temperatures
regarding the subject of commu- TA TIP
nication, we have now brought
this customer forum into being. It We are constantly asked what temperature is shown, when and where it is
will provide you with an opportu- shown, in what relation and what this means. Several temperatures are im-
nity to communicate with us and portant in thermal analysis.
other TA users regarding an appli-
cation, a question, a wish or a
suggestion for improvement. We TS: effective sample temperature
shall attempt to provide answers TR: effective reference temperature
to questions in the edition follow- TP: set value of the temperature program
ing their receipt. Wishes and sug-
gestions for improvement will be The following temperatures are, however, measured:
collected by us and if possible
implemented in one of the follow- TC: furnace temperature
ing versions. TSM: measured sample temperature
Should you or one of your col- TRM: measured reference temperature
leagues find this magazine inter- ∆T: temperature difference between sample and reference
esting, may we ask you to inform
us by FAX or letter.
The measured temperatures are converted to the effective temperatures us-
We plan to publish the magazine
ing a mathematical model determined by the calibration. The effective
biannually.
sample temperature can be determined very accurately by this means with-
out direct measurement in the sample. As neither the sample temperature
nor the temperature difference are measured with the TG50 and the
Contents TMA40, the sample temperature and the program temperature are identical
here.
TA TiP:
Different temperatures
NEW in sales program:
New TGA850 cell
Software TSW870 V3:
SW option X-terminal
SW option Model-free kinetics
Expansions/Improvements
Trade fairs, symposia
Fax Back
With our software (TSW870 Vx), a If several curves with different tem- Irrespective of the type of represen-
curve can be shown as a function of perature programs are opened, this tation, with our software (TSW870)
time, as a function of time/reference representation is meaningless. it is always possible to have the
temperature, as a function of the In the reference temperature repre- numerical results calculated in terms
reference temperature or the sample sentation, the measured values are of the sample temperature or as a
temperature. plotted as a function of the reference function of the abscissa.
In the simple time representation, no temperature. On the other hand, the following
temperature information is given. In the sample temperature represen- temperatures are shown on the
In the time/reference representation, tation, on the other hand, the curve module:
the curve is also shown as a function is plotted as a function of the sample
DSC820: TC
of time except here the program temperature. Here, the temperature
TGA850: TC
temperature is plotted above the remains practically constant during
TC15: TP (before and after the measurement TC)
time axis. The program temperature the melting process and this is
is taken from the curve which is first shown by the vertical part of the The module control window always
opened in the evaluation window. curve. shows the furnace temperature.