Set the ventilator standby, then press the Config key, the Configuration submenu appears
on the screen, you may find the Calibrat button. See figure1.
Turn the knob to select Pressure sensor and press it, see figure. 6
Calibrating procedure refer to Calibrating Flow sensor.
To calibrate flow valve when there be a difference of 30% between the value of tidal
volume monitored and the settings.
Calibrate inspiratory valve
Short connect the inspiratory port and expiratory port ,see figure 7
Figure7 Calibrate inspiratory valve
Avoid leading to the system failure because of inadvertent operation, only inputting proper
code can perform next step. Then follow the on-screen prompt strictly to calibrate flow
valve.
Input password:7294 to calibrate Inspiratory valve, set AIR and O2 0, press OK. (seldom use). It
will take about 5 minutes.
Calibrate expiratory valve
Connect patient circuit and test lung(splint simulate lung) as follow: