Medical Devices Safety Enhancement and Performance Improvement


Through a Periodic Calibration Program

Chapter  in  IFMBE proceedings · May 2019


DOI: 10.1007/978-981-10-9038-7_10

CITATIONS READS

0 27

3 authors, including:

Ali Tavakoli Golpaygani Mm Movahedi


Standard Research Institute Shiraz University of Medical Sciences
24 PUBLICATIONS   56 CITATIONS    23 PUBLICATIONS   129 CITATIONS   

SEE PROFILE SEE PROFILE

All content following this page was uploaded by Ali Tavakoli Golpaygani on 11 September 2018.

The user has requested enhancement of the downloaded file.


www.jbpe.org Technical Note
Why Should We Have a Periodic Safety Department of Biomedi-
1

cal Engineering, Stand-


and Performance Program for Medical ard Research Institiute,
Karaj, Iran
Devices
Tavakoli Golpaygani A.1

ABSTRACT
Nowadays, more than 10,000 different types of medical devices can be found in hospi-
tals.These devices used in medical centers and hospitals for monitoring and treatment
of patients require periodic safety and performance checking in order to have confi-
dence in their functioning and operation. Physicians need better accurate medical mea-
surements in order to better diagnose diseases, monitor patients and deliver treatments,
in this way failure to ensure appropriate measurements will certainly have diverse ef-
fects. Safety and performance testing of medical devices in the medical sector is a one
of the key factor in improving public health. Acquiring results of some investigations
indicate a need for new and severe regulations on periodic performance verifications
and medical equipment quality control program especially in highrisks instruments.
The metrological reliability of four high risk medical devices, Electrosurgical unit,
Defibrillator, Syringe pump and Infant incubator in use some hospitals (privates and
publics) in one of the province of Iran according to international and national standards
was evaluated. Quantitative analysis of Some parameters that impact the safety and
performance showed the amount of the obtained results in some equipment are in criti-
cal range and have higher values than standard limitations. General electrical safety
evaluations for measuring the patient leakage currents and patient auxiliary currents
carried out for all of groups,in some cases the amount of leakage currents were over
the standard limitations.
Acquiring results indicate a need for new and severe regulations on periodic perfor-
mance verifications and medical equipment quality control program especially in high
risk instruments. It is also necessary provide training courses for operating staff in the
field of meterology in medicine and what’s the critical parameters and how they can
get good accuracy results equipment.

Keywords
Metrology, Medical devices, Safety, Performance, Calibration

Introduction

N
owadays, electrical devices are being employed in a wide va- *Corresponding author:
riety of fields in medical sciences with different effects. Ac- A. Tavakoli Golpaygani
cording to the importance of this matter in relation to human Faculty of Electrical, Me-
chanical and Construc-
health, periodic verifications, safety and performance testing of medical tion Engineering,
devices in the form of quality control tests and calibration are necessary. Standard Research Insti-
This issue has started and noticed seriously since the past decade for tute, Karaj, Iran
E-mail: tavakoli.golpa@
medical devices. gmail.com
Even the best medical measuring instruments and equipment are af- Received: 14 April 2017
fected by wearing and tearing, environmental changes and other fac- Accepted: 20 July 2017

