Download as pdf or txt
Download as pdf or txt
You are on page 1of 11

REVIEW

Formulation Development and Pharmacokinetic Laboratory, Pharmacy Group, Birla Institute of Technology and Science, Pilani,
India

Comparison of various international guidelines for analytical


method validation

S. Chandran, R. S. P. Singh

Received November 9, 2005, accepted June 16, 2006


Asst. Prof. Dr. Sajeev Chandran, Formulation Development and Pharmacokinetic Laboratory, Pharmacy
Group, Birla Institute of Technology and Science, Pilani-333 031, Rajasthan, India
sajeev@bits-pilani.ac.in
Pharmazie 62: 4–14 (2007) doi: 10.1691/ph.2007.1.5064

Analytical method validation is the systematic process of establishing that an analytical method is
acceptable for its intended purpose. In general the developer or user of the method generates evi-
dence on specificity, linearity range, accuracy, precision, detection limit, quantitation limit, ruggedness
and robustness of the method for regulatory submissions or in-house application. The iterative pro-
cess of method development and validation has a direct impact on the quality of the above data. Such
validated analytical methods for qualitative or quantitative testing of drug molecules assume greater
importance when they are employed to generate quality and safety compliance data during develop-
ment and post-approval of drug products. The present paper aims to discuss salient points of the
analytical method development and validation cycle. It also attempts to compare and summarize
guidelines issued by different agencies for validation of analytical methods used for analysis of drug
substances in the pure form and in pharmaceutical formulations.

1. Introduction cal method validation can be considered as the process of


defining the analytical requirements, and confirming that
Analytical method development and validation procedures the method under consideration has performance capabil-
are vital in the discovery and development of drugs and ities consistent with what the application requires. The
pharmaceuticals. Analytical methods are used to aid in the method’s performance has to be validated and the uncer-
process of drug synthesis, screen potential drug candi- tainty of the result estimated, at a given level of confi-
dates, support formulation studies, monitor the stability of dence. It is not sufficient just to determine uncertainty, but
bulk pharmaceuticals and formulated products, and test it also should be quoted in a way that is widely recog-
final products for release. The quality of analytical data is nized, internally consistent and easy to interpret. In gener-
a key factor in the success of a drug and formulation de- al, methods for regulatory submission must include studies
velopment program. During the post approval commercial on specificity, linearity, accuracy, precision, range, detec-
production stage of bulk drugs and pharmaceutical pro- tion limit, quantitation limit, and robustness which ensure
ducts, the official or in-house test methods that have re- that the analytical methodology in question gives timely,
sulted from the analytical method development and valida- accurate, reproducible and reliable data which are ade-
tion process cycle become indispensable for reliable quate for the intended purpose of use. Validation of an
monitoring of the integrity, purity, quality, strength and analytical method cannot eliminate all the problems likely
potency of the manufactured products. There is often a to arise during implementation of the methodology but it
need to transfer methodology from one laboratory to an- does ensure that major problems are prospectively seen
other and/or to include it in official compendia. Such ex- and a mechanism to control the variability is suggested
ercises include the use of a method by large numbers of (WHO report 1992).
people, in various laboratories across the globe and on in- Method validation procedures and acceptance criteria were
struments manufactured by different manufacturers, there- for a long time matter of personal prudence until various
by causing a greater probability of decreased reproducibil- industrial committees and regulatory agencies developed
ity and reliability. These problems can be foreseen and framework guidelines for performing such validations for
avoided by thorough validation of the analytical method methods applicable to drugs and pharmaceuticals (Green
(Brown et al. 2001). 1996). The submission of analytical method validation
For an analytical result to be fit for its intended purpose it data has been made mandatory for successful submission
must be sufficiently reliable that any decision based on it of New Drug Applications (NDA) and Abbreviated New
can be taken with confidence. In the light of this, analyti- Drug Applications (ANDA). This requirement has in-

