Professional Documents
Culture Documents
Alluramonoplaneinstructionforuseeng PDF
Alluramonoplaneinstructionforuseeng PDF
INTEGRIS
Allura
Monoplane
INTEGRIS SYSTENS INSTRUCTIONS FOR USE
Release 1
English
9896 001 33152
INTEGRIS Allura
Monoplane
INSTRUCTIONS FOR USE
Release 1
English
9896 001 33152
Philips Medical Systems
Instructions for Use
Document number
9896 001 33152
4 I N T E G R I S A l l u r a M on o p la n e Release 1
Contents
Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
5 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
0- 2 C o n te n t s I N T E G R I S A l l u r a M on o p la n e Release 1
Contents
7 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
10 Appendices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
11 Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
12 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-1
0- 4 C o n te n t s I N T E G R I S A l l u r a M on o p la n e Release 1
1 Introduction
In order to identify this ‘Instructions For Use’, with the system for which it is
intended to be used, a label has been placed on the title page and indicates
the following:
• project number
• shop/COS number
• customer order number
• project destination.
9896 001 33152
Before using this ‘Instructions For Use’ with your system, ensure that the
identification labels both show the same information.
This manual is intended to assist users and operators in the safe and effective
operation of the equipment described. The ‘user’ is considered to be the body
with authority over the equipment; ‘operators’ are those persons who actually
handle the equipment. Before attempting to operate the equipment, you
must read, note and strictly observe all DANGER notices and safety
markings on the INTEGRIS Allura Monoplane.
Before attempting to operate the equipment, you must read this manual
thoroughly, paying particular attention to all WARNINGS, Cautions and
Notes incorporated in it. You must pay special attention to all the
information given and procedures described in the SAFETY section.
WA R N I N G S are directions which if not followed could cause fatal or serious injury to an operator,
patient or any other person, or could lead to a misdiagnosis.
CAUTIONS are directions which if not followed could cause damage to the equipment described in
this Manual and/or any other equipment or goods, and/or cause environmental
pollution.
This Manual describes the system in its full configuration with all options
and accessories fitted. Not every function described may be available on your
system.
This English language version of the Manual was originally drafted, approved
and supplied by Philips Medical Systems under the product part code
indicated on page 4.
sections describe specific actions and the sequence in which they must be
performed, with references to the ‘Legend’ for locations, and to the ‘System
overview’ for detailed descriptions of the functions concerned.
Installation, use and operation of this equipment is subject to the law in the
jurisdiction(s) in which the equipment is being used. Both users and
operators must only install, use and operate the equipment in such ways as
do not conflict with applicable laws, or regulations which have the force of
law.
Uses of the equipment for purposes other than those intended and expressly
stated by the manufacturer, as well as incorrect use or operation, may relieve
the manufacturer (or his agent) from all or some responsibility for resultant
non-compliance, damage or injury.
CAUTION In the United States, Federal law restricts this device to sale, distribution, and use by or
on the order of a physician.
1.4 Contra-indications
X-rays are potentially hazardous. Special precautions must be taken and/or
caution must be exercised in the following cases:
• Special consideration must be given to the protection of the embryo or
fetus during radiological examination or treatment of women known to be
pregnant.
• Sensitive body organs (e.g., lens of eye, gonads) must be shielded whenever
they are likely to be exposed to the working beam.
• Acute skin burns (patients).
• Acute hair loss (patients).
• Chronic radiation injury (staff).
9896 001 33152
1.5 Compatibility
Philips Medical Systems
Changes and/or additions to the equipment that are carried out by persons
without the appropriate training and/or using unapproved spare parts may
lead to the Philips Medical Systems warranty being voided. As with all
complex technical equipment, maintenance by persons not appropriately
qualified and/or using unapproved spare parts carries serious risks of damage
to equipment and personal injury.
1.6 Compliance
The Philips INTEGRIS Allura Monoplane complies with relevant
international and national standards and laws. Information on compliance
can be obtained from your local Philips Medical Systems representative or
from:
1.7 Training
Operators of the Philips INTEGRIS Allura Monoplane must have received
adequate training in its safe and effective use before attempting to operate the
9896 001 33152
It is vital that you read, note, and where applicable strictly observe all
DANGER notices and safety markings on the INTEGRIS Allura
Monoplane.
It is vital that you follow strictly all safety directions under the heading
SAFETY and all WARNINGS and CAUTIONS throughout this Manual, to
help ensure the safety of both patients and operators.
Only qualified and authorized personnel may operate this equipment. In this
context, qualified means those legally permitted to operate this type of
medical electrical equipment in the jurisdiction(s) in which the equipment is
being used, and authorized means those authorized by the user of the
equipment.
You can find information about the ‘User Routine Checks Program’ and the
‘Planned Maintenance Program’ in the ‘Maintenance’ section of this Manual.
Safety awareness
• Do not use the INTEGRIS Allura Monoplane for any application until you have
read, understood and learnt all the safety information, safety procedures and
emergency procedures contained in this SAFETY section. Operation of the
INTEGRIS Allura Monoplane without a proper awareness of how to use it safely
could lead to serious or fatal injury. It could also lead to clinical misdiagnosis/
incorrect treatment.
Adequate training
• Do not use the INTEGRIS Allura Monoplane for any application until you have
received adequate and proper training in its safe and effective operation. If you are
unsure of your ability to operate this equipment safely and effectively DO NOT
USE IT. Operation of this equipment without proper and adequate training could
lead to serious of fatal injury. It could also lead to clinical misdiagnosis/incorrect
treatment.
Safety devices
• Never attempt to remove, modify, override or frustrate any safety device on the
equipment. Interfering with safety devices could lead to serious or fatal injury.
WA R N I N G S • In the event of a system movement emergency, press emergency power off [130] or
[160] to block all movements except manual L-arm rotation and longitudinal
movement and tabletop float.
WA R N I N G Do not remove covers or cables from this equipment unless expressly instructed to do
so in this Manual. High electrical voltages are present within this equipment. Removing
covers or cables could lead to serious or fatal injury.
Only use this equipment in rooms or areas that comply with all applicable
laws (or regulations having the force of law) concerning electrical safety for
this type of equipment. Always electrically isolate this equipment from the
mains electrical supply before cleaning, disinfecting or sterilizing it.
WA R N I N G Do not remove covers from this equipment unless expressly instructed to do so in this
Manual. Moving parts are present within this equipment. Removing covers could lead
to serious or fatal injury.
Philips Medical Systems
WA R N I N G Flammable or potentially explosive disinfecting sprays must not be used since the
resultant vapor could ignite, causing fatal or serious injury and/or damage to
equipment.
WA R N I N G Only use extinguishers on electrical or chemical fires which are specifically labeled for
those purposes. Using water or other liquids on an electrical fire can lead to fatal or
serious injury.
WA R N I N G Do not allow any portable radio devices (such as mobile telephones) into the
examination room - whether switched on or off. Such devices could exceed EMC
Philips Medical Systems
radiation standards and, under certain conditions, interfere with the proper
functioning of the system. This could, in extreme cases, lead to fatal or serious injury.
CAUTION In the United States Federal law restricts this device to sale, distribution, and use by or
on the order of a physician.
Full use must be made of all the equipment’s radiation protection features
and of all radiation protection devices, accessories, systems and procedures
available to you as the operator.
WA R N I N G Never attempt to remove, modify, override or frustrate any safety device on the
equipment. Interfering with safety devices could lead to serious or fatal injury.
Radiation guidelines
Leakage radiation
The leakage radiation for the X-ray tube/collimator assembly is less than 100
mR/hr (0.87 mGry/hr) measured at a distance of 1 meter in any direction of
the source when operated at the leakage technique factors 125 kV, 12mA.
N OT E For information on stray radiation refer to section 9, Technical Data.
Not Applicable.
9896 001 33152
Philips Medical Systems
L-arm
Examination and
reference monitors
C-arm stand
Display unit
Acquisition console
Viewing monitor
Viewing console
36280010
INTEGRIS Allura Monoplane system
L-arm
Image
intensifier
C-arm
9896 001 33152
X-ray tube
36280020
Philips Medical Systems
The geometry segment provides total body coverage and offers a wide range
of projections for PA and AP imaging for all vascular and neuro applications.
The INTEGRIS Allura C-arm stand with 38 cm (15 inch) or 30 cm (12
inch) Image Intensifier (II) is attached to a ceiling suspended L-arm. The L-
arm is suspended from a ceiling rail which runs parallel to the longitudinal
axis of the table.
Rotation:
• from 90° LAO (+rotation) to 90° RAO (-rotation) (stand in its side
position perpendicular to the table)
• from 120° LAO (+rotation) to 185° RAO (-rotation) (stand in its head
position parallel to the table)
• variable rotation speed via tableside operation module up to 25°/sec.
Cranial/caudal angulation:
• from 90° cranial (+angulation) to 90° caudal (-angulation) (max = 90°)
(stand in its head position parallel to the table)
• from 120° cranial (+angulation) to 185° caudal (-angulation) (stand in its
side position perpendicular to the table)
• variable angulation speed via tableside operation module up to 18°/sec.
N OT E The full angulation capability is limited by the table position and patient position.
Rotational angiography:
• from 90° LAO (+rotation) to 90° RAO (-rotation) (stand in its side
position perpendicular to the table (rolling movement)) at a speed of up to
30°/sec
• from 120° LAO (+rotation) to 185° RAO (+rotation) (stand in its head
position parallel to the table (propeller movement)) at a speed of up to
55°/sec.
L-arm movements
• longitudinal movement (parking movement) of 260 cm (102.3 inch)
• motorized longitudinal movement speed of 15 cm/sec (5.9 inch/sec)
Philips Medical Systems
N OT E The maximum speeds are set in the APR’s (by Service) and cannot be changed by the
physician.
Options
Short L-arm
The short L-arm allows the C-arm to be installed in rooms with a lower than
normal ceiling height. By using the short L-arm, the minimum ceiling height
is reduced from 290 cm (114.2 inch) to 270 cm (106.3 inch).
Options
Ratchet compressor
The option comprises:
• 3 cotton compression belts 23 cm (9.1 inch) wide
• a ratchet winding mechanism on one side for symmetrical compression.
SyncraTilt
SyncraTilt enhances the accuracy and efficiency of gravity-oriented
procedures. SyncraTilt is ideal for interventional, myelography,
phlebography and head down procedures because it provides more precise
imaging of contrast medium, blood, or objects in the body. With SyncraTilt,
the isocenter is automatically located at the isocenter of rotation and
angulation of the stand. If the longitudinal position of the stand changes, the
tilt isocenter is changed to match the new stand position, ensuring that the
region of interest (ROI) is always centered.
As the table tilts, the X-ray beam automatically compensates for the
movement. When the tilt option is installed:
• manual tabletop float is only transversal
• longitudinal tabletop movement is motorized only.
The table floats even when tilted and the ROI can be tracked by panning the
tabletop. When combined with the Bolus Chase option, SyncraTilt allows
phlebography to be performed with a head-up tilted patient.
The option provides:
• maximum tilt range: -28° head down tilt (HDT) to +20° head up tilt
(HUT). For isocentric tilt, the tilt range depends on the distance between
the isocenter of the stand and the table base
• an automatic safety system with manual override
• a panning range in tilted plane equal to the standard tabletop
specifications (longitudinal movement of 100 cm (39.4 inch), lateral
movement of 36 cm (14.2 inch)).
Accessories:
• foot support
• shoulder support (not with neuro tabletop)
• ankle support
• chin support (cushion)
• 4 handgrips.
Options
Pan handle
An extension of the control facility for floating movements of the tabletop,
the pan handle is used to assist tabletop positioning at the tableside in parallel
with the standard geometry Tableside Operation module (TSO). It can be
attached anywhere to the tabletop and accessory rails without decreasing the
floating range. The pan handle is connected to the table-base connection box
The option includes a controller and APC module. The projections are
indicated on the display unit.
4.2.4 Acquisition
General
• 100 kW X-ray power pack generator
• integrated generator/examination console
• X-ray depth collimator, including two independent semi-transparent
wedged filters with automatic and manual positioning
• CCD based imaging chain with a high performance state of the art CCD
camera with a digital output.
The chains include:
9896 001 33152
Infrared receiver
The infrared receiver, for communication with the viewpad(s), is integrated
into the monitor suspension and positioned just above the monitors.
XB monitors
Two sizes of XB monitors are specified with the system:
• a 21-inch monitor, with a concave front, used in the examination room
and mounted in the monitor ceiling suspension
• a 17-inch monitor used for viewing in the control room.
X-ray tubes
The system is equipped with either an MRM or MRC X-ray tube:
9896 001 33152
MRM-GS 04 10
• MRM 04 10 Maximus ROTALIX Metal tube with grid switch and 0.4/
1.0 mm nominal focal spot values, maximal 25 and 85 kW short time load
• grid switching in pulsed fluoroscopy mode
Philips Medical Systems
MRC-GS 03-10
• MRC-GS 200 03 10 Maximus ROTALIX Ceramic tube with grid switch
and 0.3/1.0 mm nominal focal spot values, maximal 20 and 100 kW short
time load
• grid switching in pulsed fluoroscopy mode
• anode heat storage capacity: 2.4 MHU
• continuous heat dissipation: 3.5 kW
• ROT-GS 1004 tube housing for oil cooled X-ray tube with thermal safety
switch.
MRC-GS 04-07
• MRC-GS 200 04 07 Maximus ROTALIX Ceramic tube with grid switch
and 0.4/0.7 mm nominal focal spot values, maximal 30 and 67 kW short
time load
• grid switching in pulsed fluoroscopy mode
• anode heat storage capacity: 2.4 MHU
• continuous heat dissipation: 3.5 kW
• ROT-GS 1004 tube housing for oil cooled X-ray tube with thermal safety
switch.
N OT E Both MRM and MRC X-ray tubes offer additional beam filtration capabilities. See
section 9 ‘Technical data’ for additional information.
CAUTIONS In case a defective spectrabeam is detected, the user is warned that there might be a
higher dose rate than usual during fluoroscopy. The following messages will then be
displayed:
• On the frontal display unit: ‘Warning’
• On the acquisition console: ‘Spectral filter defect Call Service’.
In case a defective grid switch is detected, the user is warned that there might be a
higher dose rate than usual during fluoroscopy. The following message will then be
displayed:
• On the acquisition console: ‘Failing gridswitch Call Service’.
380
36280160
9896 001 33152
Beam limitation
The system has been provided with a beam limiting device (shutters) to limit
the extent of the X-ray field. The extent of the X-ray field is automatically
limited to just outside the edges of the maximum image reception area as
displayed on the monitor. The size of the actual maximum image reception
area is dependent on the selected II field size. The extent of the X-ray field
can be manually limited with rectangular X-ray shutters.
Fluoroscopy
There are three programmable fluoroscopy modes (3 (High), 2 (Normal) and
1 (Low)).
The configuration of these modes is dependent on whether or not the system
is configured for the following functions:
• dynamic pulsed fluoroscopy
• MRC-GS X-ray tube with SpectraBeam.
The high, normal and low fluoroscopy modes differ with regard to image
quality, dose rate and image processing (dose level, pulsed/continuous
fluoroscopy, beam filtration, noise reduction, adaptive contour enhancement
and harmonization). The modes can be selected from the imaging TSO. The
settings are configured by Service.
If the AEP meter option is present, the dose rate is indicated on the display
unit [127]. Dose reduction can be achieved by using Cu filters for X-ray
beam filtration. For filter values, refer to section 9 ‘Technical data’.
Other fluoroscopy features:
• Trace Subtract Fluoroscopy (TSF)
• fluoroscopy image storage and archiving (frame grabbing)
• fluoroscopy stopwatch.
50 HZ 8, 12.5, 25 6, 8, 12.5, 25
Options
The colour monitor is used to display colour images and data in the
examination room, for instance provided by, either:
• A hemodynamic acquisition system, or
• INTEGRIS 3D-RA.
Dual fluoro
This is a digital subsystem with additional 21 inch XB monitor allowing:
• digitally processed fluoroscopy in parallel with viewing and processing of
previously acquired images of the current and/or previous patient to
increase patient throughput and procedure efficiency
• digitally processed fluoroscopy in parallel with Trace Subtract Fluoroscopy
(TSF), providing a non-subtracted reference fluoro image for complex
interventions
The option also provides a fluoro channel in parallel to the default fluoro
channel, allowing fluoroscopy during image transfer i.e. to a CD-Medical or
remote destination. Viewing console functions are suspended during
exposures and enabled again after the exposures are made.
Examination light
The monitor suspension is provided with a bracket for mounting the
optional examination light.
• working position with lower shield only (upper shield folded down)
• parking position (shield folded under the table).
The radiation shield is provided with a special AD5-table clamping device
Philips Medical Systems
preparation as it can fully be parked under the tabletop. The shield can be
simply converted (by Service) for fitting to the nurse-side of the table. If
necessary, two protection shields can be fitted - one on either side of the
table. The option includes a docking rail which can be mounted to the wall.
Basic 25 6.25
60 Hz system:
Basic 30 7.5
1)
Or 25 at 8 bits/pixel.
2)
Or 30 at 8 bits/pixel.
Basic 25 6.25
60 Hz system:
Basic 30 7.5
Basic 25 6.25
60 Hz system:
Basic 30 7.5
Options
Storage extension
This option comprises two additional digital disks in order to extend the
image storage capacity of the basic system.
Dose/Examination report
This is an optional report facility for examination or dose data. The report
can be printed, in A4 format using the standard ASCII character set, on any
standard printer to complete the patient file. The printer is connected to the
acquisition console via a parallel (centronics) interface. The report is printed
at the touch of a button on the acquisition console at the end of the
9896 001 33152
• per run: run number, real start time, kV, mA, mAs, ms (last used value, per
channel).
Physio display
Extension for acquisition, storage and display of physiological signals.
The option provides:
• acquisition and storage of 8 channels of physio data with images
• display of one user-selected physio channel selected during image review.
Rotational Angiography
Rotational Angiography is an acquisition procedure that gives information
on the 3-dimensional structure of blood vessels through rotation of the C-
arm during image acquisition. One of the system acquisition programs is
used to select the procedure and its specific parameters. The rotational end
and start positions can be selected by the operator and stored using the store/
recall buttons on the geometry module. The procedure is controlled using
the exposure hand or foot switch, or the injector hand switch.
For subtraction rotational angiography, two acquisition runs are made. After
the first (mask) run, the stand automatically returns to its start position,
ready for the second (dye) run. The images are displayed in non-subtracted
mode during acquisition. Subtraction can be carried out during viewing.
50 Hz system:
60 Hz system:
Basic 15 15
50 Hz system:
60 Hz system:
N OT E With the ‘Dynamic acquisition’ package option installed, the rotational angiography
APRs use kV/mA exposure techniques, the other rotational angiography APRs use
9896 001 33152
Any two of the following probes can be connected to the scanner: 7.5 MHz
linear array, 3.5/5.0 MHz curved array, 3.5 MHz high-definition curved
array and 5.0/7.5 MHz curved array. Needle guide accessories can be
provided for the 7.5 MHz linear array, the 3.5/5.0 MHz curved array and the
3.5 MHz high-definition curved array probes.
4.2.6 Viewing
The viewing segment comprises:
• desktop viewing console with SyncraTouch control, providing direct access
to multiple viewing/post-processing functions for efficient operation. It
includes:
- alphanumeric keyboard and display for patient administration, a
control panel and a 17 inch XB monitor for permanent image viewing
- clinical application software packages with short response times
(optional)
• tableside viewing in the examination room using an infrared remote
controlled SyncraTouch viewpad (2 viewpads included) or the Tableside
Analysis Module (TAM).
Viewpad
Viewpad functionality can be configured to the user’s preference at
installation.
Options
CO2 trace
Software package for stacking images acquired with CO2 injections.
• calibration routines.
Functions:
• Automated Left Ventricular Analysis (ALVA)
• manual EF
• regional wall motion
• centerline wall motion
• Slager wall motion
• calibration routines
• recalculate EF
• recalculate ALVA.
Autocalibration
This option avoids the need for manual calibration for objects in the
isocenter when using the quantification software packages.
Parallel viewing
This option allows images to be viewed and processed at the viewing console
while another patient is being examined in the examination room. Parallel
viewing is temporarily suspended during fluoroscopy and/or exposures.
With the TAM connected to the patient accessory rail via a dedicated holder
the operator can perform post-processing and quantification functions
without leaving the catheterization room.
The option comprises:
• a module (TAM) with five push-buttons and a joystick for (configuration
dependent) control of image selection, processing, subtraction, photofile/
reference, text & annotation, selection for transfer and other general
functions
• holder for infrared remote control SyncraTouch viewpad
• holder for module, which can be connected to the table accessory rail.
Options
Analog output
For connecting an (analog) PMI 11 RC or PMI 3000 multi-image camera.
discs.
For safety reasons, this software package runs exclusively on a PC and does
not allow use of other software packages.
Philips Medical Systems
The user interface shows all incoming, recorded and temporarily stored cases.
Recording progress and status are shown, as well as the content of the CD-R
loaded in the system. Cases can be recorded by mouse-only operation,
automatically or by free case selection.
The option comprises:
• Inturis DICOM Recorder release 1 software, including a license for one
input connection
• 10 CD-Recordables.
On request of a clinical user, the INTEGRIS Allura system will export the
following information to the IS:
• patient data:
- patient name
- birth date
- sex
- patient ID.
• examination/request information:
- accession number
- requested procedure ID
- performed station title
- performed procedure step ID, status, start/end date and time
- performed physician’s name
- referenced image sequence.
• radiation dose:
- total time of fluoroscopy
- accumulated fluoroscopy dose
- accumulated exposure dose
- total dose
- total number of exposures
- total number of images.
Further detailed information can be found in the INTEGRIS DICOM
Conformance Statement.
The interface requires EasyLink hardware and software if the IS is not
DICOM Work List Management and Modality Performed Procedure Step
compliant.
2
80 kv 900 mA 170 ms 00 min
1
3
2 Gycm2
4a 4b
4
4c
4c
A 10
B 8
5
9
33726400
6
7
Acquisition console
N OT E To avoid malfunction, do not touch any key during the Switch-On process.
After certain errors, the system will reset automatically. Notify Service if
automatic resets occur frequently. After a system reset, most of the
parameters will have the values they had previously.
33720060
flash. The Digital Imaging System must be switched off separately.
Screen saver
9896 001 33152
function. If any other key on the Acquisition console is pressed, the screen is
restored and the key’s function is performed.
imaging technique, the main APR can be selected with [4a] and the sub-APR
with [4b]. The corresponding boxes will flash to indicate that the system is
busy with this function. After some seconds the box will be highlighted,
indicating that the system is ready to carry out this function. During this
period another main APR or sub-APR can be selected. If another APR or
sub-APR was already selected (box highlighted) the highlighting will be
switched off. If the imaging technique page is displayed, after pressing the
imaging technique function key [4c], keys [4a/b] can be used to activate
another imaging technique. When a new imaging technique is selected, the
last APR selected for that imaging technique becomes the current one, and
user adjustments made for that imaging technique become applicable.
Imaging technique.
Depending on the system configuration, the following Imaging Techniques
33720070
can be selected:
• digital vascular:
- lock-in mode
Exposure characteristics are calculated on the basis of on one test shot
with phototiming. kV/mA/ms is locked-in during the run.
• single sequential scan mode:
No integration of frame time periods. Frame rate 25 (30) images/
second.
• integrated sequential scan mode.
A specific number of frame time periods are integrated during one
exposure pulse to construct one image.
- bolus Chase mode:
• phototiming with kV control for frame rate 8 (10) images/second.
• cine technique (automatic kV/mA): frame rate 12.5 (15) images/
second.
9896 001 33152
When the related function key is pressed, the APR page belonging to the
selected imaging technique is displayed on the Acquisition console after a few
seconds and the current APR for this technique is automatically selected.
When an imaging technique is reselected, the last-used APR for that
technique, including any user adjustments to its parameters, becomes the
current APR.
When the imaging technique page is displayed, pressing the exposure hand
switch or foot switch will result in:
• display of the current APR page belonging to the current imaging
technique, and
• acquisition.
Subtraction.
Subtracted acquisition using this key is only possible when the digital
33720090
the injector hand switch. The contrast medium is injected when the system is
ready and the exposure run is made according to the currently-selected APR.
Note that acquisition can still be activated with the foot switch or hand
switch. If the injector hand switch is pressed during fluoroscopy, contrast
medium will be injected but acquisition will not be activated. Deactivate
one-knob acquisition when two-knob operation is required.
Use the foot switch or hand switch to start exposure and the injector hand
switch for injection. Consult the injector manual for details of how to
program and select injector parameters such as inject delay, flow rate,
volume, pressure etc.
Display stopwatch.
Press the |Display stopwatch| key to select the ‘Stopwatch’ function. The
33720110
N OT E In the cardiac mode, the stopwatch runs for longer than 10 minutes. When 10 minutes
is reached the stopwatch display returns to ‘0’.
33720130
Imaging Technique it is possible to toggle between two acquisition speeds
during digital acquisition using this key on the Acquisition console (or key
[157] on the Imaging module). This function can be performed more than
once during acquisition. The acquisition speed is displayed on the
Acquisition console. The current acquisition speed is the first acquisition
speed displayed.
Patient orientation
position of the patient on the table and must be set before exposure.
The following conditions are assumed:
• the top of the monitor corresponds to the patient’s head
• the bottom of the monitor corresponds to the patient’s feet
• the right side of the monitor corresponds to the patient’s left
• the left side of the monitor corresponds to the patient’s right.
The default setting is nose up and legs down (keys unlit). Four orientations
are possible:
• ‘Normal’: Patient on back with head at tabletop head
• ‘Legs-up’: Patient on back with feet at tabletop head
• ‘Nose-down’: Patient on stomach with head at tabletop head
• ‘Legs-up’ and ‘nose-down’: Patient on stomach with feet at tabletop head.
9896 001 33152
information is displayed for both channels. To cancel the function, press [7]
again or press the |Imaging technique| [4c] or |New patient| [4c] key. If
fluoroscopy or exposure is started while this page is displayed, it remains
displayed.
[9] Enter.
33720180
Viewing console
Viewing
monitor
Text
display
Viewing
console
Keyboard Mouse
33726410
• a keyboard, for selecting the control and status pages (keys [F1] to [F10]),
entering data, and selecting items on the screen with the cursor keys
• a text display and 8 keys for selecting functions on the control and status
pages. These pages are described in Section 4.4
• an intercom unit, for communication between the examination and the
control room.
18
16
17
15
33720200
Text display
shut down period (about 10 seconds) the light in [D] flashes and the
Viewing console cannot be switched on. At the end of this period, the digital
imaging system will be switched off.
[16] Talk.
See [17].
33724190
[17] Listen.
The intercom system consists of a microphone, speaker, volume control and
33724180
headset jackplug, together with two keys for selecting ‘listen’ or ‘talk’. These
two functions are not automatic and must be activated by the operator after
the system has been started up. ‘Talk’ is only active while key [16] is held
down. ‘Listen’ is activated when [17] is pressed. ’Listen’ remains active until
9896 001 33152
Keyboard
20
24
22
23
21
33720210
The keyboard is used to select control and status pages with the function keys
[F1] to [F10], to enter alphanumeric data and to select objects on the screen
with the cursor keys [20].
[21] Enter.
To confirm a keyboard entry or selection and to move to the next entry field.
[22] Delete.
To delete the previous character/space.
[23] Tab.
Not used.
• F6 System page: to set the date and time, to make a test image and to enter
‘Service mode’
• F7 Physician page: used to enter the physician names and codes with the
related disk partition and station name
• F8 Reserved
• F9 Status page: displays information on the progress of actions started
• F10 Help page: displays an overview of all configured control and status
pages.
An overview of the control and status pages is given in Section 4.4.
Mouse
The mouse is used to interact with the graphic elements (cursor etc.)
displayed on the monitors.
26
25
27
33720220
[25] Accept.
Click [25] to confirm a selected item or position.
[26] Action.
Click [26] to activate a selected function.
[27] Reject.
Click [27] to terminate a function.
Viewing keys
It is possible to assign selected functions to certain keys during installation
(see table at end of section). The installed system may therefore be controlled
by keys at other locations than those described here. Functions not available
9896 001 33152
on the console can (if configured) be activated with one of the Menu keys or
with keys on the tableside analysis module or viewpad. In general, the
indicator light between a key and its nameplate or symbol will be on if the
corresponding function is active. If the system is unable to perform a
Philips Medical Systems
function, the key indicator flashes until the function key is released. If the
current acquisition examination is the same as the current viewing
examination, the viewing console indicators show the functions activated
with the analysis module or viewpad in the examination room.
85 86 66 55 36 67 80 81 51 95 57 96 53 97
64 41
60 40
52
39 34
91 63
50
38
42
70 43
49
33728190
71 61 33 31 30 32 62 54 48 45 46 47 58
Vascular viewing keys
The relevant viewing functions [32, 33 and 91] are described in section 4.3.3
‘Viewing functions’.
For an overview of the viewing console functions see table at the end of this
section. The viewing functions [30 - 98] are described in section 4.3.3
‘Viewing functions’.
108
33728200
For an overview of the TAM functions, see the table at the end of this
section. The TAM functions [100 - 104] are described in section 4.3.3
Philips Medical Systems
‘Viewing functions’.
Mouse functions can also be performed with the following keys on the TAM.
[105] Accept.
As [25]. Press to confirm a selected item or position.
33720260
[106] Action.
As [26]. Press to activate a selected function.
33720270
[107] Reject.
As [27]. Press to terminate a function.
33720280
[108] Joystick
This is used to simulate mouse movements (moving the joystick moves the
cursor across the screen at a speed proportional to the displacement of the
joystick from its central position) and for functions with a variable parameter
(speed, contrast etc.), it can be used to increase or decrease the parameter
when the function is activated.
Viewpad
The Viewpad is a remote control device using Infrared Radiation (IR). The
IR transmitter is located on the front end of the Viewpad and, if obstructed,
no signals will be transmitted. The location of the IR receiver is fixed during
installation. A green light on the receiver indicates that the selected
command has been accepted. The Viewpad will function when packed in a
transparent sterile cover. During installation, some key functions can be
configured to user’s preference. The Viewpad is battery powered and the
batteries must be replaced regularly.
CAUTION Remove the batteries when the Viewpad will not be used for a long time.
WA R N I N G S • Do not use the Viewpad when more than one INTEGRIS system is in use in the same
room.
• Infrared signals from the Viewpad may interfere with other infrared-controlled
equipment in the room. Before using the Viewpad in a procedure, check that there
is no interference with other equipment.
Battery replacement
9896 001 33152
a b
c d
e
f g
h i
j k
l m
33723580
For an overview of the Viewpad functions, see the tables below. The
corresponding Viewing console functions are described in section 4.3.3
‘Viewing functions’.
Function overview
Selection
34/35 Reserved - - -
40 Run overview S C C
41 f Examination overview S C S
42/43 Reserved - - -
9896 001 33152
Processing
Philips Medical Systems
57 Position shutters S C -
Subtraction
60 a Subtraction on/off S C F
61 Move mask F - -
62 b New mask F C F
63 Average mask S - -
65 Run subtract C C -
66 104 Landmarking S S -
67 View trace S - -
Photofile/reference
72/73 Reserved - - -
Philips Medical Systems
76/77 Reserved - - -
80/81 Reserved - - -
85 Add text S - -
86 Annotation S - -
90 Flag image C C C
91 Flag run F C C
92 Flag heartbeat C C C
93 Flag examination C C -
94 Examination dump C C C
General
Legend
- Not possible - -
in which they were acquired, or in reverse sequence. The image numbers are
updated. If this function is activated when there is no current viewing run,
the first (last) image of the first run of the current viewing or acquisition
examination (not the photofile run) is displayed. If there is no run other than
the photofile run, the middle image of the photofile is displayed first. Images
are displayed with the current processing parameter settings. The image-
processing functions (contrast, brightness and edge enhancement) applied to
the current image are shown on the Viewing console. When a run boundary
is reached the key must be released and pressed again to go to the next or
previous run. If there is no next or previous run, pressing the key has no
effect.
Fast viewing
When the forward or reverse key is held down for longer than 0.5 s, the next
or previous images are displayed in the sequence in which they were acquired
at a rate of 3 images/s for acquisition speeds < 3 images/s, ECG-triggered
runs and photofile runs, or at the default speed for acquisition speeds > 3
images/s. The default speed is set during installation and can be increased or
decreased with [38] or [39]. If the requested display speed exceeds the
capability of the system, images are skipped. The maximum display speed for
the photofile is the maximum rate at which images can be displayed without
skipping. For VFR runs, the acquisition speed is taken to be the maximum
speed during acquisition of the run.
displayed. If the |Step to next run| [32] or |Step to previous run| [33] key is
pressed during the last/first run, no action is performed. If either key is held
down for more than 0.5 s, stepping is continuous at a rate of 1 step/s until
the key is released, or the start or the end of the examination has been
Philips Medical Systems
reached.
N OT E If [32+30] or [33+31] are pressed, the last or first run is displayed immediately,
skipping the runs in between.
The run information (number of images, current image and mask image) is
updated continuously. For a run acquired with subtraction, the system
displays a subtracted image constructed from the mask image selected during
acquisition and the ‘middle’ image of the run. Otherwise the ‘middle’ image
is displayed non-subtracted. For photofile runs the middle image is
displayed.
[34-35] Reserved.
The viewing speed for view forward/reverse [30, 31] and cyclic viewing [36,
33720400
33720410
flagged for transfer. Pressing [40] displays 20 images (5 x 4 matrix) of the
current run (16 images for a photofile) at lower resolution with the current
processing parameters for: Contrast, Brightness, Image inversion,
Subtraction and Run subtract. The image sequence and symbols used are as
follows:
a S
Image
c
33723590
20
f g
corners show the image number of the first/last image displayed. The
symbols / indicate the existence of a previous/next page. If the number
of images in the run is less than 20 some of the frames will be blank. A
rectangular frame marks the current image. The marker can be moved with
the |View forward| [30] and view reverse [31] keys to select another current
image. If the |View forward| [30] or |View reverse| [31] key is pressed for +
0.5 s, the marker moves at a rate of about 5 images/s. When the marker
reaches the first or last image it stops at that position. Pressing [30, 31] when
the current run marker is at the end or beginning of the overview page,
displays the next or previous overview page , if present. If [30 or 31] is
pressed before an overview page is completed, build-up of the next or
previous overview page will start. The center image of the next or previous
9896 001 33152
run can be selected with keys [32, 33]. During run overview, all image
processing functions are disabled. Pressing [40] again, displays the selected
(current) image of the current run at full-screen size. Pressing |Cycle through
run| [36] starts cycling.
Philips Medical Systems
N OT E S The time required to build up the overview display depends on the number of disks
installed and the matrix resolution.
h S
a
i
c
b
Image
d
33720440
20
g j
The top left and bottom right corners show the run number of the first/last
run displayed. The symbols / indicate the existence of a previous/next
page. If the number of runs in the examination is less than 20 some of the
frames will be blank. The current run is marked with a rectangular frame.
