GC-7890SD Power system Dedicated

Insulation Oil and Gas


Chromatography Analysis System
Manual
China,ShanDongJinPu Analytical
Instruments Test Co.,Ltd

Contents

1. Introduction

1.1 Main technical specification

1.2 Main performance characteristics

1.3 Working conditions

2. Keyboard operating instruction

2.1 Panel and keyboard

2.2 Keyboard and display

2.3 Boot-strap

2.4 Keyboard operation

3. Analysis of insulating oil

3.1 Separation process

3.2 Thermal conductivity detector


3.3 Hydrogen flame detector

3.4 Conversion oven

3.5 Choices of operating conditions

Attachment

Ⅰ.Introduction

GC-7890 Power system Dedicated Insulation Oil and Gas

Chromatography Analysis System Manual runs by computer automatic

control, all-keyboard operation and large-screen liquid crystal display

according to the requirements of power system«dissolved gas component

determination in insulation oil (chromatography)» ,and is characterized

by high-precision temperature control, stable and reliable performance,

high sensitivity, good reproducibility and so on.

The analyzer adopts double-pillar parallel shunt system ,equipped

with TCD, double FID and methane conversion oven, and could achieve

total analysis of nine components (H2,O2,N2,CH4,C2H2,C2H4,C2H6,CO,

CO2)of dissolved gas in insulation oil by one time.

The analyzer can be employed for internal fault detection of power

system oil and gas equipments, cooling medium analysis of

hydrogen-cooled generators, impurity analysis of sulfur hexafluoride, the


boiler flue gas analysis, gas analysis and environmental monitoring

analysis. Due to its employment not only for special analysis, but also for

general analysis, the analyzer becomes the most effective gas analytical

instrument in petroleum, chemical engineering, mining and other

systems.

1.Main technical specification

1.1Temperature control

① Temperature range

Column chamber: room temperature+15℃—420℃

Vaporizing chamber: room temperature+30℃—420℃

Conversion furnace :room temperature+30℃—420℃

Thermal conductivity detector: room temperature+30℃—420℃

Hydrogen flame detector :room temperature+30℃—420℃

② Temperature control precision: ±0.1℃

1.2 Thermal conductivity detector

① Sensitivity: minimum detectable concentration of hydrogen≤5μl/L

②Noise: ≤0.05mv

③Drift: ≤0.15mv/30min

1.3 Hydrogen flame detector

①Minimum detectable concentration of hydrocarbon≤5μl/L

②Minimum detectable concentration of CO、CO2≤5μl/L

③Noise: ≤0.05mv
④Drift: ≤0.15mv/30min

1.4 Whole machine start-up time: less than 3 hours

2.Main performance characteristics

2.1 Computer-keyboard operation

This realizes five heating zone temperature control, detector control,

over-temperature protection, "zero" degree protection, air break

protection and other functions, so as to protect the chromatographic

column and thermal conductivity rhenium tungsten filament most

effectively.

2.2 Large-screen liquid crystal display

It could display a large amount of information intuitively and clearly.

2.3 Advanced chromatography process

Based on only one injection and three-detectors analytical process, the

analyzer greatly increases detection sensitivity of hydrocarbon gas.

Double hydrogen flame setting also reduces the detection impact caused

by the gas content differences.

3.Working conditions

3.1 Normal instrument working conditions

①Environmental condition:5℃—35℃

② Relative Humidity:≤80%

③ Power supply:220V±10% 50Hz±0.5Hz


④ 3KW power provided.

⑤ No strong electromagnetic interference, corrosive gas and strong

shock in the surroundings.

3.2 Gas source

① Air purity: ensure the purity of carrier gas and hydrogen is more than

99.99%, the air to be used is purified and air generator function in dust,

oil and water removal.

②Output pressure of steel cylinder relief valve: nitrogen 0.35MPa,

hydrogen 0.25MPa, air 0.35MPa

③ Please replace steel cylinder timely when output pressure fell

below 2MPa.

3.3 Ground wire

To ensure personal safety and instrument performance, the instrument

must be reliably linked to the land. It is suggested to bury the ground wire

in the soil 1 meter below.

ⅡKeyboard operating instruction


指示灯:detector light
功能键:function key
数字键:number key
中文显示屏:Chinese display screen
检测参数键:parameter testing key

The computer-control part of HRSP-9 insulating oil gas

chromatography system exclusive for power system adopts large-scale

integrated circuits, to greatly reduce the circuit-board connectors and

facilitate installation and maintenance. The system uses advanced

software and hardware technology, therefore providing stable and reliable


performance and excellent interference immunity.

