*1 0.05 mL or more if sampling sheet B (sold sepa- Items Specifications Part No.

 Detaching the sensor


rately) is used. 1. Power OFF the meter.
*2 The closeness of agreement between measured Sampling Y046, 100 sheet-pack 3200053858
sheet B 2. Lift the sensor tongue
Instruction Manual (Operation) value and actual value of the pH 4.01 standard tip and slide the sensor
COMPACT WATER QUALITY METER solution after two-point calibration using pH 4.01 a little away from the
LAQUAtwin-pH-11, LAQUAtwin-pH-22,
and pH 7.00 standard solutions. The temperature Part Names meter.
during the calibration and measurement is the 1 Flat sensor
LAQUAtwin-pH-33 same. Two-point calibration is started from pH 2 3. Pull out the sensor all the way from the
7.00. The error of standard solutions and round- 2 Light shield cover meter.
Specifications ing error (1 digit) are not included.
1
3 Liquid junction  Inserting/removing batteries
*3 The life period if the meter is used in the backlight 6 7 10 4 Glass membrane
Model LAQUAtwin-pH-11, LAQUAtwin-pH- off mode. If the backlight is used, battery life will  Inserting the batteries
22, LAQUAtwin-pH-33 5 Lithium batteries 1. Power OFF the meter.
shorten. 8
*4 The dimensions excludes projections and the 4 6 MEAS switch 2. Slide both batteries into
Target pH
mass excludes batteries. 3 7 ON/OFF switch the battery case as
Measurement principle Glass electrode method 5
*5 IP67: no failure when immersed in water at a 8 CAL switch shown.
Minimum sample volume 0.1 mL*1 depth or 1 meter for 30 minutes. Be sure to use two Battery
Please note that the meter can not be used 9 9 Waterproof gasket CR2032 batteries, and put Battery clip
pH range 0 pH to 14 pH underwater 10 Strap eyelet them with the plus sides
Resolution LAQUAtwin-pH-11: 0.1 pH  Items in package Note (+) upwards.
(valid digits) LAQUAtwin-pH-22: 0.01 pH  Removing the batteries
LAQUAtwin-pH-33: 0.01 pH Items Quantity Press the switches 0.5 seconds or more unless 1. Power OFF the meter.
otherwise specified.
Calibration LAQUAtwin-pH-11: Up to 2 points Sensor S010 1 2. Use a ball-point pen or
LAQUAtwin-pH-22: Up to 3 points other tool to pry the batter-
LAQUAtwin-pH-33: Up to 5 points Meter 1 ies out from the clips as
Storage case 1
Initial Setup shown.
Accuracy*2 LAQUAtwin-pH-11: 0.1 pH
Batteries CR2032 2  Attaching/detaching the sensor
LAQUAtwin-pH-22: 0.01 pH  Electrode conditioning
LAQUAtwin-pH-33: 0.01 pH Standard solution pH 4 (pH 4.01) 1  Attaching the sensor
pH 7 (pH 7.00) 1 1. Power OFF the meter. Note
Tempera- Available only on LAQUAtwin-pH-33,
ture display 0 C to 50.0C Pipette 1 2. Confirm that the waterproofing gasket is  Before using the sensor for the first time or
Display Custom (monochrome) digital LCD clean and undamaged. after several days of disuse, perform electrode
Sampling sheet B (5 sheet-pack) 1 conditioning.
with backlight 3. Slide the sensor Waterproof
Instruction manual (Operation) 1 onto the meter so gasket  Perform calibration after electrode condition-
Operating 5C to 40C, 85% or less relative ing.
environment humidity (no condensation) Instruction manual (Before use) 1 that catch "A" on the
back of the meter fits
Power CR2032 batteries (2)  Consumable parts sold separately into hole "a" on the 1. Place some drops of pH 7 standard solution
Battery life sensor tongue as A to the flat sensor.
Approx. 400 h continuous operation*3 Items Specifications Part No.
a
shown. 2. Wait a few hours before use.
Outer dimen- 164  29  20 mm, Approx. 50 g*4 There is no need to switch the meter ON.
sions/mass Sensor S010, pH 3200459834 Note
3. Clean the flat sensor with running water.
Main func- Temperature compensation, water- Standard 514-4, pH 4.01 3999960108 Be careful not to twist the waterproof gasket.
tion proof*5, auto stable/auto hold, auto- solution 514-7, pH 7.00 3999960109
matic power OFF

