Professional Documents
Culture Documents
Thursday A - Reverse Osmosis
Thursday A - Reverse Osmosis
Objective: The objective of this lab was to assess the functionality of the reverse osmosis membrane
separation unit in the desalination of brackish (salt) water. The experiment aimed to determine the
separation of brackish water between the retentate and permeate streams as a function of pressure
drop across the membrane, to characterize the membrane unit for salt rejection and to evaluate the
operational cost of running the membrane unit.
Summary: In this experiment, conductivity of permeate and retentate streams at five different
pressure drop values that ranged from 300 psi to 500 psi was measured over a period of 10 minutes
with 1-minute intervals. The conductivities of salt in the retentate and permeate streams collected
were measured using the electron probe. The conductivity values determined for the permeate and
retentate streams were related to salt concentration in those streams by plotting a calibration curve
based on the known concentration of the salt solutions. In addition to determining the salt
concentrations in the permeate and retentate streams as a function of time for each pressure drop, the
values of salt rejection, which represents the percentage of salt removed by the membrane from the
feed, were also determined at each pressure drop to determine which pressure drop gave the
maximum salt rejection and to evaluate the trend of salt rejection as a function of time. Lastly, the
operational cost of running the membrane unit at each pressure drop tested over a period of one year
was calculated to determine feasibility.
Conclusions:
Desalination of brackish water to drinkable water using the membrane system can’t be
achieved due to the salt concentration in the permeate being greater than the allowable salt
concentration (0.2 mg/L) in potable water as per EPA standards
Negative slopes were obtained for the linear regression plots of osmotic pressure as a
function of retentate concentration at all values of pressure drop tested
Positive slopes were obtained for the linear regression plots of osmotic pressure as a function
of permeate concentration at all values of pressure drops tested
Salt rejection decreased over time for all pressure drop values except for the trial conducted
at 300 psi. Highest value of salt rejection was obtained at ΔP= 500 psi
Operating cost per year to run the membrane unit increased as the pressure drop increased
At the pressure drop of 500 psi, the operating cost to run the membrane unit for 8 hours per
day over a period of 1 year was calculated to be $129.5
Recommendations:
Permeate staging configuration should be used to obtain permeate stream with lower salinity
Permeate flow rate should be measured to calculate the permeability of the salt across the
membrane
Percent recovery of water can be calculated to further evaluate the efficiency of the
membrane
INTRODUCTION
increase in population, the natural supply of potable water remains relatively static and is
constrained geographically. One of the ways to increase the supply of potable water is
desalination. Desalination is the process of removing sodium chloride (salt) from saline water in
order to obtain potable water that is suitable for human consumption. Most common desalination
methods employ reverse osmosis in which salt water is forced through a membrane that allows
water molecules to pass but blocks the molecules of salt. Reverse osmosis is among one of the
Discussion of the concept of osmosis can help understand the phenomenon of reverse
osmosis. An example of a semi permeable membrane that is only permeable to water can be used
compartments in a tank, and one of the compartments has salt solution and the other has pure
water solution. Then the water will pass from the pure water compartment to the salt water
compartment, causing an increase in the liquid level in the salt water compartment. For the
system to reach equilibrium, concentrations on both sides of the membrane should be equal.
