• To prevent accidents or laboratory • For your safety, the unit will power off • A separate on/off switch for heating
fires, all magnetic stirrers feature two in case of a short-circuit, a damage to prevents unintentional heat-up.
independent safety circuits which the temperature sensor, a motor failure If heating is switched on, the on/off
switch off heating in case of any or other incidences during operation button is illuminated for visual control
overtemperature situation
• To protect you from splashes and scal- • Even if the stirrer is exposed to the
• In combination with Heat-On attach- ding from bath media, rotation speed highest temperatures, damage is ca-
ments you can convert your lab into a ramps up slowly until unit has reached tegorically ruled out – all models come
“safety zone” the pre-programmed speed with a fire-resistant aluminum die-cast
housing
• Hotplate residual heat indicator in the
digital display of the MR Hei-End is
illuminated when unit is turned off and
hotplate temperature is above 50 °C to
warn operator of potential burn hazards
( 3
• All units with digital display allow for • The unique Kera-Disk ®
coating is
easy setting and monitoring values on extremely chemical-resistant and
the illuminated readout scratch-proof and can be cleaned
easily if solvents, oil or other
• You can even work with dry ice – contaminations are spilled on the plate
a robust thermal insulation prevents
corrosion inside the housing by • The chemically resistant Kera-Disk ®
hot
preventing condensate from collecting plate allows for immediate heat
and dripping on any electronic transfer resulting in quick heat-up
components times
* 5
Sealed housing
PC board
gas
motor
In case of a short-circuit, a A separate on/off button for heating
damage or removal of the Damage to the stirrer is categorically prevents unintentional heat-up – the
temperature sensor from the ruled out even if exposed to highest button is illuminated for visual control
media, the unit powers off temperatures – all models come with
completely a fire-resistant aluminum die-cast
housing
The hermetically-sealed housing that protects the PC motor but is drained off. The other line marked in orange significantly increases the durability of the magnetic
board and the motor is marked in black. The way a liquid shows that gas can neither ingress to the PC board nor stirrer. For protection against liquid and gas, the heating
would take flowing down from the hotplate is marked in to the motor. The apparent gap is due to temperature elements are completely embedded in ceramic.
orange. As one can see, even if the liquid reaches the insulation. Because of that gap the temperature inside the
heat shields, it cannot ingress to the PC board and the magnetic stirrer stays substantially lower, which
, 7
Preferred use of the MR Series magnetic stirrers includes smooth to intense mixing and Model for standard applications without display
heating of low-viscosity fluids. Your first choice for decomposing organic and inorganic The MR Hei-Standard includes all leading safety standards and features for superior ease
substances of use and reduced cost of ownership, plus:
• The hotplate reaches a maximum temperature of 300 °C • The strong magnetic field allows for stirring even higher • For your protection an independent safety circuit will
viscosity media or volumes of 20 liters of water with switch off heating if hotplate temperature exceeds
• Safety circuits avoid an overheat situation of your ease 25 °C over set temperature
hotplate: if the temperature overshoots the heating
will be powered off immediately • Speed is adjustable from 30 respectively 100 to 1,400 rpm • Upgrade this magnetic stirrer with the optional
at an accuracy of up to ±1 % for gentle mixing temperature controller with sensor that provides
precise control of temperature to give you
temperature overshoot protection and reproducible
results
MR Hei-Tec
MR Hei-Tec
Model for higher requirements and reproducible results
MR Hei-Standard MR Hei-End The MR Hei-Tec includes all leading safety standards and features for superior ease
of use and reduced cost of ownership, plus:
• This unit features a digital display and • A separate on/off switch for heating • Upgrade this magnetic stirrer with
The StarFish system can convert your The Heat-On attachments are the allows for easy setting and monitoring of prevents an unintentional heat-up. If the optional Pt 1000 temperature
values in the illuminated readout heating is switched on the on/off button sensor for precise temperature control,
magnetic stirrer over to a time-saving industry´s safest, fastest and most is illuminated for visual control overshoot protection and reproducible
multiple-reaction workstation time-efficient method to heat and • For your protection an independent safety results
stir a sample in a round-bottom flask circuit will switch off heating if hotplate • An illuminated button indicates clearly if
temperature exceeds 25 °C over set the stirring mode is activated
temperature
MR Hei-Tec
StarFish system Heat-On system P/N 036110550
. 9
• An independent safety circuit switches off heating at • A separate on/off switch for heating prevents an unintentional heat-up.
