1083ch5 1 PDF

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 9

Pressure Measurement

5.1 Motion Balance 727


SELECTION AND APPLICATION 712 Force Balance 727
Introduction 712 Atmospheric Reference Sensors 728
Orientation Table 712 Motion Balance 728
Reference Pressures 715 Single Bellows 728
An Example 715 Dual Bellows 728
Selecting the Pressure Detector 716 Force Balance 728
Accessories 716 Bibliography 728
Intelligent Transmitters 717
Bibliography 717 5.4
BOURDON AND HELICAL PRESSURE SENSORS 731

5.2 The Bourdon Tube 732


ACCESSORIES (SEALS, SNUBBERS, CALIBRATORS, C-Bourdon Pressure Sensors 732
MANIFOLDS) 718 Spiral Bourdon Pressure Sensors 734
Helical Bourdon Pressure Sensors 734
Pulsation Dampeners and Siphons 719 Fused Quartz Helix Sensors 735
Freeze Protection 720 Bibliography 735
Chemical Seals 720
Limitations 721
5.5
Standard Seals 721
DIAPHRAGM OR CAPSULE-TYPE SENSORS 736
Process Connections 722
Self-Cleaning Seals 722 Introduction 737
Volumetric Seal Elements 723 Diaphragm Elements 737
Valve Manifolds 724 Materials and Configurations 737
Calibrations and Communicators 724 Sensor Configurations 738
Bibliography 725 Absolute Pressure Sensors 738
Motion Balance 738
Force Balance Design 739
5.3 Transmitter Ranges and Materials 739
BELLOWS-TYPE PRESSURE SENSORS 726 Atmospheric Reference 739
Introduction 726 Motion Balance Design 739
Basic Designs 727 Materials of Construction and Spans 740
Absolute Pressure Sensors 727 Slack Diaphragms 740

709
© 2003 by Béla Lipták
710 Pressure Measurement

Force Balance Designs 741 Optical Transducers 760


Features and Construction Choices 741 Smart Transmitters 760
Pressure Repeaters 741 References 761
Bibliography 742 Bibliography 761

5.6 5.8
DIFFERENTIAL PRESSURE INSTRUMENTS 743 HIGH-PRESSURE SENSORS 762
Introduction 744 Introduction 763
Measurement Error 744 Mechanical High Pressure Sensors 763
Example 745 Dead-Weight Piston Gauges 763
Smart Transmitters Reduce Error 745 Button-Type Pressure Repeater 764
D/P Instrument Designs 745 Helical Bourdon 764
Filter Status Indicators 745 Bulk Modulus Cells 764
D/P Switches 745 Pressure-Sensitive Wires 765
D/P Indicators 745 Change-of-State Detection 765
Temperature Compensation and Over-Range Dynamic Sensors 765
Protection 746 Reference 765
Ranges and Materials 746 Bibliography 765
Liquid Manometers 747
D/P Transmitters 747
Dry, Force Balance Design 747 5.9
Pneumatic Version 747 MANOMETERS 766
Suppression and Elevation 748
Introduction 767
Flat and Extended Diaphragms 748
Liquid-Sealed Designs 767
Ranges and Pressure/Temperature
Inverted Bells 767
Ratings 748
Cylindrical and Ledoux Bells 767
Wafer Elements 748
Double Bell Unit 768
Torque Tube Sensors 749
Ring Balance Manometers 768
Low-Differential Transmitters 750
Membrane Type Design 750 McLeod Vacuum Gauges 768
Bibliography 750 90 ° Rotation Type 768
Piston-Type McLeod Gauge 769
Micromanometer With Precision Needle 770
5.7 Visual Manometers 770
ELECTRONIC PRESSURE SENSORS 751 Indicator Fluids 770
Introduction 752 Liquid Barometers 770
Electrical Safety 752 Glass Tube Manometers 770
Strain Gauge Transducers 753 Well-Type Design 771
Historical Development 753 Vacuum Measurement 771
The Bonded Strain Gauge 754 Multitube and Interface Manometers 771
Temperature Compensation 754 Inclined Tube Designs 772
Detecting the Change in Resistance 754 Micromanometers 772
Transducer Designs 755 Float Manometers 772
Bonded Designs 755 Servomanometers 773
Force Balance 755 Conclusions 773
Capacitance Transducers 756 Bibliography 773
Potentiometric Transducers 757
Resonant Wire Transducers 757
Piezoelectric Pressure Sensors 757 5.10
Magnetic Transducers 758 MULTIPLE PRESSURE SCANNERS 774
Inductive Elements 758 Introduction 774
Linear Variable Differential Rotary Pressure Scanners and Distributors 775
Transformer 758 Scanning Frequency 775
Inductive Transmitters 758 Differential Pressure Scanning 775
Reluctive Elements 759 Digital Interface Units 776