Hence, the liquid level will keep on increasing in the salt water compartment until there is
enough pressure to stop the osmosis. This pressure in known as the osmotic pressure. However,
if a pressure greater than the osmotic pressure is applied to the concentrated salt solution, then
according to Le Chatelier’s principle, the direction of the flow of water will be reversed, and the
water will move against the concentration gradient, producing pure water since the membrane is
reverse osmosis membrane separation process are evolved when the water flows back against the
concentration gradient when pressure greater than the osmotic pressure is applied to the salt
solution. As the difference between the external pressure and osmotic pressure increases, more
water is able to flow back through the membrane leading to more salt rejection. However, as
more salt is rejected by the semi permeable membrane, a boundary layer is formed near the
surface of the membrane. This phenomena is commonly known as concentration polarization and
is characterized by a buildup of salt which leads to the increase in the salt concentration on the
membrane over time, which eventually reduces the separation performance of the membrane and
increases the potential for membrane fouling. However, if the reverse osmosis system is
operating at perfect conditions, then a cross-flow mechanism can cycle or flush out the
accumulated particles minimizing the leakage of salt to the permeate side leading to stable
effectively desalinating brackish water to potable water will be explored. The analysis carried out
will assess the salt concentration of the permeate and retentate streams as a function of the
osmotic pressure in addition to the characterization of salt rejection as a function of time and the
determination of the cost to operate the membrane over a period of one year. To achieve these
objectives, the experiment will be conducted at 5 different pressure drops ranging from 300-500
The RO membrane is only permeable to water, therefore it doesn’t allow salt to pass
through it creating two streams: permeate and retentate. To determine the percentage of the salt
that is rejected, the following equation will be used (Seader, et al. 2011):
(1)
Where, SR, represents salt rejection, Cpermeate is the concentration of the permeate in g/L and C tank
is the concentration of the tank (flow into the membrane) in g/L. Salt rejection determines the
effectiveness of the RO membrane in removing salt from water. The higher the value of salt
rejection, the better the system is performing. A low salt rejection value means that the
membrane requires cleaning or replacement. Hence, it is one of the main performance indicators
In addition, to the salt rejection, osmotic pressure will also be calculated since the driving
force of the separation of water and salt is the pressure gradient created by the pump as it pushes
the salt water through the semi-permeable membrane. The osmotic pressure will be calculated
(2)
Where, T is the temperature of the inlet flow to the membrane (K), C salt is the salt concentration
of the inlet salt water (g/L), and MW is the molecular weight of sodium chloride (58.44 g/mol).
Lastly, cost analysis will be conducted on the reverse osmosis membrane unit, to find the
operational cost of running the system per year. To determine the operational cost, the following
Where, the term kWh/yr will be calculated by multiplying power required to operate the reverse
osmosis unit with the number of hours per day the membrane system will be run for.
EXPERIMENTAL PROCEDURE
Materials:
In order to separate salt from brackish water using the reverse osmosis membrane unit, the
experiment utilized deionized water and salt as the experimental reagents. In addition to that,
three major pieces of equipment used to conduct the experiment were: electron probe (Omron-
122), centrifugal pump (Hydra-Cell) and a membrane separation unit (Alfa Laval), shown in the
Methods:
In this experiment an Alfa Laval membrane was used to desalinate a solution of brackish
water (35 g/L). A positive displacement pump was used to pump water through the membrane.
Permeate stream was constantly removed from the system and the retentate was recycled back to
the feed tank. Prior to starting the experiment, all valves were closed and glass shields were
positioned in front of the pump and the membrane separation unit, to avoid any safety concerns
Salt water was filled into the tank via PMP-2 to the desired water level in addition to the salt
water already present inside the feed tank. Once the water level reached the desired point, pump
PMP-1 was turned on by plugging the power cord into the power source. The reading of PI-3
was adjusted to meet the desired pressure drop value and valves, V-12 and V-13 were opened to
due to the presence of the recycle stream, the conductivity of the permeate and retentate streams
was measured at 5 different pressure drop values (300, 350, 400, 450 and 500 psi) over a time
period of 10 minutes with 1 minute intervals. The conductivity probe was used to measure the
conductivities of the sample and equation of the linear regression of the calibration curve was
When the experiment was performed, V-12 and V-13 were closed and the set point of the
flow controller, FC-1 was lowered to 0 Hz to shut off the membrane separation unit.
Prior to conducting the experiment, conductivity of known salt concentrations (0-25 g/L
with 5 g/L intervals) was measured in order to prepare a calibration curve of conductivity as a
function of salt concentration. The calibration curve is shown in Figure 2B in the Appendix
below.