4 Pharmazie 62 (2007) 1
REVIEW

creased the importance given to the validation methodol- toxicology and drug metabolism studies, purification and
ogy employed for analytical procedures (Analytical proce- identification of the parent compound and its metabolites
dures and method validation 2000). are needed. At the formulation stage the primary concern
The EN 45000 series of standards (EN 45001 : 1989) and is the stability of the drug substance. Dissolution studies
ISO/IEC Guide 25 have proposed general requirements for measure the kinetics of the appearance of the drug sub-
the competence of calibration and testing laboratories and stance as a result of derivatization, adsorption and physi-
general criteria for the operation of testing laboratories. cal encapsulation of the drug substance by the formulation
The formal recognition that a testing laboratory is compe- polymers, alteration of the conformational state of protein
tent to carry out specific tests or specific types of tests and peptide based pharmaceuticals, aggregation and hy-
under the above standards is based on nine validation drolysis of the drug molecule. The challenge here is to
parameters. Harmonized guidelines between the European understand how these factors influence the analysis, and
Union (EU), Japan and the United States have been devel- finding solutions that help in obtaining the required infor-
oped within the expert working group (Quality) of the In- mation reliably in the face of these complexities. Simi-
ternational Conference on Harmonization (ICH) of techni- larly, evaluation of the final drug product requires determi-
cal requirements for registration of pharmaceuticals for nation of physical characteristics, drug content uniformity
human use (ICH Q2A and Q2B). This document defines and release kinetics.
eight validation parameters (Q2A) and discusses their de- The quality of an analytical method developed is always
tailed methodology (Q2B). The United States Environ- appraised in terms of suitability for its intended purpose,
mental Protection Agency (USEPA) has prepared guidance recovery, requirement for standardization, sensitivity, ana-
for method development and validation for the Resource lyte stability, ease of analysis, skill subset required, time
Conservation and Recovery Act. The United States Phar- and cost in that order. It is highly imperative to establish
macoepia (USP 2003) has published specific guidelines through a systematic process that the analytical method
for method validation for evaluation of compounds. The under question is acceptable for its intended purpose. Re-
American Association of Official Analytical Chemists covery refers to the ability of the method to give a re-
(AOAC), USEPA, and other scientific organizations pro- sponse for the entire amount of analyte in the sample. It
vide multi-laboratory validated methods and have devel- is generally expressed as the percentage of reference/sam-
oped their own peer-verified method validation programs ple material that has been added to the blank. Its impor-
with detailed guidelines on parameters to be validated tance increases in the absence of a reference material.
whereas the International Union of Pure and Applied Che- Standardization is the cardinal point in method develop-
mists (IUPAC) has developed guidelines for single labora- ment. External standardization is quite common, in which
tory method validation (Green 1996). The United States the response of the analyte alone is plotted versus concen-
Food and Drug Administration (FDA) is the first regula- tration, whereas internal standardization is used when re-
tory agency that has understood the need for guidance for producibility is the problem. In internal standardization, a
analytical method validation involving biological fluids functional or isotopic analogue of the analyte that is simi-
and accordingly issued guidelines in 2001 because of their lar in physicochemical properties is added to the stan-
importance in bioavailability, bioequivalence and pharma- dards and sample prior to treatments. The ratio of re-
cokinetic studies. sponses of the standard and internal standard is plotted
Although there is general agreement on the various valida- against the concentration of the standard to obtain a cali-
tion parameters to be evaluated, diversity prevails in the bration curve (Karnes et al. 1991).
methodology employed for validation and acceptance cri- Sensitivity of the method is defined as the increase in re-
teria. The chasm between the EU, the US and Japan has sponse with unit increase in concentration. The sensitivity
been bridged by ICH but it still exists with other gui- of the method can be increased by manipulating various
dance. This review article discusses the various validation factors affecting it, like signal to noise ratio, physicochem-
parameters and compares guidelines issued by various reg- ical properties of the analyte, and the response of the in-
ulatory bodies. put transducer to analyte and composition of sample ma-
trix (Pasteelnick 1993). It is important to determine the
stability of the analyte in its sample environment in order
2. Analytical method development to access the degradative effect of sample components on
Pharmaceutical and biopharmaceutical product develop- analyte response. Ease of analysis refers to simplicity in
ment is an inherently complex process. Each stage of the sample preparation, time required for analysis and the
drug development process uses a series of analytical meth- chances of error involved in analysis. A minimal skill sub-
ods, which are developed for the specific needs of that set required in personnel and a cost effective method are
particular development stage. Since the common element always desired (Willard et al. 1995). There can be no de-
in each consecutive stage is the drug, the use of pre- marcation between analytical method validation and devel-
viously developed analytical methods is encouraged. This opment. This is an iterative process in which development
method transfer concept has two major objectives: (a) sav- is followed by validation, further changes and revalidation.
ing method development resources; and (b) providing ex-
perimental data that are comparable with previous results.
3. Analytical method validation procedure
In most cases the same analytical methods are not applic-
able throughout the whole drug development pathway. The steps involved in development, validation and deter-
Each drug development stage has a different objective, a mination of validation parameters, also termed analytical
different sample environment, different sensitivity require- performance characteristics, depend upon the type and nat-
ments, and a different impurity profile. The sample envir- ure of the analytical method. Pharmaceutical analytical
onment and/or sample matrix can affect sample prepara- methods are categorized into five general types, namely,
tion steps and degradation of the drug substance. For identification tests, potency assays, impurity tests (quanti-
example, at the early research stage purification and iden- tative), impurity tests (limits) and specific tests. The first
tification of the drug substance are the primary goals. For four tests are universal tests, but the specific tests such as

Pharmazie 62 (2007) 1 5
REVIEW

particle-size analysis and X-ray diffraction studies are The validated test method is included in the validation re-
used to determine specific properties of the active pharma- port that summarizes the results of the validation studies.
ceutical ingredient (API) or the drug product (Pasteelnick Both the validation report and test method are submitted
1993). as parts of the NDA or ANDA. The analytical method’s
A method has to be validated when it is necessary to ver- performance characteristics should be based on the in-
ify whether its performance parameters are adequate for tended use of the method and the prudent judgment exer-
use for a particular analytical problem. For example, (a) cised by the analyst. For example, if the method is to be
when a new method is developed for a specific problem; used for qualitative trace level analysis, there is no need to
(b) when indications exist that an established method is test and validate the method’s linearity over the full dy-
changing with time; (c) when an established method is namic range of the equipment. Initial parameters should be
revised to incorporate changes/improvements or to extend chosen according to the analyst’s best judgment. Finally,
it for another purpose; (d) when an established method is parameters should be agreed between the laboratory gener-
used in a different laboratory, or with different analysts or ating the data and the client using the data. Although vali-
different instrumentation; (e) to demonstrate the equiva- dation parameters have been suggested by various regula-
lence between two methods, e.g. a new method and a tory agencies, the sequence of validation is still a matter
standard. The extent of validation or revalidation required of personal preference based on past experience and the
would depend on the nature of the changes made in reap- method itself. Many authors have suggested determining
plying a method to different laboratories, instrumentation selectivity/specificity first, followed by accuracy, precision,
or operators, and the circumstances in which the method linearity and range, LOD, LOQ, robustness and rugged-
is going to be used. Some degree of validation is always ness (Green 1996). Detailed guidelines on analytical meth-
appropriate even when using apparently well-characterized od validation issued by various international agencies are
standard or published methods. summarized in Tables 3 to 8. These tables include defini-
A well-developed method should be easy to validate. As tion of validation parameters, method of determination, re-
the development of the method and the validation process commendations, method of expressing the parameter and
advance, the information gathered is captured in the de- the calculations thereof and the acceptance criteria.
sign and subsequent improvement of the method. Ideally,
the validation protocol should be written only following a
thorough understanding of the method’s capabilities and 4. Validation parameters
intended use. The validation protocol will list the accep-
4.1. Specificity
tance criteria that the method can meet. Any failure to
meet the criteria will require that a formal investigation is Specificity is the ability of the method to measure the ana-
conducted. Various steps involved in a complete method lyte response accurately in the presence of all potential
validation program are summarized in Table 1 and brief sample components, referred to as the sample matrix.
definitions of various validation performance characteris- These sample components or matrix may include placebo
tics are presented in Table 2. formulation, synthesis intermediates, excipients, degrada-