The marker can be moved with the forward and reverse keys [30, 31] to
select another current run. If the |view forward| [30] or |View reverse| [31]
key is pressed for + 0.5 s, the marker moves at a rate of about 5 images/s.
When the marker reaches the first or last image it stops at that position.
Pressing [30, 31] when the current run marker is at the end or beginning of
the overview page displays the next or previous overview page, if present. If
[30 or 31] is pressed before an overview page is completed, build-up of the
next or previous overview page will start.
For runs containing a selected image for transfer, the center image of that run
is marked accordingly. All image processing functions are disabled during
9896 001 33152
N OT E S The time required to build up the overview display depends on the number of disks
installed and on the matrix resolution.
33720450
Processing
During viewing, the operator can change the levels of contrast, brightness
and edge enhancement applied to the displayed images. In the case of a
single-shot image, the new levels are applied to that image only. In the case of
a run, the new levels are applied to all the images in that run, separately for
subtracted and non-subtracted images.
33720510
See [53].
33720520
N OT E S If a contrast, brightness or edge enhancement control key is held down for longer than
one second, the action is automatically repeated at a rate of 4 steps per second until
the key is released or the maximum or minimum level is reached.
33720530
setting with inverted polarity. The area outside the circle, added text,
annotations and electronic shutters remain unchanged. Pressing [54] again
restores the image to its default polarity.
Pressing [55] zooms the central part of the image by a factor of 2. The area
zoomed can be changed by holding down |Action| [26] and moving the
mouse. Clicking |Accept| [25] confirms the action, and the zoomed image
remains on the viewing monitor. Clicking |Reject| [27] cancels the action
and restores the original image. The required display area can be changed
with the mouse. It is possible to step through a run with the same zoom
factors.
N OT E Pan & zoom can also be performed with the joystick [108] on the Tableside analysis
module.
Moving the mouse or joystick to the right opens the vertical shutters, moving
it to the left closes them. The maximum setting for the shutters is 90%
closed. The shutters can be ‘locked’ in the chosen position by pressing
|Accept| [25, 105]. The functions for viewing images or masks in forward or
Philips Medical Systems
reverse sequence are still available with the same shutter settings. Pressing
[57] again or |Reject| [27, 107], or selecting another run for viewing
terminates the function.
they had when the run was viewed for the first time after acquisition.
Subtraction
33720580
no mask image has been selected, the first image of the run will be used as the
mask. Pressing [60] again cancels all activated subtraction functions and
redisplays the unsubtracted current image.
First select a new mask image and press new mask [62]. Pressing [63] loads a
new (blurred) mask image, which is obtained from the average of a
predefined number of images of the current run. The resulting subtracted
images are displayed on the Viewing monitor(s). If activated, an average
mask indicator () is displayed on the Viewing monitor(s) (bottom right).
The following pre-set number of images may be used to build the average
mask: 1-8, 10, 12, 14 or 16 (dependent on matrix size, see Section 10.2.3
‘Mask averaging’).
When a specific image is selected as new mask [62] (in the current run), the
corresponding average mask image is obtained by averaging the images that
precede the selected mask image. The maximum number of images that can
be used for averaging is given above. If this number is odd, the number used
is equal to the maximum - 1. Press [63] again to terminate the function and
return to the normal subtracted image. The scope of the function is restricted
to the current run. While the function is active, Cycle through run [36] and
View forward/reverse [30, 31] can be used to navigate through the run.
Higher/lower speed [38, 39] can be used to change the cycle speed.
33720640
two different runs (subtracted rotational angiography). The system displays a
subtracted image using the images of the current run (contrast run) and the
images of the next run (mask run) with the same image number as mask
image (contrast(n) – mask(n)). If the next run has an incompatible format
(i.e. matrix size or pixel-depth parameters), the system steps to the next run
(and so on). If no run with a compatible format can be found, the system
selects the contrast run as the mask run. The View forward/reverse [30, 31]
and Cycle through run [36] functions can be used to display other subtracted
images from two runs. When the run subtract function is active, some
functions change as follows:
Display speed
The maximum display speed for dynamic viewing in run subtract mode is at
least 2 images/s.
Status page
Only the contrast run is highlighted on display. When a run is selected in the
Status page, the run subtract function is switched off.
the mask run (‘Move mask’ inactive). If ‘Move mask’ [61] is active, then this
function does not operate.
number of the mask image, both contrast and mask-image numbers are set to
the highest image number in the new contrast run. If no compatible run can
be found in the desired direction, no step run is performed. If ‘Move mask’ is
active, it performs the same function as described above, except this time
only on the mask run. It does not affect the image number in the mask run
to be used as a mask (as this will be set to the same as the contrast image
number).
[66] Landmarking.
A subtracted image may lack necessary anatomical orientation. Landmarking
33728230
next image to the composite. Pressing |View reverse| [31] has no effect.
Pressing [67] again terminates the function. The trace image is removed and
the last image added is displayed. The behaviour of the View trace function
depends on the type of injection: Iodine or CO2. The type of injection can be
selected with |CO2 trace| [68] or on one of the Menus [95-97, 102]. Iodine
injection is the default when the system is switched on.
If CO2 injection is selected, the text ‘CO2‘ is added to the trace icon. If
Iodine trace-mode is selected, the resultant image will contain the lowest
pixel values of all processed images. If CO2 trace-mode is selected, the
resultant image will contain the highest pixel values of all the processed
images.
33723600
Photofile/reference
[71] and holding it down for more than 1 s. During this period the light next
Philips Medical Systems
to the key will flash. If the key is released during this period, the delete
function is cancelled. When the image has been deleted, the screen is blank.
Deleting photofile images does not usually affect the free image capacity or
free photofile capacity. It only affects the free capacity if the deleted image
was the last image in the photofile. The ‘Delete image from photofile’
function is not permitted if the current examination is scheduled for transfer.
[72-73] Reserved.
33720710
[75] Store reference 2.
Pressing [j (74), k (75)] on the viewpad ‘grabs’ (stores) an image displayed on
33720720
the examination monitor to the photofile. The stored image is displayed on
the reference monitor with its photofile number and the total number of
photofile images in the lower right corner.
The following functions are available:
• fluoro frame grab (during fluoroscopy)
• external video grab (e.g. ultrasound)
• move acquired image to reference monitor.
During viewing, images can also be stored in the photofile with the ‘Store
image in photofile’ function [70].
[76-77] Reserved.
[80-81] Reserved.
T-run can be hidden or displayed on the image by pressing [85]. The layout
Philips Medical Systems
of the text is the same as that of hard copies (see Section 4.5.3 ‘Hard copy
layout’).
[86] Annotate.
Alphanumeric characters can be added to the image in any position. Pressing
33720790
[86] displays an empty white square on the displayed image. Text can then be
added to the image via the keyboard. Initially, the size of the square is one
character, and it changes as characters are added or deleted, or their size is
changed.
The size of the characters can be increased or decreased before character entry
with:
|>|: 2 x enlarged
|>||>|: 4 x enlarged (keep |Shift| key pressed)
|>||>||>|: 8 x enlarged (keep |Shift| key pressed)
The size of the square changes accordingly. The scale can also be reduced
from large to small with |<|, |<||<| or |<||<||<| before character entry. The
square, including any annotation, can be moved over the image by holding
down |Action| [26] and moving the mouse. Clicking |Reject| [27] erases the
currently entered text and displays a new empty square for input. All
annotations can be erased by clicking |Reject| [27] immediately after starting
the Annotate function before any text is entered.
To accept the currently-entered text and annotation position:
• press |Accept| [25] to display a new text entry square
• press |Enter| [21] to leave the entered text displayed on the image.
The new text will be displayed on other images in the same run selected with
View forward/reverse [30, 31]. If a run boundary is crossed all text will be
removed. The function can be terminated by pressing [86] again.
N OT E S • Use a combination of horizontal and vertical arrows to draw a diagonal arrow.
• Image detail within the square is not shown because the corresponding image data
would be overlaid by the annotation. This must be borne in mind when positioning
the square and before pressing |Accept| [25] or |Enter| [21].
• The image area containing the image number etc. is not available for annotation.
displayed together with the rotation and angulation angles on the Viewing
monitor with this function. A maximum of 8 channels of physiological data
can be acquired at the same time. The number of acquisition channels and
the corresponding number of display channels are set during installation. It is
possible to step through all acquired channels by repeatedly pressing the
function key. An extra display channel is reserved for ‘no data acquired’ to
disable the display of physiological data. The timebase of the input signal is
modified so that the complete recorded sequence fits the monitor. ECG
signals are specially processed to preserve the exact location of the R-peaks. A
vertical bar indicates the position of the currently-displayed image in the run.
During viewing this bar moves horizontally across the displayed physiological
data. The function is disabled in overview mode and photofile viewing. The
physiological data are removed when an overview mode is selected or a
photofile image is displayed.
9896 001 33152
WA R N I N G A time delay of 10 to 300 ms may occur between the displayed image and the
corresponding displayed ECG signal. This delay is caused by, and is dependent on, the
Philips Medical Systems
N OT E If no physiological data have been acquired or the physio option is not installed, this
key can be used to display the rotation/angulation angles.
33720810
workstation or local archive connected to the system. Flagging an image that
has already been flagged deselects it. This function can also be activated in
’Run overview’ mode [40]. If the image results from an analytical program, it
must be stored in the photofile and flagged there.
Medical and High-Speed DICOM Image Interface are used. Only ’Flag run’
[91] can be used for transfer to CD-Medical and the High-Speed DICOM
Image Interface. If configured, pressing [93] flags all runs (including the
photofile) of the current viewing examination for transfer, if no runs were
flagged before. Pressing [93] for an examination in which run(s) (excluding
the photofile) are already flagged will unflag all runs (including the photofile)
of this examination. If the ’Flag examination’ function is activated, a ‘Flag’
indicator appears on the Viewing console text display and a Flag run
indicator is displayed during ’Run overview’ [40] or ’Examination overview’
[41]. The ’Flag examination’ function is disabled if another examination is
being transferred or an analytical function is busy.
N OT E S • Flagging cannot be carried out on a current examination which is scheduled for
transfer or being transferred.
• If an image, heartbeat, or entire run has been flagged, the run concerned will be
marked with a ‘Flag’ indication on the Status page (F9) [24].
• Automatic Flag run can be configured by Service.
• The Flag examination function [93] also affects the behavior of the Examination
dump function [94].
examination monitor and on the system output for the VCR. If no runs are
flagged for transfer (|Flag run for transfer| [91]), all runs including the
photofile will be replayed. If runs are flagged, only these are replayed. If
Philips Medical Systems
physiological data were acquired with the run, the latest selected channel is
displayed. During system installation, replay can be set to ‘Raw’ or
‘Processed’. If ‘Raw’ is configured, the images are replayed as acquired. If
‘Processed’ is configured, the images are replayed with the latest processing
parameters. The function is terminated when acquisition is activated or at
the end of the replay. During replay, all viewing functions are disabled.
N OT E The disk power supply, tape and other control settings must be switched on before
starting the examination dump.
General
[95] Menu 1.
See [97].
33720860
[96] Menu 2.
See [97].
33720870
[97] Menu 3.
33720880
33720000
When one of the menu keys is pressed, the monitor on the viewing console is
cleared and a menu is displayed. If a menu consists of more than 1 page,
‘Previous’ and/or ‘Next’ can be used to display the corresponding page. A
menu item can be selected by moving the cursor with the mouse or joystick
and pressing |Accept| [25 or 105]. After selection, the menu disappears, any
image present is displayed normally, and the selected function starts up. The
items in a menu can be configured during installation.
33720900
Viewing console [98], the TAM (if configured)[102] and from the menus
[95-97]:
33720910
• vascular quantification software package
• ventricular quantification software package
coronary quantification software package
33720920
•
• manual quantification software package
33720930
• myocardial quantification software package.
Any software package may consist of a combination of one or more specific
analytical functions. From top to bottom, the symbols shown on the left
33720940
(AVA).
For descriptions of the analytical functions, see Section 5.5.2 ‘Clinical
33720960
programs’.
TAM functions
programs’.
activated from the Viewing console [95-97]. Items in the TAM menu can be
configured by Service. All the functions described in this manual, as well as
the clinical programs, can be assigned to this menu.
[103]
Like key [55], this key is used to zoom a part of the image to full screen size.
33720540
Pressing [103] zooms the central part of the image by a factor of 2. The
zoomed region can be changed by pressing action [106] while moving the
joystick [108]. Pressing accept [105] terminates the procedure, leaving the
zoomed image displayed on the Viewing monitor. Pressing reject [107]
terminates the procedure and restores the original image. It is possible to step
through the run with the same zoom factor.
[104] Landmarking.
As [66]. A subtracted image may lack necessary anatomical orientation.
33728230
Landmarking (partial subtraction) allows a selectable degree of background
anatomy to be added to the subtracted image. To provide optimal image
information and to maintain vessel detail, the polarity of the landmark is the
opposite of that of the vessels. The Landmarking is only possible if
subtraction [60] is on. Pressing [104] displays an initial landmarking mask
over the image with a subtraction factor equal to 75% of the original image.
The subtraction factor can be increased or decreased by pressing |Action|
[106] and moving the joystick [108] up or down. The subtraction factor can
be set to any value between 0 and 100% . Images and masks can be viewed in
the forward or reverse direction with the same landmark. Pressing |Accept|
[105] terminates the function. The set subtraction factor will be applied to
all images in the run.
Viewpad functions
For a description of the Viewpad functions, see the descriptions of the
corresponding Viewing console functions in this section. The Viewing
console functions are listed in the tables at the end of Section 4.3.2 ‘Viewing
console, TAM and viewpad’.
Monitor allocation
Exam. Ref.
33728250
Monitor configuration
Restriction
9896 001 33152
Restrictions
• parallel viewing is temporarily suspended during exposures
• parallel fluoro is deactivated as soon as the exposure switch is activated
• Fluo
• Trace
• Acq • Fluo
• View • LIH
• LIH
33723620
Examination monitor Reference monitor Parallel monitor
Restrictions
• dual fluoro is deactivated as soon as the exposure switch is activated
• parallel fluoro is deactivated as soon as the exposure switch is activated
• viewing at the viewing console is temporarily suspended during exposures
• frame grabbing of images displayed on the parallel monitor is not possible,
it is only possible to grab images displaed on the examination monitor.
Monitor positioning
9896 001 33152
36280030
General
A typical feature of the XB monitors is the high brightness.
36270050
111 112 114
XB monitor
The power supply to the monitors is switched on and off with the system
power.
Reference setting
Brightness and contrast are factory set to a reference level. Pressing [112 +
113] simultaneously activates this reference setting (indicator light [110]
green).
For more information on these settings, refer to Section 5.1.4 ‘XB monitors’.
Daisy chain
For optimal viewing of the monitors it is important that they have identical
settings. The monitors in the examination room are provided with a daisy
chain facilitating a master/slave control for the user settings. Thus, when the
contrast setting on one monitor is altered, the contrast setting on the other
monitors is altered as well. Any of the installed monitors can be used as
master to control the others.
Room facilities
116
36280040
9896 001 33152
Examination light
a
117
36280050
a. upper handgrip
b. lower handgrip (detachable).
X-ray on indicator
9896 001 33152
118
Philips Medical Systems
36280040
Display unit
The display unit is located in the examination room below the monitors. The
display unit displays two title text lines and two information lines. The
information displayed depends on the actual system status.
Indicators:
• stand position (ROT, ANG, SID etc)
• II-format: 2 digits in [cm] or [inch] (depending on system configuration).
• dose rate (if configured): 5 digits in [mGycm2/s] or [cGycm2/s] in the
range of 0 - 99,999
• tube load indicator: green, orange or red light
• fluoroscopy mode indicator: ‘3’ (high), ‘2’ (normal), ‘1’ (low) or ‘4’
(reduced)
• fluoroscopy on/off indication (if Parallel viewing option is present)
• exposure/fluoroscopy ‘Action/warning’ indicator: displays ‘action’ or
‘warning’
• X-rays on indicator
• integral fluoroscopy time (if configured): 4 digits, resolution 0.1 minute
• ‘System starting’ and ‘X-ray disabled’ messages
• FLUORO, FLUORO-VIEW, FLUORO-DUAL.
33723640
122 126 127
Display unit
The audible tube load indicator can be set by Service at installation. If set,
during fluoroscopy or exposure a repeating 4-pulse (audible) signal (where
Trepeat = 2 s; tpulse = 125 ms and fpulse = 488 H) is generated by the Display unit
if the Tube load indicator [122] is showing red (Tube hot, wait).
dose is updated and also displayed on the acquisition console and on the
viewing monitor (Status page).
• COLLIS: Collision indicator
• WARNING/ACTION: to indicate that a message is being displayed on
the acquisition console
• Stopwatch time; if activated, the time is displayed as: TIME mm:ss.
Pan handle
The Pan handle can be attached to the tabletop or to the patient accessory
rail.
33721030
a b a c
Pan handle attached to accessory rail Pan handle attached to the tabletop
a. Handle to tighten the clamp and lock the pan handle to the rail or table.
b. Clamp to fit the pan handle to the accessory rail.
c. Clamp to fit the pan handle to the tabletop.
a c
b
a a
e
d
f a
g
h
9896 001 33152
33721040
Philips Medical Systems
Item Connections
a Earth (ground).
c Injector.
Item Connections
system.
N OT E S • The two cables for the two TSOs must be run through a grommet at the top-rear of
the table base.
• To provide access to the connectors, the back cover can be opened by pulling it on
the cable inlet side.
Geometry module
132 140
36280190
For more information about bodyguards refer to section 10.4 ‘Stand and
table movements’.
CAUTION If a second, identical geometry module is available, commands given on the module
connected to the upper connector in the connection box have priority over commands
given on the other geometry module.
9896 001 33152
N OT E The geometry functions described below are illustrated in Section 10.4 ‘Stand and
table movements’.
To reset the geometry system after an ‘emergency power off’ switch the
system off [B] and then on again [A].
N OT E Pressing |Power on and reset| [A], before switching the system off using [B], does not
terminate the emergency power off situation.
If the AD5 SyncraTilt option is installed and the tabletop is in a tilt position,
the longitudinal movement is motorized. To activate the motorized
longitudinal movement, press [131] down and then move it in the required
direction. If the geometry module is mounted at the foot end, the behaviour
of [131] is identical to the behaviour at doctor’s side. As the speed of the table
movement depends on the pressure applied to control [131], use gentle force
to ensure a smooth floating movement. The maximum transverse shift of the
AD5 is 2 x 18 cm (2 x 7 inch); the maximum longitudinal shift is 100 cm
(39.4 inch).
N OT E S • This control is a large push button that can be used as a handgrip when moving the
tabletop.
• For releasing and applying the brakes, the optional Pan handle has the same
functionality as this push button but it does not control motor speed.
longitudinal scan under manual control. This switch only operates when
[131] is not activated.
If the AD5 SyncraTilt option is installed and the tabletop is in a tilt position,
pressing switch [132] once, keeps the transverse brake on when [131] is used.
Only longitudinal table movements are then possible.
Control knob [133] is used for motorized table height adjustment. To raise
33721090
the table, push the control up (maximum height 104 cm/40.9 inch). To
lower the table, pull the control down (minimum height AD5: 76 cm/29.9
inch). The rotational brakes are released while the table moves up.
N OT E S • If the Pivot option is installed, the height-range of the table is: 78.5 - 106.5 cm (30.9
- 41.9 inch).
• If the Swivel option is installed, the height-range is 83 - 113.5 cm (32.6 - 44.6 inch).
33721100
axis of the X-ray beam.
Joystick up: SID increases (II away from patient).
Joystick down: SID decreases (II towards patient).
The actual SID is displayed on the Display unit [124].
[135] Reserved.
+ - + -
[137-138] Reserved.
N OT E Keys [139] and [140] are only available if the AD5 table is fitted with the ‘SyncraTilt’
option.
Up Down
33722690
which safeguards patient and equipment. If either the |Table tilt movement|
[139] or |Table height adjustment| [133] joystick is then used, an audible
signal warns the operator of the increased collision risk during a tilt or height
adjustment.
[141] Reserved.
SID to the values stored with |Store position 1| [142] or the pre-defined
rotational angiography end position. Pressing the right-hand side recalls the
values set with |Store position 2| [143] or the pre-defined rotational
33720750
N OT E The L-arm can also be moved manually after releasing the brakes by pressing |L-arm
rotation brake release| [190] and/or |Longitudinal movement brake release| [191] on
the L-arm.
Imaging module
150b
150c 153
33723670
158 155 156
CAUTION If a second, identical Imaging module is available, commands given on the module that
is connected to the upper connector in the connection box have priority over
commands given on the other Imaging module.
b
b
33722750
Joystick up: Vertical shutters [a] open.
Joystick down: Vertical shutters [a] close.
Joystick to the right: Horizontal shutters [b] open.
Joystick to the left: Horizontal shutters [b] close.
Pressing the control down resets the shutter positions.
[152] Reserved.
selects the next smaller field size. Pressing the right side of the key selects the
next larger field size. The field size selected is shown on the display unit
[125]. The default Image Intensifier mode (small, medium, large or other
33722770
[154] Reserved.
Joystick movements for the right wedge filter are the opposite of the
33728253
Variable frame rate (VFR) is the same function as [4c] on the Acquisition
console. During digital vascular acquisition it is possible to toggle between 2
acquisition speeds if VFR values are included in the chosen APR program.
The function can be performed more than once during acquisition. The
acquisition speed is displayed on the Acquisition console. The current
acquisition speed is the first acquisition speed (V2) displayed. If the parallel
fluoro or dual fluoro option is installed, the VFR key [157] on the imaging
module will serve to toggle between the normal fluoroscopy mode and the
parallel fluoroscopy mode.
161a
161b
161c
164
165
36270040
APC module
To reset the geometry system after an emergency power off, switch the system
off [B] and then on again [A].
N OT E Pressing |Power on and reset| [A], before switching the system off using [B], does not
terminate the emergency power off situation.
b: Sequencer mode.
33722890
c: Reference mode.
When a mode is selected the indicator next to the key switches on and the
33722900
Scratch mode
Two scratch stand positions can be stored and recalled. When scratch mode
is re-entered the previously stored positions are available provided that the
system function ‘new patient’ has not been activated.
displayed on the Display unit. When the right side of the button is pressed,
9896 001 33152
pressed until the stand stops moving (rotation and angulation). Thereafter,
Philips Medical Systems
the II will move to the stored SID unless the movement is stopped by
releasing key [165]. The operator is responsible for avoiding collisions. When
the movement is complete, the other stored position is selected and
displayed.
Sequencer mode
Stand positions from one or more service installed sequences can be recalled.
When only one sequence is available, the user function ‘select position’
becomes immediately active without displaying the sequence identification.
the right side selects the next position . The current position is displayed on
the display unit. By repeatedly pressing this key, all sequence positions can be
displayed.
pressed until the stand stops moving (rotation and angulation). Thereafter,
the II will move to the stored SID unless the movement is stopped by
releasing key [165]. The operator is responsible for avoiding collisions.
Reference mode
Allows reproduction of the stand position corresponding to the image
displayed on the reference monitor.
geometry is selected. When the right side of the button is pressed, the right
displayed reference geometry is selected. The selected position is displayed on
the display unit.
Acc
accepted by keeping [165] pressed until the stand(s) stop(s) moving (rotation
and angulation). Thereafter, the II will move to the stored SID unless the
movement is stopped by releasing key [165]. The operator is responsible for
avoiding collisions.
9896 001 33152
For angiography of the lower peripherals images are acquired as the patient is
moved in one continuous movement at a speed that matches the blood flow
in the peripheral vessels (bolus chasing). This can be achieved automatically
or interactively by the operator.
[170-171] Reserved.
172
33722950
[172] Speed control hand switch.
With interactive mode Bolus Chase the speed control hand switch is
automatically enabled when the end and start positions have been selected. If
a run is interrupted (automatic and interactive mode), the speed control
hand switch can also be used to return to the start position.
The speed control hand switch is used in interactive Bolus Chase mode to
control the scan movement. Pressing [172] starts the tabletop scan
movement. The speed is controlled by the degree to which the button is
pressed. The maximum average speed is about 15 cm/s (5.9 inch/s). As soon
as the tabletop has been moved to its end position, the movement stops and
the speed control must be released. Pressing the speed control again reverses
the scan movement. Acquisition is independent of scan direction. The
system ensures smooth starting and stopping of each movement to avoid
jolting the patient.
179 177
175
33726480
[175] Fluoroscopy.
33728040
Pressing this pedal initiates radiation. Radiation continues until the pedal is
released.
[176] Reserved.
[178] Reserved.
[179] Exposure.
33723030
Pressing this pedal initiates preparation of the X-ray tube and exposure.
Releasing the pedal ends the exposure.
185
33723050
[185] Preparation/exposure.
The two-level pushbutton on this hand switch has two functions:
1 Preparation of the X-ray tube; pressing the small button until it is flush
Philips Medical Systems
190
191
36280200
Handgrips on the L-arm (both sides)
When the brakes are released by pressing and holding down [190], the L-arm
can be moved manually with either of the two handgrips. When the L-arm is
in position, releasing [190] applies the brakes.
When the brakes are released by pressing and holding down [191], the L-arm
can be moved manually with either of the two handgrips. When the L-arm is
in position, releasing [191] applies the brakes.
N OT E Manual movements override motorized movements.
The drive accepts both erasable and non-erasable 5.25-inch disks. Optical
disks are mounted in cartridges. Images cannot be stored on a write-
protected disk. Each disk has a storage capacity of 620 Mb, sufficient for
approximately 2 x 150 images (10242) or 2 x 600 images (5122 and 1024 x
Philips Medical Systems
512). New disks must be formatted before use (Format function on the
Optical disk page). This takes about 10 minutes. Saving and retrieving
images using an optical disk are background processes. This means that the
system remains available for acquisition and viewing while images are being
saved or retrieved.
33728251
215 216 219 a 218
Optical disk drive
[225-233] Reserved.
9896 001 33152
b
c
33723860
200X Scanner
Item Part
b Probe
c Control panel
d Monitors
Control panel
243 236
237
9896 001 33152
244 239
33723890
Philips Medical Systems
The control panel for the scanner is attached to a movable arm under the
monitors. It can be moved into position for an examination and then stored
back under the monitors after use.
33720050
The indicator lamp above the probe connection on the scanner also comes
on. The examination monitor displays the ultrasound image automatically.
[G] System off.
After switching off the scanner, the examination monitor is switched back to
33720060
normal mode. The indicator lamp above the probe connection on the
scanner goes out.
N OT E The scanner will not start up if it is switched on again immediately after switching off.
To start the scanner, switch it off, wait 10 seconds and then switch it on again.
Two probes (single or dual frequency) may be connected to the scanner at the
same time. Press the key repeatedly to select the required probe and
frequency. The selected probe and frequency are displayed in the status field
on the examination monitor. The green light on the 3.5/5.0 MHz and 3.5
MHz high-definition curved array probe indicates the orientation of the scan
plane. In normal operation the position of the green light on the probe
corresponds to the left side of the image on the monitor.
Switches the ultrasound image to the examination monitor. The key lights
when the monitor is displaying ultrasound images. To return to the X-ray
display mode, initiate fluoroscopy [175, 178, 180], acquisition [179, 185] or
select a viewpad function [a-m]. Once the |Examination monitor| key [236]
is pressed it remains illuminated.
N OT E After switching the ultrasound system on, the examination monitor is automatically
selected for ultrasound display.
Not used.
This key, in conjunction with the joystick [247], is used to measure the
distance (in millimeters) between two points in the image. The procedure is
described in Section 5.3.5 ‘Apply acquisition functions as required’.
[239] Freeze/Unfreeze.
Press [239] once to freeze the current image. Press [239] again to return the
36271210
monitor to live mode. If the image is frozen, ‘F’ is highlighted in the image
status field.
[240] Biopsy line on/off.
Press [240] once to display a biopsy line (for the selected probe) in a non-
36271220
N OT E Pressing |Biopsy line on/off| [240] when the image is frozen has no immediate effect.
The biopsy line is only displayed after the image has been unfrozen.
33723990
the display to normal. The status field indicates the image polarity.
N OT E Pressing |Reverse on/off| [241] when the image is frozen has no immediate effect. The
image is only reversed after the image has been unfrozen.
changed with keys [242, 243]. The upper and lower limits of the image field
depend on the probe type and frequency. A lower frequency gives a larger
image field.
give the entire ultrasound image a consistent brightness. The normal setting
is with all of the gain and depth controls set to their center position. For
certain applications, the normal position is not desirable and the brightness
level of the ultrasound image can be adjusted with controls [244 to 246].
N OT E Always try to optimize the image using the scanner’s gain controls; the monitor
controls should not be changed.
[247] Joystick.
The joystick is used to move the on-screen measuring caliper to the desired
measuring positions.
9896 001 33152
Philips Medical Systems
F1 Schedule page
F2 Review page
F3 1)
F4 1)
F5 1)
F6 System page
F7 Physician page
F8
F9 Status Page
F10 Help page
33725140
• F1 Schedule page: used to enter and select scheduled examinations for
acquisition, also used to perform a query of the work list manager and
enable the display of more information about a scheduled patient record.
• F2 Review page: used to select acquired examinations for viewing and
enable the display of more information about an examination.
• F3-F5 Function (1)) depends on system configuration, function keys [F3
to F5] can be configured to access the Transfer page, and the Transfer page
can be configured to list all possible archiving systems. The remaining
function keys can be configured to access separate archiving system pages.
Configuration is carried out by Service. The following configurations are
possible for keyboard function keys [F3 toF5]:
- Report page: enables the user to report a selected examination to a
connected information system
- Export page: lists all examinations flagged for transfer to a configured
export station.
- ODR page: used to control all actions of an optical disk recorder
- Copy page: used to copy photofiles from completed examinations to a
remote HCU
9896 001 33152
• F7 Physician page: used to enter the physician names and codes with the
related disk partition and workstation address.
• F8 Reserved.
• F9 Status page: displays information on the progress of actions started.
• F10 Help page: displays an overview of all configured control and status
pages.
Text display
When a page containing examination data is selected, the page with the
current examination is displayed. When no current examination exists, the
first page (if there are several) is displayed. To select other pages, use the
Previous or Next function. The current acquisition examination is displayed
separately on the last line of the last page. The current examination and
active function are highlighted. To select another examination, use the
|Cursor up| or |Cursor down| keys [20, 26] and confirm the selection by
pressing |Enter| [21, 27]. Help and error messages are displayed where
applicable.
36120060
Select Add Delete Modify Query More Previous Next
The ‘Prot’ (Protection) field indicates that the patient record is protected
against modification and/or deletion. The displayed protection value can be
one of the following:
• ‘N’: not protected
Philips Medical Systems
N OT E For each protected patient record, the only information which can be changed is the
physician code.
Entries under ‘Name’ and ‘Birthdate’ are mandatory. All others can be
skipped. When the ‘Sex’ field is skipped, the system automatically enters an
‘M’ in this field. The help and error lines give instructions and error messages
to the operator.
The function line shows the selectable functions, which are (from left to
right):
• Select: select a new patient for acquisition
• Add: enter data for new examinations
• Delete: delete an examination
• Modify: change patient or examination data
• Query: start a query with the connected information system (work list
manager) by displaying the ‘Work List Query Data Entry Screen’
• More: give more information about a patient record by displaying the
‘Information Screen’
• Previous: displays previous schedule page
• Next: displays next schedule page.
36120070
The ‘Work List Query Data Entry Screen’ is used to enter data in order to
perform a query of the work list Manager. This screen appears following
selection of the ‘Query’ function on the ‘Schedule page’.
The following data fields are the minimum that must be completed in order
to perform a patient based query:
• Patient Name (maximum 32 characters)
• Patient ID (maximum 32 characters)
• Accession Number (maximum 16 characters)
9896 001 33152
Information screen
36120080
36120080
The ‘Information screen’ is used to display additional information about the
selected scheduled patient record or acquisition examination. This screen
appears following selection of the ‘More’ function on the ‘Schedule page’.
The following functions can be selected from the function line (from left to
right):
• Modify: modify scheduled patient record information (this is only possible
if the examination is not protected)
• Return: to cancel the Information Screen and return to the main Schedule
page.
For unprotected scheduled patient records, the following information can be
modified:
• Patient Name (maximum 32 characters)
• Patient ID (maximum 32 characters)
• Patient Birthdate (format configured during installation, e.g.: dd-mm-
yyyy - 10 characters)
• Patient Sex (1 character)
• Accession number (maximum 16 characters)
• Req. Procedure ID (maximum 16 characters)
• Procedure Step ID (maximum 16 characters).
9896 001 33152
Philips Medical Systems
36120050
The Review page or pages give an overview of all completed examinations
with the number of acquired runs and images. An examination can be
selected from this page for viewing, amendment or deletion.
The ‘Prot’ (Protection) field indicates that the examination is protected
against modification and/or deletion. The displayed protection value can be
one of the following:
• ‘N’: not protected
• ‘M’: protected against modification
• ‘D’: protected against deletion
• ‘Y’: protected against modification and deletion.
If a examination originates from a Radiology or Hospital Information System
(RIS or HIS), the ‘Prot’ field contains a ‘Y’.
For a completed, unprotected, examination, the following data can be
amended:
• Patient name
• Birth date
• Patient sex
• Patient ID
• Physician code.
The following functions can be selected from the function line (from left to
right):
• Select: select examination for viewing
• Delete: delete examination
9896 001 33152
Information screen
36120090
The ‘Information Screen’ is used to display additional information about the
selected scheduled patient record or acquisition examination. This screen
appears following selection of the ‘More’ function on the ‘Schedule page’.
The following functions can be selected from the function line (from left to
right):
• Modify: modify scheduled patient record information (this is only possible
if the examination is not protected)
• Return: to cancel the Information Screen and return to the main Schedule
page.
For unprotected scheduled patient records, the following information can be
modified:
• Patient Name (maximum 32 characters)
• Patient ID (maximum 32 characters)
• Patient Birthdate (format configured during installation, e.g.: dd-mm-
yyyy - 10 characters)
• Patient Sex (1 character)
• Accession number (maximum 16 characters)
• Req. Procedure ID (maximum 16 characters)
• Procedure Step ID (maximum 16 characters).
9896 001 33152
Philips Medical Systems
Report page
36120020
Report Cancel Disk A/B Previous Next
The following functions can be selected from the function line (from left to
right):
• Report: send examination report to the connected information system
• Cancel: cancels current transfer
• Disk A/B: select disk partition in multidisk systems (if configured)
• Previous: display previous report page
• Next: display next report page.
9896 001 33152
Philips Medical Systems
Export page
36120010
Save Cancel Disk A/B Previous Next
The Export page or pages display all examinations flagged for transfer,
including, if applicable, an examination with the status ’sending’. If there is
no examination with the status ’sending’ then the pct field is determined. For
each examination the patient attributes derived from the scheduled patient
record are shown. The default examination to be transferred will be the
acquisition examination. If more flagged examinations exist than can fit on
one page, then that page will be shown that contains the current acquisition
examination.
