Test Certificate

You might also like

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 3

Certificate no : JTC-PT/07-18/205 Dated : 08.07.

2018

TEST CERTIFICATE

CLIENT : Dhanvantri Hospital


ADDRESS : Vasant vihar naubasta, Kanpur
DEPARTMENT : Operation Theater
EQUIPMENT TESTED : Clean Room
TYPE OF TEST : PARTICLE COUNT TEST DATE : 08.07.2018
DESIGNED CLASS : 1000 DUE ON : 08.01.2019
INSTRUMENT USED : METONE LASOR BASED PARTICAL COUNTER
PACIFIC SCIENTIFIC INSTRUMENTS, USA MAKE
SR. NO.237 B
Date of Calibration : 24.09.2017 Due On : 24.09.2018
(of Particle counter)

TEST RESULTS

Air PARTICLES PER CUB. FT. PER MIN. REMERKS


SR. Velocity Size of Partical
AREA FPM
NO. Count Code 0.5Microns 5 Microns Avg. count0.5 Cleanliness
Micron Label Class
Location LIMIT LIMIT 1000
1000
1. O.T. 125 A 856 _

B 752 _
C 710 _ 801 1000-OK
D 895 _
E 791 _
For JOY CLEAN AIR SYSTEM Certified that the tests were carried out
Accordance with ISO 14644
Tested By : A.K. Sharma
Signature : :
Certificate no : JTC-PT/07-18/206 Dated : 08.07.2018

TEST CERTIFICATE

CLIENT : Dhanvantri Hospital


ADDRESS : Vasant vihar naubasta Kanpur
DEPARTMENT : Operation Theater
EQUIPMENT TESTED : Clean Room
TYPE OF TEST : Air Changes Test TEST DATE : 08.07.2018
NO OF FILTERS : 6 VALID UPTO : 08.07.2019
INSTRUMENT USED : Vane Anemometer , Model – AM-4201 , Sr. No-Q883095
Date of Calibration : 20.01.2018 Due On : 20.01.2019
(of Particle counter)

TEST RESULTS

Volume of Grills Grills Area Velocity Air Supply AirChanges Remarks


Room In Size in in Sq ft. in Ft./min CFM Per hr. Q x
Cubie ft. Sq ft (A) (V) Q=VxA 60/Vol

3610 2x1 2 300 600 63.49 Observed


2x1 2 280 560 Value
2x1 2 280 560 complies as
per
2x1 2 350 700 acceptance
2x1 2 380 760 Criteria
2x1 2 320 640
3820 63
(say.)
For JOY TRADING CO. Certified that the tests were carried out
Accordance with ISO 14644
Tested By : A.K. Sharma
Signature: :
Certificate no : JTC-PT/07-18/207 Dated : 08.07.2018

TEST REPORT

Test Date: 08.07.2018 Due On : 08.07.2019

Client : Dhanvantri Hospital


Address : Vasant vihar naubasta Kanpur
Area Tested : Operation Theater
Type of Filter : HEPA Filters
No Of Filters : 6
Dimensions : 610 X 610 X 150 mm (L x W x D)
Rated Capacity : 1000 Cfm
Nature of Test : Filter Integrity Test

Standard Instrument Used For Test:


Instrument Name Make Model Oil Used

Aerosol Photometer TECS INC. PH-2003 _

Aerosol Generator ATI 6C PAO Oil

Observations:
Name Filter No. Upstream Conc. Leakage found Remarks
in % (Filter Integrity)
1 100% 0.001 PASSED
2 100% 0.003 PASSED
Operation 3 100% 0.003 PASSED
Theater 4 100% 0.001 PASSED
5 100% 0.002 PASSED
6 100% 0.001 PASSED

Acceptance Criteria: Less than 0.01 % leakage accepted

Filter efficiency : 99.997% down to 0.3 Microns

Approved by