GE Energy

Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : 73-VIA-867A(PROXPAC) Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330106-05-30-10-02-05 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : MAYY442507 0 0.798 N/A N/A
PROBE OHMS : 8.3 250 1.544 8.12 3.13
CABLE MODEL # : N/A * 500 3.573 8.28 5.26
CABLE SERIAL # : N/A 750 5.644 8.02 1.96
CABLE OHMS : N/A 1000 7.65 8.20 4.19
PROX MODEL # : 131233-01 1250 9.7 7.92 0.64
PROX SERIAL # : MARY111289 1500 11.68 8.32 5.72
PROBE LOCATION: KM-7302 NDE BRG (X) 1750 13.76 8.00 1.65
PROBE GAP(NEG) : 2000 15.76 7.64 -2.92
TARGET MATERIAL: 4140 Steel * 2250 17.67 6.88 -12.58
2500 19.39 6.16 -21.73
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 20.93 5.32 -32.40
5 SYSTEM LENGTH: 5 METERS 3000 22.26 7.42 -5.72
In c re m en ta l S F (in % )

Fill in the Gap voltage

10

5
0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-5

-10
-15

-20
-25

-30

-35 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

8.06 Average Scale Factor = 8.06 mV/µm


2.414 Percent Deviation = 2.41 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : 73-VIA-868A(PROXPAC) Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330106-05-30-10-02-05 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : MAYY442508 0 0.799 N/A N/A
PROBE OHMS : 8.3 * 250 1.205 7.98 1.45
CABLE MODEL # : N/A 500 3.201 8.15 3.58
CABLE SERIAL # : N/A 750 5.239 8.00 1.70
CABLE OHMS : N/A 1000 7.24 7.76 -1.40
PROX MODEL # : 131233-01 1250 9.18 7.72 -1.91
PROX SERIAL # : MARY111280 1500 11.11 7.72 -1.91
PROBE LOCATION: KM-7302 NDE BRG (X) 1750 13.04 7.60 -3.43
PROBE GAP(NEG) : 2000 14.94 7.24 -8.01
TARGET MATERIAL: 4140 Steel * 2250 16.75 6.72 -14.61
2500 18.43 6.16 -21.73
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 19.97 5.32 -32.40
5 SYSTEM LENGTH: 5 METERS 3000 21.3 7.10 -9.78
In c re m en ta l S F (in % )

Fill in the Gap voltage

10
5

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-5

-10

-15
-20

-25
-30

-35 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.77 Average Scale Factor = 7.77 mV/µm


1.271 Percent Deviation = 1.27 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : 73-VIA-867B(PROXPAC) Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330106-05-30-10-02-05 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : MAYY442509 0 0.803 N/A N/A
PROBE OHMS : 8.1 * 250 2.334 8.17 3.84
CABLE MODEL # : N/A 500 4.377 8.16 3.63
CABLE SERIAL # : N/A 750 6.416 8.02 1.86
CABLE OHMS : N/A 1000 8.42 7.76 -1.40
PROX MODEL # : 131233-01 1250 10.36 8.00 1.65
PROX SERIAL # : MARY111285 1500 12.36 7.80 -0.89
PROBE LOCATION: KM-7302 NDE BRG (Y) 1750 14.31 7.44 -5.46
PROBE GAP(NEG) : 2000 16.17 6.88 -12.58
TARGET MATERIAL: 4140 Steel * 2250 17.89 6.28 -20.20
2500 19.46 5.40 -31.39
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 20.81 4.88 -37.99
5 SYSTEM LENGTH: 5 METERS 3000 22.03 7.34 -6.69
In c re m en ta l S F (in % )

Fill in the Gap voltage

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-10

-20

-30

-40

-50 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.78 Average Scale Factor = 7.78 mV/µm


1.144 Percent Deviation = 1.14 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : 73-VIA-868B(PROXPAC) Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330106-05-30-10-02-05 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : MAYY442510 0 0.789 N/A N/A
PROBE OHMS : 8 * 250 0.952 7.82 -0.69
CABLE MODEL # : N/A 500 2.906 8.10 2.87
CABLE SERIAL # : N/A 750 4.93 7.88 0.13
CABLE OHMS : N/A 1000 6.9 7.72 -1.91
PROX MODEL # : 131233-01 1250 8.83 7.56 -3.94
PROX SERIAL # : MARY111288 1500 10.72 7.64 -2.92
PROBE LOCATION: KM-7302 NDE BRG (Y) 1750 12.63 7.48 -4.96
PROBE GAP(NEG) : 2000 14.5 7.20 -8.51
TARGET MATERIAL: 4140 Steel * 2250 16.3 6.52 -17.15
2500 17.93 6.12 -22.24
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 19.46 5.64 -28.34
5 SYSTEM LENGTH: 5 METERS 3000 20.87 6.96 -11.61
In c re m en ta l S F (in % )

