User’s Guide
Notices U.S. Government Rights Warranty
The Software is “commercial computer THE MATERIAL CONTAINED IN THIS
Copyright Notice software,” as defined by Federal Acqui- DOCUMENT IS PROVIDED “AS IS,”
sition Regulation (“FAR”) 2.101. Pursu- AND IS SUBJECT TO BEING
© Keysight Technologies 2015 - 2016 ant to FAR 12.212 and 27.405-3 and CHANGED, WITHOUT NOTICE, IN
No part of this manual may be repro- Department of Defense FAR Supple- FUTURE EDITIONS. FURTHER, TO THE
duced in any form or by any means ment (“DFARS”) 227.7202, the U.S. MAXIMUM EXTENT PERMITTED BY
(including electronic storage and government acquires commercial com- APPLICABLE LAW, KEYSIGHT DIS-
retrieval or translation into a foreign puter software under the same terms CLAIMS ALL WARRANTIES, EITHER
language) without prior agreement and by which the software is customarily EXPRESS OR IMPLIED, WITH REGARD
written consent from Keysight Technol- provided to the public. Accordingly, TO THIS MANUAL AND ANY INFORMA-
ogies as governed by United States and Keysight provides the Software to U.S. TION CONTAINED HEREIN, INCLUD-
international copyright laws. government customers under its stan- ING BUT NOT LIMITED TO THE
dard commercial license, which is IMPLIED WARRANTIES OF MER-
Trademark embodied in its End User License CHANTABILITY AND FITNESS FOR A
Bluetooth and the Bluetooth logos are Agreement (EULA), a copy of which can PARTICULAR PURPOSE. KEYSIGHT
trademarks owned by Bluetooth SIG, be found at http://www.keysight.com/ SHALL NOT BE LIABLE FOR ERRORS
Inc., U.S.A. and licensed to Keysight find/sweula. The license set forth in the OR FOR INCIDENTAL OR CONSE-
Technologies. EULA represents the exclusive authority QUENTIAL DAMAGES IN CONNECTION
by which the U.S. government may use, WITH THE FURNISHING, USE, OR
Manual Part Number modify, distribute, or disclose the Soft- PERFORMANCE OF THIS DOCUMENT
U1281-90003 ware. The EULA and the license set OR OF ANY INFORMATION CON-
forth therein, does not require or per- TAINED HEREIN. SHOULD KEYSIGHT
Edition mit, among other things, that Keysight: AND THE USER HAVE A SEPARATE
(1) Furnish technical information WRITTEN AGREEMENT WITH WAR-
Edition 02, March 31, 2016
related to commercial computer soft- RANTY TERMS COVERING THE MATE-
Printed in: ware or commercial computer software RIAL IN THIS DOCUMENT THAT
documentation that is not customarily CONFLICT WITH THESE TERMS, THE
Printed in Malaysia provided to the public; or (2) Relinquish WARRANTY TERMS IN THE SEPARATE
to, or otherwise provide, the govern- AGREEMENT SHALL CONTROL.
Published by: ment rights in excess of these rights
Keysight Technologies customarily provided to the public to Safety Information
Bayan Lepas Free Industrial Zone, use, modify, reproduce, release, per-
11900 Penang, Malaysia form, display, or disclose commercial CAUTION
computer software or commercial com-
Technology Licenses puter software documentation. No A CAUTION notice denotes a hazard. It
The hardware and/or software additional government requirements calls attention to an operating proce-
described in this document are fur- beyond those set forth in the EULA dure, practice, or the like that, if not
nished under a license and may be shall apply, except to the extent that correctly performed or adhered to,
used or copied only in accordance with those terms, rights, or licenses are could result in damage to the product
the terms of such license. explicitly required from all providers of or loss of important data. Do not pro-
commercial computer software pursu- ceed beyond a CAUTION notice until
Declaration of Conformity ant to the FAR and the DFARS and are the indicated conditions are fully
set forth specifically in writing else- understood and met.
Declarations of Conformity for this
where in the EULA. Keysight shall be
product and for other Keysight prod-
under no obligation to update, revise or
ucts may be downloaded from the
Web. Go to http://www.keysight.com/
otherwise modify the Software. With WARNING
respect to any technical data as
go/conformity. You can then search by
defined by FAR 2.101, pursuant to FAR A WARNING notice denotes a hazard. It
product number to find the latest Dec-
12.211 and 27.404.2 and DFARS calls attention to an operating proce-
laration of Conformity.
227.7102, the U.S. government dure, practice, or the like that, if not
acquires no greater than Limited Rights correctly performed or adhered to,
as defined in FAR 27.401 or DFAR could result in personal injury or death.
227.7103-5 (c), as applicable in any Do not proceed beyond a WARNING
technical data. notice until the indicated conditions are
fully understood and met.
CAT III Category III 1000 V overvoltage CAT IV Category IV 600 V overvoltage
1000 V protection 600 V protection
– Do not exceed any of the measurement limits defined in the specifications to avoid
WARNING instrument damage and the risk of electric shock.
– Do not use the multimeter if it is damaged. Before you use the multimeter, inspect the
case. Look for cracks or missing plastic. Pay particular attention to the insulation
surrounding the connectors.
– Inspect the test leads for damaged insulation or exposed metal. Check the test leads
for continuity. Replace damaged test leads before you use the multimeter.
– Do not operate the multimeter around explosive gas, vapor, or wet environments.
– Do not apply more than the rated voltage (as marked on the multimeter) between
terminals, or between terminal and earth ground.
– Never use the multimeter in wet conditions or when there is water on the surface. If
the multimeter is wet, ensure that the multimeter is dried only by trained personnel.
– Before use, verify the multimeter’s operation by measuring a known voltage.
– When measuring current, turn off the circuit power before connecting the multimeter
in the circuit. Remember to place the multimeter in series with the circuit.
– When servicing the multimeter, use only the specified replacement parts.
– Use caution when working above 60 V DC, 30 V AC rms, or 42.4 V peak. Such voltages
pose a shock hazard.
– When using the probes, keep your fingers behind the finger guards on the probes.
– Connect the common test lead before you connect the live test lead. When you
disconnect the leads, disconnect the live test lead first.
– Remove the test leads from the multimeter before you open the battery cover.
– Do not operate the multimeter with the battery cover or portions of the cover removed
or loosened.
– To avoid false readings, which may lead to possible electric shock or personal injury,
replace the battery as soon as the low battery indicator appears and flashes.
– Only probe assemblies with Measurement Category III or IV ratings should be used for
mains measurements.
Measurement CAT III Measurements performed in the building installation. Examples are
measurements on distribution boards, circuit- breakers, wiring, including cables, bus-bars,
junction boxes, switches, socket outlets in the fixed installation, and equipment for
industrial use, and some other equipment including stationary motors with permanent
connection to the fixed installation.
