Professional Documents
Culture Documents
Energie & Process Report Pitot Tube (Mechanical Engineerig
Energie & Process Report Pitot Tube (Mechanical Engineerig
Energie & Process Report Pitot Tube (Mechanical Engineerig
Ø tube 42.6 mm
Measurement Station
3. Difference
60 476 123 479 282 %
2. Write the hand calculations for one flow out (note the number dialed above calculation)
ꝋ v = √ (2 ∆p)/ (ρ)x Atube ∆P = ½ ρv2
Tube Diameter: 42.6 mm= 0.0426 m
Atube= π x r^2
Atube= π x (0.0426 x ½ )^2 = 0.001425 m2
ρ = 1000 kg/m3 V= 5 m/s
∆P= ½ x 1000 x (5)^2 = 12500 Pa
ꝋ v= √(2 x 12500)/ (1000) x 0.001425= 0.00713 m3/s
ꝋ v = α A √ (2 ∆Pm)/ (ρ)
α= 0.8 A= 0.00088 ∆Pm= 20000 ρ = 1000 kg/m3
ꝋ v= 0.8 x 0.00088 x √(2 x 20000)/ (1000) = 0.8 x 0.00088 x √(40000)/ (1000)
= 0.8 x 0.00088 x √40 = 0.00445 m3/s
This is the hand calculation for proving that the formulas are correct compare to the practical
that has been done. The tube diameter was given and is 0.0426 m, using the tube diameter it
was able to calculate the area of the tube using the formula that is bold stated above. Once
you have the area the volume flow formula can be used to calculate the volume flow. Before
you can use the volume flow formula the delta P needs to be calculated. Using the formula
for calculating delta P, the volume flow can be calculated. This was then the volume flow in
the pitot tube. Now there is left to calculate the volume flow of the orifice plate, which will
show the final volume flow of the metering station. The calculation is correct because it has
the same answer as the calculation sheet.
3. Analysis and final conclusions
As a final conclusion you can conclude that the measured results that were found as follows:
The measurements that was done shows that the volume flow differs from the pitot tube and the
orifice plate. The reason that it differs is because the meter flange gives a different value then the
pitot tube and the meter flange is the one that is more accurate. The theory then differs a bit from
the practical. This is because the pitot tube uses an electronic pressure sensor with a digital
display and the orifice plate uses a differential pressure gauge plane. Both rely upon Bernoulli’s
equation, but use different terms. It is because the pressure is supposed to be the same in point A
and the same in point B. The Bernoulli equation actually is used when there is a difference in
pressure. There is at first in the pitot tube when the pressure is being measured 5 times in
different RPM & flow that the pressure is almost the same. After taking five measurements in
five different conditions that is mentioned above in the steps measurements were taken. The
dynamic pressure for the first was the highest and the dynamic pressure for the second and forth
were the lowest, therefore I can conclude that for a low flow there is a low dynamic pressure so
the dynamic pressure depends on the velocity of the tube. For a higher dynamic pressure there is
a higher volume flow, the volume flow then depends on the dynamic pressure. The difference
between the pressures can be caused by the air in the tube and the curves in the tube. The air
present in the tube can cause that the velocity is slower/ preventing that the flow could go faster,
therefore the difference in the pitot tube and the flange meter is lower. Difference between
pressures can also be caused by curves inside the tube, the tubes that are set in the system have
bends of 90 degrees that can have an influence on the velocity causing the velocity at those
points to have an obstacle in the way that slows its trajectory.