J Biomed Phys Eng I


Tavakoli Golpaygani A www.jbpe.org

tors, losing accuracy over time. According to critical medical devices are safe, accurate, re-
the importance to human health, systematic liable and operating at the required level of
and periodic checking of test and measuring performance. For instance the results of four
instruments, national and international stan- investigations of electrical safety and metro-
dards on a regular schedule are very essential logical reliability for four important medical
for reliable measurements. This process has devices with a high applicability in treatment
many advantages such as maximizing effec- process are presented. In each of these studies
tiveness and minimizing risk for patients and at least twenty equipment with different mod-
operators. els were used at some publics and private hos-
Safety and performance testing are required pitals were tested. A schematic of each device
to ensure the results and reports which are ac- is presented in Figure (1).
curate , reliable and safe to deliver treatment
and diagnose process. The accuracy of the 1A Case study on Electrosurgical
instrument readings provides how the medi- Equipment
cal devices can generate test results which are Electrosurgical units (ESU) are the most
very close to the true value. For example, in common type of electrical equipment in the
many illnesses such as hypertension symp- operating room. Electrosurgical unit is one
toms; a diagnostic process starts with only of the most effective surgical instruments,
one physiological parameter (blood pressure) through the better control and the management
measurement. In the other side in many situ- of complications during surgery, it improved
ations, the initial stage of disease diagnosis patient safety. The principle of electrosurgery
starts with laboratory testing of different sam- is inducing heat by high frequency electrical
ples from patients (such as blood, urine, …) currents. Coagulation, cutting, desiccation and
,then, the laboratory output is only some ana- fulguration of biological tissue are different
lytical data. These data are used in determin- effects of electrosurgery function. The correct
ing whether a patient has a disease or not. We operation of electrosurgical unit is essential to
expect a high quality service from every clini- ensure patient safety. In this case study, quan-
cal laboratory, which all test results produced titative analysis of HF leakage current mea-
by all laboratories at all times should be accu- surements, showed the amount of the obtained
rate and clinically meaningful. Medical error results in many units is in critical ranges and
measurements made by the medical devices has higher values than standard limitations es-
tend to shift the test results from the true val- pecially in ground referenced generators.
ue. Thereforean error can cause under-detec- The amount of earth resistance of hospitals
tion or over-detection of diseases. Physicians or medical centers has important effects on HF
decided based on the clinical results obtained leakage currents of electrosurgical equipment.
from the different measurements, if the results High earth resistance causes high HF leakage
are not correct and accurate, it might cause the current, as a result, the quality of earth sys-
physician to under-dose or over-dose the pa- tem of medical centers has a key role in safety
tient with incorrect prescriptions. [1, 2]. performance of an electrosurgical unit. The
achieved results in this case study represent
Material and Methods that HF leakage current of ground referenced
Recently, it has been seen from 5,000 to more generators are more than isolated generators.
than 10,000 different types of medical devices Therefore, the use of isolated generators in or-
in an average to large sized hospitals. Hospi- der to reach to high level of patients’ safety is
tals and medical centers must ensure that their preferred in medical centers [3].
II J Biomed Phys Eng
www.jbpe.org Periodic Safety and Performance Testing of Medical Devices

Figure 1: A schematic of medical devices in case studies

2A Case Study on Defibrillators lators equipment in the use at some hospitals


Equipment were evaluated. Quantitative analysis of ac-
Medical research and statistical analysis all curacy of delivered energy measurements,
over the world show generally over 135,000 represented the amount of the obtained results
people die annually following acute myocar- in many units are critical and have less value
dial infarctions as ventricular fibrillation (VF) over the standard limitations, especially in de-
or ventricular tachycardia (VT), and the only vices with poor quality battery. The delivered
effective treatment for which is early defibril- energy of defibrillators especially in frequent
lations. Defibrillation is the application of a discharge is dependent seriously on the power
predefine electrical current through the myo- battery condition. Only some devices deliv-
cardium, this phenomenon causes synchro- ered acceptable output accuracy of delivered
nous depolarization of the cardiac muscle. energy and the accuracy in poor quality bat-
This action terminates the arrhythmia, and al- tery conditions, after activation of discharge
lows normal sinus rhythm to be reestablished. alarm were low.
In this case, study on the performance and General electrical safety evaluations for
safety testing of frequent use, manual defibril- measuring the patient leakage current and pa-

J Biomed Phys Eng III


Tavakoli Golpaygani A www.jbpe.org

tient auxiliary leakage current carried out for tient auxiliary leakage current carried out for
all of the under test devices. In some cases all of the under test devices. In some cases,
the amount of leakage currents were over the the amount of leakage currents was over the
standard limitations. As a technical investiga- standard limitations.
tion, it is shown that the earth system quality Due to design and manufacturing technolo-
of hospital has a role key in electrical safety gy development of the pumps, it is required to
test and it is necessary to evaluate the quality consider supplementary professional training
of hospital earth system [4]. courses for the users to enable them to use all
available pump options and reduce the human
3A Case Study on Infusion Pumps errors [5].
Equipment
Infusion devices are used extensively in 4A Case Study on Infant Incubators
clinical settings and patients’ homes as essen- Equipment
tial tools to provid critical care, perioperative Today’s, the most of the neonatal or new-
care and management of pain. It is estimated born babies were put in infant incubators for
that 80% of hospitalized patients receive intra- overcoming the distress during the post-natal
venous (IV) therapy. Because infusion pumps care. These incubators can be defined as an
are frequently used to administer critical flu- optimal cabin environment with electronic
ids, including high-risk medications, pump equipment supplying the premature or babies
failures can have significant implications for with illness problems to be able to continue
patient safety. Hence, the reliability of infu- their vital functions. It shows more than 80%
sion pumps is extremely important. In the of neonatal babies are hospitalized in infant
third study, quantitative analysis of flow rate incubators and are treated there over a period
accuracy measurements and represented the of 3 days to 7 months. In the fourth study, a
amount of the obtained results in some units quantitative analysis of temperature, humidity
are critical and have inaccurate values, es- accuracy measurements, the stability of tem-
pecially in equipment with inappropriate IV perature, represented the amount of the ob-
set. For example, it has been seen, the usage tained results in some units are critical and are
of a wrong or nonstandard syringe or tub- not in standard limitations. Stability of tem-
ing set, which is not made based on the ap- perature is an important factor, some incidents
proved technical specifications, increases the such as apnea can result from variations in the
occurred error percent and the inaccuracy 10 incubator temperature. In some devices with
to 20 percent. The position of drip detector is repair history have been seen that selected and
an important factor in infusion pumps. If the replaced temperature sensors have not had re-
detector is installed near the injection location, quired accuracy as defined specifications in
then the error percent can increase up to 25%. manufacture’s documents or manual.
The reduction of battery power is one of the General electrical safety evaluations for
effective factors on the performance of the in- measuring the patient leakage currents and pa-
fusion pumps. It has been seen that the low tient auxiliary currents were carried out for all
quality batteries and the batteries which are of the under test devices. In some cases the
sensitive to the improper voltage could reduce amount of patient leakage currents was over
the outflow rate between 10 to 30 percent. the standard limitations, especially in two un-
General electrical safety evaluations for der test devices with recoiling of motor’s fan,
measuring the patient leakage current and pa- the patient leakage current and the noise level