Operating condition
– –20 °C to 55 °C, 0% to 80% RH (non-condensing)
Temperature
Storage condition
– –40 °C to 70 °C, 0% to 80% RH (without batteries)
Pollution degree 2
Safety compliance
– IEC/EN 61010-1:2010 (3rd Edition)
– IEC/EN 61010-2-033:2012 (First Edition)
– CAN/CSA-C22.2 No. 61010-1-12
– CAN/CSA-C22.2 No. 61010-033-12
– ANSI/UL Std. No. 61010-1-12
– ANSI/UL Std. No. 61010-033-12
EMC compliance
– IEC 61326-1:2012/EN 61326-1:2013
– ICES/NMB-001: Issue 4, June 2006
– AS/NZS CISPR 11:2004
This symbol indicates the time period during This symbol is a South Korean Class A EMC
which no hazardous or toxic substance Declaration. This is a Class A instrument
elements are expected to leak or deteriorate suitable for professional use and in
during normal use. Forty years is the expected electromagnetic environment outside of the
useful life of the product. home.
Product category
With reference to the equipment types in the WEEE directive Annex 1, this instrument is
classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Measurement Category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Regulatory Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC . . . . . . . . . .10
Product category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
1 Introduction
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Documentation map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Safety notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Preparing Your Multimeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Check the shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Removing the holster . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Installing the holster . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Installing the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Turn on your multimeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Automatic power-off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Enabling the backlight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Selecting the range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Alerts and warnings during measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Adjusting the tilt stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Remote communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Power-on options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Your Multimeter in Brief . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Rotary switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Display screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Input terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Using the Remote Switch Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Cleaning Your Multimeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
2 Making Measurements
Crest Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
3 Multimeter Features
Non-Contact AC Voltage Detection (Vsense) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Making Relative Measurements (Null) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Capturing Maximum and Minimum Values (MaxMin) . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Capturing Peak Values (Peak) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Freezing the Display (TrigHold and AutoHold) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Recording Measurement Data (Data Logging) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Performing manual logs (HAnd) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Performing interval logs (AUto) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Performing event logs (triG) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Performing export logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Reviewing Previously Recorded Data (View) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Sanitizing the Log Memories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
1 Introduction
This chapter lists the package contents for this multimeter, and teaches you how to set up
your multimeter for the first time. An introduction to all the features of the multimeter is
also given. This introduction does not cover all of the capabilities of the multimeter but
gives basic examples to help you perform basic operations on your multimeter.
1 Introduction
Documentation map
The following manuals and software related to the U1281A/U1282A Handheld Digital
Multimeter are available for download. Please visit our website at
http://www.keysight.com/find/hhTechLib for the latest version.
Check the manual edition on the first page of each manual.
User’s Guide. This manual.
Quick Start Guide. Printed copy for outdoor use, included with shipment.
Service Guide. Downloadable from http://www.keysight.com/find/hhTechLib
Keysight Handheld Meter Logger Software.
Downloadable from http://www.keysight.com/find/hhmeterlogger
Safety notes
The following safety notes are used throughout this manual. More pertinent safety notes for
using this product are located under the Safety Symbols section.
Caution denotes a hazard. It calls attention to a procedure that, if not correctly performed
CAUTION or adhered to, could result in damage to or destruction of the product. Do not proceed
beyond a caution notice until the indicated conditions are fully understood and met.
Tr
ue
R
M
S
M
ul
tim
et
e
er
o
ut
A
se
en
Vs
l
ua
D
ak z
Pe H
M
A
m
V
m
A
V
T
U
O
V
m
0V
60 V
00
T
IV 10 V
V A III AX
C AT M
C
A
0m D
44 SE M
X
FU
R
M
T CO
A
M D
A SE
10 FU T
U A
O m
V se
ns
e
2A
Tru
e RM
S Mu
ltim
ete
r
1
28
U1
to
Au
se
en
Vs
ak al
Pe Du
z
MH
mV mA
V A
T
OU
mV
V
0V
60 V
00
T IV 10
CA T III AX
CA M V
A
0m D
44 SE
X FU M
T
MA D RM CO
A
10 FU SE
T
OU
A
2
er
et
tim
ul
M
S
M
R
ue
Tr
A
82
12
U
o
ut
A
se
en
Vs
l
ua
D
Hz
ak
M
Pe
A
m
V
m
A
V
T
U
O
V
A 00 0V
m
M 10 60
V
V
A IV
X
C AT
III
T
C
V
A
FU 0m
D
M
SE
44
O
C
T
M
R
D X
SE A
FU A M
10
mT
A
U
O
A
Before you proceed with the battery installation, remove all cable connections to the
CAUTION terminals and ensure that the rotary switch is at the OFF position. Use only the battery
type specified in the datasheet.
The battery level indicator at the lower left-hand corner of the display indicates the relative
condition of the batteries. To ensure that the multimeter’s battery level indicator is
accurate, please select your battery type in the Setup menu (refer to “Changing the battery
type” on page 131 for more information). When the battery voltage falls below 3.8 V, the
multimeter will shut down automatically regardless of the Auto Power-Off setting.
Table 1-1 describes the various battery levels the indicator represents.
Battery capacity
Full capacity
2/3 capacity
1/3 capacity
(Flashing periodically) Almost empty
To avoid false readings, which could lead to possible electric shock or personal injury,
WARNING replace the battery as soon as the low battery indicator appears. Do not discharge the
battery by shorting the battery or reverse the battery polarity in any of the subjects.
To power ON your multimeter, turn the rotary switch to any position other than . The
model number of your multimeter and its firmware version are shown on the primary
display and secondary display respectively.
Firmware version
Model number
To power OFF your multimeter, turn the rotary switch to the position.
Automatic power-off
Your multimeter automatically turns off if the following actions do not take place within the
set duration (refer to “Changing the Auto Power-Off and backlight timeouts” on page 121
for more information):
– The rotary switch is moved.
– A key is pressed.
– The multimeter is in MaxMin mode (see “Capturing Maximum and Minimum Values
(MaxMin)” on page 100).
– The multimeter is in Peak Hold mode (see “Capturing Peak Values (Peak)” on
page 102).
The following actions will turn the multimeter back on after it is powered off automatically:
– Pressing any key on the keypad.
If viewing the display becomes difficult in low-light conditions, press to activate the
backlight for the LCD and the keypad.
To conserve battery life, a user-adjustable timeout controls how long the backlight stays
on. The default timeout is 15 seconds (refer to “Changing the Auto Power-Off and
backlight timeouts” on page 121 for more information).
graph. Pressing Range switches the multimeter between manual and autoranging. It also
scrolls through the available multimeter ranges when manual ranging is enabled.
Autoranging is convenient because the multimeter automatically selects an appropriate
range for sensing and displaying each measurement. However, manual ranging results in
better performance since the multimeter does not have to determine which range to use
for each measurement.
In autorange, the multimeter selects the lowest range to display the highest available
precision (resolution) for the input signal. If manual range is already enabled, press Range
Each additional press of Range sets the multimeter to the next higher range, unless it is
already in the highest range, at which point the range switches to the lowest range.