The conductivity values of the permeate and retentate streams were measured during the
lab at pressure drops ranging from 300-500 psi over a period of 10 minutes with 1 minute
intervals. The conductivity values at all respective pressure drops and time points determined for
the permeate and retentate were converted to concentration values using equation of the linear
regression line of the calibration curve shown in Appendix B. Table 1 shows the concentration
values of the permeate and retentate streams at each time point for a period of 10 minutes for the
pressure drop of 500 psi. Concentration values of retentate and permeate streams at other
The salt rejection values for all the trials conducted at various pressure drops were calculated
using equation (1). Table 2 below shows how the salt rejection changes with respect to time at a
Table 2. Salt rejection values for the trial conducted at the pressure drop of 500 psi at each
time interval
Time (min) Salt rejection
1 0.693
2 0.659
3 0.619
4 0.591
5 0.568
6 0.568
7 0.557
8 0.563
9 0.557
10 0.557
Relationship between osmotic pressure and outlet concentrations (retentate and permeate)
To calculate the osmotic pressure for each trial conducted at different pressure drop, equation (2)
was used. Since the permeate concentration was used to calculate the osmotic pressure, all linear
regressions of osmotic pressure versus permeate concentration had a slope of 22.4016 (appendix
C). Linear regression plots for osmotic pressure versus retentate concentration were also
generated and analyzed for each trial performed. Figure 1 below shows the linear regression plot
105
100
Osmotic Pressure (atm)
95
90
85
12.75 12.80 12.85 12.90 12.95 13.00 13.05
Retentate Concentration (g/L)
Figure 1. Linear regression plot for osmotic pressure vs retentate concentration (g/L) at the
pressure drop of 400 psi. The linear regression equation is:
Since the slope of the linear regression of osmotic pressure versus retentate concentration is
negative with a p-value of 0.014, it can be concluded that for this trial the retentate concentration
decreased with an increase in the osmotic pressure. However only 3 out of 5 trials conducted
showed a similar trend (i.e p-value was less than or equal to 0.05). The results for all 5 trials
conducted at pressure drops ranging from 300-500 psi are shown in table 3 below
Table 3. Slopes and their respective p-values and 95% CI obtained for the linear regression
between osmotic pressure as a function of retentate concentration at the pressure drops tested
Pressure drop (psi) Slope (atm g/L) P-value 95% CI (atm g/L)
300 - 15.2 0.251 (-43.5, 13.1)
350 - 6.3 0.017 (-11.2, -1.5)
400 - 32.5 0.014 (-56.4, -8.7)
450 - 19.8 0.05 (-39.5, -0.02)
500 - 3.1 0.794 ( -29.4, 23.2)
As stated earlier, only trials conducted at the pressure drop values of 350, 400 and 450 psi
produced linear regression slopes with p-values less than or equal to 0.05. This supports the
conclusion that osmotic pressure increases as retentate concentration decreases for each of these
trials. However a similar conclusion cannot be made for the trials conducted at the pressure drop
of 300 and 500 psi, because the linear regression slopes had p-values greater than 0.05. It should
be noted that all five linear regressions of osmotic pressure as a function of retentate
pressure and retentate concentration. This could be explained by the fact that the retentate stream
was continuously being recycled into the RO membrane feed, and thus overtime the salinity of
the feed increased. This increase in feed salinity over time caused an increase in osmotic pressure
because the concentration gradient of the solution across the membrane was continuously
increasing. The regressions and their respective residual plots for each trial conducted at pressure
drop of 300, 350, 450 and 500 psi can be found in appendix C.
