BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

BLT-F90/F80/F50/F30 maternal /
fetal monitor service manual

1
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Declaration
Manufacturers have this non-public service manual by copyright, and the right to
be treated as confidential information. The service manual is as a reference for
machine maintenance and repair only, others have no right to expose this service
manual.
This service manual is protected by laws for copyright and proprietary information.
Without the written consent of the manufacturer, this service manual may not be used
of any part for photographic, copying, or translated into another language.
Contents of the service manual are subject to be changed without notice.

Precautions, warnings and suggestive description

To avoid the possibility of damage, be sure to comply with the following safety
instructions during maintenance of this instrument
1) The monitor contains some internal parts that can be repaired by the users, but it
should be performed by technical service persons that had been authorized by the
manufacturer.
2) The monitor is not classified as therapy equipment and shall not be used at
home.
3) Some mother parameters of the monitor can resist defibrillator. The defibrillation
not touch the patient, hospital beds and the instrument.
4) Before cleaning or repairing of the instrument, make sure the power supply is cut
off.
5) Check the reusable accessories regularly to see if they are damaged, such as the
sensor and the cables.If necessary, replace and dispose them properly as medical
waste accessories.
6) Warning: Do not use under flammable gas such as anesthetic, as there will be a
risk of explosion.
7) Warning: Do not dispose the batteries in a fire as they may explode.
8) Warning: Never touch signal input or output connector and patients in the same
time.
9) Caution: The instrument must be performed by authorized and qualified service
engineers.

2
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

10) Caution: This instrument is designed for continuous work, and drip-proof
closure devices, pls take care to avoid splashing water.
11) Caution: Keep the instrument clean, and avoid oscillation.
12) Caution: Do not do high-temperature sterilization process or use electron beam
and γ radiation sterilization.
13) Caution: Electromagnetic interference - Ensuring that the use of the
environment is away from the strong electromagnetic interference, such as radio
transmitters or mobile phones and other interference.
14) Caution: Before use it, the user must check the device to make sure it does not
significantly affect safety of the patient or have any equipment damage.
15) Caution: The following safety checking must be performed by properly trained
person, who has knowledge and practical experience. Usually, it should be checked
once a year or once a month or less. If you find obvious damage, we recommend you
to replace the damaged parts before using it.
16) Do testing according to the inspection procedures of the public institution.
17) Check the device for mechanical and functional damage.
18) Make sure the safety-related labels are easily identifiable.
19) Verify that the function of the device is consistent with the description of the
manual.
20 ) Caution:After the effective lifetime, the instrument shoulb be returned to the
manufacturer for recycle use in accordance with local regulations.
21 ) Caution: After use ,the battery in accordance with local regulations for proper
disposal.
22 ) Caution: If the instrument is not used in a long time , be sure to remove the
battery in time.
23 ) Caution: The battery should be stored in a cool and dry environment .
24 ) Caution:Do not put the metal objects together with the battery in storage, to
avoid accidental short circuit.
25 ) Caution:In the premise of meeting the needs of the clinical, we recommend
that ultrasonic irradiation time of the patient should be as short as possible .
26 ) Caution:When the instrument is moved from a cold environment to a warm
and humid place, do not use it immediately.
27 ) Caution : Keep the operating environment no vibration, no corrosion, no
combustible material , and avoid excessively high or low temperature and
humidity