© 2003 by Béla Lipták


Contents of Chapter 5 711

Rotary Air Signal Distributors 776 5.13


Air Signal Distributor Manifolds 776 PRESSURE AND DIFFERENTIAL PRESSURE
Ramping Pressure Scanners 777 SWITCHES 790
Electronic Pressure Scanners 777 Pressure Switch Features 790
Low Level Multiplexer 777 Features and Configurations 790
High Level Multiplexer 778 Terminology 791
Bibliography 778 Selection 791
Differentials 791
Designs for Hazardous Area Applications 792
5.11
Differential Pressure Switches 792
PRESSURE GAUGES 779
Bibliography 793
Introduction 780
The Gauge Components 780
The Gauge Housing 780 5.14
Gasket and Lens 780 VACUUM SENSORS 795
Sensing Element 781 Introduction 796
Safety Features 781 Vacuum Gauge Classifications 796
Dials and Pointers 781 Mechanical Vacuum Gauges 797
Maximum Pointer 782 Manometers 797
Digital Displays 782 Capacitance Manometers 798
Diaphragm-Type Vacuum Gauges 782 Quartz Helix Vacuum Gauge 798
Special Features 782 Viscous Friction of Spinning Ball 799
Illumination 782 Molecular Momentum Vacuum Gauges 799
Temperature Compensation 782 Mechanical Linkage Vacuum Gauge 799
Duplex and Differential Gauges 782 Thermal Vacuum Gauges 799
Failure Causes and Reliability 783 Pirani Vacuum Gauge 800
Conclusions 783 Thermocouple Vacuum Gauge 800
Reference 783 Thermopile Vacuum Gauge 801
Bibliography 783 Convectron Vacuum Gauge 801
Ionization Vacuum Gauges 802
Hot Cathode Ionization Gauges 802
5.12 Cold Cathode Ionization Gauges 803
PRESSURE REPEATERS 785 Vacuum Gauge Calibration 804
Booster Relays 785 McLeod Vacuum Gauges 804
Flanged Repeaters 785 Calibration Reference Tubes 804
Force Balance Pressure Repeaters 786 Vacuum Controllers 804
Vacuum Repeaters 787 Aneroid Manostats 805
Button-Diaphragm Repeaters 787 Cartesian Diver Regulators 805
Pressure Repeater Applications 788 Analog Electronic Controllers 805
Repeating a Vacuum 788 Mass Flow Controllers 805
Level Measurement Applications 788 References 806
Bibliography 789 Bibliography 806

© 2003 by Béla Lipták


5.1 Selection and Application
B. G. LIPTÁK (1969) J. T. HALL (1982) J. A. NAY (1995, 2003)

Partial List of Suppliers: ABB Instrumentation (www.abb.com/us/instrumentation)


Barton Instruments (www.barton-instruments.com)
Brooks Instrument (www.brooksinstrument.com)
Dresser Instrument (www.dresserinstruments.com)
Endress + Hauser Inc. (www.us.endress.com)
Fisher controls (www.fisher.com)
Foxboro/Invensys (www.foxboro.com)
Honeywell (www.iac.honeywell.com)
Marsh Bellofram (marshbellofram.com)
Moore Industries (www.miinet.com)
Omega Engineering (www.omega.com)
Rosemount/Emerson (www.rosemount.com)
Siemens (www.sea.siemens.com)
United Electric (www.ueonline.com)
Youkogawa Corp. of America (www.yca.com)

