S5 Perfusion System
Concept 4–5
Console 6–7
3
Pumps 8 – 17
System Panel 18 – 21
Accessories 32 – 40
S5 Pump Table
7
Two height-adjustable push bars on the left and
right side of the console can be used for moving
the S5 and mounting accessories.
The S5 mast extension system can carry up to 3 mast The swivelling S5 mast extension system allows
roller pumps. The separate control units are mounted the pumps and disposables to be positioned close
away from the pump heads on the console mast sys- to the patient.
tem. The mast extension system can be attached on
the left or right side of the console and can be folded The mast holders for the accessories are compact
if the system is to be moved. A supporting castor is and easy to use.
used to stabilise the mast system. The pumps, oxy-
genator and tubing sets can be positioned close to The system facilitates shorter tubing lengths result-
the patient. ing in lower priming volumes.
9
Working position 2
Working position 1
Transport position
11
The compact holders, which were specially devel-
oped for the mast roller pump, have convenient fast
clamp connectors. They can be connected to/discon-
nected from the mast quickly and easily. The pump
housings are made of stainless-steel and very easy
Product designation Part number to clean.
S5 mast roller pump system 150
50-80-00
consists of: The MRP 85 was designed for low flow rates and
S5 mast roller pump 150 is especially suitable for paediatric and infant/
10-88-00
(connection cable supplied) neonate perfusion.
S5 control panel for mast roller
pump 150/85 (connection cable 28-95-80 The system facilitates the use of shorter tubing re-
supplied) sulting in lower priming volumes.
This established Centrifugal Pump System gives the The system comes complete with a drive unit, con-
S5 operator a high degree of flexibility. The mast as- trol panel, connection cable and emergency system.
sembly saves space. There is no need for an extra To further improve safety and ease of use we also
pump space on the console. provide a unique electrical remote-controlled tubing
clamp. If an alarm situation is triggered (e.g. level/
The Stöckert SCP Centrifugal Pump System can be bubble alarm or retrograde flow), the electric re-
upgraded without difficulty. Using an S3 to S5 adap- mote-controlled tubing clamp closes the arterial line
ter cable (see page 35), SCP Systems that are being in a fraction of a second keeping the danger of air
used with a Stöckert S3/SC heart-lung machine can delivery to a minimum.
also be connected to the S5. The S5‘s built-in moni-
toring and control units, e.g. pressure control, level
monitor and bubble detector, cut down on costs.
Pumps
13
Electrical remote-controlled
tubing clamp (for installation on
60-05-60
mast) (for S5 System exclusively)
consists of:
Tubing clamp
3-joint mast holder with fast
clamp connector | 500 mm
Connection cable with
45-12-02
S5 System plug
Electrical remote-controlled
tubing clamp (for installation on
60-05-65
mast) (for S5 System exclusively)
consists of:
Tubing clamp
3-joint mast holder with fast
clamp connector | 620 mm
Connection cable with
45-12-02
S5 System plug
The roller pumps are the centrepiece of the S5 Perfu- S5 Roller Pump
sion System. Every pump has an independent control
system and its own pump control panel that is oper- The S5 roller pump with a 150 mm diameter is used
ated using a high-contrast colour touch screen. The primarily for arterial blood flow, cardiotomy suction
pumps can be individually configured and the moni- and vent suction.
toring functions for each pump can then be linked
and displayed on the touch screen. An innovative,
durable setting knob (incremental shaft encoder) is S5 Double Roller Pump
used to electronically adjust the set rotational speed.
A maximum of five roller pumps can be placed on The S5 double roller pump combines two roller
the pump table and connected to the E/P pack. The pumps with a diameter of 85 mm in a single hous-
optimised pump head geometry (horse-shoe shaped) ing. Both roller pumps can be operated and con-
keeps pressure peaks during operation to a minimum. trolled independently. The double roller pump is es-
pecially suitable for paediatric/neonate perfusion or
as a cardioplegia pump; it can also be used for
cardiotomy and vent suction.
Pumps
15
Product Advantages
For perfect positioning relative to the surgical site, All monitoring functions and the pump‘s current
the pump heads can be rotated 180° (MRP 150) or status are displayed on the touch screen. This
240° (MRP 85). They can be rotated in increments ensures the operator can quickly respond during
of 15° (self-locking). an alarm situation.
The pump head design makes inserting the pump The extremely precise pump head ensures that the
tubing quick and easy. occlusion is exact and consistent.
In master/follower mode, the follower pump reacts Settings are entered using the individual pump‘s
to any master pump rotational speed changes. structured menus.
