Program očitava sve TIBBO portove Portovi se prijavljuju u prozoru levo i neaktivni su
2. Da bi se aktivirao port, potrebno je sledeće:
- kliknuti na odabrani port, zatim - kliknuti na settings - otvara se novi prozor za podešavanje parametara
3. Parametri vage su sledeći (za Shollex IP )
Network
M.A.C.- address - ne dirati
DHCP -0- disabled IP – podesiti prema LAN-u (10.10.10.4 – za Shollex ) Port 1000 Registration at dDNS server - 0 - disabled Auto – registration on link serv - 0 - disabled Gateway IP address - 10.10.10.100 (za Shollex ) Subnet mask - 255.255.255.0
Connection
Connection time out - 5 min
Transport protocol - TCP Link service login - 0 - disabled Inbond commands - 0 - disabled Data login - 0 - disabled
4. Routing mode 0 – server (slave)
Accept connection 0 – any IP address Notification destination - 0 - last port
5. Outband packets
max packet length - 255
max intercaracter delay - 1 start on any char - 1 - yes use start character - 0 – no start character (ASC I I code ) - 0 number of post characters – 0
Po završetku setapa pritisnuti OK i program setuje vagu.
To je seting za vagu LP 1 Ukoliko vaga posle ovakvog setinga ne radi neispravna je mrežna kartica. Potrebno je zameniti karticu. Seting LP vorks
Po otvaranju programa LP1 works
1. kliknuti na kockicu config 2. opredeliti se za opciju komunikacije TCP/IP 3. pod local IP podesiti IP adresu računara koji komunicira sa vagom 4. Gateway ne dirati 5. pod Scale IP podesiti IP adresu vage onu koju smo definisali u programu DS Manager 6. kliknuti (označiti ) tu adresu u kvadratiću levo od adrese i kliknuti na SAVE 7. otići u program na PLU 8. uneti PLU-ove, čekirati ih 9. klikom na download , program otvara podprozor 10. u novom prozoru kliknuti START 11. kada je prenos podataka gotov, pojavi se natpis TRANSFER AND (Complete) i time je posao oko prenosa gotov.