3
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

28 ) Caution:If the instrument becomes wet or there is condensation water, stop


operation.
29 ) Caution:The antenatal fetal monitor is designed firm, anti-fall seismic, and
meet clinical needs for complex work. However, users should be careful when use it ,
in particular, the ultrasound probe is a ceramic chip , and can not throw touch and
should avoid surface scratches.
30 ) Caution: Water-based ultrasound gel is recommended, as oil-based ultrasound
gel may damage the surface of the probe, so oil-based ultrasound gel should be
prohibited.
31 ) Caution: After each use , remove the ultrasound gel on surface of the ultrasonic
probe. Clean the probe and machine by damp cloth with mild disinfectant or
detergent.
32 ) Caution: Use a soft cloth moistened with a solution of sodium hypochlorite
1000ppm for disinfection.
33 ) Caution: The machine and probe and other components can not be sterilized
through steam .
34 ) Caution: The pressure probe is non- waterproof , do not use ultrasound gel, and
should be absolutely avoided inflow of various liquids .
35 ) Caution: The power cord must be inserted at three-wire outlet , do not remove
the ground , do not use bad socket ;
36 ) Caution: After use , to prevent the break, do not make the cable wrapped
around the probe.
37 ) Caution: During monitoring, do not turn off the volume, as fetal heart tone is
very important to need to be careful to hear it: the accuracy of fetal heart rate is
controlled by the device and can not be adjusted by the user . If the fetal heart rate
results unreliable , use other methods such as using a stethoscope to verify , or
contact your local dealer or manufacturer for help.

4
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Content
BLT-F90/F80/F50/F30 maternal / fetal monitor service manual....................................................... 1
Chapter 1 General Introduction......................................................................................................... 6
1.1 The main features .......................................................................................................... 6
1.2 Operational principle..................................................................................................... 7
Chapter 2 Product Introduction ...................................................................................................... 8
2.1 Main unit .......................................................................................................................... 8
2.1.1 Front view ............................................................................................................. 8
2.1.2 left side view : ................................................................................................. 10
2.1.3 Right side view:................................................................................................ 11
2.1.4 Vear view............................................................................................................. 12
2.2 Operation menu .............................................................................................................. 13
2.2.1 Main menu .......................................................................................................... 13
2.2.2 Review ................................................................................................................ 17
2.2.3 Maternal info. ...................................................................................................... 17
2.2.4 volume setup ....................................................................................................... 17
2.3 Internal structure ............................................................................................................ 18
2.3.1 Internal modulus.................................................................................................. 18
2.3.2 F90/F80 LCD panel internal structure ................................................................ 19
2.3.3 Schematic diagram .............................................................................................. 20
2.3.4 Accessories.......................................................................................................... 21
Chapter 3 Disassembly................................................................................................................. 23
3.1 Prepare : ..................................................................................................................... 23
3.2 Disassembly procedure .................................................................................................. 23
3.2.1 F90/F80 disassembly .............................................................................................. 23
3.2.2 F50/F30 disassembly ........................................................................................... 28
3.4 Exploded view............................................................................................................. 34
3.4.2 F50/F30 exploded view ....................................................................................... 36
Chapter 4 Maintenance ................................................................................................................... 37
4.1 Transport ........................................................................................................................... 37
4.2 Maintenance ...................................................................................................................... 37
4.3 Sterilization ....................................................................................................................... 37
4.4 Ultrasound Gel .................................................................................................................. 38
4.5 Clean the printer ................................................................................................................ 38
4.6 Troubleshooting ................................................................................................................ 38
4.7 Warranty and Maintenance................................................................................................ 41

5
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Chapter 1 General Introduction


Fetal monitor is a bran-new, well-designed monitor, which greatly facilitates
medical staff and pregnant women, as is easy to operate, making the daily monitoring
work easier, efficient and cost becomes lower.
BLT-F90 Fetal monitor applies to external monitoring. It contains a FHR probe, a
TOCO probe, a SpO2 probe, a blood pressure cuff, one ECG cable, a temperature
probe and a FM button for the pregnant woman. Its performance is in line with the
latest world standards. The entire machine design is simple, reliable and friendly use.