Table 1: Brief description of various activities to be performed during method validation


Step No. Activity

1 Define method development plan and gather background information


2 Develop laboratory method and generate test procedure
3 Develop validation protocol or operating procedure for the validation
4 Define application, purpose and scope of method
5 Define performance parameters and acceptance criteria
6 Define validation experiments
7 Verify relevant performance characteristics of equipment
8 Qualify materials, e.g. standards and reagents
9 Perform pre-validation experiments
10 Adjust method parameters or/and acceptance criteria, if necessary
11 Perform full internal (and external) validation experiments
12 Develop standard operating procedures for executing method routinely
13 Define criteria for revalidation
14 Define type and frequency of system suitability tests and/or analytical quality control (AQC) checks for routine use
15 Document validation experiments and approval of validation report

Table 2: Brief definition of various validation performance characteristics


Performance Characteristics Activity

Specificity Ability to measure desired analyte in a complex mixture


Accuracy Agreement between measured and real value
Precision Agreement between a series of measurements
Linearity Proportionality of measured value to concentration
Range Concentration interval where method is precise, accurate, and linear
Detection limit Lowest amount of analyte that can be detected
Quantitation limit Lowest amount of analyte that can be measured or quantified
Robustness Ability to remain unaffected by small changes in parameters
Ruggedness Reproducibility under normal but variable laboratory conditions

6 Pharmazie 62 (2007) 1
REVIEW

Table 3: Comparison of different guidelines for ‘specificity’ parameter of analytical method validation

Guidelines ICH US FDA AOAC USP IUPAC

Reference Guideline for industry- Guidance for Industry- AOAC Peer Verified General Chapter Harmonized guidelines
document text on validation of Bio analytical method Methods Program, h1225i, Validation of for single laboratory
analytical procedure validation (May 2001) Manual on policies compendial methods, validation of methods
(ICH Q2A) (Mar 1995) and procedures United States Pharma- of analysis (Nov 1999)
Guidance for Industry- (Nov 1993) copeia XXV, Rockville,
Q2B validation of MD, The United States
analytical procedures: Pharmacopeial Conven-
Methodology (ICH tion Inc. p. 2149–2152
Q2B) (Nov 1996) (2003)
Specificity
Definition Ability to assess un- Ability of an analytical Ability of a method to Ability to assess unequi- The degree to which a
equivocally the analyte method to differentiate measure only what it is vocally the analyte in method can quantify
in the presence of com- and quantify the analyte intended to measure. the presence of compo- the analyte accurately
ponents, which may be in the presence of other nents, which may be in the presence of
expected to be present. components in the expected to be present interferants.
sample.
Method  When the impurities  Analysis of blank  Reagent blanks and  When the impurities  Selectivity index
are available: Spiking samples of appropriate field blanks should are available: Spiking should be calclated.
of pure substance biological matrix be run to ensure no of pure substance Selectivity index
(drug or drug product) obtained from at least interfering compounds (drug or drug product) ¼ ban=bint
in appropriate level of six sources and test are present. in appropriate level of ban ¼ Slope of calibra-
impurities/excipients for interference.  To verify the specifi- impurities/excipients tion curve
and demonstrate the  Selectivity should city of the method for and demonstrate the bint ¼ Slope of
result is unaffected. be ensured at LLOQ the analyte(s) of inter- result is unaffected. response indepen-
 When the impurities LLOQ- Lower limit of est, results should be  When the impurities dently produced by a
are not available: quantitation tested under different are not available: potential interferant.
Comparing the test experimental condi- Comparing the test
results of sample con- tions, e.g., two differ- results of sample
taining impurities or ent analytical princi- containing impuri-
degradation product to ples or two different ties or degradation
second well-character- detection techniques. product to second
ized procedure. These  The method should be well-characterized
comparisons should able to distinguish the procedure. These
include sample under analyte from known comparisons should
relevant stress condi- interfering materials include sample under
tion. and the behavior of relevant stress condi-
 In chromatographic the analyte during tion.
method: Peak purity analysis should be  In chromatographic
test to be done by indistinguishable from method: Peak purity
diode array and mass the corresponding test to be done by
spectrophotometry. standard material in diode array and mass
the appropriate matrix. spectrophotometry.
Expression/  Proof of discrimina- Evidence that the sub- ––  Proof of discrimina-  Selectivity Index
calculation tion of analyte in the stance being quantified tion of analyte in the
presence of impurities is the intended analyte. presence of impurities,
e.g. for chromato- e.g., for chromato-
graphy chromatogram graphy chromatogram
should be submitted. should be submitted.
 Peak purity test helps  Peak purity test to de-
in demonstrating that monstrate that the peak
the peak is not attribu- is not attributable to
table to more than one more than one com-
component. ponent.
 For assay two results  For assay two results
should be compared should be compared
and for impurity tests and for impurity tests
two profiles should be two profiles should be
compared. compared.
Acceptance Not specified Not specified Not specified Not specified Not specified
criteria

tion products, process impurities, etc. The response of the to generate degradants that can be potentially generated
analyte in test mixtures containing the analyte and all po- during the normal course of analysis if the sample is sub-
tential sample matrixes is compared with the response of a jected to harsh environmental conditions. The stress condi-
solution containing only the analyte. Any contribution of tion selected should be sufficient to cause substantial de-
the matrix to the response leads to constant or proportional gradation (approximately 70–90% of its labeled purity).
systematic error and such methods are referred as non-spe- For bulk pharmaceuticals typical stress conditions em-
cific. The analyte is exposed to stress conditions sufficient ployed include heat (50  C), light (600 foot candle), acid