For a completed examination, the following data can be amended:
• Patient name
• Birth date
• Patient sex
• Patient ID
• Physician code
• Run
• Pct (percent)
• Status.
The status can be:
• scheduled: the flagged part of the examination is scheduled for transfer to
the cache buffer
• busy: the flagged part of the examination is being transferred to the cache
buffer
• sending: the flagged part of the examination is being transferred from the
cache buffer to the selected destination
• closing: the examination has been transferred to the selected destination
9896 001 33152
The following functions can be selected from the function line (from left to
right):
• Save: send selected images of the selected examination to the selected
destination
ODR page
36120040
Save Cancel Retrieve Format Disk A/B Previous Next
Copy page
36120030
Copy Cancel Disk A/B Previous Next
Transfer page
33725210
REPORT EXPORT ODR COPY
The Transfer Page gives an overview of all accessible archiving systems. The
archiving page is displayed when a system is selected.
If configured, the following can be selected from the function line (from left
to right):
• REPORT: to select the report page
• EXPORT: to select the export page
• ODR: to select the ODR page
• COPY: to select the copy page
33725120
The System page is used to set the time and date. Thereafter, the date and
time are maintained by the system. The date and time appear on the header
line of each page and are stored in the file for each examination.
Philips Medical Systems
The following functions can be selected from the function line (from left to
right):
• Set time: the INTEGRIS system time is displayed and can be set (the time
format is hh:mm)
• Set date: the INTEGRIS system date can be set (the format is configurable
during installation)
• Service: gives access to the system for service purposes
• Test img: displays a test image intended for HCU calibration
36120100
Add Delete Modify Previous Next
The Physician page can contain up to 64 physician codes with the related
disk partitions and network addresses, for use with the Schedule page. A
physician code (max. 4 characters) is an abbreviation of the physician’s name
(max. 26 characters). The code can be used on the Schedule page but the full
name of the physician will appear on films.
The following functions can be selected from the function line (from left to
right):
• Add: to enter a physician’s name, code, disk partition and network address
• Delete: to delete a physician’s name, code, disk partition and network
address
• Modify: to modify a physician’s name, code, disk partition and network
address.
• Previous: display previous physician page
• Next: display next physician page.
The ‘Prot’ (Protection) field indicates that the physician record is protected
against modification. The displayed protection value can be one of the
following:
• ‘N’: not protected
• ‘M’: protected against modification.
The disk partition is only available if a disk partition is configured, and a
9896 001 33152
33725450
Select Modify Previous Next Select Modify Previous Next
The Status page gives information on the status of the current acquisition
and viewing examinations. The status and other relevant information on the
current acquisition examination, as listed in the Schedule page, are displayed
on the left of the page. The status of the current viewing examination, as
listed in the Review page, is displayed on the right hand side, but only when
this examination is not the current acquisition examination. The information
on this page is updated automatically.
Examinations can have the following status:
• Current: currently selected for acquisition
• Examined: acquisition complete
• Scheduled: selected for archiving
• Busy: transfer in progress
• Retrieved: successfully retrieved from ODR disk
• Done: transfer successfully completed
• Error: an error has occurred that cannot be resolved by the operator
• Warning: an error has occurred that can be resolved by the operator
• Cancelled: transfer cancelled.
36271230
Digital dynamic (alter APR activated)
Focus : small
kV :
Focus : Small
kV :
36271270
01:23
Tube warm
33723160
4.5.2 Monitors
Examination/Viewing Monitor
General
• if the parallel viewing option is not installed the monitor layout is the
9896 001 33152
monitor.
- when the system is busy with another viewing examination (= parallel
viewing):
• the examination monitor shows viewing and acquisition related
information from the acquisition examination.
• the viewing monitor shows viewing related information from the
viewing examination and the viewing mode frozen icon (e.g. during
acquisition, parallel viewing etc.).
a e
f
c
33724070
d g
[a] Patient data and examination date (only if the ‘Add text’ function [85] is
activated):
Patient name, patient birth date and sex, registration number and
examination date.
[d] T-mask:
99.999
Additional examination information:
T-image:
(only if the ‘Add text’ function [64] is activated)
33723280
99.999
T-run:
HH:MM:SS
T-mask (99.99): Time between the current mask image and the first image in
the run.
T-image (99.99): Time between the current image and the first image in the
run.
T-run (HH:MM:SS): Time that the run was acquired.
[e] Hospital name and physician name (only if the ‘Add text’ function [85] is
activated):
Hospital name (2 lines); Physician name (1 line).
CO2.
9896 001 33152
CO2
Photo.
Philips Medical Systems
n1
PHOTO Only for photofile images.
33723300
n2
n1: Total number of images in photofile.
n2: Current photofile image number.
Viewing mode frozen.
Only if the ‘Parallel viewing’ option is installed and the Dual fluoroscopy or
33724220
ROT
Geometry angles.
90
33724080
ANG
60
RUN
Image information.
nl
n2
33723310
MASK IMAGE
n3 n4
Run
Not for photofile images. ‘Run’ appears twice in Run-subtract viewing mode.
n1: Current run number.
n2: Total number of images in the run (or pairs of images, for biplane
operation ). Maximum three characters.
Mask
n3: Current mask (pair) number (maximum of three characters).
Image
n4: Current image number (or the image pair number, for biplane operation)
(maximum of three characters).
33724075
Trace image.
Only if Trace subtract fluoroscopy [158] is activated.
33723190
Stopwatch.
Only if ‘stopwatch’ [4c] is activated.
33723230
M:SS
Time left.
10 seconds left for acquisition before cyclic overwriting of the acquisition
33723240
examination.
[a]
n1
Previous overview page.
Only if ‘overview’ [40, 41] is activated and the previous overview page
33723270
contains images.
n1: image number of the first image displayed.
[g] Next overview page.
Only if ‘overview’ [40, 41] is activated and the next overview page contains
33723320
n1
images.
n1 : image number of the last image displayed.
Reference monitor
33724090
b
View reference.
9896 001 33152
[b]
Displayed if viewpad function is activated.
33724100
Philips Medical Systems
n1
Photo.
PHOTO Only for photofile images.
33723300
n2
n1: Total number of images in photofile.
n2: Current photofile image number.
c
a
33724110
[a] Patient and examination data:
Patient name, patient birth date and sex, registration number and
examination date.
Image field
Status
field
Scale
format Results
field
33724115
During acquisition
During acquisition the ultrasound image is displayed on the examination
monitor as follows:
Item Explanation
F freeze (highlighted)/Unfreeze
R Reverse off.
4.6 Accessories
33724350
Rail accessory clamps Tabletop accessory clamps
33726020
Catheterization arm support Mattress
33726030
Neuro head rest Drip stand
33726040
N OT E Sterile disposable covers are not supplied and must be obtained locally.
33726060
Head fixing aids Ratchet compressor
36280060
Table mounted radiation shield Ceiling suspended radiation shield
36280070
Pan handle Examination light
36270110
9896 001 33152
Cable guides
For the optional table tilt accessories, see section 10.4.4 ‘Angio DIAGNOST
5 with SyncraTilt option’.
Philips Medical Systems
WA R N I N G S Safety awareness
Do not start up the INTEGRIS Allura Monoplane system unless you and all other
operators present have read, fully understood and know all the safety information and
emergency procedures given in Section 2, ‘Safety’ of this Manual. Operation of the
INTEGRIS Allura Monoplane system without having read, understood and learnt ALL
the safety information and procedures in Section 2, ‘Safety’ could lead to fatal or
serious injury. It could also result in clinical misdiagnosis/incorrect treatment.
You can find information about the ‘User routine checks program’ and the
‘Planned maintenance program’ in the ’Maintenance’ section of this manual.
Operator knowledge
Do not operate the INTEGRIS Allura Monoplane system with patients unless you have
a good understanding of its capabilities and functions. Using this equipment without
such an understanding may compromise its effectiveness and/or prejudice the safety of
the patient, yourself and others.
You are strongly recommended to read this manual before using the
9896 001 33152
determine the functioning of the system and the display of images can be
customized by Service to suit user requirements. Some of these parameters
can be overridden by the operator before an examination or during viewing.
Preparing the system Switching the system on/off How to switch the system on
for diagnostic imaging
(§ 5.0.1) How to switch the system off
X-ray filters -
acquisition acquisition
Transporting the - -
system (§ 5.0.7)
Conditions
CAUTIONS • To prevent malfunction, do not touch any key during the Switch-on process.
• Only switch off the X-ray system at a local mains power switch after pressing the
|Power off| [B] and digital imaging system off [D] keys on the acquisition and
9896 001 33152
viewing consoles.
1 Check that it is safe to switch the system off (i.e. no patient on the table).
2 Press |Power off| [B] on the acquisition console.
WA R N I N G S • Do not use the INTEGRIS Allura Monoplane for any medical application unless you
are certain that the Planned Maintenance Programme is up to date. See Section 7
‘Maintenance’ of these Instructions for use. If any part of the INTEGRIS Allura
Monoplane is known (or suspected) to be defective, DO NOT USE IT until a repair
has been made. Operation of the INTEGRIS Allura Monoplane with defective
components may expose the operator or a patient to radiation or other safety
hazards. This, in turn, could lead to fatal or other serious personal injury. It could
also lead to clinical misdiagnosis.
• Do not operate the INTEGRIS Allura Monoplane with patients unless you have a
good understanding of its capabilities and facilities. Using this equipment without
such an understanding may compromise its effectiveness and/or reduce the safety of
the patient, yourself and others.
Conditions
CAUTION Do NOT position the geometry module at the foot end of the table if the (optional) tilt
function is to be used.
33721020
Cable guides
A set of cable guides is supplied with the patient support. The cable guides
are fitted to the patient support accessory rail in order to support the cables
of the Tableside Operation Modules (TSO).
Conditions
When using the cable guides, they should be positioned at intervals along the
accessory rail, ensuring that enough cable guides are used to adequately
support the TSO cables.
36270110
(hook) portion of the cable guides.
Pan handle
The pan handle is used to release the tabletop brakes. Additionally the pan
handle can be used for floating the tabletop.
The pan handle can be configured for two modes of operation, such that:
• when pressed and released, the tabletop brakes are deactivated (table float
activated); when pressed and released a second time, the brakes are
reactivated.
• when pressed and held down, the tabletop brakes are deactivated; releasing
the pan handle reactivates the brakes.
Conditions
No special conditions apply.
9896 001 33152
Philips Medical Systems
33721030
a b a c
Pan handle attached to accessory rail Pan handle attached to the tabletop
1 Fit clamp [b] (not visible) to the accessory rail, positioning the rail inside the
clamp groves.
2 Turn handle [a] to tighten the clamps and lock the pan handle to the rail.
WA R N I N G If the foot switch is to be fitted with a sterile cover, it is important to take care that the
cover is not fitted too tightly. This is to ensure that when one pedal is pressed the cover
does not activate another pedal(s).
If using a sterile cover, be sure the original package of the sterile cover is not
damaged and the date is not expired.
In order to create a sterile environment, all system parts in the working area
must have a sterile cover. The exception to this is the examination light lower
handgrip, which can be replaced by a sterilized handgrip.
Sterile disposable covers are not supplied and must be obtained locally.
Conditions
A thin sheet of sterilized plastic is placed over the tabletop, TSOs and pan
handle.
The viewpad(s), radiation shield and image intensifier must be covered
separately with a sterile cover.
1 Depress the handgrip locating button and remove the handgrip from the
examination light.
9896 001 33152
Philips Medical Systems
36270120
2 Fit a sterilized handgrip, ensuring that the locating button clicks into place.
General
The Allura system is equipped with the Bodyguard object sensing system on
critical locations of the X-ray system. The system will protect the patient by
slowing down movement speeds when an object is detected within a certain
safety distance. The detection system does not prevent all collisions, but due
to the reduced movement speeds these collisions will not be harmful if they
occur.
We explicitly state that it is the responsibility of the user to prevent collisions
of the Allura system with any object.
In case of emergency the C-arm stand can be parked, providing complete free
all-round access to the patient. The floating tabletop movements allow the
patient to be removed from the X-ray beam with maximum speed and ease.
WA R N I N G S Emergencies:
• In the event of a system movement emergency (movement of the system without
any user action), press emergency power off [130] or [160] to block all movements
except manual L-arm rotation, longitudinal movement and the tabletop float.
• In the event of a clinical emergency involving a patient, do not press emergency
power off [130] or [160] or switch off the X-ray system using a local mains power-
off switch. Move the image intensifier away (either motor driven or manually, by
firmly pushing it up) and/or the tabletop, and/or rotate or park the stand to provide
clear, all-round access to the patient.
Patient safety:
• When moving the Image Intensifier towards the patient; take care that the Image
Intensifier front plate does not hit small objects.
• If during motorized movements an audible tone (beeping sound) is produced by the
system, this indicates that the bodyguard detection system does not function
properly. For safety reasons the system will automatically switch to a reduced speed
mode. Be aware that in this mode collisions can occur. Call Service.
Conditions
The following procedure describes the conditions required prior to lower
peripheral examinations in general.
1 Switch the system off by pressing |Power off| [B] on the Acquisition console.
9896 001 33152
2 Switch the system on by pressing |Power on and reset| [A] on the Acquisition
console.
Philips Medical Systems
36280290
Preferred setting
Preferred procedure
Alternative procedure
For reversed bolus chasing the procedure is the same except for procedural
steps 5 and 6:
9896 001 33152
5 Position the patient in the supine position with the feet as close as possible to
the edge of the tabletop (opposite the pedestal).
6 Place the tabletop in the fully extended position towards the C-arm such that
the II is above the bifurcation.
Philips Medical Systems
X-ray filters
CAUTION Careful positioning of the filters is essential for good image quality, especially in
peripheral procedures. The center filter must be placed as high up as possible between
the patient’s legs. The side filters must be molded to the shape of the legs so that there
are no gaps between the filters and the legs. If necessary, immobilize the patient’s legs
with restriction straps.
A set of long wedge filters is provided. The center filter is placed between the
patient’s legs with the wide part towards the feet and the two side filters are
placed as close as possible to the left and right of the legs respectively. The top
edge of the center filter is provided with marks to facilitate measurements.
Each mark has a diameter of about 5 mm at a distance of about 5 cm (2
inch).
33724540
Patient positioning
33724380
WA R N I N G S If the Pivot option is available:
• When transferring a patient, make sure that the tabletop is fully retracted towards
the foot end and that the tabletop pivot is locked in position.
• The tabletop is held in place in a groove in the locked position and can be moved if
pressure is applied. This may cause unexpected movement during patient transfer
and possible injury to the patient.
• Make sure there are always two people available (one at each side of the patient)
to transfer the patient on/off the table.
9896 001 33152
Conditions
Before positioning the patient on the table make sure that:
• the stand is in its parked position or in the P-A projection
Philips Medical Systems
• the tabletop is moved all the way towards the foot end (fully retracted).
The tabletop is then freely accessible from both sides allowing transfer of the
patient from a bed to the tabletop. If a Pivot device is available, the table can
be pivoted to increase accessibility. To pivot the tabletop, push it in the
required direction. The tabletop should be pushed at least 1.5 m (59.0 inch)
from the pivot point with the tabletop brakes on.
36280140
Liebel Flarsheim Medrad
Conditions
These condition apply only if two knob aquisition is selected.
The injector is disabled if the acquisition hand switch or foot switch is
activated before the injector hand switch. The injector is disabled until the
X-ray preparation time, plus any programmed (APR) time has expired.
9896 001 33152
Conditions
For correct patient identification at least the following information has to be
entered on the schedule page:
• patient’s name
• birth date
• sex.
36120060
The patient’s sex and the physician code can be selected from a list, by using
the |Cursor up| or |Cursor down| key [20] when in the appropriate field.
Directions will be given on the help line of the text display when the cursor is
in the selected field. One of the items is a blank line, which can be filled in by
typing a new item and then pressing |Enter| [21].
6 Enter the patient’s ID.
The cursor will move to the ‘Phys’ field.
7 Enter the physician code.
If a defined physician code is entered, the related physician name will be
added to the scheduled patient record attributes. If a non-defined physician
code (no corresponding physician record present) is entered, then instead of
the physician name, the entered physician code will be displayed on the
digital or high line rate hard copy output(s).
8 Enter the scheduled procedure time.
The scheduled procedure time, by default, will be the current system time
and will be filled in automatically, however it can be changed
9 To enter more examinations, activate ‘Add’ again and repeat procedural steps
3 to 7.
Conditions
• When there are no patients on the Schedule page and ‘More’ is selected the
message ‘No patient information’ is displayed on the error line of the
Schedule page.
• When there are no examinations on the Review page and ‘More’ is selected
the message ‘No examination information’ is displayed on the error line of
the Review page.
• It is only possible to modify information of scheduled patients in the
‘Information screen’ that has been entered manually.
9896 001 33152
1 Press F1 or F2 [24].
The Schedule or Review page is displayed.
2 Activate the ‘More’ function [15].
3 Use the |Cursor up/down| keys [20] on the keyboard to select the required
patient/examination.
4 Press |Enter| [21] to confirm the selection.
36120080
36120080
For manually entered patient information it is possible to modify some of the
fields. The following fields can be modified:
• Patient Name, ID, Birthdate and Sex
• Accession number
• Req. Procedure ID (Requested procedure ID)
• Procedure Step ID.
Fields that are protected against modification will be skipped automatically.
5 Press the |Modify| key.
6 Use the |Cursor up/down| keys [20] on the keyboard to step through the
different fields, making the modifications in each field as required.
7 Pressing |Enter| [21] on the last entry field (Station AE title) will save the
modified information.
8 If the modifications are not required, press the |Modify| key again, or press
|Return|.
All modifications are lost and the Information screen shows the original
values.
9 Press |Return| to cancel the ‘Information Screen’ and return to the Schedule
or Review page.
before the system was turned off. If no current acquisition examination was
present before power down, a new examination (‘No name entered’) is added
and selected as current acquisition examination. Fluoroscopy is now possible.
This temporary examination is removed when a new acquisition examination
Philips Medical Systems
Conditions
• If the Viewing console is not available or is switched off, press |New
patient| [4c] on the Acquisition console.
• If the Viewing console is available, it is not possible to select a new patient
from the Acquisition console.
2 Use the |Cursor up| and/or |Cursor down| keys [20] to select a scheduled
acquisition examination.
3 Confirm the selection by pressing |Enter| [21].
The name of the patient will be displayed on the Reference monitor until
another action that uses the monitor is performed. The Status page will
appear on the Viewing console text display. The examination time is the time
at which the examination is selected for acquisition.
After selection, the system checks if a photofile of the configured length (set
by Service during installation) can be created without deleting images. If this
is not the case, the prompt ‘No sufficient photofile space, continue (Yes/No)’
is displayed (default answer is ‘Yes’). If ’No’ is entered, the selection
procedure is stopped. Otherwise, the selection procedure continues, creating
a photofile as large as the remaining free storage space allows.
All newly acquired images are added to the disk partition corresponding to
the physician code of the current acquisition examination.
Fluoroscopy
Fluoroscopy is generation of images at low dose rates.
Conditions
If an error occurs, fluoroscopy will not start. When the error disappears, the
footswitch has to be released and pressed again.
Low level fluoroscopy is still possible after an emergency power off.
Before performing fluoroscopy, the following functions can be set or
modified:
• patient orientation [5, 6]
• image intensifier field size [153, 154]
• shutter adjustment [151, 152]
• wedge filter [155, 156]
• stopwatch [4c].
9896 001 33152
N OT E With only one imaging mode available at any one time and acquisition being
dominant, acquisition will overrule fluoroscopy. This reduces the risk of waste of
contrast (or overdose of contrast and X-ray) in case of inappropriate use of the
Philips Medical Systems
footswitch.
Acquisition
Digital acquisition of X-ray images. The result will be a series of images.
Conditions
Acquisition is disabled during a change of APR. Pressing the footswitch after
a change of APR directly activates X-ray exposure.
After starting the exposure, all other active system functions are stopped and
the Examination monitor is blanked. A delay bar is displayed in the bottom
left corner of the Examination monitor if a delay time is programmed. After
system preparation, the images are acquired using the selected image
technique and displayed on the Examination monitor. If the acquisition
speed exceeds the display speed at full resolution, the display will be at half
resolution.
[7]. To alter a parameter, press |Alter APR parameters| [10] (see Section
5.1.6. ‘APR handling’.
5 Before performing digital acquisition, the following functions can be set or
modified:
• patient orientation [5, 6]
• II field size [153, 154]
• shutter adjustment [151, 152]
• wedge filter [155, 156]
Single-shot acquisition
Digital acquisition of a single X-ray image.
Acquired images are added to the current patient’s examination file. The
acquired images can be viewed at any time. The Viewpad is most convenient
for this purpose.
Conditions
Single-shot acquisition is only possible for digital vascular procedures with
lock-in APRs and can only be started with the foot switch. One-knob
acquisition and programmed delay time is not possible.
WA R N I N G It is the responsibility of the operator to avoid actions that could increase the tube
load.
Tube load
The tube load is shown by one of the indicators [122] on the display unit:
green, yellow or red. If the red indicator is lit, only low and normal
fluoroscopy modes are possible. The maximum recommended run time or
tube cool-down time is displayed on the acquisition console.
Dose rate
The dose rate is monitored during acquisition and is shown on the display
9896 001 33152
unit [127] where it is updated every second. At the end of acquisition, the
accumulated dose is displayed on the Acquisition console and on the Status
page [F9] on the viewing console. If configured, a dose/examination report
can be printed [4c].
Philips Medical Systems
This Section refers to a system with factory default viewing functions. If the
viewing functions were reconfigured during system installation (see Section
4.3.2, ‘Viewing console, TAM and viewpad(s)’), it is strongly recommended
to amend the section accordingly.
When acquisition functions are activated by the operator in the examination
room, the patient annotation function [86] and analytical programs [98] are
terminated. Other programs are suspended with their current settings intact
and remain visible on screen in a suspended state. Terminated functions must
be re-activated after exiting the function that was activated in the
examination room.
Viewing can be started from the Review page or by pressing any key that
produces an image.
Conditions
If a function is activated at one user interface (e.g. Viewing console, TAM or
Viewpad), it will remain active until it is deactivated. In addition, the
activated function will only be indicated as active on the user interface from
which it was selected.
If the current acquisition examination is the same as the current viewing
examination, processing indicators on the Viewing console will refer to
functions activated from the Viewpad(s) and TAM.
Parallel viewing allows selection of images (image runs) of a patient other
than the one currently being examined.
Unless otherwise specified, the following rules apply to viewing functions
controlled by the mouse (and by the joystick and related keys on the TAM):
• processing parameters only take on the displayed value after clicking
|Accept| [25] (left mouse button). When a function key is pressed again
before clicking |Action| [26] (middle mouse button), the processing
parameters are reset to the values they had when the function was entered
• when a function is overridden or a run boundary is crossed, the function is
terminated, and the processing parameters remain active as if |Accept| [25]
had been clicked
• modified processing parameters can be reset to their acquisition values by
9896 001 33152
36120050
2 Use ‘Previous’, ‘Next’ and/or ‘Disk A/B’ to select the page containing the
viewing examination.
3 Use the |Cursor up/Cursor down| keys [20] to select the viewing
examination.
4 Confirm the selection by pressing |Enter| [21].
The middle image of the first run is displayed on the image monitor. The
Status Page of the selected examination will be displayed.
5 To select a specific run from the Status page, use the |Cursor up/Cursor
down| keys [20].
6 Confirm the selection by pressing |Enter| [21].
7 Once a run has been selected, other images from the run can be selected and
processed by using the |View forward/reverse| keys [30/31].
9896 001 33152
33726410
Viewing console
Philips Medical Systems
The following Viewing console functions can be used for image selection, see
illustration above:
To display the images of the run in the sequence in • View forward/reverse [30, 31]
which they were acquired or in reverse sequence:
To step quickly through the runs of the examination • Step to next/previous run [32, 33]
selected:
To modify the viewing speed for forward, reverse or • Higher/lower speed [38, 39]
cyclic viewing:
a b
c d
e
f g
h i
j k
l m
33723580
Viewpad
The following viewpad functions can be used for image selection for the
acquisition patient only, see illustration above:
To display the current image subtracted from the Subtraction on/off [a]
current mask:
To make the current image the new mask image: New mask [b]
To display the images of the run in the sequence in View forward/reverse [d, c]
which they were acquired or in reverse sequence:
9896 001 33152
To zoom the central area of the the currently Center zoom [g]
displayed image:
5.0.5 Post-processing
Post-processing allows the operator to optimize the quality of the images, i.e.
contrast, brightness and edge enhancement etc.
Conditions
In order to perform post-processing the required run must first be selected.
It is necessary to have good images at the start of the post-processing
procedures in order to achieve optimum quality.
Performing post-processing on one specific image affects the whole run.
To zoom a quarter of the image area by a factor of 2: Pan & zoom [55]
To override applied processing functions and reset the Override manual adjustment [58]
run to its state at acquisitions:
9896 001 33152
To display the current image subtracted from the Subtraction on/off [60]
current mask:
To select a new mask image using the View forward/ Move mask [61]
reverse function:
To make the current image the new mask image: New mask [62]
To combine in one image all images acquired during View trace [67]
the flow of contrast medium in a single image:
5.0.6 Archiving
Images obtained from the INTEGRIS system can be transferred to a HCU.
For archiving examinations to other archiving systems refer to Section 5.6
‘Archiving’.
Conditions
Before sending images to a HCU the images need to be stored in the
photofile, selection of the images has to be done manually.
During fluoroscopy/ultrasound
1 Press |Store reference 1/2| key [j, k] on the viewpad. The image is displayed
on the Reference monitor and stored in the photofile.
During acquisition/viewing
1 Complete the current acquisition examination.
The image(s) can be selected after completing the examination run.
2 If necessary, press the |Add text| key [85] to switch image information on/off.
3 If necessary, press the |Annotate| key [86] to switch text annotation on/off.
4 Use the selection keys [30 - 43] (see Section 4.3.3) on the viewing console to
select the image to be stored.
5 Press the |Store image in photofile key| [70] on the viewing console.
9896 001 33152
2 Press and hold the |Delete image from photofile| key [71] on the viewing
console until the viewing monitor is blank.
The following functions can be applied to the photofile:
To delete the currently-displayed image from Delete Image from Photofile [71]
the photofile:
Conditions
Ensure that:
• the HCU connected to the INTEGRIS system is ready for use
• images required to be transferred to the HCU are stored in the photofile.
36120030
4 Select the examination to be copied with the |Cursor up| and |Cursor down|
keys [20].
5 Confirm the selection by pressing |Enter| [21].
The following information will be added to each image or to the first image
on each film (configurable by Service during software installation):
• patient name
• birth date
• examination ID
• patient sex
• examination date
• examination time
• physician’s name
• hospital name.
Other examinations can be selected before the transfer process has been
completed. Each examination will start on a new film. The film format is
Service configurable.
Non-vascular interventional -
procedures
Cardiac procedures -
Trauma patients -
Neuro procedures -
X-ray filters and Ceiling suspended radiation shield How to position the ceiling
accessories (§ 5.1.3) suspended radiation shield
(continued)
Ratchet compression band How to attach the compression
band unit to the edge of the table
reference setting
Alter sequence -
Bolus chase
Rotational angiography
Pivot option
Pivoting the table around a fixed point (table base).
Conditions
The pivot option must be installed on the AD5 table.
1 To pivot the table, push the side of the table in the required direction. It is
advisable to push the head end of the table at least 1.5 m (59.1 inch) from
the pivot point with the tabletop brakes activated.
9896 001 33152
36280300
13° pivot position, -90° pivot position and +90° pivot position
Philips Medical Systems
SyncraTilt option
The tabletop can be tilted head-down (max. -28°) and head-up (max. +20°)
using the |Table tilt movement| [139] control. The actual tilt angle is shown
on a scale on the table column and is also given on the display unit [124].
The SID is normally indicated [124] when the tabletop is in the horizontal
position.
If the tabletop is in a tilt position, the tabletop can be panned using control
[131]. The transverse movement is manual and longitudinal movement is
motorized. To activate the motorized longitudinal movement, press |Both
tabletop brakes on/off and longitudinal movement| [131] in the required
direction. As the longitudinal panning speed depends on the side-pressure
applied to control [131], use a gentle force to ensure smooth movement.
Various isocentric tilt positions are illustrated below. Isocentric tilting is only
possible if configured (Service).
36280310
Isocentric HDT with P-A projection Isocentric HUT with P-A projection
for patient in supine position or A-P for patient in supine position or A-P
projection for patient in prone position projection for patient in prone position.
36280320
Conditions
The SyncraTilt option must be installed on the AD5 table.
When the tabletop is moved from a head-down (maximum tilt -28°) to a
Philips Medical Systems
head-up (maximum tilt +20°) position, or vice versa, the movement stops
when the tabletop is in the horizontal position. To continue the movement,
release |Table tilt movement| knob [139] then push it again in the required
direction.
1 Unlock the knob [139] by pressing it down and turning it 90° clockwise.
2 Use the |Table tilt movement| knob [139] to adjust the tilt angle of the table.
3 To avoid an accidental tilt movement, lock the knob [139] by pressing it
down and turning it 90° counter-clockwise.
Pressing the |Movement override| button [140] disables the automatic tilt
movement safety system which safeguards patient and equipment. If either
the |Table tilt movement| [139] or |Table height adjustment| [133] joystick is
then used, an audible signal warns the operator of the increased collision risk
during a tilt or height movement.
36280560
Preferred settings
N OT E The preferred settings are illustrated for the floor mounted C-arm stand but the same
settings are also used for the ceiling suspended C-arm stand.
Preferred procedure
3 Pivot the table into a locked position away from the C-arm and retract the
tabletop towards the table base for convenient transfer of the patient onto the
table.
Philips Medical Systems
4 Place the patient in the supine position with the head as close as possible to
the edge of the tabletop.
5 Pivot the table to the locked position at +13° for the right arm or -13° for the
left arm.
6 Position the arm on the arm support at an angle of about 35° with the body.
7 Pivot the L-arm to -40° for the right arm or to +40° for the left arm in order
to align the shutters with the arm.
Other procedures
As above, but with the II under the table.
36280570
Preferred setting
Preferred procedure
table.
5 Place the patient in the prone position with the head as close as possible to
the edge of the tabletop.
Philips Medical Systems
9 Set the C-arm angulation to +90° (C-arm lies in horizontal plane and
surrounds the head-end of the table) so that the orientation of the image on
the monitor matches that of the needle to facilitate accurate positioning of
the needle.
10 For further positioning of the ROI use the table movements (horizontal float
and table height).
11 After positioning the needle and injecting the contrast medium, the required
projections can be set for image acquisition.
12 Move the II as close as possible to the object.
Cardiac procedures
36280330
Preferred settings Other settings
Preferred procedure
tabletop towards the table base for convenient transfer of the patient onto the
table.
5 Place the patient in the supine position with the head as close as possible to
the edge of the tabletop.
Philips Medical Systems
Other procedures
As above but with the L-arm rotated to +90° or -90°, so that the C-arm is
perpendicular to the table.
Trauma patients
For procedures that require free access for a trolley to the C-arm stand in
cases of emergency, trauma etc.
Preferred procedure
1 Pivot the table to the +90° or -90° position (see Section 5.1.1 ‘Pivot option’).
Neuro procedures
CAUTION Always use the wedge shaped neuro head rest to position the head at the isocenter
while ensuring a convenient working height.
36280330
Preferred settings Other settings
Preferred procedure
Other procedures
As above but with the L-arm rotated to +90° or -90° so that the C-arm is
perpendicular to the table (see also ‘Rotational angiography’).
36280340
Preferred settings Other settings
Preferred procedure
WA R N I N G Make sure that no collision can occur during rotation of the C-arm, as all bodyguards
are switched off during rotational angiography acquisition.
10 Rotate the C-arm to the required end-position of the scan and store this
9896 001 33152
N OT E Positioning the C-arm in the pre-defined end or start position can be done using the
|Recall position 1 or 2| key [144] on the geometry module.
Other procedures
The procedure described above allows scanning of the ROI in a plane
perpendicular to the object. However, it is also possible to scan in a plane
with a skewed angle to the object. If this is required carry out one of the
following alternative procedures while keeping all other preferred settings
unchanged:
• pivot the L-arm only, so that it makes a skewed angle with the table, or
• keep the L-arm perpendicular to the table but angulate the C-arm, so that
it makes a skewed angle with the L-arm, or
• do both, i.e. pivot the L-arm and angulate the C-arm with respect to the
L-arm.
36280580
Preferred settings Other settings
Preferred procedure
WA R N I N G Make sure that no collision can occur during rotation of the C-arm, as all bodyguards
are switched off during rotational angiography acquisition.
10 Rotate the C-arm to the required end-position of the scan and store this
position in the Store 1 memory by pressing |Store position 1| [142].
11 Rotate the C-arm to the required start-position of the scan and store this
position in the Store 2 memory by pressing |Store position 2| [143].
N OT E Positioning the C-arm in the pre-defined end or start position can be done using the
|Recall position 1 or 2| key [144] on the geometry module.
36280350
Preferred settings Other settings
Preferred procedure
Other procedures
As above but with the L-arm in the 0° position in line with the table.
9896 001 33152
Cerebral filter
The cerebral filter is used for neuro angiographies of the head and avoids
direct radiation on the image intensifier entrance screen, thus improving the
overall image quality.
Conditions
The cerabral filter is only to be used for cerabral angiography examinations.
1 Push the cerebral filter into the rim on the X-ray tube housing.
36270140
How to remove the cerebral filter
1 Remove the cerebral filter by inserting a finger into the filter hole and lifting
the filter out of the rim on the X-ray tube housing.
Conditions
There are no special conditions for the following procedures.
a
II entrance screen.
36270160
36270170
Cover plate removed
36270180
36270190
Entrance screen grid removed
36270200
locates back into its retaining slot.
5 Secure the II entrance screen cover plate to the II housing using the four
screws and washers removed in step 1.
9896 001 33152
housing.
36270150
36270160
36270190
Cover plate removed
36270210
retaining slot.
5 Secure the II entrance screen cover plate to the II housing using the four
screws and washers removed in step 1.
The table mounted radiation shield offers the physician and staff additional
protection against scattered radiation. Maximum protection can be achieved
by using the table mounted radiation shield in conjunction with the ceiling
suspended radiation shield.
Philips Medical Systems
WA R N I N G Before using the radiation shield, check that the shielding material is not damaged.
The shield should be free from visible cracks and tears when examined using
fluoroscopy. It is strongly recommended to perform this check on a regular basis (see
Section 7, ‘Maintenance’) and whenever there is a possibility that the shield may have
been damaged.
CAUTION Always remember to remove the radiation shield from the table accessory rail before
tilting the table.
Conditions
There are no special conditions for the following procedures.
33725550
2 Slide the clamping device to the edge of
the table accessory rail.
3 Turn the knob clockwise to clamp the
shield to the table accessory rail.
