Professional Documents
Culture Documents
Torque Wrench Calibration Presentation PDF
Torque Wrench Calibration Presentation PDF
Torque Wrench Calibration Presentation PDF
© NMISA 2011
Calibration Setup
350,0 N•m
Clockwise Rotation
Reference Standard
Torque Transducer &
Readout Unit
Force Applied
© NMISA 2011
GUM Steps
© NMISA 2011
Measurement Model
© NMISA 2011
Identifying the sources of uncertainty
SCal
TSTD TUUT URes
URep
SRes
SCF
TUUT
© NMISA 2011
Quantifying the sources of uncertainty
• SCAL
• Calibration Results Table from the Calibration Certificate
© NMISA 2011
Quantifying the sources of uncertainty (2)
• SCAL
• This is the uncertainty due to the accuracy of the Reference
Standard Torque Transducer, which is not perfect
• Corrections must first be applied, or the uncertainty increased, to
take the error into account (largest error on values either side of the
calibration point was +1,0 N•m)
• The Reference Standard Torque Transducer used has a full scale of
700 N•m and was calibrated in 100 N•m steps (See calibration
certificate)
• Therefore we will have to use the polynomial equation to determine
the actual torque generated by the UUT at 350 N•m since it is a
measurement point in between 300 N•m and 400 N•m.
• Since we have to interpolate a value we will use the largest reported
uncertainty from the calibration certificate for the values on either
side of the calibration point which is ± 1 N•m.
© NMISA 2011
Quantifying the sources of uncertainty (3)
• SCAL
• Since we will be using the polynomial to interpolate a
value at 350 N•m, we DO NOT need to correct for the
+ 1 N•m error at 399,8 N•m.
• Therefore total uncertainty for the “accuracy” of the
Reference Standard Torque Transducer is ±1 N•m
• This is treated as normal at 95,45% Level of Confidence
• The divisor is the coverage factor k which for 95,45%
LOC is 2
• The degrees of freedom are always ∞ or 100 % Reliable
due to the source of traceability being accredited and
reputable.
© NMISA 2011
Quantifying the sources of uncertainty (4)
• SRES
• This is due to the resolution of the Reference Standard
Torque Transducer Readout Unit
• We must first determine the “effective resolution”
• The least significant digit displayed is 0,1 N•m
• Resolution is always treated as a Rectangular
distribution source of uncertainty and this is the range.
• The semi-range is therefore (0,1 N•m/2)=0,05 N•m
• The divisor is √3
• The degrees of freedom are always ∞ or 100 %
Reliability
© NMISA 2011
Quantifying the sources of uncertainty (5)
• SCF
• Polynomial Equation Coefficients Table from the
Calibration Certificate
POLYNOMIAL 2 3
y=a+bx+cx +dx
EQUATION
© NMISA 2011
Quantifying the sources of uncertainty (6)
• SCF
• This is the additional uncertainty which results from the
interpolation calculation to determine the torque
generated by the UUT at a point in between the
calibration points of the Reference Standard Torque
Transducer
• It is obtained directly from the calibration certificate as
the “Standard Error of the polynomial curve Fit” value =
± 0,045 N•m
• This is treated as a normal distribution at a 68,27%
Level of Confidence
• The divisor is the coverage factor k which for 68,27%
LOC is 1
• The degrees of freedom are also obtained directly from
the calibration certificate = 4
© NMISA 2011
Quantifying the sources of uncertainty (7)
• URES
• This is due to the resolution of the UUT Torque Wrench
scale. (How it influences the setting of the wrench to a
specified torque)
• Typically this would be the smallest graduation on the
UUT setting dial which for this UUT is 2 N•m
• This is the range of the rectangular distribution
• Therefore the semi-range is (2 N•m/2)=1 N•m
• The divisor for Rectangular Distributed uncertainty
contributors is √3
• The degrees of freedom for resolution is always ∞ or
100 % Reliability
© NMISA 2011
Quantifying the sources of uncertainty (8)
• UREP
• This results from the variability in the measurement
results obtained after repeating the measurement 5
times.
• It can be dealt with either as “repeatability” or as
“reproducibility”
• “Repeatability” – all conditions remained the same
during the repeated measurements
• “Reproducibility” – any one or more of the conditions
changed during the repeated measurements
• Different approaches can be used to repeat the
measurement
• Take 5 measurements at one setting sequentially
• Take 5 sets of measurements from zero to full scale
© NMISA 2011
Quantifying the sources of uncertainty (9)
Repeatability
20 % 20 % 20 % 20 % 20 % 20 %
Meas 1 Meas 2 Meas 3 Meas 4 Meas 5 Mean
60 % 60 % 60 % 60 % 60 % 60 %
Meas 1 Meas 2 Meas 3 Meas 4 Meas 5 Mean
© NMISA 2011
Quantifying the sources of uncertainty (9)
Reproducibility
20 % 20 % 20 % 20 % 20 % 20 %
Meas 1 Meas 2 Meas 3 Meas 4 Meas 5 Mean
60 % 60 % 60 % 60 % 60 % 60 %
Meas 1 Meas 2 Meas 3 Meas 4 Meas 5 Mean
© NMISA 2011
Quantifying the sources of uncertainty (9)
• UREP
• Treating as “Repeatability” (as per ISO 6789)
• We use the ESDM
• ESDM = ESD/SQRT (n) = 0,98/sqrt (5)
= 0,436348 N•m
© NMISA 2011
GUM Steps
© NMISA 2011
Categorize as type A or type B
© NMISA 2011
GUM Steps
© NMISA 2011
Uncertainty Budget
© NMISA 2011
GUM Steps
© NMISA 2011
Reporting the result
© NMISA 2011
Conclusions
• Both methods in this case prove that the UUT is well within
the allowable ± 4% of Maximum (± 14 N•m)
• Using the ESDM (In accordance with ISO 6789) results in
the smallest uncertainty (unrealistic??)
• Using the ESD (contrary to ISO 6789) results in the largest
uncertainty (realistic??)
• Always use the polynomial for calibrations using the
laboratory Reference Standard Torque Transducer
• This will correct for any error on the Reference Standard
eliminating the need to apply corrections
• This will solve the problem of the “Applied Torque” not
being exactly at the nominal values
© NMISA 2011