Salt rejection over time at pressure drops tested
0.62
0.61
0.60
0.59
Salt rejection
0.58
0.57
0.56
0.55
0.54
0.53
0 2 4 6 8 10
Time (min)
Figure 2. Salt rejection variance over time for the pressure drop of 350 psi. The linear regression
equation is . The 95% CI for the slope is
(-0.010581, -0.006371)
Looking at the trend of the salt rejection as a function of time, it is apparent that there is an
inverse relationship between salt rejection and time. The p-value (0.000) for the slope of this
regression confirms this relationship that as time increases the salt rejection decreases because
the p-value obtained is less than 0.05. The results for the remaining trials conducted are shown in
table 4 below:
Table 4. Slope values determined from the linear regression of salt rejection as a function of time
at the pressure drops tested
From the values of the slopes of the linear regression of salt rejection as a function of time at all
the trials conducted at the indicated pressure drop values illustrated in table 4 above, it can be
seen that for 4 out of 5 trials conducted at various pressure drops the salt rejection decreased as
the time passed by because the slopes obtained were negative. The p-values for all the trials
conducted at various pressure drops were lower than 0.05, meaning the relationship seen in each
regression is valid meaning salt rejection decreases with time since majority of the trials
conducted showed that trend. The observed trend can be explained by given that the retentate
stream is continuously being recycled during each experimental run conducted at different values
of pressure drop, which will cause the salt concentration of the RO membrane feed stream to
increase over time, eventually causing the amount of salt being removed in the membrane
(permeate stream) to increase. This increase in permeate concentration results in a lower salt
rejection, which support the conclusion that salt rejection decreases over time. The linear
regression plots for salt rejection as a function of time for pressure drops of 300, 400, 450 and
500 psi and their respective residual plots can be found in appendix D.
The Environmental Protection Agency (EPA) recommends that the concentration of salt in
drinking water should not exceed 20 mg/L because if the concentration of salt in water exceeds
this particular value, it may result in adverse health effects. The maximum salt rejection obtained
in this experiment was 0.69 (at the pressure drop of 500 psi), which produced permeate salt
concentrations above the recommended limit of 20 mg/L. Therefore, a cost analysis was
performed to predict the cost to run the membrane system eight hours per day for an entire year,
rather than calculating the profit that could be made by selling purified (drinkable) water. This
was done by first calculating pump power, using the amps required to run the pump at each
pressure drop and multiplying it by the outlet voltage (208V). Pump power was then multiplied
by the number of hours the pump would run over a year (2,920 hours) and the average
commercial cost of a kWh in Easton, PA (5.61¢) to determine the operational cost per year for
running the membrane unit. A linear regression of the operational cost per year versus pressure
drop values tested in the experiment was generated, shown in figure 3 below.
130
120
Cost per year (USD)
110
100
90
80
70
300 350 400 450 500
Pressure (PSI)
Figure 3. Linear regression of pump pressure (psi) versus the operational cost of running the
pump per annum. Regression equation is:
From the linear regression of operational cost per year versus the pressure drop values, it can be
concluded that the cost to run the pump increases as the pressure drop increases as seen by the
positive relationship. Since, the maximum salt rejection value was obtained at the pressure drop
of 500 psi, the reverse osmosis membrane system should be run at the pressure drop of 500 psi.
If the RO membrane unit is run at the pressure drop of 500 psi 8 hours a day for every day of the
year, the cost incurred would be approximately $129.5. The low cost makes sense because the
amount of amps required to run the pump were considerably low due to the fact that the RO
globe valve was not fully opened (only 0.2 turns). The cost per year for every pressure drop
CONCLUSIONS
The experiment aimed to characterize the functionality of the reverse osmosis membrane
in desalinating brackish water by examining how brackish water separates between the permeate
and retentate streams a function of membrane pressure, calculating the salt rejection and the
operational costs incurred to run the membrane unit over a period of one year.
Results indicate that the current reverse osmosis membrane system is inefficient in
desalinating brackish water to potable water because the permeate concentration obtained at all
the pressure drop values tested is not below the standard allowable limit of salt concentration (20
mg/L) in potable water as per the EPA regulations. The linear regression plots prepared for the
the osmotic pressure and the retentate concentration, however only for the trials conducted at the
pressure drops of 350, 400 and 450 psi, it can be concluded with 95% confidence that as the
osmotic pressure increases the retentate concentration decreases because the p values obtained
were less than 0.05. Furthermore, it was determined that the salt rejection of the membrane
decreased over time for all trials except for the one conducted at the pressure of 300 psi.