an operator-programmed temperature value above If heating is switched on the on/off button is illuminated for visual control
your set temperature
• Upgrade this magnetic stirrer with the optional Pt 1000 temperature
• An additional operational safety step is required sensor for precise temperature control, overshoot protection and
to change parameters which prevents unwanted reproducible results
changes to the actual setting
MR Hei-End
P/N 036110580
• Utilize the optional software to control and document the • Take advantage of a built-in PID controller that
entire process ensures fastest heat-up times while preventing an
overtemperature situation of your sample MR Hei-End with Pt 1000
• MR Hei-End features a digital RS 232 and analog P/N 036110592
interface for connecting the optional software easily
temperature
800 W
600 W
MR Hei-Tec with Pt 1000
P/N 036110582
time
'& 11
Resolution (°C) 1
Stirring bars Evaluation Kit
015882570 Accuracy of measurement (°C) ±1
pack of 10 pcs.
Measuring cycle (sec) 1.5
temperature
Temperature sensor EKT Hei-Con G Pt 1000, glass
Set-temperature
Support rod Silicone protective cover Stirring bars - oval shape (25x10 mm) Sensor length (mm) 210
036111270
For temperature sensor Replaceable cover against pack of 3 pcs. for 25-ml to 50-ml flasks Fast-Mode Cable length (mm) 700
length: 425 mm, Ø 13 mm splashes and dripping water
Fine-Mode Min. immersion depth (mm) 20
material: stainless steel Stirring bars - oval shape (30x10 mm)
036111280
P/N 036300880 For MR Hei-Mix D, pack of 3 pcs. for 100-ml to 250-ml flasks Ambient temperature range (°C) -5 to +60
MR Hei-Mix L, MR Hei-Tec Housing Polyamide
Support rod for bath and MR Hei-End Stirring bars - oval shape (40 mm) Weight (kg) 0.3
attachments from 3 to P/N 036306040 015882560 time
pack of 5 pcs. for 500-ml to 1,000-ml flasks
Dimension l x w x h (mm) 54 x 22 x 126
5 liter For MR Hei-Standard
P/N 036306030 P/N 036306050 Protection class (DIN EN 60529) IP 40
Stirring bars - oval shape (50 mm)
015891060 Power supply (V) 8 – 15
pack of 5 pcs. for 2,000-ml flasks
'( 13
The Heat-On attachments are suitable to replace oil baths and heating mantles in your lab Heat-On blocks Multi-Well holder and inserts
and reduce the risk of fire hazards. Moreover the attachments minimize the messy oil clear-
Each Heat-On block is a stand-alone product that can be This unique Multi-Well holder is designed to hold either
ups that result from changing oil or removing flasks from an oil bath! placed directly onto the stirring hotplate one or two inserts for flasks*.
The inserts are available for 10-ml, 25-ml, 50-ml, 100-ml and
Heat-On 250-ml block 015882060 150-ml flasks. Flask inserts also feature cut-away sides for
Heat-On 250-ml block 015882580 use with two or three-neck flasks and accept the optional
• Our attachments are by far the safest, fastest with flask sidearm cutouts safety lifting handles
and most efficient method for heating and mixing Heat-On 500-ml block 015882070
solutions in round-bottom flasks from 10 ml to 5 liter Heat-On 1-l block 015882080 * Only accepts one 150-ml insert
Heat-On 2-l block 015882090
• Heat-On fits precisely to every flask, thus ensuring Heat-On 3-l block 015882100
maximum surface contact and fast heat transfer. Heat-On 4-l block 015882105
Eliminates the mess from cleaning flasks after Heat-On 5-l block 015882300
removal from an oil bath or hazardous spills when an
oil bath slips off a stirring plate. As a result, oil as a
heat transfer media is made redundant and your lab is
a safer place with the Heat-On block system
250 ml 150 ml 300 °C 10.8 min Support rod for bath attachments 036306030 Heat-On 10-ml insert 015882010
from 3 to 5 liter Heat-On 25-ml insert 015882020
