REFERENCE OF TECHNICAL
GANTRY
Gantry Aperture 70 cm
360° Rotation Speeds 0.5,0.6,0.75,1.0,1.5,2.0
Slices 16
Gantry controls on right and left of gantry Yes
Gantry controls on front and back of gantry Yes
Maximum Scan FOV 50cm 50cm FOV
Tilt ±30°
Gantry Cooling Method Air Cooling
3D Laser positioning lights Yes
Intercom system Yes
DETECTOR
Solid State CT Detector Yes. GOS.
Detector rows in z direction 24
Slices in z direction 16
Total detector width @ isocenter 24mm
X-ray Generator
Generator power 50 kW
Three kVp Selections Three (90,120,140)
mA Range 30 mA – 420 mA
X-Ray Tube
MaximumTube Heat 5 MHU
Large focal spot 1.0 x 1.3
Small focal spot 0.5 x 1.3
Dose Management
Adhere to ALARA Principles Yes. DoseWise Principles
ACS (Automatic Current Selection) DoseRight ACS
Dynamic Dose Modulation (DDOM) DoseRight D-DOM
z-axis Dose Modulation (zDOM) DoseRight z-DOM
Display of CTDIvol and DLP during planning Yes.
Dedicated Pediatric Protocols Yes.
Iterative Reconstructions
Iterative reconstruction system iDose4
Patient Table
Scannable range 150 cm
Maximum Weight at Full Accuracy 200 kg
Carbon Fiber Top Yes
Table Accessories Yes
Remote Table control from Console up/down, in/out
Reconstruction
Maximum reconstruction rate Up to 20 ips
Concurrent reconstruction Yes.
Offline reconstruction mode Yes.
Up to six different reconstruction assignments Yes.
Reconstruction Matrix of 768² and 1024² Yes. 512², 768² and 1024²
Image Quality
Low Contrast Resolution 2mm @ 0.3% @ 40mGy surface
Absorption Range -1024 to +3072
Scanner Console
Memory ≥16GB
Hard disk capacity ≥1000GB
DVD/CD Archive 4.7 GB
Remote control from console Table: Up/down, In/out
Spiral
Continuous Spiral Ontime Up to 100 seconds
Maximum scannable range in spiral 150 cm
Image Processing/Review
Dental CT Software yes
Slab Viewer yes
Cursor for pixel value measurements yes
Angle measurements yes
Invert Window yes
RelateSlice yes
3D Small Volume Analysis yes
Curved MPR viewing yes
3D display viewing yes
Volume rendering viewing yes
Maximum Intensity Projection (MIP) yes
Minimum Intensity Projection (minIP) yes
ROI yes
Volume Calculation yes
CT Number Display yes
cine yes