Pressure detection devices can be classified on the basis of Silicon microchip technology continues steadily to reduce
the pressure ranges they can measure, on the basis of the the cost of advanced features, reducing the size and weight
design principle involved in their operation, or on the basis of hardware and improving their availability and accuracy,
of their application. In this chapter, the various categories are while extending the long-term stability their calibration.
not separated in any strict manner. Industrial instruments are Many of the sensors are available with digital communication
discussed in detail, with emphasis on the most commonly used capability, which can serve calibration, adjustment, and
devices; laboratory instruments are covered in less detail. reporting of process variables, allowing for complete plant-
wide integration.
With so many types of sensors, it might seem that making
the proper selection for a particular installation would be
INTRODUCTION
difficult and time consuming. Actually, this is not the case.
A multitude of devices are covered here for the purpose of
Each section starts with a brief summary of basic features
completeness, but for a typical industrial installation, the
applicable to the group of instruments discussed in that sec-
selection is fairly simple, and often repetitive.
tion. This information allows readers to quickly determine
whether that category of instrumentation is suitable for their
Orientation Table
application.
This chapter covers a wide range of pressure sensors, The reader should find Table 5.1a, the Orientation Table for
which can measure pressures from ultrahigh vacuums, such Pressure Detectors, of value in narrowing the choices. For each
−13
as 10 mmHg, to ultrahigh gauge pressures approaching category of sensors, this table indicates the overall pressure
400,000 PSIG (2,800 MPa). range that the category is capable of detecting. The table also
The range of the costs and inaccuracies of these instru- notes whether the unit is available for industrial on-line instal-
ments are equally broad. A simple, 1.5 in. diameter, 5% inac- lation or for laboratory use only. Although any transmitting
curate gauge might cost only $10, while a fused-quartz helix instrument can easily be provided with an inexpensive analog
sensor with an error of 0.01% and a digital readout could cost or digital local indicator, the table differentiates those sensor
$6000. The cost of pressure transmitters range from a few categories, which primarily serve as local gauges or indicators.
hundred dollars or less for the disposable models that have Also distinguished are the sensor categories, which are com-
limited features, to $2000 for smart models with built-in monly available in microprocessor-based smart configurations.
digital PID control algorithms and/or digital networking The table also indicates the type of pressure reference used.
capabilities. When the environmental pressure surrounding the instrument

712
© 2003 by Béla Lipták
TABLE 5.1a
Orientation Table for Pressure Detectors
Features Applicable Pressure Ranges

mmHg absolute “H2O PSIG

Local Readout (Gauge)

Remote Readout Trans.

Smart Units Available


(1 mmHg = 133 Pa) (1"H2O = 250 Pa) (1 PSIG = 6.9 kPa)

Laboratory or Pilot
Inline Device

Plant Device 10 -14 10 -10 10 -6 10 -3 10 -1 1 50 200 400 600 4 7 11 102 103 104 105 106
Type of Design −300 -200 -100 -10 -5 -1 ±0.1 +1 +5 +10 +100 +200 +300

Abs. Press. Motion Balance


Bellows

Abs. Press. Force Balance


Atm. Press. Ref. Motion Bal.
Atm. Press. Ref. Force Bal.
Aneroid Manostats

C-Bourdon
Bourdon

Spiral Bourdon
Helical Bourdon
Quartz Helix

Abs. Press. Motion Balance


Diaphragm

Abs. Press. Force Balance


Atm. Press. Ref. Motion Bal.
Atm. Press. Ref. Force Bal.

Strain Gauge
Capacitive Sensors
Electronic

Potentiometric
Resonant Wire
Piezoelectric
Magnetic
Optical
High-Press.
Sensors

Dead Weight Piston Gauge

5.1 Selection and Application


Bulk Modulus Cell
Manganin Cell

Inverted Bell
Manometers

Ring Balance
Float Manometer
Barometers
Visual Manometers
Micromanometers
Repeaters.

D/P Cell
Pressure

Std. Diaphragm
Button Diaphragm
Ioniza-

Hot Cathode
tion

Cold Cathode

713
© 2003 by Béla Lipták
714
TABLE 5.1a Continued
Orientation Table for Pressure Detectors

Pressure Measurement
Features Applicable Pressure Ranges

mmHg absolute “H2O PSIG

Local Readout (Gauge)

Remote Readout Trans.

Smart Units Available


(1 mmHg = 133 Pa) (1“H2O = 250 Pa) (1 PSIG = 6.9 kPa)

Laboratory or Pilot
Inline Device

Plant Device
10 -14 10 -10 10 -6 10 -3 10 -1 1 50 200 400 600 4 7 11 10 2 10 3 10 4 10 5 10 6
Type of Design -300 -200 -100 -10 -5 -1 ±0.1 +1 +5 +10 +100 +200 +300

Thermocouple
Thermal

Thermopile
Resistance Wire-Pirani
Convectron

Quartz Helix
Mechanical

McLeod
Molecular Momentum
Capacitance
Spinning Ball

- Indicates that the device uses full-vacuum reference in its operation.


