adloc-TENSIOMETER GEONEXT

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

FOR TENSION TRANSDUCER TOTCO TLC 100K DEFAULT VALUES

ARE THIS THE VALUES TO ENTER THE CALIBRATION :

4 mA = 0 ton
20mA = 45.36 tons

20 – 4 = 16
45.36 / 16 = 2.835 tons / mA METRIC units

4 mA = 0 lbm
20 mA = 100001.68 lbm API units
1. Suppose you scan 5.754 mA when the drillers WOH is 22.72 tons (Suppose
W.O.H. on slips)

5.754 – 4 = 1.754
1.754 x 2.835 = 4.97 ton

So… AdLoc Assistant :


2. Suppose you scan 9.715 mA when the drillers WOH is 80 tons (Suppose W.O.H. off
bottom)

9.715 – 4 = 5.715
5.715 x 2.835 = 16.21 tons

So… AdLoc Assistant :