Filtering Samples To On-Line Process Analyzers

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

Filtering Samples to

On-Line Process Analyzers


Guidelines for Proper Filtration and Delivery
of Samples to On-Line Process Analyzers
Filtering samples to on-line analyzers

Use these guidelines to ensure requirements for an effective filter in the desirable in many filter applications - such
proper filtration and delivery sampling line is generally recognized, it is as compressed air or water filters - filters
of samples to on-line process also important to recognize that it is usu- not specifically designed for sampling usu-
ally necessary to use a filter specifically ally are not suited for analyzers.
analyzers designed for sampling applications, rather Requirements for sample filters range so
than trying to make do with a general pur- widely that specifying a filter is best done
Ken Perrotta, Division Engineering Manager
pose or homemade filter. on a case-by-case basis. There is, how-
Allan Fish, Product Manager
Parker Hannifin Corporation The characteristics that a sample filter ever, one generalization that applies to all
Filtration and Separation Division should have, in addition to filtering out sample filter requirements: the filter must
Haverhill, Massachusetts contaminants, are: be able to separate efficiently a noncontin-
uous phase contaminant from the continu-
• The filter must not change the compo-
ous sample stream phase. Specifically, the
sition of the sample, other than to remove
filter must be able to make the following
unwanted impurities. Therefore, the choice
Increasingly precise process control strat- separations, in addition to removing solid
of filter media generally is limited to a few
egies, often linked to computer control, particles:
chemically inert materials: glass, stainless
have accelerated the use of sophisticated
steel and PTFE. • Gas samples - remove liquid droplets
on-line chemical composition analyz-
ers in plant applications. Instruments • Since the sample filter often is in a • Liquid samples - remove immiscible
such as gas and liquid chromatographs, remote or inconvenient location, it must liquid droplets and gas bubbles.
ion chromatographs, laser optic instru- be capable of operating for a reasonable Most filter media will do an adequate job
ments, atomic absorption instruments and period between scheduled maintenance of removing solid particles from liquids
specific ion analyzers, which were rari- checks. Even more important, it should not or gases, but the only practical commer-
ties in laboratories ten years ago, are now be susceptible to unscheduled problems, cial media that will separate liquids from
found routinely in plant settings. While such as filter element plugging or ruptur- gases, gas bubbles from liquids, and two
continuing miniaturization and “rugged- ing, between regular maintenance checks. immiscible liquids is resin-bonded glass
ization” of the electronics are making the • Sample filter maintenance in the field microfiber media. All recommendations
instrumentation circuitry more tolerant of usually is performed under adverse condi- in this paper are based on resin-bonded
the plant environment, a problem that will tions by personnel who are not trained glass fiber media.
not diminish is the level of contamination chemists; therefore, the filter should be
in plant samples compared to laboratory designed for easy and uncomplicated
samples. maintenance. Filter elements should be
Factors in plant operation that magnify the rugged and not susceptible to handling
difficulty of delivering acceptable samples damage; the unassembled housing should
to on-line process analyzers are: have a minimum number of loose parts,
and the housing should be designed so
• The sample stream must be processed
that it is virtually impossible to install a
and fed to the analyzer continuously, rather
filter element incorrectly.
than the batchwise method permissible in
the laboratory. • A filter should introduce minimum lag
time into the system. Lag time can be dealt
• Test frequency in the plant is far great-
with in the sample system design (slip-
er than in the laboratory. Where a labora-
stream sampling, for example), but sizable
tory analyzer might perform 100 tests a
dead volume in the filter housing should
month, an on-line plant analyzer could do
be avoided. Since large reservoir volume is
100 tests a day. In the high frequency plant
usage, trace contaminants that would not
be a problem in the laboratory can rapidly Figure 1: External reservoir for high pressure
build up and cause instrument failure. applications

Filter requirements. It is not surprising


that contaminants in the plant samples are
reported to be the most frequent cause of
problems with on-line analyzers. While the
Figure 2: Slipstream or bypass filtration Figure 3: Slipstream plus coalescing filtration