33725560
The radiation shield can be manually located into one of the following
positions:
• working position with lower and upper shield
• working position with lower shield only (upper part folded down)
• parking position (shield stowed under the table).
36280080
Lower and upper shield, Lower shield only and Parking position
9896 001 33152
33725580
33725590
How to fold the upper shield down
36280090
How to park the radiation shield
36280100
Ceiling suspended radiation shield
The ceiling suspended radiation shield offers the physician and staff
additional protection against scattered radiation. Maximum protection can
be achieved by using the ceiling suspended radiation shield in conjunction
with the table mounted radiation shield.
WA R N I N G Before using the radiation shield, check that the shielding material is not damaged.
9896 001 33152
The shield should be free from visible cracks and tears when examined using
fluoroscopy. It is strongly recommended to perform this check on a regular basis (see
Section 7, ‘Maintenance’) and whenever there is a possibility that the shield may have
been damaged.
Philips Medical Systems
CAUTION Always be aware that collisions may occur with the ceiling suspended radiation shield
during positioning of the monitor ceiling carriage and/or during positioning of the
geometry segment. Such collisions must be avoided as this may cause damage to the
equipment.
Conditions
If required, place a sterile bag, from the bottom, over the apron and lead
acrylic shield, secure the bag in the two notches on the lead acrylic shield (a).
1 Using the pivot action and up/down movement of the suspension arm,
manually move the shield to the desired position (b).
2 If required, tilt the shield to the desired position (c).
36280110
Optimal protection can be achieved by using the ceiling suspended radiation
shield and the table mounted radiation shield together.
36280150
WA R N I N G The ratchet mechanism exerts considerable mechanical force while the strap is being
tightened. Be careful to control the amount of pressure used.
Philips Medical Systems
Conditions
There are no special conditions for the following procedures.
How to attach the compression band unit to the edge of the table
1 Position the unit on the edge of the table and tighten the knurled fixing
screws underneath the unit.
33724550
a b
How to remove the compression band unit from the edge of the table
1 Slacken the knurled screws and slide the complete attachment from the table.
Head supports
Head supports are used to immobilize the pat ent’s head and for patient
comfort during the examination, this in turn reduces motion artefacts on the
images.
Conditions
A piece of anti-slip foil, cut from the reel supplied and renewed when
necessary, should be placed between the head cup and the table to prevent
the cup from slipping.
c
b
9896 001 33152
33724560
1 Place the head cup [c] at the head end of the table on a piece of non-slip foil.
Philips Medical Systems
2 If the head is to be tilted, place one or both wedges [a, b] under the head cup.
3 Position the patient so that the head lies comfortably in the cup.
33723550
Rail accessory clamps Tabletop accessory clamps
Conditions
When no other accessories are fitted to the clamps, the clamps should be
removed from the rails.
Conditions
There are no special conditions for the following procedure.
Arm supports
The arm supports are only used for patient comfort during examination.
9896 001 33152
Conditions
There are no special conditions for this procedure.
Philips Medical Systems
33724570
a b
1 Slide the long part of the support [a, b] under the patient at forearm level and
place the arm in the channel so formed.
2 Repeat procedural step 1 for the other arm.
Conditions
There are no special conditions for the following procedures.
b
33724580
1 Slide the clamp over the edge of the table to the right of the patient at
shoulder level.
2 Fix the clamp by turning the lever underneath.
3 Set the angle of the support as required by rotating it on the frictional pivot
[b].
4 If the handgrip on the top of the adjustable arm support is not required, it
may be removed after loosening the knurled nut [a], and an end stop
substituted.
9896 001 33152
The catheterization arm support is used for upper peripheral and brachio
cephalic catheterization procedures.
Conditions
The patient must be positioned on the table before sliding the
catheterization arm support under the patient’s shoulder.
WA R N I N G All accessories are intended to help secure the patient to the tabletop. When attaching
the accessories, always check correct positioning and fixing of the clamps. This is very
important to prevent the patient sliding off the table during a tilt movement.
Conditions
Accessories required during patient examination should be fitted prior to
patient positioning.
WA R N I N G The ankle straps must always be used in combination with the two handgrips as a
safety backup.
33724600
Prone position
33724610
Supine position
33724620
tabletop rail using the rail accessory
clamps.
33724630
2 Place the ring over the peg.
33724640
33724660
The straps can be used with the patient
prone or supine.
33724670
Prone position
33724680
Supine position
Conditions
The following procedure is only valid for a normal (i.e. non-neuro) tabletop.
1 Mount two clamps on the tabletop
without tightening them
9896 001 33152
33724690
33724700
The supports can be adjusted to fit a
prone or supine patient.
33724710
Prone position
N OT E The shoulder supports should not be used for examination of the neck region as they
are opaque to X-radiation. For this type of examination, it is better to use the ankle
straps to secure the patient.
33724720
Supine position
Short version
33724730
Long version
1 Fit the handgrips into the rail accessory clamps and tighten the clamps to
secure the handgrips.
33724710
5.1.4 XB and colour monitors
The XB monitors are used to display all live and reference images during
examinations. The XB monitors have a light sensor which detects the
ambient light in the examination room and automatically adapts the
brightness of the monitors.
The colour monitor is used to display colour images and data in the
examination room, for instance provided by, either:
• A hemodynamic acquisition system, or
• INTEGRIS 3D-RA.
Reference setting
The reference setting is the default (factory adjusted) brightness and contrast
setting of the monitor(s). Recalling the reference setting restores the
brightness and contrast settings of all individual monitors in the examination
room to reference setting. The reference setting can be activated on either
monitor (daisy chain).
Conditions
There are no special conditions for this procedure.
MODE
REF. SETTING 1
9896 001 33152
The reference setting is recalled for all monitors in the examination room.
contrast setting of the monitor(s). Recalling the user reference setting restores
the brightness and contrast setting of all individual monitors in the
examination room to user reference setting. The user reference setting can be
activated on either monitor (daisy chain).
Conditions
There are no special conditions for this procedure.
N OT E When adjusting the user reference setting, the ambient lighting does not need to be
taken into consideration.
MODE
USER SETTING
The user reference setting is recalled for all monitors in the examination
room (daisy chain).
MODE
USER STORE
The new user reference setting is stored for all monitors in the examination
room.
WA R N I N G S • Prepare the probe for sterile conditions as required. The probe is not shipped sterile.
9896 001 33152
electrical hazard.
CAUTIONS • Inspect the probe for cracks and other damage before use.
• Avoid pinching or kinking the probe cable.
• Disconnect the probe from the scanner before cleaning or disinfecting it.
• Do not immerse the probe in water.
• Never operate the equipment at an ambient temperature greater than 40 °C.
• Avoid installing the unit near a heater or in direct sunlight.
• To avoid serious damage to the scanner and/or probes only connect or disconnect
the probes when the scanner is switched off (when the indicator lamp above the
probe connection on the scanner is off).
Conditions
The probes must be replaced in the probe-holders (located at the back of the
control panel) or in their original case before moving the monitors and
scanner.
When using probes under sterile conditions wrap the disinfected probe in
protective sterile covers.
Use a protective sterile probe cover or sheath to cover the probe during
patient examinations. To avoid patient-to-patient contamination, dispose of
the probe cover or sheath immediately after each examination.
When using probes in an intraoperative procedure, clean and disinfect the
probe prior to use and between each examination. Disinfection is required
before the first examination and between all subsequent examinations.
1 Move the control panel from the parked position underneath the monitor(s).
2 Check the probe(s) (housing, cable and connector) and the scanner for
damage.
3 Connect each probe to the scanner with cable pointing downward.
4 Check that the connections are secure.
5 Press the |System on| button [F] on the scanner front panel.
After switching on, the system is ready for use when the key lights (after
approximately 5 seconds). The indicator above the probe connection on the
scanner is also lit. The examination monitor displays the ultrasound image
automatically.
9896 001 33152
N OT E If the system is switched off immediately after switching on, it cannot be switched on
again immediately. Wait 10 seconds before switching the system on again.
7 After acquiring images (see Section 5.1.5 ), switch off the scanner.
8 Using a sufficient quantity of gel, position the probe on the skin over the area
of interest. If necessary, use the cable hook under the swivel arm and on the
left side of the scanner to keep the probe cables properly aligned.
Needle guides
Needle guides are used to direct the needle to a specific region of interest, as
seen on the monitor.
Conditions
Clean and sterilize the needle guide prior to use and between examinations.
The needle guide is not shipped sterile.
Only use the needle guide when the probe is covered with a sterile sheath.
Handle the needle guides with care. Inspect daily to ensure that they are not
bent or otherwise distorted.
Needle guides must not be used for chorionic villi sampling, percutaneous
umbilical blood sampling, or in vitro fertilization.
WA R N I N G Do not use the needle guide if it is distorted in any way until a service engineer inspects
it.
1 Clean the needle guide before each examination to remove all visible tissue
residue, fluid or other material.
2 Scrub the needle guide with water and a mild detergent. Use pipe cleaners to
clean the lumen of the needle guide.
3 Sterilize the needle guide prior to first use and between all subsequent
examinations. There are two ways to sterilize the needle guide:
• steam autoclaving
• soaking the needle guide in CidexTM Activated Dialdehyde Solution in
accordance with the manufacturer’s instructions.
4 After sterilizing the needle guide, rinse it in sterile water.
7.5 MHz linear array probe
9896 001 33152
Philips Medical Systems
33724740
Fitting the needle guide to the 7.5 MHz linear array probe
5 Slide the needle guide holder (type II) over the probe cover onto the ‘V’
attachment on the probe.
6 Fix the holder to the probe cover by tightening the thumb screw.
7 Slide the needle guide for the required needle size into the needle guide
holder.
8 If necessary, rotate the needle guide to allow the needle to be inserted and
removed.
Other probes
33724750
Attaching the needle guide to the 5.0/7.5 MHz, 3.5/5.0 MHz or 3.5 MHz HD probe
1 If still in place, remove the black rubber ‘O’ ring from around the probe.
2 Slip the needle guide clamp (type I) onto the probe, making sure that the
locating pin engages in the locating hole.
3 Apply contact gel to the probe and insert the probe into the sterile probe
cover provided with the disposable kit.
4 Snap the needle guide holder over the probe cover onto the clamp.
5 Slide the needle guide for the required needle size into the needle guide
holder.
6 Insert the needle into the guide and rotate the guide so that it locks into the
holder.
How to switch the scanner off and park the control panel
CAUTION When positioning the arm of the scanner tray under the monitors make sure that the
probes in the probe tray do not collide with the handles underneath the monitors.
1 Press the |System off| button [G] on the scanner front panel.
The examination monitor re-displays the X-ray image.
9896 001 33152
33724940
Control panel in parked position
Conditions
When the system is switched off and switched on again, a new patient is
selected or another APR within the same imaging technique is selected, the
system will return to the default APR parameters.
4 To terminate the procedure, press the |Alter APR parameters| key [10] on the
acquisition console again.
The altered APR parameters will be stored and the indicator will flash for
about 2 seconds.
N OT E S • For a 2-factor technique, the kV and mAs values can be adjusted. When the kV value
is changed, the new mAs value is checked to ensure that it lies within the limits of
the mAs parameter. Whilst this is taking place, the kV display flashes. If the mAs
value lies outside the limits, the highest possible mAs value is highlighted and the
current value can be altered. The system itself may also decrease the current value
if it lies outside the limits.
• For a 3-factor technique, the kV, ms and mA values can be adjusted. When the kV or
ms value is changed, the new mA value is checked to ensure that it lies within the
limits of the mA parameter. Whilst this is taking place, the kV or ms display
(whichever was changed) flashes. If the mA value lies outside the limits, the highest
possible mA value is highlighted.
Alter sequence
The Alter Sequence function allows the speed and runtime of a Digital
9896 001 33152
Duration Frame-rate
T1 s V1 /s
T2 s V2 /s
T3 s V3 /s
Duration Frame-rate
Total images: N1
Free run/img: N2/N3
N1 is the total number of images acquired in the system. N2/N3 is the free
disk capacity (runs and images). T1 , T2 and T3 are the periods (seconds) in
the sequence and V1, V2 and V3 are the delay or the acquisition speed in the
corresponding period.
The ‘Total images’ figure that is displayed and is calculated from the box
values T1 to V3 can exceed the free image capacity. If so, the user is informed
with a system message on the acquisition console:
‘Total-images: 35’
‘Free run/img: 5/24’
The displayed free capacity ‘N2/N3’ is the number of runs and images which
are free for the currently selected APR.
The total images is calculated as follows:
• complete pre-programmed sequence:
- (T1*V1)+(T2*V2)+(T3*V3)
• manual pre-programmed:
- (a) (T1*V1)+(T3*V2)or
- (b) (T1*V1)++(T3*V3).
The Alter Sequence function can be terminated by pressing the |Alter APR
parameters| key [10] again. The new total number of images is then
calculated and displayed. A programming example is given at the end of this
section.
Philips Medical Systems
Selectable values
• T1 0, 1, 2, 3, ..... 19, 20
• T2 Manual, 0, 1, 2, 3, ..... 19, 20
• T3 0, 1, 2, 3, ..... 19, 20 or 0, 1, 2, 3, ......39, 40 if T2 ‘manual’,
• V1, V2, V3 Selectable values of acquisition speeds depend on pre-
programmed APR parameters and the mains supply frequency (50 or 60
Hz).
Digital-vascular [frames/s]:
Sequence display
The contents of the displayed sequence depend on the APR mode.
The following sequence displays are possible:
• normal sequence
• manual sequence
• ECG sequence
• high-speed sequence
• bolus chase
• rotational scan sequence.
Lock-in mode
Normal sequence
Duration Frame-rate
T1 s V1 /s
T2 s V2 /s
T3 s V3 /s
Total images: N1
Free run/img: N2/N3
Manual sequence
Duration Frame-rate
T1 s V1 /s
T3 s manual
V2 /s
9896 001 33152
V3 /s
Total images: N1
Free run/img: N2/N3
Philips Medical Systems
The total time of the VFR period is given as T3 sec. The actual acquisition
speed is displayed as:
V2 (frames/s); the alternative acquisition speed as V3 (frames/s).
The entry order is T3 , V2 , V3.
ECG sequence
If ‘ECG’ is programmed during installation, the wait time (T1) and the run
time (T2) can be changed.
Duration Frame-rate
T1 s delay
T2 s ECG
Total images: N1
Free run/img: N2/N3
The maximum number of images that will be acquired is:
• single ECG: 3 x T2
• dual ECG: 6 x T2
High-speed sequence
The speed values 12.5 and 25 (50 Hz) / 15 and 30 (60 Hz) cannot be dialed
or changed by the operator. These values can be programmed during
installation. For these speeds, the sequence display will show:
Duration Frame-rate
T1 s delay
T2 s V2 /s
Total images: N1
Free run/img: N2/N3
Bolus chase
Three APR modes are possible for bolus chase:
• Automatic bolus chase with subtraction (with bolus chase package only);
the system automatically selects the correct run sequence based on the
KAT (Knee Arrival Time). The KAT can be entered using Alter sequence.
• Interactive bolus chase with subtraction (with ‘bolus chase plus’ package
9896 001 33152
only); only the normal sequence can be changed using Alter sequence.
• Non-subtracted; available using the Alter sequence function (VFR and
Normal sequence).
Philips Medical Systems
K.A.T. 7s
Total images: 37
Recommended injection:
8ml/s 80 ml (10s)
Total images: 21
Free run/img: 932
KAT and total number of images can be altered by pressing enter [9]. The
recommended injection parameters for the KAT specified are displayed. The
operator must also confirm the table start and end positions by moving the
table and using the Geometry module (see Section 4.3.5 Connection box
and tableside operation modules. The table start and end positions specify a
table-travel distance (TTD). The TTD influences the range of values
admissible for the ‘total number of images’. When the Acquisition console
receives a TTD, it calculates the new range for the ‘total number of images’.
If the current total number of images falls outside this range the value of this
parameter is automatically adjusted to the maximum/minimum admissible
value and the display is updated accordingly. This happens when a TTD is
received during alter-APR. The default range (after APR selection) is 9 - 39
images.
Rotational angiography
For rotational angiography by the cine technique, the sequence block can be
altered as described for the ‘high-speed’ sequence. For rotational angiography
by the phototimer technique the sequence can be altered as described for the
‘normal’ or ‘manual’ sequence. When the operator accepts the altered values
(by pressing the alter-APR key), the new total number of images is calculated
and displayed.
If the sequence display is as follows:
Duration Frame-rate
4s 1 /s
5s 8 /s
7s 3 /s
Duration Frame-rate
4s 1 /s
7s manual
8 /s
3 /s
The operator can successively enter new values for T3 (=7), V2 (=8) and V3
(=3).
9896 001 33152
Duration Frame-rate
Philips Medical Systems
4s 1 /s
9s manual
6 /s
2 /s
With these acquisition sequence parameters, the system starts exposure (after
the test shot) at 1 frame/s over a period of 4 seconds. The system then
automatically switches to an acquisition rate of 6 frames/s for a maximum
Duration Frame-rate
T1 s V1 /s
T3 s manual
V2 /s
V3 /s
If the ‘Alter sequence’ function is active the user can enter new values for T1,
V1, T2, V2, T3 and V3 successively. If another value is dialled and entered for
T2 (other than ’Manual’) then the sequence display will change into:
Duration Frame-rate
T1 s V1 /s
T2 s V2 /s
T3 s V3 /s
T3 s = 0 initially.
Now the user can dial en enter new values for V2, T3, and V3 successively.
Duration Frame-rate
4s 1 /s
12s manual
6 /s
3/s
and with the ‘Alter sequence’ the T2 value (‘manual’) is changed to 7 and
9896 001 33152
Duration Frame-rate
Philips Medical Systems
4s 1 /s
7s 6 /s
0s 3/s
The user is now able to dial and enter new values for V2, T3, V3, T1, V1 and
T2 successively.
WA R N I N G Activating a recalled stand position may cause a collision; if necessary stop the recall
function and position the stand manually.
Scratch mode
Two scratch mode stand positions can be stored and recalled. When scratch
mode is re-entered the previously stored positions are available provided that
the system function ‘new patient’ has not been activated.
Initially, 1 is set for a position with a rotation and angulation of 0° and a SID
of 1.0 m (39.4 inch). 2 is set for -45° rotation and 0° angulation with a SID
of 1.0 m. Positions can be stored in all APC modes and are displayed on the
display unit.
Conditions
Storing a new position overwrites the previously stored position.
1 Press |Store position 1| [162] or |Store position 2| [163] on the APC module,
the stand position is stored as position 1 or 2.
2 To recall position 1 or 2 press the |Select position 1 or 2| key [164] on the
APC module.
When the left side of the button is pressed, position 1 is recalled and
displayed on the display unit. When the right side of the button is pressed,
9896 001 33152
Sequencer mode
Stand positions from one or more service installed sequences can be recalled.
Conditions
When only one sequence is available, the user function ‘select position’
becomes immediately active without displaying the sequence identification.
CAUTION The operator is responsible for avoiding collisions.
1 Press the |Select sequence| set and position key [164] on the APC module to
choose from different customer specific pre-programmed sequences.
Pressing the left side of the button selects the previous sequence and pressing
the right side selects the next sequence. The current sequence is shown on the
Display unit.
2 Press the |Accept and activate| key [165] to accept and activate the current
sequence and display the first position on the display unit.
3 Press the |Select sequence| set and position key [164] to select the required
position within the programmed sequence.
Pressing the left side of the button selects the previous position and pressing
the right side selects the next position. The current position is displayed on
the display unit. By repeatedly pressing this key, all sequence positions can be
displayed.
4 Press and hold the |Accept and activate| key [165] to accept the position and
activate stand movement.
Movement is stopped by releasing key [165].
Reference mode
Stand positions corresponding to the stored references (left and right) can be
recalled.
Conditions
When the mode is re-entered, the last stored positions are available provided
that the system function ‘New patient’ has not been activated.
1 Use the |View forward/reverse| keys [30, 31] on the viewing console to select
the required image on the examination monitor.
9896 001 33152
2 Press the |Store reference 1 or 2| key [74 or 75] to store the image in the
reference file.
The image is displayed on the reference monitor.
Philips Medical Systems
3 Press the |Select position| key [164] to select the reference stand position.
When the left side of the button is pressed, the stand position corresponding
to the image on the reference monitor is selected. If a second reference
monitor is present, pressing the right side of the button selects the stand
position corresponding to the image on that monitor. The selected position
is displayed on the display unit.
4 Press and hold the |Accept and activate| key [165] to accept the position and
activate stand movement.
- accession number
- requested procedure ID and description
- scheduled procedure step ID, description and start date
- scheduled station (Application Entity (AE)) title
Philips Medical Systems
- modality.
N OT E It will at all times be possible to enter manually patient demographics information
within the Integris in case of an emergency or in case the IS connection is down.
However, patient data received from the IS is automatically protected against
modification and cannot be modified within the Integris system. Instead, the data must
be modified in the IS itself and re-imported into the Integris system.
Each time the schedule page is selected [F1], the system automatically
performs a query to the Work list Manager (IS) for all the procedures that are
scheduled on this modality for today. The displayed work list is a merge of
the work list received from the information system and any patients which
have been entered manually. The displayed list is sorted by the Scheduled
Procedure Step start date and time.
It is also possible to perform a more specific query of the work list Manager
by selecting the ‘Query’ function, which becomes available once the schedule
page [F1] has been selected, and then entering values for a combination of
DICOM attributes, this is a user defined work list query. This type of query
can be divided into two groups:
• patient based query: query the work list for a particular patient
• broad query: query the work list for e.g. many patients.
The minimum requirement to perform a patient based query is:
• Patient name
• Patient ID
• Accession number
• Req. procedure ID.
The minimum requirement to perform a broad query is:
• Scheduled Procedure Step start date
• Modality
• Station AE title.
Conditions
Patient data must be available in the IS.
If the connection with the IS fails, within 30 seconds the message ‘No
connection with the Information System’ appears on the error line of the
schedule page. Under these circumstances, it is still possible to add a patient
record by hand following selection of the ‘Add’ function on the schedule
page.
When performing a work list Manager query:
• If there are no results on the query, the message ‘No query results’ is
displayed in the error line of the schedule page.
• If the query results exceed the maximum number of entries allowed, the
message ‘Number of results exceeds 50. Please be more specific’ is
displayed.
• If the Information System does not support the ‘Query’ function, the
message ‘This function is not supported by the information system’ is
displayed.
9896 001 33152
• When wildcards are used when not allowed, the message ‘Invalid DICOM
attribute value’ is displayed.
36120060
Select Add Delete Modify Query More Previous Next
2 Use the |Cursor up/down| keys [20] on the keyboard to select the required
patient examination.
3 Press |Enter| [21] to confirm the selection.
36120070
3 Use the |Cursor up/down| keys [20] on the keyboard, or the |Enter| key [21],
9896 001 33152
to step through the different fields. Enter the data in each field as required.
The only field that allows wildcard matching is the patient’s name.
4 Pressing |Enter| [21] on the last entry field (Station AE title) automatically
starts the query toward the work list Manager.
Philips Medical Systems
5 If the query is not required, press |Return| to cancel the ‘work list Query
Data Entry Screen’ and return to the main schedule page.
The results of the query are displayed on the schedule page. The schedule
page remains empty while the query is being performed.
Conditions
In the case of a protected examination, it is only possible to modify the
physician code.
An examination that is being saved or copied cannot be modified until the
process has been completed.
If the physician code is modified for an examination in which images have
been acquired, the disk partition will not be changed.
36120060
5 Modify the data in the selected field and press |Enter| [21].
6 Proceed through the remaining data fields by pressing |Enter| [21] repeatedly.
Conditions
Any examination can be deleted from the Schedule page (F1).
Only the oldest examination and the current examination can be deleted
from the Review page (F2).
If the current viewing examination is deleted, the current acquisition
examination will be set as current viewing examination.
If the current acquisition examination is deleted, the system automatically
selects a ‘No Name Entered’ acquisition examination as current. This text is
automatically removed when a new acquisition examination is selected.
If an examination is protected against manual deletion, the following
warning message is displayed, ‘Exam. protected against manual deletion. Do
you want to proceed (Yes/No)’ (default answer is ‘No’). Answering ‘Yes’ to
this question will result in the deletion of the examination even if it is
protected, this facility has been included for emergency situations.
1 Press F1 or F2 [24].
The Schedule or Review page is displayed with the oldest examination
highlighted.
Conditions
On the system page (F6) only the time and date can be modified by the
operator, all other items are configured by Service during installation.
33725120
Set time Set date Service Test Img IQ Tools
Conditions
36120100
Add Delete Modify Previous Next
33723200
Trace image symbol Trace subtract image symbol
N OT E For CO2 injections, the colors of the vessels are the opposite of those seen with Iodine
injections since CO2 has minimal opacity whereas Iodine has maximal opacity.
Conditions
TSF can be performed in any fluoroscopy mode.
1 Press the |Trace subtract fluoroscopy| key [158] on the imaging module, and
select the injection medium (Iodine or CO2 from one of the menus [95-97
(viewing console) or 102 (TAM, if configured)].
The Trace image symbol appears in the top right corner of the examination
monitor.
2 Initiate fluoroscopy with the foot switch [175].
3 Wait until fluoroscopy is stabilized and subtracted fluoroscopy is
automatically activated.
The image on the monitor turns grey (mask image).
9896 001 33152
Conditions
If digital exposures are made in the examination room while post-processing
or archiving is being performed at the Operator’s console, the Operator’s
console functions are temporarily frozen.
When the system is in the parallel or dual fluoroscopy mode, the fluoro
images cannot be grabbed, nor can they be recorded on video.
In the parallel fluoroscopy mode, running Quantitative Analysis software
packages will be terminated by exposures.
• image processing
• transfer run for Bolus Chase Reconstruction (BCR) (if applicable).
The steps will be described in detail for the automatic mode and the
Philips Medical Systems
interactive mode.
Automatic mode
a b c d c b a d
36280630
Standard Bolus Chase Reversed Bolus Chase
a. Start position.
b. Table travel distance; min 70 cm (27.6 inch), max 100 cm (39.4 inch).
c. End position.
d. Bolus Chase table scan direction.
WA R N I N G Before performing Bolus Chase, carefully check that patient and table will not collide
with any object.
General preparation
1 Position the C-arm perpendicular to the table with the X-ray tube under the
table.
2 Make sure that the table is in the non-rotated and non-tilted position.
3 When using a floor mounted C-arm, press the |Swivel in| key [186] to move
the table into the peripheral position.
4 Position the patient in the supine position with head as close as possible to
the edge of the tabletop (opposite the pedestal).
N OT E If reversed bolus chasing is preferred both the patient and the movement have to be
reversed and step (4) is as follows:
Position the patient in the supine position with feet as close as possible to the edge of
the tabletop (opposite the pedestal).
5 Press the |Tabletop brakes on/off and longitudinal movement| control [131]
on the geometry module to release the tabletop brakes and move the tabletop
transversely to its central position.
Make sure that the femoral heads of the patient are just visible in the lower
9896 001 33152
6
corners of the fluoroscopic image with an II input field size of 38 cm (15
inch) and, if not, reposition the C-arm, or reposition the patient on the table.
7 If the Bolus Chase Reconstruction (BCR) with EasyVision is planned after
Philips Medical Systems
the examination, position the special ruler for BCR between the legs of the
patient.
8 Position the central peripheral filter carefully between the patient’s legs with
the thin, wide end towards the feet and the thick, narrow end as high up as
possible.
9 Position the two side filters as close as possible to the left and right side of the
legs with the wider part at the feet.
These filters must be molded to the shape of the legs to avoid gaps between
the filters and the legs.
10 Immobilize the patient’s legs at knee and ankle using restraint straps.
N OT E The legs must be strapped together with the central filter only, the side filters must be
left free for molding against the legs and to avoid tilting.
For patients with genu varum (O), the knees should be slightly lifted and
supported underneath to enable them to be strapped together closely. For
patients with genu valgum (X), the same procedure is necessary to allow the
ankles to be strapped together closely.
11 Press the |Tabletop brakes on/off and longitudinal movement| key [131] on
the geometry module and move the tabletop to check filter positioning and
patient coverage fluoroscopically.
N OT E For convenience this should be done before the patient is covered with sheets.
The patient should now be prepared for the insertion of the catheter.
Determination of KAT
1 Insert the catheter and make sure that the tip is at L3/L4 level.
2 Press the |Tabletop brakes on/off and longitudinal movement| key [131] on
the geometry module and move the tabletop to bring the knee joint into the
field of view.
3 Select the KAT measure APR [4a/b].
4 Prepare the injector for a small and short test injection (e.g. 8 ml, flow 8 ml/
s, delay and rise time = 0 s).
5 Select One-knob acquisition [4c] on the acquisition console for ‘coupled
injection’.
6 Enable the injector (ready indicator ON).
7 Start an acquisition run using the injector hand switch.
8 Continue the exposures until the front of the bolus has reached the two knee
joints, do not wait for maximum contrast in the knee area; the bolus will not
necessarily arrive at both knee joints simultaneously.
9 Stop the exposures by releasing the injector hand switch.
10 Define the KAT by viewing the run and reading the ‘T-image’ on the image
with the ‘Add text’ function [85] activated.
9896 001 33152
The KAT is defined as the time when the front of the bolus just starts to
become visible at the knee joint. If there is a difference between the left and
the right KAT, the operator may, depending on the clinical situation, decide
to use the average value of the two KAT values or the KAT of the leg of most
Philips Medical Systems
interest.
If the KAT values differ too much (more than 40%) the chance that the
contrast bolus will be visible in both legs becomes smaller. In those
situations, it may help to increase the injection duration (i.e. increase the
volume setting). Another alternative is to give up the single injection
technique and to use the ‘Multiple position DSA’ method (see section
10.2.1).
K.A.T. 7s
Total images: 15
Recommended injection:
8 ml/s 80 ml (10s)
5 Prepare the injector for the contrast run and set the recommended injector
parameters as displayed on the acquisition console.
N OT E The table speed is automatically adjusted such as to attempt to follow the middle of
the bolus. The duration of injection is a crucial factor therefore. If for some reason
the amount of contrast medium has to be decreased, the flow rate must be decreased
accordingly to maintain the same duration of injection (shown in brackets on the
acquisition console display).
9896 001 33152
coverage is obtained.
1 Select the Automatic Bolus Chase APR [4a/b] on the acquisition console.
The message ‘Move tabletop to end position’ is displayed on the acquisition
console and the message ‘Action’ is displayed on the monitor display.
The indicator near the |Store position 1| key [142] on the geometry module
flashes to indicate that the end position can be entered.
1. Catheter tip
at L3/L4 level
2. Magnification
table high (max)
image intensifier close to object
4. Start position
mid L2 at upper edge of image
femoral heads in lower corners
9896 001 33152
Philips Medical Systems
3. End position
ankle joints in center of image
33728252
Mask run
In the automatic mode the operator is free to select which run (mask or
contrast) will be made first. The advantage of making the contrast run first is
that this unsubtracted first run may itself provide sufficient diagnostic
information to avoid the need for a mask run and further X-ray exposure.
However, if subtracted images are found to be required it is then necessary to
allow sufficient time for the contrast medium to leave the veins before
making the mask run in order to avoid venous over-projection after
subtraction. The advantage of making the mask run first is that there can be
no venous over-projection in the subtracted image, but unnecessary mask
exposures will have been made if the unsubtracted images are subsequently
found to provide sufficient diagnostic information in themselves.
1 Disable the injector (Ready indicator OFF).
2 Make sure that the tabletop is in its start position (indicators near keys [142,
143] are lit). If not, move the table back to its start position using the |TSD
speed control| [172].
3 Press the injector hand switch to start exposures and tabletop travel.
The injector will not start since it has been disabled.
4 Keep the injector hand switch pressed until the tabletop has returned to its
start position.
Contrast run
6 If necessary, review the run and decide whether a mask run is required (if the
procedure had started with a contrast run).
9896 001 33152
Image review
1 Select the first run and then the |Run subtract| key [65] (if applicable) on the
viewing console.
The corresponding images from the two runs will be subtracted one from the
Philips Medical Systems
other and displayed dynamically. All viewing and processing functions can be
applied to all the subtracted and non-subtracted images.
1 Use the export page (F3 to F5 [24] on the keyboard) to transfer the contrast
run to the EasyVision workstation for BCR. The mask run need not be
transferred as EasyVision provides BCR based on contrast images only.
Interactive mode
a b c d c b a d
36280630
Standard Bolus chase Reversed Bolus chase
a. Start position.
b. Table travel distance; min 70 cm (27.6 inch), max 100 cm (39.4 inch).
c. End position (see note below).
d. Bolus Chase table scan direction.
CAUTION Before performing Bolus chasing, check carefully that patient and table will not collide
with any object.
N OT E Defining the end position also sets the end position for the speed control hand switch
[172] movement. Therefore, before defining the end position, be sure that this position
is compatible with the ROI.
General preparation
Same as ‘General preparation’ for ‘Automatic mode’.
The KAT and injector parameters (if applicable) must always be determined
after general preparation and before confirmation of the end and start
positions of the tabletop.
9896 001 33152
K.A.T. 7s
Total images: 15
Recommended injection:
8 ml/s 80 ml (10s)
1 Select the Interactive Bolus Chase APR [4a/b] on the acquisition console.
Otherwise the procedure is the same as that described in the ‘Automatic
mode’. A scan range of less than 70 cm (27.6 inch) is possible in interactive
mode.
Duration Frame-rate
10s 2/s
20s 1/s
20s 0,5/s
1 If required, change the run sequence with the |Alter APR parameters| key
[10].
2 Use the |Dial wheel| [8] to change the parameters and press |Enter| [9] to
move to the next parameter.
3 Press the |Alter APR parameters| key [10] on the acquisition console to
confirm and exit the alter APR mode.
9896 001 33152
Contrast run
In the interactive mode the first run has to be the contrast run because it is
the real time visualization of the contrast bolus that allows the operator to
adjust the table speed interactively.
Philips Medical Systems
5 When the contrast bolus reaches the bottom of the image on the monitor,
start moving the table top and regulate the speed with the hand held |TSD
speed control| [172], such that the contrast bolus remains close to the
bottom of the image.
WA R N I N G Do not wait for maximal opacification, if you miss the peak of the bolus you may not
be able to catch up with it!
Image review
1 Use the |Cursor up/down| keys [20] to select the first run and then press the
9896 001 33152
From the beginning to the end of the rotational angiography, the C-arm
rotates at a maximum speed of 30°/s during the rolling movement or 55°/s
during the propeller movement. This speed is reached within 2 seconds.
Service can program a delay in the APR so that acquisition is not started
before a certain speed is reached. The last 2 seconds of rotation are needed
for the C-arm to slow down, and if the hand (or foot) switch is kept pressed
after rotation is complete, the C-arm will automatically return (at half the
rotation speed) to its start position. If the selected exposure switch is released
before the scan is finished, rotation and image acquisition are automatically
stopped and the system returns to step 3 (see below). The message ‘Confirm
end position’ is then displayed on the acquisition console (see also, Section
10.2.2 ‘Rotational angiography’).