- Indicates that the device uses atmospheric pressure reference.
- Indicates that the operating priciple used does not involve the use of reference pressure.

© 2003 by Béla Lipták


5.1 Selection and Application 715

Reference Pressures
TABLE 5.1b
Pressure Conversion Reliable reference pressures are important, because they can
To Convert to Pascals (Pa) From Multiply by be a source of error just as much as an error on the measure-
atmosphere (standard) 1.01 × 10
5 ment side can. In the case of absolute pressure sensors, the
atmosphere (technical = 1 kgf/cm )
2
9.81 × 10
4 reference chamber cannot be fully evacuated to absolute zero
5 pressure, but full vacuum is only approached within a few
bar 1.00 × 10
3
thousands of a millimeter of mercury (torr). This means that
centimeter of mercury (0°C) 1.33 × 10 a nonzero value is being treated as zero, which, when mea-
centimeter of water (4°C) 98 suring higher vacuums, can cause significant errors. The other
2
dyne/cm 0.1 potential error source is the possibility of in-leakage of atmo-
foot of water (39.2°F) 2.98 × 10
3
spheric air into the vacuum reference chamber of absolute
gram-force/cm
2
98 pressure detector.
inch of mercury (32°F) 3.39 × 10
3 In case of positive pressure detectors, if the barometric
3 pressure is the reference, atmospheric pressure variations
inch of mercury (60°F) 3.37 × 10
cause a problem. As the atmospheric pressure can vary, by
inch of water (39.2°F) 249 about 1 in. of mercury (13.6 in. or 0.345 m of water), the
inch of water (60°F) 248 resulting error can be significant if the process pressure is
2 4
kgf/cm 9.81 × 10 near atmospheric. In addition, the output signal of the sensor
kgf/m
2
9.81 can change even when the process pressure is constant. The
kgf/mm
2
9.81 × 10
6 resulting error might not be significant when detecting high
2 6 gauge pressures, but it can be a problem with compound
kip/in (ksi) 6.89 × 10
detectors.
millibar 100
A compound pressure sensor is one that operates at near
millimeter of mercury (0°C) 133 atmospheric pressures and can detect the pressures on both
2
lbf/ft 47.8 above and below atmospheric. In controlling the pressure in
2 3
lbf/in (psi) 6.89 × 10 sealed rooms (clean rooms, biohazard containment chambers,
psi 6.89 × 10
3 ordinary tight buildings, etc.) where the ventilation systems
torr (mmHg, 0°C) 133 may purposely hold the pressure above or below atmospheric
pressure, this variable reference can be a source of problems.
3
Example: 1 PSI = 6.89 × 10 Pa/PSI = 6890 Pa In selecting pressure measurement devices for such applica-
= 27.7 in. water tions, the design engineer must not ignore this and must either
= 2.31 ft. water determine that the effect of barometric pressure variation can
= 2.04 in. Hg
2
be safely neglected, or must measure and correct for this
= 0.07 kgf/cm
variation in the measurement and control systems.

is used as reference (for the measurement of gauge pressures), An Example


it is noted by A for atmospheric. When the reference is an
evacuated chamber with almost perfect vacuum sealed inside, Take the example of a chemical reactor that has to be evac-
which is used for the measurement of absolute pressures, this uated to 10 mmHg before being purged at 1 in. (25.4 mm)
type of reference is noted by V . There is no reference for H2O positive pressure. After purging, when the reaction
differential pressure devices because they just compare one starts, this reactor operates at a higher positive pressure.
pressure against another. A few measurement technologies, Due to the problem with the references, there is no single
such as the thermal and ionization type vacuum gauges, do pressure transmitter that can detect all these pressures. If a
not depend on the use of reference pressures or vacuums vacuum reference is used, the purge pressure of 1 in. H2O
within the instrument itself. over atmospheric cannot be reliably detected. If an atmo-
Table 5.1b gives a variety of pressure conversion multi- spheric reference is used, the 10 mmHg vacuum cannot be
pliers. accurately detected, because it could be subject to a 25 mmHg
In addition to the general orientation provided by Table 5.1a, error. In the past, the logical solution was to use multiple
a summary of the features of some of the pressure instruments sensors. Today, we can also use intelligent transmitters,
that are commonly used in the manufacturing, petrochemical, which are provided with multiple references and are capable
and power industries is given in Table 5.1c. Note that in actual of switching them on the basis of the batch sequence of the
practice the metric ranges are also rounded. reactor.

