BLT-F90&F80&F50&F30 Maternal Fetal Monitor Service Manual
BLT-F90&F80&F50&F30 Maternal Fetal Monitor Service Manual
BLT-F90/F80/F50/F30 maternal /
fetal monitor service manual
1
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Declaration
Manufacturers have this non-public service manual by copyright, and the right to
be treated as confidential information. The service manual is as a reference for
machine maintenance and repair only, others have no right to expose this service
manual.
This service manual is protected by laws for copyright and proprietary information.
Without the written consent of the manufacturer, this service manual may not be used
of any part for photographic, copying, or translated into another language.
Contents of the service manual are subject to be changed without notice.
To avoid the possibility of damage, be sure to comply with the following safety
instructions during maintenance of this instrument
1) The monitor contains some internal parts that can be repaired by the users, but it
should be performed by technical service persons that had been authorized by the
manufacturer.
2) The monitor is not classified as therapy equipment and shall not be used at
home.
3) Some mother parameters of the monitor can resist defibrillator. The defibrillation
not touch the patient, hospital beds and the instrument.
4) Before cleaning or repairing of the instrument, make sure the power supply is cut
off.
5) Check the reusable accessories regularly to see if they are damaged, such as the
sensor and the cables.If necessary, replace and dispose them properly as medical
waste accessories.
6) Warning: Do not use under flammable gas such as anesthetic, as there will be a
risk of explosion.
7) Warning: Do not dispose the batteries in a fire as they may explode.
8) Warning: Never touch signal input or output connector and patients in the same
time.
9) Caution: The instrument must be performed by authorized and qualified service
engineers.
2
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
10) Caution: This instrument is designed for continuous work, and drip-proof
closure devices, pls take care to avoid splashing water.
11) Caution: Keep the instrument clean, and avoid oscillation.
12) Caution: Do not do high-temperature sterilization process or use electron beam
and γ radiation sterilization.
13) Caution: Electromagnetic interference - Ensuring that the use of the
environment is away from the strong electromagnetic interference, such as radio
transmitters or mobile phones and other interference.
14) Caution: Before use it, the user must check the device to make sure it does not
significantly affect safety of the patient or have any equipment damage.
15) Caution: The following safety checking must be performed by properly trained
person, who has knowledge and practical experience. Usually, it should be checked
once a year or once a month or less. If you find obvious damage, we recommend you
to replace the damaged parts before using it.
16) Do testing according to the inspection procedures of the public institution.
17) Check the device for mechanical and functional damage.
18) Make sure the safety-related labels are easily identifiable.
19) Verify that the function of the device is consistent with the description of the
manual.
20 ) Caution:After the effective lifetime, the instrument shoulb be returned to the
manufacturer for recycle use in accordance with local regulations.
21 ) Caution: After use ,the battery in accordance with local regulations for proper
disposal.
22 ) Caution: If the instrument is not used in a long time , be sure to remove the
battery in time.
23 ) Caution: The battery should be stored in a cool and dry environment .
24 ) Caution:Do not put the metal objects together with the battery in storage, to
avoid accidental short circuit.
25 ) Caution:In the premise of meeting the needs of the clinical, we recommend
that ultrasonic irradiation time of the patient should be as short as possible .
26 ) Caution:When the instrument is moved from a cold environment to a warm
and humid place, do not use it immediately.