WA R N I N G S • Before performing rotational angiography, check carefully that the C-arm cannot
collide with the patient or any other objects.
• During the actual rotation scan movement the detection system is switched off. For
safety reasons therefore, the user should not allow any patient or table movement.
For safety reasons, after the start and end position of a rotational angiography are
programmed, do not allow any patient or table movement. Because the bodyguard
is switched off, the patient could be harmed during the high speed rotational
angiography movement.
Conditions
9896 001 33152
To improve image quality and to make sure that the propeller movement can
be carried out, position the head support directly on the tabletop (i.e. no
mattress in between).
Make sure that the table is in line with the table base at the 0° pivot position.
Philips Medical Systems
The message ‘Confirm end position’ appears on the acquisition console [4],
and the indicator below key [142] starts to flash.
2 Position the L-arm in the desired position, either manually after pressing the
|L-arm rotation brake release| button [190] and |Longitudinal movement
brake release| button [191] or under motor control using the |Motorized
movement of the frontal stand| control [145].
For a C-arm rolling movement: position the L-arm in the side-position (+90°
(nurse side) or -90° (doctor side)).
For a C-arm propeller movement: position the L-arm in the 0° (head)
position.
3 In the case of a propeller movement, ensure that the tabletop height is
adjusted sufficiently so as to enable the C-arm propeller movement.
4 Position the ROI in the isocenter of the X-ray beam.
To do this it may be necessary to use a lateral projection to adjust the height
of the tabletop and a frontal projection to adjust the tabletop position.
5 Move the C-arm to the desired end position using the |Motor-controlled
rotation and angulation| knob [136].
6 In case of a C-arm rolling movement, the C-arm will now be in the user
defined position (angulation/rotation). If either the 0°, -30° or +30°
angulation position is wanted, then the C-arm needs to be repositioned using
the |Motor-controlled rotation and angulation| knob [136]. At -30° and +30°
angulation the movement will stop automatically.
Ensure that the ROI is in the isocenter of the X-ray beam before the end
position is stored as all table movements are blocked after the end position is
stored.
7 Confirm the rotational scan end position by pressing the |Store position 1|
key [142] on the geometry module.
The indicator below key [142] stops flashing and will be extinguished, and
the indicator below key [143] starts to flash. The message ‘Confirm start
position’ is displayed on the acquisition console [4]. From this moment on,
motorized angulation, table movement and SID adjustment are disabled.
8 Move the C-arm to the desired start position using the |Motor-controlled
rotation and angulation| knob [136] and ensure that the ROI is in the
isocenter of the X-ray beam.
9 Confirm the rotational scan start position by pressing the |Store position 2|
key [143] on the geometry module.
The indicator below key [142] will be lit continuously and the indicator
9896 001 33152
11 If required, a mask run can be performed. Press and hold down the
|Preparation/exposure| button [185] on the acquisition hand switch.
This starts acquisition and rotation of the C-arm from the start position to
the end position and back again. This completes the mask run.
12 Start the rotational scan using one or two knob operation:
• One-knob operation, activated key [4c]:
1 Press the injector hand switch and keep it pressed until the C-arm has
rotated from the start position to the end position and back again. This
completes the contrast run.
• Two-knob operation, deactivated key [4c]: holding the acquisition hand
switch in one hand and the injector hand switch in the other:
1 Press the |Preparation/exposure| button [185] on the acquisition hand
switch until the small button is flush with the large button and hold the
button in this position, this initiates preparation of the X-ray tube.
2 Press the injector hand switch and keep it pressed, this initiates injection of
contrast media.
3 Press the |Preparation/exposure| button [185] on the acquisition hand
switch all the way in to initiate exposure, keep the buttons pressed until
the C-arm has rotated from the start position to the end position and back
again. This completes the contrast run.
13 If required, a mask run can be performed. Press and hold down the
|Preparation/exposure| button [185] on the acquisition hand switch.
This starts acquisition and rotation of the C-arm from the start position to
the end position and back again. This completes the mask run.
The message ‘Confirm end position’ appears on the acquisition console [4],
and the indicator below key [142] on the geometry module starts to flash.
2 For the corresponding rotational angiography scan, use the |Motorized
movement of the frontal stand| control [145] to position the L-arm in the
correct position as calibrated by Service, wait for the ‘click’ sound to make
sure that the stand is in the correct position.
For a C-arm rolling movement: position the L-arm in the side-position
(depending on the calibration, either +90° or -90°) and at the pre-defined
longitudinal position.
For a C-arm propeller movement: position the L-arm in the head-position
(0° rotation) at the pre-programmed longitudinal head position.
Make sure that the L-arm is in the correct 3D-RA calibrated side position,
and also the correct longitudinal position.
3 In case of a propeller movement, ensure that the tabletop height is adjusted
sufficiently so as to enable the C-arm propeller movement.
4 Position the ROI in the isocenter of the X-ray beam.
To do this it may be necessary to use a lateral projection to adjust the height
of the tabletop and a frontal projection to adjust the tabletop position.
5 Move the C-arm to the pre-programmed end position by pressing the left
hand side of the |Recall position 1 or 2| key [144] on the geometry module.
The indicator below key [142] flashes until the C-arm reaches the pre-
programmed end position, the indicator is then lit continuously.
6 In the case of a C-arm rolling movement, the C-arm will now be in the 0°
angulation position. If either the -30° or +30° angulation position is wanted,
then the C-arm needs to be repositioned using the |Motor-controlled
rotation and angulation| knob [136]. At -30° and +30° angulation the
movement will stop automatically.
Ensure that the ROI is in the isocenter of the X-ray beam before the end
position is stored as all table movements (except for the pivot movement) are
blocked after the end position is stored.
7 Confirm the rotational scan end position by pressing the |Store position 1|
key [142] on the geometry module.
The indicator below key [142] is extinguished and the indicator below key
[143] starts to flash. The message ‘Confirm start position’ is displayed on the
acquisition console [4]. From this moment on, motorized angulation, table
movement and SID adjustment are disabled.
8 Move the C-arm to the pre-programmed start position by pressing the right
9896 001 33152
hand side of the |Recall position 1 or 2| key [144] on the geometry module.
The indicator below key [143] flashes until the C-arm reaches the pre-
programmed start position, the indicator is then lit continuously.
9 Confirm the rotational scan start position by pressing the |Store position 2|
Philips Medical Systems
11 If required, a mask run can be performed. Press and hold down the
|Preparation/exposure| button [185] on the acquisition hand switch.
This starts acquisition and rotation of the C-arm from the start position to
the end position and back again. This completes the mask run.
12 Start the rotational scan using one or two knob operation:
• One-knob operation, activated key [4c]:
1 Press the injector hand switch and keep it pressed until the C-arm has
rotated from the start position to the end position and back again. This
completes the contrast run.
• Two-knob operation, deactivated key [4c]: holding the acquisition hand
switch in one hand and the injector hand switch in the other:
1 Press the |Preparation/exposure| button [185] on the acquisition hand
switch until the small button is flush with the large button and hold the
button in this position, this initiates preparation of the X-ray tube.
2 Press the injector hand switch and keep it pressed, this initiates injection of
contrast media.
3 Press the |Preparation/exposure| button [185] on the acquisition hand
switch all the way in to initiate exposure, keep the buttons pressed until
the C-arm has rotated from the start position to the end position and back
again. This completes the contrast run.
13 If required, a mask run can be performed. Press and hold down the
|Preparation/exposure| button [185] on the acquisition hand switch.
This starts acquisition and rotation of the C-arm from the start position to
the end position and back again. This completes the mask run.
5.3.5 Ultrasound
9896 001 33152
Conditions
Philips Medical Systems
1 Press the |Depth decrease| [243] or |Depth increase| [242] key on the
ultrasound control panel to display a smaller or larger image field.
The depth of the displayed image can be changed real time by using keys
[242 and 243]. The upper and lower limit of these selections depend on the
probe type and frequency. A lower frequency gives a larger image field.
1 Press the |Reverse on/off| key [241] on the ultrasound control panel once to
reverse the displayed image.
Pressing [241] again will return the display to normal. The status field
indicates whether or not the image is reversed.
N OT E Reverse on/off [241] has no immediate effect when the image is frozen. Reversal only
occurs once the image has been unfrozen.
1 Press the |Biopsy line on/off| key [240] on the ultrasound control panel once
to display a biopsy line (for the selected probe) in a non-frozen image.
Pressing [240] again erases the line.
N OT E Biopsy line on/off [240] has no immediate effect when the image is frozen. The biopsy
line is only displayed once the image has been unfrozen.
1 Press the |Freeze/unfreeze| key [239] on the ultrasound control panel once to
freeze the current image.
Pressing [239] again will return the image to the live mode. If the image is
frozen, ‘F’ is highlighted in the image status field.
9896 001 33152
3 Press the |Distance measurement| key [238] to fix the position of the first
cross.
The second caliper cross is displayed on the image.
4 Move the second caliper cross to the desired position using the |Joystick|
[247].
5 Press the |Distance measurement| key [238] to fix the position of the second
cross.
The distance (millimeters) is displayed in the result field.
6 Press the |Distance measurement| key [238] again.
The measurement information is erased and the second caliper cross remains
on the screen to allow further measurement.
7 Press the |Freeze/unfreeze| key [239].
The measurement procedure ends and the ultrasound image returns to
normal mode.
Viewing during - -
acquisition (§ 5.4.1)
108
The following TAM functions can be used for image selection for the
acquisition patient only, see illustrations above:
To zoom the current image: Pan & Zoom [103]. The zoomed area
can be changed with the joystick [108]
9896 001 33152
5.4.2 Ultrasound
The ultrasound images appear on the frontal examination monitor.
Conditions
To view the ultrasound images after acquisition the images need to be stored
(grabbed) into the photofile.
1 Press the |Store reference 1/2| key [j (74), k (75)] on the Viewpad.
The stored ultrasound image is displayed on the reference monitor. Its
photofile number and the total number of photofile images are shown in the
lower right corner of the reference monitor.
5.5 Post-processing
This section gives an overview of extended post-processing functions and
describes the quantative analysis programs.
Overview of extended post-processing procedures:
the image:
WA R N I N G There may be a time delay of 10 to 300 ms between the displayed image and the
corresponding ECG signal displayed. This is caused by, and depends upon, the type of
physiology system connected.
User input
The mouse is used for user input. When the mouse is active the actions of
the three mouse buttons are displayed at the bottom of the screen. By
convention, the three mouse buttons are designated, from left to right, as
|Accept| [25] (left mouse button), |Action| [26] (middle mouse button) and
|Reject| [27] (right mouse button). The |Video invert| [54], |Contrast| [45,
46], |Brightness| [48, 49] and |Edge enhancement| [51, 52] functions can be
used with a clinical program. Images can be selected using the image
selection control keys: |Cycle through run| [36], |Cycle through examination|
[37], |View forward/reverse| [30/31]. When a cursor is visible it can be
moved using the mouse. Where relevant, the image produced by a clinical
program can be stored in the photofile by pressing |Flag image| [90] or |Store
image in photofile| [70] on the Viewing console. Patient information can be
added to this image by pressing |Add text| [85] on the Viewing console. If the
image is stored in the photofile, the added patient information is also stored.
WA R N I N G S • The clinical program software has been thoroughly checked and tested prior to
delivery. The software is designed to produce a mathematical model as described in
the medical literature or medical research. Philips Medical Systems cannot be held
responsible in any way for any inaccuracies of any nature resulting from the use of
this software. If the calibration guidelines are not followed, the absolute
measurements may be inaccurate or unreliable.
• As the accuracy of the clinical programs is based upon the ROI position (being in
the isocenter), do not change the position of the table during the examination.
• When using the CF, make sure that the ROI is in the isocenter. If the ROI is not in
the isocenter, this will result in inaccurate measurements which can lead to a
hazardous situation.
9896 001 33152
N OT E User input for the clinical programs described is via the Viewing console and the
mouse. The Tableside Analysis Module (TAM) in the examination room can be used
Philips Medical Systems
instead of the mouse. Press the menu key on the TAM to display the clinical programs
available and use the joystick [108], instead of the mouse, for program selection,
cursor movement or drawing actions. The |Accept| [105], |Action| [106] and |Reject|
[107] keys are located to the right of the joystick [108]. When using the TAM, image
selection can be performed using the image selection keys on the viewpad.
5.5.3 Calibration
General
Before performing quantative analysis programs, calibration must be carried
out either manually or automated (if autocalibration option is installed).
Manual calibration is done by the operator prior to making measurements,
and can be done in two ways:
• grid calibration, for example see 5.5.5 ‘4. AVA manual calibration
procedure’
• catheter calibration, for example see 5.5.4 ‘1. Vessel diameter standard
procedure’.
Calibration for the automated procedure is periodically performed by
Service. This means that the operator only needs to select the ‘Auto
Calibration Factor (Auto CF)’ and does not need to calibrate the system. The
auto calibration factor is only valid for objects in the isocenter.
If no automated calibration has been performed, or only manual calibration
has been performed, the following screen will be superimposed on the image
displayed on the viewing monitor:
33726290
33726300
If no manual calibration has been performed, then the ‘Standard CF-Value’ is
set to ‘none’.
The operator can choose which type of calibration will be used.
Conditions
When using the automated calibration factor, it is important to ensure that
the ROI is in the isocenter. This will ensure that measurements are
performed accurately.
1 Move the tabletop to position the ROI in the isocenter (see procedure
above).
When viewing, the ROI must be displayed in the center of the monitor.
9896 001 33152
References
Procedure
PROGRAM MENU 1
PREVIOUS
NEXT
ACA Short
ACA
Vessel Diameter
Run Overview
33721970
1 Select the correct image series on which to run the Vessel Diameter program.
2 Press one of the three menu keys on the viewing console. The indicator above
the key remains on while the function remains active.
3 If necessary, select Next/Previous screen to access more items.
4 Use the left mouse button [25] to select the ‘Vessel Diameter’ program. A
menu is displayed for selection between calibration and measurement. If no
calibration factor has been determined, calibration is the only item available.
Calibration
Measurement
CF = 0.108 mm/pixel
33721980
5 Click the left mouse button [25] on the selected item to confirm the choice.
9896 001 33152
obtained by comparing the indicated catheter diameter with its known size.
• For accurate and reliable results, the use of a calibration object of a size
larger than 20 pixels is recommended. If a smaller object is used, the
message ‘Warning: size of diameter smaller than recommended’ will be
displayed.
- Click the left mouse button [25] to continue the program.
- Click the right mouse button [27] to cancel the program.
N OT E The calibration factor is expressed in millimeters/pixel of the display matrix.
Step 1
Selecting the image and zoomed area.
CF = 0.108 mm/pixel
33721990
1 Use the image selection keys on the viewing console to select an image with a
clear view of the catheter. The image is zoomed and the zoom frame can be
positioned as for the normal zoom function.
2 Click the left mouse button [25] to confirm the image selection and zoomed
area. The cursor appears on the screen and can be controlled using the
mouse.
Step 2
Drawing a calibration line.
CF = 0.108 mm/pixel
33722020
9896 001 33152
1 Position the central dot of the cursor on one edge of the catheter. This marks
the start of the calibration distance.
2 Hold down the middle mouse button [26] and drag the cursor to the
opposite edge of the catheter. A line is drawn as the mouse is moved.
Philips Medical Systems
3 If necessary, click the right mouse button [27] to reposition the start point of
the line.
CF = 0.108 mm/pixel
33722030
4 Click the left mouse button [25] to confirm the calibration line.
Step 3
Selecting the size.
A menu now appears over the image, offering French sizes (F4 to F9) and
Free Format (FF).
F4
F5
F6
F7
F8
F9
FF
2.7
CF = 0.108 mm/pixel
33722000
Since calibration has been carried out, the arrow now points to
‘Measurement’ by default.
Calibration
Measurement
CF = 0.108 mm/pixel
33722010
5 Click the left mouse button [25] on ‘Measurement’ to continue with the
measurement procedure.
Step 4
Selecting an image and zoomed area.
The following criteria are important for the image selection process:
• the arterial segment should be well filled with contrast medium.
• the obstructed vessel segment should be clearly visible, preferably without
any overlap with other vessels or side branches. The central part of the
image is zoomed automatically and the zoom frame can be positioned
using the mouse.
CF = 0.115 mm/pixel
33722040
9896 001 33152
1 Use the image selection keys on the viewing console to select an image.
2 Click the left mouse button [25] to confirm the image selection and zoomed
Philips Medical Systems
area.
Step 5
Drawing the first measurement line.
CF = 0.115 mm/pixel
33722050
The cursor appears on the image.
1 Position the cursor on one edge of the vessel to be measured.
2 Hold down the middle mouse button [26] and drag with the mouse to draw
a line to the opposite edge of the vessel.
3 Click the right mouse button [27] to reposition the starting point of the line
if necessary.
4 Click the left mouse button [25] to confirm the position of the line.
5 The first diameter, D1, (in mm) is displayed at the bottom of the image and
is the reference diameter for the percentage diameter obstruction.
Step 6
Drawing second measurement line.
D1 = 2.01 mm
CF = 0.115 mm/pixel
33722060
1 Draw a second line with the mouse as described in (1) to (3) of ‘step 5’.
2 Click the left mouse button [25] to confirm the position of the second line.
Step 7
Final display.
1
2
%D = 60 %
D1 = 2.01 mm
D2 = 0.79 mm
Cf = 0.115 mm/pixel
33722070
The final display also shows the second diameter, D2 (in mm), and the
percentage diameter obstruction (%D). The value of %D is the stenotic
index based on the diameters. The hospital name and the examination date
are displayed at the top of the image. When the light above the menu key
goes out, the program has finished. The results remain on display and can be
stored in the Photofile. Patient information and text can be added.
Step 1
Free Format menu.
This step follows on from ‘step 3’ of the Vessel Diameter Standard Procedure,
after FF (Free Format) has been selected. A menu appears for entry of the
catheter diameter in mm (3F = 1 mm).
.
0
1
2
3
4
5
6
8
9896 001 33152
9
0
OK
CF = 0.115 mm/pixel
33722080
N OT E If a size has been entered or selected previously, it will be displayed in the Enter field,
and can be selected by moving the arrow to the OK box and clicking the left mouse
button [25] to confirm.
Step 2
.
0
1
2
3
4
5
6
8
9
0
OK
CF = 0.115 mm/pixel
33722090
1 Move the arrow to the first digit of the value to be entered and click the left
mouse button [25]. Any existing Free Format size will be erased and the first
digit selected will be inserted in the Enter field.
2 Enter the next digit as described above. Up to 9 digits and a decimal point
can be entered.
3 The last-entered digit can be removed by clicking the right mouse button
[27]. Digit selection can then be resumed.
.
0
1
2
3
4
5
6
8
9
0
OK
CF = 0.115 mm/pixel
33722100
4 Once the required size (in millimeters) has been entered, move the arrow to
the OK box and click the left mouse button [25] to confirm. The calibration
9896 001 33152
factor is displayed at the bottom of the image. The menu reappears as shown
below.
Philips Medical Systems
Calibration
Measurement
CF = 0.108 mm/pixel
33722010
5 Click the left mouse button [25] to continue with the Vessel Diameter
Measurement Procedure (‘step 4’).
The AVA program can be run on images or an image series displayed in non-
subtracted or subtracted mode with the injected contrast medium displayed
as black or white. The section on the AVA Standard Procedure describes the
standard procedure for catheter calibration and measurement. The section on
the AVA Manual Selection Procedures describes manual procedures for
selecting the obstruction and reference diameters. Free Format AVA
Calibration Procedures describes calibration procedures for free format
selection to suit the size of catheter used and the calibration procedures to be
used with an object of known size are described in the section: AVA Manual
Calibration Procedure. This calibration procedure can be used when the AVA
program is run on an image series in which the catheter is not displayed or
9896 001 33152
References
1 Van der Zwet, P.M.J. et al. A new approach for the automated definition
of path lines in digitized coronary angiograms. Int. J. Cardiac Imaging, 5,
no. 2-3, 1990. pp. 75-83.
2 Van der Zwet, P.M.J. et al. An on-line system for the quantitative analysis
of coronary arterial segments. Computers in Cardiology, 1990.
Procedure
PROGRAM MENU 3
PREVIOUS
NEXT
33722110
1 Select the correct image series to which the program is to be applied. The
correct image series must be selected before proceeding.
2 Press one of the three menu keys [95, 96 or 97] to display the menu. The
indicator above key remains on while the function is active.
3 If necessary, select Previous/Next to access more items.
4 Use the left mouse button [25] to select ‘Automated Vessel Analysis’.
The AVA menu is displayed. This contains the items ‘Exit Program’,
‘Catheter Calibration’, ‘Manual Calibration’ or ‘Measurement’. If no
calibration factor has been determined, the ‘Measurement’ item will not
appear.
Exit Program
Catheter Calibration
Manual Calibration
Measurement
CF = 0.127 mm/pix
9896 001 33152
WA R N I N G If these calibration guidelines are not followed, the measurements may be inaccurate
or unreliable.
Step 1
Selecting an image.
33722130
SELECT IMAGE FOR CALIBRATION
Accept ----- Menu
1 If necessary, click the right mouse button [27] to return to the AVA menu.
2 Select an image with a clear view of the catheter and click the left mouse
button [25] to confirm the image selection.
Do not alter the image processing settings (contrast, brightness, etc) once an
image has been selected. The following prompt will appear: ‘Dark vessels on
bright background? Yes — No’.
3 Click the left mouse button [25], if the injected contrast medium or the
catheter is to be displayed as black on white, or click the right mouse button
[27], if the injected contrast medium or the catheter is to be displayed as
white on black.
Step 2
Selecting the segment for calibration.
The catheter segment used for calibration must be straight. It must not be
9896 001 33152
tapered or curved.
Philips Medical Systems
33722140
INDICATE START POINT FOR CALIBRATION
Accept ----- Menu
33722150
4 Move the center dot of the cursor to the end point of the catheter segment to
be used for calibration.
5 If necessary, click the middle mouse button [26] to select another start point.
6 Click the left mouse button [25] to confirm the end point and selected
9896 001 33152
segment.
Step 3
Automatic detection of contours.
Philips Medical Systems
Once the start and end points of the catheter segment have been defined, a
path line between these points will be detected automatically.
The following steps are performed automatically:
• the path line in the catheter segment is defined
• the contour-finding process starts
• the image is zoomed. The twofold magnification (zoom) is not performed
if the segment length exceeds a quarter of the image size
33722160
DETECTED CONTOURS
Accept Hide New segment
1 After the final contours have been displayed, click the middle mouse button
[26] to hide the detected contours and check their accuracy.
2 Click the left mouse button [25] to confirm the final catheter contours.
3 If necessary, click the right mouse button [27] to select a new segment for
calibration. The program returns to ‘step 2’.
N OT E S If the calibration factor is not sufficiently accurate, the following warning is displayed:
‘Calibration factor accuracy only xx%’ (‘xx’ is the accuracy value).
• click the left mouse button [25] to accept this accuracy and the final catheter
contours.
• click the right mouse button [27] to select a new segment for calibration if the
accuracy is not acceptable. The program returns to ‘step 2’.
• or double click the right mouse button [27] and select Manual Calibration to carry
out calibration using an object of known size (see AVA Manual Calibration
Procedure below).
Step 4
Selecting the size.
A menu is now displayed over the image listing French sizes (F4 to F9) and
Free Format (FF).
9896 001 33152
Philips Medical Systems
F4
F5
F6
F7
F8
F9
FF
1,7
33722170
Accept ----- -----
1 Move the arrow to select the desired F-size and click the left mouse button
[25]. For Free Format, see Special AVA Calibration Procedures below.
N OT E If a size has previously been entered or selected, it is displayed in a box at the bottom
of the menu.
Exit Program
Catheter Calibration
Manual Calibration
Measurement
CF = 0.204 mm/pix
SELECT MENU ITEM
33722180
The calibration factor in millimeters per pixel (display pixel) is shown in the
lower right corner of the screen and the AVA menu (‘Exit program’, ‘Catheter
calibration’, ‘Manual calibration’ and ‘Measurement’) is redisplayed. Since
calibration has now been performed, the arrow points to the Measurement
box by default.
2 Use the left mouse button [25] to select an item.
9896 001 33152
AVA Measurement
Step 5
Philips Medical Systems
Selecting an image.
1 Use the image selection keys on the viewing console to select an appropriate
image. The following criteria are important for the image selection process:
• the vessel segment should be well filled with contrast medium
• the obstructed vessel segment should be clearly visible, preferably without
any overlap with other vessels or side branches.
33722190
SELECT IMAGE FOR MEASUREMENT
Accept ----- Menu
Do not change the image processing settings (contrast, brightness, etc) once
an image has been selected.
The following prompt appears: ‘Dark vessels on bright background? Yes —
No’.
• Click the left mouse button [25] if the injected contrast medium is to be
displayed black on white.
• Click the right mouse button [27] if the injected contrast medium is to be
displayed white on black.
Step 6
Selecting the vessel segment.
To guarantee reproducible assessment of the morphology of vessel segments,
such segments should be selected in a standardized manner, i.e. from side
branch to side branch. Therefore, it is preferable to position the start point
immediately distal and the end point immediately proximal, to a branch
point.
9896 001 33152
Philips Medical Systems
33722200
1 Move the center dot of the cursor to the start point of the vessel segment to
be analyzed.
2 Click the left mouse button [25] to confirm the start point.
3 If necessary, click the right mouse button [27] to return to the AVA menu.
33722210
INDICATE END POINT OF VESSEL SEGMENT
Accept ----- New start point
4 Move the center dot of the cursor to the end point of the vessel segment to be
analyzed.
5 Click the left mouse button [25] to confirm the end point and the selected
segment.
6 If necessary, click the middle mouse button [26] to select another start point.
Once the start and end points of the vessel segment have been defined, the
message ‘Image Filtering in Progress’ is displayed and a path line between
these points will be detected automatically. The purpose of the image
filtering procedure is to ensure reliable detection of the path line in the
selected segment. If no path line can be found, the message ‘Cannot find
path line’ will be displayed and the program returns to the beginning of ‘step
6’.
Step 7
Confirming or correcting the path line.
9896 001 33152
Philips Medical Systems
33722220
PATHLINE
Accept Correct Start Point
The path line must lie entirely inside the selected vessel segment but does not
necessarily have to follow the centerline of the vessel segment precisely.
1 Click the left mouse button [25] to accept the path line and proceed with
‘step 8’.
2 Click the right mouse button [27] to return to ‘step 6’ and select a new start
point for the vessel segment.
3 If necessary, click the middle mouse button [26] to correct the path line. A
cursor appears on the image.
33722230
INDICATE CORRECTION POINT TO
TO BE
BE INCLUDED IN PATHLINE
Accept ---- Reject
4 To change the path line, place the cursor at the point to be included in the
path line (off the current path line but inside the selected vessel segment) and
click the left mouse button [25] to confirm. A new path line will then be
drawn through this point. If a new path line cannot be found, the previous
path line will be redisplayed with the message ‘Cannot find corrected path
line’. More corrections can be made by repeating the correction procedure
until a satisfactory result has been obtained. 33722240
PATHLINE
Accept Correct Initial pathline
9896 001 33152
5 Click the right mouse button [27] to restore the initial path line and return
to the beginning of ‘step 7’.
6 Or click the left mouse button [25] to confirm the path line.
Philips Medical Systems
Step 8
Automatic detection of contours.
After the path line has been defined, the contour-finding process starts.
The following steps are performed automatically:
• the contour-finding process starts. The message ‘Initial Contour Detection
in Progress’ is displayed
DETECTED CONTOURS
Accept Hide Correct
33722250
1 Once the final contours have been displayed, press the middle mouse button
[26] to hide the contours and check their accuracy.
2 Click the left mouse button [25] to accept the detected contours and proceed
to ‘step 10’.
3 Click the right mouse button [27] to correct a segment of the detected
contours.
Step 9
Correcting a contour segment.
A cursor appears on the image.
9896 001 33152
Philips Medical Systems
INDICATE DISTAL
DISTAL POINT OF ERRONEOUS
ERRONEOUS CONTOUR PART AND ERASE
33722260
1 Position the cursor over the contour, distal to the segment to be corrected
and press the right mouse button [27]. The contour will be erased until this
button is released.
N OT E Erasure always takes place towards the start point. The cursor must therefore be
placed at the end of the segment to be erased. When the cursor reaches the end of the
contour, it jumps to the start of the remaining part of the contour and erasure
continues from there.
CONTOUR CORRECTION
Accept Draw Erase
33722270
2 Draw the new contour segment by holding down the middle mouse button
[26] and dragging with the mouse.
3 Click the left mouse button [25] to close the contour segment with a straight
line and to confirm the final vessel contours. If the new contour segment
differs significantly from the original, a new contour will be calculated
automatically within a limited region.
DETECTED CONTOURS
Accept Hide Correct
33722280
been obtained.
5 Click the left mouse button [25] to confirm all corrections made.
Step 10
Philips Medical Systems
diameter stenosis which is marked on the image by a straight line. The values
of the obstruction and reference diameters in millimeters are displayed in the
top left corner.
Obs : 3.84 mm
Ref : 6.85 mm
33722290
1 Click the right mouse button [27] to return to ‘step 6’ and select a new vessel
segment.
2 Click the middle mouse button [26] if the automatically selected obstruction
and reference positions are not appropriate, or if it is necessary to analyze
more than one obstruction in the selected segment, and proceed as described
in the section on AVA Manual Selection Procedures.
3 Click the left mouse button [25] to confirm the automatically-defined
obstruction.
Step 11
Result page.
9896 001 33152
33722300
This page shows the anatomy of the obstruction along with the most relevant
geometrically-derived parameters. The black rectangular box shows the
diameter function for the analyzed vessel segment. Vessel diameter in
millimeters (vertical axis) is plotted against position along the centerline
(horizontal axis) from start to end point. The straight line through the
diameter function is the reference diameter function, defining the estimated
geometry of the vessel before disease occurred, as reconstructed from the
available diameter values. The position of the minimum obstruction
diameter and the proximal and distal boundaries are indicated by vertical
lines. These positions are also shown on the vessel segment. The areas
between the reconstructed reference contours and the luminal contours
within the obstructed region are shaded; the shaded area is a measure of the
area of the plaque as seen in this particular view. The obstruction and
reference diameters (Obs.D., Ref.D.), the percentage diameter (%D-Sten)
and the percentage area (%A-Sten) stenosis, the calibration factor (CF), the
symmetry index and the obstruction length (Obs. len.) are shown at the
bottom of the image. The percentage area stenosis is based on the
assumption of circular cross-sections at the obstruction and reference
positions. The symmetry index is 1.0 for a completely symmetric plaque area
and 0.0 for a totally eccentric area. The hospital name and the date are shown
at the top of the image.
• Click the right mouse button [27] to store the displayed result in the
photofile.
• To define another obstruction and/or reference site (return to ‘step 10’) or
to select a new vessel segment (return to ‘step 6’), click the left mouse
button [25] followed by the left button again or the right mouse button
respectively.
Step 12
Exiting the program.
1 To return to the AVA menu, click the left mouse button [25] and then the
right mouse button [27] when the result page is displayed.
Exit Program
Catheter Calibration
Manual Calibration
Measurement
CF = 0.204 mm/pix
SELECT MENU ITEM
33722310
2 Select ‘Exit program’ and click the left mouse button [25].
9896 001 33152
N OT E The program can be exited at any time by pressing the active program menu key on the
viewing console.
Step 1
Redefining the obstruction site.
This step follows on from ‘step 10’ of the AVA Standard Procedure.
Obs : 3.84 mm
Ref : 6.85 mm
33722320
1 Hold down the middle mouse button [26] and drag the reference site with
the mouse. The obstruction and reference diameters, displayed in the top left
corner of the image, are automatically updated. The reference value is
determined from the computed reference diameter function.
Obs : 4.41 mm
Ref : 6.89 mm
Step 2
Redefining the reference site.
9896 001 33152
Obs : 3.84 mm
Ref : 6.85 mm
REFERENCE SITE
Accept Redefine New segment
33722340
1 Click the left mouse button [25] to accept the automatically-calculated
reference diameter at the user-defined obstruction site and go on to step 4 in
this section.
2 Click the middle mouse button [26] to redefine the reference site.
Step 3
Changing the reference site.
The reference diameter displayed in the top left corner of the image will now
be measured between the actual arterial luminal contours. This is clearly
different from the situation described in ‘step 1’ of this section.
Obs : 4.41 mm
Ref : 4.41 mm
1 Hold down the middle mouse button [26] and drag reference site with the
9896 001 33152
mouse.
Philips Medical Systems
Obs : 4.41 mm
Ref : 6.43 mm
33722360
2 Click the right mouse button [27] to restore the automatically-calculated
reference diameter at the location of the user-defined obstruction site. The
program returns to step 2 of this section.
3 Or click the left mouse button [25] to confirm the new reference diameter.
Step 4
Result page.
A full description of the Result page is given in ‘step 11’ of the AVA Standard
Procedure.
33722370
On the result page the reference diameter will now be indicated by a broad
9896 001 33152
section AVA Standard Procedure) double click the left mouse button. To
select a new vessel segment (return to ‘step 6’ of section AVA Standard
Procedure), click the left mouse button [25] followed by the middle
button [26].
To return to the AVA menu, click the left mouse button [25] followed by the
right mouse button [27].
Step 1
Free Format menu.
This step follows on from ‘step 4’ of the AVA Standard Procedure. The menu
displayed lists a choice of French sizes or Free Format (FF).
N OT E If a size has been entered or selected previously, it is displayed in a box at the bottom
of the menu. To select this value, point to it with the mouse and click the left mouse
button [25].
F4
F5
F6
F7
F8
F9
FF
F4
1 Move the arrow to ‘FF’ (Free Format) and click the left mouse button [25]. A
menu is displayed for entry of the size in millimeters (3F = 1mm).
Step 2
Entering the size.
.
0
1
2
3
4
5
6
9896 001 33152
7
8
9
OK
To select the size last entered or selected (displayed in the ‘Enter’ field):
1 Move the arrow to the bottom box (OK) and click the left mouse button to
confirm.
.
0
1
2
3
4
5
6
7
8
9
OK
33722400
5 After the required size (in millimeters) has been entered, move the arrow to
the bottom box (OK) and click the left mouse button [25] to confirm.
Exit Program
Catheter Calibration
Manual Calibration
Measurement
9896 001 33152
Philips Medical Systems
CF = 0.204 mm/pix
33722410
PROGRAM MENU 3
PREVIOUS
NEXT
3 Use the mouse to move the arrow to ‘AVA’ and click the left mouse button
[25] to confirm. If necessary, select ‘Next/Previous’ to access more items.
The AVA menu is displayed.
9896 001 33152
Philips Medical Systems
CF = 0.204 mm/pix
SELECT MENU ITEM
33722420
Accept ----- -----
4 Move the arrow to ‘Manual calibration’ and click the left mouse button [25]
to confirm the selection.