Computer controller regulates the temperature of chromatographic

column chamber, vaporization chamber, conversion oven, hydrogen

flame and thermal conductivity detector with high precision, and is

characterized by gas cutting-off protection, over-temperature protection,

"0℃" protection, malfunction causes display and power-off data

protection(with memory function). All controllers zero digitally and

realize true computerization.

1. Panel and keyboard

2. Keyboard and display

2.1 Function keys:

Temperature parameters: used to set and display temperature

parameters. Press to set or change temperature figure of column chamber,

hydrogen flame, vaporization, thermal conductivity and conversion oven

five points. Press Enter key to set five temperature figure circularly. The

maximum figure set is 399 ℃.

Testing parameters: used to set and display the parameters of FID1

amplifier, FID2 amplifier and TCD controller.

Time parameters: used to set and display the output signal switching

time between TCD controller and FID2 amplifier.


Temperature display: multi-function keys. used to display the real-time

temperature on the five points in the normal state; used as circle key in

the alarming state and display the real-time temperature on five points

and alarming causes circularly.

Analysis on: start analysis and workstation.

Analysis off: stop analysis.

Run: used to launch heating and display the temperature figure on five

points. If too low carrier gas pressure in temperature-control state or other

malfunctions are detected while pressing this key, the instrument fails to

launch heating and the heating relay is cut off. At the same time, with the

buzzer buzzing for attention and indicating alarming causes, alarming

light flashes and thermal conductivity cell current breaks. After the

removal of malfunction causes, the heating relay recovers to turn on

automatically.

Stop: stop heating. Thermal conductivity cell current closes.

Enter: shift parameters set for a newline.

Ignition: used for ignition of hydrogen flame detector, press to ignite

automatically.

Timing on: start timing and display time.

Timing off: stop timing and display time.

Bridge current on: used to turn on thermal conductivity bridge current

in the normal state. When detecting too low carrier gas pressure or other
malfunctions, thermal conductivity bridge current and heating relay are

cut off, the buzzer buzzes for attention, and the screen displays

malfunction causes.

Bridge current off: used to cut off thermal conductivity bridge current.

↑ Zero: used to select anyone of the controller's zero and display

positive adjustment. Zero any controller’s baseline and display zeroing

position cooperated with figure keys “1”,”2”. After pressing this key,

press figure key “1” for fine adjustment and figure key “2” for coarse

adjustment. When zeroing touches the limit position, buzzer buzzes all

the time for attention. Press any other key to stop zeroing.

↓ Zero: used to select anyone of the controller's zero and display

negative adjustment. Zero any controller’s baseline and display zeroing

position cooperated with figure keys “1”,”2”. After pressing this key,

press figure key “1” for fine adjustment and figure key “2” for coarse

adjustment. When zeroing touches the limit position, buzzer buzzes all

the time for attention. Press any other key to stop zeroing.

F: used for the functions to be developed

2. figure keys:

"0-9" 10 keys totally, are used to set temperature, time, and other

parameters.

"1" and "2" are also used for the controller zeroing.

3. 5 lights totally on the instrument panel, meaning of the indicator


lights are:

Heating light: The light-on indicates heating running state.

Constant-temperature light: When the real-time temperature set in each

path is all in the range of the corresponding set temperature within ±

0.5 ℃, the light is on.(When the temperature of a way is set 0 ℃, it is

thought to be constant temperature).

Alarming light: When actual temperature in any path is above

corresponding set temperature (set not to 0 ℃)for mare than 15℃ or

carrier gas pressure is too low, alarming light is on and heating relay was

cut off to stop heating. At the same time, buzzer buzzes for attention and

the screen displays the malfunction causes .

Bridge current lights: The light-on indicates that the bridge current of

thermal conductivity detector has been connected.

Analysis light: the light-on indicates that the analysis is in process.

3. Boot-strap

For HRSP-9 insulating oil gas chromatography system exclusive for

power system, each chamber temperature, detector parameter, zeroing

and so on are all set by keyboard and displayed in Chinese characters and

figures. Every parameters have memory function.

Open the main power switch, display shows circularly:

Welcome to use HRSP-9 Power Gas Chromatography

开机液晶显示方框
indicate that self-examination has been completed, the computer is

working correctly and the keyboard can be operated.

4. Keyboard operation:

4.1. Temperature parameters setting:

Press Temperature Parameter ,then display:

设定温度方框

温 度 参 数

柱室=x x x 热导=x x x

氢焰=x x x 转化=x x x

汽化=x x x

Since then, press the figure keys to set temperature on different points,

for example, setting column chamber 95℃. Press 0,9,5, and then press

Enter to shift the cursor at hydrogen flame , in order to set the

temperature of hydrogen flame. Set the temperature required for

vaporization, thermal conductivity , conversion in turn.

