Professional Documents
Culture Documents
ECRI Hemodialysis Units
ECRI Hemodialysis Units
Hemodialysis Units
UMDNS information
This Product Comparison covers the following
device term and product code as listed in ECRI’s
Universal Medical Device Nomenclature Sys-
tem™ (UMDNS™):
• Hemodialysis Units [11-218]
Purpose
Hemodialysis units perform extracorporeal dialysis
to replace the main activity of the kidneys in patients
with impaired renal function, such as those with end-
stage renal disease (ESRD). The kidneys maintain the
body’s fluid, electrolyte, and acid/base balance, counter-
acting the destabilizing influences of metabolic activity
and a constantly changing external environment. Im-
paired kidney function causes the body to retain meta-
bolic wastes and water. Hemodialysis therapy removes
these, as well as ions and organic salts, from the blood-
stream. It does not restore renal function or promote
healing of the kidneys, but restores a reasonable state
of health by partially performing renal functions,
Arterial Blood
Extracorporeal
Dialyzer
Blood Circuit
C331UN3A
Source
Heater
To Drain
thereby minimizing further damage to other organs contaminants, such as minerals and bacteria. Water-
and physiologic systems. treatment systems typically use depth filtration, water
softeners, activated carbon filtration, reverse osmosis
Principles of operation (RO), and deionization (DI) to achieve the standard
Single-patient hemodialysis systems can be divided acceptable level of contaminants.
into three major components: the dialysate delivery
system, the extracorporeal blood-delivery circuit, and Treated water enters the dialysis machine and usu-
the dialyzer. Blood is taken via the extracorporeal ally passes through a heater and a deaerator before
circuit, passed through a dialyzer for solute and fluid being mixed with the concentrate to form dialysate.
removal, and returned to the patient. Each system has Two types of proportioning systems are used to mix the
its own monitoring and control circuits (see Fig. 1). water and concentrate: fixed-ratio controllers mix spe-
cific amounts of each, while servo-controlled systems
Dialysate delivery system monitor the conductivity of the dialysate and regulate
the delivery of concentrate to satisfy specified conduc-
The delivery system prepares dialysate — a solu- tivity and pH limits. The temperature of the dialysate
tion of purified water with an electrolyte composition is kept in the 34° to 42°C range to prevent excessive
similar to that of blood — and delivers it to the dia- cooling or heating of the blood. The temperature and
lyzer. Dialysate acts to remove metabolic wastes from conductivity sensors can initiate alarms and divert the
the blood and also acts as a source of ions to maintain dialysate away from the dialyzer if the conductivity or
the blood’s proper electrolyte and pH levels. Either temperature is not within specified limits. Some sys-
acetate or bicarbonate concentrate is included in the tems monitor other parameters, such as the pH, to
dialysate as a buffering agent. Additional water is determine dialysate status.
mixed into the dialysate to approximate normal bicar-
bonate ion blood concentrations. Extracorporeal blood circuit
To prevent long-term toxic effects, incoming water The external blood-delivery system (extracorporeal
must be treated to remove inorganic and organic blood circuit) circulates a portion of the patient’s blood
2 ©2003 ECRI. Duplication of this page by any means for any purpose is prohibited.
Hemodialysis Units
C065HN1A
pressures are monitored on both venous and arterial
lines; high- and low-pressure alarms turn off the blood
Semipermeable Membrane
pump if necessary.
Because blood tends to clot when it comes into Figure 2. The diffusion process causes an exchange
contact with foreign surfaces such as those in the of small-molecular-weight blood and dialysate ions
tubing and dialyzer, heparin, an anticoagulant, is in- across the semipermeable membrane.
fused through a syringe pump aseptically connected to
an infusion line in the arterial side of the blood circuit.
The infusion pump can be set to deliver heparin at a Balanced Concentration Results in
predetermined rate. A drip chamber on the venous side Equilibrium of Diffusable Solutes
of the blood circuit contains a clot-trapping filter to
help prevent upstream clots and other debris from
reaching the patient.
Dialyzer
©2003 ECRI. Duplication of this page by any means for any purpose is prohibited. 3
Healthcare Product Comparison System
proteins are too large to pass through the pores in the This data can then be analyzed and expressed in graph
membrane. form or be sent to a central monitoring unit, such as
the nurses’ station, where the data can be viewed
UF by pressure gradient is the primary method of and/or transferred to permanent medical records. It is
removing excess water from the blood through the also possible for home-dialysis patients to be moni-
semipermeable membrane. It occurs when water, a tored at a remote location. However, remote adjust-
small molecule, is forced across the membrane by ments to treatment are not yet possible.
hydrostatic pressure — dialysis is maintained at a
negative pressure for UF (the primary UF mechanism
in hemodialysis). Fluid removal is measured by the UF Reported problems
removal rate, which is automatically controlled in
Infections are a leading cause of morbidity and
newer units.
mortality in chronic hemodialysis patients. For exam-
High-flux dialysis, which typically shortens treat- ple, HBsAg (an indicator for the presence of hepatitis
ment time, has resulted from research with larger, B virus) has been detected on various surfaces in
high-efficiency dialyzers whose membranes’ increased hemodialysis centers, including dialysis-machine con-
surface areas and permeability permit higher fluid- trol knobs. Strict, specific policies and procedures de-
removal rates than are possible with standard dialyz- signed to reduce infection risks should be
ers. Development of thin-wall, hollow-fiber mem- implemented. These policies should address such is-
branes has enhanced UF and clearance rates. sues as sterilization and disinfection, housekeeping,
Hemodialysis machines that are to be used with high- laundry, maintenance, waste disposal, isolation pre-
flux dialyzers must be capable of operating with trans- cautions, and universal precautions.
membrane pressures (TMPs) approaching zero while
maintaining adequate control of UF. However, the use Water purification
of high-flux dialysis is still controversial: while the Adequate water purification is essential in
modality may permit faster treatment because of the hemodialysis. The quality and composition of tap
very high UF rates, the adequacy of metabolic by-prod- water varies from location to location; water that is
uct clearance has not been well established. There has considered safe for drinking is often still dangerous for
also been concern about the ability of bacteria or endo- hemodialysis. A long-term dialysis patient is exposed
toxins to penetrate the membranes used in high-flux to 450 liters of water per week, which is almost 50
dialysis. Some hospitals use high-flux dialysis to treat times the amount ingested by healthy people; as a
drug overdoses and chemical poisonings because of its result, trace amounts of elements in tap water may
ability to filter the blood very rapidly. approach toxic levels in dialysis patients. Chlorine
Another type of dialysis is continuous renal replace- and chloramines, which are added to tap water to
ment therapy (CRRT). This type of therapy allows inhibit bacterial growth, have been known to cause
tighter control of volume transfer and more even re- hemolysis. Aluminum from tap water has been cited
moval of wastes than intermittent treatment. CRRT as a possible cause of illnesses such as dialysis
is especially useful in patients who cannot tolerate the encephalopathy, bone disease, and anemia; iron can
rapid volume loss associated with intermittent ther- move across the dialyzer membrane and cause exces-
apy; in unstable patients, such as those in the intensive sive iron storage in the liver; copper can cause anemia
care unit; and in patients undergoing cardiac surgery. and metabolic acidosis; and excess lead can result in
neurologic damage. Standards have been established
Microprocessors for the quality of water used in hemodialysis systems,
but there is still some question regarding what consti-
Microprocessors in some units control alarms, sen- tutes a harmful level of impurity. Researchers are
sors, and operating functions; allow for data storage; discovering that many components currently being
and interface with clinical databases. Data down- removed by the purification system and by the dialyzer
loaded into a clinical database can be used to analyze membrane, such as zinc, are important to body meta-
information concerning a particular patient or ma- bolism. Vitamin therapy is often prescribed for dialysis
chine, or it can be used to analyze treatments involving patients.