Fill in the Gap voltage

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-5

-10

-15

-20

-25

-30 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.67 Average Scale Factor = 7.67 mV/µm


2.541 Percent Deviation = 2.54 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE PP C 3 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330101-00-08-05-02-05 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : 01L00HCC 0 0.62 N/A N/A
PROBE OHMS : 7.5 * 250 1.175 7.95 0.99
CABLE MODEL # : 330130-045-00-00 500 3.162 8.02 1.91
CABLE SERIAL # : 11C00GW0 750 5.167 8.05 2.31
CABLE OHMS : 9 1000 7.18 7.88 0.13
PROX MODEL # : 330180-50-05 1250 9.15 7.84 -0.38
PROX SERIAL # : 07I0018B 1500 11.11 7.64 -2.92
PROBE LOCATION: K-7302 1750 13.02 7.60 -3.43
PROBE GAP(NEG) : 2000 14.92 7.52 -4.45
TARGET MATERIAL: 4140 Steel * 2250 16.8 7.44 -5.46
2500 18.66 7.44 -5.46
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 20.52 7.36 -6.48
5 SYSTEM LENGTH: 5 METERS 3000 22.36 7.45 -5.29
In c re m e n ta l S F (in % )

Fill in the Gap voltage

3
2
1
0
-1 0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000

-2
-3
-4
-5
-6
-7 Displacement (in µm)
Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.81 Average Scale Factor = 7.81 mV/µm


0.762 Percent Deviation = 0.76 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
PASS Incremental Scale Factor Test = PASS
PASS

TEST RESULT = PASS


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE PP C 5 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330101-00-08-05-02-05 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : AUGY469707 0 0.619 N/A N/A
PROBE OHMS : 7.9 * 250 0.887 8.06 2.36
CABLE MODEL # : 330130-045-00-00 500 2.901 8.07 2.52
CABLE SERIAL # : 11C00GW0 750 4.918 8.09 2.77
CABLE OHMS : 9.4 1000 6.94 7.80 -0.89
PROX MODEL # : 330180-50-05 1250 8.89 7.72 -1.91
PROX SERIAL # : 07I0018B 1500 10.82 7.36 -6.48
PROBE LOCATION: K-7302 1750 12.66 7.20 -8.51
PROBE GAP(NEG) : 2000 14.46 6.92 -12.07
TARGET MATERIAL: 4140 Steel * 2250 16.19 6.76 -14.10
2500 17.88 6.48 -17.66
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 19.5 6.32 -19.70
5 SYSTEM LENGTH: 5 METERS 3000 21.08 7.03 -10.72
In c re m en ta l S F (in % )

Fill in the Gap voltage

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-5

-10

-15

-20

-25 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.65 Average Scale Factor = 7.65 mV/µm


2.795 Percent Deviation = 2.8 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE PP A 4 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : N/A (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : N/A 0 0.62 N/A N/A
PROBE OHMS : 7.8 * 250 0.641 6.83 -13.19
CABLE MODEL # : 330130-045-00-00 500 2.349 8.10 2.92
CABLE SERIAL # : 11C00GW0 750 4.374 8.03 2.01
CABLE OHMS : 9.3 1000 6.381 7.80 -0.94
PROX MODEL # : 330180-50-05 1250 8.33 7.60 -3.43
PROX SERIAL # : 07I0018B 1500 10.23 7.36 -6.48
PROBE LOCATION: K-7302 1750 12.07 7.16 -9.02
PROBE GAP(NEG) : 2000 13.86 6.80 -13.60
TARGET MATERIAL: 4140 Steel * 2250 15.56 6.60 -16.14
2500 17.21 6.28 -20.20
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 18.78 6.16 -21.73
5 SYSTEM LENGTH: 5 METERS 3000 20.32 6.77 -13.93
In c re m en ta l S F (in % )