The multimeter will display the hazardous voltage ( ) symbol as an early precaution when
the measured voltage is equal to or greater than 30 V or when OL occurs. The following
table shows the voltage measurement modes affected:
Measurement DC AC
The multimeter will display the hazardous current ( ) symbol as an early precaution when
the measured current reaches the maximum fuse rating or when OL occurs. If your
measuring current is > 10 A ~ 19.999 A, you will need to lower the current within a
30-second time limitation to avoid blowing the multimeter’s fuse.The following table shows
the current measurement modes affected:
Measurement DC AC
Input warning
To avoid circuit damage and possibly blowing the multimeter’s current fuse, do not place
CAUTION the probes across (in parallel with) a powered circuit when a lead is plugged into a current
terminal. This causes a short circuit because the resistance through the multimeter's
current terminals is very low.
The red LED on the multimeter turns on and the multimeter emits a continuous beep and
displays or in the secondary display when the test lead is inserted into
the A or mA input terminal but the rotary switch is not set to the correct current
position. The sound will continue and the red LED will remain lit until the test leads are
removed from the multimeter. If the test leads are not removed, the sound will stop after 5
seconds.
Vsense
Dual
mV MHz
V
mV
V mA
OUT
A RMT
OUT V
mA COM
This warning is intended to stop you from attempting to measure voltage, continuity,
resistance, capacitance, diode, or temperature values when the leads are plugged into a
current terminal.
The red LED on the multimeter turns on briefly and the multimeter emits a brief beep and
displays in the secondary display when the rotary switch is set to the mA or A
current position but no test leads are inserted into the A or mA input.
To PC (host)
Remote communication
You can communicate remotely with the multimeter from your PC via an IR-USB
connection (see Using an IR-USB cable) or an IR-Bluetooth connection (see Using a
Bluetooth adapter).
When configuring the IR interface of the multimeter, use the following settings as the
default:
– Baud rate: 9600 bits per second
– Parity bit: None
– Data bits: 8 data bits
– Number of Stop bits: 1 bit
You can use any of the following software to communicate with the multimeter:
– Keysight Handheld Meter Logger (for Windows PC)
– Keysight Mobile Meter (for Android or iOS devices)
– Keysight Mobile Logger (for Android or iOS devices)
Bluetooth adapter
The Keysight Handheld Meter Logger Software and its supporting documents are available
for download from http://www.keysight.com/find/hhmeterlogger.
Power-on options
Some options can be selected only while you turn the multimeter on. These power-on
options are listed in the table below. To select a power-on option, press and hold the
specified key while turning the rotary switch from to any other position. Power-on
options remain selected until the multimeter is turned off.
Key Description
Hold LCD test. All annunciators are displayed in the LCD. Press any key to exit this mode.
Simulates the Auto Power-Off (APO) mode. Press any key to turn the multimeter back on and
Hz Save resume normal operation.
Shift Enters the Setup mode. For more information, refer to Chapter 4, "Using the Setup Menu".
Esc
Triggers the exporting of data via the multimeter’s optical communication port. The reading on the
primary display is exported, according to the display update rate. Restart the multimeter to resume
MaxMin
normal operation.
Overview
Front panel
The front panel parts of your multimeter are described in this section. Click the respective
“Learn more” pages for more information on each part.
Vsense
1
U1282A True RMS Multimeter
Dual
3
mV MHz
V
mV
V mA 4
A
OUT
A RMT
OUT V
mA COM
5
3 Keypad page 36
5 Terminals page 45
Rear panel
The rear panel parts of your multimeter are described in this section. Click the respective
“Learn more” pages for more information on each part.
1 IR communication port -
Rotary switch
The measurement functions for each rotary switch position are described in Table 1-7.
Turning the rotary switch changes the measurement function and resets all other
measurement options.
NOTE Some rotary switch positions have a shifted function printed in orange. Press Esc Shift to
switch between the shifted and regular function. See page 37 for more information on the
Esc
Shift key or page 142 for more information on shifted functions.
Remove the test leads from the measuring source or target before changing the rotary
WARNING switch position.
mV MHz
V
mV
V mA
OUT
Each position of the rotary switch (shown in Figure 1-11) is described in Table 1-7.
Off page 25
A
AC, DC, or AC+DC current measurement
Keypad
The operation of each key is explained below. Pressing a key enables a function, displays a
related symbol, and emits a beep. Turning the rotary switch to another position resets the
current operation of the key. Click the respective “Learn more” pages for more information
on each function.
Dual
Vsense – While in Null mode, press again – Press Range to toggle between page 96
to view the stored reference value that low sensitivity (Lo.SE) or high and
has been saved. The display will return sensitivity (Hi.SE) page 98
to normal after 3 seconds. Vsense
– Press Hz Save to scroll through the reading, press Hz Save again for
frequency (Hz), duty cycle (%), and pulse more than 1 second.
width (ms) measurements.
– If data logging is set as
– In duty cycle and pulse width
(interval data logging), pressing
page 87
Hz measurements, press for more and
Hz Save for more than 1 second will
Save
than 1 second to switch between the page 105
positive- or negative-edge trigger. enter the interval data logging mode,
where data is logged at the interval
defined in the multimeter’s Setup.
– Press Hz Save to scroll back to the
frequency measurement mode, and hold – If data logging is set as (event
for more than 1 second to exit the data logging), pressing Hz Save for
frequency measurement function. more than 1 second will enter the
event data logging mode, where data
is logged each time a triggering
condition is satisfied.
– Enables autoranging.
– Press Range again to scroll
– During temperature measurements,
through each available measurement
range.
pressing Range changes the page 26
Range – Enables or d isables the temperature
temperature measurement unit
measurement without ambient
between Celsius (°C) and Fahrenheit
compensation mode.
(°F)
– Changes the measurement range and
restarts the maximum and minimum
peak measurements.
– Press or on Hold
Display screen
The display annunciators of your multimeter are described in this section. See also
“Measurement units” on page 43 for a list of available measurement signs and notations
and “Analog bar graph” on page 44 for a tutorial on the analog bar graph located at the
bottom of your display screen.
AC, DC, and AC+DC indication for secondary display page 57 and page 87
Tone enabled -
Measurement units
The available signs and notations for each measurement function in your multimeter are
described in Table 1-10. The units listed below are applicable to the primary display and
secondary display measurements of your multimeter.
Sign/Notation Description
200 V, A, ,
33.3 V, A, ,
An unstable bar graph and unmatched primary display when measuring DC voltage usually
means the presence of AC voltages in the circuit if the DC Filter is disabled in the Setup
mode.
[1] The analog bar graph measurement rate is approximately 30 times/second for DC voltage, current, and
resistance measurements.
Input terminals
The terminal connections for the different measurement functions of your multimeter are
described in Table 1-12. Observe the rotary switch position of your multimeter before
connecting the test leads to the connector terminals.
Ensure that the terminal connections are correct for that particular measurement function
WARNING before starting any measurement.
To avoid damaging this multimeter, do not exceed the rated input limit.
CAUTION
A RMT
OUT V
mA COM
Rotary
switch Input terminals Overload protection
position
V
1000 Vrms
V
mV
COM
V
mV
MHz
A COM
A
11 A/1000 V, fast-acting fuse
mA mA COM
440 mA/1000 V, fast-acting fuse
OUT
RMT
Remote probe terminal (see “Using the Remote Switch Probe” on page 47 for more information)
Figure 1-15 The location of the button on the Remote Switch Probe
To change the default button operation, see ““Changing the remote button function” on
page 133.