a number of patients and machines.
Water used for dialysis should be tested peri-
Microprocessors also allow large amounts of infor- odically. The monitoring frequency depends on the
mation to be stored and analyzed during treatment. water treatment used. Centers using RO or DI devices
This includes calculating flow rate, duration of the should perform chemical tests at least annually and
session, total blood processed, and blood pressure. bacteriologic tests monthly. Guidelines for facilities
4 ©2003 ECRI. Duplication of this page by any means for any purpose is prohibited.
Hemodialysis Units
using other water-treatment methods recommend containing toxic levels of aluminum. An FDA/CDC
testing at least every three months. investigation indicated that the acidified portion of a
bicarbonate-based dialysate solution was stored or me-
Water-purification equipment malfunction, poor tered through a pump constructed of aluminum. Alu-
system maintenance, and sudden changes in feed- minum from the pump had leached into the dialysate
water quality are also potential problems in obtaining concentrate during transfer to the patient. FDA recom-
acceptable water quality for dialysis. These problems mends that whenever any component of an existing
can cause purification components to become ineffec- dialysis unit is changed or updated, the design and
tual, allowing feed-water contaminants to pass components used in the water treatment, concentrate
through the purification system. In one instance, a delivery, and dialysate delivery system be reassessed.
hemodialysis center allowed patients to receive treat- FDA also recommends routine monitoring of all dialy-
ment after the exchange resins used in the purification sis patients’ blood for serum aluminum levels.
system had been exhausted. As a result, high concen-
trations of fluoride were released into patients’ blood,
causing fluoride intoxication. Poor maintenance can Dialyzer reuse
also lead to bacteriologic infestations, causing patient
infection. Occasional instances of bacterial and hepatitis B
infection have been reported in hospital dialysis units.
In addition to potentially harmful impurities from While technical improvements in materials and disin-
the municipal water supply, there have also been re- fection procedures have improved the safety of dialyzer
ports of fluid contamination during disinfection of a reuse, this practice continues to be controversial. For
unit’s water-purification system. The Centers for Dis- example, if inadequately disinfected, tubing and dia-
ease Control and Prevention (CDC) has investigated lyzer membranes can harbor infection-causing organ-
several incidents in which hemodialysis patients were isms. With repeated use, there is a higher risk of
exposed to chloramines, sodium azide, or hydrogen membrane damage, allowing organisms to pass into
peroxide and has made specific recommendations that the bloodstream. Hazards exist with certain kinds of
may reduce chemical contamination of hemodialysis reprocessing systems: toxic reactions to formaldehyde
fluids. fumes have been reported in dialysis unit operators,
and trace amounts of disinfectant may remain in the
Dialysate dialyzer after reprocessing, exposing the patient to
potentially harmful levels of the agent during dialysis.
Operator error in preparing the dialysate is possi-
ble. While acetate was once the preferred ingredient Ideally, dialyzers labeled “single use” or “dispos-
for all dialysis patients, bicarbonate dialysis has be- able” should be discarded after their initial use; this
come a common alternative because of superior post- practice eliminates the danger of microbial and/or dis-
treatment comfort. Bicarbonate dialysis requires infectant contamination and ensures an efficient dialy-
mixing two concentrates, acid and bicarbonate, with sis treatment. However, financial considerations have
treated water, and the ratio of acid to bicarbonate is forced many hospitals and clinics to reuse many kinds
critical. Because conductivity reflects the total ionic of disposables, and the reuse of dialyzers is now a
content of the dialysate but does not measure its actual widespread practice. Additionally, reprocessed dialyz-
composition, fatal errors in pH balance can bypass ers may offer patients some benefits, such as reducing
alarms by satisfying conductivity requirements. Ex- the incidence of “first-use syndrome” — an allergic re-
treme care should be taken in mixing dialysate and in action to a new, unprocessed dialyzer, which is experi-
attaching dialysate containers to the proper inlet ports enced by some patients.
on the machine. As a precaution, dialysate should be
checked for both pH and conductivity before each di- Recognizing the prevalence of dialyzer reuse, in
alysis treatment. May 1995, FDA finalized regulation concerning the
Bicarbonate buffers are susceptible to contamina- labeling of dialyzers; it now requires manufacturers of
tion from bacteria that require the salt in bicarbonate reusable dialyzers to recommend at least one method
concentrate for growth. Techniques that can retard for reprocessing. Guidelines, such as those proposed by
bacterial growth in bicarbonate media should be fol- the Association for the Advancement of Medical In-
lowed when preparing dialysate concentrate. strumentation, should be closely followed to ensure
safe dialyzer reprocessing. In addition, dialysis unit
In May 1992, the U.S. Food and Drug Administra- operators and patients should be fully informed about
tion (FDA) alerted hemodialysis providers to a poten- the adverse health effects of exposure to particular
tially hazardous situation concerning dialysate disinfectants (e.g., formaldehyde).
©2003 ECRI. Duplication of this page by any means for any purpose is prohibited. 5
Healthcare Product Comparison System
Other problems as well as facilities considering units for home use, may
realize cost savings by purchasing earlier models.
