Internal Calibration Check HPLC System (Agilent Prime LC 1260 Infinity II)
Internal Calibration Check HPLC System (Agilent Prime LC 1260 Infinity II)
: 01
AVITECH
Nutrition (P) Ltd. HPLC SYSTEM Issue Date: 07-09-2018
Revision No.: 01
(Analytical Division) (Agilent Prime LC 1260 Infinity II)
Sr. No. Set flow rate Theoretical Actual Actual flow Limit
Seconds seconds(Y) rate
1 1.0 ml/min 600 second 0.990 to1.010
2 1.5 ml/min 400 second 1.485 to 1.515
3 2.0 ml/min 300 second 1.980 to 2.020
Equation: 10 X 60
Y
(2) RETENTION TIME REPRODUCIBILITY: For DAD
Take …………….mg of Caffeine and dilute to …………. ml with mobile phase. Further dilute 1 ml to
100 ml with mobile phase (final concentration 10ppm).
1
2
3 % RSD not more than 1.0
4
5
Mean
Calibrated By Checked By
Analyst/Sr. Analyst Sr. Manager (QC)
SD
% RSD
1
2
3 % RSD not more than 2.0
4
5
Mean
SD
% RSD
Calibrated By Checked By
Analyst/Sr. Analyst Sr. Manager (QC)
1
2
3
4
5
Correlation coefficient NLT 0.990
Sr. No. Mean area of solution Area of caffeine in % Carry Over Limit
blank injection NMT 0.20%
Wave 265 266 267 268 269 270 271 272 273
length nm.
Absorbance
Limit LIMIT 272 ± 2 nm
Wave length 274 275
nm
Absorbance
Limit
Calibrated By Checked By
Analyst/Sr. Analyst Sr. Manager (QC)
m.AU % of m.AU % of
composition composition
C100
79.0 to 81 %
C80
59 to 61 %
C60
39.0to 41.0 %
C40
19.0to 21.0 %
C20
Calibrated By Checked By
Analyst/Sr. Analyst Sr. Manager (QC)
(9) RETENTION TIME REPRODUCIBILITY: For FLD
Take ……………..mg of Anthracene and dilute to …………….ml with 100% Acetonitrile. Further
dilute 1 ml to 100 ml with mobile phase.
1
2
3 % RSD not more than 1.0
4
5
Mean
SD
% RSD
1
2
3 % RSD not more than 2.0
4
5
Mean
SD
Calibrated By Checked By
Analyst/Sr. Analyst Sr. Manager (QC)
% RSD
3 60 ° C 59 to 61 ° C
Remark: Instrument meets/does not meet with acceptance criteria. The calibration is OK / Not OK.
Calibrated By Checked By
Analyst/Sr. Analyst Sr. Manager (QC)