To avoid electrical shock or damage to the multimeter, ensure that the insides of the
WARNING casing stay dry at all times.
Dirt or moisture in the terminals can distort readings. Follow the steps below to clean your
multimeter.
1 Turn the multimeter off and remove the test leads.
2 Turn the multimeter over and shake out any dirt that may have accumulated in the
terminals.
Wipe the case with a damp cloth and mild detergent — do not use abrasives or solvents.
Wipe the contacts in each terminal with a clean swab dipped in alcohol.
2 Making Measurements
Crest Factor 50
Measuring AC Voltage 51
Using the LPF Function for AC measurements (for U1282A only) 53
Measuring DC Voltage 55
Measuring AC and DC Signals 57
Using the LPF (Low Pass Filter) Function for AC+DC measurements 59
Making dB Measurements 60
Measuring Resistance 62
Measuring conductance 64
Testing for Continuity 65
Testing Diodes 68
Frequency Counter (for U1282A only) 72
Measuring Capacitance 74
Measuring Temperature 76
Temperature measurement without ambient compensation 79
Measuring AC or DC Current 80
% Scale of 4-20 mA or 0-20 mA 84
Measuring Frequency 87
Measuring pulse width 90
Measuring duty cycle 91
Square Wave Output 92
The following sections describe how to take measurements with your multimeter.
2 Making Measurements
Crest Factor
The crest factor may be determined by using this formula:
Peak value
Crest factor = --------------------------------
True rms value
You may refer to “Capturing Peak Values (Peak)” on page 102 on how to obtain the peak
values.
The crest factor may be up to 3.0 at full-scale except for the 600 V and the 1000 V range
where it is at 2.5 and 1.5 respectively, as shown in the table below:
6V 3.0 +/– 18 V
Exceeding the crest factor limit may result in an incorrect or a lower reading. Do not
WARNING exceed the crest factor limit to avoid instrument damage and the risk of electric shock.
Measuring AC Voltage
Set up your multimeter to measure AC voltage as shown in Figure 2-2. Probe the test
points and read the display.
Scrolls between
– AC V
V DC V
– AC+DC V, or
– DC V
Scrolls between
– AC mV
mV DC mV
– AC+DC mV, or
– DC mV
NOTE – Press and hold for more than 1 second to scroll through the available dual display
combinations. (Refer to “Append ix B: Dual Display Combinations Using the Dual Key”
on page 143 for more information)
– Press Hz Save to enable the frequency test mode for voltage measurements. See
“Measuring Frequency” on page 87 to learn more.
Vsense
AC
Vsense
3
Peak Auto
Dual
mV MHz
V
2
mV
V mA
V mV
A
OUT
A RMT
OUT V
mA COM
– To avoid possible electric shock or personal injury, do not use the LPF option to verify
WARNING the presence of hazardous AC voltages. AC voltage values greater than what are
indicated may be present when the LPF is enabled.
– First, make an AC voltage measurement without LPF to detect the possible presence
of hazardous voltages. Then, enable the LPF if required for measurement stability and
response speed.
Your multimeter is equipped with an AC LPF (low-pass filter) to help reduce unwanted
electronic noise when measuring AC voltage or AC frequency. The LPF can improve
measurement performance on composite sine waves that are typically generated by
inverters and variable frequency drives.
Table 2-2 Rotary switch positions allowing AC voltage measurements with LPF
AC V AC V with LPF
V
AC mV AC mV with LPF
mV
Set up your multimeter to measure AC voltage as shown in Figure 2-2. Press Esc Shift to
enable the LPF. Your multimeter continues measuring in the chosen AC mode, but now the
signal diverts through a filter that blocks voltages above 1 kHz (refer to Figure 2-3), as
shown in Figure 2-4.
1 kHz
or mV
+ Esc
Shift
V
Measuring DC Voltage
Set up your multimeter to measure DC voltage as shown in Figure 2-7. Probe the test
points and read the display.
Scrolls between
– AC V
V DC V
– AC+DC V, or
– DC V
Scrolls between
– AC mV
mV DC mV
– AC+DC mV, or
– DC mV
This multimeter displays DC voltage values as well as their polarity. Negative DC voltages
NOTE will return a negative sign on the left of the display.
– If the signal includes an AC component and the peak value exceeds the measurement
CAUTION range, you will need to lock the range that is greater than the voltage peak. Use the
maximum range of 1000 V to check the signal condition, and manually select a
suitable range for the signal.
– As the DC voltage measurement has the Normal (Series) Mode Rejection Rate (NMRR)
capability for 50/60 Hz noise, a hazardous voltage indication ( ) will be shown even if
the display is showing 0 V or < 30 V. The measurement range will automatically be set
to a higher range, and the analog bar graph will be varying faster or greater than the
displayed value.
NOTE – Press and hold for more than 1 second to scroll through the available dual display
combinations. (Refer to “Append ix B: Dual Display Combinations Using the Dual Key”
on page 143 for more information)
– Press Hz Save to enable the frequency test mode for voltage measurements. See
“Measuring Frequency” on page 87 to learn more.
Vsense
Dual
2
mV MHz
V mV V
mV
V mA
OUT
A RMT
OUT V
mA COM
Scrolls between
– AC V
V DC V
– AC+DC V, or
– DC V
Scrolls between
– AC mV
mV DC mV
– AC+DC mV, or
– DC mV
Scrolls between
– AC A
DC A
A – AC+DC A, or
– DC A
Scrolls between
– AC mA (or μA)
mA
DC mA (or μA)
– AC+DC mA (or μA), or
– DC mA (or μA)
NOTE – Press and hold for more than 1 second to scroll through the available dual display
combinations. (Refer to “Append ix B: Dual Display Combinations Using the Dual Key”
on page 143 for more information)
– Press Hz Save to enable the frequency test mode for voltage measurements. See
“Measuring Frequency” on page 87 to learn more.
Using the LPF (Low Pass Filter) Function for AC+DC measurements
Your multimeter is equipped with an AC LPF to help reduce electronic noise when
measuring a mixed signal.
1 Enable the LPF (refer to “Enabling the AC path filter (for U1282A only)” on page 138).
3 Your multimeter continues measuring in the AC+DC mode, but now the signal diverts
through a filter that blocks unwanted voltages above 1 kHz.
Making dB Measurements
Your multimeter is capable of displaying voltage as a dB value, either relative to 1 milliwatt
(dBm) or a reference voltage of 1 volt (dBV).
To set the multimeter to display values in either dBm or dBV, first set up your multimeter to
measure voltage as shown in Figure 2-2 or Figure 2-7. Probe the test points, and read the
display. Then, press and hold for more than 1 second to scroll through the options
until the voltage measurements are displayed either as a dBm value (see Figure 2-11) or a
dBV value (see Figure 2-12)
AC V AC V with LPF
V
AC mV AC mV with LPF
mV
Scrolls between
– AC V
V DC V
– AC+DC V, or
– DC V
Scrolls between
– AC mV
mV DC mV
– AC+DC mV, or
– DC mV
Press and hold for more than 1 second to scroll through the options until you exit the
dBm or dBV function. Selecting the frequency test mode ( Hz Save ) also cancels the dBm or
dBV function.