When a cost analysis was performed on the RO system, it was found that the cost per
year (USD) to run the system increased as pump pressure increased (regression line had a p-
value of 0.000). At a pump pressure of 500 psi the cost to run the system 8 hours a day, 365 days
a year, was determined to be $129.5. The amount of the cost determined can be explained by the
low ampere requirements by the pump because the RO globe valve not fully opened (only 0.2
turns).
RECOMMENDATIONS
A general trend in most of the graphs of salt rejection plotted as a function of time
showed a decrease in salt rejection with time. This trend can be explained by a number of
reasons ranging from mechanical failures to membrane fouling. Mechanical failures involve the
presence of faulty o-rings. A faulty o-ring can allow leakage of the solution through the
membrane thereby leading to decreased salt rejection. In addition to the possibility of the o-ring
being faulty, membrane fouling could also be another likely factor effecting the amount of salt
rejection observed. Given repeated use of the membrane during the semester, a potential build of
salt is possible due to concentration polarization which reduces the flow rate of the feed. This in
turn leads to a decreased salt rejection over a period of time. Appropriate replacement or cleaning
of the membrane is recommended to obtain higher salt rejection. If the value of the normalized
salt passage increases by 15%, the RO membrane needs to be cleaned in order to obtain higher
salt rejection. In this experiment the values of salt passage (1-SR) obtained for all the trials
conducted at pressure drops ranging from 300-500 psi were above 15% indicating the presence
of fouling and the need for cleaning the reverse osmosis membrane. To achieve further reduction
in permeate salinity, permeate staging configuration or two membrane units in series can be used
to increase the salt rejection and achieve a greater percent of water recovery.
Permeate flow rate can be determined by measuring the mass of the permeate stream over a
particular time period. A linear regression between the permeate flow rate and the pressure drop
across the membrane can then be performed to calculate the permeability of salt through the
membrane for future experiments which can then be used to calculate the permeate flow rates for
scaled up systems.
Figure 1A. The piping & instrumentation diagram shown above was taken from the Reverse
Osmosis HAZOP preparatory report written by Group B. The salt water flow is first pumped
from TK-2, the reservoir tank, to TK-1, the storage tank, by pump, PMP-2. Saltwater inside the
storage tank is pumped into the membrane, M-1, as the feed stream via V-3, and V-5, two flow
regulator valves. The membrane desalinates the pumped water and separates the feed stream into
the permeate and the retentate. The permeate stream flows to the drain via V-12, a flow regulator
valve. The retentate stream flows to either the drain via V-8, a flow regulator valve, or is recycled
back to the storage tank via HEX-1, the heat exchanger, and V-11, a flow regulator valve.
140
130
Osmotic Pressure (atm)
120
110
100
90
80
11.50 11.75 12.00 12.25 12.50 12.75 13.00 13.25
Retentate Concentration (g/L)
90 20
Residual
Percent
50
0
10
1 -20
-40 -20 0 20 40 80 90 100 110
Residual Fitted Value
Residual
0
1
0 -20
-20 -10 0 10 20 30 1 2 3 4 5 6 7 8 9 10
Residual Observation Order
Figure 3 C (c). Residual plot for osmotic pressure as a function of retentate concentration when
the pump is running at 450 psi
105 105
Osmotic Pressure (atm)
100 100
95 95
90 90
3.9 4.0 4.1 4.2 4.3 4.4 4.5 4.6 4.7 4.8 12.5 13.0 13.5 14.0 14.5 15.0
Permeate Concentration (g/L) Retentate Concentration (g/L)
Figure 4 C (a). Linear Regression of Osmotic Figure 4 C (b). Linear regression of osmotic
Pressure (atm) vs. Permeate Concentration (g/L) pressure (atm) vs Retentate Concentration
when pump is running at 350 psi. Regression (g/L) when pump is running at 350 psi.
equation is: Regression equation is:
.