Pharmazie 62 (2007) 1 7
REVIEW

(0.1 N HCl), base (0.1 N NaOH), and oxidant (3% H2O2). Four approaches have been suggested to determine the ac-
For formulated products conditions such as heat (40  C), curacy of analytical methods (Green 1996). The accuracy
light and humidity (60–75% RH) are often used. Methods can be determined using a single certified reference mate-
of determining specificity and its expression as prescribed rial and comparing the measured value with true value. A
by various international agencies are presented in Table 3. second approach compares the result of the proposed
A statistical approach that tests the intercept against zero method with the result of a reference/another method
using a one-sided t-test has been reported (Shah 1991; whose accuracy and precision is known. These two ap-
Bolton 1997). In the absence of primary standards the ma- proaches are futile if the certified reference material or
trix effect is tested by comparing the slopes of matrix and method is not available. In such a situation a recovery
non-matrix standards or by a standard addition method study is performed in which the analyte is spiked by
(Shah 1991). weight or volume into the matrix covering the entire line-
arity range followed by quantitation where the procedure
follows the final proposed sample preparation and re-
4.2. Linearity and range sponse measurement technique. The result then is ex-
The linearity of an analytical method refers to the ability pressed as percent recovery. The fourth approach is the
to elicit test results that are, either directly or by well-de- technique of standard additions, which can also be used to
fined mathematical transformations, proportional to the determine recovery of spiked analyte. This approach is
concentration of analyte in the sample over the entire used if it is not possible to prepare a blank sample matrix
range of interest. This is determined by measuring the re- without the presence of the analyte, for example, with lyo-
sponse of standard solutions in the range of 50 to 150% philized material, in which the speciation in the lyophi-
of target concentration. An unweighted linear least square lized material is significantly different when the analyte is
line is plotted, which assumes that all the errors occur in absent. Since the accuracy is often expressed as percen-
the Y-direction, i.e. the error in measuring response is more tage bias it may be helpful to determine whether the bias
than the error in preparation of a sample concentration, is because of random error alone. This is done using the
and that all errors are normally distributed. Though the t-test to determine whether the mean value differs signifi-
linearity characteristic of any proposed method is expected cantly from the true value. If the deviation is significant,
to include the origin of the response and the concentration then the ratio of the deviation between the mean and meas-
axes, a significantly different intercept may be acceptable ured results to the mean result is calculated as the esti-
if the accuracy of the method is not compromised (Miller mate of bias (Green 1996; Miller 1991).
et al. 1988; Miller 1991). Homocedasticity is assessed by
observing the plot of residuals versus concentration. If an
4.4. Precision
increase or decrease in variance, better known as heteroce-
dasticity, is observed weighted regression is preferred The precision of the method is defined as the degree of
(Miller et al. 1988; Miller 1991). Acceptability of linearity scatter of individual test results of multiple measurements
data is often judged by examining the correlation coeffi- of a homogenous sample. A comparison of various guide-
cient and y-intercept of the linear regression line for the lines regarding determination of precision is presented in
response versus concentration plot. A correlation coeffi- Table 6. There are four types of precision that can be de-
cient greater than 0.9999 is generally considered as evi- termined for an analytical method, namely instrument pre-
dence of acceptable fit of the data to the regression line. cision or injection repeatability, repeatability or intra-assay
The y-intercept should be less than a few percent of the precision, intermediate precision and reproducibility.
response obtained for the analyte at the target level. Though all official guidelines describe only the first three
The range of an analytical method is the interval between types of precision, there are some papers that describe all
the upper and lower levels (including these levels) over four types of precision (Green 1996).
which acceptable accuracy, linearity, and precision are ob- Instrument precision is determined by repeated measure-
tained. In practice, the range is determined using data from ment of one sample solution so as to test the performance
the linearity and accuracy studies. To avoid the possibility of the instrument used in the analytical methodology. The
of misinterpretation, two approaches have been suggested. repeatability or intra-assay precision is obtained by repeat-
In the first approach, deviation from the regression line is edly analyzing independently prepared homogenous sam-
plotted against concentration or log concentration. Devia- ples in one laboratory, by one operator, using one piece of
tion should be equally distributed between positive and ne- equipment and one set of reagents on one day. At least
gative values. Another approach suggests plotting the ratio five determinations of three concentrations at the low,
of response to concentration versus concentration or log medium and high range of calibration are performed and
concentration. The line obtained should be linear over the the % relative standard deviation (RSD) is calculated.
full range. Some analytical procedures may require non- Precision of less than 1% RSD is easily achieved in com-
linear calibration but a linear model with univariant regres- pound analysis in pharmaceutical quality control but preci-
sion is preferred (Shah 1991; Miller 1991). The range is sion decreases with the complexity of the matrix espe-
normally expressed in the same units as the test results cially with biological matrices (Green 1996).
(e.g. amount per unit volume, parts per million or percen- Intermediate precision is the precision obtained when ana-
tage) obtained by the analytical method. A comparison of lysis involves multiple analysts, multiple equipment, multi-
various guidelines for the linearity and range parameter of ple days, and multiple sets of reagents in the same labora-
analytical method validation is summarized in Table 4. tory. The objective of the determination of this precision
is to identify the various factors within a single laboratory
that will contribute to the variability of the results and to
4.3. Accuracy
find a mechanism to control them (Green 1996).
The accuracy of an analytical method is the closeness of Validation of reproducibility is important if the method is
the result obtained to the true value. Various guidelines going to be used in different laboratories. When a method
regarding determination of accuracy are listed in Table 5. is transferred from one laboratory to another it invariably

8 Pharmazie 62 (2007) 1
REVIEW

Table 4: Comparison of different guidelines for ‘linearity and range’ parameter of analytical method validation