Measuring Resistance
To avoid possible damage to your multimeter or to the equipment under test, disconnect
CAUTION the circuit power and discharge all high-voltage capacitors before measuring resistance.
Set up your multimeter to measure resistance as shown in Figure 2-14. Probe the test
points and read the display.
Vsense
3
Peak Vsense Auto
Dual
2
mV MHz
V
mV
V mA
OUT
A RMT
OUT V
mA COM
lead resistance from the measurement, hold the test lead tips together and press .
Now the resistance at the probe tips will be subtracted from all future display readings.
Measuring conductance
Small conductance values correspond to extremely high resistance values. The nS range
allows you to easily calculate and determine the resistance of components up to 100 GΩ
(0.01 nS resolution).
To measure conductance, set up your multimeter to measure resistance as shown in
Figure 2-14. Press Range until the conductance measurement is selected (nS unit shown).
Probe the test points, and read the display.
High-resistance readings are susceptible to electrical noise. Use averaging to smooth out
most of the noisy readings.
To avoid possible damage to your multimeter or to the equipment under test, disconnect
CAUTION the circuit power and discharge all high-voltage capacitors before testing for continuity.
Set up your multimeter to test for continuity as shown in Figure 2-15. Probe the test points
and read the display.
The continuity test features a beeper that sounds and a red LED that turns on as long as a
NOTE circuit is incomplete or broken. The audible and visual alert allow you to perform quick
continuity tests without having to watch the display.
In continuity, a short means a measured value is less than the threshold resistance values
listed in Table 2-8.
You can set the beeper to sound and the Vsense LED to turn on as a continuity indication
whether the circuit-under-test is less than (short) or more than or equal to (open) the
threshold resistance (refer to “Changing the continuity type” on page 129).
– Normal open: Circuit is normally open, the beeper will sound and the LED will turn on
when a short is detected.
– Normal closed: Circuit is normally closed, the beeper will sound and the LED will turn
on when an open is detected.
The audible continuity will be locked in the 600 Ω range for resistance measurement.
– The continuity function detects intermittent shorts and opens lasting as briefly as 1 ms.
NOTE A brief short or open causes the multimeter to emit a short beep and flash.
– You can enable or disable the audible and visual alert via the multimeter’s Setup. See
“Changing the continuity alert type” on page 130 for more information on the audible
alert option.
Vsense
4
U1282A True RMS Multimeter
OFF
( open)
ON
Dual ( closed)
Esc
Shift
2
mV MHz
V
mV
V mA
OUT
A RMT
OUT V
mA COM
Testing Diodes
To avoid possible damage to your multimeter or to the equipment under test, disconnect
CAUTION the circuit power and discharge all high-voltage capacitors before testing diodes.
Set up your multimeter to test diodes as shown in Figure 2-18. Probe the test points and
read the display.
– Use the diode test to check diodes, transistors, silicon controlled rectifiers (SCRs), and
NOTE other semiconductor devices. A good diode allows current to flow in one direction only.
– This test sends a current through a semiconductor junction and then measures the
junction’s voltage drop. A typical junction drop is 0.3 V to 0.8 V.
– Connect the red test lead to the positive terminal (anode) of the diode and the black
test lead to the negative terminal (cathode).
– Your multimeter can display diode forward bias of up to approximately 3.1 V. The
NOTE forward bias of a typical diode is within the range of 0.3 V to 0.8 V; however, the reading
can vary depending on the resistance of other pathways between the probe tips.
– If the beeper is enabled during a diode test, the multimeter will beep briefly for a
normal junction and sound continuously for a shorted junction below 0.050 V. See
“Changing the beep frequency” on page 127 to disable the beeper.
Reverse the probes (as shown in Figure 2-19) and measure the voltage across the diode
again. Assess the diode according to the following guidelines:
– A diode is considered open if the multimeter displays in both forward and reverse
bias modes.
Vsense
3
Peak Vsense Auto
Dual
2
mV MHz
MHz V
mV
V mA
OUT
A RMT
OUT V
mA COM
Vsense
3
Peak Vsense Auto
Dual
2
mV MHz
MHz V
mV
V mA
OUT
A RMT
OUT V
mA COM
– Use the frequency counter for low voltage applications. Never use the frequency
WARNING counter on AC power line systems.
– For input more than ± 1.8 Vp, you are required to use the frequency measurement
mode available under the current or voltage measurement instead of the frequency
counter.
The frequency counter can be set to divide a signal for megahertz (MHz) measurement. You
can use this to measure the stability of the crystal oscillator clock over time.
MHz
1 Position the rotary switch to .
2 Press Esc
Shift to select the frequency counter mode.
Vsense
AC
4
3 Peak Vsense Auto
Dual
Esc
Shift
2
mV MHz
MHz V
mV
V mA
OUT
A RMT
OUT V
mA COM
Measuring Capacitance
To avoid possible damage to the multimeter or to the equipment under test, disconnect
CAUTION circuit power and discharge all high-voltage capacitors before measuring capacitance.
Use the DC voltage function to confirm that the capacitor is fully discharged.
Set up your multimeter to measure capacitance as shown in Figure 2-22. Probe the test
points and read the display.
– The multimeter measures capacitance by charging the capacitor with a known current
NOTE for a known period of time, measuring the resulting voltage, and then calculating the
capacitance.
– is shown on the bottom left of the display when the capacitor is charging, and is
shown when the capacitor is discharging.
Vsense
– To improve measurement accuracy of small value capacitors, press with the test
leads open to subtract the residual capacitance of the multimeter and leads.
– For measuring capacitance values greater than 1000 μF, discharge the capacitor first,
then select a suitable range for measurement. This will speed up the measurement time
and also ensure that the correct capacitance value is obtained.
Vsense
3
Peak Vsense Auto + –
Dual
2 V
mV MHz
mV
V mA
OUT
A RMT
OUT V
mA COM
Measuring Temperature
Do not connect the thermocouple to electrically live circuits. Doing so will potentially
WARNING cause fire or electric shock.
– Do not bend the thermocouple leads at sharp angles. Repeated bending over a period
CAUTION of time can break the leads.
– Always set the temperature unit display per the official requirements and in
compliance with the National laws of your region.
The multimeter uses a type-K (default setting) temperature probe for measuring
temperature. To measure temperature, set up your multimeter as shown in Figure 2-24.
Probe the test points and read the display. The primary display normally shows
temperature or the message (open thermocouple). The open thermocouple message
may be due to a broken (open) probe or because no probe is installed into the input jacks of
the multimeter. Press and hold Range for more than 1 second to change the temperature
units between °C or °F
NOTE – Shorting the terminal to the COM terminal will display the temperature at the
multimeter’s terminals.
– To change the default thermocouple type from type-K to type-J, see “Changing the
thermocouple type” on page 125 for more information.
The bead-type thermocouple probe is suitable for measuring temperatures from –40 °C to
NOTE 204 °C (399 °F) in PTFE-compatible environments. Do not immerse this thermocouple
probe in any liquid. For best results, use a thermocouple probe designed for each specific
application — an immersion probe for liquid or gel, and an air probe for air measurement.
For quick measurement, use the compensation to view the temperature variation of the
thermocouple sensor. The compensation assists you in measuring relative temperature
immediately without compensating for the ambient temperature.
Vsense
3
4
Dual
Esc
Shift
mV MHz
2 V
mV
V mA
OUT
A RMT
OUT V
mA COM
1 Press Range to select compensation. This allows for a quick measurement of the
relative temperature.
2 Avoid contact between the thermocouple probe and the surface to be measured.
Vsense
3 After a constant reading is obtained, press to set the reading as the relative
reference temperature.
4 Touch the surface to be measured with the thermocouple probe and read the display.
Measuring AC or DC Current
– Before attempting any current measurement, switch off the power source of the
WARNING circuit and measure the AC or DC voltage to ensure that the power source has been
switched off.
– Never attempt an in-circuit current measurement where the open-circuit potential to
earth is greater than 1000 V. Doing so will cause damage to the multimeter and
possible electric shock or personal injury.
Set up your multimeter to measure AC or DC current as shown in Figure 2-27. Open the
circuit path to be tested. Probe the test points, and read the display.
Scrolls between
– AC A
DC A
A – AC+DC A, or
– DC A
Scrolls between
– AC mA (or μA)
mA
DC mA (or μA)
– AC+DC mA (or μA), or
– DC mA (or μA)
– To avoid blowing the multimeter’s 440 mA fuse, use the mA terminal only if you
CAUTION are sure the current is less than 440 mA. Refer to the Input warning section for
information on the alerts the multimeter uses when leads are not used correctly for
current measurements.
– Placing the probes across (in parallel with) a powered circuit when a lead is plugged
into a current terminal can damage the circuit you are testing and blow the
multimeter’s fuse. This happens because the resistance through the multimeter’s
current terminals is very low, resulting in a short circuit.
– To measure current, you must open the circuit-under-test, then place the multimeter in
NOTE series with the circuit.
– Turn off power to the circuit. Discharge all high-voltage capacitors. Insert the black test
lead into the COM terminal. Insert the red test lead in an input appropriate for the
measurement range.
– If you are using the mA terminal, set the rotary switch to mA for current below
– Press and hold for more than 1 second to scroll through the available dual display
combinations (refer to “Append ix B: Dual Display Combinations Using the Dual Key”
on page 143 for more information).
– Press Hz Save to enable the frequency test mode for current measurements. See
“Measuring Frequency” on page 87 to learn more.
Vsense
3
LOAD
Peak Vsense Auto
AC
Dual
mV MHz
V
2 mV
V mA
mA A
A
OUT
A RMT
OUT V
mA COM
position to mA and set up your multimeter to measure DC current by following the steps
listed in the Measuring AC or DC Current section. Press and hold for more than 1
second to scroll through the options until (or ) is shown on the right side of the
display.
Scrolls between
– AC mA (or μA)
mA
DC mA (or μA)
– AC+DC mA (or μA), or
– DC mA (or μA)
– The 4-20 mA current loop output from a transmitter is a type of electrical signal that is
NOTE used in a series circuit to provide a robust measurement signal that is proportional to
the applied pressure, temperature, or flow in process control. The signal is a current
loop where 4 mA represents the zero percent signal and 20 mA represents the 100
percent signal.
– The % scale for 4-20 mA or 0-20 mA in this multimeter is calculated using its
corresponding DC mA measurement. The multimeter will automatically optimize the
best resolution for the selected measurement.
The analog bar graph displays the current measurement value. (In the example above,
8 mA is represented as 40% in the 0-20 mA % scale and 25% in the 4-20 mA % scale.)
999.99%
~ 600 mA[a]
9999.9%
Use the % scale with a pressure transmitter, a valve positioner, or other output actuators to
measure pressure, temperature, flow, pH, or other process variables.
Vsense
3
Dual
4
mV MHz
V
mV
2
V mA
mA
A
OUT
A RMT
OUT V
mA COM
Pressure transmitter
Measuring Frequency
Never measure the frequency where the voltage or current level exceeds the specified
WARNING range. Manually set the voltage or current range if you want to measure frequencies
below 20 Hz.
AC V AC V with LPF
V
AC mV AC mV with LPF
mV
Scrolls between
– AC V
V DC V
– AC+DC V, or
– DC V
Scrolls between
– AC mV
mV DC mV
– AC+DC mV, or
– DC mV
Scrolls between
– AC A
DC A
A – AC+DC A, or
– DC A
Scrolls between
– AC mA (or μA)
mA
DC mA (or μA)
– AC+DC mA (or μA), or
– DC mA (or μA)
To measure frequency, rotate the switch to one of the primary functions allowing frequency
measurements highlighted in Table 2-15. Press Hz Save , then probe the test points and
read the display.
Pressing Range controls the input range of the voltage or ampere function, not the
frequency range.
The frequency of the input signal is shown in the primary display. The voltage or ampere
value of the signal is shown in the secondary display. The bar graph does not indicate
frequency but indicates the voltage or ampere value of the input signal.
– Measuring the frequency of a signal helps detect the presence of harmonic currents in
NOTE neutral conductors and determines whether these neutral currents are the result of
unbalanced phases or non-linear loads.
– Frequency is the number of cycles a signal completes each second. Frequency is
defined as 1/Period. Period is defined as the time between the middle threshold
crossings of two consecutive, like-polarity edges, as shown in Figure 2-32.
– The multimeter measures the frequency of a voltage or current signal by counting the
number of times the signal crosses a threshold level within a specified period of time.
– To obtain the best measuring results for frequency measurements, please use the AC
measuring path.
90%
50%
10%
+ Width - Width
Period
Press Hz Save to scroll through the frequency, pulse width, and duty cycle measurements.
Press Hz Save to scroll back to the frequency measurement mode, and hold for more
than 1 second to exit the frequency measurement function.
The pulse width function measures the amount of time a signal is high or low, as shown in
NOTE Figure 2-32. It is the time from the middle threshold of the rising edge to the middle
threshold of the next falling edge. The measured waveform must be periodic; its pattern
must repeat at equal time intervals.
1 To measure pulse width, position the rotary switch to one of the functions allowing
frequency measurements shown in Table 2-15.
2 Press Hz Save until the measurements are shown in the millisecond (ms) unit. Probe the
test points and read the display.
The pulse width of the input signal is shown in the primary display. The voltage or ampere
value of the signal is shown in the secondary display. The bar graph does not indicate pulse
width but indicates the voltage or ampere value of the input signal.
The pulse width polarity is displayed to the left of the pulse width value. indicates a
positive pulse width and indicates a negative pulse width. To change the polarity being
Press Hz Save to scroll through the frequency, pulse width, and duty cycle measurements.
Press Hz Save to scroll back to the frequency measurement mode, and hold for more
than 1 second to exit the frequency measurement function.