1.1 The main features

Display: 12.1” TFT-LCD


Fetal heart rate, uterine pressure, fetal activities monitoring data and curves, fetal
signs, doctors event flag, time, alarm status display in the same screen.
While it is detecting the fetal parameters, such as the fetal heart rate, uterine pressure,
fetal activities, it can also monitor mother's ECG, noninvasive blood pressure, oxygen
saturation, pulse, respiration, body temperature, at the same time.
With the built-in twins function modules, it can realize twins monitoring function
with an extra fetal monitoring probes.
With the unique function to identify the normal range of fetal heart, the monitoring
curve can be apprehended at a glance.
The automatic monitoring mode allow users to set up parameters of working mode
according to their personalized habit.
To improve monitoring sensitivity, it can support clinical event identification, and
identification of fetal stimulator action.
With the functions to store and playback the monitoring data, users can store and
manage the historical data.
With the built-in thermal recorder and 152mm Z type printing paper, users can print
in real-time, or print the historical reviewing data, and can also print with regular
time.

6
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

With the optional CTG evaluation module, users can print the evaluation result in
real-time and in segmented, for users’ analysis and reference.
With the built-in network interface, it can connect to the central station.
With built-in pluggable and rechargeable lithium battery.
Sound and light alarms in three levels, and the alarm volume is adjustable.

1.2 Operational principle

BLT-F90 fetal monitor detect fetal heart rate by non-invasive doppler ultrasound.
We know that a certain frequency of the ultrasonic obstacles encountered in the
dissemination of reflection to occur, if the object is stationary, then the reflected wave
frequency and the transmitted wave at the same frequency, once the movement of
objects, the reflected frequency will change, and the greater the rate of movement of
objects , the greater the change in frequency, this effect is called the doppler effect. So
when we put the ultrasound probe on the maternal abdomen, the fetal heart beat
ultrasonic probe with respect to movement, the transmitted wave after reflection
across the heart generates frequency offset, from which we can calculate the fetus
heartbeat frequency per minute.
To get the SpO2 and pulse rate of pregnant women, through the red light from LED
and infrared light emitted by the outer periphery of the body parts, it can detect the
pulse wave generated by pulsating arterial blood flow caused by changes in light
absorption by the photoelectric detection circuit, and detect the light absorption rate
before and after pulsing, then we can calculate the SpO2 accordingly. Therefore, if
there is no pulsing or pulsing of pregnant women is too weak, the measurement can
not be performed.
The monitor has the noninvasive blood pressure measurement by Oscillometry,
namely an inflatable cuff occlusion of the arteries, blood flow is blocked, the cuff is
deflated start and detect blood vessels appear systolic, mean and diastolic blood
pressure, its accuracy is better than Korotkoff, and invasive blood pressure monitoring
were positively correlated.

7
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Chapter 2 Product Introduction


2.1 Main unit

2.1.1 Front view

13

5 6 7 8 9 10 11 12

14

8
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

13
5/6

7, 8,9,10,11,12

3
14

No. Item No. Item


1 Alarm lamp 8 Alarm pausing/acknowledging
2 Display 9 FHR1/ FHR2 Switch
3 On/Off 10 Clinical event marking
4 Printer 11 TOCO Reset

5 Power indicating lamp 12 Print button


6 Charging lamp 13 Trim Knob
7 NIBP 14 Printing paper

9
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

2.1.2 left side view :

4
5

3,4,5

No. Item No. Item


1 Handel 4 FHR2 sensor connector
2 Accessories set-down location 5 Fetus waker connector

10
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

3 Ultrasonic triple-sensor connector

2.1.3 Right side view:


view:

3
4

No. Item No. Item


1 ECG connector 3 NIBP connector
2 SpO2 connector 4 TEMP connector

11
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

2.1.4 Vear view

3 4 5 6 7

No. Item No. Item


1 Front panel bracket 5 COM port
2 heat emission hole 6 Equipotential grounding terminal
3 Network interface 7 AC power input connector
4 USB interface

12
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

2.2 Operation menu

2.2.1 Main menu

Screen layout Fetal screen,fetal cascade,fetal & maternal,big numerics,maternal screen