Allergic or anaphylactic reactions in response to
device materials and the mode of dialyzer sterilization Machines acquired for home dialysis must be easy
have been reported; improving the biocompatibility of to operate, incorporating good human factors design.
components by the reduction of, for example, plasti- Because power, water pressure, and temperature may
cizer and trace-metal leaching is a major concern of not be as well regulated in homes as in hospitals and
researchers and manufacturers. In addition, ethylene most dialysis centers, factors such as limited accept-
oxide, the most common sterilant used in dialyzer able ranges of supply voltage and water temperature
manufacturing, has been shown to cause allergic reac- become issues in selecting a unit. Reliability is of
tions; in response, some manufacturers are using special concern for machines that are to be used in the
gamma radiation and steam sterilization. home because there is usually no backup unit or repair
technician available; if the unit ceases to function in
The rapid fluid removal caused by high-efficiency the middle of treatment, the consequences can be seri-
dialyzers can produce strong hypotensive reactions in ous. Home-dialysis training programs are available for
patients who suffer from severe cardiovascular disease patients and their relatives.
or who have retained more than 5 kg of fluid between
dialysis treatments (this retained fluid must be re- Most hemodialysis units accept various brands of
moved to maintain proper electrolyte concentrations, disposable accessories (e.g., tubing sets, bags). How-
pH, and blood pressure). The cardiovascular systems ever, some units require the use of proprietary dispos-
of these patients cannot compensate for the sudden ables, which can increase the operational costs per
and extreme loss of fluid volume from the blood; con- procedure. Facilities with multiple hemodialysis units
sequently, blood pressure falls sharply. should make sure that all units can accept the same
brands of disposables. Some suppliers will offer dis-
Dialysis disequilibrium syndrome is experienced by counts on disposables when they are purchased in bulk
the majority of people undergoing treatment on a quantities.
standard dialysis unit. Symptoms range from mild
attacks of malaise and drowsiness to convulsions, Disposables are a significant operating cost, and
coma, and death; the suspected cause is, again, the that cost can vary greatly depending on supplier dis-
inability of the vascular system to adjust to the change counts and incentives. ECRI’s PriceGuide™ service
in fluid volume during dialysis. benchmarks the price you pay for single-use medical
products. PriceGuide compares your hospital’s current
Problems can also result from hemodialysis cathe- pricing data with the national or regional average and
ters and blood lines. Complications with catheters lowest recorded prices paid. For more information,
have contributed to venous stenosis, venous and cathe- contact ECRI.
ter thrombosis, and fibrin sheath formation. Kinked
dialysis blood tubing has caused acute hemolysis.
Stage of development
Clinicians and researchers are seeking ways to im-
Purchase considerations prove the quality of life for patients with renal failure,
If only one or two units are being purchased to especially when transplantation is not a feasible alter-
supplement an existing stock, then staff familiarity native. One trend is toward prescription dialysis, in
and experience with a particular manufacturer’s ma- which the dialysate is fine-tuned to the individual
chines can be an important factor. In general, clini- needs of the patient. Sodium therapy and a renewed
cians will require less training on and will be more interest in bicarbonate dialysis are examples of this
comfortable with new models from a manufacturer trend. Manufacturers have responded by offering
whose equipment they are currently using. Clinical variable and programmable systems, which allow in-
engineers would also benefit from the experience dividualized patient therapy.
gained in servicing earlier models from the same
Although the procedure has drawbacks (see the
manufacturer. An additional advantage is that the
discussion in Reported Problems), the reuse of single-
components of the new machines may be identical to
use dialyzers for the same patient, which has been
those of earlier models, reducing the need to store
performed since maintenance dialysis began, is now
additional parts and thereby simplifying inventory.
widely practiced, as hospitals and treatment centers
Facilities that need units for limited chronic or acute have found it increasingly necessary to maximize use
care applications not requiring recent innovations of their equipment and resources. Because dialyzer
(e.g., variable sodium, regular or profiled UF control), reuse significantly lowers the cost per treatment, safe
6 ©2003 ECRI. Duplication of this page by any means for any purpose is prohibited.
Hemodialysis Units
and effective methods for recycling dialyzers are in “Burned out bulb” contributed to dialysis patient
great demand, and a number of companies have death. Biomed Saf Stand 1995 Mar 15;25(5):38.
emerged that develop and manufacture dialyzer re-
processing machines. Formaldehyde had been the CDC advises cohorting to prevent outbreaks: staff
most common disinfectant, but because of safety con- segregated also a key. Hosp Infect Control 1995 Jul;
cerns on the part of the Occupational Safety and 22(7):89-90.
Health Administration, it is now rarely used. Disinfec-
tants such as sodium hypochlorite and peracetic acid Collins AJ, Keshaviah PR. Are there limitations to
are currently used. Test kits that detect trace amounts shortening dialysis treatment? Trans Am Soc Artif
of disinfectant are available with reprocessors. The Intern Organs 1988 Jan-Mar;34(1):1-5.
future of home dialysis may include single-patient
hemodialysis machines that can incorporate dialyzer Cummings NB, Klahr S, eds. Chronic renal disease:
reprocessors. causes, complications and treatment. New York:
Plenum; 1985.
While medical research is attempting to define the
limits of the shortest, most efficient dialysis procedure, Dialysis centers hit by bloodborne outbreaks. Hosp
technological development is following a parallel Infect Control 1995 Jul;22(7):87-9.
course. Smaller, lighter, battery-powered units are
being designed that can dialyze almost as quickly as Feldman HI, Kinosian M, Bilker WB, et al. Effect of
the high-efficiency models. Almost completely auto- dialyzer reuse on survival of patients treated with
matic, these machines contain microprocessors that hemodialysis. JAMA 1996 Aug 28;276(8):620-5.
monitor and control all of the unit’s sensors, alarms,
and operating functions. One manufacturer has devel- Friedman EA, Lundin AP 3rd. Environmental and
oped a dialysis system that incorporates an intradia- iatrogenic obstacles to long life on hemodialysis
lytic blood-volume monitor for hemodynamic [editorial]. N Engl J Med 1982 Jan 21;306(3):167-9.
surveillance.
Ghezzi PM, Sanz-Moreno C, Gervasio R, et al. Techni-
Researchers are trying to produce a biocompatible,
cal requirements for rapid high-efficiency therapy
infection-proof synthetic material that can be molded
in uremic patients. Paired filtration-dialysis (PFD)
into tubing, implanted in large vessels, and attached
with a two-chamber technique. Trans Am Soc Artif
to the extracorporeal blood circuits for use as a perma-
Intern Organs 1987 Jul-Sep;33(3):546-50.
nent access channel to the blood. This artificial vessel
is necessary to extend the length of time that patients Hakim RM, Depner TA, Parker TF 3rd. Adequacy of
can receive dialysis treatment because veins and arter- hemodialysis. Am J Kidney Dis 1992 Aug;20(2):
ies become thrombosed after repeated damage to their 107-23.
walls and because blood flow is diminished; continued
puncturing of these vessels could cause their collapse Hornberger JC, Garber AM, Chernew ME. Is high-flux
or result in the formation of emboli. dialysis cost-effective? Int J Technol Assess Health
Advancement in membrane technology is also under Care 1993 Winter;9(1):85-96.
way. Researchers are trying to create more efficient
and more biocompatible membranes for use during Keshaviah P, Luehmann D, Ilstrup K, et al. Technical
dialysis. Research into new synthetic materials as requirements for rapid high-efficiency therapies.
well as bioartificial membranes is being conducted. Artif Organs 1986 Jun;10(3):189-94.
