Backfill Techniques

Syringe Backfill Technique Single-Cone Backfill Technique (Ideal for Medium and Large Canals)
• The speed of extrusion is set on the control • Downpack through the master cone as usual,
panel of the System-B/Elements unit. After but do not perform the one-second separation
pre-heating (45 seconds) the forward toggle burst. Instead, allow the plugger to cool in the
switch on the handpiece is pressed until mate- canal for about 10 seconds (2 clicks from the
rial extrudes out of the needle tip, to “prime” System-B/Elements unit).
the needle. Note: when using RealSeal set the
heat on your System-B/Elements unit to “S” for • Break the plugger loose by rotating it back and
synthetic. forth with apical pressure, and then while
rotating it, tease out the still-cold plugger.
• Place the heated needle into the canal and wait
a full 5 seconds, allowing the needle to reheat • Place an AutoFit™ backfill cone (the same size
after being cooled by contact with the dentin. as the plugger used for the downpack) coated
with sealer into the empty backfilling space
• After the 5 second pause, with the needle lightly and move it in and out 3 -4 times to make sure
held in place, press one of the handpiece toggle that sealer coats the backfill space. The area
switches (back button for medium speed; left by the plugger will exactly match the
forward button for faster speed) and begin to shape of the backfill cone.
extrude the gutta percha.
• Sear the cone off at the orifice level with the
• After the extruded material fills the backfill System-B/Elements electric heat plugger.
space ahead of the needle, the backpressure of
the extruded gutta percha will move the needle • Place the rigid stainless steel end of the hand
back out of the canal. It’s important that you plugger against the gutta percha, and with a
resist the temptation to pull the needle out of firm sustained pressure, condense the coronal
the canal... let the extruded gutta percha back mass at the orifice level.
it out. Take your time, and allow about 5-10
seconds for the needle to reach the orifice level. • This technique is also ideal for removing voids
created during an extruder backfill. Adjust the
• Using the rigid stainless steel end of the appro- stop on the CW heat plugger so that it reaches
priate CW hand plugger, give the warm gutta beyond the existing void. Plunge the heated
percha a very firm condensation push. You can plugger through the void, cool for 10 seconds,
eliminate a void 4 mm’s away if you apply then remove the cold plugger and complete
enough pressure here. the backfill using the steps outlined above.

• If you are using RealSeal, light cure the coronal • If you are using RealSeal, light cure the coronal
surface for 40 seconds. surface for 40 seconds.