System setup Language select English / Chinese

Display mode Monitor / DEMO

Brightness 1 to 5 level

Factory default

System time Y/M/D/H/M/S

Color setup 10 colors

Print setup Print speed 1 cm/min, 2 cm/min, 3cm/min

Print density 1 to 5 level

Timing print off,10, 20, 30, 40 min

Print score On / off

Score select FHR1, FHR2

13
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Fetal param. Trace offset 20bpm, 30bpm, 40bpm

FHR channel FHR1, FHR2

Alarm delay On / off

TOCO reset 0, 5, 10, 15

FM count Auto / Manual

FM strength trace On / off

FM threshold 30, 40, 50, 60, 70

Fetal alarm FHR1 Alarm level, alarm limited


setup

FHR2 Alarm level, alarm limited

Default setup

Maternal alarm SpO2 Alarm level, alarm limited


setup

HR Alarm level, alarm limited

SYS Alarm level, alarm limited

DIA Alarm level, alarm limited

MEAN Alarm level, alarm limited

RESP Alarm level, alarm limited


14
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

TEMP Alarm level, alarm limited

Default setup Alarm level, alarm limited

ECG setup Lead type 3/5 lead

Primary I,II,III,AVR,AVL,AVF,V-

Wave gain 0.25X,0.5X,1X,2X,4X

Wave speed 6.25 mm/s,12.5 mm/s, 25mm/s

ECG mode AC filter, EMG filter,Drift filter

NIBP setup Measure mode Auto / Manual

Measure interval 3,5,10,15,30,60,90,120,240min

Initial inflation 100-280mmHg

Unit mmHg,kPa

Reset

SpO2 setup Wave speed 6.25 mm/s, 12.5 mm/s, 25mm/s

Average time 4s,8s,16s

15
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

RESP setup Wave gain 0.25X,0.5X,1X,2X,4X

Wave speed 6.25 mm/s, 12.5 mm/s, 25mm/s

RESP lead RA_LL,RA_LA

Apnea alarm time 10-60s

TEMP setup Unit ℃,F

Machine info. Software version

Device no.

Machine Standard signal

AC filter freq. 50HZ, 6HZ

Manometer test

Overpressure test (H)

Overpressure test (S)

Air leakage test

NIBP calibrate 100-254 mmHg

Touch screen calibrate

Device no.
16
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Factory setup Maternal param. On / off

Touch screen On / off

English On / off

Chinese On / off

Scoring module On / off

2.2.2 Review

Enter menu to review and manager the data.

2.2.3 Maternal info.

Name, Age, Gest. Week, Gest. Day, Fetal no.

Enter menu to check maternal information.

2.2.4 volume setup

FHR volume setup Off,1 to 5 level

Alarm volume setup Off, 1 to 3 level

Enter menu to set the FHR volume and alarm volume.

17
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

2.3 Internal structure

2.3.1 Internal modulus

F90/F80 Internal modulus:

1 12 11

3 13
2
14
5 4 15

10
6
7

9
8

F50/F30 Internal modulus:

5 10

7 9
8

18
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

No. Item No. Item


1 Speaker 9 Paper storehouse
2 SMPS 10 System board
3 Battery interface 11 Multi-parameter module(F90)
4 3 in 1 transducer interface 12 ECG interface(F90)
5 Waker interface 13 SpO2 interface(F90)
6 Fatal monitor module 14 NIBP interface(F90)
7 Motor 15 TEMP interface(F90)
8 Printer

2.3.2 F90/F80 LCD panel internal structure

5 7
6

19
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

7
5 6

No. Item No. Item


1 Alarm lamb board 5 Encoder board
2 Inverter (F90/F80) 6 Keypad
3 Touch screen interface 7 On/off switch board
4 LCD 8 Interface board (F50/F30)

2.3.3 Schematic diagram


COM
Multi-parameter module
Battery SMPS I/O interface board USB
(F90/F50)
J9 J20 J10 J1
RJ 45