Fill in the Gap voltage

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-5

-10

-15

-20

-25 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.46 Average Scale Factor = 7.46 mV/µm


5.21 Percent Deviation = 5.21 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE PP A 3 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330101-00-08-05-02-05 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : AUGY470402 0 0.626 N/A N/A
PROBE OHMS : 7.9 * 250 1.015 8.16 3.68
CABLE MODEL # : 330130-045-00-00 500 3.055 8.25 4.85
CABLE SERIAL # : 11C00GW0 750 5.118 8.25 4.80
CABLE OHMS : 9.3 1000 7.18 8.00 1.65
PROX MODEL # : 330180-50-05 1250 9.18 8.00 1.65
PROX SERIAL # : 07I0018B 1500 11.18 7.72 -1.91
PROBE LOCATION: K-7302 1750 13.11 7.60 -3.43
PROBE GAP(NEG) : 2000 15.01 7.44 -5.46
TARGET MATERIAL: 4140 Steel * 2250 16.87 7.44 -5.46
2500 18.73 7.28 -7.50
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 20.55 7.24 -8.01
5 SYSTEM LENGTH: 5 METERS 3000 22.36 7.45 -5.29
In c re m en ta l S F (in % )

Fill in the Gap voltage

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-2

-4

-6

-8

-10 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.93 Average Scale Factor = 7.93 mV/µm


0.762 Percent Deviation = 0.76 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE PP C 4 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330101-00-08-05-02-05 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : AUGY469710 0 0.628 N/A N/A
PROBE OHMS : 7.6 * 250 0.664 7.45 -5.36
CABLE MODEL # : 330130-045-00-00 500 2.526 8.22 4.45
CABLE SERIAL # : 11C00GW0 750 4.581 8.36 6.18
CABLE OHMS : 9.1 1000 6.67 8.16 3.68
PROX MODEL # : 330180-50-05 1250 8.71 8.00 1.65
PROX SERIAL # : 07I0018B 1500 10.71 7.92 0.64
PROBE LOCATION: K-7302 1750 12.69 7.80 -0.89
PROBE GAP(NEG) : 2000 14.64 7.68 -2.41
TARGET MATERIAL: 4140 Steel * 2250 16.56 7.68 -2.41
2500 18.48 7.64 -2.92
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 20.39 7.80 -0.89
5 SYSTEM LENGTH: 5 METERS 3000 22.34 7.45 -5.38
In c re m en ta l S F (in % )

Fill in the Gap voltage

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-2

-4

-6 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.95 Average Scale Factor = 7.95 mV/µm


1.017 Percent Deviation = 1.02 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE PP C 6 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330101-00-08-05-02-05 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : AUGY469708 0 0.623 N/A N/A
PROBE OHMS : 7.8 * 250 1.068 8.14 3.43
CABLE MODEL # : 330130-045-00-00 500 3.103 8.24 4.65
CABLE SERIAL # : 11C00GW0 750 5.162 8.23 4.60
CABLE OHMS : 9.1 1000 7.22 8.12 3.18
PROX MODEL # : 330180-50-00 1250 9.25 7.76 -1.40
PROX SERIAL # : 07J00JP 1500 11.19 7.80 -0.89
PROBE LOCATION: K-7302 1750 13.14 7.48 -4.96
PROBE GAP(NEG) : 2000 15.01 7.36 -6.48
TARGET MATERIAL: 4140 Steel * 2250 16.85 7.24 -8.01
2500 18.66 7.12 -9.53
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 20.44 7.04 -10.55
5 SYSTEM LENGTH: 5 METERS 3000 22.2 7.40 -5.97
In c re m en ta l S F (in % )

Fill in the Gap voltage

6
4

2
0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-2

-4
-6

-8
-10

-12 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.89 Average Scale Factor = 7.89 mV/µm