Separating liquids from gases. Glass ficient volume to hold all the liquid that is gravity between the two liquids is less than
microfiber filter tubes efficiently separate expected to be collected during any period 0.1 units (for example, if an oil suspended
suspended liquids from gases. The filter of unattended operation. To drain liquid in water has a specific gravity between
fibers capture the fine liquid droplets sus- while the filter is operating at pressure or 0.9 and 1.1), the separation time for the
pended in the gas and cause the droplets vacuum conditions, valve #1 is closed and coalesced phase may be impractically long.
to run together to form large drops within valve #2 is opened. In that case, if there is only a small quan-
the depth of the filter tube. The large tity of suspended liquid, the filter tube can
If the filter is under vacuum, the external
drops are then forced by the gas flow to be used until saturated with the suspended
reservoir is a practical method of collect-
the downstream surface of the filter tube, liquid and then changed.
ing coalesced liquid for manual draining
from which the liquid drains by gravity. periodically. Alternatively, if any external Another practical problem with liquid-liq-
This process is called “coalescing”. Since vacuum source is available, such as an as- uid separations is that small quantities of
the coalesced liquid drains from the tubes pirator, the liquid may be drained continu- impurities can act as surface-active agents
at the same rate that liquid droplets enter ously from the housing drain port. and interfere with the coalescing action.
the tubes, the tubes have an unlimited life For that reason it is not possible to predict
when coalescing liquids from relatively Separating two liquid phases. In prin-
accurately the performance of a liquid-liq-
clean gases, and the filters operate at their ciple, glass microfiber filter tubes separate
uid coalescing filter, and each system must
initial retention efficiency even when wet suspended droplets of a liquid which is
be tested on site. The general guidelines
with liquid. The flow direction is inside-to- immiscible in another liquid by the same
for the system to start testing are to use
outside to permit the liquid to drip from the process as they separate droplets of liquid
25 micron filter tubes and flow inside-out
outside of the filter to the housing drain. form a gas. The liquid droplets suspended
at very low flow rates. If the suspended
in the continuous liquid phase are trapped
The filter tube grade should be selected liquid is lighter than the continuous phase
on the fibers and run together to form
for maximum liquid drainage rate, rather the housing should be oriented so that the
large drops, which are then forced through
than maximum filtration efficiency rating. drain port is up.
the filter to the downstream surface. The
Since liquid drainage rate decreases with large drops separate from the continuous Removing gas bubbles from liquids. Glass
increasing filtration efficiency rating, this is liquid phase by gravity difference, settling microfiber filter tubes readily remove sus-
a clear case where overspecifying filter ef- if heavier than the continuous phase and pended gas bubbles from liquid, eliminat-
ficiency will lead to unsatisfactory results. rising if lighter. The coalescing action of ing the need for de aeration tanks, baffles
If liquid is carried into the filter in slugs glass microfiber filters is effective with or other separation devices. Flow direction
rather than dispersed as droplets in the aqueous droplets suspended in oil or other through the filter is outside-to-inside, and
gas, a filter that is properly sized for hydrocarbons, and also with oil in water the separated gas bubbles rise to the top
steady-state conditions can be flooded suspensions. of the housing and are vented through the
and permit liquid carryover. If slugging drain port. If slipstream sampling is used
In practice, however, liquid-liquid sepa-
of liquid is expected, a filter with a rela- (see below), the separated bubbles are
rations are much more difficult than
tively large bowl should be selected to swept out of the housing with the bypassed
liquid-gas separations. The specific gravity
provide adequate liquid holding capacity, liquid. Filter tubes rated at 25 micron are a
difference between two liquids is always
and positive provisions should be made good choice for gas bubble separation.
less than between a liquid and a gas, and
to drain the liquid automatically from the therefore, a longer phase separation time Slipstream or bypass sampling. Instru-
bowl of the housing as fast as it accu- is needed. Either the filter housing must be ment sample use rates are invariably quite
mulates. An automatic float drain can be oversized or the flow rate greatly reduced low, yet it is essential to minimize lag time
used if the pressure is in the 10 to 400 psig to avoid carryover of the coalesced phase. in the sample system. Since analyzers
range. Above 400 psig, the upper limit for As a rule of thumb, flow rate for liquid- often are located some distance from the
commercially-available float drains, the liquid separation should be no more than sampling point, samples usually are trans-
possibilities are: a constant bleed drain, a one-fifth the flow rate for solid-liquid ported to the analyzer at a relatively high
valve with an automatic timed actuator, or separation. Even at low flow rates, if the flow rate to minimize lag time. The sample
an external reservoir with manual valves specific gravity difference between the two is divided at the analyzer, with the analyzer