FHR
J21
LCD
TOCO FHR1
Touch screen
Fetal System J7
FM J2
monitor J8 board Inverter
FHR2 module
J6
Keypad Alarm lamp board
FM
J5 Encoder board

On/off
J12/13 J23 J24 J5
Scoring Motor Paper detecting Printer
module module 20
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

2.3.4 Accessories

BLT-F90 fetal/ maternal monitor including a 3 in 1 sensor (FHR1, TOCO & FM), Coupling
Reagent, printing paper, ECG cable, SpO2 probe, cuff and PVC tube. Besides, FHR2 sensor, fetal
waker & TEMP sensor are optional.

3 in 1 sensor

ECG cable SpO2 probe

PVC tube cuff

21
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Fetal waker FHR2 sensor

TEMP sensor

22
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Chapter 3 Disassembly
3.1 Prepare :

Tool : Philips screw drive

3.2 Disassembly procedure

3.2.1 F90/F80 disassembly

Note : please take out the battery before disassemble the machine.

1. Remove the screws which fix the machine and then open the machine.

23
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

2. Unplug the LCD cable and keypad cable.(Note: remove the screws before unplug the LCD
cable )

3. Unplug the wires which connect to the system board, and then remove the screws which fix
the system board.

system board

24
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

4. Unplug the wires which connect to the fetal function module, and then remove the screws
which fix the module.

fetal function module

5. Unplug the wires which connect to the multi-parameter module, and then remove the screws
which fix the module;

25
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

multi-parameter module

6. Remove the tube, and then unscrew the screws which fix the pump & valve set.

pump & valve set

7. Unplug the wires which connect to the power supply module, and then remove the screws
which fix the module.

26
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Power supply module

8. Remove the screws and take out the stepping motor.

Stepping motor

9. Remove the screws and then take out the printer head.

printer head

27
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

10. Remove the screws which fix the LCD panel.

11. Remove the screws which fix the LCD and then take out the LCD panel.

3.2.2 F50/F30 disassembly


disassembly

Note: please unplug the power code and remove the battery before disassemble the machine.

28
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

1. Remove the screws which fix the machine, and then open then machine; before separate the
front panel, please unplug the cables which connect to the system board.

2. Disassemble the front panel:

2.1 Disassemble the keyboard: remove the screws and then take out the keyboard;

29
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Screws and
keyboard
cable

Keyboard and silica gel keypad

2.2 Disassemble the encoder board: remove the screws and then take out the encoder board.

Encoder board knob

Encoder wire

2.3 Disassemble the interface board: remove the screws and then take out the interface board.

30
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Interface board

2.4 Disassemble the LCD panel: remove the screws and take out the LCD pnael;

LCD panel

2.5 Disassemble alarm lamp board and power indicator lamp: remove the screws and then
take out the boards.

31
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Alarm lamp board & wire

Power indictor board & wire


3. Disassemble the internal bards:

3.1 Disassemble the fetal function module: unplug the cables and remove the screws, then
take out the fetal function module;

32
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Fetal function module

3.2 Disassemble the fetal function module: unplug the cables and remove the screws, then
take out the system board;

System board

3.3 Disassemble the stepping motor and printing head: remove the screws and take out the
stepping motor and printing head;

33
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Printing head

Stepping motor

3.4 Exploded view

3.4.1 F90/F80 exploded view

34
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

No. Item No. Item


1 Pad 13 Stepping motor plate
2 Printer cover 14 Wall mount bracket
3 Printer gate 15 Printing paper box
4 Speaker 16 Printer plate
5 Speaker plate 17 Thermal head
6 Front cover 18 Printer roller
7 Knob 19 Side connector panel
8 Alarm lamp cover 20 LCD fixing plate
9 System board 21 Sensor plate
10 Damping shaft 22 LCD
11 Side connector plate 23 Power supply module plate
12 Stepping motor 24