Step 1
Selecting the image and zoomed area.
After Manual Calibration has been selected, the image is zoomed
automatically. The image can be panned by moving the mouse.
1 Use the image selection keys on the viewing console to select an image.
2 Click the right mouse button [27] to return to the AVA menu.
3 Click the left mouse button [25] to confirm the image selection and zoomed
9896 001 33152
area.
Step 2
Drawing a calibration line.
The cursor displayed on the screen can be controlled using the mouse.
Philips Medical Systems
33722440
1 Hold down the middle mouse button [26] and drag with the mouse to draw
a calibration line of known length.
2 Click the right mouse button [27] to reposition the start point of the line.
3 Click the left mouse button [25] to confirm the calibration line.
N OT E S If the calibration factor is not sufficiently accurate, a warning is displayed: ‘Calibration
factor accuracy only xx%’ (where ‘xx’ is the accuracy value).
- click the left mouse button [25] to accept this accuracy.
- or click the right mouse button [27] to reject the accuracy and return to the AVA
menu. Accuracy can be improved by using a larger object for calibration.
Step 3
Entering the size.
A menu is now displayed over the image allowing entry of any millimeter
value for the size of the calibration object.
.
0
1
2
3
4
5
6
7
8
9
OK
To select the size last entered or selected (displayed in the ‘Enter’ field):
1 Move the arrow to the bottom box (OK) and click the left mouse button
[25] to confirm.
To enter a new value:
2 Move the arrow to the first digit of the value to be entered and click the left
mouse button [25]. The selected digit is entered in the ‘Enter’ field, erasing
the previous free format size.
3 Move the arrow to the next digit to be entered and click the left mouse
button [25]. Proceed in this way until the required size has been entered.
A maximum of nine digits can be entered, with the decimal point in any
position.
4 If necessary, click the right mouse button [27] to remove the last digit
entered.
33722460
5 After the required size (in millimeters) has been entered, move the arrow to
the bottom box (OK) and click the left mouse button [25] to confirm.
CF = 0.207 mm/pix
SELECT MENU ITEM
33722470
box by default.
N OT E If the image series containing the calibration object is the same as that on which AVA is
to be performed, select Measurement. Click the left mouse button [25] to confirm the
choice and proceed with ‘step 5’ of ‘AVA Measurement’ in the AVA Standard
Procedure.
Step 4
Selecting the program.
CF = 0.207 mm/pix
SELECT MENU ITEM
33722480
Accept ----- -----
1 Use the mouse to point the arrow at the Exit Program box and click left
mouse button [25] to confirm.
2 Select the image series on which AVA is to be performed.
3 Press one of the three menu keys [95, 96 or 97]. The indicator above the key
remains on while the function remains active.
PROGRAM MENU 3
PREVIOUS
NEXT
6 Select ‘Measurement’ and click the left mouse button [25] to confirm.
7 Go on to ‘step 5’ of ‘AVA Measurement’ in the AVA Standard Procedure.
Philips Medical Systems
References
Procedure
• calibration:
- step 1 Selecting the program
- step 2 Selecting an image and zoomed area
- step 3 Drawing a calibration line
- step 4 Selecting free format
- step 5 Selecting the size of the calibration line.
2 Automated left ventricular analysis procedure:
• selecting the program
• defining the ED contour:
- step 1 Selecting an image
- step 2 Defining the valve plane
- step 3 Defining the position of the apex
- step 4 ED contour.
• defining the ES contour:
- step 1 Selecting an image.
- step 2 Defining the valve plane
- step 3 Defining the position of the apex
- step 4 ES contour.
• analysis results:
- step 1 First Result Page.
- step 2 Entering height and weight. Second Result Page.
- Step 3 Entering heart rate. Third Result Page.
- Step 4 Defining the epicard wall. Fourth Result Page.
- Step 5 Result Overview.
3 Fine tuning:
• manual contour correction.
4 ALVA recalculation:
• performing the ALVA recalculation.
5 ALVA properties:
• volume calculation
• BSA input
• text overlay in photofile.
Calibration
N OT E The accuracy with which the calibration is performed has a direct influence on the
reliability of the left ventricular volumes calculated by the automated left ventricular
analysis program. Large deviations from expected volumes are generally due to
inaccurate calibration.
9896 001 33152
distortion
• the image intensifier field size must not be greater than 9 inches
• make sure that the geometry of the X-ray stand and table is the same as
that used during the left ventricular exposure, this involves choosing the
same image intensifier field and the same geometric enlargement. In other
words, the calibration object must be in the same position as the heart was
during image acquisition.
Step 1
33725220
1 Press one of the three menu keys [95, 96 or 97] on the Viewing console. The
key remains lit as long as the function is active.
2 Use the left mouse button [25] to select the item ‘Calibration’.
3 If necessary, select Next/Previous screen to access more items.
After the calibration program has been selected, the image is zoomed
automatically. The image can be panned using the mouse.
Step 2
Selecting an image and zoomed area.
Image : Select -- -- --
33721120
9896 001 33152
N OT E The ‘message line’ at the bottom of clinical images indicates the current actions of the
mouse buttons. For example, in the above image, the left mouse button is used for the
Philips Medical Systems
‘Select’ command and the middle and right buttons have no action.
1 Use the image selection keys on the viewing console to select an image.
2 Click the left mouse button to confirm the image selection and zoomed area.
The cursor displayed on the screen can be controlled using the mouse.
Step 3
Drawing a calibration line.
Diameter
Diameter :: Accept
Accept Draw
Draw Reject
Reject
33721130
1 Hold down the middle mouse button [26] and drag with the mouse to draw
a calibration line of known length.
2 Click the right mouse button [27] to reposition the starting point.
3 Click the left mouse button [25] to confirm the calibration line. A menu is
now displayed over the image listing French sizes (F4 to F9) and free format
(FF).
Step 4
Selecting free format.
33721140
1 Move the arrow to ‘FF’ (free format) and click the left mouse button [25]. A
9896 001 33152
box menu is displayed for entering the size (in millimeters) of the calibration
object line.
N OT E The last item on the menu is the size last entered or selected. When using standard size
objects, this value can be selected directly by pointing to it with the mouse and clicking
Philips Medical Systems
Step 5
Selecting the size of the calibration line.
To select the size last entered or selected (displayed in the ‘Enter’ field):
1 Move the arrow to the bottom box (OK) and click the left mouse button to
confirm.
33721150
To enter a new value:
1 Move the arrow to the first digit of the value to be entered and click the left
mouse button [25]. The selected digit is entered in the ‘Enter’ field, erasing
the previous free format size.
2 Move the arrow to the next digit to be entered and click the left mouse
button [25]. Proceed in this way until the required size has been entered. A
maximum of nine digits can be entered, with the decimal point in any
position.
3 If necessary, click the right mouse button [27] to remove the last digit
entered.
Enter (mm) : 5
33721160
9896 001 33152
4 After the required size (in millimeters) has been entered, move the arrow to
the bottom box (OK) and click the left mouse button [25] to confirm.
Philips Medical Systems
Enter (mm) : 50
33721170
Calibration is now complete. The calibration factor, CF, is displayed on the
image in millimeters per pixel (display pixel).
CF = 0.145 mm/pixel
33721180
33751190
1 Press one of the three menu keys [95, 96 or 97]. The corresponding indicator
remains on while the function is active.
2 Use the left mouse button [25] to select the item ‘automated left ventricular
analysis (ALVA)’.
3 If necessary, select Next/Previous screen to access more items.
Step 1
Selecting an image.
Exit
Measurement with standard CF
Measurement with Auto CF
1 Select the end diastolic (ED) image using the image selection keys on the
9896 001 33152
viewing console. Only cardiac cycles free from extrasystolic contractions are
appropriate.
2 Click the left mouse button [25] to confirm selection of the ED image. A
Philips Medical Systems
Step 2
Defining the valve plane.
33721210
Define the valve plane in this RAO view as follows:
1 Place the cursor at one end of the aortic valve line.
2 Click the middle mouse button [26] to define this point as one of the end
points of the valve line.
3 Define the other end point of the valve line contour by moving the mouse
while holding [26] down.
4 Click the left mouse button [25] to accept the valve line.
5 If necessary, click the right mouse button [27] to redefine the first end point
of the valve line.
Step 3
Defining the position of the apex.
.
1 Move the cursor to the apex point and click the left mouse button [25] to
Philips Medical Systems
Step 4
ED contour.
33721230
1 Press and hold down the middle mouse button [26] to temporarily remove
the contour line from the image and check its accuracy.
2 Click the left mouse button to confirm the ED contour if no correction is
necessary.
Step 1
Selecting an image.
9896 001 33152
1 Use the image selection keys to select the end systolic (ES) image following
the current ED image.
2 Click the left mouse button [25] to confirm this selection.
The cursor is displayed on the image and the valve line and apex can be
defined in the same way as for the ED image (steps 2 and 3). The ES contour
is then calculated. The contour is displayed and the cursor appears in the
center of the image.
Step 4
ES contour.
33721250
1 Press and hold down the middle mouse button [26] to temporarily remove
the contour line from the image and check its accuracy.
2 Click the left mouse button to confirm the ES contour if no correction is
necessary.
The procedure for manual correction is described in the section on
‘Automated left ventricular analysis fine tuning’.
Analysis results
Step 1
First ‘result’ page.
EF : 55% a
EDV : 172.0 ml
ESV : 77.0 ml
SV : 94.9ml
CF : 0.159 mm/px
The asterisk [a] denotes that the EF value is based on a non-default (user
Philips Medical Systems
• ‘Next’: continues the ALVA procedure and generates the next result page
• ‘Results’: shows the calculated results on the results overview
• ‘Store’: stores the ‘result’ page in the photofile. Depending on the ALVA
configuration (see Section 5 ‘ALVA properties’) the result page will be
stored with or without patient data.
1 If desired, click the right mouse button [27] to store the first ‘result’ page in
the photofile.
2 Click the left mouse button [25] to continue.
ALVA procedure
Required
inputs Back
1st "result"page Results overview
Next Quit
Next
3rd "result"page
Results
Heart rate
Next
4th "result"page
Results
Epicard wall
or next
33725240
Step 2
9896 001 33152
33725250
2 Use the mouse to select the patient’s height in meters.
To enter a patient height of 1.80 m:
1 Move the arrow to position ‘1’ and click the left mouse button [25].
2 Repeat step 1 for the decimal point and other digits.
3 Confirm the entry by moving the arrow to ‘OK’ and clicking the left mouse
button [25]. The patient’s weight is now requested.
EF : 55%
SV : 94.9 ml
BSA : 2.00 m2
EDV(I) : 86.0 ml /m2
ESV(I) : 38.5 ml /m2
SV(I) : 47.5 ml /m2
EDV : 172.0 ml
ESV : 77.0 ml
CF : 0.159 mm/px
33725260
Second result page
This page shows the ED image, with the longest axes of the ED and ES
contours, and the value of the ejection fraction (EF [%]). The end diastolic
volume (EDV [ml]), end systolic volume (ESV [ml]), stroke volume (SV
[ml]), and calibration factor (CF [mm/pixel]) are also displayed. In addition,
the body surface area (BSA [m2]) and the BSA-index volumes EDV(I),
ESV(I), and SV(I) are displayed in ml/m3.
1 If desired, click the right mouse button [27] to store the second ‘result’ page
in the photofile.
Step 3
Entering heart rate.
1 Click the left mouse button [25] to obtain the third ‘result’ page. Before
displaying this page the system requests additional input from the user.
9896 001 33152
2 Use the mouse to enter the heart rate in beats per minute.
EF : 55%
HR : 70 BMP
C0 : 6.64 1/min
CI : 3.32 1/min
33725270
Third result page
This page shows the ED image, with the longest axes of the ED and ES
contours, and the value of (1)the ejection fraction (EF [%]). Since calibration
has been performed and the heart rate (HR [bpm]) has been entered, the
cardiac output (CO [l/min]) is displayed.
The cardiac index (CI [1/min.m2]) is also displayed, since the BSA has either
been entered or calculated.
1 If desired, click the right mouse button [27] to store the third ‘result’ page in
the photofile.
Step 4
Defining the epicard wall.
1 Press the left mouse button [25] to obtain the fourth ‘result’ page. Additional
user input is required before this page can be displayed.
2 Drag with the mouse while holding down the middle mouse button [26] to
Philips Medical Systems
define the epicard wall. If necessary, click the right mouse button to redefine
the outline of the epicard wall.
33721330
3 Click the left mouse button to confirm the position of the epicard wall.
EF : 55 %
WT : 9.1 mm
WV : 225 ml
WM : 239 g
WS : 4.90
This page shows the ED image, with the longest axes of the ED and ES
contours, the epicard wall contour, and the value of the ejection fraction (EF
[%]). The calculated wall thickness (WT [mm]), wall volume (WV [ml]),
wall mass (WM [g]), and normalized wall stress (WS) are also displayed.
1 If desired, click the right mouse button [27] to store the fourth ‘result’ page
into the photofile.
2 Click the left or middle mouse button [25 or 26] to obtain the Results
9896 001 33152
overview.
Step 5
Results overview.
Philips Medical Systems
EF : 55 %
EDV : 172.0 ml
ESV : 77.0 ml
SV : 94.9 ml
EDV(I) : 86.0 ml/m2
ESV(I) : 38.5 ml/m2
SV(I) : 47.5 ml/m2
CO : 6.64 1/min
CI : 3.32 1/(min m2)
WT : 9.1 mm
WV : 225 ml University Medical Center
WM : 239 g Dr. Cardiologist
WS : 4.90 18-01-1998
CF : 0.159 mm/px
BSA: 2.00 m2 John Doe
05-01-1951
HR : 70 bpm H123456789
Volume calculation method:
Area Length
Volume correction method:
Vol.= 0.787 Vol.- 3.759
33725290
(user defined) Back Quit Store
Results overview
If all ‘result’ pages have been generated, this page shows the ED image, with
the longest axes of the ED and ES contours, the epicard wall contour
together with all calculated results (from ‘result’ pages 1 to 4). It also
indicates the methods used for volume calculation and volume correction,
and gives the patient data. If results overview is selected after the first, second
or third ‘result’ page, only those results that are currently available are
displayed.
To access other pages:
• click the left mouse button [25] to view the first ‘result’ page
• click the middle mouse button [26] to exit the ALVA program
• click the right mouse button [27] to store the Results overview in the
photofile.
3. Fine tuning
Correct Contour : Accept Draw Erase Correct Contour : Accept Draw Erase
33721350
Correct Contour : Accept Draw Erase Correct Contour : Accept Draw Erase
4. ALVA recalculation
If required, the volumes and other parameters can be recalculated without
Philips Medical Systems
33725220
Performing the ALVA recalculation
1 Select the correct image series for which the ALVA recalculation is to be
performed. The volumes for this series are then recalculated using the
calibration factor that has been used during the ALVA procedure.
2 Press one of the three menu keys to display the menu. The indicator remains
on while the function is active.
3 Use the left mouse button [25] to select ALVA recalculation. The ALVA
program will begin at the first ‘result’ page, showing the results based on the
contours previously defined.
4 Click the left mouse button [25] to obtain the next ‘result’ page; Additional
user input is required only for those items not previously entered during the
standard ALVA procedure.
• Click the middle mouse button [26] to obtain the results overview or
• Click the right mouse button [27] to store the ‘result’ page in the
photofile.
5. ALVA properties
The operator can use the ALVA properties section to change the
configuration settings used in the ALVA program.
The configuration settings are:
• volume calculation method:
- area length or Simpson
- volume correction formula (= regression equations).
9896 001 33152
• BSA input:
- BSA value or heigth and weight.
• text overlay in photofile.
Philips Medical Systems
33725300
Changing the ALVA properties
1 Press one of the three menu keys to display the menu. The indicator remains
on while the function is active.
2 Use the left mouse button [25] to select ALVA properties. The ALVA
properties main screen is displayed.
ALVA PROPERTIES
Exit program
Volume Calculation
BSA input
Text overlay in photofile
Volume calculation
Select ‘Volume calculation’ to change the method used for volume
9896 001 33152
ALVA PROPERTIES
33725320
Method Correction formula Menu
The volume calculation page
N OT E Mathematical models are used to calculate the left ventricular volumes. As these
models only approximate the complex shape of the left ventricle (Vmeasured) regression
equations are applied to correct the volumes. These regression equations are based on
data obtained from postmortem casts. The format of the regression equation for the
volume calculation is:
V = (slope value) * Vmeasured + (offset value)
Both the slope and offset values are regression factors. These factors are dependent on
the model used and the projection view during acquisition. The default regression
equations for both the Area length method and the Simpson’s rule method used in the
ALVA program are valid for RAO 30° projections (ref. 3).
1 Click the left mouse button [25] to change the method of volume
calculation.
ALVA PROPERTIES
• Click the left mouse button [25] to select the Area length method. The
volume calculation page is displayed (Area length method previously
selected).
• Click the right mouse button [27] to cancel. The volume calculation page
is displayed.
• Click the middle mouse button [26] to select the Simpson method. The
volume calculation page is displayed showing the new method of
calculation to be used.
ALVA PROPERTIES
33725340
CHANGE VOLUME CALCULATION METHOD OR FORMULA
Method Correction formula Menu
WA R N I N G If regression factors other than the default factors are used, the calculated left
ventricular volumes could be incorrect. It is therefore strongly recommended
to use the program default regression factors only.
If the regression factors are changed, an asterisk “*” is displayed after the EF
value in the ALVA program.
N OT E When a volume calculation method is selected all user defined regression factors are
lost and replaced by the default regression factors.
1 Click the middle mouse button [26] to change the correction formula.
ALVA PROPERTIES
• Click the left mouse button [25] to use the default correction formula
belonging to the selected volume calculation method. The volume
calculation page is displayed.
• Click the right mouse button [27] to cancel. The volume calculation page
is displayed.
• Click the middle mouse button [26] to change the correction formula
belonging to the selected volume calculation method. A box menu is
displayed for entering the slope value.
ALVA PROPERTIES
.
1
2
3
4
5
6 Volume calculation method: Simpson´s rule
7 Volume correction formula:
8 Volume = 0.737 Volume - 4.649 (default)
9
OK
Slope Value 0.737000
33725360
Enter Slope Value: Insert ------- Remove
2 Move the arrow to the digit or decimal point to be inserted and click the left
mouse button [25].
3 Repeat step 2 to enter the complete slope value. If necessary, click the right
mouse button [27] to delete the last digit entered.
4 Confirm the new slope value by moving the arrow to ‘OK’ and clicking with
the left mouse button [25]. A box menu is displayed for entering the offset
value.
ALVA PROPERTIES
-
.
1
2
3
4
5
6 Volume calculation method: Simpson´s rule
7 Volume correction formula:
8 Volume = 0.813 Volume - 4.649 (user defined)
9
OK
Offset Value -4.649000
33725370
5 Use the mouse to enter the offset value. If the offset value to be entered is a
negative value, first select and enter the ‘minus’ sign and then the digits or
decimal point.
6 Confirm the entry by moving the arrow to ‘OK’ and clicking with the left
Philips Medical Systems
ALVA PROPERTIES
CORRECTION FORMULA
33725380
Accept Modify -------
7 If necessary, click the middle mouse button [26] to modify the new
correction formula.
8 Click the left mouse button [25] to accept the new correction formula. The
ALVA properties main screen is displayed showing the updated values.
ALVA PROPERTIES
Exit program
Volume Calculation
BSA input
Text overlay in photofile
BSA input
The ‘BSA input’ page can be used to select either ‘BSA’ or ‘Height/weight’ as
input for the second ‘result’ page in the ALVA program.
1 Move the arrow to ‘BSA input’ on the ALVA properties main screen and click
with the left mouse button [25]. The BSA input page is displayed.
9896 001 33152
Philips Medical Systems
ALVA PROPERTIES
33725400
SELECT BSA INPUT METHOD IN ALVA
BSA Value Height/Weight Menu
The BSA input page
2 Click the left mouse button [25] to select BSA as input for the ALVA
program or click the middle mouse button [26] to select height/weight as
input for the ALVA program.
3 Click the right mouse button [27] to return to the ALVA properties main
screen.
ALVA PROPERTIES
Exit program
Volume Calculation
BSA input
Text overlay in photofile
screen and click with the left mouse button [25]. The ‘Text overlay in
photofile’ page is displayed.
Philips Medical Systems
ALVA PROPERTIES
33725420
SELECT TEXT OVERLAY IN PHOTOFILE
Yes No Menu
The ‘text overlay in photofile’ page
2 Click the left mouse button [25] to store the text overlay with the ALVA
‘result’ pages in the photofile or click the middle mouse button [26] to store
no text overlay with the ALVA ‘result’ pages in the photofile.
3 Click the right mouse button [27] to return to the ALVA properties main
screen.
ALVA PROPERTIES
Exit program
Volume Calculation
BSA input
Text overlay in photofile
the contraction model described in the medical literature. The method aims
to describe the displacement between end diastole and end systole of
particular points on the left ventricular wall. The calculations are considered
valid for the same standard RAO projection that is used for the global
ejection fraction calculation. This clinical program uses the left-ventricular
outlines resulting from the left ventricular (EF) analysis programs thus
avoiding additional user interaction and ensuring correct application
conditions.
Reference
Sheehan, F.H. Advantages and applications of the centerline method for
characterizing regional ventricular function. Circulation, 74, (2), pp. 293-
305, 1986.
Standard procedure
33721110
1 Select the correct image series with which to run the CWM program. The
CWM function uses the information for the image series currently displayed,
so the correct image series must be selected before proceeding.
2 Check that one of the left ventricular (EF) analysis programs has been
applied to the image series.
3 Press one of the three menu keys to display the menu. The indicator remains
on while the function is active.
4 Use the left mouse button [25] to select the centerline wall motion function.
No further user action is required once the CWM program has started.
5 Use the left mouse button [25] to view the next/previous screens.
When the indicator above the menu key goes out, the program has finished.
The results are displayed and can be stored in the Photofile.
The final display shows: 9896 001 33152
Philips Medical Systems
33721550
Top left
The ED image used for left ventricular (EF) analysis with the contours
accepted in the analysis. The centerline is determined between the ED and
the ES contours. 100 equidistant chords perpendicular to this centerline are
calculated. The program ensures that chords do not cross one another. Only
50 chords are displayed.
Top right
This quarter of the display contains a graph of normalized chord lengths
plotted against wall location. The vertical axis represents the chord lengths
normalized to the end diastolic circumference. In other words, the chord
length is expressed as a percentage of the end diastolic circumference. The
horizontal axis represents the locations of the measurement points. The ED
and ES points are matched by the chords. The apex point is indicated on the
horizontal axis. The gray band represents the wall motion parameters for a
normal patient population, collected by the Thorax Center, Erasmus
University and the University Hospital Dijkzigt, Rotterdam, the
Netherlands. The gray band shows the average normal value ± 2 standard
deviations.
Bottom left
This quarter shows the deviations of the local wall motion parameters
compared with the values obtained for the normal patient population. The
gray band indicates the average normal values ± 2 standard deviations (95%
probability). The apex point is indicated on the horizontal axis. If the graph
of wall motion remains within the gray band for the normal population, it
9896 001 33152
Bottom right
This quarter of the display gives general information.
Line 1: Name of the hospital
Philips Medical Systems
Reference
Wong, W.H., Kirkeeide, R.L. and Gould, K.L.; Computer applications in
angiography; in Cardiac Imaging and Image Processing; Collins, S.M and
Skorton, D.J. (eds); McGraw-Hill Book Company; New York 1986; 232-
233.
Standard procedure
33721110
1 Select the correct image series to be analyzed by the RWM program. The
RWM program uses the information for the image series currently displayed,
so the correct image series must be selected before proceeding.
2 Check that one of the left ventricular (EF) analysis programs has been
applied to the image series.
9896 001 33152
3 Press one of the three menu keys to display the menu. The indicator remains
on while the function is active.
4 Use the left mouse button [25] to select the regional wall motion function.
No further user action is required once the RWM function has started.
Philips Medical Systems
5 Use the left mouse button [25] to view the next/previous screens.
When the indicator above the menu key goes out, the program has finished.
The results are displayed and can be stored in the Photofile.
The final display shows:
33721570
Top left
The ED image used for left ventricular (EF) analysis with the contours
accepted during the analysis. The mathematical center of gravity is
determined for both the ED and ES contours. In each contour, the right
mitral valve point is determined and the wall between the left aortic valve
and the right mitral valve points is then divided into the 5 anatomical wall
segments. These segments are shown for the ED contour only. In the
displays, the segments are numbered in clockwise sequence for reference.
Top right
This quarter of the display contains a graph of radial shortening, plotted
against wall location. The vertical axis represents the percentage displacement
of corresponding ED and ES wall locations with reference to the center of
gravity, in other words, the difference in length between the ED and ES radii
divided by the length of the ED radius. The horizontal axis represents the
locations of the wall points. The ED and ES are matched between their left
aortic valve and right mitral valve points. The dashed line indicates the
divisions between the anatomical segments. The numbers at the bottom are
the segment numbers as shown in the top-left quarter of the display.
Bottom left
This quarter of the display gives the numerical results. First, the EF
percentage calculated by the left ventricular (EF) analysis program is
displayed, representing a volume ratio. The second block shows the average
percentage shortening for each of the 5 anatomical segments. The numbers
are the segment numbers used in the top left quarter of the display and the
9896 001 33152
Bottom right
Philips Medical Systems
References
Standard procedure
PROGRAM MENU 2
PREVIOUS
NEXT
1 Select the correct image series to be analyzed by the SWM program. The
SWM program uses the information for the image series currently displayed,
so the correct image series must be selected before proceeding.
2 Check that one of the left ventricular (EF) analysis programs has been
9896 001 33152
4 Use the left mouse button [25] to select Slager wall motion. No further user
action is required once the SWM program has started.
5 Use the left mouse button [25] to view the next/previous screens.
When the indicator above the menu key goes out, the program has finished.
The results are displayed and can be stored in the Photofile.
The final display shows:
33721590
Top
The top half of the display shows a composite graph of CREF values for the
20 segments. CREF values (regional contribution to global EF) are derived
from systolic wall displacement data and left ventricular long-axis shortening.
The individual anterior and posterior CREF values of the patient are
superimposed and connected by straight lines. To compare the quantitative
results with those provided by the usual visual interpretation, the left
ventricular boundary is divided into 5 anatomical regions, denoted
anterobasal, anterolateral, apical, diaphragmatic and posterobasal. The
segments are assigned to these regions and the CREF values for the regions
are plotted as well. The gray band represents the wall motion parameters for a
normal patient population, collected by the Thorax Center, Erasmus
University and the University Hospital Dijkzigt, Rotterdam, the
Netherlands. The gray band shows the average normal value ± 2 standard
deviations.
Bottom left
This shows the ED image chosen for left ventricular (EF) analysis with the
contours accepted during the analysis. Left ventricular segmental wall
motion is computed along 20 straight lines, calculated from a mathematical
expression derived from anatomical landmark trajectories in normal patients.
The 20 lines result from 20 well-defined ED contour points or segments, 10
anterior and 10 posterior. The point or segment numbers are plotted along
the contour. A center of contraction is defined for each pair of 2 opposite ED
contour points.
9896 001 33152
Bottom right
This section of the display shows the ejection fraction as previously
computed from the contours, the name of the hospital, examination date, the
patient’s name and date of birth, and the identification code entered with the
patient’s name
Philips Medical Systems
The ACA Standard Procedure is described first for catheter calibration and
then for measurement. The section on ACA Manual Selection Procedures
describes the manual procedures for selecting the obstruction and reference
diameters. The section on ACA Final Display gives a detailed description of
the final display. The ACA short procedure, which allows measurement with
minimal user interaction, is also described. The section on Free Format ACA
Calibration Procedures describes special calibration procedures, with free
format selection to suit the size of catheter used.
References
Procedure
1 Select the correct image series on which the program is to operate. The
correct image series must be selected before proceeding.
2 Press one of the three menu keys [95, 96 or 97] to display the menu.The
indicator remains on while the function is active.
9896 001 33152
Philips Medical Systems
PROGRAM MENU 1
PREVIOUS
NEXT
ACA Short
ACA
Vessel Diameter
Run Overview
33721600
3 If necessary, select ‘Previous/Next’ to access more items.
4 Use the left mouse button [25] to select ‘ACA’.
The ACA menu is displayed. This contains the following items: ‘Exit
program’, ‘Catheter calibration’ and ‘Measurement’. If no calibration factor
has been determined, the Measurement item will not appear.
Exit Program
Catheter Calibration
Measurement
CF = 0.108 mm/pix
SELECT MENU ITEM
Accept ---- ----
33721610
• the catheter segment chosen for calibration may be straight or curved, but
must not taper.
WA R N I N G If these calibration guidelines are not followed, the absolute measurements may be
inaccurate or unreliable.
Step 1
Selecting an image.
33721620
1 Select an image with a clear view of the catheter.
2 Click the left mouse button [25] to confirm the image selection.
3 If necessary, click the right mouse button [27] to return to the ACA menu.
Step 2
Selecting the segment for calibration.
The catheter segment used for calibration may be straight or curved, but
must not taper.
1 Move the mouse until the central dot of the cursor is positioned at the start
point of the catheter segment to be used for calibration.
2 Click the left mouse button [25] to confirm the start point.
Philips Medical Systems
33721640
3 Place the center dot of the cursor at the end point of the catheter segment to
be used for calibration.
4 Click the left mouse button [25] to confirm the end point and selected
segment.
5 If necessary, click the right mouse button [27] to select another start point.
Step 3
Automatic detection of contours.
After the start and end points of the catheter segment have been selected the
message ‘Image filtering in progress’ is displayed and a path line between
these points will be detected automatically. The purpose of the image
filtering procedure is to ensure reliable detection of a path line in the catheter
segment. If no path line can be found, the message ‘Cannot find path line’
will be displayed and the program returns to the beginning of ‘step 2’.
The following steps are performed automatically:
• the path line in the catheter segment is defined
• the contour-finding process starts
• the image is zoomed. The twofold magnification (zoom) is not performed
if the segment length exceeds a quarter of the image size
• the initially-detected contours are displayed along the catheter segment
and the message ‘Final Contour Detection in Progress’ appears. The
initially-detected contours serve as a model for final contour detection.
The final contour-detection procedure is based on the assumption that the
two contours of a non-tapering catheter are parallel.
9896 001 33152
Philips Medical Systems
DETECTED CONTOURS
Accept Hide New Segment
33721650
1 Once the final contours are displayed, press the middle mouse button [26] to
hide the detected contours and check their accuracy.
2 Click the left mouse button [25] to confirm the final catheter contours.
3 Click the right mouse button [27] to select a new segment for calibration.
The program returns to ‘step 2’.
Step 4
Selecting the size.
A menu is now displayed over the image listing French sizes (F4 to F9) and
free format (FF).
F4
F5
F6
F7
F8
F9
F9
2.7
1 Move the arrow to select the desired F-size and click the left mouse button
9896 001 33152
[25]. For Free Format, see the description of the Special ACA calibration
procedure.
N OT E If a size has previously been entered or selected, it is displayed in a box at the bottom
Philips Medical Systems
of the menu.
Exit Program
Catheter Calibration
Measurement
CF = 0.108 mm/pix
SELECT MENU ITEM
Accept ---- ----
33721670
The calibration factor in millimeters per pixel (display pixel) is displayed in
the lower right corner and the ACA menu (Exit Program, Catheter
Calibration, Measurement) is redisplayed. Since calibration has now been
performed, the arrow points to the Measurement box by default.
2 Use the left mouse button [25] to select an item.
ACA measurement
Step 5
Selecting an image.
1 Use the image selection keys on the viewing console to select an appropriate
image of the coronary arterial segment to be analyzed.
The following criteria are important for image selection:
• the arterial segment should be well filled with contrast medium
• to avoid motion blur, the image should preferably be selected in the
diastasis or end diastolic phase
• the obstructed coronary segment should be clearly visible, preferably
without any overlap with other vessels or side branches.
2 Click the left mouse button [25] to confirm image selection. A cursor
appears for selecting the coronary arterial segment to be analyzed.
3 If necessary, click the right mouse button [27] to return to the ACA menu.
Step 6
Selecting the coronary segment.
To guarantee reproducible assessment of the morphology of coronary
segments, such segments should be selected in a standardized manner,
according to the recommendations of the AHA (Ref.4), i.e. from side branch
to side branch. Therefore, it is preferable to position the start point
immediately distal to a branch point and the end point immediately
proximal to a branch point or at the end of the coronary artery. The start
point should always be proximal to the end point, i.e. the direction from start
point to end point is the same as the direction of the coronary blood flow.
The reason for this constraint is that the computation of the stenotic flow
reserve is dependent on direction.
33721690
1 Place the center dot of the cursor at the start point of the arterial segment to
be analyzed.
2 Click the left mouse button [25] to confirm the start point.
3 If necessary, click the right mouse button [27] to return to the ACA menu.
9896 001 33152
4 Place the center dot of the cursor at the end point of the arterial segment to
be analyzed.
5 Click the left mouse button [25] to confirm the end point.
6 If necessary, click the right mouse button [27] to select another start point.
After the start and end points of the coronary segment have been selected the
message ‘Image filtering in progress’ is displayed and a path line between the
start and end points will be detected automatically. The purpose of the image
filtering procedure is to ensure reliable detection of a path line in the
coronary segment. If the image selected is the same as that used for
calibration, image filtering will not be performed. If no path line can be
found, the message ‘Cannot find path line’ will be displayed and the program
returns to the beginning of ‘step 6’.
Step 7
Confirming or correcting the path line.
The path line must lie entirely inside the selected arterial segment but does
not need to follow the centerline of the arterial segment precisely.
PATHLINE
Accept Correct Startpoint
33721710
1 Click the left mouse button [25] to accept the path line and proceed with
‘step 8’.
2 Click the right mouse button [27] to return to ‘step 6’ and select a new start
point for the arterial segment.
3 If necessary, click the middle mouse button [26] to correct the path line. A
cursor appears on the image.
33721720
4 Position the cursor at the point to be included in the path line (off the path
line but inside the selected arterial segment) and click the left mouse button
[25] to confirm. A new path line will then be drawn through this point. If a
new path line cannot be found, the previous path line will be redisplayed
with the message ‘Cannot find corrected path line’.
PATHLINE
Accept Correct Initial Pathline
33721730
Step 8
Automatic detection of contours.
The contour-finding process starts as soon as the path line has been defined.
Philips Medical Systems
• the initially-detected contours are displayed along the arterial segment and
the message ‘Final Contour Detection in Progress’ appears.
DETECTED CONTOURS
Accept Hide Correct
33721740
1 Once the final contours have been displayed, press the middle mouse button
[26] to hide the detected contours and check their accuracy in relation to the
actual vessel boundaries.
2 Click the left mouse button [25] to accept the detected contours and
continue with ‘step 12’.
3 To correct a segment of the detected contours, click the right mouse button
[27].
Step 9
Correcting the contours.
A cursor appears on the image.
