This certificate of analysis summarizes the results of quality testing on a batch of 2000 disposable trocars. The trocars were tested against specifications for various quality characteristics including outward appearance, size, flexibility, sealing performance, corrosion resistance, matching performance, sterilization, and EO residues. The results showed that the trocars met all standard requirements and passed all tests.
This certificate of analysis summarizes the results of quality testing on a batch of 2000 disposable trocars. The trocars were tested against specifications for various quality characteristics including outward appearance, size, flexibility, sealing performance, corrosion resistance, matching performance, sterilization, and EO residues. The results showed that the trocars met all standard requirements and passed all tests.
This certificate of analysis summarizes the results of quality testing on a batch of 2000 disposable trocars. The trocars were tested against specifications for various quality characteristics including outward appearance, size, flexibility, sealing performance, corrosion resistance, matching performance, sterilization, and EO residues. The results showed that the trocars met all standard requirements and passed all tests.
Model Type III (1) 10 Quantity 2000 Lot. No. 15070301 Test time 2015-07-07-2015-07-16 According to Product Standard published by Hangzhou Kangji Medical Instrument Co., Ltd
No. Item Standard requirement Result Conclusion
The outer surface should be smooth, no burr and 8/8 OK 1 Outward flaw, also unpolluted and no foreign matters. Pass The tip of the optical needle should be transparent. 8/8 OK OD: 012.5±1 mm 12.5012.52 Pass 2 Size(cannula (1)10) ID: (1)11±1mm 11.0011.02 Pass L: 100+20mm 100 Pass Size(optical needle OD: 010.7±1mm 10.70-4 0.71 Pass Cr-)
(1)10) L: 145±20mm 145 Pass
The sealing valve and gas barrier valve should be 4 Flexibility(D10) open and closed flexible and no clamping 13/13 OK Pass stagnation. The sealing valve of the cannula should have good Sealing 5 sealing performance and no leakage with 4kPa air 13/13 OK Pass performance( 10) pressure. The gas barrier valve of the cannula should have Gas barrier 6 good sealing performance and bubbles should be 13/13 OK Pass performance( 10) less than 20pcs in lmin with 4kPa air pressure. Corrosion The metal parts should meet the requirements of 7 13/13 OK Pass resistance(010) IS013402:1995 Matching The cannula and needle's matching space should be 8 13/13 OK Pass performance(010) not more than 0.3mm and no clamping stagnation. The product should be sterilized by a confirmed 9 Sterile OK Pass sterile process. 10 EO residues <10 tig/g 8.87 [tg/g Pass