Residual Plots for Osmotic Pressure (atm)
Normal Probability Plot Versus Fits
99
5.0
90
2.5
Residual
Percent
50 0.0
-2.5
10
-5.0
1
-10 -5 0 5 10 90 93 96 99 102
Residual Fitted Value
Residual
2 0.0
-2.5
1
-5.0
0
-7.5 -5.0 -2.5 0.0 2.5 5.0 1 2 3 4 5 6 7 8 9 10
Residual Observation Order
Figure 4 C (c). Residual plot for osmotic pressure as a function of retentate concentration when
the pump is running at 350 psi
105
105
100
Osmotic Pressure (atm)
100
Osmotic Pres sure (a tm)
95 95
90 90
85 85
3.8 3.9 4.0 4.1 4.2 4.3 4.4 4.5 4.6 12.75 12.80 12.85 12.90 12.95 13.00 13.05
Permeate Concentration (g/L) Retentate Concentration (g/L)
Figure 5 C (a). Linear Regression of Osmotic Figure 5 C (b). Linear regression of osmotic
Pressure (atm) vs. Permeate Concentration (g/L) pressure (atm) vs Retentate Concentration
when pump is running at 400 psi. Regression (g/L) when pump is running at 400 psi.
equation is: Regression equation is:
Residual Plots for Osmotic Pressure (atm)
Normal Probability Plot Versus Fits
99
90 5.0
Residual
2.5
Percent
50
0.0
10 -2.5
1 -5.0
-8 -4 0 4 8 90.0 92.5 95.0 97.5 100.0
Residual Fitted Value
2.5
Residual
2
0.0
1 -2.5
0 -5.0
-5.0 -2.5 0.0 2.5 5.0 7.5 1 2 3 4 5 6 7 8 9 10
Residual Observation Order
Figure 5 C (c). Residual plot for the linear regression of osmotic pressure as a function of
retentate concentration when the pump is operating at 400 psi.
130 130
125 125
Osmotic Pressure (atm)
120 120
115 115
110 110
105 105
4.6 4.8 5.0 5.2 5.4 5.6 5.8 6.0 13.1 13.2 13.3 13.4 13.5 13.6 13.7 13.8 13.9
Permeate Concentration (g/L) Retentate Concentration (g/L)
Figure 6 C (a). Linear regression of osmotic Figure 6 C (d). Linear regression of osmotic
Pressure (atm) vs permeate concentration (g/L) pressure (atm) vs retentate concentration (g/L)
When pump is operating at 300 psi. Regression when pump is running at 300 psi. Regression
Equation is: equation is:
Residual Plots for Osmotic Pressure (atm)
Normal Probability Plot Versus Fits
99
10
90
5
Residual
Percent
50 0
-5
10
-10
1
-20 -10 0 10 20 110.0 112.5 115.0 117.5 120.0
Residual Fitted Value
Frequency 5
2
Residual
0
1 -5
-10
0
-12 -8 -4 0 4 8 12 1 2 3 4 5 6 7 8 9 10
Residual Observation Order
Figure 6 C (c). Residual plot for the linear regression of osmotic pressure as a function of
retentate concentration when the pump is operating at 300 psi
105 105
100 100
Osmotic Pressure (atm)
Osmotic Pressure (atm)
95 95
90 90
85 85
80 80
75 75
70 70
3.0 3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 12.0 12.2 12.4 12.6 12.8 13.0 13.2 13.4
Permeate Concentration (g/L) Retentate Concentration (g/L)
Figure 7 C (a). Linear regression of osmotic Figure 7 C (b). Linear regression of osmotic
Pressure (atm) vs permeate concentration (g/L) pressure (atm) vs retentate concentration (g/L)
When pump is operating at 500 psi. Regression when pump is operating at 500 psi. Regression
Equation is: equation is:
Residual Plots for Osmotic Pressure (atm)
Normal Probability Plot Versus Fits
99
10
90
0
Residual
Percent
50
-10
10
-20
1
-20 -10 0 10 20 93 94 95 96
Residual Fitted Value