1 Position the cursor over the contour, distal to the part to be corrected and
press the right mouse button [27]. The contour will be erased until this
Philips Medical Systems
button is released.
N OT E Erasure always takes place towards the start point. The cursor must therefore be
placed at the end of the segment to be erased. When the cursor reaches the end of the
contour, it will jump to the start of the remaining part of the contour and erasure will
continue from there.
CONTOUR CORRECTION
Accept Draw Erase
33721760
2 Draw the new contour segment by holding down the middle mouse button
[26] and dragging the mouse.
3 Click the left mouse button [25] to close the contour section with a straight
line and to confirm the final arterial contours. If the new contour segment
differs significantly from the original, a new contour will be calculated
automatically within a limited region.
DETECTED CONTOURS
Accept Hide Correct
33721770
Step 10
9896 001 33152
centerline. This defines the diameter function. This function is used to derive
reference contours, which represent the best approximation to the original
size of the vessel before disease occurred. The obstruction and reference
diameters are automatically calculated for the point of maximum percentage
diameter stenosis, whose location is indicated by a straight line, and are
displayed in millimeters in the upper left corner.
33721780
1 Click the right mouse button [27] to return to ‘step 6’ to select a new vessel
segment.
2 Click the middle mouse button [26] and carry out the ACA Manual
Selection Procedure if the automatically selected obstruction and reference
positions are not appropriate, or if it is necessary to analyze more than one
obstruction in the selected segment.
3 Click the left mouse button [25] to confirm the automatically-defined
obstruction.
Step 11
Result pages.
33721790
RESULTS
RESULTS: Continue Second Page Store Continue Second Page Store
First result page Second result page
• Use the middle mouse button [26] to select the first and second result
pages.
9896 001 33152
• Click the right mouse button [27] to store the result page in the photofile.
• To redefine another obstruction and/or reference site (return to ‘step 10’)
or to select a new vessel segment (return to ‘step 6’), click the left mouse
button [25] and then click the left or middle mouse button respectively. A
Philips Medical Systems
full description of the result pages is given in the section ‘ACA final
display’.
Step 12
Exiting the program.
1 To display the ACA menu, click the left mouse button [25] and then the
right mouse button [27] when the first or second result page is displayed.
Exit Program
Catheter Calibration
Measurement
CF = 0.107 mm/pix
SELECT MENU ITEM
Accept ---- ----
33721800
2 Select Exit Program and click the left mouse button [25].
N OT E The program can be exited at any time by pressing the active Menu key [95, 96 or 97]
on the Viewing Console.
Step 1
Redefining the obstruction site.
9896 001 33152
1 Hold down the middle mouse button [26] and drag the obstruction site with
Philips Medical Systems
the mouse. The obstruction and reference diameters, displayed in the top
left corner of the image, are updated. The reference value is determined from
the computed reference diameter function.
33721820
2 Click the right mouse button [27] to restore the automatically-calculated
(initial) obstruction site and return to ‘step 10’ of the ACA Standard
Procedure.
3 Or click the left mouse button [25] to confirm the new obstruction site.
Step 2
Redefining the reference site.
At this stage it is still possible to restore the previous automatically-calculated
(initial) diameters by clicking the right mouse button [27] and return to ‘step
10’ of the ACA Standard Procedure.
REFERENCE SITE
Accept Redefine Initial sites
33721830
Step 3
Changing the reference site.
The reference diameter displayed in the top left corner of the image will now
be measured between the arterial luminal contours. This is clearly different
from the situation described in ‘step 1’ above.
33721840
1 Hold down the middle mouse button [26] and drag with the mouse to
change the position of the reference site.
Step 4
Result pages.
9896 001 33152
Philips Medical Systems
33721860
RESULTS
RESULTS: Continue Second page Store Continue First page Store
On the first result page the reference diameter will now be indicated by a
broad line which is shaded to distinguish it from the obstruction diameter.
On the first and second result page the parameters that are directly affected
by the manual selection are marked with an asterisk (*).
• Click the middle mouse button [26] to select the first and second result
page.
• Click the right mouse button [27] to store the result page in the photofile.
• To redefine another obstruction and/or reference site (return to ‘step 10’ of
the ACA Standard Procedure) double click the left mouse button [25]. To
select a new vessel segment (return to ‘step 6’ of the ACA Standard
Procedure) click the left mouse button [25] followed by the middle button
[26].
• To return to the ACA menu, click the left mouse button [25] and then
click the right mouse button [27].
The result pages are described in detail in the section on ACA Final Display.
33721790
RESULTS
RESULTS: Continue Second Page Store Continue Second Page Store
This page shows the anatomy of the obstruction with the most relevant,
geometrically-derived parameters. The black rectangular box shows the
diameter function for the analyzed coronary segment. The diameter in
millimeters (vertical axis) is plotted against position along the centerline,
from start to end point (horizontal axis). The straight line through the
diameter function is the reference diameter function, defining the estimated
geometry of the vessel before disease occurred, as reconstructed from the
available diameter values.
The position of the minimum obstruction diameter and the proximal and
9896 001 33152
distal boundaries are indicated by vertical lines. These positions are also
shown on the vessel segment. The area between the reconstructed reference
contours and the luminal contours within the obstructed region is shaded;
this area is a measure of the area of the atherosclerotic plaque as seen in this
particular view. The percentage diameter (%D-Sten) and percentage area
Philips Medical Systems
33721870
RESULTS
Continue First Page Store
Top Left
The top-left quadrant shows a graph of the Stenosis Flow Reserve (SFR) for
this lesion.
33721880
Several points should be kept in mind when considering SFR values, SFR
reflects solely the effect of the measured arterial morphology on flow. To
make SFR specific for the measured lesion morphology, other parameters
affecting the coronary flow and coronary flow reserve, e.g. heart rate, aortic
pressure, additional coronary stenoses, are fixed for the determination of
SFR.
9896 001 33152
Coronary pressure (distal to the stenosis) is plotted along the Y axis and
coronary flow (relative to resting flow) along the X axis. The downward
curved line shows the projected fall in coronary pressure as flow through the
lesion is increased by vasodilation of the distal myocardial bed. Coronary
pressure continues to fall until the distal bed is dilated to its maximum
extent, a condition represented by the slanting straight line (the pressure-flow
relationship for the distal bed at maximum vasodilation). The intersection of
the curve and the straight line is the maximum relative flow through the
lesion, i.e. the SFR (Ref.5).
Bottom left
The bottom-left quadrant gives an overview of all geometrical parameters:
obstruction diameter and circular cross-sectional area, obstruction length,
reference diameter and circular cross-sectional area, percentage diameter and
area stenosis, plaque area and symmetry index (1.0 is completely symmetric
and 0.0 totally eccentric).
33721890
Top right
The top-right quadrant shows some of the underlying numerical data. The
total pressure drop across moderate to severe stenoses is derived from two
sources: frictional losses along the entrance and throat regions of the stenosis,
defined by the Poisseuille resistance; and losses caused by the sudden
expansion of the flow as it emerges from the stenosis exit, defined by the
turbulent resistance. Both resistances can be computed from the geometry of
the obstruction (Ref.5). The normal flow is calculated from the average
cross-section of the vessel and an assumed average flow velocity of 20 cm/s.
The table gives the pressure drop over the stenosis (mmHg) for the various
normalized flow values up to the stenotic flow reserve value. 9896 001 33152
Philips Medical Systems
33721900
Bottom right
The bottom-right quadrant shows the hospital name, the date and the
patient identification.
RESULTS
33721910
1 Select the correct image series to which the ACA Short program is to be
applied. The correct image series must be selected before proceeding.
2 Press one of the three menu keys to display the menu. The indicator above
Philips Medical Systems
PROGRAM MENU 1
PREVIOUS
NEXT
ACA Short
ACA
Vessel Diameter
Run Overview
33721920
3 If necessary, select Next/Previous to access more items.
4 Use the left mouse button [25] to select the ACA Short program. The ACA
menu is displayed.
Exit Program
Catheter Calibration
Measurement
CF = 0.108 mm/pix
SELECT MENU ITEM
Accept ---- ----
33721610
Step 1
Selecting an image.
1 Use the image selection keys on the viewing console to select an appropriate
image of the coronary arterial segment to be analyzed.
The following criteria are important for the image selection process:
• the arterial segment should be well filled with contrast medium
33721680
2 Click the left mouse button [25] to confirm image selection. A cursor
appears for selecting the coronary arterial segment to be analyzed.
3 If necessary, click the right mouse button [27] to return to the ACA menu.
Step 2
Selecting the coronary segment.
To guarantee reproducible assessment of the morphology of coronary
segments, such segments should be selected in a standardized manner,
according to the recommendations of the AHA (Ref.4), i.e. from side branch
to side branch. Therefore, it is preferable to position the start point
immediately distal to a branch point and the end point immediately
proximal to a branch point or at the end of the coronary artery. The start
point should always be proximal to the end point, i.e. the direction from start
point to end point should be the same as the direction of the coronary blood
flow. The reason for this constraint is that the computation of stenotic flow
reserve is dependent on direction.
9896 001 33152
Philips Medical Systems
1 Move the cursor so that its center dot is at the start point of the arterial
segment to be analyzed.
2 Click the left mouse button [25] to confirm the start point.
3 If necessary, click the right mouse button [27] to return to the ACA menu.
33721700
4 Move the cursor so that its center dot is at the end point of the arterial
segment to be analyzed.
5 Click the left mouse button [25] to confirm the end point and the selected
segment.
6 If necessary, click the right mouse button [27] to select another start point.
Step 3
Automatic detection of contours.
Once the start and end points of the coronary segment have been selected,
the message ‘Image filtering in progress’ is displayed and a path line between
the start and end points will be detected automatically (The purpose of the
image filtering procedure is to ensure reliable detection of a path line in the
coronary segment. If the image is the same as that used for calibration, image
filtering will not be performed). If no path line can be found, the message
‘Cannot find path line’ will be displayed and the program returns to the
beginning of ‘step 3’.
The following steps are performed automatically:
• the contour-finding process starts. The message ‘Initial Contour Detection
in Progress’ is displayed
• the image is zoomed.The twofold magnification (zoom) is not performed
if the segment length exceeds a quarter of the image size. Final contour
detection in the non-magnified image will have a detrimental effect on the
accuracy of the derived parameters
• the initially-detected contours are displayed along the catheter segment
and the message ‘Final Contour Detection in Progress’ appears.
9896 001 33152
Step 4
Result pages.
33721790
RESULTS
RESULTS: Continue Second Page Store Continue Second Page Store
• Click the middle mouse button [26] to select the first and second result
pages.
• Click the right mouse button [27] to store the result in the photofile.
• To select a new vessel segment (return to ‘step 2’), click the left mouse
button [25] and then click the middle mouse button [26].
The Result Pages are described in detail in the section on ACA Final Display.
Step 5
Exiting the program.
1 To display the ACA menu, click the left mouse button [25] and then the
right mouse button [27] when the first or second result page is displayed.
Exit Program
Catheter Calibration
Measurement
CF = 0.107 mm/pix
SELECT MENU ITEM
Accept ---- ----
33721800
2 Select ‘Exit program’ and click the left mouse button [25].
9896 001 33152
N OT E The program can be exited at any time by pressing the active Program Menu key on
the Viewing Console.
Step 1
Free Format menu.
This step follows on from ‘step 4’ of the ACA Standard Procedure (ACA
Catheter Calibration). A menu appears over the image offering the choice of
French sizes or Free Format (FF).
N OT E If a size has been entered or previously selected, it is displayed in a box at the bottom
of the menu. To select this value, point to it with the mouse and click the left mouse
button [25].
F4
F5
F6
F7
F8
F9
FF
2.7
33721930
1 Move the arrow to ‘FF’ (free format) and click left mouse button [25]. A box
menu is displayed for entering the size (diameter in millimeters) of the
calibration object.
Step 2
Entering the size.
.
0
1
2
3
4
5
6
8
9
0
OK
To select the size last entered or selected (displayed in the ‘Enter’ field):
1 Move the arrow to the bottom box (OK) and click the left mouse button to
confirm.
Philips Medical Systems
.
0
1
2
3
4
5
6
8
9
0
OK
33721950
To enter a new value:
1 Move the arrow to the first digit of the value to be entered and click the left
mouse button [25]. The selected digit is entered in the ‘Enter’ field, erasing
the previous free format size.
2 Move the arrow to the next digit to be entered and click the left mouse
button [25]. Proceed in this way until the required size has been entered.
A maximum of nine digits can be entered, with the decimal point in any
position.
3 If necessary, click the right mouse button [27] to remove the last digit
entered.
.
0
1
2
3
4
5
6
8
9
0
OK
4 After the required size (in millimeters) has been entered, move the arrow to
the bottom box (OK) and click the left mouse button [25] to confirm.
redisplayed together with the calibration factor. Since calibration has now
been performed, the arrow points to the Measurement box by default. The
arrow can be moved by moving the mouse.
Exit Program
Catheter Calibration
Measurement
CF = 0.108 mm/pix
SELECT MENU ITEM
Accept ---- ----
33721670
5 Use left mouse button [25] to confirm the choice.
5.6 Archiving
Overview of the (extended) archiving procedures:
Examination finished - -
Recording on a Video - -
Cassette Recorder
(VCR) (§ 5.6.6)
Philips Medical Systems
Report examples
The ‘Flag’ information field on the Status page indicates that either an image,
heartbeat or a complete run has been flagged, but does not indicate which
one of them has been flagged. To check whether it is an image, heartbeat or
run that has been flagged, use the ‘Run overview’ or ‘Examination overview’
key [40, 41] on the viewing console.
Conditions
Before transferring images, runs or examinations to an archiving system, they
have to be flagged.
If automatic run flagging is configured (Service), the whole examination will
be flagged. Images, runs or examinations that are not required to be
transferred need, therefore, to be unflagged.
All flagging keys [90 - 94] toggle between on and off.
2 If not highlighted, press the function key [15] corresponding to the ‘Select’
function.
The ‘Select’ function is then highlighted.
3 Use the cursor keys [20] to select the examination containing the images to
be flagged.
1 If configured, press the |Flag examination| key [93] on the viewing console.
The entire examination is flagged for transfer and the ‘Flag’ message appears
next to all runs, including the photofile, on the Status page.
2 To unflag the examination, press the |Flag examination| key [93] on the
viewing console.
The ‘Flag’ message disappears from all runs, including the photofile, on the
Status page.
1 Use the cursor keys [20] to select the run for flagging/unflagging.
2 Press |Enter| [21].
The run is selected and displayed on the viewing monitor(s).
3 To flag/unflag the run, press the |Flag run| key [91].
After flagging/unflagging, the ‘Flag’ message is displayed at, or cleared from,
the end of the selected run line.
4 To flag/unflag other runs, use the |Step to next/previous run| keys [32, 33]
and repeat procedural step 3.
1 Use the |Step to next/previous run| keys [32, 33] to select the run containing
the image to be flagged.
2 Use the |View reverse/forward| keys [31, 30] to select the image to be
flagged/unflagged.
3 To flag/unflag the image, press the |Flag image| key [90].
4 Repeat procedural steps 1 to 3 to flag/unflag other images.
1 Use the |Step to next/previous run| keys [32, 33] to select the run containing
the heartbeat to be flagged.
9896 001 33152
2 Use the |View reverse/forward| keys [31, 30] to select the heartbeat to be
flagged/unflagged.
3 To flag/unflag the heartbeat, press the |Flag heartbeat| key [92].
Philips Medical Systems
or EasyVison
R4.3
onwards
DICOM
DICOM
COMPLIANT
EQUIPMENT
36280170
The INTEGRIS Allura system allows transfer to an external (DICOM)
station using the High-Speed DICOM Image Interface.
An external (DICOM) station can be:
• Picture Archiving and Communication System (PACS)
• (DICOM) viewer
• EasyVision workstation
• (DICOM) image server
• Local archive (to be referred to as ‘Secondary archiving’)
• Other DICOM compliant equipment.
Since a local archive (such as IDR, ODR, HCU) has been identified as an
external (DICOM) station, the same explanation is valid for the local
archive.
Philips Medical Systems
If the system is configured for automatic run flagging (Service), all current
acquisition examination runs and the photofile are flagged during
acquisition. Instead of using the automatic ‘Flag run’ function [91],
individual images may be flagged using the ‘Flag image’ function [90] (if
configured). If subtracted images are flagged individually, the last selected
mask is also sent.
Conditions
this was not the case, the transfer process has to be repeated.
If the selected external (DICOM) station is not ready, then the system will
retry every two minutes.
Do not switch off the INTEGRIS system before the transfer process is
complete. The status of the transfer process can be checked by selecting the
Export page (using function keys F3, F4 or F5 [24]).
Do not switch off the external (DICOM) station before the transfer process
is complete.
36120010
Save Cancel Disk A/B Previous Next
2 If not highlighted, press the function key [15] corresponding to the ‘Save’
function.
3 Use the |Cursor up/down| keys [20] to select an examination for export.
4 Press the |Enter| key [21].
The selected examination is transferred to the external (DICOM) station, via
the intermediate buffer.
During the transfer process, the message ‘TRANSFER IN PROGRESS TO:’
followed by the station name and then ‘SYSTEM LOCKED’ is displayed on
the viewing/examination monitors, and the ‘Status’ field on the export page
will display ‘Busy’ or ‘Sending’.
2
After a few moments, the message ‘TRANSFER ABORTED’ is displayed on
all monitors.
During the cancel operation, the ‘Status’ field on the export page will display
Philips Medical Systems
Conditions
The Information system (IS) must be able to receive the data to be exported.
If the IS does not accept the ‘Report’ request the message ‘This function is
not supported by the information system’ is displayed on the error message
line of the Report page.
If the connection with the IS fails or another error situation occurs, the
message ‘No connection with the information system’ is displayed on the
error message line of the Report page.
As the data exported to the IS will indicate to which station name images
have been exported last, it is strongly recommended to export patient data to
the IS immediately after the images have been exported to the (long term)
image archive. This will maximize, but not guarantee, consistency between
the IS and the image archive.
During the transfer process, the message ‘TRANSFER IN PROGRESS TO:’
followed by the station name and then ‘SYSTEM LOCKED’ is displayed on
the viewing/examination monitors, and the ‘Status’ field on the Report page
9896 001 33152
will display ‘Busy’. Normal operation of the acquisition system is not possible
during transfer to the IS.
36120020
Report Cancel Disk A/B Previous Next
2 If not highlighted, press the function key [15] corresponding to the ‘Report’
function.
3 Use the |Cursor up/down| keys [20] on the keyboard to select the required
patient examination to be reported.
4 Press the |Enter| key [21].
The patient examination data is then exported to the IS.
Following successful completion of the transfer, the ‘Status’ field on the
Report page will display ‘Reported’.
36120060
Select Add Delete Modify Query More Previous Next
2 If not highlighted, press the function key [15] corresponding to the ‘Modify’
function.
The ‘Modify’ function is then highlighted.
3 Use the |Cursor up/down| keys [20] to select the required examination.
4 Press the |Enter| key [21] to confirm.
5 Move to the physician code (Phys) field by pressing the |Enter| key [21].
6 In the physician code field, use the |Cursor up/down| keys [20] to select the
required physician code, with corresponding external (DICOM) station,
from the list.
7 Press the |Enter| key [21] to confirm.
Each side of an optical disk has a capacity for approximately 150 images of
10242 pixels or 600 images of 5122 pixels. The average transfer time for 20
high-resolution (10242) images is less than 5 minutes. An optical disk written
on a release ‘n’ system can only be read by a system of release ‘n’ or higher.
Conditions
9896 001 33152
Write protection
Philips Medical Systems
33723830
The cartridge has a write-protect switch on each side. To protect one side of a
cartridge, set its switch to the protect position. To protect both sides of the
cartridge, set both switches to the protect position.
When handling ODR cartridges:
• do not remove the disk from the cartridge
• do not touch the disk surface through the optical access door
• use, handle and store in clean office conditions
• do not smoke while handling the optical disk. Smoke particles are a prime
source of disk contamination
• do not eat or drink while handling the optical disk. Liquids spilled on the
cartridge can seriously contaminate the disk surface which may then need
to be reconditioned.
New disks must be formatted before they can be used for image storage.
If no optical disk or a write-protected disk is inserted, the message ‘Please
insert an unprotected optical disk, press enter when done’ will be displayed
on the viewing console.
The system will check to ensure that all flagged items can fit on the optical
disk. If there is insufficient space, the system displays the message
‘Insufficient disk space, save examination on another disk’ on the viewing
console.
9896 001 33152
If the disk inserted was formatted on a system with an earlier release number,
the message ‘Incompatible disk, save examination on another disk’ is
displayed on the viewing console.
Philips Medical Systems
1 To insert a cartridge, slide it into the disk cartridge insertion slot [a] of the
ODR with the side to be read or written upwards.
LED indicator [216] flashes (green) when the drive is empty.
The cartridge will be partially ejected from the disk cartridge insertion slot
[a] and can be removed by hand.
N OT E S • Cartridges cannot be ejected during a save, retrieve or erase process.
• The cartridges are mechanically keyed so that they cannot be incorrectly inserted.
• An error message is displayed if an incompatible cartridge is inserted.
36120040
Save Cancel Retrieve Format Disk A/B Previous Next
2 If not already highlighted, press the ‘Format’ key [15] on the viewing console
to select the ‘Format’ function.
If the inserted disk contains examinations, the following prompt is displayed:
‘The optical disk contains examination items: continue format? Y/N’ If the
inserted disk contains data but no examinations, the following prompt is
displayed: ‘The optical disk contains data, but no examination items:
continue format? Y/N’
3 Type: N to terminate the function, or type: Y to continue the format process.
One side of the disk is then formatted.
3 Use the ‘Previous, Next’ and/or ‘Disk A/B’ functions to select the page with
the flagged examinations for transfer.
4 Use the cursor keys [20] to select the examination to be transferred to the
optical disk.
5 Confirm the selection by pressing |Enter| [21].
The system will start saving the flagged items of the selected examination to
the optical disk. If desired, more examinations can be scheduled for transfer
to the optical disk.
1 Press F3, F4 or F5 [24] to select the ODR page. Ensure that the correct
optical disk is loaded in the recorder
2 If not already highlighted, press the ‘Retrieve’ key [15] on the viewing
console to select the ‘Retrieve’ function.
If an acquisition examination is in progress the system responds with the
prompt ‘Exposures will be disabled, are you sure’. The reason for this is that
the system will close the current patient and open a new patient to store the
retrieved examination in. After retrieval is complete, a new patient is opened.
3 Type: Y to continue.
The system then checks to ensure that there is a valid optical disk in the disk
drive. If not, it displays the message ‘Please insert correct optical disk, press
enter when done’.
The system now checks the optical disk, reads information and displays the
prompt ‘Please select the examination to be retrieved’.
4 Select the examination to be retrieved with the cursor keys [20] and press
|Enter| [21].
Only one examination can be retrieved at a time. If more examinations need
to be retrieved, repeat from procedural step 2.
Retrieved examinations are stored in the last selected disk partition.
If the system does not have enough storage capacity to retrieve the
examination, it responds with the message ‘Not enough space to retrieve
examination’. Other examinations must be deleted to create space (‘Delete’
function on Review page).
Conditions
If a VCR is installed and configured in the APR for the selected imaging
technique, the recorder is automatically started and stopped by the system.
This can be configured for both fluoroscopy and/or exposures. Ensure that
the VCR is in “stand-by” mode before acquisition is started.
Real-time recording during fluoroscopy or acquisition:
Conditions
All report data is cleared after selection of a new patient.
The local printer (paper) has to be installed and ready for use.
Reports are printed with standard ASCII characters (no special characters are
available).
If an item is not available it is left blank.
The following restrictions apply to the miscellaneous image techniques:
• the generated data is different for 2-knob and 3-knob techniques
• the angulation, rotation and SID values are unknown
• the plane field (PL) is left blank on a monoplane system and for
monoplane image techniques.
9896 001 33152
1 Press [4c] on the acquisition console to print the configured (Service) report.
Philips Medical Systems
Report examples
33725800
Example of dose report
Cumulative total exposure dose Sum of fluoroscopy and exposure dose values.
Per run:
• Run number -
• exposure parameters
- kV Last value for each run.
- mA Last value for each run.
- mAs Last value for each run.
- ms Last value for each run.
Examination report
9896 001 33152
Philips Medical Systems
33725810
Cumulative fluoroscopy -
time
Per run:
• Run number –
• APR name APR name (sub-APR name for vascular purposes, maximum of 20
characters).
• Runspeed The maximum run speed is used for sequential blocks and VFR.
• Exposure parameters
- kV Last value for each run.
- mA Last value for each run.
- mAs Last value for each run.
- ms Last value for each run.
- Angulation/rotation For rotation angio, the last value for each run [degrees].
- Source image distance Last value for each run [cm].
- II format used [cm].
2 If necessary, press the |Add text| key [85] to switch text annotation off.
3 Use the |View forward/reverse| keys [30, 31] on the viewing console to select
the relevant image.
4 Press the |Store image in photofile| key [70] on the viewing console.
The image is stored in the photofile.
5 If deactivated, press the |Add text| key [85] to switch text annotation on.
1 If necessary, press the |Add text| key [85] to switch text annotation off.
2 Press the |Store image in photofile| key [70] on the viewing console.
The image is stored in the photofile.
3 If deactivated, press the |Add text| key [85] to switch text annotation on.
33724890
4 Use the Mouse or Joystick [108] to select the ‘Image transfer’ function and
press the |Accept| key [25 or 105].
The following menu appears on the viewing and examination monitors:
9896 001 33152
Philips Medical Systems
33724900
5 Use the Mouse or Joystick [108] to select the file name to which the image
will be transferred and press the |Accept| key [25 or 105].
For certain errors the system will reset automatically. Contact Service if
automatic resets occur frequently. The following tables show all error and
warning messages valid for the INTEGRIS systems:
Errors
Message Action
9896 001 33152
B: Bolus chase SPD controller not ready for use Call Service
Bolus chase speed controller not ready for use Call Service
Message Action
B: Table step and scan mv. not ready for use Call Service
B: TSO mod. and brake footsw. not ready for use Call Service
Message Action
Frontal stand not ready for use, set SID to 1 m Call Service
Lateral stand not ready for use, set SID to 1 m Call Service
Message Action
Table step and scan movement not ready for use Call Service
TSO mod. and brake footswitch not ready for use Call Service
Warning: bodyguard faulty or too close at start-up, risk. Increase distance from bodyguard
Philips Medical Systems
Message Action
Warnings
Message Action
Available disk time Warning 10 seconds before first exposure (current patient)
will be overwritten.
Messages
User message
• please insert an unprotected optical disk, press |Enter| [21] when done
• too many images in this examination to fit on optical disk, unflag items
• exposures will be disabled, are you sure
• please insert correct optical disk
• please select the examination to be retrieved
• not enough space to retrieve examination
• net error.
7.0 Introduction
To ensure satisfactory operation the Philips INTEGRIS Allura Monoplane
requires periodic planned maintenance and routine user checks. These are
essential to keep the equipment operating safely, effectively and reliably.
Planned maintenance may only be carried out by qualified and authorized
service technicians. In this context, qualified means those qualified to work
on this type of medical electrical equipment in the jurisdiction in which the
equipment is being used, and authorized means those authorized by the user
of the equipment. Philips is able to provide a full planned maintenance and
repair service on either a call basis or a contract basis. Contact your Philips
Service Organization.
PCBs and racks Ensure secure fitting and check for dust and Yearly
corrosion
Table cleanliness Hygiene and to ensure safe and optimum life Daily or as
of equipment needed
N OT E Visual and/or audible checks during routine use. Refer to Service documentation.
If applicable, the following procedure should be carried out for the lateral
channel as well.
1 Press (frontal) |Shutter adjustment| [151] to reset the shutters.
2 Position the tabletop horizontally and adjust it to maximum height.
3 Position the stand (in the frontal position) with the X-ray beam
perpendicular to the tabletop.
4 Position two lead rulers crosswise on the tabletop and use tape to attach the
rulers.
5 Use joystick (frontal) |SID adjustment| [134] to move the II as close as
possible to the rulers.
6 Press |II field size| [153] to select an appropriate II field size.
7 Initiate fluoroscopy by pressing footswitch [175] or ([178]).
8 Use the |Tabletop brakes on/off and longitudinal movements| knob [131] to
position the center of the intersection of the two lead rulers to the center of
the image.
9 Initiate single-shot acquisition by pressing footswitch [177] and write down
the ruler values (A1 to D1), corresponding to the edges of the image (see
illustration below).
9
10
C1
11
12
13
14
15
8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
16
17
18
A1 B1
19
20
21
22
23
D1
33728247
24
C2
10
11
12
13
14
15
9896 001 33152
7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
16
17
18
A2 B2
19
20
Philips Medical Systems
21
22
23
33728248
D2
24
13 Determine the distance [X] between the focal spot and the tabletop by
deducting the II to tabletop distance from the SID [124].
14 For each edge (A to D) calculate the following:
WA R N I N G To avoid electric shock, always isolate the equipment from the mains electrical supply
prior to cleaning, desinfecting or sterilizing it.
CAUTION Never allow water or other liquids to leak into the equipment as this may cause short-
circuits or corrosion.
7.3.1 Cleaning
Enameled parts and aluminum surfaces should only be wiped clean with a
damp cloth and mild detergent, and then rubbed down with a dry woolen
cloth. Never use corrosive cleaning agents, solvent or abrasive detergents or
polishes. If you are uncertain of the nature of a cleaning agent, do not use it.
Chrome parts should only be cleaned by rubbing down with a dry woolen
cloth. Do not use abrasive polishes. To preserve the finish, use non-abrasive
wax.
7.3.2 Disinfection
All parts of the equipment, including accessories and connecting cables, can
be disinfected by wiping them with a cloth dampened with disinfectant.
Never use corrosive or solvent disinfectants. If you are in any doubt about the
nature of a disinfecting agent, do not use it.
9896 001 33152
CAUTION Disinfecting a medical equipment room by means of sprays is not recommended since
the vapor can penetrate the equipment causing electrical short-circuits or corrosion.
Philips Medical Systems
operator must be satisfied that all traces of the vapor have dispersed before
switching the equipment on again. Disinfection techniques for both the
equipment and the room must comply with all applicable laws and
regulations which have the force of law within the jurisdiction in which the
equipment is located.
9896 001 33152
Philips Medical Systems
8.0 Introduction
Philips Medical Systems is concerned to help protect the natural
environment and to help ensure continued safe and effective use of the
INTEGRIS Allura Monoplane system through proper support, maintenance
and training. Philips equipment is therefore designed and manufactured to
comply with relevant guidelines for environmental protection. As long as the
equipment is properly operated and maintained it presents no risk to the
environment. However, the equipment may contain materials which could
be harmful to the environment if disposed of incorrectly. Use of such
materials is essential for the implementation of certain functions and for
meeting certain statutory and other requirements. This section of the
Manual is directed mainly at the user of the equipment or system, the body
with legal authority over the equipment. Operators are not usually involved
in disposal, except in the case of certain batteries (see paragraph 8.3 below).
Once the equipment has been passed on to a new user, the previous user may
still receive important safety-related information, such as bulletins and field
change orders. In many jurisdictions there is a clear duty on the previous user
to communicate such safety-related information to new users. Previous users
who are not able or prepared to do this should inform Philips Medical
Systems about the new user, so that Philips Medical Systems can provide the
new user with safety-related information.
WA R N I N G Do not dispose of the INTEGRIS Allura Monoplane system (or any parts of it) with
industrial or domestic waste. The system may contain materials such as lead, tungsten
or oil, or other hazardous substances that can cause serious environmental pollution. It
is advisable to contact your Philips Service Organization before disposing of the
INTEGRIS Allura Monoplane system.
replace the batteries, open the cover on the rear of the remote control,
remove the old batteries and place new batteries in the position indicated in
the battery compartment.
Battery type: Philips ‘PENLITE’ LR03.
Philips Medical Systems
CAUTION Always remove the batteries if the remote control will not be used for some time.
9.1 Specifications
9.1.2 Mains
• 380 V ± 10%, 50 and 60 Hz, 3 phase.
• 440 V ± 10%, 50 and 60 Hz, 3 phase.
• 480 V ± 10%, 50 and 60 Hz, 3 phase.
MRC-GS 200 03 10
• Grid-switch version for enhanced pulsed fluoroscopy image quality.
• Focal spot: 0.3/1.0.
• Power: 20/100 kW.
Philips Medical Systems
MRC-GS 200 04 07
• Grid-switch version for enhanced pulsed fluoroscopy image quality.
• Focal spot: 0.4/0.7.
• Power: 30/67 kW.
• Anode heat storage capacity: 2400 kHu.
• Continuous heat dissipation: 3500 W.
N OT E For each of the tables below, the column headed ‘Additional filter (Programmed)’ is
filled in by Service, if the default filter values are changed, during system installation.
Philips Medical Systems
Fluoroscopy modes
3 0
2 1, if MRC-GS
1, if MRM-GS
1 1, if MRC-GS
1, if MRM-GS
7 - 0
8 - 0
1 Cerebral 0
2 Neck 0
3 Cerebral fast 0
9896 001 33152
4 - 0
5 - 0
6 - 0
Philips Medical Systems
7 Lungs 0
8 Lungs (1024) 0
9 Lungs non-subtracted 0
10 Arch 0
11 - 0
12 - 0
13 Abdomen AP 0
14 Abdomen LAT 0
15 High speed 0
16 Kidney (1024) 0
17 - 0
18 - 0
19 Iliac 0
20 Upper Extr. 0
21 Lower Extr. 0
22 Foot/Hand 1
23 - 0
24 - 0
28 Rot. Scan 0
29 - 0
30 K.A.T 0
31 IV Neck 0
32 IV Arch 0
33 IV Abdomen 0
34 IV Iliac 0
35 IV Upper legs 0
36 IV Lower legs 0
Fluoroscopy APR’s
9896 001 33152
4 Emergency continuous 0 0
5 Low continuous 0 0
6 Normal continuous 0 0
7 Low pulsed 2 2
8 Normal pulsed 1 1
9 Quality pulsed 0 0
10 Low Vasc. 2 2
11 Normal Vasc. 2 2
12 High Vasc. 1 1
TV chain:
• CCD camera with proprietary digital output
• CCIR: 625/1249 lines for 50 Hz systems
• EIA: 525/1069 lines for 60 Hz systems.
9.1.11 XB monitor
• Mains voltage: 90 - 264 V.
• Mains frequency: 45 - 66 Hz.
• Maximum power consumption: 190 W.
• Weight:
9896 001 33152
- 24 kg (17-inch monitor)
- 33 kg (21-inch monitor).
• Size (width x height x depth):
- 17-inch monitor: 42.6 x 44.6 x 49.5 cm (16.7 x 17.6 x 19.5 inch)
Philips Medical Systems
Monitor suspension
Maximum load applied on a monitor position: 40 kg.
