Professional Documents
Culture Documents
Don't Read It.
90
Though the method is accurate it has some limitations such as weighing of large
flask reducing the sensitivity of the balance, requirement of high degree of
accuracy in maintaining the constant temperature, for each measurement whole
assembly has to be set-up making the procedure laborious and time consuming.
One of the most accurate methods of density measurement of solutions is the
magnetic float method designed by Lamb and Lee304 and then modified
considerably by others305,306 to obtain the results accurate to 0.1 ppm in the density
measurements. In this method, a glass float with a small bar magnet is dipped in a
91
cell containing the solution whose density is to be measured, with the help of
platinum weights (Figure 2.2). A solenoid is wound around die cell through which
the current can be passed. Any change in the density of the solution alters the
position of the sinker float in the solution. By passing the current through the
solenoid one can bring the float back to
the original position. The changes in
density can be calculated by measuring Solution
container
the current required to bring the float
Magnetic
back to the original position. This — ~~Float
and convenient for the density measurements of liquids. In this method the
measurement of natural vibrational frequency of the oscillating tube containing
liquid under investigation is made. The natural vibrational frequency of the tube is
related to the density of the liquid inside it by
d = A + Bx2
where d is the density of the liquid, A and B are the instrumental constants and x is
the oscillation period of the tube.
2.1.3 Densitometer: Instrumentation
Figure 2.4 shows the general block diagram of the ANTON PAAR digital
densitometer describing the oscillator excitation, amplitude control, and time lapse
measuring and data reception circuitry which was described by Leopold in
detail.311 The motion of transducer, attached to vibrating U-tube (figure 2.5),
induces an alternating voltage in the sensor, proportional to its time dependant
velocity. The signal generated is amplified and used in two circuits.
Figure 2.4. Block diagram of ANTON PAAR digital density meter (DMA 602/60)
forced harmonic oscillator is maintained and phase shifts are made negligible by
setting a large bandwidth (3 - 10000 Hz) for the excitation and amplitude control
loop relative to the resonance frequency of the tube when filled with liquid
covering the density range 0 - 3000 kg m‘3. However, this chosen bandwidth
results in an uncertainty of 10 ps in
detecting the zero voltage crossing
of the velocity proportional
amplified sensor voltage. Thus, the
sensitivity of the electronic
detection circuitry defines the small
time unit for the time-lapse
measurements. The amplified signal
sensor is concurrently impressed on
the Schmidt trigger. A start
command, which may be effected
by manual depression of a push Figure 2.5. Sketch of a vibrating tube
button automatically following a densitometer
Out of these two units the first one contains a ‘U’ tube for holding sample.
The ‘U’ tube is completely enclosed in a glass envelope, which contains high
thermal conductivity gas. A glass jacket through which a Thermostated liquid is
circulated to reduce the temperature equilibration time surrounds the envelope. In
this model forced harmonic oscillator of fixed amplitude is obtained by the use of
an electro-optical sensor circuit. A phototransistor is irradiated by IR light emitted
by light emitting diode. A small vane, coated with magnetic material, is attached to
the apex of the vibrating ‘LF tube. In its motion the vane chops the light beam,
thereby modulating the phototransistor output. This signal, when amplified and
limited, is fed back to control the magnetization of the ferrite material, which in
turn controls the motion of the oscillator.
As soon as the instrument is switched on the sample tube starts to oscillate
and the instrument is ready for use. If at the time of filling of the sample tube
illumination is turned on, the photoelectric part of the oscillating system will not
operate and the density measurement is suspended. After the illumination is turned
off the instrument is ready for measurement.
A syringe holder is used to hold the syringe while it is being deployed for
filling the sample tube. While loading the vibrating tube with liquid sample, the
precaution is taken that the introduction of the liquid must be made slowly enough
to enable the sample liquid to properly wet the walls of the sample tube. The
meniscus of the liquid while filling the sample tube must be concave and not
convex, to avoid the trapping of micro air bubbles so that the period readings
obtained are stable. The sample tube is completely filled when the liquid meniscus
passes the upper enlarged portion of the sample tube. The opening of the upper part
of the sample tube is to be closed off with Teflon stopper and the illumination is to
be turned off. Once the illumination of the sample is turned off, the tube starts
oscillating and the reading x (tau) will be displayed on the display unit after the
interval of selected time period (k).