One technology involves the incorporation of cultured
renal tubular cells and glomerular cells layered onto a Meers C, Morton AR, Toffelmire EB. Dialysis access
synthetic membrane. morbidity with high-efficiency dialysis. Dial Trans-
plant 1993 Jun;22(6):324-5, 329, 352.
©2003 ECRI. Duplication of this page by any means for any purpose is prohibited. 7
Healthcare Product Comparison System
Nose Y. Why do we kill so many patients on hemodia- Wittich E. Reusing dialysers in the care of renal pa-
lysis in the U.S.? [editorial]. Artif Organs 1993 tients. Prof Nurs 1995 Jul;10(10):663-5.
Nov;17(11):893-4.
Standards and guidelines
Paolini F, Mancini E, Bosetto A, et al. Hemoscan™: a
dialysis machine-integrated blood volume monitor. Note: Although every effort is made to ensure that the
Int J Artif Organs 1995 Sep;18(9):487-94. following list is comprehensive, please note that other
applicable standards may exist.
Parker TF 3rd. Technical advances in hemodialysis Agency for Health Services and Technology Assess-
therapy. Semin Dial 2000 Nov-Dec;13(6):372-7. ment. Hemodialyzer reuse: considerations of safety
and costs [technology assessment report]. 11RE.
Petersen J, Hyver S, Yeh I. High flux and cellulosic
1991.
membranes. Biocompatibility and in vivo clear-
ances. Trans Am Soc Artif Intern Organs 1987 Jul- American National Standards Institute/Association
Sep;33(3):265-8. for the Advancement of Medical Instrumentation.
Hemodialyzer blood tubing [standard]. ANSI/AAMI
Pru CE, Cuervo C, Ardila M, et al. Hepatitis C trans- RD17-RC01. 1984 (revised 1995).
mission through dialysis machines. ASAIO J 1994 Reuse of hemodialyzers [recommended practice].
Jul-Sep;40(3):M889-91. 2nd ed. ANSI/AAMI RD47-1993. 1993.
Rudnick JR, Arduino MJ, Bland LA, et al. An outbreak Safe current limits for electromedical apparatus
of pyrogenic reactions in chronic hemodialysis pa- [standard]. 3rd ed. ANSI/AAMI ES1-1993. 1985 (re-
tients associated with hemodialyzer reuse. Artif Or- vised 1993).
gans 1995 Apr;19(4):289-94. Association for the Advancement of Medical Instru-
mentation. AAMI standards and recommended
Shaldon S, Koch KM. Biocompatibility in hemodia-
practices: dialysis — volume 3. 7th ed. DSBK4-005-
lysis: clinical relevance in 1995. Artif Organs 1995
MM. 2001.
May;19(5):395-7.
Reuse of hemodialyzer blood tubing [technology in-
Sweet SJ, McCarthy S, Steingart R, et al. Hemolytic formation report]. TIR6-152-EC. 1989 (revised
reactions mechanically induced by kinked hemodia- 1990).
lysis lines. Am J Kidney Dis 1996 Feb;27(2):262-6.
Canadian Society of Nephrology. Clinical practice
U.K. Medical Devices Agency. Haemodialysis equip- guidelines for the delivery of hemodialysis. J Am
ment: Braun Dialog. Evaluation 1996 Nov; No. 306. Soc Nephrol 1999 Jun;10(6):S287-321.
Canadian Standards Association. Extracorporeal cir-
U.K. Medical Devices Agency. Haemodialysis equip- cuits for use with hemodialysers, hemofilters, and
ment: Fresenius 4008E. Evaluation 1994 Dec; No. hemoconcentrators [standard]. 2nd ed. CAN/
223. CSA-Z364.1.1/Z364.1.2-94. 1984 (reaffirmed 2001).
U.K. Medical Devices Agency. Haemodialysis equip- Fluid supply and monitoring systems for hemodia-
ment: Hospal Integra. Evaluation 1996 Dec; No. lysis [standard]. 2nd ed. Z364.2.1-94. 1986 (reaf-
305. firmed 2001).
Water treatment equipment and water quality re-
U.K. Medical Devices Agency. Haemodialysis equip- quirements for hemodialysis [standard]. 2nd ed.
ment: Hospal Monitral SC 30 and BSM 32-SC. Z364.2.2-94. 1986 (reaffirmed 2001).
Evaluation 1994 Dec; No. 225.
Center for Devices and Radiological Health. Guidance
U.K. Medical Devices Agency. Haemodialysis equip- for hemodialyzer reuse labeling [guideline]. 1995.
ment: review issue. Evaluation 1992 Feb; No. 115. European Committee for Standardization. Hemodia-
lysers, hemodiafilters, hemoconcentrators and their
U.K. Medical Devices Agency. Haemodialysis equip-
extracorporeal circuits [draft standard]. prEN
ment: review issue. Evaluation 1994 Dec; No. 224.
1283:1996. 1996.
Vlchek DL, Burrows-Hudson S. Quality assurance International Electrotechnical Commission. Medical
guidelines for hemodialysis devices. Rockville (MD): electrical equipment — part 1: general require-
Department of Health and Human Services; 1991. ments for safety [standard]. IEC 60601-1-1 (1988-
FDA 91-4161. 12). 1988 (revised 2000).
8 ©2003 ECRI. Duplication of this page by any means for any purpose is prohibited.
Hemodialysis Units
Medical electrical equipment — part 1: general re- Importance of pretreating municipal water used for
quirements for safety. Amendment 1 [standard]. dialysis [User Experience Network™]. 1996 Feb-
IEC 60601-1-am1 (1991-11). 1991. Mar;25(2-3):110-1.
Medical electrical equipment — part 1: general re- Undetected venous line needle dislodgement during
quirements for safety. Amendment 2 [standard]. hemodialysis [hazard]. 1998 Nov;27(11):404.
IEC 60601-1-am2 (1995-03). 1995.
Health Devices Alerts
Medical electrical equipment — part 1-1: general
requirements for safety. Collateral standard: safety This Product Comparison lists Health Devices Alerts
requirements for medical electrical systems. 2nd ed. (HDA) citations published since the last update of this
IEC 60601-1-1 (2000-12). 1992 (revised 2000). report. Each HDA abstract is identified by an Acces-
Medical electrical equipment — part 1: general re- sion Number. Recalls and hazard reports include de-
quirements for safety. Section 1. Collateral stan- scriptions of the problem involved; abstracts of other
dard: safety requirements for medical electrical published articles are referenced by bibliographic in-
systems. Amendment 1. IEC 60601-1-1-am1 (1995- formation. HPCS subscribers can call the Hotline for
11). 1995. additional information on any of these citations or to
request more extensive searches of the HDA database.