27 ) Caution : Keep the operating environment no vibration, no corrosion, no
combustible material , and avoid excessively high or low temperature and
humidity
3
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
4
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Content
BLT-F90/F80/F50/F30 maternal / fetal monitor service manual....................................................... 1
Chapter 1 General Introduction......................................................................................................... 6
1.1 The main features .......................................................................................................... 6
1.2 Operational principle..................................................................................................... 7
Chapter 2 Product Introduction ...................................................................................................... 8
2.1 Main unit .......................................................................................................................... 8
2.1.1 Front view ............................................................................................................. 8
2.1.2 left side view : ................................................................................................. 10
2.1.3 Right side view:................................................................................................ 11
2.1.4 Vear view............................................................................................................. 12
2.2 Operation menu .............................................................................................................. 13
2.2.1 Main menu .......................................................................................................... 13
2.2.2 Review ................................................................................................................ 17
2.2.3 Maternal info. ...................................................................................................... 17
2.2.4 volume setup ....................................................................................................... 17
2.3 Internal structure ............................................................................................................ 18
2.3.1 Internal modulus.................................................................................................. 18
2.3.2 F90/F80 LCD panel internal structure ................................................................ 19
2.3.3 Schematic diagram .............................................................................................. 20
2.3.4 Accessories.......................................................................................................... 21
Chapter 3 Disassembly................................................................................................................. 23
3.1 Prepare : ..................................................................................................................... 23
3.2 Disassembly procedure .................................................................................................. 23
3.2.1 F90/F80 disassembly .............................................................................................. 23
3.2.2 F50/F30 disassembly ........................................................................................... 28
3.4 Exploded view............................................................................................................. 34
3.4.2 F50/F30 exploded view ....................................................................................... 36
Chapter 4 Maintenance ................................................................................................................... 37
4.1 Transport ........................................................................................................................... 37
4.2 Maintenance ...................................................................................................................... 37
4.3 Sterilization ....................................................................................................................... 37
4.4 Ultrasound Gel .................................................................................................................. 38
4.5 Clean the printer ................................................................................................................ 38
4.6 Troubleshooting ................................................................................................................ 38
4.7 Warranty and Maintenance................................................................................................ 41
5
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
6
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
With the optional CTG evaluation module, users can print the evaluation result in
real-time and in segmented, for users’ analysis and reference.
With the built-in network interface, it can connect to the central station.
With built-in pluggable and rechargeable lithium battery.
Sound and light alarms in three levels, and the alarm volume is adjustable.
BLT-F90 fetal monitor detect fetal heart rate by non-invasive doppler ultrasound.
We know that a certain frequency of the ultrasonic obstacles encountered in the
dissemination of reflection to occur, if the object is stationary, then the reflected wave
frequency and the transmitted wave at the same frequency, once the movement of
objects, the reflected frequency will change, and the greater the rate of movement of
objects , the greater the change in frequency, this effect is called the doppler effect. So
when we put the ultrasound probe on the maternal abdomen, the fetal heart beat
ultrasonic probe with respect to movement, the transmitted wave after reflection
across the heart generates frequency offset, from which we can calculate the fetus
heartbeat frequency per minute.
To get the SpO2 and pulse rate of pregnant women, through the red light from LED
and infrared light emitted by the outer periphery of the body parts, it can detect the
pulse wave generated by pulsating arterial blood flow caused by changes in light
absorption by the photoelectric detection circuit, and detect the light absorption rate
before and after pulsing, then we can calculate the SpO2 accordingly. Therefore, if
there is no pulsing or pulsing of pregnant women is too weak, the measurement can
not be performed.
The monitor has the noninvasive blood pressure measurement by Oscillometry,
namely an inflatable cuff occlusion of the arteries, blood flow is blocked, the cuff is
deflated start and detect blood vessels appear systolic, mean and diastolic blood
pressure, its accuracy is better than Korotkoff, and invasive blood pressure monitoring
were positively correlated.
7
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
13
5 6 7 8 9 10 11 12
14
8
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
13
5/6
7, 8,9,10,11,12
3
14
9
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
4
5
3,4,5
10
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
3
4
11
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
3 4 5 6 7
12
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Brightness 1 to 5 level
Factory default
13
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Default setup
Primary I,II,III,AVR,AVL,AVF,V-
Unit mmHg,kPa
Reset
15
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Device no.
Manometer test
Device no.