(see Figure 1). The external reservoir can liquids is less than 0.1 units (for example, using the portion it requires (usually a very
be constructed of pipe or tubing with suf- if an oil suspended in water has a specific small fraction of the total sample), and the
balance recycled to the process or vented. sampling if the filter is to coalesce and located as close to the analyzer as pos-
continuously drain suspended liquid from sible to minimize the chance of condensa-
If the sample filter is located in the
a gas stream or to coalesce liquid droplets tion between the filter and the analyzer.
low-flow line to the analyzer, it will have
from a liquid stream. As noted previously, Additional precautions that can be taken
good life between filter element changes
the coalesced liquid is removed in the form to avoid downstream condensation include
because the solids loading rate is very
of large drops from the downstream side cooling the sample below ambient tem-
low; however, the filter must be carefully
of the filter. Therefore, the coalescing filter perature upstream of the coalescing filter,
selected to avoid introducing unaccept-
requires two outlet ports, one for the dry and heating the line gradually between the
able lag time. If the filter is located in the
gas and one for the liquid drain. To com- filter and the analyzer.
high-flow portion of the sample system, its
bine coalescing and slipstream filtration, a
effect on sample lag time can be relatively Sampling liquid effluent streams. Liquid
filter housing would need four ports - two
low, but the life between filter changes effluent analyzers usually deal with aque-
for inlet and bypass and two for filtered gas
may be inconveniently short because the ous streams having a high solids content.
and coalesced liquid - which is not a prac-
element is filtering a much greater volume In addition, the analyzers are often located
tical design. Therefore, slipstreaming plus
of material than the analyzer is using. in remote areas of the plant and are in-
coalescing requires two stages of filtration
frequently serviced. Therefore, the sample
Ideally, a filter should be located at the (see Figure 3). The second (coalescing)
filter system must have long life between
point where the low-flow stream is with- stage must be located in the sample line
filter tube changes even in a high solids
drawn to the analyzer (see Figure 2). This to the analyzer, and should be as small
situation. The general recommendation
arrangement permits the main volume of as possible to minimize lag time. If the
for this requirement is a two-stage filter
the filter to be swept continuously by the quantity of suspended liquid is not large, a
system, a 75 micron prefilter followed by
high flow rate system, thus minimizing lag miniature in-line disposable filter unit may
a 25 micron final filter. The filters should
time. At the same time, only the low-flow be considered as a trap for the suspended
be oversized as much as possible without
stream to the analyzer is filtered, thus liquid, to be replaced when saturated.
casing excessive lag time. Plastic filter
maximizing filter life.
Stack gas sampling. When sampling hot, housings are usually a good choice.
A slipstream filter requires inlet and outlet wet gas, such as stack gas, a filter capable
Sampling high pressure steam or conden-
ports at opposite ends of the filter element of withstanding the gas temperature can
sate. Measurements are often required
to allow the high flow rate of the by-passed be installed in the stack at the beginning
in high pressure boiler systems of steam
material to sweep the surface of the filter of the sample line to prevent solids from
and condensate conductivity, specific ion
element and the filter reservoir, and a third entering the gas sample line. After the
concentrations and feedwater additive
port connected to the low flow rate line to sample is cooled, a coalescing filter is used
concentrations. In a continuous sampling
the analyzer, which allows filtered samples to remove suspended liquids before the
system, the high pressure steam or con-
to be withdrawn from the filter reservoir. sample goes to the analyzer (see Figure
densate is cooled to below 100°F and then
If bubble removal from a liquid is a re- 4). Flow direction is inside-to-outside.
reduced to near atmospheric pressure for
quirement; this function may be combined Filter housings with Pyrex glass bowls are
metering to the analyzers. Filtration is re-
with slipstream filtration, since the recom- often used in this application to permit a
quired upstream of the pressure reducing
mended flow direction for bubble removal visual check of the liquid level in the filter
valves, to prevent pitting of the valve seats
is outside-to-inside, and the separated housing. Since there is often a consider-
by suspended particles and to eliminate
bubbles will be swept out of the housing by able amount of liquid present at this point,
variations in flow rate to the analyzers.
the bypass stream. In this case the liquid steps must be taken to drain the housing
to ensure that liquid does not build up and A stainless steel filter housing with the
feed should enter at the bottom of the
carry downstream to the analyzer. appropriate pressure rating and 25 micron
housing and the bypass liquid exit at the
filter tube is recommended. Since the
top of the housing. The liquid coalescing filter should be
analyzer system is often located some
A special problem arises in slipstream distance from the sampling point, slip-
stream filtration is usually required. Figure
5 shows a sampling system in operation at
a steam generating facility.

