Professional Documents
Culture Documents
R2
R2
R2
The objectives of this studies are to assess the quality of surface waters. Besides, to determine
the amount of dissolved oxygen in waste water samples. The studies are to establish the
concentration of organic matter in waste water samples. Lastly, to understand the
characteristics of DO contained in water and wastewater.
1
Table of Contents
Abstract ...................................................................................................................................... 1
Introduction ................................................................................................................................ 3
Discussion .................................................................................................................................. 8
Conclusion ................................................................................................................................. 9
2
Introduction
There are about 40 million organic compounds known in the environment which cannot
be defined individually with considerable analytical effort and in short time. Therefore, the so-
called sum parameters are used. These parameters reflect effect and material characteristics of
one or more substances. The most popular sum parameter in waste water analysis are the BOD
(biochemical oxygen demand), COD (chemical oxygen demand), TOD (total oxygen demand)
and TOC (total organic carbon). The TOC reflects the organic pollution on the basis of a direct
carbon determination. The other parameters are based on oxygen, which is required to reduce
or to oxidise the samples' substances. (Thomas, 1998)
3
process of sewage treatment is not considered: The nitrification. Obviously, a WWTP can only
be controlled and monitored reliably by total BOD measurements, instead of the determination
of the insufficient BOD5. In the real sense, BOD measurements are respiration measurements.
Due to their rapidity, respiration measurements are preferred for online analysis. Provided that
the conditions are known, respiration measurements [mg/(l*min)] can be converted into BOD
measurements [mg/l]. Due to the 5-day analysis time and the measurement of the carbonaceous
BOD instead of total BOD, the BOD5 is not suitable to assess the current capacity of the waste
waters' degradation by bacteria. The BOD5 is only a time-delayed information about the
pollution of waste water and cannot be used for optimization or control of a WWTP.
Alternatively, BOD analysers or respiration analysers can be used, which enable measurements
within 5 to 60 minutes. Thus, estimations can be made promptly for the biodegradability of the
waste water and its behaviour in the plants. The market is dominated by short-time respiration
measurements by O2 electrodes. Due to the low oxygen solubility of the waste water, the small
respirations are extrapolated to the final result by using factors. Essentially, only the easy
biodegradable substances are detected this way. Alternatively, LAR AG provides an online
respirometer that operates like a miniature waste water treatment plant: The BioMonitor. This
BOD analyser degrades nitrogen and carbon compounds in special waste water cascades, so
that the significant total BOD is determined. Due to the multi-stage cascade structure also
difficult to degrade components are reliably detected. The oxygen required for the degradation
is measured by an O2 sensor. Especially advantageous is the use of the plant's activated sludge,
whereby the conditions at the WWTP are simulated reliably. (Thomas, 1998)
4
mixing and sedimentation. Due without proper mixing process, the precipitation reaction that
occurs will be less than perfect. (G. J. Schroepfer, 1964)
Fenton reaction is a reaction that has been widely known is able to reduce the COD value.
The principle is the formation of free radicals that are created from the reaction between Fenton
reagent is FeSO4 with Hydrogen peroxide. Fenton reaction is what will be the forerunner of
the birth of AOP system. (G. J. Schroepfer, 1964)
5
5. Advanced Oxidation Process, The latest COD Reducing Method
AOP is the latest technology that is able to reduce the COD value of the wastewater. Even
of COD with a value above 100 thousand though. AOP was created from the added refinement
of Fenton reaction in the presence of ozone injection so adding reproduce free hydroxyl finally
able to oxidize chemicals in water. Excess AOP process is very fast reaction speed. If the
processing by microbiological reduction in COD can take days. Then the AOP process only
takes a matter of hours or even minutes. Another plus is also located in the area used for this
system is very small is compared with other systems. Moreover, this process does not require
a lot of chemicals to be added in the wastewater. (G. J. Schroepfer, 1964)
The latter method is widely used in the finishing process or post treatment after primary
treatment process. Usually used as an activated carbon filter. Activated carbon will absorb
organic substances, ozone or chlorine substances remaining on the processing results. So safe
effluent to be discharged into the environment. Filtration using activated carbon is also
commonly used in water treatment processes to remove odors and reduce the chemicals in the
water. (G. J. Schroepfer, 1964)
Determination of COD
To determine the COD, 10cm3 of 0.25N standard potassium dichromate solution was
added to 20cm3 of wastewater sample in a 250ml round-bottom flask. 1g of silver sulphate and
40cm3 of concentrated sulphuric acid were added in small portions with thorough swirling,
until the silver sulphate was completely dissolved. A few glass beads were added to serve as
anti-bumping aid, and the flask was connected to reflux condenser. The mixture was heated
gently for 10 minutes, after which the content of the flask was now cooled. 50cm3 of distilled
water was flushed through the condenser, and the cooling was completed under running tap
water. 2 drops of indicator solution were added and the resulting mixture titrated with
standardized ferrous ammonium sulphate solution until there was a change of colour from
yellow-green via blue-green to reddish brown. A blank value was determined in the same way
with 20cm3 of distilled water. The COD values of the respective samples were calculated using
equation (1).
6
Equation 1
1000(𝑎 − 𝑏)𝐶
𝐶𝑂𝐷 =
𝑉
In equation (1), a and b are the respective volumes of ferrous ammonium sulphate used for the
blank and sample (cm3 ), V is the volume of sample (cm3 ), and c the normality of ferrous
ammonium sulphate. (G. J. Schroepfer, 1964)
Experimental procedure
1- put a few clean glass beads to COD flask. Then put 20 ml of sample in flask,
add 0.4g mercuric sulfate and add 10 ml of potassium dichromate solution
6- dilute the mixture and titrate the excess dichromate with ferrous ammonium
sulfate solution using ferroin indicator, two or three drops of the indicator are
added to the sample just before titration.
7- Take as the end point the sharp color change from blue-green to reddish
brown.
7
Experimental Results
After performing the procedures and obtain the needed measurements which will be used
now in the calculations. The measurements showed that:
Equation 2 COD
COD=
𝑁 𝑠𝑜𝑑𝑖𝑢𝑚 𝑡ℎ𝑖𝑜𝑠𝑢𝑙𝑝ℎ𝑎𝑡𝑒 ×𝑑𝑉 ×𝑤𝑡.𝑜𝑓 𝑜𝑥𝑦𝑔𝑒𝑛
𝑣𝑜𝑙𝑢𝑚𝑒 𝑜𝑓 𝑠𝑎𝑚𝑝𝑙𝑒
0.1 ×(20−7.5)×8
COD= = 500 𝑚𝑔/𝑙
20
Discussion
After we make the calculations and the chemical oxygen demand (COD) had been
calculated it must be compared to the allowable standard limit of the COD in the water. As
its mentioned in the water standard quality that the allowable limit for COD in the water is
250 mg/L. As the allowable limit is 250 mg/L so the COD obtained isn’t acceptable
according to the water standards quality, so it’s obtained that after obtaining the
measurements and perform the needed calculations it obtained that the COD is 500 mg/L.
8
Conclusion
(BOD) and (COD) are the foremost normally used parameters for the characterization
of wastewaters. each of those parameters have blessings and downsides, and therefore
the alternative typically depends on several factors like the fundamental quantity needed to
see each of them. it's essential to get a correlation between BOD and COD
for numerous effluent treatment plants, to assist within the style and operation of
those plants. Wherever the allowable limit is 250 mg/l however it had been complete to be five
hundred mg/l.
9
References
10