User Com: Different Temperatures

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

USER COM

1
Information for users of May 1995
METTLER TOLEDO thermal analysis systems

Dear customer
The aim of this customer maga-
Different
zine is to provide direct contact
between ourselves and you. After
several approaches from you
temperatures
regarding the subject of commu- TA TIP
nication, we have now brought
this customer forum into being. It We are constantly asked what temperature is shown, when and where it is
will provide you with an opportu- shown, in what relation and what this means. Several temperatures are im-
nity to communicate with us and portant in thermal analysis.
other TA users regarding an appli-
cation, a question, a wish or a
suggestion for improvement. We TS: effective sample temperature
shall attempt to provide answers TR: effective reference temperature
to questions in the edition follow- TP: set value of the temperature program
ing their receipt. Wishes and sug-
gestions for improvement will be The following temperatures are, however, measured:
collected by us and if possible
implemented in one of the follow- TC: furnace temperature
ing versions. TSM: measured sample temperature
Should you or one of your col- TRM: measured reference temperature
leagues find this magazine inter- ∆T: temperature difference between sample and reference
esting, may we ask you to inform
us by FAX or letter.
The measured temperatures are converted to the effective temperatures us-
We plan to publish the magazine
ing a mathematical model determined by the calibration. The effective
biannually.
sample temperature can be determined very accurately by this means with-
out direct measurement in the sample. As neither the sample temperature
nor the temperature difference are measured with the TG50 and the
Contents TMA40, the sample temperature and the program temperature are identical
here.
TA TiP:
Different temperatures
NEW in sales program:
New TGA850 cell
Software TSW870 V3:
SW option X-terminal
SW option Model-free kinetics
Expansions/Improvements
Trade fairs, symposia
Fax Back

USER COM May 95 1


When the different temperatures are plot-
ted as a function of time, the following pic-
ture results for a calibrated cell:

With our software (TSW870 Vx), a If several curves with different tem- Irrespective of the type of represen-
curve can be shown as a function of perature programs are opened, this tation, with our software (TSW870)
time, as a function of time/reference representation is meaningless. it is always possible to have the
temperature, as a function of the In the reference temperature repre- numerical results calculated in terms
reference temperature or the sample sentation, the measured values are of the sample temperature or as a
temperature. plotted as a function of the reference function of the abscissa.
In the simple time representation, no temperature. On the other hand, the following
temperature information is given. In the sample temperature represen- temperatures are shown on the
In the time/reference representation, tation, on the other hand, the curve module:
the curve is also shown as a function is plotted as a function of the sample
DSC820: TC
of time except here the program temperature. Here, the temperature
TGA850: TC
temperature is plotted above the remains practically constant during
TC15: TP (before and after the measurement TC)
time axis. The program temperature the melting process and this is
is taken from the curve which is first shown by the vertical part of the The module control window always
opened in the evaluation window. curve. shows the furnace temperature.

A curve and its four possible representations

2 USER COM May 95


The new TGA850 thermogravimetric
module is here!
The TGA850 module is available
from May and expands the thermo-
gravimetric product range in the top-
of-the-line performance segment. As
with the DSC820, a modular concept
has been selected here. This allows
you to assemble the instrument con-
figuration optimum for your needs.
An expansion (e.g. with sample ro-
bot) can also be implemented at a
later date.
The measuring cell incorporates
state of the art METTLER TOLEDO
weighing technology in the form of
the ultramicro balance. You can thus
rely on the results. With the SDTA
operating possibility, you can gener-
ate the DTA curve following the The features of the TGA850 module at a glance:
measurement. The SDTA (Single • large measurement range 2g
Differential Thermal Analysis) sig-
• high resolution 0.1 µg
nal is calculated from the measured
• minimal noise (RMS) < 1 µg
sample temperature and is a standard
• parallel guided balance no positioning error
feature. This allows the analysis of
• temperature range room temperature –1100 °C
effects which do not cause mass
• high temperature accuracy +/– 0.25 °C
changes. It can be seen simply and
• horizontal furnace minimum disturbing influence of
quickly whether the reaction is exo-
the purge gas
thermic or endothermic.
• gas-tight measuring cell definable measurement environment
• SDTA measurement physical transformations are visible
With the optional sample robot, you
• coupling possibility (TG-MS) analysis of decomposition products
can increase your productivity enor-
• modular expansion open for the future
mously. At the highest level of auto-
• computer control simple operation using the latest
mation, sample feed, performance of
computer software
the measurement, evaluation and
result assessment up to the color
printout are all undertaken automati-
cally. As with the DSC820, the sam-
ple robot can be loaded with 34 sam-
ples. Each sample can be assigned
not only a different method but also
a different evaluation.
If need be, the crucible can be fitted
with a lid and positioned on the sam-
ple turntable. Depending on the
method, the lid can be removed by
the sample robot immediately before
the measurement. As a result, the
sample exchanges virtually no mois-
ture with the surroundings between
weighing in and performance of the
experiment.
Measured TGA Curie and SDTA melting temperatures