Guidelines ICH US FDA AOAC USP IUPAC

Linearity and range


Definition Linearity: Ability Calibration/Standard Linearity –– Defines Linearity: Ability Not explicitly defined
(within range) to obtain Curve: Relationship the ability of the meth- (within range) to obtain
test results, which are between response and od to obtain test results test results, which are
directly proportional to known concentration of proportional to the con- directly proportional to
the concentration analyte. centration. the concentration
(amount) of analyte in (amount) of analyte in
the sample. Range: the sample. Range: In-
Interval between the terval between the upper
upper and lower con- and lower concentration
centration (amounts) of (amounts) of analyte in
analyte in the sample the sample including the
including the concen- concentrations for which
trations for which suit- suitable level of accu-
able level of accuracy, racy, precision, and line-
precision, and linearity arity has been demon-
has been demonstrated. strated.
Method Drug (different dilu- Sufficient number of Not specified Drug (different dilution)  Visual examination
tion) and/or separately standards in the ex- and/or separately of residual plot
weighed synthetic mix- pected matrix is taken weighed synthetic mix-  Application of sta-
ture. Measurement of and response measured. ture. tistical testing (sig-
response and plot re- Plot of concentration vs. Measurement of re- nificance testing)
sponse vs. concentra- response is plotted and sponse and plot re-  Test for lack of fit
tion of analyte and de- linearity is demon- sponse vs. concentration can be weighted/
monstration of linearity strated. Number of of analyte and demon- simple regression
by standards depends on: stration of linearity by
 Visual inspection of  Anticipated range of  Visual inspection of
plot analytical value plot
 Appropriate statisti-  Nature of analyte/  Appropriate statistical
cal methods response relationship methods
Conc. of sample de-
pends on expected con-
centration range
Recommen-  Minimum of 5 con-  One blank sample 4 conc. levels are to be  Minimum of 5 con-  Six or more calibra-
dation centrations are re- (matrix sample with- selected. centrations are recom- tion standards
|fflfflfflfflfflfflfflffl{zfflfflfflfflfflfflfflffl}

commended for line- out internal standard  1/2 mended for linearity evenly spaced over
arity (IS)  1 3 Days For Range: the range of interest.
For Range:  One zero sample  3/2  Assay of drug/fin-  Range: 0–150% or
 Assay of drug/fin- (matrix sample þ IS)  2 ished product: 80– 50–150% of target
ished product: 80–  6–8 non-zero sam- 120% of test concen- concentration de-
120% of test con- ples covering ex- tration. pending on which
centration. pected range includ-  For content unifor- of this is more suit-
 For content unifor- ing LLOQ. mity: 70–130% of able
mity: 70–130% of test concentration  Calibration standard
test concentration  For dissolution test- should be run at
 For dissolution test- ing: 20% over least in duplicate.
ing: 20% over specified range Preferably in tripli-
specified range  For impurity: from re cate or more
 For impurity: from porting level to 120%
reporting level to of specification.
120% of specifica-
tion.
Expression/  Correlation coeffi- Not specified Not specified  Correlation coeffi- Not specified
calculation cient, y-intercept, cient, y-intercept,
slope of regression slope of regression
line, residual sum of line, residual sum of
squares. squares.
Acceptance Not specified At LLOQ: Not more Not specified Not specified Not specified
criteria than 20% deviation
from nominal value
Other than LLOQ: 15%
deviation from nominal
value
At least 4 out of 6 non-
zero sample should
meet the above criteria
including LLOQ and
highest concentration.

Pharmazie 62 (2007) 1 9
REVIEW

Table 5: Comparison of different guidelines for ‘accuracy’ parameter of analytical method validation

Guidelines ICH US FDA AOAC USP IUPAC

Accuracy
Definition Expresses the closeness Closeness of mean test Closeness of the deter- Closeness of test results Trueness is the close-
of agreement between results obtained by mined value to the true obtained by that method ness of agreement be-
the value, which is ac- method to the true value value. to true value. tween a test result and
cepted either as a con- (concentration of ana- accepted reference
ventional true value or lyte). value of the property
the value found. (Also being measured. Smal-
referred as trueness). ler bias means greater
trueness
Method  Application of pro-  Replicate analysis of  Use of certified re-  Application of proce- Bias is determined by
cedure to analyze samples containing ference materials dure to analyze syn- comparing response of
synthetic mixture of known amounts of  Use of reference thetic mixture of method to a reference
known purity. analyte. method of known known purity. like,.
 Comparison of result uncertainty  Comparison of result  Use of certified
with already estab-  Use of recovery with already estab- reference material.
lished procedure. from spiked samples lished procedure.  Use of reference
 Accuracy may be  Accuracy may be in- material
inferred once preci- ferred once precision,  Use of reference
sion, linearity and linearity and specifi- methods
specificity have been city have been estab-  Use of spiking/
established. lished. recovery
Significance testing is
recommended.
Recommen- Minimum of nine Three concentration Three concentration Minimum of nine deter- Not specified
dation determinations levels covering entire levels throughout the minations
 Low concentration range range  Low concentration of
of range  3 repli-  Low concentration of  Low concentration range  3 replicates
cates range  5 replicates of range  Medium concentra-
 Medium concentra-  Medium concentra-  Medium concentra- tion of range  3
tion of range  3 tion of range  5 re- tion of range replicates
replicates plicates  High concentration  High concentration of
 High concentration  High concentration of of range range  3 replicates
of range  3 repli- range  5 replicates  Amount added for
cates (According to CDER re- fortification should
view for HPLC method: be a substantial frac-
80, 100, 120% of target tion or more than
concentration  tripli- the amount present
cate) in the sample.
Expression/  Percent recovery by Mean –– True value % Recovery ¼ (mean  Percent recovery by Not specified
calculation the assay of known concentration fortified the assay of known
added amount of –– mean concentration added amount of ana-
analyte unfortified)/Known in- lyte
 Mean – Accepted crement in concentra-  Mean –– Accepted
true value with tion true value with confi-
confidence interval dence interval
Acceptance Not specified  At LLOQ: Mean Not specified Not specified Not specified
criteria value should be with-
in 20% of actual
value
 Other than LLOQ:
Mean value should be
within 15% of
actual value.