IV J Biomed Phys Eng


www.jbpe.org Periodic Safety and Performance Testing of Medical Devices

of fan were more than limited amounts [6]. access for operators and users.
3. The benefits of a calibration service is un-
Discussion known and in many situation is regarded as a
Safety and performance testing of medical purchased document.
devices in the medical sector is a one of the So, it is also necessary to provide training
key factor in improving public health. For the courses on the fundamental of operation and
best confidence in their functioning and op- performance parameters for medical staff in
eration, the medical devices used in medical the field of metrology in medicine and how
centers and hospitals for monitoring and treat- they can get good accuracy results especially
ment of patients require periodic safety and in high risk medical devices.
performance checking. Physicians need bet- Finally, it is obvious, responsibility, educa-
ter accurate medical measurements in order to tion, commitment, management support and
diagnose diseases better, monitor patients and sufficient resources are the strongest drivers
deliver treatments. for developing the successful culture of main-
Even an error in a simple medical device tenance and calibration in technical medical
such as clinical thermometers to measure body management.
temperature, sphygmomanometers to measure
blood pressure, weighing scales to measure Acknowledgment
body weight etc. can cause medical decision The author would like to express his thanks
making to shift from the normal decision. In to all experts participated in the study from
some condition, medical errors might cause biomedical engineering department of hospi-
patients pain bearing permanent or temporary tals and medical centers.
disability and in some state even cause death.
In another view, this matter can cause increas- Conflict of Interest
ing health costs. Increasing the number of fol- None
low-up tests, medication and treatment for a
patient who has been incorrectly diagnosed to References
have a certain disease will be incurred health  1. Khandpour RS. Handbook of Biomedical
costs. Instrumentation. New York: Mc Graw Hill;
2014.
Conclusion  2. Noblitt J. Introduction to Biomedical En-
Acquiring results in four discussed case gineering Technology. Biomed Instrum
studies indicate a need for new and severe reg- Technol. 2008;42:446.
ulations on periodic performance verifications  3. Tavakoli Golpaygani A, Movahedi MM,
and medical equipment quality control and Reza M. A Study on Performance and
calibration. This matter is more significant in Safety Tests of Electrosurgical Equipment.
high-risks medical devices. In many hospitals, J Biomed Phys Eng. 2016;6:175-82.
there are some deficiencies related to the op- PubMed PMID: 27853725; PubMed Cen-
erator and users of medical devices, some of tral PMCID: PMC5106550.
them are noticed below.  4. Tavakoli Golpaygani A, Movahedi M, Reza
1. They thought that a calibrated device M, Hassani K. A study on performance
works without any errors for a long time and and safety test of infusion pump devices.
no need recalibration in defined intervals. 2017.
2. The calibration certificates are kept in the  5. Tavakoli Golpaygani A, Movahedi M, Reza
outside of the operating area and difficult to M, Hassani K. A study on performance

J Biomed Phys Eng V


Tavakoli Golpaygani A www.jbpe.org

and safety test of infusion pump devices. and Reza M.“A study on performance and
Biomedical Research. 2017;28. safety test of infant incubators devices” J.
 6. Tavakoli Golpaygani A., Movahedi M.M. Of Clinical Engineering. 2018, In Press.