Professional Documents
Culture Documents
Erbe ICC-200 - Instruction Manual
Erbe ICC-200 - Instruction Manual
MAITEINANCE MANUAL
CONTENT
1. FUSE PROTECTOR.......................................3
2. MAINTENANCE.........................................4
You are highly appreciated for purchasing our PLX102 Mobile X-ray Equipment. In order to
guarantee that you can use this machine correctly and would make full use of it, please
carefully read below maintenance manual before you repair and maintain this equipment.
1. Fuse protector
FU3 BLX 1A 1
FU4 BLX 1A 1
left rear of the controller, and FU3, FU4 are assembled on the right front. Between the two
groups, there is a division plate and FU3, FU4 are assembled on the plate with the stickers of
their codes.
Attention: Make sure the power is off before replacing the fuse protector!
If group FU3 (1 A), FU4 (1 A) are needed to change, we only need unscrew 2 fixed screws on
the front plate of the controller and then put down the plate and you will see BLX plug socket
on the slight right front. Please choose right fuse socket on basis of the stickers FU3 or FU4
according to the fault and open screw cover, and then start to replace the fuse.
If group FU1(20 A), FU2 (20 A) are needed to change, we should not only unscrew 2 fixed
screws on the front plate of the controller but also unscrew 4 fixed screws on the four corners
of the plate and 2 fixed screws in the controller, and then remove the controller part and you
will see RL1-15 plug socket on the left rear of the controller. Please choose right fuse socket
on basis of the stickers FU1 or FU2 according to the fault and open screw cover, and then
When replacement is completed, please tighten the screw cover and screws and assemble
the controller part as well as the front plate. Check all components are assembled correctly
without the extra screws and now this equipment can be turn on.
2. Maintenance
2. With long time use, the degree of the mechanical security will be declined. In order to
prevent the harm to patients or the operators, we should always check whether the
managing files.
4. Operator should be well trained and be able to complete the routine examination.
5. In order to prevent the fatigue reaction of the spring, the X-ray source assembly should be
Daily inspection ● Check whether the knobs, switches and remote control unit work
normally or not.
● Check whether the X-ray source assembly appears oil leakage.
Weekly inspection ● Check whether there is not common noise when X-ray source
assembly is exposing.
4
● Check the resistance when the equipment is connected with ground
inspection ● Check whether all the screws of mechanical parts are loose.
● Check the whole X-ray source assembly and the beam limiting device.
machine with the cotton fabric soaked in detergent or disinfectant, otherwise, it will accelerate
1. When needed removing the beam limiting device for some reasons, you should unscrew
the plug from the plug socket of the picture 14, and then screw the locknut of picture 1
counterclockwise to make it released from the X-ray tube assembly. Should you instal the
Figure 1
5
3.2 How to change the lamp in the beam limiting device
After power on the beam limiting device, even adjust the button, there is still no light
from it, the possibility is that the lamp is not in normal work. Please see the figure 4, take the
plug of beam limiting device from the socket, measure the resistance Rx1 with multimeter.
Test the foot 1 and foot 2 of the beam limiting device socket. If they are in normal state,
please find other reasons; if not, take off the 2-M3 screw from the back cover board of
Figure2. After taking off the back cover board, take off the 2-M3 screw, and the lens hood, and
Turn off the 2-M3 screw from the magnetic base, take off the lamp. If the lamp is broken,
please change a new one. And then spin tight the 2-M3 screw. (Please note that don’t spin to
hardly in cose of making it broken. After finishing spinning, try to pull it out lightly, it’s ok if you
can not pull it out). Spin the plug of beam limiting device to the beam limiting device socket (in
figure 4), and then press time-limiting switch (see figure 3). The lamp should be lighting for
35s and then turn off automatically. Install other spare partes, please notice to install all of
Figure 2
6
Figure 3
● The resistance of power supply is ● Check whether the power supply satisfies
7
● Choose “local control” from “Local
● Choose “Remote control” from “Local
Remote control/Remote control”;
control/Remote control”;
control of no ● Far distance, or prevent by metal;
● Change position of operation people;
use ● Remote control switch is our of
● Change A23 battery
power
Press “Power ● No power suply in socket; ● Check the power supply socket;
on”, no power ● Power supply wire is off; ● Check and repair the power supply wire;
supply ● Power supply fuse is broken ● Check, and change FU1, FU2 and FU3