encounters: analysts with different experience and thor- lence the results from other laboratories are compared
oughness, differences in room temperature and humidity, against the primary laboratory, whereas in analytical
equipment with different characteristics, and also varia- equivalence a range of acceptable results is chosen prior
tions in the nature and quality of supplies, materials, con- to study (USFDA 2001).
sumables, and instrument conditions (e.g. in HPLC mobile
phase composition, pH, flow rate of the mobile phase,
4.5. Limit of detection
column specifications). Reproducibility is determined by
measuring a homogenous sample in multiple laboratories Limit of detection (LOD) is the lowest concentration of
with the object of verifying that the same results are ob- analyte that can be reliably detected using the method but
tained when the methodology is transferred to other la- not necessarily quantified. The LOD of a method should
boratories. Statistical equivalence and analytical equiva- be established quite early in the method development-vali-
lence are used to judge the acceptability of results dation process and its determination should be repeated
obtained from different laboratories. In statistical equiva- using the specific wording of the final procedure. In the

10 Pharmazie 62 (2007) 1
REVIEW

Table 6: Comparison of different guidelines for ‘precision’ parameter of analytical method validation

Guidelines ICH US FDA AOAC USP IUPAC

Precision
Definition The precision of analy- Describes the closeness Degree of agreement of Degree of agreement Closeness of agreement
tical procedures expres- of individual measures measurements under among individual test between test results ob-
ses the closeness of of an analyte when the specific conditions. results when the meth- tained under stipulated
agreement (degree of procedure is applied od is applied repeat- conditions.
scatter) between series repeatedly to multiple edly to multiple sam-
of measurements ob- aliquots of single plings of a
tained from multiple homogenous volume of homogenous sample.
sampling of the same biological matrix
homogenous sample
under the prescribed
conditions.
Method Determination of % Analytical procedure is Determination of % Determination of % Determination of %
relative standard devia- applied repeatedly to relative standard devia- relative standard devia- relative standard devia-
tion (RSD) of response multiple aliquots of tion (RSD) of multiple tion (RSD) of response tion (RSD) or % coeffi-
of multiple aliquots single homogenous aliquots of multiple aliquots cient of variation (CV)
volume of biological and F-test for normally
matrix. distributed error is ap-
plied.
Recommen- Repeatability (Same Three concentration 4 conc. levels are to be Repeatability (Same Minimum of two con-
dation operating condition levels covering entire selected. operating condition centrations
over short interval of range  1/2 over short interval of  At or near highest
time): Minimum of  Low concentration  1 time): concentration in range
nine determinations of range  5 repli-  3/2 Minimum of nine de-  At or near lowest con-
over the entire range cates  2 terminations over the centration in range
 Low concentration  Medium concentra- Repeatability: to be entire range
of range  3 repli- tion of range  established within  Low concentration
cates 5 replicates laboratory of range  3 repli-
 Medium concentra-  High concentration  Different days cates
tion of range  of range  5 repli-  Different analysts  Medium concentra-
3 replicates cates  Different calibration tion of range 
 High concentration Intra batch precision curves 3 replicates
of range  3 repli- (repeatability):  Different batches of  High concentration
cates Measures precision reagents of range  3 repli-
(or) during single run  Different matrices. cates
 At target concentra- Inter batch precision Reproducibility: (or)
tion  6 determina- (reproducibility): Between labs.  At target concentra-
tions Measures precision tion  6 determina-
Intermediate precision with time. tions
(within laboratory var- Intermediate precision
iation): (within laboratory var-
 Different Days iation):
 Different Analysts  Different Days
 Different Equipment  Different Analysts
etc.  Different Equipment
etc.
Expression/ Standard deviation, Coefficient of variation RSD Standard deviation, SD
calculation RSD and confidence RSD and confidence {(stot ¼ sr2/n þ srun2)1/2
interval interval where n ¼ no. of repeat
results} or RSD
Acceptance Not specified  At LLOQ: Not specified Not specified Not specified
criteria CV 20%
 Other than LLOQ:
15%

case of methods requiring technology transfer it is impor- photometric methods LOD is determined using the rela-
tant to test the LOD of the method on different instru- tion 3.3s/S where s is the standard deviation of the re-
ments of similar models to those used in other labora- sponse and S is the slope of the calibration curve. The
tories. Depending on the nature of the method, various standard deviation of the response can be obtained either
approaches have been suggested to determine the detec- by measuring the standard deviation of the blank response
tion limit. or by calculating the residual standard deviation of the
Simple visual examination may be adequate for a non-in- regression line or by calculating the standard deviation of
strumental method. In the case of chromatographic meth- the y-intercept of the regression line or Sy/x, i.e. the stand-
ods that exhibit constant background noise, it can be esti- ard error of the estimate (ICH Q2A, ICH Q2B, 1996). A
mated based on signal-to-noise ratio. The LOD in such comparison of various guidelines for the limit of detection
cases will correspond to the concentration at which the parameter in analytical method validation is summarized
response signal-to-noise ratio is 3. In the case of spectro- in Table 7.

Pharmazie 62 (2007) 1 11
REVIEW

Table 7: Comparison of different guidelines for ‘detection limit’ parameter of analytical method validation

Guidelines ICH US FDA AOAC USP IUPAC

Detection Limit
Definition Lowest amount of analyte Explicitly not Lowest content that can Lowest amount of ana- Smallest amount of conc.
in the sample, which can described be measured with reason- lyte in the sample, which of analyte in the sample
be detected but not neces- able statistical certainty can be detected but not that can be reliably dis-
sarily quantitated under necessarily quantitated tinguished from zero.
stated experimental condi- under stated experimental
tions. conditions.
Method 1. By visual evaluation Not described Based on more than 20 For non-instrumental: Not specified
2. Based on S/N ratio blank readings. Analysis of sample with
 Applicable to proce- known concentration of
dure, which exhibit analyte and by establish-
baseline noise. ing minimum concentra-
 Actual lowest concen- tion at which analyte can
tration of analyte be reliably detected.
detected in compared For instrumental: Process
with blank response for non-instrumental can
3. Based on S.D. of be adopted. Detection
response and slope limit should be suffi-
LOD ¼ 3.3s/s ciently low for analysis
s ¼ Slope of calibration of samples with known
curve concentration of analyte
s ¼ S.D. of response; can above and below the
be obtained by required detection limit.
 Standard deviation of
blank response
 Residual standard
deviation of the
regression line
 Standard deviation of
the y-intercept of the
regression line
 Sy/x i.e. standard
error of estimate
Expression/  If based on visual Not described The mean value of the Not specified Not specified
calculation examination or S/N matrix blank readings
ratio –– relevant chro- (n  20) plus three
matogram is to be standard deviations of
presented. the mean, expressed in
 If by calculation/extra- analyte concentration.
polation –– estimate is
validated by analysis of
suitable no. of samples
known to be near or
prepared at detection
limit.
Acceptance S/N ratio > 2–3; Not Not described Not specified Not specified Not specified
criteria specified in other cases

