered miniscrews showed 65% higher placement purposes. However, dental CT such as cone-beam CT
torque than the cylindrical type. This coincides with the can be a useful technique for a noninvasive assessment
result from a previous study that used artificial bone.8 of BMD in estimating the stability of a miniscrew at the
As the tapered screw was placed, the torque along with preoperative stage.
the deformation of the surrounding bone increased with
increasing diameter. In this study, a similar bone density
was calculated at the placement area of each screw CONCLUSIONS
type. This meant that screw type affected the initial The BMD of the cortical bone, screw position, and
stability of the miniscrew because it was placed into screw type have a compound influence on the primary
some areas with low bone density or thinner cortical bone. stability of a miniscrew. Cone-beam CT might be a use-
A significant difference in placement torque was ful and noninvasive way to assess the BMD for estimat-
found according to the screw type, but no difference ing the stability of a miniscrew at the preoperative stage.
American Journal of Orthodontics and Dentofacial Orthopedics Cha et al 79
Volume 137, Number 1

REFERENCES dental quantitative CT prior to dental implant placement in


1. Lee SY, Cha JY, Yoon TM, Park YC. The effect of loading time on cadaver mandibles: pilot study. Radiology 2002;224:247-52.
the stability of mini-implant. Korean J Orthod 2008;38:149-58. 16. Carano A, Melsen B. Implants in orthodontics. Interview. Prog
2. Park YC, Chu JH, Choi YJ, Choi NC. Extraction space closure Orthod 2005;6:62-9.
with vacuum-formed splints and miniscrew anchorage. J Clin 17. Beer A, Gahleitner A, Holm A, Tschabitscher M, Homolka P. Cor-
Orthod 2005;39:76-9. relation of insertion torques with bone mineral density from dental
3. Im DH, Kim YS, Cho MA, Kim KS, Yang SE. Interdisciplinary quantitative CT in the mandible. Clin Oral Implants Res 2003;14:
treatment of Class III malocclusion using mini-implant: problem- 616-20.
oriented orthodontic treatment. Korean J Orthod 2007;37:305-14. 18. Cann CE. Quantitative CT for determination of bone mineral
4. Motoyoshi M, Matsuoka M, Shimizu N. Application of orthodon- density: a review. Radiology 1988;166:509-22.
tic mini-implants in adolescents. Int J Oral Maxillofac Surg 2007; 19. Cheng SJ, Tseng IY, Lee JJ, Kok SH. A prospective study of the
36:695-9. risk factors associated with failure of mini-implants used for or-
5. Park HS, Jeong SH, Kwon OW. Factors affecting the clinical thodontic anchorage. Int J Oral Maxillofac Implants 2004;19:
success of screw implants used as orthodontic anchorage. Am J 100-6.
Orthod Dentofacial Orthop 2006;130:18-25. 20. Kido H, Schulz EE, Kumar A, Lozada J, Saha S. Implant diameter
6. Melsen B, Costa A. Immediate loading of implants used for ortho- and bone density: effect on initial stability and pull-out resistance.
dontic anchorage. Clin Orthod Res 2000;3:23-8. J Oral Implantol 1997;23:163-9.
7. Ivanoff CJ, Sennerby L, Johansson C, Rangert B, Lekholm U. 21. Sowden D, Schmitz JP. AO self-drilling and self-tapping screws in
Influence of implant diameters on the integration of screw rat calvarial bone: an ultrastructural study of the implant interface.
implants. An experimental study in rabbits. Int J Oral Maxillofac J Oral Maxillofac Surg 2002;60:294-9.
Surg 1997;26:141-8. 22. Hitchon PW, Brenton MD, Coppes JK, From AM, Torner JC.
8. Song Y, Cha J, Hwang C. Evaluation of insertion toruqe and pull- Factors affecting the pullout strength of self-drilling and self-
out strength of mini-screws according to different thickness of tapping anterior cervical screws. Spine 2003;28:9-13.
artificial bone. Korean J Orthod 2007;37:5-15. 23. Kim HJ, Yun HS, Park HD, Kim DH, Park YC. Soft-tissue and
9. Motoyoshi M, Hirabayashi M, Uemura M, Shimizu N. Recom- cortical-bone thickness at orthodontic implant sites. Am J Orthod
mended placement torque when tightening an orthodontic mini- Dentofacial Orthop 2006;130:177-82.
implant. Clin Oral Implants Res 2006;17:109-14. 24. Koistinen A, Santavirta SS, Kroger H, Lappalainen R. Effect
10. Motoyoshi M, Yano S, Tsuruoka T, Shimizu N. Biomechanical of bone mineral density and amorphous diamond coatings on
effect of abutment on stability of orthodontic mini-implant. A insertion torque of bone screws. Biomaterials 2005;26:
finite element analysis. Clin Oral Implants Res 2005;16:480-5. 5687-94.
11. Cattaneo PM, Dalstra M, Melsen B. Analysis of stress and strain 25. Seebeck J, Goldhahn J, Stadele H, Messmer P, Morlock MM,
around orthodontically loaded implants: an animal study. Int J Schneider E. Effect of cortical thickness and cancellous bone den-
Oral Maxillofac Implants 2007;22:213-25. sity on the holding strength of internal fixator screws. J Orthop
12. Cattaneo PM, Dalstra M, Melsen B. The finite element method: Res 2004;22:1237-42.
a tool to study orthodontic tooth movement. J Dent Res 2005; 26. Ciarelli MJ, Goldstein SA, Kuhn JL, Cody DD, Brown MB. Eval-
84:428-33. uation of orthogonal mechanical properties and density of human
13. Heidemann W, Gerlach KL, Grobel KH, Kollner HG. Influence of trabecular bone from the major metaphyseal regions with mate-
different pilot hole sizes on torque measurements and pullout rials testing and computed tomography. J Orthop Res 1991;9:
analysis of osteosynthesis screws. J Craniomaxillofac Surg 674-82.
1998;26:50-5. 27. Norton MR, Gamble C. Bone classification: an objective scale of
14. Brown GA, McCarthy T, Bourgeault CA, Callahan DJ. Mechani- bone density using the computerized tomography scan. Clin Oral
cal performance of standard and cannulated 4.0-mm cancellous Implants Res 2001;12:79-84.
bone screws. J Orthop Res 2000;18:307-12. 28. Ueda M, Matsuki M, Jacobsson M, Tjellstrom A. Relationship
15. Homolka P, Beer A, Birkfellner W, Nowotny R, Gahleitner A, between insertion torque and removal torque analyzed in fresh
Tschabitscher M, et al. Bone mineral density measurement with temporal bone. Int J Oral Maxillofac Implants 1991;6:442-7.

You might also like