© 2003 by Béla Lipták


716 Pressure Measurement

TABLE 5.1c
Pressure Detector Errors, Ranges, and Costs
Type Range Inaccuracy Approx. Cost
General-purpose Bourdon-tube 15–10,000 PSIG 2% $100
indicator (1–690 bars)
High-accuracy test gauge Low vacuum to 3000 PSIG 0.1% to 0.01% $300–$6,000
(Low vacuum to 207 bars)
Bourdon/spiral case-mounted Low vacuum to 50,000 PSIG 0.50% $1,200
indicator/recorder (Low vacuum to 3450 bars)
Spring-and-bellow case-mounted Low vacuum to 50 PSIG 0.50% $1,600
recorder (Low vacuum to 3.5 bars)
Nested capsular case-mounted 10–90 PSIG, 0.50% $1,600
recorder (0.7–6.2 bars)
Low-pressure bell case-mounted −0.1 to 0.1 in. H2O 2% $2,200
indicator (−3 to 3 mm H2O)
Beam-mounted strain gauge 0–1000 PSIG 0.25% $800
(sensor only) 4–20-mA DC output (0–69 bars)
Piezoresistive transducer 0–5000 PSIG 0.50% $500
4–20-mA DC output (0–365 bars)
“Smart” piezoresistive transmitter 0–6000 PSIG 0.10% $1,200–$2,000
4–20-mA DC output (0–414 bars)
“Smart” field communicator for remote — $1000–$3,000
calibration and configuring of “smart”
transmitter
Capacitive sensor/transmitter 1 in. H2O–6000 PSIG 0.2% 1000
(25 mm H2O–414 bars)

SELECTING THE PRESSURE DETECTOR transmitted, the differential pressure or the electronic sensors
described in Sections 5.6 and 5.7 should be considered. The
When local pressure indication is required and the process high-pressure sensors, described in Section 5.8, are the rec-
pressure range is between 0 to 10 in. H2O (2.6 kPa) and 0 to ommended choices for pressures from 20,000 PSIG (140
100,000 PSIG (690 MPa), the conventional pressure sensors, MPa) up to 400,000 PSIG (2,800 MPa).
which are described in Sections 5.3, 5.4, 5.5, and 5.11 can Multiple pressure sensors, including scanners and
be considered. The local pressure gauges, described in Section multiplexers, are discussed in Section 5.10. Pressure rep-
5.11, can have ranges from 10 in. H2O (2.6 kPa) up to 100,000 eaters capable of repeating pressures from full vacuum to
PSIG (690 MPa). 10,000 PSIG (69 MPa) are described in Section 5.12.
For the measurement of near-atmospheric pressures, Pressure and differential pressure switches for applications
the bellows diaphragm sensors and manometers (Sections at up to 20,000 PSIG (138 MPa) pressures can be found
5.3, 5.4 and 5.9) are the most likely choices. Similarly, for in Section 5.13.
local vacuum measurement down to 1 mmHg (0.13 kPa),
the diaphragm, the bellows-type, and the vacuum manom-
eters (Sections 5.3, 5.5, and 5.9) will give satisfactory
−12
performance. Vacuum sensors, ranging from 10 to 760 Accessories
mmHg, are discussed in Section 5.14. See Figure 5.14a for
a summary of all the available vacuum sensors and their The pressure detectors are often provided with various acces-
ranges. sory items (discussed in Section 5.2), which serve to protect
Where remote transmission is required, the force balance them from process conditions and environmental effects, are
or motion balance transmitters (Sections 5.3, 5.4, 5.5, and provided to reduce maintenance. The most common causes of
5.7) will handle most applications. They can detect vacuums failure or maintenance problems include plugging, vibration,
down to 1 mmHg (0.13 kPa) absolute and gauge pressures freezing, corrosion, excessive temperatures, and hard-to-
up to 100,000 PSIG (690 MPa). When small, near-atmospheric handle process materials. The various protection devices dis-
or high pressures up to 200,000 PSIG (1,400 MPa) are to be cussed in Section 5.2 can assist in making the installation

