Installation Checklist: XP-100 or XP-300
Installation Checklist: XP-100 or XP-300
Installation Date/Month/Year
completed on:
Facility name:
Instrument serial
No.
Person in charge
of installation
Remarks:
Sysmex Corporation
2nd edition March 25, 2013
XP Installation Checklist
Ⅰ Unpacking
1 Unpacking
1) Is there any damage to the instrument? □
2) Are all accessories included? □
Ⅱ Installation
1 Installation □
Ⅲ Connection
1 Connection of reagent and waste tubes
1) Has CELLPACK been connected to the instrument? □
2) Is the tube length between the instrument and CELLPACK
no more than 2m? □
3) Has STROMATOLYSER-WH been connected to the
instrument? □
4) Have reagent and waste containers been placed in
replaceable positions? □
5) Have waste tubes been secured? □
(2/3)
ⅣPerformance Check
1 Checking connection again
1) Have tubes and cords been connected correctly? □
6 Checking HGB background count level (Refer to Service Manual, Chapter 4.)
1) Is HGB background count level within the acceptable
range?
a) HGB background convert value
(Acceptable range: 2000 +/-200) □
TB 201342
XP Installation Checklist
(3/3)
13 Checking data
1) Are the mean values within the displayed values when 2 or
more levels of controls are analyzed 5 times?
* Note: Ensure that analysis is performed in the control mode.
(Attach analysis data.) Control
Lot No.
□ X 10^2/uL (or 10 3/uL)
WBC TARGET X 10^2/uL (or 10 3/uL)
LIMIT X 10^2/uL (or 10 3/uL)
□ X 10^4/uL (or 10 6/uL)
RBC TARGET X 10^4/uL (or 10 6/uL)
LIMIT X 10^4/uL (or 10 6/uL)
□ g/dL
HGB TARGET g/dL
LIMIT g/dL
□ %
HCT TARGET %
LIMIT %
□ fl
MCV TARGET fl
LIMIT fl
□ pg
MCH TARGET pg
LIMIT pg
□ g/dl
MCHC TARGET g/dl
LIMIT g/dl
□ X 10^4/uL (or 10 3/uL)
PLT TARGET X 10^4/uL (or 10 3/uL)
LIMIT X 10^4/uL (or 10 3/uL)
14 Shutdown
1) Has the instrument been shut down properly? □
Ⅴ Final check
1 Exterior check
1) Are there any dust, stains or visible scratches? □
2) Is the instrument level on the installation platform? □
Edition Date of
No. revision Details Remarks
1st 4/27/2012 This is the 1st edition.
edition
2nd 3/25/2013 The following note has been added under "11 TB 201342
edition Entering setting values" on page 2/3.