LAQUAtwin series (pH) Operation 1


Basic Operation 2. Open the light shield cover and place some Measurement When the measured value is stable, stops
drops of the standard solution on the flat blinking and the displayed value is locked with
 Power ON sensor taking care to cover the entire flat  Sample setting and displayed simultaneously.
1. Press and hold the ON/ sensor. 1. Open the light shield cover and put some 3. Document the displayed value.
OFF switch. Rinsing the sensor with the standard solution drops of sample on the flat sensor to cover
the entire flat sensor. 4. Press the MEAS switch.
The power is switched ON, beforehand will provide a more accurate cali-
The auto hold function is deactivated and
and the meter model num- bration as it will reduce sample crossover con- 2. Close the light shield cover. disappears.
ber is displayed on the LCD. tamination.  Measurement mode Be sure to perform this step before starting the
 Power OFF 3. Close the light shield cover and press the The auto stable (AS) mode and the auto hold next measurement. Or, you may mistake the
1. Press and hold the ON/OFF switch. CAL switch. (AH) mode can be selected. Refer to "  Mea- displayed hold value for the next measured
The power is switched OFF. and blink, and the calibration value is surement mode change" (page 4) for the opera- value.
displayed. tion to set the measurement mode.
Calibration  Auto stable (AS) mode Note
Light shield cover
Calibration is required before measurement. This is the default setting. appears when  If a measured value is out of the specified
Use standard solution within the measurement the measured value meets the stability criteria. measurement range, "Or" is displayed for
range in the specifications. If the value changes, disappears. upper range and "Ur" is displayed for under
Tip 1. Confirm that the meter is in the measure- range.
ment mode, and place a sample on the sen-  When you have a problem with the calibration
 Calibration values are saved even if the meter
sor. or measurement, refer to frequently asked
is switched OFF. After the calibration is complete, and questions.
 Calibration value is rewritten if calibration is stop blinking and the measured value is dis- When the read value meets
repeated using the same standard solution. played. the stability criteria,
 Select the pH standard solutions close to the The calibration value at 25C is displayed for 1 appears and the reading is  Measurement display change
pH value of sample when the sample is known. s and the display returns to the measurement locked. The display mode switches as follows by press-
 Perform three-point calibration using pH 4.01, mode automatically. 2. Document the displayed value when ing the MEAS switch in the AS mode.
pH 7.00 (6.86), and pH 10 (9.18) standard solu- 4. Open the light shield cover and remove the appears. LAQUAtwin-pH-11 and LAQUAtwin-pH-22:
tions when the sample is unknown. standard solution. Then remove moisture If the read value does not meet the stability cri-  Between pH and voltage alternately
on the sensor by gently dabbing with a soft teria, disappears and the reading changes LAQUAtwin-pH-33:
 Calibration points tissue. with time. Among pH, voltage, and temperature
The number of calibration points is dependent on This completes the 1st point calibration.  Auto hold (AH) mode
the meter model. 5. To perform 2nd point calibration, repeat appears when the measured value meets
Maintenance
 LAQUAtwin-pH-11:
steps 2. to 4. the stability criteria. The reading then locks  Storage
Up to two-point calibration
 LAQUAtwin-pH-22:  Calibration error and will not change until the MEAS switch is 1. Clean the sensor with tap water.
Up to three-point calibration pressed for the next measurement. 2. Dab gently with soft tissue or cloth to
If blinks and Er4 (error dis-
 LAQUAtwin-pH-33: play) appears, the calibration 1. Confirm that the meter is in the measure- remove moisture on the sensor and meter.
Up to five-point calibration has failed. ment mode, and place a sample on the sen-
 Multi-point calibration Perform electrode conditioning. sor. Note
Check that the correct standard solution is used, 2. Press the MEAS switch. Especially be sure to treat the flat sensor gently
1. Select the standard from USA or NIST.
and repeat calibration after cleaning the sensor. to prevent damaging it.
(Refer to "  Buffer standard change" If the calibration repeatedly fails when using the
The auto hold function is acti-
(page 4).) correct standard solution(s), the sensor may vated.
USA is selected by the default. blinks until the mea- 3. Close the light shield cover and the slide
have deteriorated. Replace the sensor with new
one. sured value has stabilized. cap before storing the meter.