3
0
Frequency
Residual
2
-10
1
-20
0
-25 -20 -15 -10 -5 0 5 10 1 2 3 4 5 6 7 8 9 10
Residual Observation Order
Figure 7 C (c). Residual plot for the linear regression of osmotic pressure as a function of
retentate concentration when the pump is operating at 500 psi
Appendix D- Linear regression plots of salt rejection as a function of time for pressure drops of
300, 400, 450 and 500 psi and the respective linear regression residual plots
0.550 90
0.025
Residual
Percent
50
0.525 0.000
Salt rejection
10
1
-0.050 -0.025 0.000 0.025 0.050 0.450 0.475 0.500 0.525 0.550
0.500 Residual Fitted Value
2 0.025
Residual
0.450
1 0.000
0 2 4 6 8 10 0
Time (min) -0.02 -0.01 0.00 0.01 0.02 0.03 0.04 1 2 3 4 5 6 7 8 9 10
Residual Observation Order
Figure 8 D (a). Linear regression of salt Figure 8 D (b). Residual plot for the linear
rejection as function of time at the membrane regression of salt rejection as a function of
pressure of 300 psi. Regression equation is: time at the pressure drop of 300 psi
Residual Plots for Salt rejection
0.63
Normal Probability Plot Versus Fits
99
0.62
0.02
90
0.01
Residual
0.61
Percent
50
0.00
0.60
Salt rejection
10 -0.01
1 -0.02
0.59 -0.04 -0.02 0.00 0.02 0.04 0.56 0.57 0.58 0.59 0.60
Residual Fitted Value
Frequency
0.01
Residual
0.56 2 0.00
1 -0.01
0.55
0 2 4 6 8 10 0 -0.02
Time (min) -0.02 -0.01 0.00 0.01 0.02 1 2 3 4 5 6 7 8 9 10
Residual Observation Order
Figure 9 D (a). Linear regression of salt Figure 9 D (b). Residual plot for the linear
Rejection as a function of time at the regression of salt rejection as a function of
pressure drop of 400 psi. Regression time at the pressure drop of 400 psi
Equation is:
Residual
Percent
50
0.00
0.62
10
Salt rejection
-0.02
0.60 1
-0.050 -0.025 0.000 0.025 0.050 0.50 0.55 0.60 0.65
Residual Fitted Value
0.58
Histogram Versus Order
0.56
0.04
2.0
Residual
0.52 1.0 0.00
0.5
0.50 -0.02
0 2 4 6 8 10 0.0
Time (min) -0.03 -0.02 -0.01 0.00 0.01 0.02 0.03 0.04 1 2 3 4 5 6 7 8 9 10
Residual Observation Order
Figure 10 D (a). Linear regression of salt Figure 10 D (b). Residual plot for the linear
Rejection as a function of time at the regression of salt rejection as a function of
Pressure drop of 450 psi. Regression time at the pressure drop of 450 psi
Equation is:
Residual Plots for Salt rejection
0.70
Normal Probability Plot Versus Fit s
99 0.04
90 0.02
Residual
Percent
0.65 50 0.00
-0.02
Salt rejection
10
1 -0.04
-0.050 -0.025 0.000 0.025 0.050 0.54 0.57 0.60 0.63 0.66
Residual Fitted Value
0.60
Histogram Versus Order
0.04
3
0.02
Frequency
2
Residual
0.55
0.00
1
-0.02
0 2 4 6 8 10 0 -0.04
Time (min) -0.03 -0.02 -0.01 0.00 0.01 0.02 0.03 0.04 1 2 3 4 5 6 7 8 9 10
Residual Observation Order
Figure 11 D (a). Linear regression of salt Figure 11 D (b). Residual plot for the linear
Rejection as a function of time at the regression of salt rejection as a function of
pressure drop of 500 psi. time at the pressure drop of 500 psi
Regression equation is:
Appendix E.
Table 1E. Yearly operational cost of RO pump at pressure drops of 300, 350, 400, 450 and 500
psi