» For best results we recommend round-bottom 500 ml 300 ml 300 °C 16.4 min
Retort clamp 015882540 Heat-On 50-ml insert 015882030
flasks that comply with DIN EN ISO 4797 and
1,000 ml 600 ml 300 °C 21.1 min
Boss head SilentCrusher M / Heat-On 015882550 Heat-On 100-ml insert 015882040
2,000 ml 1,200 ml 300 °C 35.1 min Heat-On 150-ml insert 015882050
DIN EN ISO 24450
3,000 ml 1,800 ml 300 °C 47.3 min Flask Stand & Clamp Kit 015882510
(including Support rod for bath attachments
4,000 ml 2,400 ml 300 °C 51.0 min
from 3 to 5 liter, Retort clamp, Boss head
5,000 ml 3,000 ml 300 °C 75.5 min
SilentCrusher M / Heat-On)
'* 15
Base plate
Base plate with optional handles fits securely onto the top plate maintaining good
Universal 5-way clamps
thermal contact with the heated surface
The StarFish clamp allows you to hold glassware of virtually any size
and comes with a choice of Silicone rubber or Velcro support straps
StarFish base plate (145 mm) Each clamp features 5 telescopic arms which can be extended and locked in
P/N 015890020 place to suit your needs. Using different straps on separate clamps allows
you to grip the flask neck with the Silicone rubber strap for lifting, whilst the
StarFish base plate handles (pair) 5-way clamp 5-way clamp condenser above slides through the Velcro strap
P/N 015890060 (Velcro) (Silicone strap and handle)
P/N 015890310 P/N 015892140
', 17
Hei-Mix D / Hei-Mix L
• Top plate is insulated from the motor • You can even work with dry ice - a robust
and protects your heat-sensitive samples thermal insulation prevents corrosion MR Hei-Mix D
from any thermal damage inside the housing which minimizes P/N 036110500
dripping of condensate on the electronic
• To treat your sample gently and protect components MR Hei-Mix L (without display)
it from damage the rotation speed ramps P/N 036110480
up slowly until it has reached the pre- • These models come with a top plate
programmed speed material of finest-quality stainless steel
MR Plug & Play P/N 036110588 • 5-way clamp (silicone strap & handle)
• 2 x Water Manifold
Hei-Tec • 1 x Gas/Vacuum Manifold
Hei-Mix S
MR Plug & Play • 650-mm split rod
This package includes: P/N 015859600
• Magnetic Stirrer MR Hei-Tec Multi-Well Hei-Standard
• Stir quickly and efficiently with • Ideal for titrations due to white- • Temperature sensor Pt 1000 (V4A)
an extended speed range up to colored top plate • Support rod for Pt 1000 This package includes:
2,200 rpm • Clamp for Pt 1000 • Magnetic Stirrer MR Hei-Standard
• Long-lasting polyamide housing P/N 036110573 • EKT Hei-Con temperature controller
• Space-saving unit for efficiently and PVDF top plate material with sensor (stainless steel)
mixing sample sizes up to 5 liter. • Support rod for EKT Hei-Con
Small footprint of L 140 / W 126 / • Clamp for EKT Hei-Con
H 80 mm only, and the top plate • Heat-On Multi-Well system including
comes with a diameter of Multi-Well holder and the following
MR Hei-Mix S 104 mm inserts: 2 x 25 ml, 2 x 50 ml, 2 x 100 ml
P/N 036110460 • 2 x Retort clamp, 2 x Boss head
P/N 015859620
'. 19
MR MR MR MR MR MR
Model
Hei-Mix S Hei-Mix L Hei-Mix D Hei-Standard Hei-Tec Hei-End
Speed range (rpm) 0 - 2,200 100 - 1,400 100 - 1,400 100 - 1,400 100 - 1,400 30 - 1,400
Cer ti f i cat e
Speed accuracy (%) ±5 ±2 ±2 ±2 ±2 ±1
Sensor breakage protection – – – with EKT Hei-Con with Pt 1000 with Pt 1000
To confirm the ability for
Temperature control – – – Micro controller Micro controller Micro controller
25 °C over
±5
25 °C over
±5
10 - 25 °C
continuous operation
Safety circuit hotplate (°C) – – – hotplate hotplate over nominal
temperature temperature temperature of the MR Series Magnetic Stirrers with hotplate
Stirring capacity, max. (H2O) (l) 5 20 20 20 20 20
S c h w a b a c h , J a n u a r y 2 0 1 3
20 21