4.2 Detector parameters setting:

Press Testing Parameter, then display:


检 测 参 数

灵敏度Ⅰ=X

灵敏度Ⅱ=X

衰减=X

桥 流=XXX

NOTES:

①The sensitivity of FID Ⅰcan only be set to1,2,3,4.

②The sensitivity of FID Ⅱcan only be set to 2, 3, 4, the attenuation can

be set to 0,1,2,3.

③ TCD bridge current is set to 200mA at most and default 200mA is set

for more than 200mA.

4.3 Time parameter setting

Press Time Parameter, then display:

时 间 参 数

T1=XX:XX t1= XX:XX

T2= XX:XX t2= XX:XX


4. 4 Temperature operation:

After the temperature on every point is set, just press Run (heating

light is on at the same time).after the temperature in every path touches

constant temperature, the constant temperature light is on.

When carrier gas is not open or carrier gas pressure is too low on

pressing Run or the gas is cut off during the running, the heating power is

cut off, with alarming light flashing and buzzer buzzing for attention

intermittently. At the same time the screen displays as follows:

载气压力太低!

When the platinum wire in any chamber breaks or displaying temperature

is above set temperature for more than 15 ℃, the heating power is cut off,

with alarming light flashing and buzzer buzzing for attention

intermittently. At the same time the screen displays as follows:

XX 温度超温!

When the platinum wire in any chamber is short-circuited or displaying

temperature is 0℃, the heating power is cut off, with alarming light
flashing and buzzer buzzing for attention intermittently. At the same time

the screen displays as follows:

XX 温度零度!

When carrier gas is not open or carrier gas pressure is too low or the gas

is cut off during the running with over-temperature and platinum faults,

the heating power is cut off, with alarming light flashing and buzzer

buzzing for attention intermittently. At the same time the screen displays

the malfunction causes circularly:

4.5 Stop heating.

Press stop, to stop heating and cut off thermal conductivity bridge current.

And all indicator lights extinguish.

4.6. Current bridge on and off:

When thermal conductivity current is not added ,press Bridge current

on, the following tips are displayed:


XXXmA 桥流确定要加吗?

And then press the key once more. If bridge current indicator light is on,

it indicates that bridge current has been added. If the indicator light of

thermal conductivity bridge current is off and the buzzer buzzes for

attention intermittently, it indicates that the bridge current has not been

added. At this time, the screen displays the malfunction causes. In case of

multiple malfunctions, the screen displays different malfunction causes

circularly, just like pressing Run.

press Bridge current off to stop the thermal conductivity bridge current ,

and the indicator light of bridge current extinguishes.

4.7 Zeroing:

↑ Zero, ↓ zero are both double-functional keys, not only to choose

any controller's zeroing, but also to choose the controller’s zeroing

direction. At this time screen displays: (FID1 zeroing, FID2 zeroing,

TCD zeroing)
氢 焰 调 零 [Ⅰ]

细调

粗调

press"1" for fine adjustment, press "2" for coarse adjustment, and thus

zero amplifier. When zeroing by pressing"1" or "2" , the indicating arrow

shifts in the direction of the arrow at the same time. when adjusting to the

end, buzzer alarms for attention for a long time and press any key to stop.

Note: HRSP-9 insulating oil gas chromatography system exclusive

for power system is characterized by parameter memory function. Once

confirmed, temperature parameter, detector parameter, time parameter

and zeroing result are not influenced by shutdown or power-off. Just

press Run to run the instrument next time.

Ⅲ Insulation oil analysis

1 Separation process

Chromatographic column separation system applies two-dimensional

chromatography technique, adopting double-pillar parallel shunt column.

Samples gas is injected from injection port and introduced into


chromatographic column by carrier gas. In the column box, the carrier

gas is divided into two channels by shunt tee. One enters the GDX-50

column to separate C1 C2 and detect hydrocarbons in the FID Ⅰ.And the

other enters the TDX-01 column to separate H2(O2)、CO、CH4 、CO2

and detect H2 (O2) in the TCD, then enters conversion furnace to

transform CO, CO2 into CH4 and detect CH4 in FIDⅡ.

(refer to page 14 for shunt column system flow chart)

peak-emerging order:

GDX-502 column: CH4 、C2H4 、C2H6 、C2H2

TDX-01 column: H2(O2)、CO、CH4 、CO2

Determined components:

TCD: H2(O2)

FID Ⅰ: CH4 、C2H4 、C2H6 、C2H2

FID Ⅱ: CO、CH4 、CO2

2 thermal conductivity detector

2.1. technical performance

① TCD tank: Semi-proliferation structure, quadri-armed rhenium

tungsten filament, 100Ω in room temperature.

