SPECIFICATIONS INFUSION PUMP TE-135/TE-135C

Product Name TERUFUSION INFUSION PUMP TE-135


Model TE – 135
Infusion Set Designated TERUMO IV sets for pump
Pumping System Peristaltic Finger Method
Control Method Volumetric Control
Delivery Rates STANDARD mode: 1 – 999 mL/h ( 1 mL/h step )
MICRO mode: 0,1 – 99.9 mL/h ( 0.1 mL/h step )
Delivery Limit 1 – 9999 mL ( 1 mL step ) or
“ - - - - “ ( Delivery Limit Free )
Volume Delivered STANDARD mode : 0 to 9999 mL ( 1 mL step )
MICRO mode : 0.0 to 99.9 mL ( 0.1 mL step )
100 to 9999 mL ( 1 mL step )
Flow Accuracy ± 10 % Conditions Infusion Set: Designated TERUMO
IV sets for pump.
• Solutions : Water / Saline
• Test Time : Second Hour
• Delivery Rate : 1.0 mL/ h or more
• Temperature : 25 o Celcius
With a designated Infusion Set water or saline,
measured at 1 hours after start-up and subsequent
every hour.
( Actual Flow Accuracy depends on the type of
solution )
Occlusion Pressure 30 – 140 kPa ( 0.30 – 1.43 kgf/cm2 )
Purge Approx 500 mL/h
Alarms Battery, Air-in-line, Occlussion, Door, Infusion
Completion ( Keep Open Rate Function ), Repeat
( activated if the cause of the initial alarm remains for
about 2 minutes ), Start Reminder ( activated if about
2 minutes wasted after infusion become available ),
Battery Error, Delivery Rate/Delivery Limit Comparison
alarms.
Alarm when using Drip Sensor ( Option ):
Empty, Flow Error ( Free Flow )
( An alarm during infusion is recognized by the
relevant alarm indicator or display, the operation
indicator, a buzzer etc. The pump is stopped by any
alarm other than the Battery ( except Shutdown ) and
Infusion Completion Alarms
Safety Functions • Panel Lock : Disables the user to change the D
Rate, D Limit values when the door is opened *
• Tube Clamp : Automatically presses and shuts the
IV tubing when the door is opened
• Key Lock : Disables the switches except
( START ), ( KEY LOCK ), & ( POWER ). This
function is deactivated when any alarm is
occurred *
• Air in-line Detection Sensitivity : adjust the
sensitivity of the air-in-line detector in 2 steps
(Low/High;default: Low )
• Occlusion Detection Sensitivity: Adjust in 10 steps
( default: 5 )
• Maintenance Timer : Turns on the maintenance
timer indicator after a specified time passes
( Default: not used )
• Upper Limit of D Rate : Limit the maximum D Rate
value withinthe predefined range
Additional Function • Volume Delivered Clear : Reset the volume
delivered to “ 0 “ when pressing VOLUME
DELIVERED switch for 2 second
• Alarm Volume : Adjust the alarm volume in four
steps ( Small / Medium / Large / Step
( escalating ), default : Medium
• Shutdown : Turns the pump off after alarming if
the internal battery voltage lower to a certain level
during operation on the battery
• History : Activates the History mode to display the
past operation record
• Stand-by : Deactivates the Start Reminder alarm
for 20 minutes with a long press of the
( STOP/SILENCE ) switch
• Cable Disconnection Detection : Detect the power
cable disconnection from the pump and gives an
alarm *
• Buzzer Sound Option : Switches the alarm buzzer
to a continuous sound
• Infusion Mode Limit : Disables the user to change
MICRO mode
Operating Conditions Temperature : 5 – 400 Celcius
Humidity : 20 – 90 %
Storage Conditions Temperature : - 20 – 450 Celcius
Humidity : 10 – 95 %
Power Supply AC : 100 – 240 Volt, 50/60 Hz
DC : 12 – 15 Volt/DC
Power Consumption 28 VA ( Watt )/ AC 100 – 240 Volt
11 Watt ( DC Power )
Classification Class I and Internal Power Supply
Type CF, IPX1, Continuous operation
Dimensions 100 mm (W) x 201 mm ( H) x 187 mm ( D )
Weight Approx 1.7 kg
Accessories AC Power Cable, Pole Clamp, Time Lag Fuse ( 2 pcs ),
Instruction Manual
For Model TE-135C only:
Nurse Call Output Connectable with nurse call via the computer
interface/nurse call connector.
Contact capacity: DC 12 Volt, 1 Ampere
Computer Interface Data output available via RS232 or RS485 connector
Accessories Computer Interface Protocol
Optional Accessories • DC Cable
( Sold Separately ) • Drip Sensor ( Code No: TE*975 )