4.6. Limit of quantitation using the relationship LOQ ¼ 10s/S where s is the stan-
dard deviation of the response and S is the slope of the
The limit of quantitation (LOQ) is the concentration at and
calibration curve. s is obtained from the standard deviation
above which the analyte can be reliably quantitated with a
of response of blanks or by residual standard deviation or
previously defined level of certainty. A summary of var-
regression line or the standard deviation of the intercept.
ious guidelines for determination of the limit of detection
parameter of analytical method validation is presented in
Table 8. The LOQ is determined by reducing the analyte
concentration until a level is reached where the precision 4.7. Robustness
of the method is unacceptable. If the required precision at Robustness is the ability of the method to remain unaf-
the LOQ is already specified, then the %RSD of six deter- fected by small changes in method parameters carried out
minations at decreasing concentration is plotted against deliberately or otherwise during the validation/usage of
concentration. The quantitation limit is determined by ex- analytical methodology. These method parameters may be
trapolating from the graph (ICH Q2A, ICH Q2B, 1996). evaluated one factor at a time or simultaneously as part of
Like LOD, the quantitation limit can be determined by dif- a factorial experiment. Obtaining data on the effects of
ferent approaches. In case of non-instrumental techniques these parameters may allow a range of acceptable values
even a visual examination may suffice. If the method exhi- to be included in the final method procedure. The factors
bits background noise this can be determined based on a that are generally studied are the pH of the mobile phase
signal-to-noise ratio of 10. This can also be estimated or solvent, buffer concentration, small changes in solvent

12 Pharmazie 62 (2007) 1
REVIEW

Table 8: Comparison of different guidelines for ‘quantitation limit’ parameter of analytical method validation

Guidelines ICH US FDA AOAC USP IUPAC

Quantitation Limit
Definition Lowest amount of ana- The lowest amount of The limit of quantita- Lowest amount of ana- Not defined
lyte in a sample, which analyte that can be tion is the lowest lyte in a sample, which
can be quantitatively quantitatively deter- amount of analyte in a can be quantitatively de-
determined with suita- mined with suitable pre- sample, which can be termined with suitable
ble precision and accu- cision and accuracy also quantitatively deter- precision and accuracy.
racy. called LLOQ (Lower mined with precision
limit of quantification). and accuracy appropri-
ate to analyte and ma-
trix considered.
Method 1. By visual evaluation Preparation of standard Not specified 1. By visual evaluation Not recommended;
2. Based on S/N ratio curve and lowest conc. 2. Based on S/N ratio only recommends
 Applicable to proce- on the calibration curve  Applicable to proce- expressing uncertainty
dure, which exhibits should be accepted as dure, which exhibits of measurement as
base line noise. LLOQ if it satisfies fol- base line noise. function of concentra-
 Low conc. of analyte lowing condition.  Low conc. of analyte tion.
is compared with  Response at LLOQ ¼ is compared with
blank 5 Response by blank
3. Based on S.D. of blank 3. Based on S.D. of re-
response and slope  Analyte peak should sponse and slope
LOQ ¼ 10s/s be identifiable dis- LOQ ¼ 10s/s
s ¼ Slope of calibration crete and reproducible s ¼ Slope of calibration
curve with precision of curve
s ¼ S.D. of response; 20% and accuracy of s ¼ S.D. of response;
can be obtained by 80–120% can be obtained by
 Standard deviation  Standard deviation of
of blank response blank response
 Residual standard  Residual standard
deviation of the deviation of the
regression line regression line
 Standard deviation  Standard deviation of
of the y-intercept of the y-intercept of the
the regression line regression line
Sy/x i.e. standard error Sy/x i.e. standard error
of estimate of estimate
Recommen- Limit should be vali- Not specified Not specified Not specified Not specified
dation dated by analysis of
suitable no. of samples
known to be near or
prepared at quantitation
limit.
Expression/  Limits of quantita- Not specified Mean value of the ma- Expressed as analyte Not specified
calculation tion and method trix blank reading plus concentration
used for determining 10 standard deviations (% or ppb)
should be presented. of the mean, expressed
 Expressed as analyte in analyte concentra-
concentration tion.
Acceptance Not specified Not specified Not specified Not specified Not specified
criteria

system, temperature, and injection volume etc. (ICH Q2B 4.9. System suitability
1996; USP 1995).
The system has to be tested for its suitability for the in-
tended purpose. During the early stages of the method
4.8. Ruggedness development process some of the more sophisticated sys-
Ruggedness is not defined by ICH guidelines. It is defined tem suitability tests may not be practical due to the lack
by the USP as the degree of reproducibility of test results of experience with the method. In the early stages of de-
obtained by analysis of the same samples under a variety velopment it may be useful to perform some additional
of conditions. It involves analysis of aliquots of homoge- system suitability tests to evaluate system performance un-
nous lots in different laboratories by different analysts un- der different method conditions. This information will
der different operational and environmental conditions. help to develop an appropriate system suitability test strat-
The degree of reproducibility of test results is determined egy in the future. As more experience is acquired for this
as a function of the assay variables. IUPAC considers the method, a more sophisticated system suitability test may
effect of change of instrument, operator, brand of reagent, be necessary. For this, all critical factors that will signifi-
concentration of reagent, pH of solution and time allowed cantly affect the method performance need to be identified
(run time) for completion of the process (AOAC 1993; and they should remain within the specified limits or the
Thompson et al. 2002). critical factors should be manipulated to change the sys-