Figure 4: Stack gas sampling Figure 5: High pressure steam sampling


Hydrophobic membrane sampling. Many
online instruments are susceptible to cor-
rosion and skewed analysis from any water
and moisture contamination. Gas chro-
matographs, mass spectrometers, oxygen
analyzers and other sensitive on-line
instruments require complete removal of
all moisture. Instrument sensitivity levels
range from PPM to PPB and “percent
level” component concentrations.
(Note: For the membrane to operate cor-
As a result, it is good practice to install rectly, there must be a bypass flow.)
a hydrophobic membrane filter in line to
the instrument to protect and safeguard it Figure 6
from any moisture contamination. A stain-
less steel filter housing with a hydrophobic
membrane allows the sample gas to flow
on the upstream side of the membrane
and exit through the outlet port on the
downstream side (see Figure 6). Entrained
moisture will not flow through the mem-
brane and will exit out the by-pass port
on the upstream side of the membrane
completely protecting the instrument from
moisture.
If the sample gas contains excessive
amounts of moisture and particulates it is
recommended to use a stainless steel filter (Note: For the membrane to operate cor-
rectly, there must be a bypass flow.)
housing that incorporates a coalescing
prefilter and the hydrophobic membrane
(see Figure 7). The coalescing prefilter will
protect the membrane from premature
Figure 7
blinding and extend its useful life.

1/4” NPT Minimal panel space, permanent line For this application, it is recommended
Outlet Port
mount sampling. As more and more a stainless steel filter designed to be
sample lines and instruments are added to horizontally mounted at a 10 degree angle
instrument sheds, the need to utilize the be installed to minimize the amount of foot
space in the most efficient way has become print on the panel (see Figure 8). This de-
critical. In addition, the need to maintain sign incorporates the inlet ,outlet and drain
sample filters without having to “break” ports all in the head of the filter enabling
the sample line and expose it to ambient filter element change outs without disrupt-
conditions has also become quite impor- ing the sample line.
tant to most facilities. By not having to
1/4” NPT “break” sample line the need to flush the
Drain Port
lines prior to start up is eliminated.
Figure 8: Horizontal mounting

Parker Hannifin Corporation Filtering Samples to On-Line Analyzers


Filtration and Separation Division ©Copyright Parker Hannifin Corporation 2013 September 2013
242 Neck Road
Haverhill, MA 01835
1-800-343-4048, 1-978-858-0505

www.parker.com/balston

You might also like