35
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

3.4.2 F50/F30 exploded view

No. Item No. Item


1 Pad 2 Printer cover
3 Printer gate 4 Speaker
5 Speaker plate 6 Front cover
7 Knob 8 Alarm lamp cover
9 System board 10 Damping shaft
11 Side connector plate 12 Stepping motor
13 Stepping motor plate 14 Wall mount bracket
15 Printing paper box 16 Printer plate
17 Thermal head 18 Printer roller
19 Speaker fixing plate 20 Fetal function module
21 ON/OFF button 22 LCD
23 Alarm lamp board

36
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Chapter 4 Maintenance
4.1 Transport
To meet clinical use, design of the fetal monitor is very strong, but we nee to use it
with caution, and pay attention to maintenance, especially the ultrasound probe is a
ceramic chip, collision is not allowed.

4.2 Maintenance

In addition to cleaning, there is no additional maintenance for the fetal monitor.


Keep the outer surface of the instrument clean and no dust. Clean the outer surface of
the instrument (including LCD display) with a dry, soft cloth. If necessary, clean the
casing with soap or water dampened soft cloth.
Clean the redundant ultrasound gel with a soft cloth. Use soap and water only to clean
the probe.
Caution: Do not use strong solvents, such as acetone.
Caution: Do not use abrasive materials (such as steel wool or silver polish).
Caution: Do not allow any liquid into the cabinet, nor can any part of the instrument
immersed in the liquid.
Caution: Do not spill liquid into the instrument when cleaning it.
Caution: Do not allow any cleaning solution left in the instrument's surface.
Note: The probe surface can be wiped with 70% alcohol or isopropyl, and get dry
naturally or with a clean, dry cloth.

4.3 Sterilization

Clean the outer surface and the probe according to the method described above, and
then wipe the surface of the probe with 70% alcohol or isopropyl. wipe the surface of
the probe with a dry and soft cloth, to remove any residual liquid substances.
Caution: Do not sterilize the device or the probe by low temperature steam or other
methods.

37
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

4.4 Ultrasound Gel


Recommend the use of water-based onboard ultrasound gel. Never use oil-based
ultrasound gel.If the probe is damaged by the use of oil-based ultrasound gel,
warranty of the device will be automatically suspended.

4.5 Clean the printer


The thermal print head can work for more than 20 years without stoppage, this is
only the gurranty of electrical aspects. It is related largely to the paper used and
cleanliness of the operating environment. When the print is not clear or there are some
blank areas, the first step is cleaning the print head, methods and steps are as follows:
1. Turn off power supply of the monitor.
2. Turn on the printer panel.
3. Put the cotton swab with ethanol into the thermal element portion of the print
head (thin black thermal band partially visible in the print head), and move
around to gently wipe it, wipe several times especially in the area of print is not
clear. A few minutes after finished cleaning, tun on the device to confirm..
4. If the phenomenon is not completely eliminated, repeat step 3 until satisfaction.

4.6 Troubleshooting
Fetal monitor is with quality and reliability, if any problems, please follow this table to check and troubleshooting.

Malfunction Reason Solution

No display after turn on


Power code poor connect; Check the power code
the machine
FHR sensor poor connect; Re-connect the sensor;
The probe in a wrong position, cannot find To adjust the probe location;
the HR Add coupling reagent;
FHR abnormal
No or just a little coupling reagent Adjust the sensor location after signal
Fetal / gravida move recovery;
Sensor is defective Change a good sensor.
sensor poor connect; Re –connect the sensor;
The probe in a wrong position; Adjust the sensor location ;
TOCO abnormal
No TOCO reset; TOCO reset
Gravida no uterine contraction Wait for uterine contraction.
No volume Volume is set to be 0 Set the volume.
No printing curve line Re-install the printer paper (thermal
Wrong position of printer paper
when printing, or FHR surface facing up )
Wrong position of thermal surface
curve line in TOCO zone,