Pharmazie 62 (2007) 1 13
REVIEW

tem performance favorably. Therefore, the system suita- 5. Conclusion


bility strategy not only consists of the tests and limits but
The efficient development and validation of analytical
also the approach used to optimize the system perfor-
methods are critical elements in the development of phar-
mance when the original test result exceeds the limit. In
maceuticals and ensuring regulatory compliance. Analy-
addition, if the method demands high method sensitivity, a
tical method validation is an important tool for ensuring
detector sensitivity solution may be required to demon-
the performance of the method. Various guidelines by dif-
strate suitable signal to noise ratio from the system exhi-
ferent regulatory bodies and organizations disagree on dif-
biting baseline noise (USFDA 1993).
ferent points. Though ICH guidelines have resolved the
Numerous approaches may be used to set the limits for
differences between Europe, the USA and Japan, organiza-
system suitability tests. This depends on experience with
tions like IUPAC and AOAC still have differences on some
the method, material available and personal preference.
points. There should be an effort to put forward uniform
Default values for system suitability for HPLC methods
guidelines for validation throughout the world and to cre-
are: capacity factor should be more than 2, injection preci-
ate a similar platform for acceptance criteria.
sion should have %RSD less than 1%, resolution factor
should be greater than 2, tailing factor should be less than References
2% and theoretical plate should be more than two thou-
Analytical Procedures and Method Validation (2000), Fed Reg 52: 776–
sand (USFDA 1993). 777.
AOAC Peer Verified Methods Program (1993), Manual on policies and
procedures, Arlington, VA.
4.10. Stability Bolton S (1997) Pharmaceutical Statistics: Practical and Clinical Applica-
tion, 3rd edition, Marcel Dekker, New York, pp. 153, 216–269.
During the earlier validation studies, the method developer Brown R, Caphart M, Faustino R, Frankewich E, Gibbs R, Lentzinger E,
gained some information on the stability of reagents, mo- Lunn G, Rajagopalan R, Chiu Y, Sheinen E (2001) Analytical Proce-
bile phases, standards, and sample solutions. For routine dures and Method Validation: Highlights of FDA’s draft guidance,
testing in which many samples are prepared and analyzed LCGC (www.chromatography online.com), 19: 74.
General Chapter No. 1225, Validation of Compendial Methods, United
each day, it is often essential that solutions are stable en- States Pharmacopoeia XXIII, National Formulary, XVIII (1995) Rock-
ough to allow for delays such as instrument breakdowns or ville, MD, The United States Pharmacopeial Convention, Inc, 1710.
overnight analyses using auto-samplers. Stability has not Green JM (1996) A practical guide to analytical method validation. Anal
been given due importance in ICH guidelines but the US Chem 305A309A.
International Conference on Harmonization (ICH) of Technical Require-
FDA has discussed stability parameters for biosamples. It ments for the Registration of Pharmaceuticals for Human use, Validation
is important to determine the stability of an analyte in a of Analytical Procedures, ICH Q2A (1996), Geneva.
particular matrix by comparison with freshly prepared stand- International Conference on Harmonization (ICH) of Technical Require-
ards. It recommends freeze thaw stability for three freeze ments for the Registration of Pharmaceuticals for Human use, Validation
of Analytical Procedures: Methodology, ICH Q2B (1996) Geneva.
thaw cycles. It also recommends short-term temperature Karnes T, Shiu G, Shah VP (1991) Bioanalytical method validation. Pharm
stability, which is done at room temperature for 4–24 h Res 8: 421–426.
based on the expected duration for which samples will be Miller JC, Miller JN (1988) Basic statistics methods for analytical chemis-
maintained at room temperature for the intended study. try part 1-Statistics of repeated measures Analyst 113: 1351–1356.
Miller JN (1991) Basic statistical methods for analytical chemistry, part 2-A
Long-term stability should also be assessed over a time Calibration and regression methods Analyst 116: 3–14.
period greater than the time difference between the date of Shah VP, Midha KK, Dighe S, McGilveray IJ, Skelly JP, Yacobi A, Layl-
first sample collection and the date of last sample analy- off T, Viswanathan CT, Cook CE, McDowall RD, Pittman KA, Spector S
sis. The concentration of all the stability samples should (1991) Analytical methods validation: Bioavailability, bioequivalence and
pharmacokinetic studies. Conference report. Eur J Drug Metab Pharma-
be compared to the mean of back-calculated values for the cokinet 16: 249–255.
standards at the appropriate concentration from the first Thompson M, Ellison SLR, Wood R (2002) Guidelines for Single Labora-
day of long term stability testing. tory Validation of Methods of Analysis. Pure Appl Chem 74: 835–855.
The stability of stock solutions of drug and the internal US FDA Guidance of Industry-Bioanalytical Method Validation (2001),
# 5600, Fishers Lane, Rockville, MD 20857.
standards should be evaluated at room temperature for at US FDA Technical Review Guide: Validation of Chromatographic Methods
least 6 h and if the stock solution is stored then the rele- (1993) Center for Drug Evaluation and Research (CDER), Rockville,
vant stability should be checked. The stability of pro- MD.
cessed samples, including their residence time in an auto- United States Pharmacopoeia (2003) United States Pharmacopeial Conven-
tion, 26th Edition, Rockville, MD.
sampler known as postoperative stability, is to be deter- WHO Expert Committee report on specifications for pharmaceutical pre-
mined (USFDA 2001). parations (1992) 32nd Report. 32: 117–121.
At this point, the limits of stability should be tested. Sam- Willard HH, Meritt LL, Dean JA, Settle FA (1995) Instrumental Method of
ples and standards should be tested over at least a 48 h Analysis, CBS Publishers, New Delhi, 7th Edn., 2–17.
Pasteelnick LA (1993) Analytical methods validation. In: Nash RA, Berry
period, and the quantitation of components should be de- IR (eds.) Pharmaceutical process validation, 2nd ed., Marcel Dekker Inc.,
termined. If the solutions are not stable over 48 h, storage New York, p. 411–428.
conditions or additives should be identified that can im-
prove stability.

14 Pharmazie 62 (2007) 1

You might also like