The duty cycle (or duty factor) of a repetitive pulse train is the ratio of the positive or
NOTE negative pulse width to the period expressed as a percentage, as shown in Figure 2-32.
The duty cycle function is optimized for measuring the on or off time of logic and switching
signals. Systems such as electronic fuel injection systems and switching power supplies are
controlled by pulses of varying width, which can be checked by measuring duty cycle.
1 To measure duty cycle, position the rotary switch on one of the functions allowing
frequency measurements shown in Table 2-15.
2 Press Hz Save until the measurements are displayed as a percentage (%). Probe the test
points and read the display.
The duty cycle percentage of the input signal is shown in the primary display. The voltage
or ampere value of the signal is shown in the secondary display. The bar graph does not
indicate duty cycle but indicates the voltage or ampere value of the input signal.
The pulse polarity is displayed to the left of the duty cycle value. indicates a positive
pulse and indicates a negative pulse. To change the polarity being measured, press and
Press Hz Save to scroll through the frequency, pulse width, and duty cycle measurements.
Press Hz Save to scroll back to the frequency measurement mode, and hold for more
than 1 second to exit the frequency measurement function.
2 Press Esc
Shift to switch between duty cycle and pulse width for the primary display.
Vsense
Dual 1 2 3 4
oscilloscope
mV MHz
V
mV
V mA
A
2
OUT
OUT
10A MAX CAT IV 600V
FUSED CAT III 1000V
440mA
FUSED MAX
A RMT
OUT V
mA COM
3 Multimeter Features
The following sections describe the additional features available in your multimeter.
3 Multimeter Features
– You are advised to test on a known live circuit within the rated AC voltage range of
WARNING this product before and after each use to ensure that Vsense works.
– Voltage could still be present even if there is no Vsense alert indication. Do not rely on
Vsense with shielded wires. Never touch live voltage or conductors without the
necessary insulation protection or switching off the voltage source.
– Vsense may be affected by differences in socket design, insulation thickness, and
insulation type.
You are advised to measure voltage by using test leads through the voltage measurement
CAUTION function after using Vsense, even if there is no alert indication.
Vsense
1 Press and hold to enable Vsense (on any position of the rotary switch except
and OUT ).
If the multimeter senses the presence of AC voltage, the multimeter will beep and the red
NOTE
Vsense LED at the top of the multimeter will turn on. The symbol will also be displayed
on the LCD. The audible and visual alert allows you to easily sense the presence of AC
voltage in close proximity.
2 Press Range to toggle the Vsense’s sensitivity between high sensitivity (Hi.SE) or low
sensitivity (Lo.SE).
Vsense
When Vsense is disabled, the multimeter will return to the primary function of the current
NOTE rotary knob position regardless of what function it was in before Vsense was enabled.
2
AC voltage source
AC
3
Vsense
+
Vsense
beeper
U1282A True RMS Multimeter sounds
1
> 1 sec. Peak Vsense Auto
Vsense
Dual
mV MHz
V
mV
V mA
OUT
A RMT
OUT V
mA COM
Null can be set for both auto and manual range settings, but an overload reading cannot
NOTE be stored as a null value.
Vsense
1 To activate the relative mode, press the key. The measurement value at the time
that when Null ( ) is enabled, is stored as the reference value.
Vsense
2 Press again to view the stored reference value. The display will return to normal
after a brief moment.
Vsense
3 To disable the Null function, press while the stored reference value is shown.
For any measurement function, you can directly measure and store the null value by
Vsense
pressing with the test leads open (nulls the test lead capacitance), shorted (nulls the
test lead resistance), or across a desired null value circuit.
– In resistance measurement, the multimeter will read a non-zero value, even when the
NOTE two test leads are in direct contact, because of the resistance of these leads. Use the
null function to zero-adjust the display.
– For DC voltage measurements, the thermal effect will influence the accuracy of the
Vsense
measurements. Short the test leads and press when the displayed value is stable
to zero-adjust the display.
Vsense
Press to enable
the Null function.
The display
Vsense
returns to
Press
normal after 3
seconds
Flashing
Vsense
Press again to
disable the Null
function.
2 Press MaxMin again to scroll through the Max, Min, Avg, or present (MaxMinAvg) input
values.
3 The elapsed time is shown on the secondary display. Press Hold to restart the
recording session.
– Changing the range manually will also restart the recording session.
NOTE
– If an overload is recorded, the averaging function will be stopped. is shown in place
of the average value.
– The APO (Auto Power-Off) function is disabled when MaxMin is enabled.
– The maximum recording time is 99999 seconds (27 hours, 46 minutes, 39 seconds).
is shown if the recording exceeds the maximum time.
4 Press MaxMin for more than 1 second to disable the MaxMin function.
This mode is useful for capturing intermittent readings, recording minimum and maximum
readings unattended, or recording readings while equipment operation keeps you from
observing the multimeter display.
The average value displayed is the true arithmetic mean of all readings taken since the
start of recording. The average reading is useful for smoothing out unstable inputs,
calculating power consumption, or estimating the percentage of time a circuit is active.
1 To activate the peak mode, press the MaxMin key for more than 1 second.
2 Press MaxMin again to display the maximum (HoldMax) or minimum (HoldMin) peak
values along with their respective time stamps.
3 If (overload) is shown, press the Range key to change the measurement range. This
action will also restart the measurement.
5 Press MaxMin for more than 1 second to disable the Peak function.
When the peak value of the input signal goes below the recorded minimum value or above
the recorded maximum value, the multimeter beeps and records the new value.
At the same time, the elapsed time since the peak recording session was started is stored
as the recorded value’s time stamp.
Start to capture
Press Hold
to reset the recording
Press MaxMin to switch session.
between the maximum and
minimum peak values.
– The Setup menu entry is disabled. (Refer to “Changing the variation count” on
page 119)
In TrigHold operation mode, pressing Hold will manually trigger the holding of the
next measured value. The icon flashes before the display is updated.
AutoHold operation
Pressing Hold activates the AutoHold if the following conditions are met:
– The Hold Setup menu entry is enabled. (Refer to “Changing the variation
count” on page 119)
The AutoHold operation monitors the input signal and updates the display, and if enabled,
emits a beep whenever a new stable measurement is detected. The AutoHold mode will be
triggered when the input signal varies more than a selected adjustable (AutoHold
threshold) variation count (default 50 counts).
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
If the reading value is unable to reach a stable state, the reading value will not be updated.
NOTE
A manual log stores an instance of the measured signal each time Hz Save is pressed for
more than 1 second. See page 106.
An interval log stores a record of the measured signal at a user-specified interval. See
page 106.
An event log stores a record of the measured signal each time a trigger condition is
satisfied. See page 108.
A export log stores a record of the measured signal and exports it via the multimeter’s
optical communication port each time Hold is pressed. See page 109.
Export ( ) 100
Manual ( ) 100
Interval ( ) 10000
Before starting a recording session, set up the multimeter for the measurements to be
recorded.
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Available options: , , or .
1 Press Hz Save for more than 1 second to store the present input signal value and
function. and the log entry number are displayed at the top of the display. The
display will return to normal after a short while (around 1 second).