The temperature equilibrium can be conveniently followed over several
read out cycles until the x value becomes constant (± 1 x 10"6), indicating complete
temperature equilibrium. At this time, value of x is to be recorded.
96
dm = A + Bx2
dyivta A + Bx
108.766 x 10'5 = A + B (1.286304)2
0.997047 = A + B (1.683578)2
A = -1.3956007
B = 0.84414357
By using the values in the table 2.1 the values of A and B (instrumental
constants) were calculated at 25° C. These A and B constants are temperature
dependent. These constants were used for the calculations of the densities of the
solutions.
For each density measurement of a solution being studied at a given
temperature, the A and B constants were determined separately by calibrating the
instrument with air and water, i.e., for every measurement of density of solution
the x values for air and water were recorded. As a check of authenticity of the
density measurements, the measurements for known systems were carried out and
compared with the best literature data312 reported. The graphs of -103 (do - d)
97
against concentration for aqueous alkali halide solutions at 25° C are reproduced in
figure 2.7 along with literature data.312
TABLE 2.1 Values of A and B constants at 25° C and 5k time scale and
calculations of densities.
Sample x at 5k t Mean Density at 25° C
/g cm'3
Air 1.286304
1.286305
1.286304 1.286304 108.7664 x 10'5
1.286304
1.286304
Water 1.683578
1.683578
1.683577 1.683578 0.997047
1.683579
1.683578
45
40 • This work
o Literature
35
•? 30
o
(90
25
o
V
3 20
x
15
10
0
0.5 1 1.5
m /mol kg'1
(a)
30
• This work
25 o Literature
V°
O
tuo
-§ 15
3
x
o
10
(b)
50
45 • This work
o Literature
40
U>
LA
x(d~do)/gcnT
u>
o
to
LA
fN
©
10
LA
LA
bs
©
ts
be
O
©
o
*“
ere.
(C)
99
vO
o
• This work
o Literature
oin
10 x (d-do) /g cm '
m
O
o
rs
o
•°
©
(N
o
00
■
o
d
<r
I
on
O
"J
N
00
oo o
, UI3 8/ (Op-p) X
ON
on o ©
01
o O
U)
on
©
O
©
(N
©
o
o '
rr
o,
oo.
3
(5
Figure 2.7. Comparison of densities of alkali halides determined in this work with
the literature312 (a - NaCl, b - KC1, c - NaBr, d - KBr, e - CsBr).
100
2.1.6 Accuracy
Accuracy of the instrument is mostly dependent on the temperature control
since the A and B constants are temperature dependent. The densitometer is
capable of measuring the densities with the accuracy ±1 * 10'3 kg m‘3 provided the
temperature of the sample is maintained to the accuracy about ± 0.001° C or better.
The temperature accuracy of the present work is about ± 0.02° C. The accuracy of
density measurements in the present work was estimated to be of the order of ± 5 x
10'3kgm'3.
Electrical Connector
Syringe Ports
Cell Cover
Teflon Seal
Syringe Guides
Teflon Support
Thermistors
Support Rods
Glass Beaker
Head Thermostat
Monitor Window
Cell Cover
Front Panel
Mirror
Thermistors
Class Beaker
Figure 2.11. Side view with basic design of K-7000 vapor pressure osmometer
104
At the time of each measurement, at least 5-6 readings were taken for the fixed
time settings and the averaged value was used for further processing. Since
KNAUER K-7000 vapor pressure osmometer is a commercial instrument and it
works above the ambient temperature, one cannot do the measurements at 25° C or
below this. Thus it was necessary to have a proper cooling assembly so that the
surrounding of the instrument is kept colder than the working temperature. For this
we have designed and fabricated the special refrigerated thermostated house
working between 0 to 40° C and the instrument was kept in it (figure 2.9). The
accuracy in osmotic coefficient (<j>) measurements was found to be better than ± 1 x
10'3. Comparison of the osmotic coefficient NaCl and KC1 in water obtained in this
work with the literature data313 is shown in figure 2.12a and 2.12b, respectively,
which shows the authenticity of the osmotic vapor pressure measurements done in
this work.
