XK3100 SERIES
INDICATOR
Catalog
CHAPTER 1 Summary
The XK3100-B2 Indicator is a new product of our company. It is a kind of general weighing
indicator. It adopted the AD IC of famous semiconductor factory in USA. It has high precision,
high stability, high credibility, zero code adjust and widely of the input signal, upper and lower
limit setting and alarm, peak test (selectable). The indicator can be made many kinds of
electronic scale connected with load cells easily such as platform scale, ground scale, vehicle
scale, oil scale and animal scale. It is a perfect indicator which can be used in business
industry and etc.
Main Function of XK3100-B2:
1. With AC/DC power supply, 6V 4Ah no-maintenance battery and high efficiency recharging circuit
2. Low power cost, can be used for 200 hours after recharging(only with one load cell and
not open the backlight)
3. Auto adjust of zero code (tare automatically), input signal range: - 6 mv ----- +21 mv
4. Calibrate with digital keyboard (locked by hardware and software)
5. 6 digits LCD display, 24 mm high of symbol, backlight can be selected Zero auto track
6. Several divisions and point selecting, calibrate with any point
7. Low voltage check, calibration self-check, overload warning and error information
CHAPTER 3. Installation
1. The front of indicator
Keyboard
Display
Figure 3-1
2. The back of indicator
Power plug
Figure 3-2
3. Load cell connected with indicator
-V -IN
+IN
Figure 3-3
Note: +V, -V (1, 2): Load cell’s excitation
+S, -S (1, 2): Load cell’s feedback signal
+IN, -IN (3,4): Load cell’s output signal
Shield (5): the load cell’s shield cable (connected with GND of AC plug)
E-105
CHAPTER 4. Calibration
When you calibrate the indicator, the load cell’s shield cable must connected with GND,
otherwise the indicator reads will have errors. (Figure 3-3)
Calibration should be operated by measure technician is as follows:
1. Weight calibration
STEP OPERATION DISPLAY NOTICE
Power on, press “calibration” and “ Enter” together when self-checking, loosen when hear the
1
buzzer.
2. Press Tare If you want to quit without save settings. It keeps the old.
3. Power off the indicator after calibration, check the data have been saved or not. If not
2. Calibration parameter
2. The indicator use currency RS232 UART signal every 10BIT is a data frames, the format of
frames as follows
Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit
1 2 3 4 5 6 7 8 9 10
BIT1 start bit
BIT2~BIT9 data bit
BIT10 stop bit
CHAPTER 6. Operation
1. Keyboard instruction:
(1) Cali. To modify some calibration parameter.
(2) Mode To select function: Weighing, Counting & percentum.
(3) Unit To select unit: kg or lb.
(4) Set To select up and low limits range in weighing model.
To sample the weight in counting and percentum model.
(5) Zero To make weight zero when nothing on the platform but the display is not zero.
Press this key 2-4 seconds, the backlight will be switched on or off.
(6) Tare To delete the tare.
(7) ↑ To input number.
E-105
2. Operation instruction:
When power on,the indicator will be in weighing model. The “Zero” flag and “kg” flag
disappear. You can press Mode to select model: weighing, counting, percentum.
Tare:
Press Tare to make the weight window display “Zero” after the weight is stable. The “Tare”
flag
disappears. The weight is Net. Remove the poise, the displays is negative. Press Tare again,
the displays is zero. The “Tare” flag disappear.
Note: (1). Tare function can set and set continuously only the weight is over zero and stable.
(2). Tare range: 99999 (if the display data is over 99999, not include the decimal, tare
function is ineffective)
Zero:
Press Zero to make display zero.(Zero range: <=4%FS, it is invalid under tare state.)
Weighing mode
Select Unit: Press Unit to select the unit between “kg” and “lb”. The corresponding flag
disappears.
Counting mode
Press Set to sample weight in the counting model. It displays “SAP X” (X is the sample
number).
1. Put enough samples on platform and press ↑ to select samples number: 10, 20, 50,100 &
200,500,1000. Press Enter confirm.
2. The window displays “LOAD-C” after the sample number is selected. Press Enter to finish
sample.
3. There are two instance of unit weight lacking:
a. When the weight window displays “-SLAC-”, it means the unit weight of sample is less
than 80% of division. You still can count but the counting may be inaccurate. The signal
will disappear after about 3 seconds.
b. When the weight window displays “-CSL-”, it means the unit weight of sample is not
enough at all. Please press Set to re-sample. Press Enter to exit and return weighting.
Press Set again it will exit from the sample state.
Percentum mode
1. Press Set in the percentum state to display “LOAD-P”. There are two ways to sample:
a. Put enough samples on the platform and then press Enter to finish sample.
b. Press Unit then the scale displays “000000”. Press ↑ and Enter to input the weight
value.
E-105
2. When the sample weighed less than 0.1% FS., the window displays “-CSL-”. It means the
sample is not enough. You should re-sample. Press Set to re-sample. Press Enter to exit
return weighing.
Press Set again it will exit from the sample state
Caution: The up and low limit range and alarm ways will be saved, different unit VS different
alarm range and ways.