LAQUAtwin series (pH) Operation 2 CODE: GZ0000472467


 Temperature sensor adjustment 1. Press and hold the CAL and ON/OFF Appendix Question Answer
Temperature sensor adjustment is available on switches for over 3 seconds when the
meter is switched OFF to Initialize calibra-  Frequently asked questions Er4 is dis- Please note that if you press the
LAQUAtwin-pH-33.
To perform accurate measurement with correc- tion. played during CAL switch in mV or temperature
After a moment of all segment indication, the Question Answer the calibra- display mode, Er4 is displayed. This
tion for temperature effects, follow the steps
below. Normally this is not necessary. software version is displayed. How can I Perform 2-point calibration. If cali- tion is because there is no calibration
1. Ready a reference thermometer, and allow And then, the display check the bration error occurs, the sensor has facility available for these modes.
the meter and reference thermometer to changes as shown right. sensor's con- deteriorated. Replace the sensor. Er1 is dis- The internal IC in the meter may be
reach to room temperature. 2. Press the CAL switch. dition? played soon defective. Perform meter initializa-
2. Set the display mode to temperature refer- All calibration data is reset. Can I mea- This meter cannot measure a sam- power ON. tion.
ring to "  Measurement display change" When the initialization of cali- sure high or ple with temperatures outside the If Er1 is still displayed after the initial-
(page 2). bration data is complete, End low tempera- meter's operating temperature range ization, the internal IC in the meter is
3. Press the CAL switch. appears. ture sam- (5C to 40C). defective. Replace the meter with a
The meter displays the setting screen for tar- The meter automatically switches OFF. ples? The difference between the sample new one (the meter cannot be
get temperature.  Initializing the settings temperature and ambient tempera- repaired).
4. Press the MEAS switch to adjust the dis- All setup choices are erased. The meter is reset ture increases the measurement Er2 is dis- The internal IC in the meter is defec-
played temperature on the meter to match to the factory default values. error. Perform measurement after played right tive. Replace the meter with a new
the temperature indicated by the reference 1. Press and hold the MEAS, CAL and ON/OFF the sample reaches the ambient after power one (the meter cannot be repaired).
thermometer. switches for over 3 seconds when the temperature. ON.
Pressing the MEAS switch increases the dis- meter is switched OFF to enter the initial- The mea- If lights steadily in AH mode, the
ization. Er3 is dis- The internal IC in the meter is defec-
played temperature. After the displayed tem- sured value measured value is locked. played right tive. Replace the meter with a new
perature reaches 40°C, it returns to 5°C. After a moment of all segment indication, the does not Press the MEAS switch to unlock the
software version is displayed. after power one (the meter cannot be repaired).
5. Press the CAL switch again to apply the change after value. ON.
And then, the display
displayed value to the adjustment. changing the If the value does not change after
changes as shown right.
The adjustment starts. The adjusted value sample. unlocking, the sensor may be dam-
blinks with and displayed. 2. Press the CAL switch. aged. Replace the sensor.
After the adjustment is complete, the adjusted All calibration data is reset.
When the initialization of “Or” or “Ur” The measured value may be out of
value stops blinking with MEAS and dis- blinks in the specified measurement range.
played. settings is complete, End
appears. value mea- Measure a standard solution to
If Er4 (error display) appears, the adjustment surement. check, and if “Or” or “Ur” still blinks,
has failed. Retry the above steps increasing The meter automatically switches OFF.
replace the sensor.
the time spent on the step 1.
If the adjustment repeatedly fails, the sensor blinks The measured temperature is not
may have deteriorated. Replace the sensor during mea- within the specified operating tem-
with new one. surement. perature (5C to 40C).
 Initializing calibration data If the ambient temperature is within
Initialize calibration in the following cases. the specified range and blinks,
 To delete the calibration data
replace the sensor.
 If the number of points for the last calibration is The meter Check that the batteries are inserted
uncertain. does not properly. If the battery voltage is low,
 After the sensor is replaced. power ON. replace them both with new ones at
the same time.

LAQUAtwin series (pH) Operation 3 CODE: GZ0000472467


Setup Mode  Setup completion
The setup mode allows the user to customize the meter to his specific needs.
To enter the setup mode, press and hold the MEAS and ON/OFF switches for over 3 seconds when the
meter is switched OFF. All the LCD segments appear and then the meter enters the setup mode.
Tip END indication (1 s) Powered OFF
 To have the changes apply, you need to go through the entire steps from “Setup mode entry” to
“Setup completion” shown below. To leave a setting as it is, just press CAL switch in the setting.
 To exit the setup mode with no change of settings, press the ON/OFF switch earlier than pressing
CAL switch in the last step but one, or the “Backlight setting” step.

 Setup mode entry

ON/OFF switch
MEAS switch
All segment Software ver. CAL switch
indication (2 s) indication (1 s)

 Buffer standard change


The standards of pH buffer can be changed.

USA buffer NIST buffer


(default)
 Measurement mode change
The measurement mode can be switched.

* Measurement display change is available in


the AS mode. Refer to "  Measurement
display change" (page 2).
AS mode* (default) AH mode

 Backlight setting
The backlight can be switched to ON or OFF.

31, Miyanonishi-cho, Kisshoin Minami-ku, Kyoto,


601-8306, Japan
Backlight ON Backlight OFF http://www.horiba-adt.jp
(default)
For any questions regarding this product, please contact
your local agency, or inquire from the following website.
http://global.horiba.com/contact_e/index.htm

LAQUAtwin series (pH) Operation 4 CODE: GZ0000472467