Calibration Report of Quad Tech 1689-9750 RLC Digibridge

MPC Control #: CA3505 Serial Number: 4326715

Asset ID: CA3505 Calibration Date: January 19, 2019

Section 1 - Frequency Accuracy Test


Lower Upper Uncertainty
Function Tested Nominal Value As Found As Left Result
Limit Limit (±)
100.00 Hz 99.9900 Hz 99.9992 Hz 99.9992 Hz 100.0100 Hz PASS 0.006 Hz
120.00 Hz 119.9880 Hz 119.9990 Hz 119.9990 Hz 120.0120 Hz PASS 0.006 Hz
999.99 Hz 999.890 Hz 999.991 Hz 999.991 Hz 1000.090 Hz PASS 0.006 Hz
Signal Frequency 9.9999 kHz 9.99890 kHz 9.99991 kHz 9.99991 kHz 10.00090 kHz PASS 0.00006 kHz
20.000 kHz 19.9980 kHz 19.9998 kHz 19.9998 kHz 20.0020 kHz PASS 0.0006 kHz
49.999 kHz 49.9940 kHz 49.9995 kHz 49.9995 kHz 50.0040 kHz PASS 0.0006 kHz
99.998 kHz 99.9880 kHz 99.9991 kHz 99.9991 kHz 100.0080 kHz PASS 0.0006 kHz

Section 2 - Resistance Accuracy Test


Measure Rate : Slow
Lower Upper Uncertainty
Frequency Nominal Value As Found As Left Result
Limit Limit (±)
1.0000 Ω 0.9988 Ω 1.0001 Ω 1.0001 Ω 1.0012 Ω PASS 0.00006 Ω
10.000 Ω 9.988 Ω 10.001 Ω 10.001 Ω 10.012 Ω PASS 0.0006 Ω
@ 1 kHz 100.00 Ω 99.88 Ω 99.99 Ω 99.99 Ω 100.12 Ω PASS 0.006 Ω
1000.0 Ω 998.8 Ω 999.9 Ω 999.9 Ω 1001.2 Ω PASS 0.06 Ω
10.000 kΩ 9.988 kΩ 9.999 kΩ 9.999 kΩ 10.012 kΩ PASS 0.0001 kΩ

Section 3 - Capacitance Accuracy Check


Lower Upper Uncertainty
Frequency Nominal Value As Found As Left Result
Limit Limit (±)
10.000 pF 9.988 pF 9.999 pF 9.999 pF 10.012 pF PASS 0.011 pF
100.00 pF 99.88 pF 99.99 pF 99.99 pF 100.12 pF PASS 0.11 pF
1000.0 pF 998.8 pF 1000.0 pF 1000.0 pF 1001.2 pF PASS 1.1 pF
@ 1 kHz
10.000 nF 9.988 nF 9.999 nF 9.999 nF 10.012 nF PASS 0.011 nF
100.00 nF 99.88 nF 99.99 nF 99.99 nF 100.12 nF PASS 0.11 nF
1000.0 nF 998.8 nF 999.9 nF 999.9 nF 1001.2 nF PASS 1.1 nF

Section 4 - Inductance Accuracy Check


Lower Upper Uncertainty
Frequency Nominal Value As Found As Left Result
Limit Limit (±)
1.0000 mH 0.9988 mH 0.9999 mH 0.9999 mH 1.0012 mH PASS 0.0002 mH
10.000 mH 9.988 mH 9.998 mH 9.998 mH 10.012 mH PASS 0.003 mH
@ 1 kHz
100.00 mH 99.88 mH 99.99 mH 99.99 mH 100.12 mH PASS 0.02 mH
1.0025 H 1.0013 H 1.0024 H 1.0024 H 1.0037 H PASS 0.0002 H

Confidential: Controlled Document. Page 1 of 2


Cannot be reproduced without the approval of the Quality Department. TQDG-FM-002 rev02
Calibration Report of Quad Tech 1689-9750 RLC Digibridge

MPC Control #: CA3505 Serial Number: 4326715

Asset ID: CA3505 Calibration Date: January 19, 2019

Statements of Pass or Fail Conformance


The uncertainty of measurement has been taken into account when determining compliance with specification, as per ILAC-G8:03/2009.
All measurements and test results guard banded to ensure the probability of false-accept does not exceed 2% in compliance with ANSI/NCSL Z540.3-2006.

The status of compliance with the acceptance criteria is reported as:


PASS - Compliant with specification
FAIL - Not compliant with specification.
FAILZ - The measured value is not within the acceptance and specified tolerance limits. However, a portion of the expanded uncertainty of measurement
at 95% is within the specified tolerance.
The expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k=2, which for a normal distribution
corresponds to a coverage probability of approximately 95%, unless otherwise stated.
This calibration report complies with ISO/IEC 17025:2017 and ANSI/NCSL Z540.3 Method 5-Guard Bands based on Expanded Uncertainty.
- End of Calibration Report -