1.00
O'
O'
O
OO
ON
r-
(</>)
osmotic coefficient
xo
Gn
o
o
*•
vO
£
-
0.93
0.92
0.91
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8
m 1/2 /mol kg'1
m 1/2/mol kg1
nr\P0-P)t
^ Wl
where (Po -P) is a pressure drop down the tube of length 7’ and ‘ V is the volume
of the liquid flowing through the tube in 7’ seconds. When the flow of liquid of
density ‘tf is vertically downward through the capillary then,
nrAhmdgt
M
where (Po - P) = h^dg, where hm is the average height of the flowing liquid and g
is the acceleration due to gravity. The equation was derived by Poisuelle on the
basis of following important assumptions:
a) The flow is everywhere parallel to the axis of the tube.
b) The flow is steady; the initial disturbances due to the accelerations from
rest have been damped out.
c) There is no slip at the walls of the tube.
d) The fluid is incompressible.
e) The fluid will flow when subjected to the smallest shearing force, the
viscous resistance being proportional to the velocity gradient.
These conditions are fulfilled for the flow of liquids through the capillaries of
small and uniform bore. Capillary viscometers can be broadly grouped278 as
under.
(1) Viscometers for absolute viscosity measurements
(2) Viscometers for relative viscosity measurements
a. Applied pressure viscometers
b. Kinematic viscometers
The instruments in section (1) are normally only used for the establishment
of primary standards and require very accurate control of applied pressure as well
as precise knowledge of the capillary dimensions. Type (2) viscometers are those
in which measurements are made relative to a primary standard liquid (obtained
with type (1)) or to derived secondary standards. Variable pressure instruments
under subsection (2)a typified by the Bingham viscometer use an externally
109
applied pressure to force standard and test liquids through their capillaries. The
second class, (2)b, as exemplified by the Ostwald viscometer, utilizes the head of
the liquid itself to furnish the pressure differential across the capillary. The
inherent simplicity of these latter kinematic viscometers accounts for their almost
universal use in electrolyte studies.
23.3 Design of viscometers
Generally the Ostwald, Cannon Fenske,319 Tuan-Fuoss,316 Jones Vezeys
and Ubbelohde320 viscometers are used for the viscosity measurements. These
viscometers are common capillary viscometers called kinematic viscometers.
Study of sources of errors in different kinds of capillary viscometers led to the
conclusion that the Ubbelohde viscometer (Figure 2.13a) is more suitable for
aqueous solutions while the Tuan-Fuoss viscometer (Figure 2.13b) is better for
non-aqueous volatile liquids. The use of viscometers like Ostwald type gives rise
to following systematic errors.
1) The volume of liquid which is filled in the viscometer requires to be known
very accurately to know the exact pressure head.
2) Due the surface tension the liquid in the upper bulb A (figure 2.13a)
experiences a force in the upward direction. The upward pull varies as the
liquid volume changes during its flow. This effect changes the effective
pressure head and unless a suitable correction in its construction is made, it
introduces error in the measurement of flow time.
3) Some liquid stick to the walls of the upper bulb A leading to the reduction
of the effective volume of the liquid.
By considering all these drawbacks in the present work, the Ubbelohde
viscometer has been used, the design of which is shown in figure 2.13. The
precision of the side tube at the lower bulb B in figure 2.13 makes the pressure
head independent of the volume of liquid contained in the viscometer.
23.4 Kinetic energy correction
The effect arises when the applied pressure is not solely dissipated by
viscous forces, some being used to accelerate the liquid in the entrance reservoir
from the rest to the velocity distribution characteristics of the liquid when obeying
Poiseulle’s Law. Except at very low rates of flow, when there may be a
compensating effect at the exit of the capillary, this extra kinetic energy is
dissipated as heat in the exit reservoir. The correction is therefore dependant on the
110
D1
>
A A
Tw
2$cc J 3.6
------ -
w«ft%
flow rate and beconanr negligible at very low values. Its calculation is given in
detail by Barr321 and leads to the expression:
mdV2
Peff =(Po~P)-
JlW
where peS is the pressure actually used in overcoming viscous forces, d is the
density, and m is a numerical coefficient (~ 1) introduced to take account of
particular shape of the capillary ends. Therefore in equation {po - p) must be
replaced by peff as given by above equation
7tr4/ . , mdV
If mean vertical height of the pressure head is hm then the pressure (po-p)~ h„dg.