Medical electrical equipment — part 1: general re-
quirements for safety. Section 2. Collateral stan- A4665 FDA has designated Class II Recall No. Z-
dard: electromagnetic compatibility — requirements 0362-02 certain Baxter Healthcare Model System 1000
and tests. IEC 60601-1-2 (2002-04). 2002. single-patient hemodialysis delivery systems. The
systems may experience a spontaneous dialysate-
Medical electrical equipment — part 2: particular proportioning-ratio change, which can result in severe
requirements for the safety of haemodialysis equip- patient acid-base disturbances. The distributor initiated
ment. IEC 60601-2-16 (1989-02). a recall by Urgent Product Recall letter dated November
International Organization for Standardization. Ex- 12, 2001. Verify that you have received the November 12,
tracorporeal blood circuit for haemodialysers, 2001, letter and recall information sheet from Baxter.
haemofilters and haemoconcentrators [standard]. Baxter is actively developing a software upgrade to
1st ed. ISO 8638:1989. 1989. detect a spontaneous dialysate-proportioning-ratio
Haemodialysers, haemofilters and haemoconcen- change. To receive the software upgrade, complete the
trators [standard]. 1st ed. ISO 8637:1989. 1989. recall information sheet and return it to Baxter by fax
at (727) 544-3025. For further information, contact
Office of Technology Assessment. Hemodialysis equip- Baxter Instrument Services by telephone at (800) 553-
ment and disposables industry [technology assess- 6898 Monday through Friday between 8:30 a.m. and
ment report]. NTIS #PB85-154953. 1984. 6:00 p.m. Eastern standard time within the U.S. or at
(847) 948-4770 outside the U.S. Source: FDA En-
Citations from other ECRI publications forcement Rep 2002 Mar 6; Distributor.
Health Devices A4731 FDA has designated Class II Recall No. Z-
Peritoneal dialysis compared with hemodialysis. 1986 1030-2 certain Gambro Phoenix hemodialysis systems.
Feb-Mar;15(2-3):34-50. The units may measure dialysate conductivity incor-
rectly because of an electronic component failure. The
Hemodialysis water purification [User Experience distributor initiated a recall by telephone call and fax
Network™]. 1988 Aug;17(8):247. dated April 9, 2002. Verify that you have received the
Cobe Centry 2 and Centry 2Rx hemodialysis units April 9, 2002, telephone call and fax from Gambro.
[hazard]. 1988 Oct;17(10):313-4. Identify and isolate any affected product in your inven-
Air embolism associated with hemodialysis [hazard]. tory. For further information, contact Gambro by tele-
1989 Nov;18(11):406. phone at (800) 525-2623. Source: FDA Enforcement
Rep 2002 May 15.
Hemodialysis machines [evaluation]. 1991 Jun;
20(6):187-232. A4974 Baxter Healthcare Corp has issued a Safety
Alert dated November 15, 2002, about certain Merid-
Hemodialysis machines [evaluation update]. 1991 Dec; ian hemodialysis instruments. The renalin/formalin
20(12):467. disinfection cycle of the units, which is used to prepare
Header cracks related to reprocessing of single-use the devices for long-term storage, may not run com-
hemodialysis dialyzers [hazard]. 1995 Aug-Sep; pletely because of electromagnetic interference from
24(8-9):359. the instrument screen when the Meridian enters the
©2003 ECRI. Duplication of this page by any means for any purpose is prohibited. 9
Healthcare Product Comparison System
sleep mode after being filled with renalin or formalin. Supplier information
Verify that you have received the November 15, 2002,
Safety Alert and confirmation form from Baxter. Identify
Baxter
any affected product in your inventory. Return the con-
firmation form to Baxter by fax at (847) 473-6961. For Baxter Healthcare Corp
further information, contact Baxter Instrument Services Renal [106389]
by telephone at (800) 553-6898, then select option 3, 1620 Waukegan Rd
option 2, and option 2 again, within the U.S., from McGaw Park IL 60085-6730
Monday through Friday, 8:30 a.m. to 6:00 p.m. Eastern Phone: (847) 473-6303, (800) 284-4060
standard time. International customers should contact Fax: (847) 473-6961
their local area Baxter support. Source: Manufac- E-mail: onebaxter@baxter.com
turer. Internet: http://www.kidneydirections.com
D5503 Great Britain. Medical Devices Agency. Dialy-
sis equipment. London: Department of Health; 2001 Bellco
Sep. 4 p. (Medical devices advice; no. AN2000[01])
Bellco SpA [331072]
D5517 FDA has designated Class II Recall No. Z- via Camurana 1
0456-2 certain Prisma dialysis system control units. I-41037 Mirandola MO
The control units’ display value of actual patient fluid Italy
removal can be incorrect. The distributor initiated a Phone: 39 (053) 529111
recall by letter dated October 2, 2001. The firm states Fax: 39 (053) 529407
that no affected product remains on the market. No E-mail: bellco@bellcospa.it
further action is required of customers. Source: FDA Internet: http://www.bellcospa.it
Enforcement Rep 2002 Jan 16; Distributor.
39880 Delarocque-Astagneau E, Baffoy N, Thiers V, B Braun
et al. Outbreak of hepatitis C virus infection in a
B Braun Medical Inc
hemodialysis unit: potential transmission by the
Renal Therapies Div [393207]
hemodialysis machine? Infect Control Hosp Epidemiol
824 12th Ave
2002 Jun;23(6):328-34.
PO Box 4027
Health Devices Inspection and Preventive Maintenance Bethlehem PA 18018-0027
System Phone: (610) 691-5400, (800) 848-2066
Fax: (610) 691-1547
Hemodialysis units. Procedure no. 413. E-mail: rtd@bbraun.com
Internet: http://www.bbraunusa.com
Health Technology Trends
B Braun Melsungen AG [178137]
Hemodialysis inadequate for many patients, say ex- Europagebaude Pfieffewiesen
perts. 1994 Feb;6(2):4-5. Postfach 1120
D-34299 Melsungen
Dialysis patient safety hinges on water treatment sys- Germany
tems [ECRI Forum]. 1994 Sep;6(5):8. Phone: 49 (5661) 710
Fax: 49 (5661) 714567
Healthcare Risk Control
E-mail: info@bbraun.com
Dialysis. 1996;1:Self-assessment questionnaires 10. Internet: http://www.bbraun.com
10 ©2003 ECRI. Duplication of this page by any means for any purpose is prohibited.