16
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
English On / off
Chinese On / off
2.2.2 Review
17
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
1 12 11
3 13
2
14
5 4 15
10
6
7
9
8
5 10
7 9
8
18
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
5 7
6
19
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
7
5 6
FHR
J21
LCD
TOCO FHR1
Touch screen
Fetal System J7
FM J2
monitor J8 board Inverter
FHR2 module
J6
Keypad Alarm lamp board
FM
J5 Encoder board
On/off
J12/13 J23 J24 J5
Scoring Motor Paper detecting Printer
module module 20
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
2.3.4 Accessories
BLT-F90 fetal/ maternal monitor including a 3 in 1 sensor (FHR1, TOCO & FM), Coupling
Reagent, printing paper, ECG cable, SpO2 probe, cuff and PVC tube. Besides, FHR2 sensor, fetal
waker & TEMP sensor are optional.
3 in 1 sensor
21
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
TEMP sensor
22
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Chapter 3 Disassembly
3.1 Prepare :
Note : please take out the battery before disassemble the machine.
1. Remove the screws which fix the machine and then open the machine.
23
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
2. Unplug the LCD cable and keypad cable.(Note: remove the screws before unplug the LCD
cable )
3. Unplug the wires which connect to the system board, and then remove the screws which fix
the system board.
system board
24
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
4. Unplug the wires which connect to the fetal function module, and then remove the screws
which fix the module.
5. Unplug the wires which connect to the multi-parameter module, and then remove the screws
which fix the module;
25
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
multi-parameter module
6. Remove the tube, and then unscrew the screws which fix the pump & valve set.
7. Unplug the wires which connect to the power supply module, and then remove the screws
which fix the module.
26
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Stepping motor
9. Remove the screws and then take out the printer head.
printer head
27
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
11. Remove the screws which fix the LCD and then take out the LCD panel.
Note: please unplug the power code and remove the battery before disassemble the machine.
28
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
1. Remove the screws which fix the machine, and then open then machine; before separate the
front panel, please unplug the cables which connect to the system board.
2.1 Disassemble the keyboard: remove the screws and then take out the keyboard;
29
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Screws and
keyboard
cable
2.2 Disassemble the encoder board: remove the screws and then take out the encoder board.
Encoder wire
2.3 Disassemble the interface board: remove the screws and then take out the interface board.
30
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Interface board
2.4 Disassemble the LCD panel: remove the screws and take out the LCD pnael;
LCD panel
2.5 Disassemble alarm lamp board and power indicator lamp: remove the screws and then
take out the boards.
31
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
3.1 Disassemble the fetal function module: unplug the cables and remove the screws, then
take out the fetal function module;
32
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
3.2 Disassemble the fetal function module: unplug the cables and remove the screws, then
take out the system board;
System board
3.3 Disassemble the stepping motor and printing head: remove the screws and take out the
stepping motor and printing head;
33
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Printing head
Stepping motor
34
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
35
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
36
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Chapter 4 Maintenance
4.1 Transport
To meet clinical use, design of the fetal monitor is very strong, but we nee to use it
with caution, and pay attention to maintenance, especially the ultrasound probe is a
ceramic chip, collision is not allowed.
4.2 Maintenance
4.3 Sterilization
Clean the outer surface and the probe according to the method described above, and
then wipe the surface of the probe with 70% alcohol or isopropyl. wipe the surface of
the probe with a dry and soft cloth, to remove any residual liquid substances.
Caution: Do not sterilize the device or the probe by low temperature steam or other
methods.
37
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
4.6 Troubleshooting
Fetal monitor is with quality and reliability, if any problems, please follow this table to check and troubleshooting.
38
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Common troubleshooting
Malfunction Solutions
Check if the power line is connected, the power socket switch is turned on,
No display on the LCD screen
the power indicator light is on and power is switched on.
1. Check if the NIBP unit and measurement mode are properly
set, if the size of the cuff is appropriate and the cuff is properly
placed.
2 The patient must not move during measurement. The patient is
speaking or not calm during the measurement. The cuff is
pressed by the patient or external objects.
Inaccurate NIBP
3. Check if there is a leakage in the cuff, air pipe or connector.
4. Check if the physical status of the patient affects measurement
result. When the patient is in shivering, convulsion, shock, low
body temperature, the measurement result will be inaccurate.
5. Try measuring repeatedly for several times, or make a
measurement on the other limb, or replace another cuff.
39
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
40
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
41
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual
Address: No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone, Zhuhai ,