0.254 Percent Deviation = 0.25 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE PP D 3 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330101-00-08-05-02-05 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : JULY460609 0 0.627 N/A N/A
PROBE OHMS : 7.8 * 250 0.632 5.92 -24.78
CABLE MODEL # : 330130-045-00-00 500 2.112 8.09 2.77
CABLE SERIAL # : 11C00GW0 750 4.134 8.17 3.84
CABLE OHMS : 9.3 1000 6.177 8.13 3.33
PROX MODEL # : 330180-50-05 1250 8.21 7.88 0.13
PROX SERIAL # : 07I0018B 1500 10.18 7.80 -0.89
PROBE LOCATION: K-7302 1750 12.13 7.64 -2.92
PROBE GAP(NEG) : 2000 14.04 7.44 -5.46
TARGET MATERIAL: 4140 Steel * 2250 15.9 7.24 -8.01
2500 17.71 7.28 -7.50
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 19.53 7.08 -10.04
5 SYSTEM LENGTH: 5 METERS 3000 21.3 7.10 -9.78
In c re m en ta l S F (in % )

Fill in the Gap voltage

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-5

-10

-15

-20

-25

-30 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.63 Average Scale Factor = 7.63 mV/µm


3.05 Percent Deviation = 3.05 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE PP D 4 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : N/A (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : N/A 0 0.627 N/A N/A
PROBE OHMS : 7.8 * 250 0.737 8.26 4.90
CABLE MODEL # : 330130-045-00-00 500 2.801 8.18 3.99
CABLE SERIAL # : 11C00GW0 750 4.847 8.21 4.35
CABLE OHMS : 9.2 1000 6.9 8.04 2.16
PROX MODEL # : 330180-50-05 1250 8.91 7.92 0.64
PROX SERIAL # : 07I0018B 1500 10.89 7.76 -1.40
PROBE LOCATION: K-7302 1750 12.83 7.56 -3.94
PROBE GAP(NEG) : 2000 14.72 7.36 -6.48
TARGET MATERIAL: 4140 Steel * 2250 16.56 7.28 -7.50
2500 18.38 7.08 -10.04
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 20.15 7.16 -9.02
5 SYSTEM LENGTH: 5 METERS 3000 21.94 7.31 -7.07
In c re m en ta l S F (in % )

Fill in the Gap voltage

6
4

2
0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-2

-4
-6

-8
-10

-12 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.91 Average Scale Factor = 7.91 mV/µm


0.508 Percent Deviation = 0.51 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : 73-VIA-864A Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330101-00-08-05-02-02 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : JULY460605 0 0.628 N/A N/A
PROBE OHMS : 7.7 * 250 0.694 7.95 0.99
CABLE MODEL # : 330130-045-00-00 500 2.681 8.08 2.72
CABLE SERIAL # : 11C00GW0 750 4.702 8.15 3.58
CABLE OHMS : 9.2 1000 6.74 8.00 1.65
PROX MODEL # : 330180-50-05 1250 8.74 7.88 0.13
PROX SERIAL # : 07I0018B 1500 10.71 7.64 -2.92
PROBE LOCATION: KG-7302 WHEEL DE BRG (X) 1750 12.62 7.48 -4.96
PROBE GAP(NEG) : 2000 14.49 7.24 -8.01
TARGET MATERIAL: 4140 Steel * 2250 16.3 7.24 -8.01
2500 18.11 7.00 -11.05
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 19.86 6.96 -11.56
5 SYSTEM LENGTH: 5 METERS 3000 21.6 7.20 -8.51
In c re m en ta l S F (in % )

Fill in the Gap voltage

6
4
2
0
-2 0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000

-4
-6
-8
-10
-12
-14 Displacement (in µm)
Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.8 Average Scale Factor = 7.8 mV/µm


0.889 Percent Deviation = 0.89 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE A 2 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330101-00-08-05-02-02 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : JULY460603 0 0.628 N/A N/A
PROBE OHMS : 7.6 * 250 0.676 7.74 -1.70
CABLE MODEL # : 330130-045-00-00 500 2.61 8.12 3.23
CABLE SERIAL # : 11C00GW0 750 4.641 8.16 3.63
CABLE OHMS : 9.1 1000 6.68 7.92 0.64
PROX MODEL # : 330180-50-05 1250 8.66 7.84 -0.38
PROX SERIAL # : 07I0018B 1500 10.62 7.68 -2.41
PROBE LOCATION: KG-7302 1750 12.54 7.44 -5.46
PROBE GAP(NEG) : 2000 14.4 7.24 -8.01
TARGET MATERIAL: 4140 Steel * 2250 16.21 7.24 -8.01
2500 18.02 6.88 -12.58
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 19.74 7.00 -11.05
5 SYSTEM LENGTH: 5 METERS 3000 21.49 7.16 -8.98
In c re m en ta l S F (in % )

Fill in the Gap voltage

6
4
2
0
-2 0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000

-4
-6
-8
-10
-12
-14 Displacement (in µm)
Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.77 Average Scale Factor = 7.77 mV/µm


1.271 Percent Deviation = 1.27 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE D 1 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330101-00-08-05-02-02 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : JULY460610 0 0.627 N/A N/A
PROBE OHMS : 7.6 * 250 0.672 7.59 -3.58
CABLE MODEL # : 330130-045-00-00 500 2.569 8.20 4.24
CABLE SERIAL # : 11C00GW0 750 4.62 8.24 4.70
CABLE OHMS : 9.1 1000 6.68 8.16 3.68
PROX MODEL # : 330180-50-05 1250 8.72 7.88 0.13
PROX SERIAL # : 07I0018B 1500 10.69 7.72 -1.91
PROBE LOCATION: KG-7302 1750 12.62 7.56 -3.94
PROBE GAP(NEG) : 2000 14.51 7.48 -4.96
TARGET MATERIAL: 4140 Steel * 2250 16.38 7.28 -7.50
2500 18.2 7.28 -7.50
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 20.02 7.12 -9.53
5 SYSTEM LENGTH: 5 METERS 3000 21.8 7.27 -7.67
In c re m en ta l S F (in % )

Fill in the Gap voltage

6
4

2
0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-2

-4
-6

-8
-10

-12 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.85 Average Scale Factor = 7.85 mV/µm


0.254 Percent Deviation = 0.25 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
PASS Incremental Scale Factor Test = PASS
PASS

TEST RESULT = PASS


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE D 2 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : N/A (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : N/A 0 0.624 N/A N/A
PROBE OHMS : 7.3 * 250 0.658 7.24 -8.06
CABLE MODEL # : 330130-045-00-00 500 2.467 8.02 1.91
CABLE SERIAL # : 11C00GW0 750 4.472 8.06 2.47
CABLE OHMS : 8.6 1000 6.488 7.97 1.25
PROX MODEL # : 330180-50-00 1250 8.48 7.88 0.13
PROX SERIAL # : 07J007JP 1500 10.45 7.56 -3.94
PROBE LOCATION: KG-7302 1750 12.34 7.52 -4.45
PROBE GAP(NEG) : 2000 14.22 7.36 -6.48
TARGET MATERIAL: 4140 Steel * 2250 16.06 7.20 -8.51
2500 17.86 7.12 -9.53
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 19.64 7.08 -10.04
5 SYSTEM LENGTH: 5 METERS 3000 21.41 7.14 -9.32
In c re m en ta l S F (in % )

Fill in the Gap voltage

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-2

-4

-6

-8

-10

-12 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.7 Average Scale Factor = 7.7 mV/µm


2.16 Percent Deviation = 2.16 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE PP C 2 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : N/A (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : N/A 0 0.624 N/A N/A
PROBE OHMS : 7.9 * 250 0.633 6.22 -20.91
CABLE MODEL # : 330130-045-00-00 500 2.189 8.22 4.45
CABLE SERIAL # : 11C00GW0 750 4.244 8.20 4.14
CABLE OHMS : 9.2 1000 6.293 8.15 3.53
PROX MODEL # : 330180-50-00 1250 8.33 7.92 0.64
PROX SERIAL # : 07J007JP 1500 10.31 7.80 -0.89
PROBE LOCATION: KG-7302 1750 12.26 7.52 -4.45
PROBE GAP(NEG) : 2000 14.14 7.44 -5.46
TARGET MATERIAL: 4140 Steel * 2250 16 7.20 -8.51
2500 17.8 7.16 -9.02
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 19.59 6.96 -11.56
5 SYSTEM LENGTH: 5 METERS 3000 21.33 7.11 -9.66
In c re m en ta l S F (in % )

Fill in the Gap voltage

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-5

-10

-15

-20

-25 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.68 Average Scale Factor = 7.68 mV/µm


2.414 Percent Deviation = 2.41 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE PP C 1 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : N/A (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : N/A 0 0.628 N/A N/A
PROBE OHMS : 7.7 * 250 1.909 8.16 3.63
CABLE MODEL # : 330130-045-00-00 500 3.948 8.32 5.67
CABLE SERIAL # : 11C00GW0 750 6.027 8.13 3.33
CABLE OHMS : 9 1000 8.06 8.04 2.16
PROX MODEL # : 330180-50-05 1250 10.07 7.76 -1.40
PROX SERIAL # : 07I0018B 1500 12.01 7.64 -2.92
PROBE LOCATION: KG-7302 1750 13.92 7.44 -5.46
PROBE GAP(NEG) : 2000 15.78 7.36 -6.48
TARGET MATERIAL: 4140 Steel * 2250 17.62 7.16 -9.02
2500 19.41 7.16 -9.02
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 21.2 4.72 -40.03
5 SYSTEM LENGTH: 5 METERS 3000 22.38 7.46 -5.21
In c re m en ta l S F (in % )

Fill in the Gap voltage

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-10

-20

-30

-40

-50 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.86 Average Scale Factor = 7.86 mV/µm


0.127 Percent Deviation = 0.13 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE PP A 5 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : N/A (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : N/A 0 0.626 N/A N/A
PROBE OHMS : 7.8 * 250 1.241 8.09 2.77
CABLE MODEL # : 330130-045-00-00 500 3.263 8.18 3.99
CABLE SERIAL # : 11C00GW0 750 5.309 8.16 3.74
CABLE OHMS : 9.2 1000 7.35 7.92 0.64
PROX MODEL # : 330180-50-05 1250 9.33 7.80 -0.89
PROX SERIAL # : 07I0018B 1500 11.28 7.52 -4.45
PROBE LOCATION: KG-7302 1750 13.16 7.40 -5.97
PROBE GAP(NEG) : 2000 15.01 7.24 -8.01
TARGET MATERIAL: 4140 Steel * 2250 16.82 6.96 -11.56
2500 18.56 6.96 -11.56
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 20.3 6.76 -14.10
5 SYSTEM LENGTH: 5 METERS 3000 21.99 7.33 -6.86
In c re m en ta l S F (in % )

Fill in the Gap voltage

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-5

-10

-15

-20 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.79 Average Scale Factor = 7.79 mV/µm


1.017 Percent Deviation = 1.02 %
5 Maximum allowable percentage deviation for a 5 meter system = 5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE D1 D2 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330102-00-75-10-02-00 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : AUGY 468554 0 0.635 N/A N/A
PROBE OHMS : 7.7 * 250 0.803 8.14 3.43
CABLE MODEL # : 330130-080-00-05 500 2.838 8.17 3.84
CABLE SERIAL # : MARN 216888 750 4.881 8.12 3.13
CABLE OHMS : 10.4 1000 6.91 8.04 2.16
PROX MODEL # : 330180-90-05 1250 8.92 7.68 -2.41
PROX SERIAL # : 08E00AX5 1500 10.84 7.72 -1.91
PROBE LOCATION: 7302 1750 12.77 7.36 -6.48
PROBE GAP(NEG) : 2000 14.61 7.40 -5.97
TARGET MATERIAL: 4140 Steel * 2250 16.46 7.20 -8.51
2500 18.26 7.32 -6.99
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 20.09 7.16 -9.02
9 SYSTEM LENGTH: 9 METERS 3000 21.88 7.29 -7.33
In c re m en ta l S F (in % )

Fill in the Gap voltage

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-2

-4

-6

-8

-10 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.83 Average Scale Factor = 7.83 mV/µm


0.508 Percent Deviation = 0.51 %
6.5 Maximum allowable percentage deviation for a 9 meter system = 6.5%
PASS Incremental Scale Factor Test = PASS
PASS

TEST RESULT = PASS


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : ZONE D1 D1 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330102-00-78-10-02-05 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : FEBG 408650 0 0.639 N/A N/A
PROBE OHMS : 7.7 * 250 0.701 7.85 -0.28
CABLE MODEL # : 330130-080-00-05 500 2.663 8.08 2.67
CABLE SERIAL # : MARN 216888 750 4.683 8.11 3.02
CABLE OHMS : 10.2 1000 6.71 7.88 0.13
PROX MODEL # : 330180-90-05 1250 8.68 7.96 1.14
PROX SERIAL # : 08E00AX5 1500 10.67 7.72 -1.91
PROBE LOCATION: 7302 1750 12.6 7.80 -0.89
PROBE GAP(NEG) : 2000 14.55 7.64 -2.92
TARGET MATERIAL: 4140 Steel * 2250 16.46 7.76 -1.40
2500 18.4 7.84 -0.38
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 20.36 7.84 -0.38
9 SYSTEM LENGTH: 9 METERS 3000 22.32 7.44 -5.46
In c re m e n ta l S F (in % )

Fill in the Gap voltage

4
3
2
1
0
-1 0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000

-2
-3
-4
-5
-6 Displacement (in µm)
Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.88 Average Scale Factor = 7.88 mV/µm


0.127 Percent Deviation = 0.13 %
6.5 Maximum allowable percentage deviation for a 9 meter system = 6.5%
PASS Incremental Scale Factor Test = PASS
PASS

TEST RESULT = PASS


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : NO 2 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330102-00-75-10-02-00 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : AUGY 468552 0 0.641 N/A N/A
PROBE OHMS : 8.2 * 250 0.714 8.05 2.26
CABLE MODEL # : 330130-080-00-05 500 2.726 8.32 5.72
CABLE SERIAL # : MARN 216888 750 4.806 8.14 3.38
CABLE OHMS : 10.4 1000 6.84 8.04 2.16
PROX MODEL # : 330180-90-05 1250 8.85 7.80 -0.89
PROX SERIAL # : 08E00AX5 1500 10.8 7.68 -2.41
PROBE LOCATION: 7302 1750 12.72 7.44 -5.46
PROBE GAP(NEG) : 2000 14.58 7.44 -5.46
TARGET MATERIAL: 4140 Steel * 2250 16.44 7.24 -8.01
2500 18.25 7.32 -6.99
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 20.08 7.28 -7.50
9 SYSTEM LENGTH: 9 METERS 3000 21.9 7.30 -7.24
In c re m en ta l S F (in % )

Fill in the Gap voltage

8
6

4
2

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-2
-4

-6
-8

-10 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.86 Average Scale Factor = 7.86 mV/µm


0.127 Percent Deviation = 0.13 %
6.5 Maximum allowable percentage deviation for a 9 meter system = 6.5%
PASS Incremental Scale Factor Test = PASS
PASS

TEST RESULT = PASS


GE Energy
Bently Nevada™ Asset Condition Monitoring LINEARITY CHECK SHEET

CUSTOMER NAME: SHELL REFINING COMPANY / PLANT NAME: SHELL REFINING COMPANY
P/O NO: 4504418513 / CONTRACT NO: N/A / EQUIP. NO.: ______________

TRANSDUCER INFORMATION CALIBRATION DATA


POINT I. D. : NO 1 Displacement GAP Incremental SF Incremental SF
PROBE MODEL # : 330102-00-75-10-02-00 (µm) (-V) (mV/µm) (%)
PROBE SERIAL # : AUGY 468551` 0 0.643 N/A N/A
PROBE OHMS : 8.1 * 250 0.664 7.22 -8.26
CABLE MODEL # : 330130-080-00-05 500 2.469 8.10 2.92
CABLE SERIAL # : MARN 216888 750 4.494 8.26 5.01
CABLE OHMS : 10.6 1000 6.56 8.08 2.67
PROX MODEL # : 330180-90-05 1250 8.58 7.88 0.13
PROX SERIAL # : 08E00AX5 1500 10.55 7.56 -3.94
PROBE LOCATION: 7302 1750 12.44 7.72 -1.91
PROBE GAP(NEG) : 2000 14.37 7.32 -6.99
TARGET MATERIAL: 4140 Steel * 2250 16.2 7.44 -5.46
2500 18.06 7.24 -8.01
7.87 PROBE SCALE F.: 7.87 mV/µm 2750 19.87 7.56 -3.94
9 SYSTEM LENGTH: 9 METERS 3000 21.76 7.25 -7.84
In c re m en ta l S F (in % )

Fill in the Gap voltage

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
-2

-4

-6

-8

-10 Displacement (in µm)


Figure 1: Incremental Scale Factor (ISF)
DC Voltage (in -V)

25

20

15

10

0
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000
Displacement (in µm)
Figure 2: Direct (GAP) voltage

7.77 Average Scale Factor = 7.77 mV/µm


1.271 Percent Deviation = 1.27 %
6.5 Maximum allowable percentage deviation for a 9 meter system = 6.5%
FAIL Incremental Scale Factor Test = FAIL
FAIL

TEST RESULT = FAIL