Hemodialysis Units
©2003 ECRI. Duplication of this page by any means for any purpose is prohibited. 11
Healthcare Product Comparison System
Conductivity range, mS/cm: Determines the electrolyte NEMKO — Norges Elektriske Materiellkontroll
concentration in dialysate by measuring its electro-
lytic conductivity in millisiemens (reciprocal mil- PC — Personal computer
liohms) per centimeter. SEMKO — Svenska Elektriske Materielkontrollan-
Rx dialysis: Many contemporary hemodialysis ma- stalten
chines enable operators to vary the final sodium and
bicarbonate contents somewhat independently, cus- SEV — Schweizerischer Electrotechnischer Verein
tomizing them for each patient. TCP/IP — Transmission Control Protocol/Internet
Bypass indicator: Signals the diversion of dialysate into Protocol
a drain, a flowmeter (e.g., to determine the UF rate),
etc., via a route that bypasses the typical dialysate TUV — Technischer Ueberwachungs Verein
delivery circuit. UF — Ultrafiltration
Blood circuit, arterial/venous pressure, mm Hg: Ranges
refer to the blood pressure in the extracorporeal UL — Underwriters Laboratories
blood circuit. VDE — Verband Deutscher Electrotechniker
Blood pump range, mL/min: Refers to the blood flow,
which is controlled by the blood pump. Note: The data in the charts derive from suppli-
ers’ specifications and have not been verified through
Abbreviations: independent testing by ECRI or any other agency.
BP — Blood pressure Because test methods vary, different products’ specifi-
cations are not always comparable. Moreover, products
CD-ROM — Compact disc — read-only memory and specifications are subject to frequent changes.
ECRI is not responsible for the quality or validity of
CE — Communaute Europeen
the information presented or for any adverse conse-
CE mark — Conformite Europeene mark quences of acting on such information.
CRT — Cathode ray tube When reading the charts, keep in mind that, unless
otherwise noted, the list price does not reflect supplier
CSA — Canadian Standards Association discounts. And although we try to indicate which
ETL — ETL Testing Laboratories features and characteristics are standard and which
are not, some may be optional, at additional cost.
FDA — U.S. Food and Drug Administration
For those models whose prices were supplied to us
GLEM — Groupement des Laboratoires d’Essais in currencies other than U.S. dollars, we have also
des Materiels de Technique Medicale listed the conversion to U.S. dollars to facilitate com-
parison among models. However, keep in mind that
IEC — International Electrotechnical Commission
exchange rates change often.
IU — International unit
Need to know more?
IV — Intravenous For further information about the contents of this
LCD — Liquid crystal display Product Comparison, contact the HPCS Hotline at +1
(610) 825-6000, ext. 5265; +1 (610) 834-1275 (fax); or
LED — Light-emitting diode hpcs@ecri.org (e-mail).
12 ©2003 ECRI. Duplication of this page by any means for any purpose is prohibited.
Hemodialysis Units
©2003 ECRI. Duplication of this page by any means for any purpose is prohibited. 13
Healthcare Product Comparison System
ALARM ACTIONS
Blood-leak detector Stops blood pump, Stops blood pump, Stops blood pump, Stops blood pump,
clamps venous line, clamps line, clamps line, UF clamps line, UF
dialyzer isolation UF turned off turned off turned off
Air/foam detector Stops blood pump, Stops blood pump, Stops blood pump, Stops blood pump,
clamps venous line, clamps line, bypass clamps line, UF clamps line, UF
dialyzer isolation turned off turned off
Conductivity Bypass Bypass Bypass Bypass
Arterial/venous
pressure Stops blood pump, Stops blood pump, Stops blood pump, Stops blood pump,
clamps venous line, bypass clamps line, UF clamps line, UF
dialyzer isolation turned off turned off
Temperature,
high/low Bypass, high limit Bypass Bypass Bypass
stops heater
H x W x D, cm (in) 153.4 x 47.6 x 77.5 132.6 x 55.9 x 38.1 152.4 x 43.2 x 40.6 152.4 x 43.2 x 40.6
(60.4 x 18.7 x 30.5) (52.2 x 22 x 15) (60 x 17 x 16); (60 x 17 x 16);
base 68.6 (27) deep base 68.6 (27) deep
FLOOR SPACE,
m2 (ft2) 0.37 (4.1) 0.33 (3.6) 0.3 (3.2) 0.3 (3.2)
WEIGHT, kg (lb) 105 (232) 104.8 (231) 80.6 (179) 80.6 (179)
PLANNING & PURCHASE
List price Not specified Not specified Not specified Not specified
14 ©2003 ECRI. Duplication of this page by any means for any purpose is prohibited.
Hemodialysis Units
©2003 ECRI. Duplication of this page by any means for any purpose is prohibited. 15
Healthcare Product Comparison System
MICROPROCESSOR TYPE Pentium II IBM Pentium II IBM Pentium user inter- 68302 Motorola
compatible *** compatible *** face
Storage Treatment, config, See Formula; uses Optional diskette Not specified
autotest, failure silicon disk on chip drive system †
Interface RS232 opto-isolated RS232 opto-isolated Optional RS232 Built-in RS232
ALARM ACTIONS
Blood-leak detector Dialyzer bypass, Dialyzer bypass, Stops blood pump, Stops blood pump,
dialysis time dialysis time clamps venous line, clamps line,
interrupted interrupted bypass UF turned off
Air/foam detector Dialyzer bypass, Dialyzer bypass, Stops blood pump, Stops blood pump,
dialysis time dialysis time clamps venous line, clamps line,
interrupted interrupted bypass UF turned off
Conductivity Dialyzer bypass, Dialyzer bypass, Bypass, stops blood Bypass
dialysis time dialysis time pump, clamps venous
interrupted interrupted line
Arterial/venous
pressure Dialyzer bypass, Dialyzer bypass, Stops blood pump, Stops blood pump,
dialysis time dialysis time clamps venous line, clamps line,
interrupted interrupted bypass UF turned off
Temperature,
high/low Stops heater, by- Stops heater, by- Stops blood pump, Bypass
pass, dialysis time pass, dialysis time clamps venous line
interrupted interrupted
H x W x D, cm (in) 174 x 50 x 75.5 174 x 50 x 75.5 5.1 x 6.3 x 14.2 114.3 x 50.8 x 50.8
(68.5 x 19.6 x 29.7) (68.5 x 19.6 x 29.7) (2 x 2.5 x 5.6) (45 x 20 x 20)
FLOOR SPACE,
m2 (ft2) 0.37 (4) 0.37 (4) 0.37 (4) 0.37 (4)
WEIGHT, kg (lb) 81 (178.5) 85 (187.3) 101 (223) 69.9 (154)
PLANNING & PURCHASE
List price Not specified Not specified $18,500 $16,140
16 ©2003 ECRI. Duplication of this page by any means for any purpose is prohibited.