Rotation:
• from 90° LAO (+rotation) to 90° RAO (-rotation) (stand in its side
position perpendicular to the table)
• from 120° LAO (+rotation) to 185° RAO (-rotation) (stand in its head
position parallel to the table)
• variable rotation speed via tableside operation module up to 25°/sec.
Angulation:
• from 90° cranial (+angulation) to 90° caudal (-angulation) (max = 90°)
(stand in its head position parallel to the table)
• from 120° cranial (+angulation) to 185° caudal (-angulation) (stand in its
side position perpendicular to the table)
• variable angulation speed via tableside operation module up to 18°/sec.
Rotational angiography:
• from 90° LAO (+rotation) to 90° RAO (-rotation) (stand in its side
position perpendicular to the table (rolling movement)) at a speed of up to
30°/sec
• from 120° LAO (+rotation) to 185° RAO (-rotation) (stand in its head
position parallel to the table (propeller movement)) at a speed of up to
55°/sec.
L-arm
• longitudinal movement (parking movement) of 260 cm (102.3 inch)
• motorized longitudinal movement speed of 15 cm/sec (5.9 inch/sec)
• rotation movement over 180° (2 x 90°)
• motorized rotation speed of 12°/sec.
N OT E S • Under certain circumstances the tilt-center automatically coincides with the iso-
center of stand rotation and angulation.
• Under certain circumstances the stand angulation is synchronized with the tilt
9896 001 33152
movement.
Options
• Pivot for table base:
Philips Medical Systems
- Maximum pivot range from -90° to +90° with locked positions at 0°,
-13°/+13°, and -90°/+90°
- Increases table height by 2.5 cm (0.98 inch).
• Bolus chase.
- Illumena.
• MEDRAD
- Mark V ProVis.
N OT E Transfer speeds mentioned above depend on the local situation (network load and
external station).
12 34545 (')*%+( ",-./!
6
0
0 '
00
#
(*7/--.88&
200
0,5mm
180
1,0mm
160 1,5mm
140
'
120
100
80
60
40
Correct Contour : Accept Draw Erase
9
20
0 2 3
10 10 10
!" #$%&
24
N OT E As you can see in the illustration, the protective device lowers the dose by at least an
order of magnitude.
1 66 87 85 76
2 38 69 64 53
3 19 49 47 32
90 0 64 33 34 46
1 39 27 24 31
2 19 20 15 17
70 0 35 6.0 13 22
1 18 4.1 7.2 12
N OT E The dose values are approximately the same for continuous fluoroscopy as for pulsed
fluoroscopy.
10.0 Introduction
This Section provides the user with additional background and supporting
information.
20kVp.
Exposure time
The exposure time must be measured with an oscilloscope connected to
measuring points EH116:X25 (KVAV) and EH116:X29 (GND). 1 V equals
20 kVp. The exposure time is the time for which the measured kV remains
equal to or greater than 75% of its maximum value.
mAs
Measured with a digital mAs meter connected to measuring points
EHX1002 and EHX1003. Indicated values, meter readings and control
settings are accurate to within the following limits:
mAs (radiography) 0.5 - 999 mAs ± 5 % or ± 0.5 mAs, if texp 10 ms and I x t 5 mAs
Philips Medical Systems
kV (continuous 40 - 110 kV ± 5 % or ± 1 kV
fluoroscopy)
Another disadvantage is that for each step the sudden acceleration and
deceleration of the table movement is very uncomfortable for the patient.
Philips Medical Systems has decided not to implement the stepping method.
Philips Medical Systems
The interactive mode gives the physician full control over the table
movement, injection protocol, frame rates and run duration.
With a hand held speed controller the table scan speed can be matched to the
speed of the contrast bolus, displayed in real time on the monitor screen.
The proper speed is achieved if the front of the bolus remains located at the
bottom of the monitor screen at the successive images in the run.
If DSA is required, a second (mask) run must be made without a contrast
injection. During the second run the images are acquired at exactly the same
locations as in the first run, by automatically repeating the recorded table
speed profile and exposure frame rate as used in the first run. With run
subtract the DSA images can be visualized.
The automatic mode applies an automatic table speed profile so that the user
only needs to give the start command.
A range of pre-programmed table speed profiles are available for selection.
Proper selection requires some advance information about the speed of the
blood in the particular patient. Therefore an integral part of the whole
automatic bolus chase procedure is to determine the so called Knee Arrival
Time (KAT) by means of a small test injection above the bifurcation. A very
common KAT value is 6 – 7 seconds, but for various pathological reasons
values between 4 and 20 seconds may be observed.
It is good angiographic practice to adapt the injection rate of the contrast
medium to the expected basal flow rate of the blood at the site of injection.
In this way the best balance is obtained between the chance of reflux and
local dilution at the injection site. So, each KAT program that can be selected
by the user has its own speed profile and optimised injection protocol advice.
In addition, the frame rate is automatically adjusted in such a way that with
the given speed profile the exposures are made at nearly equidistant positions.
Another integral part of the bolus chase procedure is to adapt to various
patients sizes by user selections of start and end positions of the table
movement.
The applied table speed profile in the automatic mode is the result of an
internal automatic optimisation calculation after the user selection of the
KAT value and the begin and end positions.
The optimisation is based on the following assumptions:
• patient positioning is as advised (see section 5.3.3 ‘Bolus chase’)
• catheter tip location is as advised (see section 5.3.3 ‘Bolus chase’)
• injection duration is as advised during KAT selection.
9896 001 33152
N OT E Deviations from these advised settings may lead to less optimal results.
stand returns automatically to its starting position ready for the second (dye)
run. During acquisition, the images are displayed in non-subtracted mode.
Subtraction can be carried out during viewing. The actual stand position
(‘ROT’ and ‘ANG’ angles) for each image can be displayed on the viewing
monitor when viewing a steady image (i.e. not during a cycle). The operator
can use this information to position the stand in order to have the same view
of an area of particular interest (see also ‘Display physiological data’ [87]).
For more information see Section 5.3.4 ‘Rotational Angiography’.
Recommended use:
To improve image quality it is strongly recommended to create an average
mask from no more than 4 images (10242 matrix) or 16 images (5122
matrix).
10.2.4 Ultrasound
Linear array and curved array technology
footprint and a sectorial image with a wide, far field. The large number of
transducer elements makes it possible to focus at four different depths, with
three focal points. The probe automatically selects the optimal focus for the
image depth.
Image resolution
Each picture element is defined by two parameters: the focus setting and the
number of echo-lines per image (frame rate). Optimal image resolution is
achieved by defining each picture element within the image with respect to
the chosen clinical application.
Focus
The focus of an ultrasound image is defined with respect to axial, lateral and
transverse resolution.
Axial resolution
Axial resolution is the minimum separation of two objects along the beam
axis (Z axis) for which two separate echoes can be identified. The axial
resolution is determined by the probe frequency: the higher the frequency of
the probe, the better the axial resolution over a greater depth range. However,
the maximum penetration decreases as the frequency increases.
Lateral resolution
Lateral resolution is the minimum separation of two objects placed at the
same depth (X axis) for which two separate echoes can be identified. The
major determinant of lateral resolving ability is beam width. Beam width
varies with distance from the probe, and is dependent on probe geometry, its
frequency, and focusing characteristics. Increasing the number of focal points
also increases the lateral resolution. As the number of focal points increase,
the frame rate decreases.
Slice thickness
The slice thickness resolution is not directly visible in the image, but it is
substantially related to the spatial resolution. The spatial resolution
corresponds to the smallest distinguishable separation between two reflecting
surfaces. With a linear or curved array probe, fixed focusing is possible only
in the Y direction (thickness of the slice).The form and material of the
acoustic lens control the focus in this direction.
Frame rate
A picture element is also defined by the number of received echo lines per
image (frame rate). The more echo lines received, the better the image that
can be calculated by the Scanner. The frame rate is displayed at the left side
of the screen.
Probes
9896 001 33152
33723870
a c/d b
Probes available for scanner
Two probes can be connected to the scanner at one time. Since the Scanner
can handle single-frequency as well as dual-frequency probes, the operator
may choose from four different frequencies (two dual-frequency probes
connected).
Linear 7.5 • • • • •
array
Curved 5.0/7.5 • • - • •
array
Curved 3.5/5.0 • • - - -
array
HD curved 3.5 • • - - -
array
Abdominal
For abdominal ultrasound examinations, the depth of the object being
visualized determines which probe is used.
Paediatric
Use the 5.0 MHz curved array [b or c] for abdominal scanning. The 3.5
MHz curved array [c] or 3.5 MHz high-definition curved array [d] may also
be used. The 7.5 MHz linear array [a] or 7.5 MHz curved array [b] gives the
best results when examining superficial objects.
Object depth:
Depth Probe freq. Probe type
(cm) (MHz)
Intra-operative
The 7.5 MHz linear array [a] may be used the for intra-operative
visualization of structures in the 0 - 6 cm range. For this technique, the probe
must be covered with a sterile sheath.
Small objects
Use the 7.5 MHz linear array [a] or 7.5 MHz curved array [b] for
visualization of structures in the 0 - 6 cm range.
Peripheral vessels
Use the 7.5 MHz linear array [a] or 7.5 MHz curved array [b] for
visualization of structures in the 0 - 6 cm range.
Needle guides
There are two types of needle guide attachments:
Type I: Needle guide attachment for the 5.0/7.5 MHz and 3.5/5.0 MHz
curved array probes and for the 3.5 MHz high definition curved array probe
(probes [b, c and d]).
Type II: Needle guide attachment for the 7.5 MHz linear array probe (probe
[a]).
33723880
g e f h
Type I needle guide attachment Type II needle guide attachment
WA R N I N G The needle guides are not designed for use with neurological/surgical patients.
Philips Medical Systems
33725440
Type B ‘applied part’ symbol
max
155
max
155
155
max
155
33722850
10.3.2 Automatic/manual wedge control (Cardiac APR)
The position of the wedge filter is determined according to a curve related to
the stand rotation (LAO (+rotation)/RAO (-rotation) projection). The
default position of the wedge filter depends on the LAO (+rotation)/RAO (-
rotation) projection. If both the RAO (-rotation) and LAO (+rotation)
projections are less than 20°, the default position of the wedge filter is +20%
clockwise over 40% of the radius (these values are valid for a 23 cm (9 inch)
Image Intensifier field and an SID of 1 meter (39.4 inch)) (see illustration).
9896 001 33152
100%
60%
+ + +
40% 40%
Philips Medical Systems
33722820
LAO/RAO ±20° 20° < LAO 120° 20° < RAO 120°
N OT E Under conditions where the wedge filter would not be seen, e.g. if either the SID is
greater than 1 meter (39.4 inch) and/or the Image Intensifier field is smaller than 23 cm
(9 inch), the filter is automatically set at ± 45° clockwise over >10% of the radius.
The automatically set wedge filter positions can be manually adjusted. The
figures below show the effects of moving [155] forward, backward, and to the
right respectively; moving the control to the left has the opposite effect to
moving it to the right.
+ +
155
156
+ +
155
156
+ +
max
or
155
156
+ +
33722830
Hysteresis
The algorithm incorporates 10° of hysteresis to ensure filter stability. This
means that when the X-ray tube is rotated within 10°, the position of the
wedge filter remains constant.
9896 001 33152
Manual mode
In manual mode, the wedge filter no longer follows the movement of the
stand. Pressing [155] down enters manual mode, the stand position may be
Philips Medical Systems
changed, but the wedge filter position does not change. Pressing [155] down
a second time returns the system to automatic mode and the wedge filter
moves to the position fixed by the stand position.
155
33723700
156
[155] controls the wedge filter for standard (monoplane) cardiac procedures.
The control allows translation and rotation of the semitransparent wedge
filter. Pressing [155] sets the wedge filter to manual mode. In this mode
[155] controls the wedge filter:
Joystick up: wedge rotates clockwise
Joystick down: wedge rotates counterclockwise
Joystick left: wedge moves out or inwards (depending on wedge filter
position (see illustrations above))
Joystick right: wedge moves in or outwards (depending on wedge filter
position (see illustrations above))
Pressing [155] again, resets the wedge filter to automatic mode (standard
(monoplane) cardiac procedures). Pressing [156] sets the wedge filter to
vascular mode. In the vascular mode [155] controls the left wedge filter and
[156] controls the right wedge filter (see [155, 156]). Pressing [156] again,
resets the wedge filter to automatic mode (standard (monoplane) cardiac
procedures).
N OT E Press [156] to switch from vascular to cardiac mode (automatic/manual mode).
9896 001 33152
Philips Medical Systems
36280590
Rotation and angulation angles are indicated on the display unit [123].
The system is provided with several safety devices to help the operator avoid
collisions and carry out the appropriate movements.
N OT E The safety features apply to motorized movements only.
Bodyguards
A bodyguard is not a switch device but a device that senses distance and
controls the maximum permitted speed of the movement. All motorized
movements of the C-arm and table are controlled by the bodyguard system.
The system will protect the patient by slowing down movement speeds when
an object is detected within a certain safety distance. The detection system
does not prevent all collisions, but due to the reduced movement speeds these
collisions will not be harmful if they occur.
N OT E If the bodyguard sensor is covered with a cloth and the cloth becomes wet, it will be
detected as an object and block motorized movement. It is advisable to use plastic
covers rather than a cloth as fluid is not absorbed by plastic and cannot therefore
block the motorized movement.
33726750
g1 g2 g3
9896 001 33152
d. Table base.
e. Ceiling carriage.
f. Pivot point.
+90°
b c d
33726760
-90°
With the stand in the neuro/cardiac position (position [g2]) the L-arm can
be pivoted around the table, when the table is in its fully extended position
and in its mid-lateral position, without collision.
CAUTION Check that there is no risk of collision with the tabletop before pivoting the L-arm.
36280600
+90°
33726780
b c d
-90°
36280360
+90° position, 0° position and -90° position
Move [145] up/down (see note) to rotate the L-arm to one the three
predefined working positions. Pushing [145] up, rotates the L-arm away
from the operator. Pulling [145] down, rotates the L-arm towards the
operator.
26280380
[145] up/down Rotation movement of the L-arm
36280390
[145] down, L-arm in the -90° position [145] up, L-arm in the +90° position
Move [145] right/left (see note) to move the L-arm from parking position to
the automatic stop at the neuro/cardiac position. Move [145] further to the
right/left (see note) to move the L-arm to the automatic stop at the (lower)
peripheral position and vice versa.
N OT E Movements depend on the operator’s position (doctor/nurse side of table). The
illustration below shows the movement when the operator is at the doctor’s side of the
table.
9896 001 33152
Philips Medical Systems
36280400
36280410
[145] left, L-arm in park position [145] right, L-arm in peripheral position
C-arm movements
b
c
36280130
C-arm bodyguards
The C-arm is mounted in a saddle attached to the L-arm in which it can slide
(angulation) up to 90° in both directions.
The saddle is pivoted on the L-arm, allowing a ‘propeller movement’ of 120°
clockwise and 185° counterclockwise as seen from the table. Both C-arm
movements are performed under motorized control.
9896 001 33152
WA R N I N G Take care to avoid collisions when moving the C-arm with the motor drive.
Motorized movements are controlled by the joysticks [134, 136 and 145] or
by the APC function (see illustrations). Automatic power limitation helps to
ensure safe operation but care must be taken to avoid collisions of any kind.
Provided that the orientation switch on the tableside control module is
correctly set, the functions of the C-arm joysticks are always logically
oriented with respect to the patient, irrespective of the position of the L-arm.
Vascular application
C-arm rotation
36280420
[136] up/down C-arm rotation
36280430
[136] up, +90° (max.) [136] down, -90° (max.)
C-arm angulation
36280450
[136] left, -55° (max.) (see note) [136] right, +60° (max.) (see note)
Cardiac application
C-arm rotation
36280460
[136] up/down C-arm rotation
36280470
9896 001 33152
[136] up, 120° LAO (max.) [136] down, 185° RAO (max.)
C-arm angulation
Philips Medical Systems
36280480
[136] left/right C-arm angulation
36280620
[136] left, +50° cranial (max.) (see note) [136] right, +50° caudal (max.) (see note)
Combined movements
36280490
36280510
Basic position 2 [136] left/up,[136] down, combined rotation
combined rotation
and angulation
WA R N I N G In the event of a clinical emergency involving a patient, do not press emergency power
off [130] or [160] or switch off the X-ray system using a local mains power-off switch.
Move the image intensifier away (either motor driven or manually, by firmly pushing it
up) and/or the tabletop, and/or rotate or park the stand to provide clear, all-round
access to the patient.
Motorized II movements
36280520
36280530
[134] up, SID = 119.5 cm (47 inch) [134] down, SID = 89.5 cm (35.2 inch)
Motorized movement
36290220
[133] up/down Table height adjustment
33726970
[133] up, tabletop at 104 cm [133] down, tabletop at 76 cm
(40.9 inch)(max.) (29.9 inch)(min.)
N OT E In case the tabletop is moved up or down and moves into the area of a bodyguard
sensor, either on the X-ray tube housing or image intensifier, the table movement is
reduced and finally stopped before a collision would occur.
9896 001 33152
Philips Medical Systems
Pivot
33723500
AD5 mounted on the pivot
36280300
-13° pivot position, -90° pivot position and +90° pivot position
33723520
AD5 with SyncraTilt option.
If the tilt option is installed, three extra functions are available on the
geometry module:
• motorized |Longitudinal movement| [131]
9896 001 33152
The tilt angle is indicated on the display unit [124] and on the scale (-28° to
Philips Medical Systems
Motorized movement
Table tilt
36290210
[139] left/right Table tilt adjustment at the isocenter of rotation
and angulatuion
36280540
[139] left, -28° HDT (max.) with [139] right, +20° HUT (max.) with
ceiling suspended C-arm in park position ceiling suspended C-arm in park position
Synchronized tilt
36280310
[139] left, synchronized tilt 0° to -28° [139] right, synchronized tilt 0° to +20°
HDT (max) (see note) HUT (max) (see note)
9896 001 33152
SyncraTilt accessories
Special accessories including a footrest and a set of shoulder straps are
Philips Medical Systems
c f
d
e
33723540
a. Footrest.
b. Ankle straps.
c. Handgrips - rail version (2 pieces).
d. Handgrips - tabletop version (2 pieces).
e. Shoulder supports (not with neuro tabletop).
f. Chin support.
Bolus chase
A dynamic digital acquisition technique that allows the flow of contrast
medium to be precisely followed through the lower peripherals in a
continuous movement with just one contrast injection. Bolus chase is
possible in two modes: automatic and interactive, both subtracted and non-
subtracted.
Bolus chase supports a variable table scan movement (70 cm (27.6 inch) up
9896 001 33152
to and including 100 cm (39.4 inch)), which is selected by the operator and
is related to the length of the patient’s legs, and a variable number of
exposures (9 up to and including 39) for patient dose reduction.
Philips Medical Systems
Bodyguards
A bodyguard is not a switch device but a device that senses distance and
controls the maximum permitted speed of the movement. All motorized
movements of the C-arm and table are controlled by the bodyguard system.
The system will protect the patient by slowing down movement speeds when
an object is detected within a certain safety distance. The detection system
does not prevent all collisions, but due to the reduced movement speeds these
collisions will not be harmful if they occur.
N OT E If the bodyguard sensor is covered with a cloth and the cloth becomes wet, it will be
detected as an object and block motorized movement. It is advisable to use plastic
covers rather than a cloth as fluid is not absorbed by plastic and cannot therefore
block the motorized movement.
Disk partition
The image storage capacity of the system can be divided by Service (during
software installation) into disk partitions, giving each partition a part of the
available image storage capacity. So a disk partition can be considered as a
logical disk having the same speed characteristics belonging to the number of
disks in the subsystem.
A maximum of two partitions can be defined. The minimum size of one
partition is limited to 30% of the entire storage capacity.
The partition in which the images of an examination are stored is determined
by the partition code related to a physician code in the examination data.
0 3
33720020
1
1 Doctor side.
2 Nurse side.
3 Foot-end.
The doctor side is the right side of the table, which is defined as the right side
of the patient. The patient lies on his/her back on the table with feet pointing
towards the bottom of the table. With this patient orientation, the left side of
the table is the so-called nurse side and the foot end is the side towards which
the patient’s feet are pointing.
Dynamic reference
Each acquisition examination has up to two Dynamic reference files.
Heartbeats that form part of an exposure run are stored in the Dynamic
reference file when the ‘Store reference’ function is used. Storage and recall of
9896 001 33152
Dynamic reference images follow the same principle as for Static reference
images.
Philips Medical Systems
Examination
An examination consists of a number of examination attributes such as
patient name, examination ID and examination date, entered by the operator
in an examination schedule, a number of exposure runs, one photofile and
zero or one run report. When an examination has been scheduled for
acquisition, images can be acquired and will be collected in a file for that
examination.
Examination report
The examination report function prints dose or examination information
concerning the acquisition examination on a printer connected to the
acquisition console. The function provides either a dose or an examination
report (configurable during installation). Since printing is a background
process, the next acquisition examination can be selected as soon as the print
process has been started.
Geometric terms
Angulation (ANG) Angulation (ANG) Movement around the axis perpendicular to the
patient at an angle to the horizontal plane equal to the rotation angle. So
when the rotation is zero, angulation is a movement around the lateral
axis. Angulation clockwise around the lateral axis is denoted as positive
or plus.
Anterior/Posterior When the patient is in supine position: AP is when the X-ray tube is
(AP), Posterior/ above the patient, PA is when the X-ray tube is under the patient.
Anterior (PA)
Caudal (CAUD) The image intensifier is angulated in the direction of the patient’s feet with
patient in head-to-stand position (i.e. positive angulation).
Cranial (CRAN) The image intensifier is angulated in the direction of the patient’s head
(i.e. negative angulation).
Isocenter Point in space around which both the rotation and angulation movements
take place.
Left Anterior X-ray beam perpendicular to the longitudinal axis of the patient in supine
Oblique (LAO) head-to-stand position with the image intensifier on the left side of the
patient (i.e. positive rotation).
Rotation (ROT) Movement around the axis parallel to the patient, i.e. around the
longitudinal axis. Clockwise rotation is denoted as positive or plus.
9896 001 33152
Right Anterior X-ray beam perpendicular to the longitudinal axis of the patient in supine
Oblique (RAO) head-to-stand position with the image intensifier on the right side of the
patient (i.e. negative rotation).
Philips Medical Systems
Source-Image Distance between the X-ray tube focus and the input screen of the
Distance (SID) imaging device.
Heartbeat
A heartbeat is defined as a sequence of consecutive images of an exposure run
starting at an Rpeak and ending at the next Rpeak of the recorded ECG signal
from that exposure run. If the ECG signal is not recorded, the boundaries of
the heartbeat are -0.5s and +0.5s from the current image. For biplane runs,
the heartbeat is determined from the frontal channel.
Image
The maximum number of images that can be stored in the system is
dependent on the matrix resolution and the number of disks installed. A
typical image attribute is the ‘ flagged for transfer’ attribute. The maximum
number of flagged images per run is 20.
Imaging technique
Selectable imaging techniques: Digital Dynamic, Digital Vascular, etc. The
imaging techniques available in a system depend on the installed options.
Joystick directions
up right
left
down
36280180
Patient orientation
Four orientations are possible:
• ’Normal’: Patient on his/her back with head at tabletop head
• ‘Legs-up’: Patient on his/her back with feet at tabletop head
• ‘Nose-down’: Patient on his/her stomach with head at tabletop head
• ‘Legs-up’ and ‘Nose-down’: Patient on his/her stomach with feet at
tabletop head.
Peripheral run
During a peripheral run with Bolus Chasing the patient (i.e. the tabletop)
9896 001 33152
Photofile
Each examination has one photofile. A photofile can have between 1 and a
maximum of 999 photo images in it (depending on the photofile length
configured). The photofile is always regarded as the first run of an
examination.
For the acquisition examination the maximum photofile space will be equal
to the remaining free photo image space of the disk partition the acquisition
examination resides in. At the moment another scheduled patient record is
selected as acquisition examination then the remaining free space of the
photofile of the old acquisition examination will be made equal (if possible)
to the configured photofile length (e.g. 20 photo images).
During review, the user is still able to store images in the photofile until it is
full, then the warning ‘photofile full’ will be given to the user.
The free photofile space of an examination retrieved from ODR is equal (if
possible) to the configured photofile length (e.g. 20 photo images).
The photo images in the photofile will always be flagged for transfer to a
HCU. If ‘Auto run flag for photofile’ is configured ‘ON’ then the photo
images in the photofile will be flagged for transfer to any other connected
device as well (e.g. ODR, Generic archive).
Photo images
Photo images have a fixed matrix size of 1280 x 1024 8 bit pixels (50 Hz).
Only on 60 Hz CCD based imaging systems it will be 1280 x 960 8 bit
pixels.
Photo images are always processed images with burned in graphics, there is
no further image processing possible (e.g. contrast, brightness).
The following functions can be applied to photofile images:
• add text
• annotation
• vessel diameter
• calibration.
Reference files
An acquisition examination can have one or two reference files. A reference
file can have between 1 and 999 reference images in it (depending on the
reference file length configured).
For the acquisition examination the maximum reference file space will be
equal to the remaining photo image space of the disk partition the
acquisition examination resides in.
Reference images
Reference images have a fixed matrix size of 1280 x 1024 8 bit pixels. Only in
60Hz CCD based imaging systems it will be 1280 x 960 8 bit pixels.
Reference images are always processed images with burned in graphics. There
9896 001 33152
are often also that interesting that they have to be archived as well. To achieve
this, reference images are kept in the photofile as well. To avoid unnecessary
copying of images, the actual images will only be saved once. This means that
a reference file is actually a different view of the images in a photo file.
Traditionally in cardio and vascular systems the use of a photofile and
reference files differs. These differences between cardio and vascular systems
were mainly caused by historical reasons. Reference images in vascular
systems were always put in a photofile and were never overwritten. The
length of the reference file and photofile was exactly the same.
Reference (formerly called roadmap) images in cardio systems however were
never put in the photofile, the reference file was a circular image buffer of
limited configurable length (typically 5 images).
Examples 1 and 2 show a possible vascular application.
• The photofile and reference file length were both configured to 4 images.
• In the first example the Reference-1 file is filled with 3 images, which are
stored in the photofile (P1, P3, P6) as well. After e.g. using the Store-
Photo function, P4 in photofile was filled, which led to an automatic
extend of the photofile with another four free places (P5 - P8).
• In the second example the photofile has been extended for a second and
third time.
The length of the reference file is exactly the same as the number of reference
images in it.
Grey boxes indicate occupied
Plain boxes indicate free
P1 P2 P3 P4 P5 P6 P7 P8
C R1-1 R2-1 R1-2 R2-2 R1-3
1 2 3
33725820
Grey boxes indicate occupied
Plain boxes indicate free
33725830
A. Reference-1 File
B. Reference-2 File
C. Photofile
* Configured Photofile and Reference File size (4)
1. First extend (4)
2. Second extend (4)
3. Third extend (4)
Rotational angiography
Images are acquired during rapid rotation of the C-arm around the patient.
9896 001 33152
Exposure run
Philips Medical Systems
For image processing parameters different sets will be maintained, one for
image processing during acquisition, one for image processing during review
(this set can be modified by the user), both for subtracted and non-
subtracted display. For biplane runs different sets are maintained for the
frontal and lateral channel.
The maximum number of exposure runs for the system is 999. The
maximum number of heart beats, flagged for transfer, per run is 5.
When the available free space falls below 5000 there are two possibilities:
• The indicator counts down from 5000 in steps equal to the number of
exposures in the last acquired run and shows the remaining free space for
exposures.
• The indicator counts down from 5000 in steps of 1000 after each acquired
run or single shot exposure. A maximum of 5 runs can be acquired (a
single shot exposure counts as a run).
Viewing examination
In order to review an examination (other than the acquisition examination),
the user has to select an examination from the ’done list’ for viewing. The
’done list’ is a list of all finished examinations including the current
acquisition examination. This selected examination will be defined as the
viewing examination.
There can only be one viewing examination at any time. All review functions
performed with the viewing console (if present) operate on the viewing
9896 001 33152
examination.
Unless a viewing examination other than the acquisition examination is
explicitly selected, all reviewing functions performed with the viewing
console will be related to the acquisition examination.
Philips Medical Systems
Working area
The area around the isocenter of the stand, where digital acquisition is
possible.
11.2 Abbreviations
General
HR Heart Rate
Legend
80 kv 900 mA 170 ms 00 min
2
Acquisition Console 1
2 Gycm2 3
A Power on and reset
B Power off
1 Display of generator related parameters 4a 4b
2 Reset fluoroscopy time buzzer
4
3 X-ray control 4c
4c
4 Display and selection keys for acquisition related parameters
4 a/b Select imaging technique or APR (main/sub)
4 c Acquisition function keys A 10
33728241
5
• New patient • Variable frame rate (VFR) 6 9
7
• Subtraction • Print dose/examination report
• One- or two knob acquisition
5 Legs up Acquisition console
6 Nose down (patient prone)
7 Display information
8 Dial wheel
9 Enter
10 Alter APR parameters
18
Viewing Console
C
C Digital imaging system power on
D Digital imaging system power off
15 Function keys D
16
Intercom
17
16 Talk 15
17 Listen
33727850
18 Volume control
20 Cursor keys
21 Enter
22 Delete
23 Tab
24 Function keys
• F1 Schedule Page • F6 System Page
• F2 Review Page • F7 Physician Page
20
• F3 - F5 • F8 Reserved
P h i l i p s M e d i c a l S y s t e m s 9896 001 33152
24
- Report page • F9 Status Page
- Export page • F10 Help Page
- Optical disk (ODR) page
- Copy page 22
- Transfer page 23
21
33727860
Keyboard
Mouse Photofile/reference/
25 Accept 70 Store image in photofile
26
26 Action 71 Delete image from photofile
27 Reject 72 Reserved 25
27
73 Reserved
Viewing keys 74 Store reference 1
Selection 75 Store reference 2
30 View forward 76 Reserved
31 View reverse 77 Reserved
32 Step to next run 78 Recall reference 1
33 Step to previous run 79 Recall reference 2
33727870
34 Reserved 80 Reserved
35 Reserved 81 Reserved
36 Cycle through run Mouse
33728256
52 Edge enhancement decrease 98 Analytical functions
53 Edge enhancement level indicator 71 61 33 31 30 32 62 54 48 45 46 47 58
54 Select video invert Tableside Analysis Module (TAM) Viewing keys
55 Pan & zoom 100 Automated Vessel Analysis (AVA)
56 Center zoom 101 Pixel shift (as [64])
57 Position shutters 102 TAM Menu (as [95-97])
58 Override manual adjustments (OMA) 103 Pan & zoom (as [55]) 100 101 102 103 104
65 Run subtract
66 Landmarking
67 View trace
68 CO2 trace
33728200
105 106 107
Viewpad a b
a Subtraction on/off (as [60])
c d
b New mask (as [62])
e
c View reverse (as [31])
d View forward (as [30]) f g
g Center zoom l m
h Contrast decrease (as [46])
i Contrast increase (as [45])
j Store reference 1 (as [74])
k Store reference 2 (as [75])
33723580
l View reference 1 (as [78]
m View reference 2 (as [79]) Viewpad
36270050
111 112 114
XB monitor
P h i l i p s M e d i c a l S y s t e m s 9896 001 33152
36270060
111 112 114
Display unit
120 Intercom ‘listen’ indicator
121 X-radiation ON indicator
33723640
122 Tube load indicator
123 Actual rotation/angulation angles
124 Source-image distance (SID) / Tilt angle indicator 122 126 127
36280190
133 144 139 134 145
141 Reserved
142 Store position 1
Monoplane geo vascular module (Tilt option)
143 Store position 2
144 Recall position 1 or 2
145 Motorized movement of the frontal stand
150a 151 157
Imaging Module
150 Fluoroscopy mode selection:
a HIGH
b NORMAL
150b
c LOW
151 Shutter adjustment 150c 153
152 Reserved
153 II field size
154 Reserved
155 Position left wedge filter
33723670
156 Position right wedge filter
157 Variable frame rate (VFR)/Dual/Parallel fluoroscopy 158 155 156
158 Trace-subtract fluoroscopy (TSF)
Vascular monoplane imaging module
Auto Position Control (APC) module
160 Emergency power off
161 APC mode selection
a Scratch mode 162 163 160
b Sequencer mode
c Reference mode
162 Store position 1
161a
163 Store position 2
P h i l i p s M e d i c a l S y s t e m s 9896 001 33152
165
33727960
APC module
172
TSD speed control
170 Reserved
171 Reserved
172 TSD speed control 179 177
33728243
176 Reserved
177 Single-shot exposure/parallel viewing
178 Reserved
TSD speed control Acquisition foot switch
179 Exposure
185
33723050
Acquisition hand switch
190
P h i l i p s M e d i c a l S y s t e m s 9896 001 33152
191
36280200
Handgrips for L-arm stand
33723820
215 216 219 a 218
216 Rewrite mode indicator
217 WORM mode indicator Optical disk recorder (ODR)
218 Eject optical disk
219 Manual eject screw
225-233 Reserved
239 Freeze/unfreeze
240 Biopsy line on/off
241 Reverse on/off
242 Depth increase 243 236
243 Depth decrease
237
244 Near gain control
245 Far gain control 244 239
246 Total gain control
247 Joystick
33728020
245 246 247 238
I-2 I n de x I N T E G R I S A l l u r a M on o p la n e Release 1
S TAM functions 4-49
Safety 2-1 Text & annotation 4-45
Electrical safety 2-3 Viewpad functions 4-50
Emergency procedures 2-2 Viewpad
Explosion safety 2-4 Battery replacement 4-31
Fire safety 2-4 W
Important safety directions 2-1 Wedge filters 10-10
Mechanical safety 2-3 X
Mobile telephones etc 2-4 XB Monitors
Screen saver 4-22 Reference setting 5-49
Second geometry module 4-5 User reference setting 5-49
Second TSO imaging module 4-9 XB monitors 4-6
Slager wall motion 5-158 Preparation 5-49
SpectraBeam 4-8 X-ray ON indicator lamp 4-6
Stand movements 10-15
Standard line rate video output 4-17
Status page 4-86
Storage media 4-69
Optical Disk Recorder (ODR) 4-69
Switches 4-67
Acquisition foot switch 4-67
Acquisition hand switch 4-68
SyncraTrak (APC) 4-5
SyncraVision 4-9
System page 4-84
T
Table movements 10-24
Tableside Analysis Module (TAM) 4-17
Tabletop Shifting Device 4-66
Speed control hand switch 4-67
Technical data 9-1
Trace Subtract Fluoroscopy 4-64
Transfer page 4-84
Type B symbol 10-8
U
Ultrasound image 4-93
Display 4-93
V
9896 001 33152
Viewpad 4-16
Viewing console 4-26
System and error messages 6-5
Text display 4-74
Viewing functions 4-35
General 4-48
Photofile/reference 4-44
Processing 4-39
Selection 4-35
Selection for transfer 4-46
Subtraction 4-41
I-4 I n de x I N T E G R I S A l l u r a M on o p la n e Release 1
9896 001 33152