Each pump can be operated in continuous or If two Override keys are pressed simultaneously,
pulse mode. the monitoring functions assigned to a pump are
overridden.
Large digital rotational speed/flow rate displays
provide a constant overview of the current pump Every pump has a stop link function. This function
speed. can be connected to the arterial pump. If the arte-
rial pump stops, the stop link pump also stops.
17
S5 Blank Module
21
Pressure 1 ECC-Timer
Pump 1
100 %
pressure to a value below 200 mmHg and the 424 mmHg 0:00 Pump 1
75:55
0.00 75 %
0 rpm
pressure continues to rise until it reaches a dis-
l/min 3/8"
Pressure 2
80 mmHg 35:05
played value of 424 mmHg, the alarm is triggered. Temperature 1 Temperature 2
Pressure 1
Pump 1 (1) stopped
Pressure 1 ECC-Timer
Pump 1
When the cause of the pump stop has been re- 105 mmHg 0:00 0.00 75 %
l/min 3/8" 0 rpm
75:55
moved and the pressure in the arterial line drops Pressure 2
80 mmHg 35:05
below 200 mmHg (in this example to 105 mmHg), Temperature 1 Temperature 2
the roller pump restarts, and the red alarm signals 20.5 °C 10.4 °C
100
inform the operator that a pump stop took place.
%
Pressure 1
Pump 1 (1) had stopped
Temperature 1 ECC-Timer
0.0 °C 0:00
Temperature 2 75:55
0.0 °C 35:05
Temperature 4 Timer 1
25.4 °C 5:00:00
S5 Pressure Monitor
This function has two pressure controls. Two pres- The double pressure monitor is used for:
sure parameters (alarm/stop and warning/control)
with both visual and acoustic alarm can be set for Measuring and displaying the pressure in the ex-
each of the controls. The sensor module has two tracorporeal circuit. The display range extends
transducer connectors. from -200 mmHg to +800 mmHg. The values can
be displayed in either mmHg or kPa.
Pressure 1 ECC-Timer
20.5 °C 10.4 °C
The temperature monitor allows the simultaneous If the temperature limit is reached, the visual and
measurement and display of up to four tempera- acoustic alarms are triggered.
tures. One temperature sensor module channel is
reserved for cardioplegia control. An outstanding feature of the measuring channels
is their measurement precision.
An upper and a lower temperature limit can be
set. The values are set on the control and display Temperature probes (listed under “Accessories”)
module on the system panel. are required to operate the monitor.
Pressure 1 ECC-Timer
20.5 °C 10.4 °C
Accessories (optional)
Temperature probes for direct
measurement in the oxygenator
(optional)
For COBE/DIDECO oxygenators 45-03-10
For BAXTER oxygenators 45-03-11
For MEDTRONIC oxygenators 45-03-30
Accessories (optional)
Vinyl probe for oesophageal or rectal temperature, Ø 4.8 mm 45-03-01
Thin, flexible probe (used as 45-03-01), Ø 2.8 mm , 165 mm long 45-03-02
Thin, flexible nylon probe with epoxy tip (used as 45-03-01), Ø 3.2 mm, 75 mm long 45-03-03
Stainless steel surface probe for skin, oral and auxiliary measurements 45-03-04
Surface probe: one side made of epoxy, the other made of stainless steel, for use with temperature
45-03-05
connectors
Rigid bar probe for fluids, 115 mm long, Ø 4.0 mm 45-03-07
Connector for temperature probe 45-03-05: 1/2” 31-15-01
Connector for temperature probe 45-03-05: 3/8” 31-15-02
Connector for temperature probe 45-03-05: 1/4” 31-15-03
Modules and Sensors
S5 Level Monitor
The level monitor controls the blood level in the The sensor is based on the effect the liquid level
oxygenator/reservoir. has on high-frequency, electromagnetic waves in an
electric oscillating circuit. Using this method, inter-
The level sensor is mounted on special dis- ference caused by electromagnetic or external light
posable, adhesive mounting strips which are at- sources and residual liquid on the oxygenator‘s wall
tached to the reservoir or oxygenator. The sensor is avoided.
can be used with most hardshell oxygenators and
reservoirs.
25
Bubble Level
3/8"
1/2"
Cardioplegia
0.000 l 0.000 l
00:00 1/4"
21 mmHg
24.5 C
This monitoring device detects air bubbles and The 3-joint mast holder with a fast clamp connec-
micro-bubbles in the extracorporeal circuit. tor allows the bubble sensor to be perfectly posi-
tioned at any point in the tubing system.
