Professional Documents
Culture Documents
Perkinelmer Aanalyst 300: Location: Friedland 202
Perkinelmer Aanalyst 300: Location: Friedland 202
Perkinelmer Aanalyst 300: Location: Friedland 202
The AA-300 is equipped with an automatic six-lamp turret and is programmed to perform sequential multielement analysis. The AA-300 is
equipped with lamps for the following elements: Ag, Al, As, Ba, Ca, Cd, Cr, Cu, Fe, K, Na, Mg, Ni, Pb and Zn The Iron(Fe) and Copper(Cu) lamps
are the most commonly used.
by Golda Yelaun
1. Check the location of exhaust vent (should be located at least 4 meters from the Flame).
2. Check the pressure of the AA Gas Purifier assembly. Make sure the output pressure is between 51 and 65 psig. ( on the wall , behind the
instrument).
3.
3. Turn on the proper gas supplies and set pressures, recommended in the handbook.
by Golda Yelaun
1. With the Flame still burning aspirate deionized water for five minutes.
2. In the Flame Control window, click on Flame off.
3. Turn off the burner gas supplies to the spectrometer.
4. Exit the software, including the main application and the Windows software.
5. Switch off the Spectrometer.
6. Switch off the computer.
7. Empty the burner drain vessel. Dispose the waste solutions according the Safety Regulations.
Instrument Calibration
1. Curve correction: Quantitative measurements in Atomic Absorption is based on Beer’s Law, which states that concentration is
proportional to absorbance: C = kA. The Perkin-Elmer AA-300 allows either linear or non-linear calibration using multiple calibration
standards.
2. Selecting the number and concentration of standards: Selecting the number and the concentration of calibration standards is very
important. If the analyte concentration of all samples to be analyzed falls within the linear range, one calibration standard should be used.
The top of the linear range for the most elements is between 0.20 and 0.30 absorbance units. If the analyte concentration in the samples
is expected to exceed the linear range, either two or three calibration standards should be used.
3. How to select calibration standards properly:
a. Make up three standards: The first one (S1) should be on the top of the linear range. The concentration of the second standard
(S2) should be 3X the concentration of the first. The concentration of the third standard (S3) should be 6X the concentration of
the first standard (S1) (6:3:1 rule).
b. Preparing the calibration standards: Calibration standards are prepared from certified stock Standards for calibration. A variety of
stock standards are usually provided at concentration level at 1000mg/l, and are available directly from Perkin-Elmer as well as
from other commercial sources.
4. Creating a calibration curve:
a. Set up the spectrometer in the concentration mode. Auto zero on Blank.
b. In the Manual Analysis Control window click Analyze Blank and then Analyze Standards.
c. Click on Analyze Sample.
d. Check the calibration curve. If the calibration curve in the linear range and some standards appear to be off it, check them and
d.
by Golda Yelaun
1. Preparing the calibration standards. Calibration standards are prepared from stock solution 1000 PPM Fe (Atomic Absorption standard).
We prepared three calibrations standards: 2ppm, 5ppm, 10ppm.
2. In the Manual Analysis Control window we have to click Analyze Blank, then Analyze Standard.Since the Method is already stored all
three standards (2,5,10 ppm) will be analyzed.
3. The instrument will create the plot Absorbance versus Fe-concentration automatically.
4. The calibration curves have to be linear. If any of the standards appear to be off the calibration curve, we have to check the standards
and repeat the calibration.
Before analyzing prepared samples we have to put QC sample and Blank for checking the instrument.
1. Analyzing Reagent Blank (absolute value have to be in the Range 0.2 mg/l)
2. Analyzing QC standard- 5ppm Fe (have to be closed to 5mg/l)
3. Analyzing prepared samples.(Concentration of the samples have to be in the Linear Dinamic Range – less than 10 ppm Fe)
4. To the end of the Run (around 10 samples) to check again Reagent Blank and 5ppm Fe.
Web Resources
Misc AA300 PerkinElmer info