Precaution
Printer
Pump
Upper case
O2 sensor
Solenoid valve
Front case
NDIR module
Display PCB
4GAS ANALYZER
3pin Com 1
(Display PCB)
Display PCB
4/5GAS 2GAS
(Pump) (Solenoid)
Printer PCB
(RS232C port)
(Key)
Pump
Coil Resistance : 1.8~2.7 KΩ (220v)
Coil Resistance : 0.3~0.6 KΩ (110v)
( Display PCB)
In (Main filter)
In
Out (Water outlet) (Solenoid valve 2)
Out
Check the rubber separation (NDIR module in)
and to be torn
- It controls the inflow of exhaust gas and air into NDIR module.
- While operating pump, measured value of the vacuum measured by
vacuum gauge (by placing the gauge at INLET) should turn out over
15cmhg.
Power supply
(Display power)
2pin connector(Fan)
- This power supply is designed for the use of AC220/240V or AC110V only.
Please verify the power sources prior to the usage.
- DISPLAY OFF ( error )
If the power socket is pass and the fan is operated, the power supply is normal.
If the power socket is pass and the fan isn’t operated, the power supply isn’t normal.
Solenoid valve
At the time of zero point adjustment
2 (Pump in)
3 (Dust filter)
- At measuring mode, close the INLET entrance by end of the finger, then
putting finger at the low part of ZERO filter. If the finger feels vacuum,
the Solenoid valve is defective.
Printer
Check if the cover is Make sure
closed correctly the correct side
Printer PCB
correct direction
- Please check the direction of the filters and L type nipple (flow low).
- Please check the O ring of the housing (leak fail /low values).
O2/NOx sensor
O2 sensor (3pins)
Check the DC mV
(side pins)
NOx upgrade
NOx PCB
(upgrade kits)
O2 sensor
NOx sensor
(upgrade kits)
4GAS 5GAS
Troubleshooting
Troubleshooting
Fault Verification
(or if it’s not solved yet)
Unable to power switch on Check the power cord and the fuse. Replace the power supply.
Check the probe, hoses, main filter “L”
Flow ERR Replace the blocked parts.
type nipple, filters blocked.
(Flow error)
Check if the probe filter is reversed. Change the direction of the probe filter.
Check the remaining gas or water in the
Zero ERR(error) Replace the NDIR module.
NDIR module.
Check the connection status of probe,
Leak Fail Replace the pump or solenoid.
hoses, O rings and filters.
Please do the leak test.
Measuring values almost Check the connection status of probe, Replace the filters.
stay in “0” or low values hoses and filters blocked, main filter O ring Replace the pump or solenoid.
Check the pump or solenoid if it is failed.
λ or O2 values is high Check the O2 sensor or leak. Replace the O2 /NOx sensor
NOx value is low Check the NOx sensor (with NOx gas) (Life span: 1~2 years).
Check the calibration gas values again.
Gas calibration: Fail Replace the NDIR module.
HC(C3H8 *0.5) should be input.
PRT(Print) ERROR Check the connector and cable of printer. Replace the printer or printer PCB.