This function requires a sensor module and the cor-
responding sensor. The scope of delivery includes a If air bubbles are detected, a visual and acoustic
bubble detector that can be used for tubing sizes alarm is triggered and the connected pump stops.
from 3/8“ to 1/2“. Alternatively, 3/8“ and 1/4“ sen-
sors are available. You can set three different bubble detection alarm
thresholds (4 mm, 5 mm und 6.5 mm Ø).
Bubble Level
3/8"
1/2"
Cardioplegia
0.000 l 0.000 l
00:00 1/4"
21 mmHg
24.5 C
S5 Cardioplegia Control
This unit can be used with an S5 roller pump or an The integrated timer automatically starts during a
S5 double pump to deliver cardioplegic solutions or pump stop regardless of operational mode and re-
blood cardioplegia during an operation. The operator cords the ischaemia time.
can choose between two operating modes that can
be selected in the menu of the control and display The (total) volume delivered since the start of
module. cardioplegia is accumulated and displayed.
Manual Operation Using two roller pumps (e.g. both roller pumps of
The operator can start and stop the pump. The dose a double pump), a cardioplegic solution can be
27
volume to be delivered counts up on the volume dis- delivered in a finely dosed, defined mixture. The
play (beginning at 0). mixture ratio is kept constant.
Gas blender
Air+O2 FiO2 CO2
30.4 37.0 °C 1 2
32.5 40.0 °C
17 14.0 °C
Accessories (included)
Cable for connection to the
45-12-02
S5 System
Accessories (optional)
Mast holder with fast clamp
55-91-50
connector
Console holder for gas blender
(only compatible with 4 and 25-40-70
5 position consoles)
Mast-Mounted Venous Line Clamp
The mast-mounted venous line clamps allow very The tubing diameter can be adjusted quickly and
precise regulation of the venous return flow. precisely using the coarse and fine adjustment
The tubing diameter is set directly on the clamp. knobs on the clamp.
This type of clamp can be used to advantage when
the system arrangement allows the operator direct The line clamp is self-locking in every position.
access to the clamp.
The logarithmical characteristic curve of the line
Features of the venous line clamp: clamp means the tubing size can be quickly reset.
The venous line clamp can be attached anywhere In an emergency situation, the clamp can be
on the mast system. opened quickly with a single hand movement.
The venous line clamp with a mechanical remote An overview of the remote-controlled venous line
control combines the advantages of a safe and relia- clamp‘s advantages:
ble mechanical clamping of the venous line with the
ease of use of an individually positionable control The line clamp‘s design allows it to be perfectly
unit that is always close to hand. positioned on the venous line for every case.
Because of its lightweight design and secure position- The separate control unit and line clamp facilitates
ing on the venous line, the clamp head can be placed a flexible assembly and guarantees extreme ease
31
close to the venous reservoir without additional of use.
support. The routing of the venous tubing is not
negatively affected nor is there kinking. If the clamp Unlike an electronic line clamp, this clamp does
head is fixed to another part of the venous tubing not have to be calibrated and is always ready for
system, it might well be necessary to support the use.
clamp head with a joint holder.
The structural design of the line clamp means that
The 1-metre Bowden cable connects the clamp head the clamped line can be visually monitored at all
with the control unit and transfers settings entered times.
on the control unit immediately to the clamping
lever in the clamp head. The control unit is mounted The use of proven mechanical components guaran-
on one of the S5‘s push bars (right or left – depend- tees a low level of maintenance and repair.
ing on the system arrangement and/or ease of use).
The coarse and fine setting knobs are used for setting
the tubing diameter and regulating the venous return
flow quickly and precisely.
S5 Pump Spacer
33
S5 Drawer Module
powerful luminosity
longevity
energy-efficiency
Cable Holders
35
S3 to S5 Adapter Cable
Perfusionist‘s Chair
The Heater-Cooler System facilitates the precise and The two-chamber pumps do not lead to a pressure
fast regulation of the patient‘s blood temperature build-up in the heat exchanger and facilitate the
regardless of the hot/cold water supply in the operat- simultaneous supply of two water circuits, e.g.
ing theatre. heat exchanger and cooling/heating pads.
This standalone device has one water tank and two An independent safety system stops the water
water circuits. It is operated and monitored from its temperature reaching critical values.
own control panel. An ice container that can be used
to quickly lower a patient‘s temperature or to cool Turning the ice container on intermittently short-
liquids (e.g. cardioplegic solutions) is available as an ens the cooling off periods.
option.
The pump suction stage ensures that the heat ex-
Product Advantages: changer and tubing are easily purged.
Accessories (optional)
Telescope mast for heater-cooler 16-09-20
Ice container 16-02-60
Heating-cooling blanket for adults
16-10-50
(55 x 150 cm)
Heating-cooling blanket for
16-10-51
children (55 x 90 cm)
Tubing set for heating-cooling
16-10-55
blanket
1 adapter connector set for blanket 16-10-10
Accessories
Heater-Cooler System 3T
Gas blender
Air+O2 FiO2 CO2
The Data Management System (DMS) facilitates the An extensive integrated statistical analysis and
automatic and complete collection and evaluation evaluation function for the complete database
of data during and after open-heart surgery when the
heart-lung machine is used. Data can be transferred to and from the clinic‘s
EDP.
The DMS includes:
Using the DMS software with its user-friendly
A DataPad (panel PC) that uses cutting-edge tech- Windows interface is very easy.
nology
Data is recorded automatically during the opera-
S5 interface module tion, and additional data may be entered manual-
ly at any time.
DMS software for data processing before, during
and after operations The DMS also processes data from external de-
vices such as blood analysis devices, patient mon-
The DMS software is based on a server online con- itors etc. online (a list of connectable external de-
cept. Data created during operations is stored on a vices is available on request). These external
main server. The online part of the DMS is used to devices communicate with the computer via data
store data during surgical procedures and operation interfaces.
of the heart-lung machine. The data transfer between
the computers is carried out via disk or network. Essential data is graphically displayed on the
screen making trend analysis possible at all
The advantages of the DMS with its server online times.
concept includes the following:
An operator-specific perfusion protocol is auto-
The data from all operations is stored in a central matically created from the stored data. The perfu-
database. sion is displayed as a graph and a table.
It avoids storage of inconsistent data. The operator can use the Setup Wizard to very sim-
ply create their own user-specific configuration.
It facilitates an identical configuration of all com-
puters in a hospital.
39
Easy to use
Note:
If you would like to receive further information
about DMS software, please contact us for a detailed
description of the programme.
Fibrillator
Accessories (included)
External power supply 26-16-10
Electrode, both sides conducting 28-10-15*
Accessories (optional)
Electrode, one side conducting 28-10-20*
Electrode, for children 28-10-25*
Potential equalization cable (5 m) 45-10-50
Operating conditions
41
Operating temperature + 10 °C ... + 40 °C
Storage temperature 0 °C ... + 40 °C
Relative humidity (operating and storing) 30% ... 75%
1.2 Masts
Maximum permissible load Mast extension system (optional)
Maximum total load on mast system 45 kg Maximum load on the telescope mast 40 kg
Maximum load on a mast 20 kg (1)
Maximum load on the vertical mast 11.5 kg
Maximum load on the infusion rack 5 kg
1.3 Pumps
Double roller Mast roller Mast roller 2 mast roller
Roller pump 150
pump 85 pump 150 pump 85 pumps 85
Height 285 mm 257 mm 289 mm 237 mm 237 mm
Width 180 mm 180 mm 178 mm 116 mm 260 mm
Depth 485 mm 485 mm 299 mm 175 mm (3)
200 mm (5)
Weight 15 kg 12 kg 11.9 kg (2) 5 kg (2) 11 kg (4)
Concentricity
Pump raceway 0.03 mm 0.03 mm
Occlusion symmetry 0.03 mm 0.03 mm
Occlusion rollers 0.015 mm 0.015 mm
(1)
max. swivel arm 200 mm; (2) with fast clamp connector; (3) without fast clamp connector; (4) with double holder; (5) without double holder
Deviation of speed slave pump max. 1 percentage point of the flow ratio setting
Display and control module Control module for mast roller pumps
Height 125 mm 260 mm
Width 179 mm 190 mm
Depth 48 mm 100 mm
Weight 0.5 kg 3.5 kg (6)
2. Electrical Specifications
2.1 Electronics and Power Pack
Input voltages 100 V ~to 240 V~; 50 / 60 Hz
Permissible mains voltage fluctuation ± 10%
Maximum power consumption (standard equipment) 1000 W
(6)
with holder
Technical Specifications
Level
Alarm limit (level sensor) for oxygenators/reservoirs made
of rigid polycarbonate, wall thickness at sensor position Level display of the sensor holder ±10 mm
3 mm max.
Pressure
Measurement range -200 mmHg to +800 mmHg
Resolution 1 mmHg
Cardioplegia
Pressure measurement range -200 mmHg to +800 mmHg 43
Resolution 1 mmHg
Temperature monitor
Display range 0 °C to +50 °C
Timer
Counting range 0 - 999 min 59 sec
Timer (optional)
Counting range 0 - 10 h (up and down)