© 2003 by Béla Lipták


5.1 Selection and Application 717

less sensitive to such effects and can reduce the required tasks Bibliography
of periodic servicing, testing, calibration, and maintenance.
Benedict, R.P., Fundamentals of Temperature Pressure and Flow Measure-
ments, 3rd ed., New York: John Wiley & Sons, 1984.
Intelligent Transmitters Cooper, F.G., “Low Pressure Sensing and Control,” InTech, September 1992.
Dally, J.W., Riley, W.F., and McConnell, K.G., Instrumentation for Engi-
Microprocessor based pressure and differential pressure neering Measurements, 2nd ed., New York: John Wiley & Sons, 1993.
transmitters are widely available today. Such a transmitter Dyer, S.A., Wiley Survey of Instrumentation and Measurement, New York:
includes an input circuit referred to as an analog-to-digital John Wiley & Sons, 2001.
(A/D) converter that converts the sensor input into a digital Faust, E., “Digital Pressure Indicators versus Manometers,” Measurements
& Control, April 1993.
signal before it is sent to the microprocessor. The micro-
Ferree, S.R., “Sensors: Simple to Smart to Systems,” InTech, November 1991.
processor performs the manipulations of ranging, lineariza- Figliola, R.S. and Beasley, D.E., Theory and Design for Mechanical Mea-
tion, error checking, and conversion and either transmits surements, 3rd ed., New York: John Wiley & Sons, July 2000.
the reading digitally or sends the resulting value to the Gray, H., “Capacitive Pressure Transducers,” Measurements and Control,
output digital-to-analog converter (D/A), which converts the June 1991.
signal back to an analog signal of 4–20 mA DC, 0–1 V DC, Hughes, T.A., Pressure Measurement, EMC series, downloadable PDF,
Instrumentation, Systems, and Automation Society, Research Triangle
or 0–10 V DC. Park, NC, 2002.
Just as microprocessors have evolved in sophistication, Instrumentation, Systems, and Automation Society, Pressure (video, VHS,
so have A/D and D/A converters, increasing their resolution PAL & NTSC), Research Triangle Park, NC, 2002.
from 8-bit up to the 18-bit, which has been used in the better Instrumentation, Systems, and Automation Society, Pressure: Indicators and
transmitters since the beginning of 2000. These advanced Transmitters, CD-ROM, Research Triangle Park, NC, 2002.
Instrumentation, Systems, and Automation Society, Process Measurement
transmitters also check their own calibration on every mea- Series, CD-ROM, Research Triangle Park, NC, 2002.
surement cycle and incorporate self-diagnostics, while being Jager, R., “A Smart Transmitter for Differential Pressure Measurement,”
configurable by using simple personal computer (PC) software. M&C News, September 1992.
The reconfiguration process is not only quick and convenient, Johnson, D., “Pressure Sensing: It’s Everywhere,” Control Engineering,
but also tends to lower inventories by making the transmitters April 2001
Juanarena, D.B., “Pressure Measurement in Cryogenic Environments,”
interchangeable.
Instrumentation, Systems, and Automation Society Conference, Houston,
The benefits of remote setup, configuration, and access TX, October 1992.
to diagnostics has resulted in a dramatic increase in the use Lewis, J.C., “Pressure Sensing: A Practical Primer,” InTech, December 1989.
of proprietary protocols that are supported by many of the Loth, S.A., “Picking the Perfect Pressure Transducer,” InTech, July 1988.
larger manufacturers. A wide variety of intelligent pressure Nishikawa, T. et al., Intelligent DP Transmitter Using a Resonant Single-
Crystal Sensor, Instrumentation, Systems, and Automation Society
transmitters are available on the market today. Some common
1993 Technical Conference, Chicago, September 19–24, 1993.
features of the leading models include not only digital and Paros, I.M., “Fiber-Optic Resonator Pressure Transducers,” M&C News,
analog outputs, but also multiple ranges, remote zero and September 1992.
span settings, configuration push buttons, PC software and “Reluctive and LVDT Pressure Transducers,” Measurements and Control,
Hand Held Configurators. The available bus and network September 1992.
protocols include Highway Addressable Remote Transducer Stockdale, R., “Advanced Pressure Sensor Designs Benefit Control Sys-
tems,” Control Engineering, June 1991.
(HART), Foundation Fieldbus, Profibus, Ethernet, or just Studebaker, P., “Engineers Find New Uses for Pressure Transmitters,” Con-
4–20 mA. Some field locations will benefit from local indi- trol, February 1994.
cation and this feature is optional with most manufacturers. Weiss, J., “Sensors: Toward the Next Generation,” InTech, September 1993.

© 2003 by Béla Lipták

You might also like