Hemodialysis Units
WHERE MARKETED Canada, Mexico, USA Worldwide,except USA Worldwide,except USA Worldwide,except USA
FDA CLEARANCE Yes No No No
CE MARK (MDD) Not specified Yes Yes Yes
DIALYSATE DELIVERY
Proportioning
system Fixed ratio Volumetric, Volumetric, Volumetric,
Fresenius, Cobe, Fresenius, Cobe, Fresenius, Cobe,
other ratios other ratios other ratios
Comfort control,
°C 35-39 35-39 35-39 35-39
Temperature alarm
limits, °C ±2 of set 33.5-40 33.5-40 33.5-40
Conductivity range,
mS/cm 13-16 12.8-15.7 ** 12.8-15.7 ** 12.8-15.7 **
Flow, mL/min 100, 200, 300, 400, 0, 300, 500, 800 0, 300, 500, 800 0, 300, 500, 800
500, 600, 700, 800
Transmembrane
pressure, mm Hg -520 to +60 -60 to +520 -60 to +520 -60 to +520
Rx dialysis
Bicarbonate Yes Adjustable std range Adjustable std range Adjustable std range
24 to 40 mmol/L 24 to 40 mmol/L 24 to 40 mmol/L
Sodium therapy Yes Adj std range, 125- Adj std range, 125- Adj std range, 125-
150 mmol/L, profiled 150 mmol/L, profiled 150 mmol/L, profiled
Ultrafiltration
removal rate, L/hr 0-4, 8 UF profiles 0-4, profiled UF 0-4, profiled UF 0-4, profiled UF
pH monitor No Not needed Not needed Not needed
BYPASS INDICATOR Yes Shown by alarm Visual Visual
BLOOD CIRCUIT
Arterial pressure,
mm Hg -280 to +320 and -300 to +280 -300 to +280 -300 to +280
-80 to +500
Venous pressure,
mm Hg -80 to +500 -60 to +520 -60 to +520 -60 to +520
Blood pump range,
mL/min 0-600 15-600 10-600 10-600
Heparin pump range,
mL/hr 0.1-9.9 0-10 (20, 30, 50 mL 0-10 (20, 30, 50 mL 0-10 (20, 30, 50 mL
syringe) syringe) syringe)
DISINFECTION
Method Chemical and heat Chemical, thermo- Chemical, thermo- Chemical, thermo-
chemical, hot rinse chemical, hot rinse chemical, hot rinse
DISPLAY TYPE Color LCD LED LED, LCD,10.4" color LED, LCD,10.4" color
DISPLAYED PARAMETERS
Dialysate pressure Yes Not displayed Not displayed Not displayed
Transmembrane
pressure Yes Yes Yes Yes
Conductivity Yes Yes Yes Yes
Flow rate Yes No Yes Yes
Elapsed time Yes Yes Yes Yes
Remaining time Yes Yes Yes Yes
Prescribed time Yes No Yes Yes
Other Pie chart of time Standard: effective Standard: effective Standard: effective
elapsed blood flow and blood flow and blood flow and
dialysis time, dialysis time, dialysis time,
UF and Na profiles *** UF and Na profiles *** UF and Na profiles ***
©2003 ECRI. Duplication of this page by any means for any purpose is prohibited. 17
Healthcare Product Comparison System
Arterial/venous
pressure Stops blood pump, Stops blood pump, Stops blood pump, Stops blood pump,
clamps line, clamps line, UF clamps line, UF clamps line, UF
UF turned off turned off *** turned off *** turned off ***
Temperature,
high/low Bypass Bypass, visual and Bypass, visual and Bypass, visual and
audible alarm audible alarm audible alarm
H x W x D, cm (in) 114.3 x 50.8 x 50.8 133 x 49 x 34 (52.4 140 x 59 x 36 (55.1 133 x 49 x 34 (52.4
(45 x 20 x 20) x 19.3 x 13.4); x 23.2 x 14.2); x 19.3 x 13.4);
base 63 (24.8) deep base 62 (24.4) deep base 63 (24.8) deep
FLOOR SPACE,
m2 (ft2) 0.37 (4) 0.30 (3.2) 0.33 (3.6) 0.30 (3.2)
WEIGHT, kg (lb) 69.9 (154) 80 (176.4) 92 (202.9) 80 (176.4)
PLANNING & PURCHASE
List price $21,140 Not specified Not specified Not specified
18 ©2003 ECRI. Duplication of this page by any means for any purpose is prohibited.
Hemodialysis Units
Comfort control,
°C 35-39 30-40 30-38 34-40
Temperature alarm
limits, °C 28-41 28-41 29-34; 40.1 max 33-41
Conductivity range,
mS/cm Na 130-160 mmol/L ** Na 130-150 mmol/L ** 20-40 mEq/L (bicarb) *** 13-17
Flow, mL/min 300-700 300-700 300-600 : 300-800 350-1,000,
increments of 50
Transmembrane
pressure, mm Hg -100 to +500 -200 to +500 0 to +700 0 to +450
Rx dialysis
Bicarbonate Yes Yes Yes Yes
Ultrafiltration
removal rate, L/hr 0-4 0-4 0-2.4 0.1-4
pH monitor Optional Optional Yes Yes
BYPASS INDICATOR Yes Yes Yes Yes
BLOOD CIRCUIT
Arterial pressure,
mm Hg -500 to +50 -700 to +750 -400 to +52 -400 to +150
Venous pressure,
mm Hg -700 to +750 -700 to +750 -52 to +400 0 to +450
Blood pump range,
mL/min 0-500 0-500 50-500 10-500
Heparin pump range,
mL/hr 0-10 0-10 0-5.0 0.5-9.9
DISINFECTION
Method Chemical, citric, Chemical, citric, Chemical Chemical, citric,
heat heat heat
DISPLAY TYPE LED, LCD LED, LCD CRT screen LCD
DISPLAYED PARAMETERS
Dialysate pressure No No Yes Yes
Transmembrane
pressure Yes Yes Yes No
Conductivity Yes Yes Yes Yes
Flow rate No Yes Yes Yes
Elapsed time Yes Yes Yes Yes
Remaining time Yes Yes Yes Yes
Prescribed time Yes Yes Yes Yes
Other None specified None specified None specified None specified
©2003 ECRI. Duplication of this page by any means for any purpose is prohibited. 19
Healthcare Product Comparison System
ALARM ACTIONS
Blood-leak detector See footnote * See footnote * Stops blood pump, Stops blood pump,
clamps line, reduces clamps line, reduces
UF rate to minimum UF rate to minimum
Air/foam detector Stops blood pump, Stops blood pump, Stops blood pump, Stops blood pump,
arterial & venous arterial & venous clamps line, UF rate clamps line, UF rate
clamps shut clamps shut to zero to zero
Conductivity See footnote ** See footnote ** Bypass Bypass
Arterial/venous
pressure Stops blood pump, Stops blood pump, Stops blood pump, Stops blood pump,
clamps shut clamps shut clamps venous line, clamps venous line,
UF rate to zero UF rate to zero
Temperature,
high/low Fluid bypass Fluid bypass Stop heater, bypass; Bypass
heater shuts down at
43.5°C
H x W x D, cm (in) 120 x 57 x 63 130 x 57 x 75 136 x 57 x 75 144 x 61 x 71
(47.2 x 22.4 x 24.8) (51.2 x 22.4 x 29.5) (53.5 x 22.4 x 29.5) (56.6 x 24 x 27.9)
FLOOR SPACE,
m2 (ft2) 0.36 (3.9) 0.43 (4.6) 0.43 (4.6) 0.43 (4.6)
WEIGHT, kg (lb) 60 (132.3) 80-87 (176.4-191.8) *** 102 (225) wet 120 (264)
PLANNING & PURCHASE
List price £12,950 £16,000 : 14,900 $18,200-21,350 : £21,500
(US$20,676) (US$25,553 : 23,796) $18,650-21,800 (US$19,622)
Training On-site, free, On-site, free, On-site On-site
unlimited unlimited
Warranty 1 year 1 year 1 year or 2,000 hr 6 months
Delivery time, ARO 3-4 weeks 3-4 weeks Not specified Not specified
Year first sold 1997 1998 : 1997 1987 : 1998 2001
Fiscal year January to December January to December January to December January to December
OTHER SPECIFICATIONS None specified. None specified. Power-failure alarm; Power-failure alarm;
digital control; BP monitor;
BP monitoring integrated ionic
displayed onscreen; K+/V clearance
optional software monitoring
for dialysis and displayed onscreen;
therapy management; compensated blood
single needle. flow displayed on
Meets requirements machine; bicarb
of ETL and UL. cartridge; optional
software for
dialysis and therapy
management; single
needle.