38
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

TOCO curve line in FHR


zone
Printing not clear, or Printing density too light; Reset the printing density;
some area cannot print Use the poor printing paper; Change printing paper
empty Printer head has dust. Clean the printer head.
Printing paper doesn’t is a good
The printing location of Adjust the location of the paper;
location;
curve line not correct. Replace the right paper.
Use another brand printing paper

Common troubleshooting
Malfunction Solutions
Check if the power line is connected, the power socket switch is turned on,
No display on the LCD screen
the power indicator light is on and power is switched on.
1. Check if the NIBP unit and measurement mode are properly
set, if the size of the cuff is appropriate and the cuff is properly
placed.
2 The patient must not move during measurement. The patient is
speaking or not calm during the measurement. The cuff is
pressed by the patient or external objects.
Inaccurate NIBP
3. Check if there is a leakage in the cuff, air pipe or connector.
4. Check if the physical status of the patient affects measurement
result. When the patient is in shivering, convulsion, shock, low
body temperature, the measurement result will be inaccurate.
5. Try measuring repeatedly for several times, or make a
measurement on the other limb, or replace another cuff.

39
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

1. Test the grounding of the monitor casing.


2. Inaccurate HR caused by great interference of the ECG occurs
when the diagnosis mode is selected for the ECG monitoring
mode.
3. The HR of the patient with pacemaker is inaccurate.
4. Observe if the ECG cable and the patient move, the electrode is
Inaccurate HR
properly connected.
5. Observe the ECG on the display screen to check if there is
abnormal QRS comprehensive wave.
6. Observe the ECG waveform on the display screen to check the
monitor gives out a pulse sound when there is R wave.
7. Try to display HR by use of pulse rate of Plath first.
1. Check if the ECG mode is correct. It should be set in surgery
mode during monitoring of surgery, and may be set in
monitoring mode during ordinary monitoring. If the ECG
waveform of the patient has big shift or big disturbance while
in monitoring mode, it should be set in surgery mode.
2. Check if the ECG electrode is placed firmly and the position
is correct. Before placing the electrode, the skin of the patient
ECG waveform shift or big should be cleaned and the grease or sweat on skin should be
disturbance wiped off with alcohol. If necessary, shave off the hair at the
place where the electrode is to be placed, abrade the stratum
corneum of skin of the patient, and clean with alcohol.
3. Check if the power socket protection is properly grounded.
For big disturbance, equal-potential ground wires should be
connected.
4. Replace a new ECG cable to check whether this problem is
solved or not.
1. Check if the SPO2 probe is fully cut in or disconnected, the
sensor of the SPO2 probe is broken or disconnected.
2. The patient moves, or the SPO2 probe and BP cuff are placed on
SpO2 could not be measured.
the same limb.
3. Try to move and place the SPO2 probe on the different places.

40
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Display is normal during the


switch on and self-testing, but
Switch off the monitor immediately, plug up the power plug, and
it could not enter the master
contact the after-sales service department or the authorized
screen.
distributor of Company.
The LCD screen could not
display function modules.
Switch off the monitor immediately, plug up the power plug, and
Dark display, the others are
contact the after-sales service department or the authorized
normal.
distributor of Company.
No display on one part of the
LCD, the other is normal.
Other abnormity occurs.

4.7 Warranty and Maintenance


All products from our company, have one year warranty and maintenance
during the whole lifetime.
About the service, maintenance and any other issues related to this or any of
our other product related questions, please contact us.
In case you want to return the product, clean it according to local disinfection
and sterilization procedures, and provide the appropriate documentation that state
the products current status. Please send this file attached to the box of product.

41
BLT-F90/F80/F50/F30 maternal/fetal monitor service manual

Product name: maternal/fetal monitor

Product type: BLT-F90/F80/F50/F30

Manufacturer: Guangdong Biolight Meditech Co., Ltd.

Address: No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone, Zhuhai ,
P.R.China
Fax: +86-756-3399919

Post code: 519085

Toll-free consultation hot line: +86-400-8818-233