Calibration of Equipments: By: Shravan Kumar
Calibration of Equipments: By: Shravan Kumar
Reactors
Storage tanks
Vessels
Objective :
To develop a standard procedure for the
equipment (Volume) calibration (e. g.
Reactor, Receiver and Tank)used in
manufacturing.
Responsibility:
1. Production Chemist
2. Production Technician
3.A DI water flow meter and two clean hose pipe for the performing the
calibration activity.
4.DI water flow meter should be calibrated.
1. Reactor calibration
Agitator
Maximum Working
capacity
(80% of the equipment
volume )
TP
Minimum Touch
stirrable volume
volume Dead
volume
Calibration of equipment can be divide into three
parts
1. Dead volume :
3. Fill the water till the water come out from the manhole .
Line
Over indicator
flow
volume
Visual
capacity
Dead
volume
FOR TANKS & VESSELS
Procedure for find the dead volume of tanks & vessels:
3.Fill the DI water till the level of line indicator reaches at one centimeter
in LI.
4.Note the final reading.
6.Repeat the activity till the LI fill completely (working capacity of the equipment ).
Procedure for over flow volume of the Tanks & Vessels: