Coil 10 DPA Lumbar Fig. 127: Coil 10 DPA Lumbar

Remove the lower frame and unscrew the screws indicated by


arrows.

WARNING DO NOT TOUCH THE BRASS SCREWS THAT


FIX THE COIL BODY TO ITS BASE AND
INDICATED IN THE NEXT FIGURE BY THE
DENIED SIGN!!!

Fig. 126: Coil 10 DPA Lumbar: how to open it

– C12 and C13: disconnect the coil and measure 0.6 V (diode
voltage) using the multi meter in diode function
– Giallo Cable: connect the coil and measure the powers supply
(15 V)
– Verde Cable: recognition resistor of the coil
– Rosso Cable: connect the coil and move the Varicap1 value by
software (Under ARAS, select <Test> <Manual> <Coil tuning>)
and measure the voltage variations
– Blu Cable: connect the coil and move the Varicap2 value by
software (Under ARAS, select <Test> <Manual> <Coil tuning>)
and measure the voltage variations
Gantry Connectors
Every coil uses the same connector to receive and send information
to the system; this connector is fixed to the upper right side of the
Magnet.

Tab. 11: Resistance coil recognition values

Coil Type Value


1 Shoulder Linear 427 Ω
2 Knee 274 Ω
3 Hand/Wrist 352 Ω
4 Foot/Ankle 583 Ω
6 Flex 505 Ω
7 Shoulder DPA 299 Ω
8 Lumbar/Thoracic 324 Ω
9 Neck 377 Ω
10 DPA Lumbar 403 Ω

Refer to the figures present in the System chapter for the Magnet
connection schemes.