The calibration report summarizes the results of testing the accuracy of measurements for frequency, resistance, capacitance, and inductance of a Quad Tech 1689-9750 RLC Digibridge with serial number 4326715. All measurements were within specified limits, with uncertainties accounted for, and the device passed calibration on January 19, 2019. Measurement results are presented in tables listing nominal values, measurement limits, results, and uncertainties. The device was found to be compliant with specifications.
The calibration report summarizes the results of testing the accuracy of measurements for frequency, resistance, capacitance, and inductance of a Quad Tech 1689-9750 RLC Digibridge with serial number 4326715. All measurements were within specified limits, with uncertainties accounted for, and the device passed calibration on January 19, 2019. Measurement results are presented in tables listing nominal values, measurement limits, results, and uncertainties. The device was found to be compliant with specifications.
The calibration report summarizes the results of testing the accuracy of measurements for frequency, resistance, capacitance, and inductance of a Quad Tech 1689-9750 RLC Digibridge with serial number 4326715. All measurements were within specified limits, with uncertainties accounted for, and the device passed calibration on January 19, 2019. Measurement results are presented in tables listing nominal values, measurement limits, results, and uncertainties. The device was found to be compliant with specifications.
Lower Upper Uncertainty Frequency Nominal Value As Found As Left Result Limit Limit (±) 1.0000 mH 0.9988 mH 0.9999 mH 0.9999 mH 1.0012 mH PASS 0.0002 mH 10.000 mH 9.988 mH 9.998 mH 9.998 mH 10.012 mH PASS 0.003 mH @ 1 kHz 100.00 mH 99.88 mH 99.99 mH 99.99 mH 100.12 mH PASS 0.02 mH 1.0025 H 1.0013 H 1.0024 H 1.0024 H 1.0037 H PASS 0.0002 H
Confidential: Controlled Document. Page 1 of 2
Cannot be reproduced without the approval of the Quality Department. TQDG-FM-002 rev02 Calibration Report of Quad Tech 1689-9750 RLC Digibridge
MPC Control #: CA3505 Serial Number: 4326715
Asset ID: CA3505 Calibration Date: January 19, 2019
Statements of Pass or Fail Conformance
The uncertainty of measurement has been taken into account when determining compliance with specification, as per ILAC-G8:03/2009. All measurements and test results guard banded to ensure the probability of false-accept does not exceed 2% in compliance with ANSI/NCSL Z540.3-2006.
The status of compliance with the acceptance criteria is reported as:
PASS - Compliant with specification FAIL - Not compliant with specification. FAILZ - The measured value is not within the acceptance and specified tolerance limits. However, a portion of the expanded uncertainty of measurement at 95% is within the specified tolerance. The expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage probability of approximately 95%, unless otherwise stated. This calibration report complies with ISO/IEC 17025:2017 and ANSI/NCSL Z540.3 Method 5-Guard Bands based on Expanded Uncertainty. - End of Calibration Report -