When all entries are occupied, will be shown in the secondary display when Hz Save
1 Press Hz Save for more than 1 second to start interval log mode.
and the log entry number are displayed at the top of the display. Subsequent
readings are automatically recorded into the multimeter’s memory at the interval
specified in the Setup menu.
2 Press Hz Save for more than 1 second to exit the interval log mode.
The maximum number of readings that can be stored for the interval log is 10000 entries.
When all entries are occupied, will be shown in the secondary display when Hz Save
When the interval log recording session is running, all other keypad operations are
NOTE
disabled; except for Hz Save , which, when pressed for more than 1 second, will stop and exit
the recording session. Furthermore, APO (Auto Power-Off) is disabled during the recording
session.
Ensure that is selected as the data logging option in the multimeter’s Setup.
Event logs are used only with the following modes:
– TrigHold and AutoHold (page 104)
– MaxMin recording (page 100)
– Peak recording (page 102)
Event records are triggered by the measured signal satisfying a trigger condition set by the
measurement function used in the following modes:
AutoHold When the input signal changes more than the variation count.
2 Press Hz Save for more than 1 second to start event log mode.
and the log entry number are displayed at the top of the display. Subsequent
readings are automatically recorded into the multimeter’s memory every time a trigger
condition specified in Table 3-2 is satisfied. The display will return to normal after a
short while (around 1 second).
3 Press Hz Save for more than 1 second to exit the event log mode.
The maximum number of readings that can be stored for the event log is 10000 entries.
When all entries are occupied, will be shown in the secondary display when Hz Save
1 Press Hold during measurement to store the present input signal value.
a In TrigHold mode, and are displayed at the top of the display, along with
. The and will disappear after a short while (around 1 second), but
the will remain.
b In AutoHold mode, and are displayed at the top of the display, along with
. The and will disappear after a short while (around 1 second), but
the will remain.
When all entries are occupied, will be shown in the secondary display when Hz Save
Viewing data stored in the multimeter’s memory is performed through the Esc
Shift key.
1 Press Esc
Shift for more than 1 second to enter the multimeter’s View mode. Press
Esc again to scroll through the manual (H), interval (A), event (E), or export (
Shift )
previously stored records.
If nothing has been recorded, the secondary display will show horizontal lines instead.
iii Press to view the next stored entry. The index number increases by one.
iv Press to view the previous stored entry. The index number decreases by
one.
3 Press Hz Save to delete the last stored entry, or press for more than 1 second to clear all
entries for the selected log type.
4 Press Esc
Shift for more than 1 second to exit the View mode.
Keysight U1281A/U1282A User’s Guide 111
3 Multimeter Features
When all entries are cleared, press and hold for more than 1 second to sanitize the log
entries.
The data sanitization operation may take up to 1-2 minutes to complete. Do not press any
CAUTION keys or turn the rotary switch until the data sanitization operation is completed.
The following sections describe how to change the preset features of your multimeter.
4 Multimeter Setup Options
Legend Description
Press Esc Shift while turning the rotary switch from the OFF position to any ON
position to access the Setup menu.
Shift
Esc
Press and hold Esc
Shift until the multimeter restarts to exit the Setup menu.
Vsense
Press or at each menu item to change the preset settings. The last
digit of the menu item (in the primary display) will flash to indicate that you can
Vsense now change the values shown in the primary display.
Vsense
Hz Save While the menu item is flashing, press Hz Save to save your changes.
The multimeter will automatically exit the Setup menu after 30 seconds of inactivity.
NOTE
When editing numerical values, use Hold to position the cursor on a numerical
digit.
When the cursor is positioned over a digit, use the and keys to change the
numerical digit.
Vsense
When you have completed your changes, save the new numerical value by pressing
Hz Save . (Alternatively, if you wish to discard the changes you made, press Esc
Shift .).
NOTE Press and hold the MaxMin button for more than 1 second to select the temperature unit
menu for settling.
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
To enable Smooth:
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
tiME (1 to 99999) s 1s
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
dbrEF (1 to 9999) Ω 50 Ω
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
This Setup item is locked for certain regions. Always set the temperature unit display per
CAUTION the official requirements and in compliance with the National laws of your region.
This setting is used with temperature measurements (see page 76). Four combinations of
displayed temperature unit(s) are available:
– Celsius only: Temperature measured in °C.
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
3200 Hz, 3268 Hz, 3339 Hz, 3413 Hz, 3491 Hz, 3572 Hz,
bEEP 3657 Hz, 3746 Hz, 3840 Hz, 3938 Hz, 4042 Hz, 4151 Hz, 3840 Hz
4267 Hz, or oFF
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
3 Press or to set the startup sound type. Select to disable the startup
sound.
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
rESEt YES or no no
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
3 Press or to select .
5 The multimeter will beep once and return to the first Setup menu item ( ).
All the settings will be reset to their default values except the temperature units.
NOTE
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
Number Function
b1 MaxMin/Peak
b2 ΔNull/Vsense
b3 Range/Auto
b4 Hz/Log
b5 Backlight/Dual
b6 Shift/View
b7 Hold
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
To perform high-speed measurements of voltage, current, resistance, and diode, set the
NOTE data refresh rate to 40 times per second. This will produce more than double the
measurement speed for these measurements.
CMRR and NMRR rejections are not applicable for this high-speed measurement mode.
Under this specialized condition, this mode provides reduced accuracy results at 40
readings per second.
The count resolution for the 40 times per second data refresh rate is 6,000, and the
resolution for the 5 times per second data refresh rate is 60,000.
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
inPUt 10 MΩ or 1000 MΩ 10 MΩ
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
1 Press Esc
Shift while turning the rotary switch to access the Setup menu.
Vsense
The table below lists the functions shown in the primary display when the Esc
Shift key is
pressed, with respect to the multimeter’s rotary switch position. Press Esc
Shift to cycle
through the available shift functions.
A Appendix A: Shift Functions Using the Shift Key
Square wave output (duty cycle mode) Square wave output (pulse width mode)
OUT
The table below lists the functions shown in the secondary display when the key is
pressed and held for more than 1 second, with respect to the multimeter’s rotary switch
position. Press for more than 1 second to cycle through the available dual display
combinations.
B Appendix B: Dual Display Combinations Using the Dual Key
AC+DC voltage
dBm measurement
measurement
(dBm mV)
(AC+DC mV)
AC+DC voltage
AC voltage
measurement
measurement (AC mV)
(AC+DC mV)
AC+DC voltage
DC voltage
measurement
measurement (DC mV)
(AC+DC mV)
Resistance
- - -
measurement (Ω)
Continuity test ( ) - - -
Capacitance
- - -
measurement (F)
Temperature
- - -
measurement (°C/°F)
AC+DC current
AC coupling frequency
measurement
measurement (Hz)
(AC+DC μA/mA)
AC+DC current
DC current measurement
measurement
(DC μA/mA)
(AC+DC μA/mA)
AC+DC current
AC coupling frequency
A measurement
measurement (Hz)
(AC+DC A)
AC+DC current
DC current
measurement
measurement (DC A)
(AC+DC A)