20 ©2003 ECRI. Duplication of this page by any means for any purpose is prohibited.
Hemodialysis Units
Comfort control,
°C NA 34-39.5 34-39.5 34-39.5
Temperature alarm
limits, °C NA 32, 41 32, 41 32, 41
Conductivity range,
mS/cm NA 13-17 13-17 13-17
Flow, mL/min NA 350-1,000,increments 350, 500, 750, 1,000 350, 500, 750, 1,000
of 50
Transmembrane
pressure, mm Hg NA 0 to +450 0 to +450 0 to +450
Rx dialysis
Bicarbonate Yes Yes Yes Yes
Ultrafiltration
removal rate, L/hr NA 0.1-4 0.1-3 0.1-3
pH monitor NA Yes Yes Yes
BYPASS INDICATOR NA Yes Yes Yes
BLOOD CIRCUIT
Arterial pressure,
mm Hg -250 to +50 -350 to +150 -350 to +150 -350 to +150
Venous pressure,
mm Hg -50 to +350, access 10-450 10-450 10-450
Blood pump range,
mL/min 10-180, return 10-500 50-700 50-700
Heparin pump range,
mL/hr 0.5-5.0, anticoag- 0.5-9.9 0.5-9.9 0.5-9.9
ulant
DISINFECTION
Method NA See footnote † See footnote † See footnote †
©2003 ECRI. Duplication of this page by any means for any purpose is prohibited. 21
Healthcare Product Comparison System
MICROPROCESSOR TYPE Not specified Not specified Not specified Not specified
Storage Not specified Flash Prom plus Floppy disk Floppy disk
CD-ROM
Interface Touch screen Serial comm inter- Serial communication Serial communication
face and Ethernet interface (CCM) *** interface (CCM) ***
LOSS-OF-WATER ALARM NA Yes Yes Yes
ALARM ACTIONS
Blood-leak detector Yes Stops blood pump, Stops blood pump, Stops blood pump,
clamps line, UF min clamps line, UF min clamps line, UF min
Air/foam detector Yes Stops blood pump, Stops blood pump, Stops blood pump,
clamps line clamps line clamps line
Arterial/venous
pressure Yes, access/return ART-P/VEN-P stops ART-P/VEN-P stops ART-P/VEN-P stops
blood pump, VEN-P blood pump, VEN-P blood pump, VEN-P
clamps venous line clamps venous line clamps venous line
Temperature,
high/low NA Stops heater, Stops heater, Not specified
bypass/heater on, bypass/heater on,
bypass bypass
H x W x D, cm (in) 147.3 x 191 x 191 140 x 60 x 65 140 x 60 x 65 140 x 60 x 65
(58 x 75 x 75) (55 x 23.6 x 25.6) (55 x 23.6 x 25.6) (55 x 23.6 x 25.6)
FLOOR SPACE,
m2 (ft2) 1.9 (6.3) 0.39 (4.2) 0.39 (4.2) 0.39 (4.2)
WEIGHT, kg (lb) 22.7 (50) 108 (237) 108 (237) 108 (237)
PLANNING & PURCHASE
List price $25,000 Not specified Not specified Not specified
22 ©2003 ECRI. Duplication of this page by any means for any purpose is prohibited.
Hemodialysis Units
Transmembrane
pressure, mm Hg -100 to +500 -100 to +500 Not specified
Rx dialysis
Bicarbonate Yes Yes Not specified
Ultrafiltration
removal rate, L/hr 0.00; 0.10 to 4.00 0-4 0-3
pH monitor No No No
BYPASS INDICATOR Visual Not specified Yes
BLOOD CIRCUIT
Arterial pressure,
mm Hg -300 to +300 No -300 to +300
Venous pressure,
mm Hg -200 to +500 -200 to +400 -400 to +400
Blood pump range,
mL/min 40-600 40-500 0-600
Heparin pump range,
mL/hr 0.0-10.0 0-9.9 0-9.9
(20, 30 mL syringe)
DISINFECTION
Method Chemical, thermo- Sodium hypochloride Chemical, heat
chemical, hot rinse concentrate
DISPLAY TYPE LCD 15.1", XGA LCD screen, LED seg LED
DISPLAYED PARAMETERS
Dialysate pressure Yes Yes Yes
Transmembrane
pressure Yes Yes Yes
Conductivity Yes Yes Yes
Flow rate Yes Yes Yes
Elapsed time Yes Yes Yes
Remaining time Yes No Yes
Prescribed time Yes Yes Yes
Other None specified None specified None specified
©2003 ECRI. Duplication of this page by any means for any purpose is prohibited. 23
Healthcare Product Comparison System
Arterial/venous
pressure Stops blood pump, Stops blood pump Stops blood pump,
clamps venous line clamps line, UF
turned off, bypass
Temperature,
high/low Bypass, stops Bypass, stops Stops blood pump,
heater, high-temp- heater, high-temp- clamps line, UF
erature alarm erature alarm turned off, bypass
H x W x D, cm (in) 146 x 39 x 48 122 x 30 x 37 141 x 38 x 36
(57.5 x 15.4 x 18.9) (48 x 11.8 x 14.6) (55.5 x 15.1 x 14.2)
FLOOR SPACE,
m2 (ft2) 0.39 (1.3) 0.14 (1.5) 0.14 (1.5)
WEIGHT, kg (lb) 85 (187.4) 65 (143.3) 105 (231.5)
PLANNING & PURCHASE
List price Not specified Not specified Not specified
24 ©2003 ECRI. Duplication of this page by any means for any purpose is prohibited.