H.Kangari O.D.
Keratometry
The keratometer is an instrument that is
used to measure the front surface (approx
2 - 4 mm radius) curvature of the cornea.
Keratometry
Information that provides are:
The radii of curvature of the cornea.
Keratometry
Optical Principles
Cornea is a convex refracting surface
In order to find the refracting power of the
cornea, we need to reflect an object of a
known size at a known distance off the
corneal surface. Then determine the size
of the reflecting image with measuring
telescope and calculate the refractive
power of the cornea based on the
refractive index of n= 1.3375
Keratometry
The keratometer usually shows the corneal curvature in
diopters of power or in millimeters and diopters.
Keratometry
Optics
Primary Care Optometry-Grosvenor page 224
Keratometry
FIGURE 1
Keratometry
Keratometer Mechanical
Adjustments and Parts
A. Focusing Knob
B. Rotating Grip for Locating
the Axis
C. Chin Rest
D. Locking Knob
E. Elevating Knob
EP. Eyepiece
G. Head Rest Adjusting Knob
H. Occluding Shield
HR. Head Rest
J. Chin Rest Adjusting Knob
K. Leveling Sight
M. Horizontal Measuring Drum
N. Vertical Measuring Drum
S. Axis Scale
Keratometry
Bausch & Lomb Keratometer
One position keratometer: it measures two
meridians at the same time
Keratometry
PROCEDURE
1)Eyepiece Adjustment : turn the
eyepiece (EP) counterclockwise fully.
Place a white sheet of paper in front of the
telescope part. Turn the eye piece
clockwise until the specified target in the
instrument becomes in sharp focus.
Keratometry
2)Adjust the height of the instrument and
align the outer canthus of the patient with
the leveling sight sign (item K in Figure 2)
Keratometry
4)Ask the patient to look at the center of
the instrument. If you are aligned, you can
see the circles in the eye. The patient can
see the picture of their eye in the
instrument.
Keratometry
5)Adjust the focusing knob( item A in Fig
2) such that you get a single image of the
central circle.(A,B in Fig 3)
Keratometry
7) Use the horizontal alignment knob(M) to bring
the crosses on each other.(E in Fig 3)
Keratometry
FIGURE 3
Keratometry
Javal-Schiotz Keratometer
Two position Keratometer
Keratometry
Procedure
Follow steps 1-4 of the above instrument
5) Adjust the instrument to see a clear image of
the rectangle and staircase targets
Keratometry
6)Rotate the instrument to align the line in
rectangle with the line in the staircase
target.
7)
Turn the measuring knob until staircase
and rectangle just overlap.
Example 42.00@180
45.00@90
-3.00x180 Mires Clear & Regular