Method 10019

NITRITE, Low Range, Test ‘N Tube (0–0.500 mg/L NO2--N)


Diazotization Method For water, wastewater, and seawater
USEPA approved for wastewater analysis*

1. Enter the stored 2. Press: 63 ENTER 3. Insert the COD/TNT 4. Fill a Test ‘N Tube
program number Adapter into the cell NitriVer® 3 Nitrite vial
The display will show
for nitrite nitrogen holder by rotating the with 5 mL of sample.
- mg/L, NO2-N and the
(NO2 -N), Test ‘N Tube. adapter until it drops Cap and shake to
ZERO icon.
into place. Then push dissolve powder. This is
Press: PRGM
Note: For alternate forms down to fully insert. the prepared sample.
The display will show: (NO2-, NaNO2), press the
CONC key. Note: For increased
PRGM ? performance, a diffuser
band covers the light path
Note: For most accurate holes on the adapter. Do
results, perform a Reagent not remove the diffuser
Blank Correction using band.
deionized water (see
Section 1).

5. Press: 6. When the timer 7. Clean the outside of 8. Place the blank in the
TIMER ENTER beeps, fill an empty the vials with a towel. vial adapter.
Test ‘N Tube vial with Push straight down on
A 20-minute reaction Note: Wipe with a damp
5 mL of sample towel and follow with a dry the top of the vial until it
period will begin.
(the blank). one to remove fingerprints seats solidly into the
Note: A pink color will and other marks. adapter.
develop if nitrite is present.
Note: Do not move the vial
from side to side as this can
cause errors.

* Federal Register, 44(85) 25505 (May 1, 1979).

343
NITRITE, Test 'N Tube, continued

9. Cover the sample cell 10. Press: ZERO 11. Place the prepared 12. Tightly cover the
tightly with the sample in the adapter. sample cell with the
The cursor will move to
instrument cap. Push straight down on instrument cap.
the right, then the
display will show: the top of the vial until it Press: READ
seats solidly into the
0.000 mg/L NO2-N adapter. The cursor will move to
the right, then the result
Note: If the reagent blank Note: Do not move the vial
correction is on, the display from side to side as this can in mg/L nitrite nitrogen
may flash “limit.” See (or an alternate form)
cause errors.
Section 1. will be displayed.
Note: Standard Adjust may
be performed using a
prepared standard
(see Section 1).

Sampling and Storage


Collect samples in clean plastic or glass bottles.
Store at 4 °C (39 °F) or lower and analyze within 48 hours. Warm
to room temperature before running the test.
Do not use acid preservatives.
Remove suspended solids by filtration.
Accuracy Check
Standard Solution Method
Pipet 5.00 mL of a fresh Hach standard, 250 mg/L as NO2--N into
a Class A 250-mL volumetric flask. Dilute to the line with
deionized water to make a 5.00-mg/L intermediate standard.
Pipet 10.00 mL of the 5.0-mg/L intermediate standard into a
Class A 500-mL volumetric flask. Dilute to the line with
deionized water to make a 0.100 mg/L NO2--N standard solution.
Prepare immediately before use.
Run the test using the 0.100 mg/L NO2--N standard in place of the
sample. Results should be between 0.090 and 0.110 mg/L NO2--N.

344
NITRITE, Test 'N Tube, continued
Method Performance
Precision
In a single laboratory, using a standard solution of 0.250 mg/L
nitrite nitrogen and two representative lots of reagent with the
instrument, a single operator obtained a standard deviation of
±0.004 mg/L NO2--N.

Estimated Detection Limit


The estimated detection limit for program 63 is 0.006 mg/L NO2-
-N. For more information on derivation and use of Hach’s
estimated detection limit, see Section 1.
Interferences
Interfering
Interference Levels
Substance

Antiminous ions Interfere by causing precipitation


Auric ions Interfere by causing precipitation
Bismuth ions Interfere by causing precipitation
Chloroplatinate Interfere by causing precipitation
ions
Cupric ions Cause low results
Ferric ions Interfere by causing precipitation
Ferrous ions Cause low results
Lead ions Interfere by causing precipitation
Mercurous ions Interfere by causing precipitation
Metavanadate ions Interfere by causing precipitation
Nitrate Very high levels of nitrate (>100 mg/L nitrate as N) appear
to undergo a slight amount of reduction to nitrite, either
spontaneously or during the course of the test. A small
amount of nitrite will be found at these levels.
Silver ions Interfere by causing precipitation
Strong oxidizing Interfere at all levels
and reducing
substances

Summary of Method
Nitrite in the sample reacts with sulfanilic acid to form an
intermediate diazonium salt. This couples with chromotropic acid
to produce a pink-colored complex directly proportional to the
amount of nitrite present.

345
NITRITE, Test 'N Tube, continued
REQUIRED REAGENTS
Description Cat. No.
NitriVer® 3 Nitrite, Low Range Test ‘N Tube Reagent Set (50 tests) ...............................26083-45
Includes:
(50) NitriVer® 3 Nitrite Test ‘N Tube Vials.......................................................................................................*
Vials, 6 x 100 mm, 6/pkg ................................................................................................22758-06
Caps, for 22758-06 vials, 6/pkg......................................................................................22411-06
Deionized water, 100-mL....................................................................................................272-42

REQUIRED APPARATUS
Quantity Required
Description Per Test Unit Cat. No.
COD/TNT Adapter ..............................................................1 ....................... each....... 48464-00
Test Tube Rack .................................................................. 1-3...................... each....... 18641-00
Pipet, TenSette, 1 to 10 mL..................................................1 ....................... each....... 19700-10
Pipet Tips for 19700-10 TenSette Pipet ...............................1 ................... 50/pkg....... 21997-96

OPTIONAL REAGENTS
Nitrite Standard Solution, 250 mg/L as NO2--N .......................................500 mL ....... 23402-49
Water, deionized............................................................................................... 4 L........... 272-56

OPTIONAL APPARATUS
Flask, volumetric, 250 mL............................................................................. each....... 14574-46
Flask, volumetric, 500 mL............................................................................. each....... 14574-49
Pipet, volumetric, Class A, 10.00 mL............................................................ each....... 14515-38

For Technical Assistance, Price and Ordering


In the U.S.A. call 800-227-4224
Outside the U.S.A.—Contact the Hach office or distributor serving you.