Therefore we have
_ ur*hmdgt mdV
T1_ Wl ~SuIt
mdV
where a'= and p'= are the constants for apparatus. These can be
SVl Sul
calculated from the dimensions of the apparatus. For applying kinetic energy
correction for the relative viscosities, we can write
r\ a'dt-$'(d/t)
flo a'^o ~$'(d0/t0)
where do and to are the density and efflux time for solvent respectively, we have,
q c/(a7-p7Q
flo d0{a't0-^'/t0)
To make necessary corrections in relative viscosity for the loss of kinetic energy
Kaminsky has derived the following equation
flo ^0^0
112
(
I 1 1 1
where K = \ + k +k 4
v'J t2t2
*0*
and K = —------
w Kg
where K is called the kinetic energy correction factor. K varies with the difference
between t and to. When k « t] the correction factor is negligible.
The kinetic energy correction can be reduced most easily by decreasing the
rate of flow and having a small efflux volume. The time of flow for a given
viscometer can be increased by decreasing the radius of the capillary bore but there
is a practical limit to this course because the presence of dust and fibbers can cause
non-reproducibility in very fine bores. The compromise usually adopted in
measurement to employ capillary radii in the range 0.03 to 0.08 cm.
2.3.5 Surface tension correction
An error due to surface tension may arise from the adhesion of liquid to the
walls of the bulbs immediately above the capillary and in the exit reservoirs. This
has the effect of altering the hydrostatic pressure head of the liquid. In relative
measurements its magnitude will be dependent on the difference in surface
tensions of the calibrating and test liquids. The surface tensions of aqueous
solutions of electrolytes and of low viscous liquids change very little with
concentration, for example the value of 2M sodium chloride solution at 25° C is
only 5% higher than that of pure water. However, if non-aqueous solutions are
being calibrated against water, it must be taken into account; for example the
surface tension of ethyl alcohol is less than a third of that of water.278
The Ubbelohde viscometer is reputed to eliminate the surface tension effect
by the provision of the suspended level at the top of the bulb B which is so
designed as to compensate for the forces operating in the bulb A.
The Tuan-Fuoss viscometer is shown in figure 2.13b. The instrument is
designed so that a series of dilutions can be made on a sample of solute without
opening the viscometer between check runs at a given concentration. In this way,
evaporation errors are eliminated, and contamination of the viscometer contents by
dust or moisture are minimized. The viscometers fit into racks in the thermostat
(298.00 ± 0.005 K) which position them accurately vertical, and which permit
rapid removal and remounting. The working volume is refilled by taking the
113
CONTACT THERMOMETER
’ THERMOMETER! t/1**)
stirrer
■ECKMAN THERMOMETER
~* TO DRIVING
HEATER
InigH^MT IT
IlMRtMPUnP IAI
.cormtuRiHC
INSULATION
CLABSAqi'ARRJM
VISCOMETER.1
VISCOMETER}
copper-tubing fixed inside the experimental bath using another German made U-10
ultra-thermostat or cryostat. The temperature of the bath was measured by means
of standard thermometer (1/10° C). The small fluctuations were noted by Beckman
thermometer. The temperature of the bath was controlled to be better than ± 0.005°
C. Further the movable stand for the thermostat has been also fabricated to avoid
the space problem in the laboratory. The photograph of actual working assembly
for the viscosity measurements is shown in figure 2.15.
115
Figure 2.16. Leveling device for fixing and positioning of the viscometers.
viscosity (%) of all samples at 25° C. At other temperatures the time of flow of
water for the corresponding temperature was used to calculate the relative viscosity
of the samples at that temperature. It was essential to use a Tuan-Fuoss viscometer
for CCI4 solutions (the flow time for pure CCI4 is 659.1 s at 298.15 K), as CCI4
evaporates, causing variation in the time of flow when an Ubbelohde viscometer is
used. 18 Since the Tuan-Fuoss type viscometer was constructed using stoppered
standard joints, the evaporation of CCI4 could be prevented and a constant time of
flow value could be determined. All the flow time measurements reported in this
work are averaged over three separate measurements of the same solutions. The
calibration of a Ubbelohde viscometer has been described earlier. 14 The
uncertainty in viscosity measurements with reference to water is estimated using
the guide to the expression of uncertainty in measurements (International
Also, the viscosity of 20% sucrose solution was determined at 298.15 K and the
value 1.6953 mPa s is obtained which agrees well with the literature278 value of
1.695 mPa s. These reproducibility tests of the literature data determine the
accuracy and precision of the experimental work we have carried-out.
119
- iy jc ,(mo\ dm'3)
(v\r
Vc /(mol dm'3)172