USER COM May 95 3


Further improvements
The TSW870
TA software – TA basic version
As a result of various suggestions Calibration of the heat flow has been
yet another step from our circle of customers, we have
made many improvements. The vari-
extended so that it can be performed
analogous to the temperature cali-
forward ous windows can now be started in
the following languages:
bration with several different sub-
stances.
English Further, a total calibration has been
The latest version V3 is available German implemented which in the case of
from May. In addition to two further French DSC and TGA determines all param-
options, X-terminal and model-free In addition to the US keyboard, eters by means of a single measure-
kinetics, the basic version and the Swiss, French and German keyboards ment. This allows simple and rapid
other options have been expanded are supported. To facilitate manual calibration of the modules. For the
by new possibilities. calculations, you can now power up greatest accuracy, you can naturally
an on-screen calculator. still use the single calibrations.
1st option: X-terminal operation
On an X-terminal (data input station
with mouse, keyboard and screen),
a UNIX PC (e.g. TAS810) or an
MS-Windows PC with X-terminal
emulation, additional evaluation
windows can be opened. For the first
time ever, this allows several per-
sons to work simultaneously with
one and the same software (and also
with the same database!). Only such
a feature allows genuine multiuser
operation.

2nd option: Model-free kinetics


The new way to calculate kinetics is
based on the theory of the Russian
scientist Dr. Vyazovkin and makes it
is possible for the first time to per-
form kinetic evaluations and draw
corresponding conclusions without
previous selection of a physico-
chemical model. Until today, the Here, three TG stages are evaluated collectively
model, usually nth order, had first to
be chosen and then the measurements
used to calculate the parameters.
At the very best, the results were only
as good as the degree to which the
chosen model reflected reality.
With the model-free kinetics, the
step which is associated with errors,
namely the model selection, has
been dropped. In addition, for the
first time complex reactions can also
be investigated very simply. These
kinetics are suitable for both DSC
and TG measurements.

Epoxy resin curing not nth order

USER COM May 95 4


New printing possibilities allow the SW option algorithm has been accelerated many
printout of a window or the entire Automatic evaluation times so that with curves based on
screen contents in black and white Following feedback from a cus- many measured points the envelope
on the HP DeskJet. tomer, the Eval macro now contains curve can be plotted within a few
Calculated curves still retain the in- information on whether limits within seconds.
formation of the original experiment a definable range should set auto- The manual plotting of lines (or base-
in V3.0. Consequently, the database matically or fixed. If a reaction does lines) has been made more user
relations also exist for such curves not always appear at the same place, friendly. You can now draw the
and search operations (e.g. by meth- the expansion still allows fully auto- baseline in the manner of a rubber
od or user) are possible. matic evaluation. A wide range with- band and correct it by moving back
A new feature is the deletion of entire in which the evaluation should be the pointer.
«trees» in the database. In other automatically sought is simply
words, you can not only delete sin- defined in the Eval macro. As the SW option TGA
gle objects as before, but entire, effects which appear are not always With the TGA option, the content
related data structures (e.g. every- of the same magnitude, the plot is calculations have been expanded by
thing that user X has generated) automatically scaled on the ordinate the drying loss. As a result, the dry-
→ methods → experiments and plotted exactly in the aspect ing loss, which appears between
→ (curves) → evaluations ratio which the evaluate window had weighing in and performance of the
Whether curves should be deleted or on generation of the Eval macro. experiment, is included in the first
not can be decided for each individ- mass stage and, depending on the
ual case. SW option DOS Data import settings, reported as such.
The DOS data import has been ex-
The evaluation window has been panded. In addition to the TA72 files, We hope that these expansions and
expanded by many minor but useful mass spectrometer files from Balzers improvements will facilitate your
possibilities. With a double-click on (MS Cube) can be imported. work with the software.
the ordinate you can switch between
the relative and absolute representa- SW option Mathematics
tion. You can now also use the nu- The Mathematics option has been
meric key block. The peak tangents expanded by the Fourier analysis
and the hatching can be switched off evaluation. This can be used for an
in the integration via the settings. even simpler analysis of periodic For further questions, please contact
With horizontal step, the inflection curves (DLTMA or ADSC – alter- your local METTLER TOLEDO
point is also calculated. nating DSC). The envelope curve company.

5 USER COM May 95