② Power system: constant current source mode power supply, 0-200mA

set, 1mA every gear.

③ Protection system: protecting rhenium tungsten filament by cutting

off constant current source timely on expiration and/or


over-temperature .

2, Points for attention when using thermal conductivity cell detector

When employing thermal conductivity cell detector, it is important to

bear in mind to pump in carrier gas before boot-strap, in order to prevent

the rhenium tungsten filament from being oxidized and destroyed.

Bridge current should not be set too large. When nitrogen or argon is

used as carrier gas, the bridge current should be set between 60mA

and 90mA. When hydrogen or helium is used as carrier gas, the current

bridge should be set between 60mA and 180mA.

The purity of carrier gas should be above 99.99%.Otherwise, the

sensitivity will reduce and rhenium tungsten filament is inclined to be

oxidized, resulting in shortened expectancy. Column must be replaced

after the leak in order to ensure air tightness. Much attention should be

placed on column joints and injection port vaporization pad part.

3. Hydrogen flame detector

3.1 Technical performance

Structure: drum-type collector

Nozzle: corundum nozzle

Ignition: platinum filament automatic ignition

Polarization voltage: 200V DC(direct current)voltage.

3.2 Points for attention when using hydrogen flame detector

①Gas flow and ratio


The ratio of nitrogen and hydrogen: usually between 1:1 to 2:1.

Air flow: usually not less than 300ml/min

② Detector temperature

Generally not lower than 120 ℃, in order to prevent condensation of

water vapor and pollution of combustion products.

③ Aging of the columns

Columns need fully aging, so as to reduce baseline noise and drift

caused by pollution. High temperature aging is recommended before

analysis.

④ Ignition

When the five-way temperature is stable and "constant temperature"

light is on, press the "ignition" button gently. Place bright metal above

ions head, and it demonstrates the success of ignition if there is water

vapor. Or increase the hydrogen flow appropriately.

4 Conversion oven

Separated CO, CO2 enter conversion oven charged with

hydrogen ,transform into methane catalyzed by nickel catalyst and

response in FID Ⅱ.Conversion process is as follows

5 Choices of operating conditions

5.1, Carrier gas flow


Carrier gas flow impact on the time of column separation and

components retention , and column flow ratio recommended.

GDX-502 column flow: TDX-01 column flow = 1:1.9

5.2 Other gas flow

Hydrogen flame burning hydrogen: 25ml/min-30ml/min

Combustion-supporting air:380-450ml / min

Hydrogen in the conversion oven comes from "hydrogen Ⅱ", and

serves as combustion hydrogen in FID Ⅱ.

5.3 Temperature

Column chamber: generally 60 ℃

Thermal conductivity: generally 120 ℃

Hydrogen flame: generally 90 ℃

conversion: generally between 360 ℃ and 380 ℃

5.4, Injection volume

generally between 0.5ml and 1ml

ⅣPoints for attention

4.1. Run the instrument and heat when nitrogen, hydrogen and air are all

pumped in.

4.2. Ignite to increase bridge current after the instrument run and is heated

constant temperature.

4.3. Don’t close gas source until conversion temperature falls below
120℃ after stopping heating.

4.4. Make sure to cut off the gas and keep room temperature when

installing or disassembling chromatographic column, and replace for a

new graphite pad when installing column every time.

4.5. Replace for a new gas injection pad after about 100 times use, and

make sure to cut off the gas and keep room temperature when replacing.

4.6.Perform aging treatment if the emergences of unstable baseline,

peak-emerging tailing or non-ideal peak separation occur. Methods are as

follows:

Set 150℃ of column oven, 180℃ of vaporization, 180℃ of

detection, 0℃ of thermal conductivity, 390℃ of conversion. Press

running button to heat 8 hours! Take care not to increase bridge

current! ! ! Set all the temperature to the normal use temperature to heat

after aging treatment, and then run.

4.7. Check whether the security behind the instrument is open if the

instrument displays nothing or fails to heat after boot-strap.

附页:柱分流柱系统流程图
柱分流柱系统流程图:column shunt system flow chart
载气:carrier gas
氢气:hydrogen
空气:air
热导放空:thermal conductivity emptying
转化炉:conversion oven
稳压阀:pressure maintaining valve
稳流阀:flow maintaining valve
针型阀:needle valve
压力开关:pressure switch
汽化器:carburetor
压力表:pressure gauge