USER’S MANUAL
Foreword
This manual is made for the IR Communicator 3 equipped with the 1.0 or higher software
version. We recommend, that the manual is stored near to the IR Communicator 3.
On guarantee and service matters, contact the DETECTOR Oy service division. We are
at Your service on all questions about our equipment.
DETECTOR OY
Telekatu 8
FI-20360 Turku
FINLAND
www.detector.fi
detector@detector.fi
IR Communicator 3 User’s manual Code no. 2071 1112
Rev.1.0
Page 3/13
CONTENT
1. BEGIN .......................................................................................................................................... 4
1.1. Keypad ................................................................................................................................ 4
1.2. Read DGT ........................................................................................................................... 5
1.3. Change the view .................................................................................................................. 6
1.4. Parameters explained .......................................................................................................... 7
1.5. Change DGT parameters .................................................................................................... 8
2. CALIBRATION ............................................................................................................................. 9
2.1. Calibrate ZERO ................................................................................................................... 9
2.1. Calibrate SPAN ................................................................................................................. 10
2.2. Calibrate current output 0 mA - 4 mA - 20 mA ................................................................... 11
3. TERMINAL ..................................................................................................................................... 12
Code no. 2071 1112 IR Communicator 3 User’s manual
Rev. 1.0
Page 4/13
1. BEGIN
IR Communicator 3 is maintenance device for DGT series of gas detectors. Only
personnel authorized by Detector Oy is allowed to carry out the maintenance. Incorrect
maintenance may cause irreversible damage to the gas detector!
1.1. Keypad
ON power on / off
ALPHA
LOCK
input method letters / numbers
1.
DETECTOR OY
IR-COMMUN ICATOR 3
S/N 12345 678901 2
V. 1.00
ON
WELCOME !
PRESS ENT ER FOR
REVERSE D ISPLAY
PRESS OTH ER KEY
TO CONTIN UE
2.
3.
30.01.13 12.00
DGTeex V 1.00
S/N 12345 67
OXYGEN
READ VALUE 020 .97 %
OUTPUT 17 .420 m A
SCALE LO 000.00
SCALE HI 025.00
LIMIT: 01 9.00
DELAY: 00 030. s
LOCOUNTS 20609.
HICOUNTS 25237.
CALIBRATI ON GAS :
021.00 %
---MORE- --
Code no. 2071 1112 IR Communicator 3 User’s manual
Rev. 1.0
Page 6/13
30.01.13 12.00
DGTeex V 1.00
S/N 12345 67 ---MORE---
OXYGEN CALIBR. 3 0.01.1 3
VALUE 020 .97 % SENSOR. 3 0.01.1 3
OUTPUT 17 .420 m A MUX SETTI NG: -
SCALE LO 000.00 PROGRAM D GT
SCALE HI 025.00 ---BACK- --
LIMIT: 01 9.00 ---BACK---
DELAY: 00 030. s
LOCOUNTS 20609.
HICOUNTS 25237.
CALIBRATI ON GAS :
021.00 %
---MORE- --
30.01.13 12.00
DGTt2 V 4.01
S/N 12345 67
FREONS
VALUE 000 00. PP M
OUTPUT 04 .000 m A
SCALE LO 000.00
SCALE HI 025.00
LIMIT: -- ---
DELAY: -- ---. s
Sensor A parameter A-PARAM. 04650.
R0 01690. Sensor R0 parameter
CALIBRATI ON GAS :
01000. PPM
---MORE- --
Code no. 2071 1112 IR Communicator 3 User’s manual
Rev. 1.0
Page 8/13
ALPHA
Tip Change input method letters / numbers
LOCK
3. ENTER
5.
2. CALIBRATION
1.
2. ● 0 ppm / %
● 1 l/min
or ambient
ENTER “BEEP!” = OK
● no pressure
= gas flows in
and out
● ≥ 1 min
● stable
reading
Code no. 2071 1112 IR Communicator 3 User’s manual
Rev. 1.0
Page 10/13
1. Set calibration gas concentration at main view, see 1.5. Change DGT parameters
2.
3. ● Use ONLY
test gas
recommended
by Detector Oy ENTER “BEEP!” = OK
● 1 l/min
● no pressure =
gas flows in and
out
● ≥ 1 min
● stable reading
IR Communicator 3 User’s manual Code no. 2071 1112
Rev.1.0
Page 11/13
1. 4 mA CALI BRATIO N
2. Measure the output current. See installation manual for test point location.
decrease increase
4.
ENTER “BEEP!” = OK
Code no. 2071 1112 IR Communicator 3 User’s manual
Rev. 1.0
Page 12/13
1.
20mA CALI BRATIO N
2. Measure the output current. See installation manual for test point location.
decrease increase
4.
ENTER “BEEP!” = OK
IR Communicator 3 User’s manual Code no. 2071 1112
Rev.1.0
Page 13/13
3. TERMINAL
Certain special parameters can be accessed through terminal only. Detector Oy will
give instructions when needed.